ISO/ASTM 51310:2004
(Main)Practice for use of a radiochromic optical waveguide dosimetry system
Practice for use of a radiochromic optical waveguide dosimetry system
ISO 51310:2004 covers the procedures for handling, testing and using a radiochromic optical waveguide dosimetry system to measure absorbed dose in materials irradiated by photons in terms of absorbed dose in water.
Pratique de l'utilisation d'un système dosimétrique à guide d'ondes optiques radiochromiques
General Information
Relations
Standards Content (Sample)
INTERNATIONAL ISO/ASTM
STANDARD 51310
Second edition
2004-06-15
Practice for use of a radiochromic optical
waveguide dosimetry system
Pratique de l’utilisation d’un système dosimétrique à guide
d’ondes optiques radiochromiques
Reference number
ISO/ASTM 51310:2004(E)
© ISO/ASTM International 2004
---------------------- Page: 1 ----------------------
ISO/ASTM 51310:2004(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe’s licensing policy, this file may be printed or viewed but shall
not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe’s licensing policy. Neither the ISO Central
Secretariat nor ASTM International accepts any liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies
and ASTM members. In the unlikely event that a problem relating to it is found, please inform the ISO Central Secretariat or ASTM
International at the addresses given below.
© ISO/ASTM International 2004
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means, electronic or mechanical,
including photocopying and microfilm, without permission in writing from either ISO at the address below or ISO’s member body in the country of the
requester. In the United States, such requests should be sent to ASTM International.
ISO copyright office ASTM International, 100 Barr Harbor Drive, PO Box C700,
Case postale 56 • CH-1211 Geneva 20 West Conshohocken, PA 19428-2959, USA
Tel. +41 22 749 01 11 Tel. +610 832 9634
Fax +41 22 749 09 47 Fax +610 832 9635
E-mail copyright@iso.org E-mail khooper@astm.org
Web www.iso.org Web www.astm.org
Published in the United States
ii © ISO/ASTM International 2004 – All rights reserved
---------------------- Page: 2 ----------------------
ISO/ASTM 51310:2004(E)
Contents Page
1 Scope . 1
2 Referenced documents . 1
3 Terminology . 1
4 Significance and use . 2
5 Apparatus . 2
6 Performance check of instrumentation . 2
7 Calibration of the dosimetry system . 3
8 Procedure . 3
9 Characterization of each batch of dosimeters . 3
10 Application of dosimetry system . 4
11 Minimum documentation . 4
12 Measurement uncertainty . 4
13 Keywords . 4
Bibliography . 5
Figure 1 Block diagram of the instrument described in section 5 . 2
© ISO/ASTM International 2004 – All rights reserved iii
---------------------- Page: 3 ----------------------
ISO/ASTM 51310:2004(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
Draft International Standards adopted by the technical committees are circulated to the member bodies for
voting. Publication as an International Standard requires approval by at least 75% of the member bodies
casting a vote.
ASTM International is one of the world’s largest voluntary standards development organizations with global
participation from affected stakeholders. ASTM technical committees follow rigorous due process balloting
procedures.
A pilot project between ISO and ASTM International has been formed to develop and maintain a group of
ISO/ASTM radiation processing dosimetry standards. Under this pilot project, ASTM Subcommittee E10.01,
Dosimetry for Radiation Processing, is responsible for the development and maintenance of these dosimetry
standards with unrestricted participation and input from appropriate ISO member bodies.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. Neither ISO nor ASTM International shall be held responsible for identifying any or all such patent
rights.
International Standard ISO/ASTM 51310 was developed by ASTM Committee E10, Nuclear Technology and
Applications, through Subcommittee E10.01, and by Technical Committee ISO/TC 85, Nuclear Energy.
iv © ISO/ASTM International 2004 – All rights reserved
---------------------- Page: 4 ----------------------
ISO/ASTM 51310:2004(E)
Standard Practice for
Use of a Radiochromic Optical Waveguide Dosimetry
1
System
This standard is issued under the fixed designation ISO/ASTM 51310; the number immediately following the designation indicates the
year of original adoption or, in the case of revision, the year of last revision.
