kSIST FprEN IEC 60601-2-44:2026
(Main)Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography
- Status
- Not Published
- Public Enquiry End Date
- 31-Jan-2026
- Technical Committee
- IEMO - Electrical equipment in medical practice
- Current Stage
- 5020 - Formal vote (FV) (Adopted Project)
- Start Date
- 23-Jul-2026
- Due Date
- 10-Sep-2026
- Completion Date
- 23-Jul-2026
kSIST FprEN IEC 60601-2-44:2026 is the CLC/IEC particular standard for medical electrical equipment used as X-ray equipment for computed tomography (CT). It specifies basic safety and essential performance requirements for CT scanners, including radiation output control, dose display and recording, patient positioning, emergency stopping, and related documentation for manufacturers, test laboratories, and clinical users.
What does kSIST FprEN IEC 60601-2-44:2026 specify?
kSIST FprEN IEC 60601-2-44:2026 applies to the basic safety and essential performance of CT SCANNERS, including the X-RAY GENERATORS used in them. The scope is focused on head and/or body imaging with an X-ray source and imaging detector in a common toroidal protective enclosure.
| Part of the document | What it covers |
|---|---|
| Clause 201 | CT-specific scope, object, collateral standards, terms, general requirements, testing, classification, marking, protection against hazards, radiation control, PEMS, construction, ME SYSTEMS, EMC, and CT scanner requirements for radiotherapy treatment planning |
| Clause 202 | Electromagnetic compatibility requirements and tests |
| Clause 203 | Radiation protection requirements for diagnostic X-ray equipment, including dose display, dose checks, reporting, beam limitation, and positioning |
| Annex AA | Choosing LOADING FACTORS for tests |
| Annex CC | The CTDI concept and the relationship between CTDI and CTDI100 |
| Annex DD | Measuring CTDI |
| Annex EE | Template for listing REFERENCE CLINICAL PROTOCOLS |
kSIST FprEN IEC 60601-2-44:2026 excludes interventional systems covered by IEC 60601-2-43, cone-beam CT integrated into radiotherapy systems covered by IEC 60601-2-68, dental cone-beam CT covered by IEC 60601-2-63, radiotherapy simulators, and systems that form the image by a source other than an X-RAY TUBE. In practice, that helps a buyer or specifier decide whether this is the right CT standard for the equipment in question.
What are the key requirements of kSIST FprEN IEC 60601-2-44:2026?
kSIST FprEN IEC 60601-2-44:2026 builds on IEC 60601-1 and then adds CT-specific safety, dose-management, and documentation requirements. The most practical parts are the emergency stop functions, radiation control, dose reporting, and positioning accuracy.
Safety functions and system control
Clause 203.101 requires emergency interruption means to stop LOADING, with accessible controls near the PATIENT SUPPORT or GANTRY and at or near the CONTROL PANEL. Clause 203.107 adds automatic termination if timer failure or data-collection failure occurs, and it requires visible indication and manual reset before another scan can start. For a manufacturer, this means the CT SCANNER must be designed so a scan can be stopped quickly and predictably, not just switched off in software.
Clause 203.104 requires connections for external INTERLOCKS that can stop X-RADIATION or prevent it from starting. Clause 203.105 requires mobile ME EQUIPMENT with an incorporated battery to prevent unauthorized powered movement and X-RAY generation while still allowing battery charging. Clause 203.106 requires means to limit delivered energy using fixed or pre-selected LOADING FACTORS and MODES OF OPERATION, and the operator must be able to terminate LOADING during a scan. In practice, these provisions are about safe access control and safe interruption of exposure.
Protocols, AEC, and dose management
Clause 203.106.2 requires REFERENCE CLINICAL PROTOCOLS for the intended uses of the CT SCANNER, and those protocols must not be overwritable. The user information must list them, and each protocol must describe the CT CONDITIONS OF OPERATION and related dose estimates. Annex EE gives a template for that list, which matters when a hospital wants to standardize protocols and document them clearly.
Clause 203.106.3 requires Automatic Exposure Control (AEC) as an alternative to manual selection of CT CONDITIONS OF OPERATION. For whole-body CT SCANNERS, XYZ exposure control shall be available. The INSTRUCTIONS FOR USE must explain how AEC works, when to use it, when not to use it, and how switching between AEC and manual exposure is controlled. This is important because dose and image quality depend on how the system adapts to patient size and anatomy.
