Surgical drapes, gowns and clean air suits, used as medical devices, for patients, clinical staff and equipment - Part 1: General requirements for manufacturers, processors and products

This standard specifies information to be supplied to users and third party verifiers, in addition to the usual labelling of medical devices (see EN 980 and EN 1041), concerning manufacturing and processing requirements. This standard gives general guidance on the characteristics of single-use and reusable surgical gowns, surgical drapes and clean air suits used as medical devices for patients, clinical staff and equipment. It is intended to prevent the transmission of infective agents between patients and clinical staff during surgical and other invasive procedures.
Surgical masks, surgical gloves, packaging materials, foot and head wear and incision drapes are not covered by EN 13795. Requirements for medical gloves are given in the EN 455 series of European Standards and packaging materials are covered by the EN 868 series. Requirements for surgical masks and head coverings will be specified in future CEN/TC 205 standards.
EN 13795 does not cover requirements for flammability of products used in laser surgery. Suitable test methods for flammability and resistance to penetration by laser radiation, together with an appropriate classification system, are given in EN ISO 11810. Additional essential requirements that apply to surgical clothing and drapes are covered by other European Standards.

Operationsabdecktücher, -mäntel und Rein-Luft-Kleidung zur Verwendung als Medizinprodukte für Patienten, Klinikpersonal und Geräte - Teil 1: Allgemeine Anforderungen für Hersteller, Aufbereiter und Produkte

Diese Norm legt die in Ergänzung zur üblichen Kennzeichnung von Medizinprodukten (siehe EN 980 und
EN 1041) zu liefernden Informationen für die Anwender und die Prüfer einer Dritten Seite hinsichtlich der Anforderungen
an die Herstellung und Wiederaufbereitung fest. Diese Norm bietet allgemeine Leitlinien für die
Eigenschaften von Einweg- und Mehrweg-Operationsmänteln, Abdecktüchern und Rein-Luft-Kleidung zur
Verwendung als Medizinprodukte für Patienten, Klinikpersonal und Geräte. Sie dient dazu, die Übertragung
infektiöser Agenzien zwischen den Patienten und dem Klinikpersonal während chirurgischer und sonstiger
invasiver Eingriffe zu verhindern.
Gesichtsmasken, Operationshandschuhe, Verpackungsmaterial, Fußbekleidung und Kopfbedeckungen sowie
Inzisionsfolien werden in Reihe EN 13795 nicht behandelt. Anforderungen an Operationshandschuhe sind in
den Europäischen Normen der Reihe EN 455 festgelegt, und Verpackungsmaterialien sind von den Normen
der Reihe EN 868 abgedeckt. Anforderungen an Gesichtsmasken und Kopfbedeckungen werden in weiteren
Normen des CEN/TC 205 festgelegt.
Reihe EN 13795 befasst sich nicht mit Anforderungen an die Entflammbarkeit von in der Laserchirurgie verwendeten
Produkten. Geeignete Prüfverfahren für die Entflammbarkeit und den Widerstand gegen die Durchdringung
durch Laserstrahlung, zusammen mit einer geeigneten Klassifizierung, sind in EN ISO 11810 angegeben.
Zusätzliche grundlegende Anforderungen, die für Operationsmäntel und -abdecktücher gelten, sind in
weiteren Europäischen Normen festgelegt.

Champs chirurgicaux, casaques et tenues de bloc, utilisés en tant que dispositifs médicaux pour les patients, le personnel et les équipements - Partie 1: Exigences générales pour les fabricants, les prestataires et les produits

