Water quality - General requirements and performance test procedures for water monitoring equipment - Measuring devices

This standard is Part 2 of a two part series of standards defining the technical requirements for testing the performance of water monitoring equipment. This standard defines general requirements and performance test procedures for automated measuring devices (AMD). These devices may be permanently installed or portable and are capable of making continuous real-time re measurements in water of chemicals or physical parameters.
Automated measuring devices are used extensively in the industrial process, water industries and by regulators for compliance and surveillance monitoring and process control purposes. The standard supports the Industrial Emissions Directive, Marine Directive and the Water Framework Directive.
The test parameters are performance characteristics that identify the capability of an AMD to provide reliable measurements. They include: bias, repeatability and errors caused by variations in linearity, drift, interferents, salinity compensation, output impedance, supply voltage, ambient temperature, relative humidity, incident light, sample temperature, sample flow-rate, sample pressure. The overall measurement reliability of an AMD is calculated in the form of a combined performance characteristic expressed as measurement uncertainty.

Wasserbeschaffenheit - Allgemeine Anforderungen und Testverfahren zur Leistungsprüfung von Geräten zum Wassermonitoring - Messgeräte

Dieses Dokument legt allgemeine Anforderungen und Leistungsprüfungsverfahren für tragbare und fest installierte Messgeräte fest, die in einer Inline  oder Online Betriebsposition verwendet werden, um physikalische und chemische Bestimmungsgrößen in Wasser zu messen. Es deckt keine chemischen Test-Kits und Laboranalysegeräte ab.
Die allgemeinen Anforderungen umfassen bestimmte Merkmale, die für den Gebrauch des Geräts, zur Information des Anwenders und für die zugehörigen Dokumente notwendig sind.
Die Prüfverfahren legen einheitliche Verfahren fest, die beim Bestimmen von wichtigen Leistungskenngrößen von MDs verwendet werden. Die Leistungsprüfungen umfassen unter Labor- und Feldbedingungen durchgeführte Prüfungen. Es ist zu beachten, dass bestimmte Prüfverfahren für einige Geräte nicht geeignet sind.
Statistische Verfahren für die Auswertung der Prüfdaten sind definiert.
Beispielwerte für Leistungskenngrößen für eine Auswahl an MDs zum Monitoring von Abläufen von Abwasser und Vorflutern sind detailliert in Anhang A als Hilfestellung angegeben.
Dieses Dokument fordert von den Herstellern eines MD die Bereitstellung von mehr technischen Daten zur Verifizierung als EN ISO 15839:2006 [7]. Deshalb ist EN ISO 15839 [7] eine bessere Hilfestellung für Hersteller, die ein neues Gerät charakterisieren möchten, während dieses Dokument den Schwerpunkt eher auf die Anwenderanforderungen für die Verifizierung der Herstellerangaben legt.

Qualité de l'eau - Exigences générales et modes opératoires d'essai de performance pour les équipements de surveillance de l'eau - Dispositifs de mesure

La présente Norme européenne spécifie les exigences générales et les modes opératoires d'essai de performance pour les dispositifs de mesure (DM) portables et fixes qui sont utilisés in situ, en ligne et en ligne avec boucle de prélèvement afin de mesurer les caractéristiques physiques et chimiques à déterminer dans l'eau. Elle exclut les dispositifs « at-line », tels que les tests chimiques en kit, et les dispositifs hors ligne, tels que les analyseurs de laboratoire.
Les exigences générales comprennent les installations fonctionnelles dont les DM ont besoin pour répondre aux besoins des utilisateurs pour leurs applications, ainsi que les informations qui doivent être incluses dans les documents associés.
Les modes opératoires d'essai spécifient les méthodes uniformes à utiliser pour déterminer les caractéristiques de performance clés des DM. Les essais de performance comprennent des essais effectués dans des conditions de laboratoire et des conditions réelles.
Des procédures statistiques sont définies pour l'évaluation des données d'essai. Il est reconnu que pour certains dispositifs, certains modes opératoires d'essai ne sont pas applicables.
Des exemples de valeurs de caractéristiques de performance pour une sélection de DM destinés à la surveillance des effluents d'eaux usées et des eaux réceptrices sont fournis à l'Annexe A à titre d’information.
La présente Norme européenne exige que le fabricant d'un DM fournisse davantage de données techniques pour la vérification que ne l'exige l'EN ISO 15839:2006 [5]. Par conséquent, l'EN ISO 15839 sera d'une plus grande utilité pour les fabricants qui souhaitent caractériser un nouveau dispositif, tandis que la présente Norme européenne est davantage axée sur les exigences de l'utilisateur relatives à la vérification des allégations du fabricant.