Clause 203.108 introduces DOSE NOTIFICATION and DOSE ALERT functions. The CT SCANNER must allow users to set thresholds, warn before a scan if an estimated dose index exceeds the threshold, require operator confirmation, and record the event in DICOM CT RADIATION dose structured report (SR) templates from ISO 12052. For CT GUIDED INTERVENTIONAL PROCEDURES, ALTERNATE DOSE ALERT BEHAVIOUR must be provided so repeated alerts do not unnecessarily interrupt the procedure.
Clause 203.108.3 adds checking before saving protocols and audit capability. That means protocol changes are not just stored - they are checked against dose thresholds and can be reviewed later. For a quality manager, this is the part that supports protocol governance and dose auditing.
Dose display, reporting, and imaging geometry
Clause 203.112 requires CTDI and DLP to be displayed on the CONTROL PANEL before scanning, with the CTDI phantom diameter shown as well. If CT CONDITIONS OF OPERATION vary during a scan, the expected values must be displayed as time-weighted averages. Clause 203.113 requires SSDE and Dw to be displayed and recorded in accordance with IEC 62985. Clause 203.114 then requires dose summary information to be saved as both a DICOM structured report and a DICOM image. In practice, the document ties console display, local records, and structured reporting together.
Clause 203.115 requires the GEOMETRIC EFFICIENCY IN THE Z DIRECTION to be displayed, with a warning, when it is below 70 %. Clauses 203.116 and 203.117 cover light markers and reconstructed section indication, including positioning accuracy. Clause 203.119 covers PATIENT SUPPORT positioning accuracy using IEC 61223-3-5:2019, with a person-equivalent load of not less than 135 kg. These clauses matter because patient setup and image localization affect both image quality and radiation use.
What terms does kSIST FprEN IEC 60601-2-44:2026 define?
- CT SCANNER - X-RAY EQUIPMENT for computer reconstruction of cross-sectional images of the head and/or body, including support parts and ACCESSORIES.
- PATIENT SUPPORT - The tabletop or head holder that supports the PATIENT so the irradiated body part can be positioned or moved.
- TOMOGRAPHIC PLANE - The geometric plane perpendicular to the z-axis at the ISOCENTRE of the X-RAY FIELD.
- DOSE PROFILE - The dose represented as a function of position along a line.
- GEOMETRIC EFFICIENCY IN THE Z-DIRECTION - The share of the total z-axis dose profile that lies within the acquisition range, expressed as a percentage.
- CT PITCH FACTOR - The ratio of PATIENT SUPPORT travel per X-ray source rotation to the NOMINAL TOTAL ACQUISITION BEAM WIDTH.
- REFERENCE CLINICAL PROTOCOL - A manufacturer-supplied protocol set for a CT application; the standard requires these to be listed and protected from overwriting.
- DOSE ALERT VALUE - A user-set dose threshold that triggers a pre-scan alert and operator confirmation.
Who uses kSIST FprEN IEC 60601-2-44:2026?
kSIST FprEN IEC 60601-2-44:2026 is used by CT scanner manufacturers, X-ray generator suppliers, test laboratories, hospitals, and regulatory or certification teams. Manufacturers use it for design, risk management, labeling, software functions, and dose reporting. Test labs use it for compliance checks, acceptance-style measurements, and functional testing.
Clinical imaging departments use it when setting up REFERENCE CLINICAL PROTOCOLS, reviewing dose display behavior, and documenting CTDI, DLP, and SSDE information. Quality managers and buyers use it to compare CT systems on safety functions, dose-management features, and the content of accompanying documents.
What changed in kSIST FprEN IEC 60601-2-44:2026 from the previous edition?
kSIST FprEN IEC 60601-2-44:2026 is the fourth edition and it replaces the third edition published in 2009 together with Amendment 1:2012 and Amendment 2:2016. The foreword states that it is a technical revision.
