Medical electrical equipment - Part 2-64: Particular requirements for the basic safety and essential performance of light ion beam medical equipment

This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of LIGHT
ION BEAM ME EQUIPMENT, hereafter referred to as ME EQUIPMENT, used for treatment of PATIENTS.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
This particular standard, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively
to the manufacture and some installation aspects of LIGHT ION BEAM ME EQUIPMENT
– intended for RADIOTHERAPY in human medical practice, including those in which the
selection and DISPLAY of operating parameters can be controlled automatically by
PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS),
– that, in NORMAL USE, deliver a RADIATION BEAM of LIGHT IONS having ENERGY PER NUCLEON in
the range 10 MeV/n to 500 MeV/n,
and
– intended to be
– for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED
PERSONS by OPERATORS having the required skills for a particular medical application,
for particular SPECIFIED clinical purposes maintained in accordance with the
recommendations given in the INSTRUCTIONS FOR USE,
– subject to regular quality assurance performance and calibration checks by a QUALIFIED
PERSON.
NOTE 1 In this particular standard, all references to installation refer to installation in the RESPONSIBLE
ORGANIZATION’S premises.
NOTE 2 In this particular standard, all references to ABSORBED DOSE refer to ABSORBED DOSE in water.
NOTE 3 Information regarding x-ray image guidance can be found in IEC 60601-2-68 (under development).
NOTE 4 IEC 61217 gives guidance on the designation of ME EQUIPMENT movements, the marking of scales, their
zero positions and the direction of movement with increasing value (see 201.7.4.101).

Medizinische elektrische Geräte - Teil 2-64: Besondere Festlegungen für die Sicherheit einschließlich der wesentlichen Leistungsmerkmale von Leichtionen-Bestrahlungseinrichtungen

Appareils électromédicaux - Partie 2-64: Exigences particulières pour la sécurité de base et les performances essentielles des appareils de thérapie par faisceau ionique lumineux

L'IEC 60601-2-64:2014 s'applique à la sécurité de base et aux performances essentielles des appareils électromédicaux par Faisceau d'Ions Légers, désignés ci-après par le terme appareils EM, utilisés pour le traitement des patients. La présente norme particulière, y compris les essais de type et les essais sur le site, s'applique respectivement au fabricant et à certains aspects d'installation des appareils EM par Faisceau d'Ions Légers destinés à être utilisés à des fins de radiothérapie dans la médecine humaine, y compris les appareils dont le choix et l'affichage des paramètres de fonctionnement peuvent être contrôlés automatiquement par des sous-systèmes électroniques programmables qui, en utilisation normale, délivrent un faisceau de rayonnement d'ions légers dont l'énergie par nucléon se situe dans la gamme 10 MeV/n à 500 MeV/n.
Mots clé: médecine nucléaire, dosimétrie de radiations.

Medicinska električna oprema - 2-64. del: Posebne zahteve za osnovno varnost in bistvene lastnosti medicinske opreme za lahkoionsko terapijo

Ta mednarodni standard se uporablja za OSNOVNO VARNOST in BISTVENE LASTNOSTI ELEKTROMEDICINSKE OPREME ZA LAHKOIONSKO TERAPIJO (v nadaljevanju: ELEKTROMEDICINSKA OPREMA), ki se uporablja za zdravljenje PACIENTOV. Če je točka ali podtočka namenjena samo obravnavi ELEKTROMEDICINSKE OPREME ali ELEKTROMEDICINSKIH SISTEMOV, bo to zapisano v naslovu in vsebini točke ali podtočke. Če ni zapisano, točka ali podtočka obravnava ELEKTROMEDICINSKO OPREMO in ELEKTROMEDICINSKE SISTEME. Ta standard z vključitvijo TIPSKIH PRESKUSOV in PRESKUSOV NA LOKACIJI se uporablja za proizvodnjo in določene vidike namestitve OPREME ZA LAHKOIONSKO TERAPIJO – namenjeno RADIOTERAPIJI v človeški medicini, vključno s tisto, kjer lahko PROGRAMLJIVI ELEKTRONSKI PODSISTEMI (PESS) samodejno nadzirajo izbiro in PRIKAZ obratovalnih parametrov,
– ki pri OBIČAJNI UPORABI oddaja RADIACIJSKE ŽARKE LAHKIH IONOV z ENERGIJO NA NUKLEON v razponu od 10 MeV/n do 500 MeV/n in
– je namenjena
– OBIČAJNI UPORABI, pri čemer jih pod vodstvom ustrezno licenciranih ali KVALIFICIRANIH OSEB upravljajo OPERATERJI z zahtevanimi veščinami za določeno medicinsko uporabo, za določene OPREDELJENE klinične namene v skladu s priporočili, podanimi v NAVODILIH ZA UPORABO,
– je predmet rednih pregledov zagotavljanja kakovosti, delovanja in umerjanja, ki jih izvaja KVALIFICIRANA OSEBA.
OPOMBA 1: V tem standardu se vsa sklicevanja na obrat nanašajo na obrat v prostorih ODGOVORNE ORGANIZACIJE.
OPOMBA 2: V tem standardu se ABSORBIRANA DOZA nanaša na ABSORBIRANO DOZO v vodi.
OPOMBA 3: Informacije glede smernic za rentgensko slikanje lahko najdete v standardu IEC 60601-2-68 (v pripravi).
OPOMBA 4: Standard IEC 61217 podaja smernice za označevanje premikov ELEKTROMEDICINSKE OPREME, označevanje tehtnic, njihovih osnovnih pozicij in smeri premikov z višanjem vrednosti (glejte točko 201.7.4.101).