Kakovost vode - Splošne zahteve in postopki preskušanja zmogljivosti opreme za monitoring vode - Merilniki

Ta standard je 2. del skupine standardov iz dveh delov, ki določa tehnične zahteve za preskušanje zmogljivosti opreme za monitoring vode. Ta standard določa splošne zahteve in postopke preskušanja zmogljivosti za samodejne merilnike (AMD). Te naprave so lahko trajno nameščene ali prenosne in omogočajo stalne vnovične meritve kemikalij ali fizičnih parametrov v vodi v realnem času. Samodejni merilniki se obsežno uporabljajo v industrijskem procesu, vodni industriji in regulatorjih za namene nadzora skladnosti in nadzornega monitoringa ter za nadzor procesov. Ta standard podpira direktivo o industrijskih emisijah, direktivo o morski strategiji in okvirno direktivo o vodah. Preskusni parametri so zmogljivostne lastnosti, ki označujejo zmožnost samodejnega merilnika, da zagotovi zanesljive meritve. Vključujejo: sistematični pogrešek, ponovljivost in napake, ki jih povzročajo razlike v linearnosti, odstopanje, moteče komponente, kompenzacijo slanosti, izhodno impedanco, napajalno napetost, temperaturo okolice, relativno vlažnost, vpadno svetlobo, vzorčno temperaturo, vzorčni pretok in vzorčni tlak. Celotna merilna zanesljivost samodejnega merilnika se izračuna v obliki kombinirane zmogljivostne lastnosti, ki je izražena kot merilna negotovost.

General Information

Status
Published
Public Enquiry End Date
02-Mar-2017
Publication Date
10-Apr-2019
Technical Committee
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
04-Feb-2019
Due Date
11-Apr-2019
Completion Date
11-Apr-2019

Buy Standard

Standard
SIST EN 17075:2019
English language
70 pages
sale 10% off
Preview
sale 10% off
Preview

e-Library read for
1 day

Standards Content (sample)

SLOVENSKI STANDARD
SIST EN 17075:2019
01-maj-2019
Kakovost vode - Splošne zahteve in postopki preskušanja zmogljivosti opreme za
monitoring vode - Merilniki
Water quality - General requirements and performance test procedures for water
monitoring equipment - Measuring devices
Wasserbeschaffenheit - Allgemeine Anforderungen und Testverfahren zur
Leistungsprüfung von Geräten zum Wassermonitoring - Messgeräte

Qualité de l'eau - Exigences générales et modes opératoires d'essai de performance

pour les équipements de surveillance de l'eau - Dispositifs de mesure
Ta slovenski standard je istoveten z: EN 17075:2018
ICS:
13.060.45 Preiskava vode na splošno Examination of water in
general
SIST EN 17075:2019 en,fr,de

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN 17075:2019
---------------------- Page: 2 ----------------------
SIST EN 17075:2019
EN 17075
EUROPEAN STANDARD
NORME EUROPÉENNE
December 2018
EUROPÄISCHE NORM
ICS 13.060.45
English Version
Water quality - General requirements and performance
test procedures for water monitoring equipment -
Measuring devices

Qualité de l'eau - Exigences générales et modes Wasserbeschaffenheit - Allgemeine Anforderungen und

opératoires d'essai de performance pour les Testverfahren zur Leistungsprüfung von Geräten zum

équipements de surveillance de l'eau - Dispositifs de Wassermonitoring - Messgeräte

mesure
This European Standard was approved by CEN on 18 June 2018.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this

European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references

concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN

member.