The main changes named in the document are:
- new requirements for calculation, display and recording of SSDE and of CTDIvol for scan projection radiogram (SPR)
- revised requirements for CT scanners used for radiotherapy planning
- revised dose-check requirements, especially for interventional procedures
- revised dose-statement requirements in the accompanying documents
- a new requirement for a DICOM image-based dose survey
- new terms and revised terms for CTDI and DLP, including SPR-related definitions
- new requirements for REFERENCE CLINICAL PROTOCOLS, AEC with XYZ modulation, prevention of unlimited scanning, MAR, and dose summary reporting
- revised requirements for emergency stopping, dose indicators, RTP alignment, loading state indication, and tomographic plane or section indication
- an informative annex for the content of REFERENCE CLINICAL PROTOCOLS
Which standards are used with kSIST FprEN IEC 60601-2-44:2026?
Clause 201.2 names the following normative references:
- IEC 60601-1 - the base standard for basic safety and essential performance of medical electrical equipment
- IEC 61223-3-5:2019 - acceptance tests and imaging performance of CT X-ray equipment
- ISO 12052:2017 - DICOM workflow, data management, and structured reporting
- IEC 60336:2020 - focal spot dimensions and related characteristics of medical X-ray tube assemblies
- IEC 62985:2019 - methods for calculating SSDE for CT
- IEC 61674:2024 - dosimeters used in diagnostic X-ray imaging
Clauses 202 and 203 also apply IEC 60601-1-2:2014+AMD1:2020 and IEC 60601-1-3:2008+AMD1:2013+AMD2:2021 as modified. The document states that IEC 60601-1-8, IEC 60601-1-9, IEC 60601-1-10, IEC 60601-1-11, and IEC 60601-1-12 do not apply.
What does the kSIST FprEN IEC 60601-2-44:2026 document contain?
kSIST FprEN IEC 60601-2-44:2026 contains a detailed structure of CT-specific clauses, figures, a test pattern table, and informative annexes. It includes figures for the coordinate system, tomographic plane alignment, reconstructed section indication, extra-focal radiation, and stray-radiation measurement.
It also contains practical test and documentation content, such as a CTDI test pattern table, a procedure for measuring CTDI, a graphical dose-profile statement, and a template for REFERENCE CLINICAL PROTOCOLS. Annex CC explains why CTDI and CTDI100 differ and how beam width and phantom length affect the values, which is useful when interpreting dose figures across different CT systems.
Relations
- Effective Date
- 18-Feb-2025
- Effective Date
- 18-Feb-2025
- Effective Date
- 18-Feb-2025
- Effective Date
- 18-Feb-2025
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
kSIST FprEN IEC 60601-2-44:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography". This standard covers: Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography
Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography
kSIST FprEN IEC 60601-2-44:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.50 - Radiographic equipment. The ICS classification helps identify the subject area and facilitates finding related standards.
kSIST FprEN IEC 60601-2-44:2026 has the following relationships with other standards: It is inter standard links to SIST EN 60601-2-44:2009/A2:2016, SIST EN 60601-2-44:2009/A1:2014, SIST EN 60601-2-44:2009/A11:2012, SIST EN 60601-2-44:2009. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
kSIST FprEN IEC 60601-2-44:2026 is associated with the following European legislation: EU Directives/Regulations: 2017/745, 93/42/EEC; Standardization Mandates: M/432, M/575. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
kSIST FprEN IEC 60601-2-44:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 48 outputs
- 49 201.12.1 Accuracy of controls and instruments
- 50 201.12.1.101 Accuracy of RADIATION output
- 51 201.12.1.102 Accuracy of recorded CT EXAMINATION data
- 52 201.13 Hazardous situations and fault conditions
- 53 201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS)
- 54 201.15 Construction of ME EQUIPMENT
- 55 201.16 ME SYSTEMS
- 56 201.17 ELECTROMAGNETIC COMPATIBILITY of ME EQUIPMENT and ME SYSTEMS
- 57 201.101 Requirements for CT SCANNERS providing images for RADIOTHERAPY 58 TREATMENT PLANNING (RTP)
- 59 202Electromagnetic compatibility – Requirements and tests
- 60 203 General requirements for RADIATION protection in diagnostic X-ray equipment
- 61 203.4 General requirements
- 62 203.4.1 Statement of compliance
- 63 203.5 ME EQUIPMENT identification, marking and documents
- 64 203.5.2 ACCOMPANYING DOCUMENTS
- 65 203.5.2.2 Dosimetric calibration
- 66 203.5.2.3 General requirements for the reference of subassemblies 67 and ACCESSORIES
- 68 203.5.2.4 INSTRUCTIONS FOR USE
- 69 203.6 RADIATION management
- 70 203.6.2 Initiation and termination of the IRRADIATION
- 71 203.6.2.1 Normal initiation and termination of the IRRADIATION
- 72 203.6.2.2 Safety measures against failure of normal termination of the 73 IRRADIATION
- 74 203.6.3 RADIATION dose and RADIATION QUALITY
- 75 203.6.3.1 Adjustment OF RADIATION dose and RADIATION QUALITY
- 76 203.6.3.2 Reproducibility of the RADIATION output
- 77 203.6.4 Indication of operational states
- 78 203.6.4.1 Indication of the X-RAY SOURCE ASSEMBLY selected