General Information

Status
Published
Publication Date
16-Aug-2015
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
15-Jun-2015
Due Date
20-Aug-2015
Completion Date
17-Aug-2015

Buy Standard

Standard
EN 60601-2-64:2015 - BARVE
English language
58 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 60601-2-64:2015
01-september-2015
0HGLFLQVNDHOHNWULþQDRSUHPDGHO3RVHEQH]DKWHYH]DRVQRYQRYDUQRVWLQ
ELVWYHQHODVWQRVWLPHGLFLQVNHRSUHPH]DODKNRLRQVNRWHUDSLMR
Medical electrical equipment - Part 2-64: Particular requirements for the basic safety and
essential performance of light ion beam medical equipment
Medizinische elektrische Geräte - Teil 2-64: Besondere Festlegungen für die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale von Leichtionen-
Bestrahlungseinrichtungen
Appareils électromédicaux - Partie 2-64: Exigences particulières pour la sécurité de base
et les performances essentielles des appareils de thérapie par faisceau ionique lumineux
Ta slovenski standard je istoveten z: EN 60601-2-64:2015
ICS:
11.040.60 Terapevtska oprema Therapy equipment
SIST EN 60601-2-64:2015 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 60601-2-64:2015

---------------------- Page: 2 ----------------------

SIST EN 60601-2-64:2015


EUROPEAN STANDARD EN 60601-2-64

NORME EUROPÉENNE

EUROPÄISCHE NORM
May 2015
ICS 11.040.60

English Version
Medical electrical equipment - Part 2-64: Particular requirements
for the basic safety and essential performance of light ion beam
medical electrical equipment
(IEC 60601-2-64:2014)
Appareils électromédicaux - Partie 2-64: Exigences Medizinische elektrische Geräte - Teil 2-64: Besondere
particulières pour la sécurité de base et les performances Festlegungen für die Sicherheit einschließlich der
essentielles des appareils électromédicaux par faisceau wesentlichen Leistungsmerkmale von Leichtionen-
d'ions légers Bestrahlungseinrichtungen
(IEC 60601-2-64:2014) (IEC 60601-2-64:2014)
This European Standard was approved by CENELEC on 2014-10-08. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.


European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2015 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
 Ref. No. EN 60601-2-64:2015 E

---------------------- Page: 3 ----------------------

SIST EN 60601-2-64:2015
EN 60601-2-64:2015
Foreword
The text of document 62C/594/FDIS, future edition 1 of IEC 60601-2-64 prepared by
SC 62C "Equipment for radiotherapy, nuclear medicine and radiation dosimetry" of IEC/TC 62
"Electrical equipment in medical practice" was submitted to the IEC-CENELEC parallel vote and
approved by CENELEC as EN 60601-2-64:2015.

The following dates are fixed:
• latest date by which the document has to be (dop) 2015-11-29
implemented at national level by
publication of an identical national
standard or by endorsement
(dow) 2018-05-29
• latest date by which the national
standards conflicting with the
document have to be withdrawn

Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such
patent rights.

This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association, and supports essential requirements of EU Directive.

For the relationship with EU Directive 93/42/EEC, see informative Annex ZZ, which is an integral part
of this document.