This European Standard exists in three official versions (English, French, German). A version in any other language made by

translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management

Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,

Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,

Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,

Turkey and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels

© 2018 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 17075:2018 E

worldwide for CEN national Members.
---------------------- Page: 3 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)
Contents Page

European foreword ....................................................................................................................................................... 4

Introduction .................................................................................................................................................................... 5

1 Scope .................................................................................................................................................................... 6

2 Normative references .................................................................................................................................... 6

3 Terms and definitions ................................................................................................................................... 6

4 Symbols ............................................................................................................................................................ 12

5 Principles ........................................................................................................................................................ 12

6 General requirements ................................................................................................................................ 13

6.1 Requirements for MDs ............................................................................................................................... 13

6.2 Requirements for MDs associated documents .................................................................................. 14

7 Performance characteristics .................................................................................................................... 15

7.1 Performance characteristics determined by laboratory testing ................................................ 15

7.2 Performance characteristics determined by field testing ............................................................. 16

8 Performance testing .................................................................................................................................... 17

8.1 Quality requirements for testing ............................................................................................................ 17

8.2 General requirements for testing ........................................................................................................... 18

8.3 Test conditions .............................................................................................................................................. 19

8.4 Reporting ........................................................................................................................................................ 20

9 Laboratory test procedures ...................................................................................................................... 20

9.1 Guidance for establishing a test plan .................................................................................................... 20

9.2 Verification by inspection ......................................................................................................................... 21

9.3 Performance tests ........................................................................................................................................ 21

9.3.1 Response time ............................................................................................................................................... 21

9.3.2 Bias, linearity, repeatability and LOQ ................................................................................................... 25

9.3.3 Interference effects ..................................................................................................................................... 25

9.3.4 Ambient temperature and relative humidity .................................................................................... 29

9.3.5 Sample temperature ................................................................................................................................... 30

9.3.6 Sample flow-rate .......................................................................................................................................... 30

9.3.7 Sample pressure ........................................................................................................................................... 31

9.3.8 Output impedance ....................................................................................................................................... 31

9.3.9 Supply voltage ............................................................................................................................................... 32

9.3.10 7-day drift ....................................................................................................................................................... 33

9.3.11 Loss of power ................................................................................................................................................. 33

9.3.12 Warm-up drift................................................................................................................................................ 34

10 Field test procedures .................................................................................................................................. 34

10.1 Objective of field test .................................................................................................................................. 34

10.2 Planning for the field test .......................................................................................................................... 35

10.3 Error under field test conditions ............................................................................................................ 37

10.4 Response time ............................................................................................................................................... 37

10.5 Variation in sensitivity ............................................................................................................................... 38

10.6 Maintenance ................................................................................................................................................... 38

10.7 Up-time ............................................................................................................................................................ 39

---------------------- Page: 4 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)

Annex A (informative) Example values for performance characteristics for a selection of

MDs for monitoring waste water effluents and receiving waters ............................................... 41

Annex B (normative) Evaluation of performance test data ........................................................................ 47

Annex C (informative) Example calculations ................................................................................................... 53

Annex D (informative) Uncertainty of the reference quantity and error calculations .................... 62

Annex E (informative) Example test report ..................................................................................................... 64

Bibliography ................................................................................................................................................................. 68

---------------------- Page: 5 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)
European foreword

This document (EN 17075:2018) has been prepared by Technical Committee CEN/TC 230 “Water

analysis”, the secretariat of which is held by DIN.

This European Standard shall be given the status of a national standard, either by publication of an

identical text or by endorsement, at the latest by May 2019, and conflicting national standards shall be

withdrawn at the latest by May 2019.

Attention is drawn to the possibility that some of the elements of this document may be the subject of

patent rights. CEN shall not be held responsible for identifying any or all such patent rights.

According to the CEN-CENELEC Internal Regulations, the national standards organisations of the

following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,

Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia,

France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta,

Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,

Turkey and the United Kingdom.
---------------------- Page: 6 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)
Introduction

This document defines general requirements and test procedures for verifying the performance of

measuring devices (MDs) used to monitor the quality of a wide range of waters including drinking

waters, waste waters, and natural waters. It covers both portable measuring devices (PMDs) and fixed

position measuring devices (FMDs). These devices include: sensors, single and multi-parameter

instruments, discrete and batch instruments, probes and sondes. It excludes chemical test kits. For the

purposes of this document the acronym MD(s) is used except where it is necessary to be specific about

the particular type (e.g. PMDs, FMDs) or component of a MD (e.g. sensor).

This document is associated with EN 16479 [1] which covers automated sampling devices (samplers)

for water and waste water.

The general requirements include several features that are necessary to meet users’ applications and

information that has to be included in associated documents.

The performance tests comprise testing carried out under laboratory and field conditions They are

designed to determine, in a systematic and consistent way, the capability of MDs to make reliable

measurements. The testing focuses on key performance characteristics. Statistical procedures are

defined for evaluation of the test data.