- 79 203.6.4.2 Indication of LOADING STATE
- 80 203.6.4.3 Indication of LOADING FACTORS and MODES OF OPERATION
- 81 203.6.5 AUTOMATIC CONTROL SYSTEM
- 82 203.6.6 Scattered RADIATION reduction
- 83 203.6.7 Imaging performance
- 84 203.6.7.2 System performance
- 85 203.6.7.2.1 Quality Assurance
- 86 203.6.7.2.2 Performance information
- 87 203.6.7.3 NOMINAL FOCAL SPOT value
- 88 203.6.7.4 RADIATION DETECTOR or X-RAY IMAGE RECEPTOR
- 89 203.6.7.101 Metal Artifact Reduction
- 90 203.7 RADIATION QUALITY
- 91 203.7.1 HALF-VALUE LAYERS and TOTAL FILTRATION in X-RAY EQUIPMENT
- 92 203.7.2 Waveform of the X-RAY TUBE VOLTAGE
- 93 203.7.3 Indication of FILTER properties
- 94 203.7.4 Test for FILTRATION by irremovable materials
- 95 203.7.5 Test for ADDED FILTERS and materials
96 203.7.6 Test for HALF-VALUE LAYER ......................................................................... 46 97 203.8 Limitation of the extent of the X-RAY BEAM and relationship between X-RAY 98 FIELD and IMAGE RECEPTION AREA............................................................................ 46 99 203.8.1 General ..................................................................................................... 46 100 203.8.4 Confinement of extra-focal RADIATION ......................................................... 46 101 203.8.5 Relationship between X-RAY FIELD and IMAGE RECEPTION AREA .................... 47 102 203.9 FOCAL SPOT TO SKIN DISTANCE ................................................................................. 47 103 203.10 ATTENUATION of the X-RAY BEAM between the PATIENT and the X-RAY 104 IMAGE RECEPTOR .................................................................................................... 48 105 203.11 Protection against RESIDUAL RADIATION ....................................................... 48 106 203.12 Protection against LEAKAGE RADIATION ........................................................ 48 107 203.12.1 General ..................................................................................................... 48 108 203.12.3 Statement of reference LOADING conditions ................................................ 48 109 203.13 Protection against STRAY RADIATION ........................................................... 48 110 203.13.1General ............................................................................................................. 48 111 203.13.101 Statements in the ACCOMPANYING DOCUMENTS .............................. 49 112 203.101 Emergency TERMINATION of X-RADIATION ..................................................... 49 113 203.102 Visual indication ........................................................................................ 50 114 203.103 Indication of operational READY STATE ........................................................ 50 115 203.104 Connection of external INTERLOCKS ............................................................ 50 116 203.105 Charging mode INTERLOCK ......................................................................... 51 117 203.106 Control of RADIATION output ........................................................................ 51 118 203.106.1 General control functions ........................................................................... 51 119 203.106.2 REFERENCE CLINICAL PROTOCOLS ................................................................. 51 120 203.106.3 AUTOMATIC EXPOSURE CONTROL (AEC) ........................................................ 52 121 203.106.3.1 General ...................................................................................... 52 122 203.106.3.3 Information to be provided ....................................................... 52 123 203.107 Safety measures against excessive X-RADIATION........................................ 53 124 203.108 Dose Checks (DOSE NOTIFICATION, DOSE ALERT) .......................................... 54 125 203.108.1 DOSE NOTIFICATION ..................................................................................... 54 126 203.108.2 DOSE ALERT ................................................................................................ 55 127 203.108.2.1 General ....................................................................................... 55 128 203.108.2.2 DOSE ALERT for CT GUIDED INTERVENTIONAL PROCEDURES ............... 56 129 203.108.3 Checking Prior to Saving Protocols and Audit Capability ............................ 57 130 203.109 Information regarding perfusion scanning .................................................. 57 131 203.110 Dose statements ........................................................................................ 57 132 203.110.1 CTDI .................................................................................................... 57 100 133 203.110.2 CTDI ............................................................................................... 59 free air 134 203.111 DOSE PROFILE statement ............................................................................ 60 135 203.112 Display and recording of CTDI and DLP ................................................ 