Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment

This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
DENTAL UNITS, DENTAL PATIENT CHAIRS, DENTAL HANDPIECES and DENTAL OPERATING LIGHTS,
hereafter referred to as DENTAL EQUIPMENT.
Excluded are amalgamators, sterilizers and dental X-ray equipment.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
within the scope of this standard are not covered by specific requirements in this standard
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.

Medizinische elektrische Geräte - Teil 2-60: Besondere Festlegungen an die Sicherheit einschließlich der wesentlichen Leistungsmerkmale von Dental-Geräten

Appareils électromédicaux - Partie 2-60: Exigences particulières pour la sécurité de base et les performances essentielles des équipements dentaires

La CEI 80601-2-60:2012 s'applique à la sécurité de base et aux performances essentielles des Unités Dentaires, des Fauteuils Dentaires pour patient, des Outils à Main Dentaires et des Scialytiques. En sont exclus les vibreurs d'amalgame, les stérilisateurs et les équipements dentaires à rayonnement X.

Medicinska električna oprema - 2-60. del: Posebne zahteve za osnovno varnost in bistvene lastnosti zobozdravstvene opreme

Ta mednarodni standard se uporablja za OSNOVNO VARNOST in BISTVENE LASTNOSTI DENTALNIH ENOT, STOLOV ZA DENTALNE PACIENTE, DENTALNIH ROČNIH PRIPOMOČKOV IN DENTALNE OPERACIJSKE RAZSVETLJAVE, v nadaljevanju: ZOBOZDRAVSTVENA OPREMA. Izključeni so mešalni aparati, sterilizatorji in zobozdravstvena rentgenska oprema. Če je točka ali podtočka namenjena samo obravnavi ELEKTROMEDICINSKE OPREME ali ELEKTROMEDICINSKIH SISTEMOV, bo to zapisano v naslovu in vsebini točke ali podtočke. Če ni zapisano, točka ali podtočka obravnava ELEKTROMEDICINSKO OPREMO in ELEKTROMEDICINSKE SISTEME.
Posebne zahteve tega standarda ne vključujejo temeljnih TVEGANJ pri nameravani fiziološki funkciji ELEKTROMEDICINSKE OPREME in SISTEMOV, ki spadajo na področje tega standarda, razen v poglavjih 7.2.13 in 8.4.1 splošnega standarda. OPOMBA: Glej tudi točko 4.2 splošnega standarda.

General Information

Status
Withdrawn
Publication Date
16-Aug-2015
Withdrawal Date
10-Apr-2023
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
11-Apr-2023
Due Date
04-May-2023
Completion Date
11-Apr-2023

Relations

Buy Standard

Standard
EN 80601-2-60:2015
English language
38 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 80601-2-60:2015
01-september-2015
0HGLFLQVNDHOHNWULþQDRSUHPDGHO3RVHEQH]DKWHYH]DRVQRYQRYDUQRVWLQ
ELVWYHQHODVWQRVWL]RER]GUDYVWYHQHRSUHPH
Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and
essential performance of dental equipment
Medizinische elektrische Geräte - Teil 2-60: Besondere Festlegungen an die Sicherheit
einschließlich der wesentlichen Leistungsmerkmale von Dental-Geräten
Appareils électromédicaux - Partie 2-60: Exigences particulières pour la sécurité de base
et les performances essentielles des équipements dentaires
Ta slovenski standard je istoveten z: EN 80601-2-60:2015
ICS:
11.060.20 =RERWHKQLþQDRSUHPD Dental equipment
SIST EN 80601-2-60:2015 en
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 80601-2-60:2015