L’EN 13795-1 spécifie les informations à fournir aux utilisateurs et aux tiers vérificateurs en plus de l’étiquetage
standard des dispositifs médicaux (voir l’EN 980 et l’EN 1041), concernant les exigences relatives à la fabrication et
au traitement. Cette norme donne des indications générales sur les caractéristiques des champs chirurgicaux, des
casaques et des tenues de bloc à usage unique et réutilisables utilisés en tant que dispositifs médicaux pour,
les patients, le personnel et les équipements. Elle vise à prévenir la transmission d’agents infectieux entre les patients
et le personnel chirurgical au cours des interventions chirurgicales et autres interventions invasives.
L’EN 13795 ne traite pas des masques chirurgicaux, des gants chirurgicaux, des matériaux d’emballage, des
surchaussures, des articles coiffants et des champs à inciser. Les exigences relatives aux gants médicaux sont
traitées dans la série des Normes européennes EN 455, et les matériaux d’emballage sont abordés dans la série
EN 868. Les exigences relatives aux masques chirurgicaux et aux articles coiffants seront spécifiées dans des
normes élaborées ultérieurement par le CEN/TC 205.
L’EN 13795 n’aborde pas les exigences relatives à l’inflammabilité des produits utilisés en chirurgie laser.
Les méthodes d’essai appropriées à l’inflammabilité et à la résistance à la pénétration par rayonnement laser, ainsi
qu’un système de classification adéquat, sont traités dans l’EN ISO 11810. D’autres exigences essentielles
applicables aux vêtements et aux champs chirurgicaux sont traitées par d’autres normes européennes.

Operacijska pokrivala, pregrinjala in plašči ter čista oblačila, ki se uporabljajo kot medicinski pripomočki za paciente, zdravstveno osebje in opremo - 1. del: Splošne zahteve za proizvajalce, predelovalce in izdelke

General Information

Status
Withdrawn
Publication Date
12-Oct-2009
Withdrawal Date
31-May-2011
Technical Committee
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
01-Jun-2011
Due Date
24-Jun-2011
Completion Date
01-Jun-2011

Relations

Buy Standard

Standard
EN 13795-1:2003+A1:2009
English language
15 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Operationsabdecktücher, -mäntel und Rein-Luft-Kleidung zur Verwendung als Medizinprodukte für Patienten, Klinikpersonal und Geräte - Teil 1: Allgemeine Anforderungen für Hersteller, Aufbereiter und ProdukteChamps chirurgicaux, casaques et tenues de bloc, utilisés en tant que dispositifs médicaux pour les patients, le personnel et les équipements - Partie 1: Exigences générales pour les fabricants, les prestataires et les produitsSurgical drapes, gowns and clean air suits, used as medical devices, for patients, clinical staff and equipment - Part 1: General requirements for manufacturers, processors and products11.140Oprema bolnišnicHospital equipmentICS:Ta slovenski standard je istoveten z:EN 13795-1:2002+A1:2009SIST EN 13795-1:2003+A1:2009en,fr,de01-november-2009SIST EN 13795-1:2003+A1:2009SLOVENSKI
STANDARD



SIST EN 13795-1:2003+A1:2009



EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EN 13795-1:2002+A1
July 2009 ICS 11.140 Supersedes EN 13795-1:2002English Version
Surgical drapes, gowns and clean air suits, used as medical devices, for patients, clinical staff and equipment - Part 1: General requirements for manufacturers, processors and products
Champs chirurgicaux, casaques et tenues de bloc, utilisés en tant que dispositifs médicaux pour les patients, le personnel et les équipements - Partie 1: Exigences générales pour les fabricants, les prestataires et les produits
Operationsabdecktücher, -mäntel und Rein-Luft-Kleidung zur Verwendung als Medizinprodukte für Patienten, Klinikpersonal und Geräte - Teil 1: Allgemeine Anforderungen für Hersteller, Aufbereiter und Produkte This European Standard was approved by CEN on 2 October 2002 and includes Amendment 1 approved by CEN on 13 June 2009.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CEN member into its own language and notified to the CEN Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre:
Avenue Marnix 17,
B-1000 Brussels © 2009 CEN All rights of exploitation in any form and by any means reserved worldwide for CEN national Members. Ref. No. EN 13795-1:2002+A1:2009: ESIST EN 13795-1:2003+A1:2009