The range of measurements over which the test procedures will be applied, the test range, is not

specified. It is for the MD manufacturer and/or the user to decide on the test range. Similarly, it is for

the MD manufacturer and/or the user to decide on the intended uses (applications) which will inform

the design of the field trial.

Water monitoring equipment is widely used for compliance monitoring purposes under national and

European regulations. This document supports the requirements of the following EU Directives:

— Industrial Emissions Directive (2010/75/EU) [2];
— Water Framework Directive (2000/60/EC) [3];
— Marine Strategy Framework Directive (2008/56/EC) [4];
— Drinking Water Directive (98/83/EC) [5];

— Technical Specifications for Chemical Analysis and Monitoring of Water Status (2009/90/EC) [6].

---------------------- Page: 7 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)
1 Scope

This document specifies general requirements and performance test procedures for portable and fixed

position measuring devices that are used in an in-line or online operating position to measure physical

and chemical measurands in water. It excludes chemical test kits and laboratory analysers.

The general requirements include functional facilities that MDs need to meet users’ applications and

information that needs to be included in associated documents.

The test procedures specify uniform methods to be used when determining key performance

characteristics of MDs. The performance tests comprise testing carried out under laboratory and field

conditions. It is recognized that for some devices certain test procedures are not applicable.

Statistical procedures are defined for evaluation of the test data.

Example values for performance characteristics for a selection of MDs for monitoring waste water

effluents and receiving waters are detailed in Annex A for guidance.

This document requires the manufacturer of a MD to provide more technical data for verification than

does EN ISO 15839:2006 [7]. Consequently, EN ISO 15839 [7] will be of greater assistance to

manufacturers wishing to characterize a new device whereas this document is more focussed on user

requirements for the verification of manufacturer’s claims.
2 Normative references

The following documents are referred to in the text in such a way that some or all of their content

constitutes requirements of this document. For dated references, only the edition cited applies. For

undated references, the latest edition of the referenced document (including any amendments) applies.

EN ISO 5814:2012, Water quality - Determination of dissolved oxygen - Electrochemical probe method

(ISO 5814:2012)
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.

ISO and IEC maintain terminological databases for use in standardization at the following addresses:

— IEC Electropedia: available at http://www.electropedia.org/
— ISO Online browsing platform: available at http://www.iso.org/obp
3.1
measuring device

device, used in an in-line or on-line operating position, which continuously (or at a given frequency)

gives an output signal proportional to the value of one or more measurands in waters which it

measures
Note 1 to entry: The device can be portable or fixed in position.

Note 2 to entry: The term “on-line measuring device” is often used for a MD used in an online position.

[SOURCE: EN ISO 15839:2006 [7], 3.26, modified]
---------------------- Page: 8 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)
3.2
portable measuring device
PMD

measuring device that can be moved from one measuring point to another and used in an in-line or on-

line operating position
3.3
fixed measuring device
FMD

measuring device that can be fixed in position and used in an in-line or on-line operating position

3.4
sensor

electronic device that senses a physical condition or chemical compound and delivers an electronic

signal proportional to the observed characteristic
[SOURCE: ISO/IEC 19762:2016 [8], 06.02.08]
3.5
in-line measuring device
in situ measuring device

system of automatic measurement in which at least the sensor is sited in the body of water

[SOURCE: ISO 6107-2:2006[9], 54, modified — term “analysis” replaced by term “measuring device”

and within definition “analysis” replaced by “measurement”]
3.6
on-line measuring device

system of automatic measurement in which the sample is taken from the body of water through a probe

to the measuring device by means of an appropriate conduit
Note 1 to entry: Sometimes referred to as an extractive measuring device.

[SOURCE: ISO 6107-2:2006+A1:2012 [10], 71, modified — term “analysis” replaced by term “measuring

device”, within definition “analysing equipment” replaced by “measuring device” and Note 1 to entry

added]
3.7
percentage error
error in measurement expressed as a percentage of the reference value
3.8
measurement bias
bias
estimate of a systematic measurement error
[SOURCE: ISO/IEC Guide 99:2007 [11], 2.18]
3.9
measurement repeatability
repeatability
precision under a set of repeatability conditions of measurement
[SOURCE: ISO/IEC Guide 99 [11], 2.21]
---------------------- Page: 9 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)

Note 1 to entry: In this standard the ability of a MD to provide closely similar indications for repeated

measurements of the same measurand under the same conditions of measurement.
3.10
limit of quantification
LOQ

stated multiple of the limit of detection, for example two or three times the limit of detection, at a

concentration of the measurand that can reasonably be determined with an acceptable level of accuracy

and precision

Note 1 to entry: Limit of quantification can be calculated using an appropriate standard or sample, and may be

obtained from the lowest calibration point on the calibration curve (excluding the blank). See ISO 6879 [12].

