Standard Guide for Thermal Performance Testing of Cryogenic Insulation Systems

SIGNIFICANCE AND USE
5.1 A key aspect in understanding the thermal performance of cryogenic insulation systems is to perform tests under representative and reproducible conditions, simulating the way that the materials are actually put together and used in service. Therefore, a large temperature differential across the insulation and a residual gas environment at some specific pressure are usually required. Added to these requirements are the complexities of thickness measurement at test condition after thermal contraction, verification of surface contact and/or mechanical loading after cooldown, and measurement of high vacuum levels within the material. Accounting for the surface contact resistance can be a particular challenge, especially for rigid materials (32). The imposition of a large differential temperature in generally low density, high surface area materials means that the composition and states of the interstitial species can have drastic changes through the thickness of the system. Even for a single component system such as a sheet of predominately closed-cell foam, the composition of the system will often include air, moisture, and blowing agents at different concentrations and physical states and morphologies throughout the material. The system, as tested under a given set of WBT, CBT, and CVP conditions, includes all of these components (not only the foam material). The CVP can be imposed by design or can vary in response to the change in boundary temperatures as well as the surface effects of the insulation materials. In order for free molecular gas conduction to occur, the mean free path of the gas molecules must be larger than the spacing between the two heat transfer surfaces. The ratio of the mean free path to the distance between surfaces is the Knudsen number (see Guide C740 for further discussion). A Knudsen number greater than 1.0 is termed the molecular flow condition while a Knudsen less than 0.01 is considered a continuum or viscous flow condition. Testing of...
SCOPE
1.1 This guide provides information for the laboratory measurement of the steady-state thermal transmission properties and heat flux of thermal insulation systems under cryogenic conditions. Thermal insulation systems may be composed of one or more materials that may be homogeneous or non-homogeneous; flat, cylindrical, or spherical; at boundary conditions from near absolute zero or 4 K up to 400 K; and in environments from high vacuum to an ambient pressure of air or residual gas. The testing approaches presented as part of this guide are distinct from, and yet complementary to, other ASTM thermal test methods including C177, C518, and C335. A key aspect of this guide is the notion of an insulation system, not an insulation material. Under the practical use environment of most cryogenic applications even a single-material system can still be a complex insulation system (1-3).2 To determine the inherent thermal properties of insulation materials, the standard test methods as cited in this guide should be consulted.  
1.2 The function of most cryogenic thermal insulation systems used in these applications is to maintain large temperature differences thereby providing high levels of thermal insulating performance. The combination of warm and cold boundary temperatures can be any two temperatures in the range of near 0 K to 400 K. Cold boundary temperatures typically range from 4 K to 100 K, but can be much higher such as 300 K. Warm boundary temperatures typically range from 250 K to 400 K, but can be much lower such as 40 K. Large temperature differences up to 300 K are typical. Testing for thermal performance at large temperature differences with one boundary at cryogenic temperature is typical and representative of most applications. Thermal performance as a function of temperature can also be evaluated or calculated in accordance with Practices C1058 or C1045 when sufficient information on the temperature profile and ...

General Information

Status
Published
Publication Date
14-Mar-2024
Technical Committee
Drafting Committee
Current Stage
Ref Project