60 vol 136 203.115 GEOMETRIC EFFICIENCY IN THE Z DIRECTION .................................................. 62 137 203.116 Indication and position of the TOMOGRAPHIC PLANE AND TOMOGRAPHIC 138 SECTION(S) ............................................................................................................. 62 139 203.117 Indication and position of the reconstructed section(s) ............................... 63 140 203.119 Indication and position of the PATIENT SUPPORT .......................................... 63 141 Annex AA (informative) Choosing LOADING FACTORS for tests ............................................ 64 142 Annex BB (informative) ....................................................................................................... 64 143 Annex CC (informative) The CTDI concept in IEC 60601-2-44: Relationship 100 144 between CTDI and CTDI ......................................................................................... 65 100 ∞ 145 CC.1 Evolution of CTDI in editions of IEC 60601-2-44 .............................................. 66 100 146 CC.2 Introduction and justification of a new definition for CTDI ................................. 67 100 147 CC.3 Underestimations of CTDI ................................................................................... 68 148 CC.4 Use of CTDI to estimate dose for object sizes other than those of the vol 149 CTDI phantoms ..................................................................................................... 69 150 Annex DD (informative) Measuring CTDI ................................................................ 69 free air 151 ANNEX EE (informative) Reference Template for the Content for Listing of REFERENCE 152 CLINICAL PROTOCOLS ........................................................................................................ 72 153 154 Figure 201.101 – Coordinate system ..................................................................................... 16 155 Figure 201.102 – Illustration of {N T}, R and {N T} + R ..................................................... 18 156 Figure 201.101.2.1 – Vertical alignment of TOMOGRAPHIC PLANE ............................................ 36 157 Figure 201.101.2.2 – Alignment of the PATIENT SUPPORT and its movement direction ............. 37 158 Figure 201.101.2.3 – Rotational alignment of the reconstructed CT images ........................... 38 159 Figure 203.8.4.1 – Zone of extra-focal RADIATION .................................................................. 47 160 Figure 203.102 – Minimum dimensions for STRAY RADIATION measurement ............................ 49 161 Figure CC.1 – CTDI versus beam width along z .................................................................. 67 w 162 163 Table 203.101 − Test pattern for CTDI ....................................................................... 60 free air 164 Table CC.1 – Ratios of CTDI by phantom length ................................................................... 69 167 INTERNATIONAL ELECTROTECHNICAL COMMISSION 168 ____________ 169 170 MEDICAL ELECTRICAL EQUIPMENT – 171 172 Part 2-44: Particular requirements for the basic safety and essential 173 performance of X-ray equipment for computed tomography 174 175 176 FOREWORD 177 1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising 178 all national electrotechnical committees (IEC National Committees). The object of IEC is to promote international 179 co-operation on all questions concerning standardization in the electrical and electronic fields. To this end and 180 in addition to other activities, IEC publishes International Standards, Technical Specifications, Technical Reports, 181 Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their 182 preparation is entrusted to technical committees; any IEC National Committee interested in the subject dealt with 183 may participate in this preparatory work. International, governmental and non-governmental organizations liaising 184 with the IEC also participate in this preparation. IEC collaborates closely with the International Organization for 185 Standardization (ISO) in accordance with conditions determined by agreement between the two organizations. 186 2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international 187 consensus of opinion on the relevant subjects since each technical committee has representation from all 188 interested IEC National Committees.
189 3) IEC Publications have the form of recommendations for international use and are accepted by IEC National 190 Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC 191 Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any 192 misinterpretation by any end user.
193 4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications 194 transparently to the maximum extent possible in their national and regional publications. Any divergence between 195 any IEC Publication and the corresponding national or regional publication shall be clearly indicated in the latter. 196 5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity 197 assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any 198 services carried out by independent certification bodies.