Endorsement notice
The text of the International Standard IEC 60601-2-64:2014 was approved by CENELEC as a
European Standard without any modification.
In the official version, for Bibliography, the following note has to be added for the standard indicated:

IEC 60601-2-68 NOTE Harmonized as EN 60601-2-68.
2

---------------------- Page: 4 ----------------------

SIST EN 60601-2-64:2015
EN 60601-2-64:2015
Annex ZA
(normative)

Normative references to international publications
with their corresponding European publications

The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.

NOTE 1 When an International Publication has been modified by common modifications, indicated by (mod), the relevant
EN/HD applies.

NOTE 2 Up-to-date information on the latest versions of the European Standards listed in this annex is available here:
www.cenelec.eu

Publication Year Title EN/HD Year
Annex ZA of EN 60601-1:2006 applies except as follows:
Replacement:
IEC 60601-1-2 2014 Medical electrical equipment - EN 60601-1-2 2014
Part 1-2: General requirements for basic
safety and essential performance -
Collateral standard: Electromagnetic
disturbances - Requirements and tests
Addition:
IEC 60601-1 2005 Medical electrical equipment - EN 60601-1 2006
 Part 1: General requirements for basic + corr. March 2010
 safety and essential performance
+A1 2012 + A1 2013
+ A1/AC 2014
+A12 2014
IEC 60601-2-1 2009 Medical electrical equipment - - -
Part 2-1: Particular requirements for the
basic safety and essential performance of
electron accelerators in the range 1 MeV to
50 MeV
IEC 60601-2-11 2013 Medical electrical equipment - EN 60601-2-11 2015
Part 2-11: Particular requirements for the
basic safety and essential performance of
gamma beam therapy equipment
IEC/TR 60788 2004 Medical electrical equipment - Glossary of - -
defined terms
IEC 61217 2011 Radiotherapy equipment - Coordinates, EN 61217 2012
movements and scales
ISO/IEC 14165-321 2009 Information technology - Fibre Channel - - -
Part 321: Audio-Video (FC-AV)

3

---------------------- Page: 5 ----------------------

SIST EN 60601-2-64:2015
EN 60601-2-64:2015
Annex ZZ
(informative)

Coverage of Essential Requirements of EU Directives
This European Standard has been prepared under a mandate given to CENELEC by the European
Commission and the European Free Trade Association, and within its scope the Standard covers all
relevant essential requirements given in Annex I of EU Directive 93/42/EEC of 14 June 1993
concerning medical devices.
Compliance with this standard provides one means of conformity with the specified essential
requirements of the Directive concerned.
WARNING: Other requirements and other EU Directives can be applied to the products falling within
the scope of this standard.

4

---------------------- Page: 6 ----------------------

SIST EN 60601-2-64:2015




IEC 60601-2-64

®


Edition 1.0 2014-09




INTERNATIONAL



STANDARD




NORME



INTERNATIONALE
colour

inside










Medical electrical equipment –

Part 2-64: Particular requirements for the basic safety and essential

performance of light ion beam medical electrical equipment




Appareils électromédicaux –

Partie 2-64: Exigences particulières pour la sécurité de base et


les performances essentielles des appareils électromédicaux par faisceau

d'ions légers











INTERNATIONAL

ELECTROTECHNICAL

COMMISSION


COMMISSION

ELECTROTECHNIQUE

PRICE CODE
INTERNATIONALE

CODE PRIX XA


ICS 11.040.60 ISBN 978-2-8322-1825-9



Warning! Make sure that you obtained this publication from an authorized distributor.

Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.

® Registered trademark of the International Electrotechnical Commission
Marque déposée de la Commission Electrotechnique Internationale

---------------------- Page: 7 ----------------------

SIST EN 60601-2-64:2015
– 2 – IEC 60601-2-64:2014  IEC 2014
CONTENTS
FOREWORD . 3
INTRODUCTION . 5
201.1 Scope, object and related standards . 6
201.2 Normative references . 8
201.3 Terms and definitions . 9
201.4 General requirements . 14
201.5 General requirements for testing of ME EQUIPMENT . 14
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 15
201.7 ME EQUIPMENT identification, marking and documents . 15
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 18
201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS . 19
201.10 Protection against unwanted and excessive radiation HAZARDS . 24
201.11 Protection against excessive temperatures and other HAZARDS . 45
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 45
201.13 HAZARDOUS SITUATIONS and fault conditions for ME EQUIPMENT . 46
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 46
201.15 Construction of ME EQUIPMENT . 46
201.16 ME SYSTEMS . 46
201.17 ELECTROMAGNETIC COMPATIBILITY of ME EQUIPMENT and ME SYSTEMS . 46
201.101 ELECTRONIC IMAGING DEVICES (EID) . 47
206 Usability . 47
Annexes . 50
Annex B (informative) Sequence of testing . 50
Annex I (informative) ME SYSTEMS aspects . 50
Bibliography . 51
Index of defined terms used in this particular standard. 52