---------------------- Page: 2 ----------------------

SIST EN 80601-2-60:2015


EUROPEAN STANDARD EN 80601-2-60

NORME EUROPÉENNE

EUROPÄISCHE NORM
May 2015
ICS 11.040.01

English Version
Medical electrical equipment - Part 2-60: Particular requirements
for the basic safety and essential performance of dental
equipment
(IEC 80601-2-60:2012)
Appareils électromédicaux - Partie 2-60: Exigences Medizinische elektrische Geräte -- Teil 2-60: Besondere
particulières pour la sécurité de base et les performances Festlegungen an die Sicherheit einschließlich der
essentielles des équipements dentaires wesentlichen Leistungsmerkmale von Dental-Geräten
(IEC 80601-2-60:2012) (IEC 80601-2-60:2012)
This European Standard was approved by CENELEC on 2015-04-14. CENELEC members are bound to comply with the CEN/CENELEC
Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the
same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.



European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2015 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
 Ref. No. EN 80601-2-60:2015 E

---------------------- Page: 3 ----------------------

SIST EN 80601-2-60:2015
EN 80601-2-60:2015 (E)

Foreword
The text of document 62D/964/FDIS, future edition 1 of IEC 80601-2-60, prepared by SC 62D,
"Electromedical equipment", of IEC/TC 62, "Electrical equipment in medical practice" and SC 6 "Dental
equipment" of ISO/TC 106 "Dentistry" was submitted to the IEC-CENELEC parallel vote and approved by
CENELEC as EN 80601-2-60:2015.

The following dates are fixed:
(dop) 2016-01-14
• latest date by which the document has
to be implemented at national level by
publication of an identical national
standard or by endorsement
(dow) 2018-04-14
• latest date by which the national
standards conflicting with the
document have to be withdrawn

Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC [and/or CEN] shall not be held responsible for identifying any or all such patent
rights.

This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association, and supports essential requirements of EU Directive(s).

For the relationship with EU Directive see informative Annex ZZ, which is an integral part of this
document.
Endorsement notice
The text of the International Standard IEC 80601-2-60:2012 was approved by CENELEC as a European
Standard without any modification.
In the official version, for Bibliography, the following notes have to be added for the standards indicated:
IEC 61810-7:2006 NOTE  Harmonized as EN 61810-7:2006 (not modified).
ISO 13732-1:2006 NOTE  Harmonized as EN ISO 13732-1:2008 (not modified).
ISO 17664:2004 NOTE  Harmonized as EN ISO 17664:2004 (not modified).
ISO 7494-2:2003 NOTE  Harmonized as EN ISO 7494-2:2003 (not modified).
ISO 21530:2004 NOTE  Harmonized as EN ISO 21530:2004 (not modified).

2

---------------------- Page: 4 ----------------------

SIST EN 80601-2-60:2015
 EN 80601-2-60:2015 (E)

Annex ZA
(normative)

Normative references to international publications
with their corresponding European publications

The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.

NOTE  When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.

Annex ZA of EN 60601-1:2006 applies with the following exceptions:

Publication Year Title EN/HD Year

Replacement:

IEC 60664-1 2007 Insulation coordination for equipment within EN 60664-1 2007
low-voltage systems -
Part 1: Principles, requirements and tests


IEC 60825-1 - Safety of laser products - EN 60825-1 -
Part 1: Equipment classification and
requirements


Addition:

IEC 60601-2-2 2009 Medical electrical equipment - EN 60601-2-2 2009
 Part 2-2: Particular requirements for basic + A11 2011
safety and essential performance of high
frequency surgical equipment and high
frequency surgical accessories


IEC 60601-2-22 2007 Medical electrical equipment - - -
Part 2-22: Particular requirements for basic
safety and essential performance of surgical,
therapeutic and diagnostic laser equipment


IEC 60601-2-57 2011 Medical electrical equipment - EN 60601-2-57 2011
Part 2-57: Particular requirements for basic
safety and essential performance of non-laser
light source equipment intended for
therapeutic, diagnostic, monitoring and
cosmetic/aesthetic use


IEC 60664-4 2005 Insulation coordination for equipment within EN 60664-4 2006
low-voltage systems - + corr. October 2006
Part 4: Consideration of high-frequency
voltage stress