199 6) All users should ensure that they have the latest edition of this publication.
200 7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and 201 members of its technical committees and IEC National Committees for any personal injury, property damage or 202 other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and 203 expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC Publications. 204 8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is 205 indispensable for the correct application of this publication.
206 9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of patent 207 rights. IEC shall not be held responsible for identifying any or all such patent rights.
208 IEC 60601-2-44 has been prepared by subcommittee 62B: Medical imaging equipment, 209 software, and systems, of IEC technical committee 62: Medical equipment, software, and 210 systems.
211 This fourth edition cancels and replaces the third edition published in 2009, Amendment 1:2012 212 and Amendment 2:2016. This edition constitutes a technical revision.
213 This particular standard has been prepared to provide, based on IEC 60601- 214 1:2005+A1:2012+A2:2020 (3.2 edition) and its collaterals, a complete set of safety 215 requirements for ME EQUIPMENT for computed tomography.
216 This edition 4 of IEC 60601-2-44 includes several changes to account for the evolved technical 217 state-of-the-art of ct-scanner and to harmonize this standard with other international and 218 national standards and regulations. The main changes are:
219 − a new requirements for calculation, display and recording of SSDE (according to IEC 220 62985), and of CTDIvol for scan projection radiogram (SPR) 221 − a revision of the requirements for ct scanner used for radiotherapy planning, 222 − a revision of requirements for dose checks (dose notification, dose alert), in particular 223 for interventional procedures 224 − a revision of requirements for dose statements in the accompanying documents 225 − a new requirement for a DICOM image based dose survey 226 In addition, several other changes and new requirements have been included 227 In detail, this particular standard compared to the IEC 60601-2-44:2009+A1:2012+A2:2016 (3.2 228 edition) includes the following changes:
229 a) New definition of terms for 230 − RECONSTRUCTED SECTION THICKNESS, 231 − NOMINAL RECONSTRUCTED SECTION THICKNESS, 232 − NOMINAL TOTAL ACQUISITION BEAM WIDTH, 233 − READY STATE, 234 − INSTRUCTIONS FOR USE, 235 − REFERENCE CLINICAL PROTOCOL, 236 − CTDI DOSIMETRY PHANTOM, 237 − ALTERNATE DOSE ALERT BEHAVIOR, 238 − CT GUIDED INTERVENTIONAL PROCEDURE, 239 − CTDI (Z), VOL 240 − SSDE(z), 241 − PATIENT SUPPORT 242 b) Terms revised for CTDI and DLP, new definitions for CTDI and DLP for scan mode VOL VOL 243 (SPR) 244 c) New notes regarding ESSENTIAL PERFORMANCE 245 d) Requirements deleted for SENSITIVITY PROFILE STATEMENT 246 e) New requirements for 247 − specification in ACCOMPANYING DOCUMENTS 248 − Warning against unexpected movement of the PATIENT SUPPORT 249 − CTDIVOL and DLP for SPR 250 − Display/Recording of SSDE 251 − Dose Summary as DICOM image and DICOM Dose SR 252 − REFERENCE CLINICAL PROTOCOLS 253 − AEC with XYZ-modulation 254 − prevention of unlimited scanning 255 − Metal Artifact Reduction (MAR) in Imaging performance 256 f) Requirements revised for 257 − Linear movements of the PATIENT SUPPORT and gantry in Emergency 258 stopping devices 259 − Deterministic Effects 260 − Dose Checks 261 − System performance 262 − RTP (alignments, typ. modes, etc.) 263 − user information (CTDI100, etc.) 264 − Dose statements 265 − indicator lights 266 − Safety measures against excessive X-radiation 267 − Indication and position of the tomographic Plane and tomographic 268 section(s) 269 − Indication of loading state 270 − axial scan modes for CTDI measurements 271 − Lightmarker (203.117): Indication of an (offset) reference plane instead of 272 indication of the tomographic plane possible 273 g) Several changes to account for state-of-the-art and for clarification 274 h) Add informative Annex for Reference Template for the Content for Listing of 275 REFERENCE CLINICAL PROTOCOLS 276 The text of this International Standard is based on the following documents:
Draft Report on voting 62B/XXX/FDIS 62B/XXX/RVD 277 278 Full information on the voting for its approval can be found in the report on voting indicated in 279 the above table.