EN 13795-1:2002+A1:2009 (E) 2 Contents page Foreword . 3Introduction . 41Scope . 42Terms and definitions . 43Information to be supplied by the manufacturer or processor . 74Manufacturing and processing requirements . 95Testing requirements . 10Annex A (informative)
Comfort . 11Annex B (informative)
Adhesion for fixation and wound isolation . 12Annex C (informative)
Prevention of infection in the operating theatre . 13Annex ZA (informative)
!!!!Relationship between this European Standard and the Essential Requirements of EU Directive 93/42/EEC on Medical Devices"""" . 14Bibliography . 15 SIST EN 13795-1:2003+A1:2009



EN 13795-1:2002+A1:2009 (E) 3
Foreword This document (EN 13795-1:2002+A1:2009) has been prepared by Technical Committee CEN /TC 205 "Non-active medical devices", the secretariat of which is held by DIN. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by January 2010, and conflicting national standards shall be withdrawn at the latest by March 2010. This document includes Amendment 1, approved by CEN on 2009-06-13. This document supersedes EN 13795-1:2002. The start and finish of text introduced or altered by amendment is indicated in the text by tags !". This document has been prepared under a mandate given to CEN by the European Commission and the European Free Trade Association, and supports essential requirements of EU Directive(s).
For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of this document. Annexes A, B and C are informative. This document includes a Bibliography. EN 13795 is expected to consist of the following parts under the general title "Surgical drapes, gowns and clean air suits, used as medical devices, for patients, clinical staff and equipment": Part 1: General requirements for manufacturers, processors and products Part 2: Test methods Part 3: Performance requirements and performance levels Originally EN 13795 was also to include Part 3: Test method for resistance to dry microbial penetration and Part 4: Test method for resistance to wet microbial penetration. However, it has been decided that these parts will now be developed by the Vienna Agreement/CEN lead route in conjunction with ISO/TC 94/SC 13. As a result, what was to have been EN 13795-3 will be published as EN ISO 22612 Clothing for protection against infectious agents – Test method for resistance to penetration by biologically contaminant dust through protective clothing materials, what was to have been EN 13795-4 will be published as EN ISO 22610 Clothing for protection against infectious agents – Test method for determination of penetration by bacteria through protective clothing materials and what was to have been EN 13795-5 will be published as EN 13795-3. According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard : Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom. SIST EN 13795-1:2003+A1:2009



EN 13795-1:2002+A1:2009 (E) 4
Introduction The transmission of infective agents during invasive surgical procedures can occur in several ways (see annex C). !Surgical drapes, including the intended use as a sterile field, gowns and clean air suits are used to minimize the spread of infective agents to and from patients’ operating wounds, thereby helping to prevent post-operative wound infections (see Annex C)." The performance required of coverings for patients, clinical staff and equipment varies with, for example, the type and duration of the procedure, the degree of wetness of the operation field, the degree of mechanical stress on the materials and the susceptibility of the patient to infection. The use of surgical gowns with resistance to the penetration of liquids can also diminish the risk to the operating staff from blood-borne infective agents carried in blood or body fluids. The EN 13795 series of European Standards, together with EN ISO 22610 and EN ISO 22612, is intended to assist the communication between users, manufacturers and third party verifiers with regard to material or product characteristics. It focuses on relevant Essential Requirements arising from the Medical Device Directive 93/42/EEC. The general requirements and guidance in EN 13795-1 are expected to be of help to manufacturers, test houses and users when designing, processing, assessing and selecting products. It is the intention of EN 13795 to ensure the same level of safety from single-use and reusable surgical clothing and drapes throughout their useful life. 1 Scope This standard specifies information to be supplied to users and third party verifiers, in addition to the usual labelling of medical devices (see EN 980 and EN 1041), concerning manufacturing and processing requirements. This standard gives general guidance on the characteristics of single-use and reusable surgical gowns, surgical drapes and clean air suits used as medical devices for patients, clinical staff and equipment. It is intended to prevent the transmission of infective agents between patients and clinical staff during surgical and other invasive procedures. Surgical masks, surgical gloves, packaging materials, foot and head wear and incision drapes are not covered by EN 13795. Requirements for medical gloves are given in the EN 455 series of European Standards and packaging materials are covered by the EN 868 series. Requirements for surgical masks and head coverings will be specified in future CEN/TC 205 standards. EN 13795 does not cover requirements for flammability of products used in laser surgery. Suitable test methods for flammability and resistance to penetration by laser radiation, together with an appropriate classification system, are given in EN ISO 11810. Additional essential requirements that apply to surgical clothing and drapes are covered by other European Standards. 2 Terms and definitions For the purposes of this European Standard, the following terms and definitions apply. 2.1 clean air suit suit intended and shown to minimize contamination of the operating wound by the wearer's skin scales carrying infective agents via the operating room air thereby reducing the risk of wound infection NOTE Unlike the suit usually worn in the operation room, the clean air suit is designed to reduce the operating room air contamination by personnel. 2.2 cleanliness freedom from unwanted foreign matter SIST EN 13795-1:2003+A1:2009