IEC 61180-1 - High-voltage test techniques for low-voltage EN 61180-1 -
equipment -
Part 1: Definitions, test and procedure
requirements


IEC 61180-2 - High-voltage test techniques for low-voltage EN 61180-2 -
equipment -
Part 2: Test equipment


IEC 61810-1 2008 Electromechanical elementary relays - EN 61810-1 2008
+ corr. February 2010 Part 1: General requirements


IEC 62471 - Photobiological safety of lamps and lamp EN 62471 -
systems


ISO 1942 - Dentistry - Vocabulary EN ISO 1942 -


3

---------------------- Page: 5 ----------------------

SIST EN 80601-2-60:2015
EN 80601-2-60:2015 (E)
Publication Year Title EN/HD Year

ISO 7785-2 - Dental handpieces - EN ISO 7785-2 -
Part 2: Straight and geared angle handpieces

4

---------------------- Page: 6 ----------------------

SIST EN 80601-2-60:2015
 EN 80601-2-60:2015 (E)

Annex ZZ
(informative)

Coverage of Essential Requirements of EU Directives
This European Standard has been prepared under a mandate given to CENELEC by the European
Commission and the European Free Trade Association, and within its scope the Standard covers all
relevant essential requirements given in Annex I of EU Directive 93/42/EEC of 14 June 1993
concerning medical devices.
Compliance with this standard provides one means of conformity with the specified essential
requirements of the Directive concerned.
WARNING: Other requirements and other EU Directives can be applied to the products falling within
the scope of this standard.

5

---------------------- Page: 7 ----------------------

SIST EN 80601-2-60:2015

---------------------- Page: 8 ----------------------

SIST EN 80601-2-60:2015



IEC 80601-2-60


Edition 1.0 2012-02




INTERNATIONAL



STANDARD




NORME



INTERNATIONALE











Medical electrical equipment –

Part 2-60: Particular requirements for the basic safety and essential performance

of dental equipment




Appareils électromédicaux –

Partie 2-60: Exigences particulières pour la sécurité de base et les performances


essentielles des équipements dentaires













INTERNATIONAL

ELECTROTECHNICAL

COMMISSION


COMMISSION

ELECTROTECHNIQUE

PRICE CODE
INTERNATIONALE

CODE PRIX V


ICS 11.040.01 ISBN 978-2-88912-914-0



Warning! Make sure that you obtained this publication from an authorized distributor.

Attention! Veuillez vous assurer que vous avez obtenu cette publication via un distributeur agréé.

---------------------- Page: 9 ----------------------

SIST EN 80601-2-60:2015
– 2 – 80601-2-60 © IEC:2012
CONTENTS
FOREWORD . 3
201.1 Scope, object and related standards. 5
201.2 Normative references . 6
201.3 Terms and definitions . 7
201.4 General requirements . 8
201.5 General requirements for testing of ME EQUIPMENT . 8
201.6 Classification of ME EQUIPMENT and ME SYSTEMS . 9
201.7 ME EQUIPMENT identification, marking and documents . 9
201.8 Protection against electrical HAZARDS from ME EQUIPMENT . 10
201.9 Protection against mechanical HAZARDS of ME EQUIPMENT and ME SYSTEMS . 14
201.10 Protection against unwanted and excessive radiation HAZARDS . 16
201.11 Protection against excessive temperatures and other HAZARDS . 16
201.12 Accuracy of controls and instruments and protection against hazardous
outputs . 20
201.13 HAZARDOUS SITUATIONS and fault conditions . 20
201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) . 21
201.15 Construction of ME EQUIPMENT . 21
201.16 ME SYSTEMS. 21
201.17 Electromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS . 21
201.101 Cordless HAND-HELD and foot-operated control devices . 21
Annexes . 21
Annex AA (informative) Particular guidance and rationale . 22
Bibliography . 31
Index of defined terms used in this particular standard. 32