[SOURCE: ISO 6107-2:2006+A1:2012 [10], 61]
3.11
linearity

degree to which there is a straight-line relationship between the (mean) result of measurement (signal)

and the quantity (concentration) of the component to be determined
[SOURCE: EN ISO 11885:2009 [13], 3.9, modified]
3.12
drift
slow change of output, at a constant input, of a measuring system
[SOURCE: EN ISO 15796:2005 [14], 2.8]
3.13
response time (t )

time interval between the instant when a continuous measuring device is subjected to an abrupt change

in measurand value and the instant when the readings cross the limits of (and remain inside) a band

defined by the 90 % and the 110 % of the difference between the initial and final value of the abrupt

change
[SOURCE: EN ISO 15839:2006 [7], 3.3, modified]
3.14
performance characteristics
set of parameters describing the performance of a MD
[SOURCE: EN ISO 15839:2006 [7], 3.27, modified]
---------------------- Page: 10 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)
3.15
measurement error
error of measurement
error
measured quantity value minus a reference quantity value
Note 1 to entry: The concept of “measurement error” can be used both:

a) when there is a single reference quantity value to refer to, which occurs if a calibration is made by

means of a measurement standard with a measured quantity value having a negligible measurement

uncertainty or if conventional quantity value is given, in which case the measurement error is known;

and

b) if a measurand is supposed to be represented by a unique true quantity value or a set of true quantity

values of negligible range, in which case the measurement error is not known.

Note 2 to entry: Measurement error should not be confused with production error or mistake.

[SOURCE: ISO/IEC Guide 99:2007 [11], 2.16]
3.16
measurement uncertainty
uncertainty of measurement
uncertainty

non-negative parameter characterizing the dispersion of the quantity values being attributed to a

measurand, based on the information used

Note 1 to entry: The parameter may be, for example, a standard deviation called standard measurement

uncertainty (or a specified multiple of it), or the half-width of an interval, having a stated coverage probability.

Note 2 to entry: In general, for a given set of information, it is understood that the measurement uncertainty is

associated with a stated quantity value attributed to the measurand. A modification of this value results in a

modification of the associated uncertainty.

Note 3 to entry: Measurement uncertainty includes components arising from systematic effects, such as

components associated with corrections and the assigned quantity values of measurement standards, as well as

the definitional uncertainty. Sometimes estimated systematic effects are not corrected for but, instead, associated

measurement uncertainty components are incorporated.

Note 4 to entry: Measurement uncertainty comprises, in general, many components. Some of these may be

evaluated by Type A evaluation of measurement uncertainty from the statistical distribution of the quantity values

from series of measurements and can be characterized by standard deviations. The other components, which may

be evaluated by Type B evaluation of measurement uncertainty, can also be characterized by standard deviations,

evaluated from probability density functions based on experience or other information.

[SOURCE: ISO/IEC Guide 99:2007 [11], 2.26]
3.17
standard measurement uncertainty
standard uncertainty of measurement
standard uncertainty
measurement uncertainty expressed as a standard deviation
[SOURCE: ISO/IEC Guide 99:2007 [11], 2.30]
---------------------- Page: 11 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)
3.18
combined standard measurement uncertainty
combined standard uncertainty

standard measurement uncertainty that is obtained using the individual standard measurement

uncertainties associated with the input quantities in a measurement model

Note 1 to entry: In case of correlations of input quantities in a measurement model, covariances must also be

taken into account when calculating the combined standard measurement uncertainty, see also

ISO/IEC Guide 98-3:2008 [15], 2.3.4.
[SOURCE: ISO/IEC Guide 99:2007 [11], 2.31]
3.19
expanded measurement uncertainty
expanded uncertainty

product of a combined standard measurement uncertainty and a factor larger than the number one

Note 1 to entry: Expanded measurement uncertainty is termed “overall uncertainty” in paragraph 5 of

Recommendation INC-1 (1980) (see the GUM) and simply “uncertainty” in IEC documents.

Note 2 to entry: The factor depends upon the type of probability distribution of the output quantity in a

measurement model and on the selected coverage probability.