EN 13795-1:2002+A1:2009 (E) 5 NOTE Such matter can be micro-organisms, organic residues or particulate matter. 2.2.1 cleanliness — microbial freedom from population of viable micro-organisms on a product and/or a package NOTE In practical use microbial cleanliness is often referred to as “bioburden“. 2.2.2 cleanliness — particulate matter freedom from particles that are contaminating a material and can be released but are not generated by mechanical impact 2.3 critical product area product area with a greater probability to be involved in the transfer of infective agents to or from the wound, e.g. front and sleeves of surgical gowns 2.4 drapeability ability of a material to conform to a given shape or object 2.5 fabric cloth made from yarn or fibres by weaving, knitting and/or other types of binding or manufacture 2.6 fixation adhesion of a surgical drape to the patient for the purpose of wound isolation NOTE See annex B. 2.7 infective agent micro-organism that has been shown to cause a wound infection or that might cause infection in a member of the surgical team or the patient 2.8 linting release of fibre fragments and other particles during handling and use NOTE These fragments and particles are originally from the fabric itself. 2.9 manufacturer natural or legal person with responsibility for the design, manufacture, packaging and labelling of a device before it is placed on the market under his own name, regardless of whether these operations are carried out by that person himself or on his behalf by a third party NOTE From the Medical Device Directive 93/42/EEC. 2.10 processor natural or legal person who processes reusable product items so that their performance complies with the requirements of this standard NOTE A processor who places a product on the market is a manufacturer in the sense of this standard. 2.11 resistance to liquid penetration ability of material to resist the penetration of liquid(s) from one side of the material through to the other SIST EN 13795-1:2003+A1:2009



EN 13795-1:2002+A1:2009 (E) 6 2.12 resistance to microbial penetration ability of material(s) to withstand penetration of micro-organisms from one side through to the other 2.12.1 dry penetration effect of a combination of air movement and mechanical action by vibration on microbial penetration in dry condition 2.12.2 wet penetration effect of combination of wetness, pressure and rubbing on microbial penetration 2.13 reusable product product intended by the manufacturer to be reprocessed and reused 2.14 single-use product product intended by the manufacturer to be used for only one surgical procedure before disposal ! 2.15 sterile field area created by sterile surgical drape material where aseptic technique is practised NOTE A sterile field can be practised e.g. on a back table." 2.16 surgical drape drape covering the patient or equipment to prevent transfer of infective agents 2.17 surgical gown gown worn by a member of a surgical team to prevent transfer of infective agents 2.18 surgical procedure surgical intervention penetrating skin or mucosa, performed by a surgical team 2.18.1 clean operation operation performed on uninfected, non-traumatised tissue and in which respiratory, alimentary or genito-urinary tracts are not entered 2.18.2 infection-prone operation operation where the nature of the operation or the condition of the patient is such that infection can occur with minimal contamination 2.18.3 invasive surgical procedure procedure that reaches the inside of the body through the body surface 2.19 water-vapour resistance water-vapour pressure difference between the two faces of a material divided by the resultant evaporative heat flux per unit area in the direction of the gradient. The evaporative heat flux can consist of both diffusive and convective components SIST EN 13795-1:2003+A1:2009



EN 13795-1:2002+A1:2009 (E) 7 3 Information to b
...