Figure AA.1 – Example of APPLIED PARTS for DENTAL EQUIPMENT . 23
Figure AA.2 – Calculation of LEAKAGE CURRENT . 24
Figure AA.3 – Insulation problem of commutator DENTAL ELECTRICAL MOTOR . 25
Figure AA.4 – Loading fan construction. 29
Figure AA.5 – Load diagram with loading fan . 30

Table 201.101 – Test voltages for solid insulation for SECONDARY CIRCUITS according to
201.8.5.2 . 10
Table 201.102 – Determination of TENSILE SAFETY FACTOR . 15
Table 201.103 – Mass distribution . 16
Table 201.104 – Allowable maximum temperatures for DENTAL HANDPIECE . 17

---------------------- Page: 10 ----------------------

SIST EN 80601-2-60:2015
80601-2-60 © IEC:2012 – 3 –
INTERNATIONAL ELECTROTECHNICAL COMMISSION
____________

MEDICAL ELECTRICAL EQUIPMENT –

Part 2-60: Particular requirements for the basic safety
and essential performance of dental equipment

FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity
assessment services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any
services carried out by independent certification bodies.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 80601-2-60 has been prepared by a Joint Working Group of
subcommittee 62D: Electrical equipment in medical practice of IEC technical committee 62:
Electrical equipment in medical practice and subcommittee 6: Dental equipment of ISO
technical committee 106: Dentistry.
The text of this particular standard is based on the following documents:
FDIS Report on voting
62D/964/FDIS 62D/984/RVD

Full information on the voting for the approval of this particular standard can be found in the
report on voting indicated in the above table. In ISO, the standard has been approved by 16
P-members out of 17 having cast a vote.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.

---------------------- Page: 11 ----------------------

SIST EN 80601-2-60:2015
– 4 – 80601-2-60 © IEC:2012
In this standard, the following print types are used:
– Requirements and definitions: roman type.
– Test specifications: italic type.
– Informative material appearing outside of tables, such as notes, examples and references: in smaller type.
Normative text of tables is also in a smaller type.
– TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS
NOTED: SMALL CAPITALS.
In referring to the structure of this standard, the term
– “clause” means one of the seventeen numbered divisions within the table of contents,
inclusive of all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);
– “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all
subclauses of Clause 7).
References to clauses within this standard are preceded by the term “Clause” followed by the
clause number. References to subclauses within this particular standard are by number only.
In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
Directives, Part 2. For the purposes of this standard, the auxiliary verb:
– “shall” means that compliance with a requirement or a test is mandatory for compliance
with this standard;
– “should” means that compliance with a requirement or a test is recommended but is not
mandatory for compliance with this standard;
– “may” is used to describe a permissible way to achieve compliance with a requirement or
test.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title
indicates that there is guidance or rationale related to that item in Annex AA.
A list of all parts of the IEC 60601 series, published under the general title: Medical electrical
equipment, can be found on the IEC website.
The committee has decided that the contents of this publication will remain unchanged until
the stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data
related to the specific publication. At this date, the publication will be
• reconfirmed,
• withdrawn,
• replaced by a revised edition, or
• amended.

---------------------- Page: 12 ----------------------

SIST EN 80601-2-60:2015
80601-2-60 © IEC:2012 – 5 –
MEDICAL ELECTRICAL EQUIPMENT –

Part 2-60: Particular requirements for the basic safety
and essential performance of dental equipment