Figure 201.101 – PATIENT SUPPORT movements . 48
Figure 201.102 – Diagram illustrating example RADIATION HEAD components and
possible PATIENT position for NON-PRIMARY RADIATION REQUIREMENTS . 49
Figure 201.103 – Diagram illustrating distance along PATIENT plane to measure
NON-PRIMARY RADIATION ABSORBED DOSE . 49

Table 201.101 – Data required in the technical description to support Clause 201.10
SITE TEST compliance . 17

---------------------- Page: 8 ----------------------

SIST EN 60601-2-64:2015
IEC 60601-2-64:2014  IEC 2014 – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________

MEDICAL ELECTRICAL EQUIPMENT –

Particular requirements for the basic safety
and essential performance of LIGHT ION BEAM ME EQUIPMENT

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and
non-governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates
closely with the International Organization for Standardization (ISO) in accordance with conditions determined
by agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International Standard IEC 60601-2-64 has been prepared by subcommittee 62C: Equipment
for radiotherapy, nuclear medicine and radiation dosimetry, of IEC technical committee 62:
Electrical equipment in medical practice.
The text of this standard is based on the following documents:
FDIS Report on voting
62C/594/FDIS 62C/600/RVD

Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.

---------------------- Page: 9 ----------------------

SIST EN 60601-2-64:2015
– 4 – IEC 60601-2-64:2014  IEC 2014
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the numbered divisions within the table of contents, inclusive of all
subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
A list of all parts of the IEC 60601 series, published under the general title Medical electrical
equipment, can be found on the IEC website.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.

IMPORTANT – The 'colour inside' logo on the cover page of this publication indicates
that it contains colours which are considered to be useful for the correct
understanding of its contents. Users should therefore print this document using a
colour printer.

---------------------- Page: 10 ----------------------

SIST EN 60601-2-64:2015
IEC 60601-2-64:2014  IEC 2014 – 5 –
INTRODUCTION
The use of LIGHT ION BEAM ME EQUIPMENT for RADIOTHERAPY purposes may expose PATIENTS to
danger if the ME EQUIPMENT fails to deliver the required dose to the PATIENT, or if the ME
design does not satisfy standards of electrical and mechanical safety. The ME
EQUIPMENT
EQUIPMENT may also cause danger to persons in the vicinity if the ME EQUIPMENT itself fails to
contain the RADIATION adequately or if there are inadequacies in the design of the TREATMENT
ROOM.
This particular standard establishes requirements to be complied with by MANUFACTURERS in
the design and construction of LIGHT ION BEAM ME EQUIPMENT for use in RADIOTHERAPY; it does
not attempt to define their optimum performance requirements. Its purpose is to identify those
features of design that are regarded, at the present time, as essential for the safe operation of
such ME EQUIPMENT; it places limits on the degradation of ME EQUIPMENT performance beyond
which it can be presumed that a fault condition exists and where an INTERLOCK then operates
to prevent continued operation of the ME EQUIPMENT.
Clause 201.10 contains limits beyond which INTERLOCKS prevent, INTERRUPT or TERMINATE
IRRADIATION in order to insure that ESSENTIAL PERFORMANCE is maintained and to avoid an
unsafe condition. TYPE TESTS that are performed by the MANUFACTURER, or SITE TESTS, which
are not necessarily performed by the MANUFACTURER, are SPECIFIED for each requirement. It
should be understood that, before installation, a MANUFACTURER can provide a compliance
certificate relating only to TYPE TESTS. Data available from SITE TESTS should be incorporated
in the ACCOMPANYING DOCUMENTS, in the form of a SITE TEST report, by those who test the ME
EQUIPMENT at installation.
Closely related to this standard is IEC 62667 which is currently being developed. It specifies
test methods and reporting formats for performance tests of LIGHT ION BEAM ME EQUIPMENT for
use in RADIOTHERAPY, with the aim of providing uniform methods of doing so. The annex of
IEC 62667 provides forms for presenting performance values, measured per the methods
SPECIFIED.

