Practice for use of a dichromate dosimetry system

ISO/ASTM 51401:2013 covers the preparation, testing, and procedure for using the acidic aqueous silver dichromate dosimetry system to measure absorbed dose to water when exposed to ionizing radiation. The system consists of a dosimeter and appropriate analytical instrumentation. For simplicity, the system will be referred to as the dichromate system. The dichromate dosimeter is classified as a type I dosimeter on the basis of the effect of influence quantities. The dichromate system can be used as either a reference standard dosimetry system or a routine dosimetry system. ISO/ASTM 51401:2013 is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation processing, and describes a means of achieving compliance with the requirements of ISO/ASTM Practice 52628 for the dichromate dosimetry system. It is intended to be read in conjunction with ISO/ASTM Practice 52628. ISO/ASTM 51401:2013 describes the spectrophotometric analysis procedures for the dichromate system. It applies only to gamma radiation, X-radiation/bremsstrahlung, and high energy electrons.

Pratique de l'utilisation d'un système dosimétrique au dichromate

General Information

Status
Published
Publication Date
04-Nov-2013
Current Stage
9020 - International Standard under periodical review
Start Date
15-Jan-2024
Completion Date
15-Jan-2024
Ref Project

Relations

Buy Standard

Standard
ISO/ASTM 51401:2013 - Practice for use of a dichromate dosimetry system
English language
6 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO/ASTM
STANDARD 51401
Third edition
2013-11-15
Practice for use of a dichromate dosim-
etry system
Pratique de l’utilisation d’un système dosimétrique au dichromate
Reference number
ISO/ASTM 51401:2013(E)
© ISO/ASTM International 2013

---------------------- Page: 1 ----------------------
ISO/ASTM 51401:2013(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe’s licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing.
In downloading this file, parties accept therein the responsibility of not infringing Adobe’s licensing policy. Neither the ISO Central
Secretariat nor ASTM International accepts any liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies
and ASTM members. In the unlikely event that a problem relating to it is found, please inform the ISO Central Secretariat or ASTM
International at the addresses given below.
© ISO/ASTM International 2013
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means, electronic or
mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or ISO’s member body in the
country of the requester. In the United States, such requests should be sent to ASTM International.
ISO copyright office ASTM International,100 Barr Harbor Drive, PO Box C700,
Case postale 56 • CH-1211 Geneva 20 West Conshohocken, PA 19428-2959, USA
Tel. +41 22 749 01 11 Tel. +610 832 9634
Fax +41 22 749 09 47 Fax +610 832 9635
E-mail copyright@iso.org E-mail khooper@astm.org
Web www.iso.org Web www.astm.org
Published in Switzerland
ii © ISO/ASTM International 2013 – All rights reserved

---------------------- Page: 2 ----------------------
ISO/ASTM 51401:2013(E)
Contents Page
1 Scope. 1
2 Referenced documents. 1
3 Terminology. 2
4 Significance and use. 2
5 Effect of influence quantities. 2
6 Interferences . 2
7 Apparatus. 3
8 Reagents. 3
9 Preparation of dosimeters. 3
10 Calibration of the dosimetry system . 4
11 Application of dosimetry system . 5
12 Minimum documentation requirements. 6
13 Measurement uncertainty. 6
14 Keywords. 6
Figure 1
Relative response of dichromate dosimeter as a function of irradiation temperature.Afit of the data using Eq 1 yields
−5
fit parameters as follows: b = 1.021;b = −6.259 × 10 ;b = 1.806. 3
0 1 2
Figure 2
Responseofhigh-rangedosimeterintermsof∆A as a function of absorbed dose to water. A least-squares third order
−4 −2
polynomial fit (see Eq 4) of the data yields fit parameters as follows:b = 7.515 × 10 ;b = 1.745 × 10 ;
0 1
−6 −7
b = 3.485 × 10 ;b = −2.765 × 10 . 5
3
2
Figure 3
Responseofthelow-rangedosimeterintermsof∆A as a function o
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.