Note 3 to entry: The term “factor” in this definition refers to a coverage factor.

[SOURCE: ISO/IEC Guide 99:2007 [11], 2.35]
3.20
combined performance characteristic

combination of individual performance characteristics expressed as an expanded measurement

uncertainty (with a coverage factor of 2)
3.21
test range
measuring range over which the MD is tested
3.22
output

reading, or a digital, analogue, wired or wireless electrical signal, generated by a MD in response to a

measurand
3.23
rated operating condition

operating condition that must be fulfilled during measurement in order that a measuring instrument or

measuring system perform as designed

Note 1 to entry: Rated operating conditions generally specify intervals of values for a quantity being measured

and for any influence quantity.
[SOURCE: ISO/IEC Guide 99:2007 [11], 4.9]
---------------------- Page: 12 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)
3.24
interferent

physical, biological or chemical property of the sample, excluding the measurand, that affects the output

signal
[SOURCE: EN ISO 15839:2006 [7], 3.16, modified]
3.25
adjustment of a measuring system
adjustment

set of operations carried out on a measuring system so that it provides prescribed indications

corresponding to given values of a quantity to be measured

Note 1 to entry: Types of adjustment of a measuring system include zero adjustment of a measuring system,

offset adjustment, and span adjustment (sometimes called gain adjustment).

Note 2 to entry: After an adjustment of a measuring system, the measuring system usually needs to be

recalibrated.

Note 3 to entry: Adjustment of a measuring system should not be confused with calibration, which is a

prerequisite for adjustment.
[SOURCE: ISO/IEC Guide 99:2007 [11], 3.11]
3.26
reference method

method to be used to obtain the measurand value of the test waters, against which the readings from

the MD under test can be compared
3.27
measurand
quantity intended to be measured
[SOURCE: ISO/IEC Guide 99:2007 [11], 2.3, modified]
3.28
reference operating condition

operating condition prescribed for evaluating the performance of a measuring instrument or measuring

system or for comparison of measurement results
[SOURCE: ISO/IEC Guide 99:2007 [11], 4.11]
3.29
flow cell

housing within which a sensor can be held and through which the test water can be directed

3.30
up-time

fraction of the total time for which usable measuring data are available from the MD

3.31
warm-up period

interval between switching on power to the measurement circuit and the instant when the MD

produces a stable value when measuring a stable measurand
---------------------- Page: 13 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)
4 Symbols
For the purposes of this document, the following symbols apply.
b Bias
t Response time
X Change in error due to variations in linearity
S Repeatability
X Change in error due to variations in interferents
X Change in error due to incorrect salinity compensation
X Change in error due to variations in incident light
X Change in error due to variations in ambient temperature
X Change in error due to variations in relative humidity
X Change in error due to variations in sample temperature
X Change in error due to variations in sample flow-rate
X Change in error due to variations in sample pressure
X Change in error due to variations in output impedance
X Change in error due to variations in supply voltage
X Change in error due to variations in drift
W Warm–up drift
U Combined performance characteristic
5 Principles

The general requirements are based on experience of users’ needs when operating a MD in online or in-

line measuring positions in a range of applications.

The performance characteristics are parameters that identify the capability of a MD to provide reliable

measurements. They are determined as measurement uncertainty and expressed as percentage error

(see 3.7) except for bias and repeatability which are expressed as relative values in percentage. For low

concentration measurements where the reading approaches zero the percentage error value becomes

inapplicable. At low concentrations the uncertainty should be expressed as an absolute value. The

tables at Annex A list example values relevant to monitoring waste water effluents and receiving

waters.
Measurement reliability includes:

— the MD's measurement bias, LOQ, linearity, drift, repeatability and response time;

— operational influences arising from variations in supply voltage, output impedance, sample

temperature, sample flow-rate and sample pressure;

— environmental influences arising from variations in ambient temperature, relative humidity,

interferents and incident light.
---------------------- Page: 14 ----------------------
SIST EN 17075:2019
EN 17075:2018 (E)

The overall measurement reliability of a MD is captured by bringing the individual performance

characteristics together in the form of a combined performance characteristic expressed as

measurement uncertainty.

The laboratory performance tests are designed to determine the values for the specified performance

characteristics in a systematic and consistent way. Laboratory testing is used to determine each

performance characteristic in turn in a controlled environment.

Statistical procedures are defined for evaluation of the laboratory test data to calculate the individual

performance characteristic m
...

Questions, Comments and Discussion