---------------------- Page: 11 ----------------------

SIST EN 60601-2-64:2015
– 6 – IEC 60601-2-64:2014  IEC 2014
MEDICAL ELECTRICAL EQUIPMENT –

Particular requirements for the basic safety
and essential performance of LIGHT ION BEAM ME EQUIPMENT



201.1 Scope, object and related standards
1
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of LIGHT
ION BEAM ME EQUIPMENT, hereafter referred to as ME EQUIPMENT, used for treatment of PATIENTS.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
This particular standard, with the inclusion of TYPE TESTS and SITE TESTS, applies respectively
to the manufacture and some installation aspects of LIGHT ION BEAM ME EQUIPMENT
– intended for RADIOTHERAPY in human medical practice, including those in which the
selection and DISPLAY of operating parameters can be controlled automatically by
PROGRAMMABLE ELECTRONIC SUBSYSTEMS (PESS),
– that, in NORMAL USE, deliver a RADIATION BEAM of LIGHT IONS having ENERGY PER NUCLEON in
the range 10 MeV/n to 500 MeV/n,
and
– intended to be
– for NORMAL USE, operated under the authority of appropriately licensed or QUALIFIED
PERSONS by OPERATORS having the required skills for a particular medical application,
for particular SPECIFIED clinical purposes maintained in accordance with the
recommendations given in the INSTRUCTIONS FOR USE,
– subject to regular quality assurance performance and calibration checks by a QUALIFIED
PERSON.
NOTE 1 In this particular standard, all references to installation refer to installation in the RESPONSIBLE
ORGANIZATION’S premises.
NOTE 2 In this particular standard, all references to ABSORBED DOSE refer to ABSORBED DOSE in water.
NOTE 3 Information regarding x-ray image guidance can be found in IEC 60601-2-68 (under development).
NOTE 4 IEC 61217 gives guidance on the designation of ME EQUIPMENT movements, the marking of scales, their
zero positions and the direction of movement with increasing value (see 201.7.4.101).
201.1.2 Object
Replacement:
____________
1
The general standard is IEC 60601-1:2005 and IEC 60601-1:2005/AMD1:2012, Medical electrical equipment –
Part 1: General requirements for basic safety and essential performance

---------------------- Page: 12 ----------------------

SIST EN 60601-2-64:2015
IEC 60601-2-64:2014  IEC 2014 – 7 –
The object of this particular standard is to establish particular BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements for LIGHT ION BEAM ME EQUIPMENT in the range 10 MeV/n to
500 MeV/n and to SPECIFY tests to check compliance to those requirements.
NOTE The adoption of this standard helps to ensure that the ME EQUIPMENT
– maintains PATIENT safety during ME EQUIPMENT movements and failure of the SUPPLY MAINS;
– delivers the pre-selected RADIATION TYPE, ENERGY PER NUCLEON, LIGHT ION species, and ABSORBED DOSE;
– delivers pre-selected LIGHT ION BEAMS to the PATIENT, by utilizing LIGHT ION BEAM modifying devices, etc.,
without causing unnecessary risk to the PATIENT, the OPERATOR, other persons or the environment.
201.1.3 Collateral standards
Addition:
Collateral standards published after the date of publication of this standard shall only apply
subject to further amendment to this standard.
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
IEC 60601-1-6 applies as modified in Clause 206. IEC 60601-1-3, IEC 60601-1-8,
2 3
IEC 60601-1-9 and IEC 60601-1-10 do not apply. All other published collateral standards in
the IEC 60601-1 series apply as published.
NOTE Collateral standards published after the date of publication of this standard will only apply subject to further
amendment to this standard.
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard or collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and ESSENTIAL
PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the IEC 60601-1-2 collateral
standard, 203.4 in this particular standard addresses the content of Clause 4 of the
IEC 60601-1-3 collateral standard, etc.). The changes to the text of the general standard are
specified by the use of the following words:
“Replacement” means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
____________
2
IEC 60601-1-9, Medical electrical equipment – Part 1-9: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for environmentally conscious design
3
 IEC 60601-1-10, Medical electrical equipment – Part 1-10: General requirements for basic safety and essential
performance – Collateral Standard: Requirements for the development of physiologic closed-loop controllers

---------------------- Page: 13 ----------------------

SIST EN 60601-2-64:2015
– 8 – IEC 60601-2-64:2014  IEC 2014
“Addition” means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
“Amendment” means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term “this standard” is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
Clause 2 of the general standard applies, except as follows:
Replacement:
IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic
safety and essential performance – Collateral Standard: Electromagnetic compatibility –
Requirements and tests
Addition:
IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic
safety and essential performance
IEC 60601-1:2005/AMD1:2012
IEC 60601-2-1:2009, Medical electrical equipment – Part 2-1: Particular requirements for the
basic safety and essential performance of electron accelerators
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.