Biological evaluation of medical devices — Part 23: Tests for irritation — Amendment 1: Additional in vitro reconstructed human epidermis models

Évaluation biologique des dispositifs médicaux — Partie 23: Essais d'irritation — Amendement 1: Modèles supplémentaires d'épiderme humain reconstruit in vitro

General Information

Status
Published
Publication Date
30-Apr-2025
Current Stage
6060 - International Standard published
Start Date
01-May-2025
Due Date
27-Oct-2025
Completion Date
01-May-2025
Ref Project

Relations

Buy Standard

Standard
ISO 10993-23:2021/Amd 1:2025 - Biological evaluation of medical devices — Part 23: Tests for irritation — Amendment 1: Additional in vitro reconstructed human epidermis models Released:1. 05. 2025
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


International
Standard
ISO 10993-23
First edition
Biological evaluation of medical
2021-01
devices —
AMENDMENT 1
Part 23:
2025-05
Tests for irritation
AMENDMENT 1: Additional in vitro
reconstructed human epidermis models
Évaluation biologique des dispositifs médicaux —
Partie 23: Essais d'irritation
AMENDEMENT 1: Modèles supplémentaires d'épiderme humain
reconstruit in vitro
Reference number
ISO 10993-23:2021/Amd.1:2025(en) © ISO 2025

ISO 10993-23:2021/Amd.1:2025(en)
© ISO 2025
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
ISO 10993-23:2021/Amd.1:2025(en)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO document should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 194, Biological and clinical evaluation of medical
devices, in collaboration with the European Committee for Standardization (CEN) Technical Committee CEN/
TC 206, Biological and clinical evaluation of medical devices, in accordance with the Agreement on technical
cooperation between ISO and CEN (Vienna Agreement).
A list of all parts in the ISO 10993 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

iii
ISO 10993-23:2021/Amd.1:2025(en)
Biological evaluation of medical devices —
Part 23:
Tests for irritation
AMENDMENT 1: Additional in vitro reconstructed human
epidermis models
Introduction, fifth paragraph
Replace the paragraph with the following paragraphs:
Traditionally, tests in small animals have been performed prior to testing on humans to help predict
human responses. More recently, in vitro tests as well as human tests have been added as adjuncts
or alternatives. For skin irritation testing of neat chemicals, in vitro tests were developed using
[31]
reconstructed human epidermis (RhE) models . The method was adapted for detection of irritant
chemicals in medical device extracts. The results of a large interlaboratory study that tested two types
of RhE models showed that these models can also be used to detect the presence of irritant chemicals
extracted from polymeric materials [polyvinylchloride (PVC) and silicone] commonly used in the
[6]
manufacture of medical devices . This method was found to be equally sensitive in the detection of
low concentrations of some strong irritant compounds when compared to the human patch testing and
[14]
intracutaneous rabbit test . Therefore, a stepwise approach for irritant testing can start with the in
vitro RhE model.
In 2023, two new type RhE models listed in OECD 439 were adopted for medical devices after they
demonstrated equivalent predictive capacity compared to the two other RhE models in interlaboratory
[42][43][44]
studies .
Clause 4, second paragraph
Replace the paragraph with the following paragraphs:
Preference for in vitro tests instead of in vivo tests in accordance with ISO 10993-2, shall be considered,
with replacement of the latter as new in vitro tests are scientifically validated and qualified for use with
medical devices and become reasonably and practicably available. The results of a large interlaboratory
study that tested two types of RhE models showed that these models can also be used to detect the
presence of irritant chemicals extracted from polymeric materials [polyvinylchloride (PVC) and
[6]
silicone] commonly used in the manufacture of medical devices . This method was found equally
sensitive to detect low concentrations of some strong irritant compounds when compared to the human
[14]
patch testing and intracutaneous rabbit test . Therefore, the in vitro irritation test shall be performed
before animal testing or human patch test is considered.
In 2023, two new type RhE models listed in OECD 439 were adopted for medical devices after they
demonstrated equivalent predictive capacity compared to the two other RhE models in catch-up
[42][43][44]
validation studies .
6.1, first paragraph
Replace the paragraph with the following paragraphs:

ISO 10993-23:2021/Amd.1:2025(en)
The in vitro method with RhE models for testing irritation was developed specifically to detect skin
[3][12]
irritation potential for neat chemicals (see OECD 439). The method was adapted and validated with
[5][6][12][13][17][18][19].
two RhE models for detection of irritant chemicals in medical device extracts This
method was found equally sensitive to detect low concentrations of some strong irritants in extracts
from polymeric medical materials (PVC and silicone) when compared to the human patch testing and
[14]
intracutaneous rabbit test . Hence, the RhE test as described in this document can replace the in vivo
rabbit test for irritation by skin exposure and by intracutaneous (intradermal) administration.
Two new type RhE models listed in OECD 439 were adopted for medical devices in 2023 after
[42][43][44]
demonstrating equivalent predictive capacity in interlaboratory studies .

6.2.2, eight paragraph
Replace the paragraph with the following:
For known cases of test-compounds requiring specific controls, some chemicals can directly reduce
the MTT reagent (e.g. electrophiles, test articles with high pH), while other chemicals can directly
colour the tissue or the cells. Such test sample properties can only interfere if sufficient amounts of
the chemical are still present on the tissue at the end of the exposure period. In these cases, a special
procedure allowing the quantification of the “true” MTT reduction should be applied. A protocol for the
determination of possible interactions with MTT is provid
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.