Relations

Buy Standard

Guide
ASTM C1774-24 - Standard Guide for Thermal Performance Testing of Cryogenic Insulation Systems
English language
25 pages
sale 15% off
Preview
sale 15% off
Preview
Guide
REDLINE ASTM C1774-24 - Standard Guide for Thermal Performance Testing of Cryogenic Insulation Systems
English language
25 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: C1774 − 24
Standard Guide for
Thermal Performance Testing of Cryogenic Insulation
1
Systems
This standard is issued under the fixed designation C1774; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope accordance with Practices C1058 or C1045 when sufficient
information on the temperature profile and physical modeling
1.1 This guide provides information for the laboratory
are available.
measurement of the steady-state thermal transmission proper-
1.3 The range of residual gas pressures for this Guide is
ties and heat flux of thermal insulation systems under cryo-
-7 +3 -5
from 10 torr to 10 torr (1.33 Pa to 133 kPa) with different
genic conditions. Thermal insulation systems may be com-
purge gases as required. Corresponding to the applications in
posed of one or more materials that may be homogeneous or
cryogenic systems, three sub-ranges of vacuum are also de-
non-homogeneous; flat, cylindrical, or spherical; at boundary
-6 -3 -4
fined: High Vacuum (HV) from <10 torr to 10 torr (1.333
conditions from near absolute zero or 4 K up to 400 K; and in
Pa to 0.133 Pa) [free molecular regime], Soft Vacuum (SV)
environments from high vacuum to an ambient pressure of air
-2
from 10 torr to 10 torr (from 1.33 Pa to 1,333 Pa) [transition
or residual gas. The testing approaches presented as part of this
regime], No Vacuum (NV) from 100 torr to 1000 torr (13.3 kPa
guide are distinct from, and yet complementary to, other
to 133 kPa) [continuum regime].
ASTM thermal test methods including C177, C518, and C335.
A key aspect of this guide is the notion of an insulation system,
1.4 Thermal performance can vary by four orders of mag-
not an insulation material. Under the practical use environment
nitude over the entire vacuum pressure range. Effective thermal
of most cryogenic applications even a single-material system
conductivities can range from 0.010 mW/m-K to 100 mW/
2
can still be a complex insulation system (1-3). To determine
m-K. The primary governing factor in thermal performance is
the inherent thermal properties of insulation materials, the
the pressure of the test environment. High vacuum insulation
standard test methods as cited in this guide should be con-
systems are often in the range from 0.05 mW/m-K to 2
sulted.
mW/m-K while non-vacuum systems are typically in the range
from 10 mW/m-K to 30 mW/m-K. Soft vacuum systems are
1.2 The function of most cryogenic thermal insulation
generally between these two extremes (4). Of particular de-
systems used in these applications is to maintain large tem-
mand is the very low thermal conductivity (very high thermal
perature differences thereby providing high levels of thermal
resistance) range in sub-ambient temperature environments.
insulating performance. The combination of warm and cold
For example, careful delineation of test results in the range of
boundary temperatures can be any two temperatures in the
0.01 mW/m-K to 1 mW/m-K (from R-value 14,400 to R-value
range of near 0 K to 400 K. Cold boundary temperatures
144) is required as a matter of normal engineering applications
typically range from 4 K to 100 K, but can be much higher
for many cryogenic insulation systems (5-7). The application
such as 300 K. Warm boundary temperatures typically range
of effective thermal conductivity values to multilayer insula-
from 250 K to 400 K, but can be much lower such as 40 K.
tion (MLI) systems and other combinations of diverse
Large temperature differences up to 300 K are typical. Testing
materials, because they are highly anisotropic and specialized,
for thermal performance at large temperature differences with
must be done with due caution and full provision of supporting
one boundary at cryogenic temperature is typical and repre-
2
technical information (8). The use of heat flux (W/m ) is, in
sentative of most applications. Thermal performance as a
general, more suitable for reporting the thermal performance of
function of temperature can also be evaluated or calculated in
MLI systems (9-11).
1.5 This guide covers different approaches for thermal
1
This guide is under the jurisdiction of ASTM Committee C16 on Thermal performance measurement in sub-ambient temperature envi-
Insulation and is the direct responsibility of Subcommittee C16.30 on Thermal
ronments. The test apparatuses (apparatus) are divided into two
Measurement.
categories: boiloff calorimetry and electrical power. Both
Current
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: C1774 − 13 (Reapproved 2019) C1774 − 24
Standard Guide for
Thermal Performance Testing of Cryogenic Insulation
1
Systems
This standard is issued under the fixed designation C1774; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This guide provides information for the laboratory measurement of the steady-state thermal transmission properties and heat
flux of thermal insulation systems under cryogenic conditions. Thermal insulation systems may be composed of one or more
materials that may be homogeneous or non-homogeneous; flat, cylindrical, or spherical; at boundary conditions from near absolute
zero or 4 K up to 400 K; and in environments from high vacuum to an ambient pressure of air or residual gas. The testing
approaches presented as part of this guide are distinct from, and yet complementary to, other ASTM thermal test methods including
C177, C518, and C335. A key aspect of this guide is the notion of an insulation system, not an insulation material. Under the
practical use environment of most cryogenic applications even a single-material system can still be a complex insulation system
2
(1-3). To determine the inherent thermal properties of insulation materials, the standard test methods as cited in this guide should
be consulted.
1.2 The function of most cryogenic thermal insulation systems used in these applications is to maintain large temperature
differences thereby providing high levels of thermal insulating performance. The combination of warm and cold boundary
temperatures can be any two temperatures in the range of near 0 K to 400 K. Cold boundary temperatures typically range from
4 K to 100 K, but can be much higher such as 300 K. Warm boundary temperatures typically range from 250 K to 400 K, but can
be much lower such as 40 K. Large temperature differences up to 300 K are typical. Testing for thermal performance at large
temperature differences with one boundary at cryogenic temperature is typical and representative of most applications. Thermal
performance as a function of temperature can also be evaluated or calculated in accordance with Practices C1058 or C1045 when
sufficient information on the temperature profile and physical modeling are available.
-7 +3 -5
1.3 The range of residual gas pressures for this Guide is from 10 torr to 10 torr (1.33 Pa to 133 kPa) with different purge
gases as required. Corresponding to the applications in cryogenic systems, three sub-ranges of vacuum are also defined: High
-6 -3 -4 -2
Vacuum (HV) from <10 torr to 10 torr (1.333 Pa to 0.133 Pa) [free molecular regime], Soft Vacuum (SV) from 10 torr to
10 torr (from 1.33 Pa to 1,333 Pa) [transition regime], No Vacuum (NV) from 100 torr to 1000 torr (13.3 kPa to 133 kPa)
[continuum regime].
1.4 Thermal performance can vary by four orders of magnitude over the entire vacuum pressure range. Effective thermal
conductivities can range from 0.010 mW/m-K to 100 mW/m-K. The primary governing factor in thermal performance is the
pressure of the test environment. High vacuum insulation systems are often in the range from 0.05 mW/m-K to 2 mW/m-K while
non-vacuum systems are typically in the range from 10 mW/m-K to 30 mW/m-K. Soft vacuum systems are generally between
these two extremes (4). Of particular demand is the very low thermal conductivity (very high thermal resistance) range in
sub-ambient temperature environments. For example, careful delineation of test results in the range of 0.01 mW/m-K to 1 mW/m-K
1
This guide is under the jurisdiction of ASTM Committee C16 on Thermal Insulation and is the direct responsibility of Subcommittee C16.30 on Thermal Measurement.
Current edition approved Sept. 1, 2019March 15, 2024. Published October 2019April 2024. Originally approved in 2013. Last previous edition approved in 20132019 as
C1774 – 13.C1774 – 13 (2019). DOI: 10.1520/C1774-13R19.10.1520/C1774-24.
2
The boldface numbers in parentheses refer to the list of references at the end of this standard.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
C1774 − 24
(from R-value 14,400 to R-value 144) is required as a matter of normal engineering applications for many cryogenic insulation
systems (5
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.