201.1 Scope, object and related standards
1)
Clause 1 of the general standard applies, except as follows:
201.1.1 Scope
Replacement:
This International Standard applies to the BASIC SAFETY and ESSENTIAL PERFORMANCE of
DENTAL UNITS, DENTAL PATIENT CHAIRS, DENTAL HANDPIECES and DENTAL OPERATING LIGHTS,
hereafter referred to as DENTAL EQUIPMENT.
Excluded are amalgamators, sterilizers and dental X-ray equipment.
If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
ME SYSTEMS only, the title and content of that clause or subclause will say so. If that is not the
case, the clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as relevant.
AZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS
H
within the scope of this standard are not covered by specific requirements in this standard
except in 7.2.13 and 8.4.1 of the general standard.
NOTE See also 4.2 of the general standard.
201.1.2 Object
Replacement:
The object of this particular standard is to establish BASIC SAFETY and
ESSENTIAL PERFORMANCE requirements for DENTAL EQUIPMENT (as defined in 201.3.202.)
201.1.3 Collateral standards
Addition:
This particular standard refers to those applicable collateral standards that are listed in
Clause 2 of the general standard and Clause 201.2 of this particular standard.
2) 3)
IEC 60601-1-3, IEC 60601-1-9 and IEC 60601-1-10 do not apply. All other published
collateral standards in the IEC 60601-1 series apply as published.
___________
1)
The general standard is IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for
basic safety and essential performance.
2)
IEC 60601-1-9:2007, Medical electrical equipment – Part 1-9: General requirements for basic safety and
essential performance – Collateral Standard: Requirements for environmentally conscious design
3)
IEC 60601-1-10:2007, Medical electrical equipment – Part 1-10: General requirements for basic safety and
essential performance – Collateral Standard: Requirements for the development of physiologic closed-loop
controllers

---------------------- Page: 13 ----------------------

SIST EN 80601-2-60:2015
– 6 – 80601-2-60 © IEC:2012
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards may modify, replace or delete requirements
contained in the general standard and collateral standards as appropriate for the particular
ME EQUIPMENT under consideration, and may add other BASIC SAFETY and
ESSENTIAL PERFORMANCE requirements.
A requirement of a particular standard takes priority over the general standard.
For brevity, IEC 60601-1 is referred to in this particular standard as the general standard.
Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this particular standard corresponds to that of
the general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content
of Clause 1 of the general standard) or applicable collateral standard with the prefix “20x”
where x is the final digit(s) of the collateral standard document number (e.g. 202.4 in this
particular standard addresses the content of Clause 4 of the IEC 60601-1-2 collateral
standard, 203.4 in this particular standard addresses the content of Clause 4 of the
IEC 60601-1-3 collateral standard, etc.). The changes to the text of the general standard are
specified by the use of the following words:
"Replacement" means that the clause or subclause of the general standard or applicable
collateral standard is replaced completely by the text of this particular standard.
"Addition" means that the text of this particular standard is additional to the requirements of
the general standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable
collateral standard is amended as indicated by the text of this particular standard.
Subclauses, figures or tables which are additional to those of the general standard are
numbered starting from 201.101. However, due to the fact that definitions in the general
standard are numbered 3.1 through 3.139, additional definitions in this standard are
numbered beginning from 201.3.201. Additional annexes are lettered AA, BB, etc., and
additional items aa), bb), etc.
Subclauses, figures or tables which are additional to those of a collateral standard are
numbered starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for
IEC 60601-1-2, 203 for IEC 60601-1-3, etc.
The term "this standard" is used to make reference to the general standard, any applicable
collateral standards and this particular standard taken together.
Where there is no corresponding clause or subclause in this particular standard, the clause or
subclause of the general standard or applicable collateral standard, although possibly not
relevant, applies without modification; where it is intended that any part of the general
standard or applicable collateral standard, although possibly relevant, is not to be applied, a
statement to that effect is given in this particular standard.
201.2 Normative references
Clause 2 of the general standard applies, except as follows:

---------------------- Page: 14 ----------------------

SIST EN 80601-2-60:2015
80601-2-60 © IEC:2012 – 7 –
Replacement:
IEC 60664-1:2007, Insulation coordination for equipment within low-voltage systems – Part 1:
Principles, requirements and tests
IEC 60825-1, Safety of laser products – Part 1: Equipment classification and requirements
Addition:
IEC 60601-2-2:2009, Medical electrical equipment – Part 2-2: Particular requirements for the
basic safety and essential performance of high frequency surgical equipment and high
frequency surgical accessories
IEC 60601-2-22:2007, Medical electrical equipment – Part 2-22: Particular requirements for
basic safety and essential performance of surgical, cosmetic, therpeutic and diagnostic laser
equipment
IEC 60601-2-57:2011, Medical electrical equipment – Part 2-57: Particular requirements for
the basic safety and essential performance of non-laser light source equipment intended for
therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
IEC 60664-4:2005, Insulation coordination for equipment within low-voltage systems – Part 4:
Consideration of high-frequency voltage stress
IEC 61180-1, High-voltage test techniques for low-voltage equipment – Part 1: Definitions,
test and procedure requirements
IEC 61180-2, High-voltage test techniques for low-voltage equipment – Part 2: Test
equipment
IEC 61810-1:2008, Electromechanical elementary relays – Part 1: General requirements
IEC 62471, Photobiological safety of lamps and lamp systems
ISO 1942, Dentistry – Vocabulary
ISO 7785-2, Dental handpieces – Part 2: Straight and geared angle handpieces
201.3 Terms and definitions
For the purposes of this document, the terms and definitions given in IEC 60601-1:2005,
IEC 60601-2-2:2009 and ISO 1942 apply, except as follows:
NOTE An index of defined terms is found beginning on page 31
Addition:
201.3.201
DENTAL ELECTRICAL MOTOR
handheld part of the DENTAL HANDPIECE electrically powered by the DENTAL UNIT
201.3.202
DENTAL EQUIPMENT
ME EQUIPMENT with any combination of DENTAL HANDPIECES, DENTAL UNITS,
DENTAL PATIENT CHAIRS and DENTAL OPERATING LIGHTS

---------------------- Page: 15 ----------------------

SIST EN 80601-2-60:2015
– 8 – 80601-2-60 © IEC:2012
201.3.203
DENTAL HANDPIECE
handheld instrument used in dentistry for use in PATIENT treatment and connected to the
DENTAL UNIT
201.3.204
DENTAL OPERATING LIGHT
device designed for use by an OPERATOR for illuminating the oral cavity, consisting of a
luminaire and one or more lamps
201.3.205
DENTAL PATIENT CHAIR
device designed to support and position the PATIENT for treatment and therefore provided with
a range of movements
201.3.206
DENTAL UNIT
device through which electrical power and/or various fluids or gasses are supplied to a
number of DENTAL HANDPIECE S and devices
Note to entry It is usually fitted with conveniently oriented instrument holders and controls, and consists of
interconnected sub-units of DENTAL EQUIPMENT and instruments providing a functional unit for dental use.
201.3.207
OPERATOR SIDE OF DENTAL HANDPIECE
part of DENTAL HANDPIECE which is designed to be handheld by the OPERATOR in NORMAL USE
201.3.208
PATIENT SIDE OF DENTAL HANDPIECE
part of DENTAL HANDPIECE which is designed to be introduced into the oral cavity where all
parts of the DENTAL HANDPIECE within 80 mm of the tip shall be considered as an APPLIED PART
201.4 General requirements
Clause 4 of the general standard applies, except as follows:
201.4.3 ESSENTIAL PERFORMANCE
Addition:
DENTAL EQUIPMENT does not have ESSENTIAL PERFORMANCE.
201.5 General requirements for testing of ME EQUIPMENT
Clause 5 of the general standard applies, except as follows:
201.5.3 Ambient temperature, humidity, atmospheric pressure
Amendment of item a):
After the ME EQUIPMENT to be tested has been set up for NORMAL USE (according to 5.7), tests
are performed within the range of environmental conditions indicated in the technical
description (see 7.9.3.1) but
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.