280 The language used for the development of this International Standard is English.
281 This document was drafted in accordance with ISO/IEC Directives, Part 2, and developed in 282 accordance with ISO/IEC Directives, Part 1 and ISO/IEC Directives, IEC Supplement, available 283 at http://www.iec.ch/members_experts/refdocs. The main document types developed by IEC 284 are described in greater detail at www.iec.ch/standardsdev/publications.
285 In this document, the following print types are used:
286 • requirements and definitions: roman type; 287 • test specifications: italic type; 288 • informative material appearing outside of tables, such as notes, examples and references: in smaller type. 289 Normative text of tables is also in a smaller type; 290 • TERMS DEFINED IN CLAUSE 3 OF IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 AND 291 IEC 60601-1:2005/AMD2:2020, IN THIS DOCUMENT OR AS NOTED: SMALL CAPITALS.
292 In referring to the structure of this document, the term 293 • "clause" means one of the eighteen numbered divisions within the table of contents, 294 inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.); 295 • "subclause" means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all 296 subclauses of Clause 7).
297 References to clauses within this document are preceded by the term "Clause" followed by the 298 clause number. References to subclauses within this document are by number only.
299 In this document, the conjunctive "or" is used as an "inclusive or" so a statement is true if any 300 combination of the conditions is true.
301 The verbal forms used in this document conform to usage described in Clause 7 of the ISO/IEC 302 Directives, Part 2. For the purposes of this document, the auxiliary verb:
303 • "shall" means that compliance with a requirement or a test is mandatory for compliance with 304 this document; 305 • "should" means that compliance with a requirement or a test is recommended but is not 306 mandatory for compliance with this document; 307 • "may" is used to describe a permissible way to achieve compliance with a requirement or 308 test.
309 An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title 310 indicates that there is guidance or rationale related to that item in Annex AA.
311 A list of all parts of the IEC 60601 and IEC 80601 series, published under the general title:
312 Medical electrical equipment, can be found on the IEC website.
313 The committee has decided that the contents of this document will remain unchanged until the 314 stability date indicated on the IEC website under webstore.iec.ch/?ref=menu in the data related 315 to the specific document. At this date, the document will be 316 • reconfirmed, 317 • withdrawn, 318 • replaced by a revised edition, or 319 • amended.
320
IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates that it contains colours which are considered to be useful for the correct understanding of its contents. Users should therefore print this document using a colour printer.
323 MEDICAL ELECTRICAL EQUIPMENT – 324 325 Part 2-44: Particular requirements for the basic safety and essential 326 performance of X-ray equipment for computed tomography 327 328 329 330 201.1 Scope, object and related standards 1) 331 Clause 1 of the general standard applies, except as follows:
332 201.1.1 Scope 333 Replacement:
334 This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of CT 335 SCANNERS, hereafter also referred to as ME EQUIPMENT.
336 If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to 337 ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the 338 case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant. 339 NOTE 1 See also 4.2 of the general standard.
340 The scope of this document is limited to CT SCANNERS intended to be used for head and/or body 341 imaging characterised by an ENCLOSURE of the X-ray source(s) and imaging detector(s) in a 342 common protective cover in the shape of a toroid. It includes safety requirements for the X-RAY 343 GENERATORS used in CT SCANNERS, including those where HIGH-VOLTAGE GENERATORS are 344 integrated with an X-RAY TUBE ASSEMBLY.
345 NOTE 2 This standard does not intend to address interventional systems as described in IEC 60601-2-43 and cone- 346 beam CT scanners integrated into a radiotherapy systems as described in IEC 60601-2-68 and dental cone-beam 347 CT scanners as described in IEC 60601-2-63.
348 NOTE 3 The body includes cardiac, extremities, neck, etc.
349 The scope of this International Standard excludes RADIOTHERAPY SIMULATORS and systems 350 where the image is created by a source other than an X-RAY TUBE.
351 201.1.2 Object 352 Replacement:
353 The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL 354 PERFORMANCE requirements for CT SCANNERS as defined in 201.3.201, to ensure safety, and to 355 specify methods for demonstrating compliance with those requirements, for CT SCANNERS.
...