1. Scope E 958 Practice for Measuring Practical Spectral Bandwidth
of Ultraviolet-Visible Spectrophotometers
1.1 This practice covers the procedures for handling, test-
E 1026 Practice for Using the Fricke Reference Standard
ing, and using a radiochromic optical waveguide dosimetry
Dosimetry System
system to measure absorbed dose in materials irradiated by
2.2 ISO/ASTM Standards:
photons in terms of absorbed dose in water.
51261 Guide for Selection and Calibration of Dosimetry
1.2 This practice applies to radiochromic optical waveguide
3
Systems for Radiation Processing
dosimeters that can be used within part or all of the specified
51400 Practice for Characterization and Performance of a
ranges as follows:
3
High-Dose Radiation Dosimetry Calibration Laboratory
1.2.1 The absorbed dose range is from 1 to 10 000 Gy for
51707 Guide for Estimating Uncertainties in Dosimetry for
photons.
3
Radiation Processing
1.2.2 The absorbed dose rate is from 0.001 to 1000 Gy/s.
2.3 International Commission on Radiation Units and
1.2.3 The radiation energy range for photons is from 0.1 to
4
Measurements (ICRU) Reports:
10 MeV.
ICRU Report 14 Radiation Dosimetry: X-Rays and Gamma
1.2.4 The irradiation temperature range is from –78 to
Rays with Maximum Photon Energies Between 0.6 and 50
+60°C.
MeV
1.3 This standard does not purport to address all of the
ICRU Report 17 Radiation Dosimetry: X–Rays Generated
safety concerns, if any, associated with its use. It is the
at Potentials of 5 to 150 kV
responsibility of the user of this standard to establish appro-
ICRU Report 34 The Dosimetry of Pulsed Radiation
priate safety and health practices and determine the applica-
ICRU Report 60 Fundamental Quantities and Units for
bility of regulatory limitations prior to use.
Ionizing Radiation
2. Referenced documents
3. Terminology
2
2.1 ASTM Standards:
3.1 Definitions:
E 170 Terminology Relating to Radiation Measurements
3.1.1 analysis wavelength—wavelength used in a spectro-
and Dosimetry
photometric instrument for the measurement of optical absor-
E 275 Practice for Describing and Measuring Performance
bance or reflectance.
of Ultraviolet, Visible, and Near Infrared Spectrophotom-
3.1.2 calibration curve—graphical representation of the
eters
dosimetry system’s response function.
E 668 Practice for the Application of Thermoluminescence-
3.1.3 dosimeter batch—quantity of dosimeters made from a
Dosimetry (TLD) Systems for Determining Absorbed Dose
specific mass of material with uniform composition, fabricated
in Radiation-Hardness Testing of Electronic Devices
in a single production run under controlled, consistent condi-
E 925 Practice for the Periodic Calibration of Narrow Band-
tions and having a unique identification code.
Pass Spectrophotometers
3.1.4 dosimetry system—system used for determining ab-
sorbed dose, consisting of dosimeters, measurement instru-
1
This guide is under the jurisdiction of ASTM Committee E10 on Nuclear ments and their associated reference standards, and procedures
Technology and Applications and is the direct responsibility of Subcommittee
for the system’s use.
E10.01 on Dosimetry for Radiation Processing, and is also under the jurisdiction of
ISO/TC 85/WG 3.
Current edition approved April 5, 2004. Published June 15, 2004. Originally
e1
published as ASTM E 1310–89. Last previous ASTM edition E 1310–98 . ASTM
E 1310–94 was adopted by ISO in 1998 with the intermediate designation ISO
15559:1998(E). The present International Standard ISO/ASTM 51310:2004(E)
replaces ISO 15559 and is a minor revision of the last previous edition ISO/ASTM
51310:2002(E).
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
3
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Annual Book of ASTM Standards, Vol 12.02.
4
Standards volume information, refer to the standard’s Document Summary page on Available from the International Commission on Radiation Units and Measure-
the ASTM website. ments, 7910 Woodmont Ave., Suite 800, Bethesda, MD 20814, U.S.A.
© ISO/ASTM International 2004 – All rights reserved
1
---------------------- Page: 5 ----------------------
ISO/ASTM 51310:2004(E)
3.1.5 measurement quality assurance plan—documented 4.2 In the application of a specific dosimetry system,
program for the measurement process that ensures on a absorbed dose is determined by use of a calibration curve
continuing basis that the overall uncertainty meets the require- traceable to national or international standards.
ments of the specific application. This plan requires traceability 4.3 The absorbed dose determined is usually specified in
to, and consistency with, nationally or internationally recog- water. Absorbed dose in other materials may be determined by
nized standards. applying the conversion factors discussed in ISO/ASTM Guide
3.1.6 net response, DR— radiation–induced change in the 51261.
relationship of measured absorbance at a specific wavelength
NOTE 1—For a comprehensive discussion of various dosimetry meth-
determined by subtracting the pre–irradiation response, R ,
0
ods applicable to the radiation types and energies discussed in this
from the post–irradiation response, R:
practice, see ICRU Reports 14, 17, and 34.
DR 5 R 2 R (1)
0
4.4 These dosimetry systems commonly are applied in the
industrial radiation processing of a variety of products, for
with:
example, the sterilization of medical devices and radiation
A
l
R 5 (2)
processing of foods (4-6).
A
lref
A
l
5. Apparatus
R 5 [ #
0 0
A
lref
5.1 The following shall be used to determine absorbed dose
and where:
with radiochromic optical waveguide dosimetry systems:
A = optical absorbance at the analysis wavelength, l,
l 5.1.1 Dosimeters—A batch or portion of a batch of radio-
and
chromic optical waveguide dosimeters.
A = optical absorbance at a reference wavelength, l .
lref ref
5.1.2 Spectrophotometer or Photometer—An instrument,
either a spectrophotometer equipped with a special dosimeter
3.1.7 optical waveguide—device that contains an optical
holder and associated coupling optics (see Ref 7 for an
path at a high index of refraction relative to the material
example), or a modified photometer (see Fig. 1 for a block
enclosing the optical path.
diagram of an instrument that uses a reference wavelength),
3.1.8 radiochromic optical waveguide—specially prepared
having documentation covering analysis wavelengths, accu-
optical waveguide containing ingredients that undergo an
racy of wavelength selection, absorbance determination, spec-
ionizing radiation–induced change in photometric absorbance.
tral bandwidth, and stray light rejection.
This change in absorbance can be related to absorbed dose in
5.1.3 Holder, to position the dosimeter reproducibly in the
5
water (1, 2).
measuring light beam.
3.1.9 reference wavelength, l —wavelength selected for
ref
comparison with the analysis wavelength. This wavelength is
6. Performance check of instrumentation
chosen to minimize effects associated with optical coupling
6.1 Check and document the performance of the photometer
and other geometric variations in the dosimeter.
or spectrophotometer (see ASTM Practices E 275, E 925,
3.1.10 response function—mathematical representation of
E 958, and E 1026). Use reference standards traceable to
the relationship between dosimeter response and absorbed dose
national or international standards, unless the photometer’s or
for a given dosimetry system.
spectrophotometer’s design precludes such use.
3.2 Definitions or other terms used in this standard that
6.1.1 When using a photometer, check and document the
pertain to radiation measurement and dosimetry may be found
accuracy of the absorbance scale at intervals not to exceed one
in ASTM Terminology E 170. Definitions in E 170 are com-
month during periods of use, or whenever there are indications
patible with ICRU 60; that document, therefore, may be used
of poor performance.
as an alternative reference.
4. Significance and use
4.1 The radiochromic optical waveguide dosimetry system
provides a means of measuring absorbed dose in materials.
Under the influence of ionizing radiation such as photons,
chemical reactions take place in the radiochromic optical
waveguide creating and/or modifying optical absorbance bands
in the visible region of the spectrum. Optical response is
determined at selected wavelengths using the equations in
3.1.6. Examples of appropriate wavelengths for the analysis for
specific dosimetry systems are provided by their manufacturers
and in Refs (1) through (5).
5
The boldface numbers in parentheses refer to the bibliography at the end of this
practice. FIG. 1 Block Diagram of the Instrument Described in Section 5
© ISO/ASTM International 2004 – All rights reserved
2
---------------------- Page: 6 ----------------------
ISO/ASTM 51310:2004(E)
6.1.2 When using a spectrophotometer, check and document 8.1.3 Visually inspect the dosimeters for imperfections (for
the precision and bias of the wavelength sca
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.