ASTM E2335-08
(Guide)Standard Guide for Laboratory Monitors
Standard Guide for Laboratory Monitors
SIGNIFICANCE AND USE
These guidelines are intended to enable a laboratory monitor to perform onsite assessments of laboratory facilities. Accepted facilities will then participate in an interlaboratory test protocol to establish precision and bias for a particular ASTM standard.
This assessment is intended to determine that all of the participants have the necessary equipment, an understanding of the test method, and the minimum level of skill necessary to gather data which are to be used to establish precision and bias for the particular standard.
This guide is not intended to be used as a tool to qualify or accredit laboratories to perform any tests. Refer to the Referenced Documents for further guidance on accreditation requirements.
SCOPE
1.1 This guide provides a general outline, for use by a laboratory monitor, to assess the qualifications of a laboratory which has requested to participate in a specific ASTM interlaboratory test.
1.2 The preliminary assessment is based on observations made before initiation of any interlaboratory tests.
1.3 The subcommittee, which appoints the laboratory monitor, specifies the minimum requirements which an organization should meet to qualify as a participant for the interlaboratory test. If a pretest is to be included, see details in Section 13.
1.4 This fire standard can not be used to provide quantitative measures.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory requirements prior to use.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E2335 − 08 AnAmerican National Standard
Standard Guide for
1
Laboratory Monitors
This standard is issued under the fixed designation E2335; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
This guide is for use by laboratory monitors who observe interlaboratory testing performed by
participatingorganizationsthatvalidatethelegitimacyoftestmethodsandalsoestablishabaseofdata
fromwhichprecisionandbiasstatementsaretobewritten.Interlaboratorytestingisdonetodetermine
the validity of a test standard. Such testing requires diligence by adhering to the protocol established
for performing the test in order to minimize variability which often exists between laboratories.
Because laboratories differ in size, organization, management, personnel, facilities and environment
there is always a potential for introducing variability in test results.
Laboratory monitors confirm that the participating facilities adhere to the test protocol as closely as
possible, thereby improving the chances of a successful interlaboratory study. The laboratory monitor
observes and assesses but does not perform the functions of an auditor or trainer; the laboratory
monitor does not provide accreditation of a laboratory. Laboratory monitors should demonstrate
expertise in both the content and the intent of the test method. The laboratory monitor is responsible
to render a report to the appointing subcommittee. It is incumbent upon the laboratory monitor to
maintain objectivity when working with participating laboratories.
Full-scale product fire tests are expensive, especially full-scale tests of bedding and home
furnishings. The development of this standard is intended to ensure that the precision and bias
developed for E05 standards demonstrates a high level of statistical validity.
This guide does not contain procedures for analysis of the test data obtained from an interlaboratory
study. The user is directed to Practice E691, if the laboratory study includes six laboratories or more,
and to Practice E2653 if the number of laboratories in the interlaboratory study is at least three, but
less than six.
Using a laboratory monitor as part of an interlaboratory test program is optional. However, if a
monitor is used, all of the laboratories in the project are to be included.
1. Scope 1.2 The preliminary assessment is based on observations
made before initiation of any interlaboratory tests.
1.1 This guide provides a general outline, for use by a
laboratory monitor, to assess the qualifications of a laboratory
1.3 The subcommittee, which appoints the laboratory
which has requested to participate in a specific ASTM inter-
monitor, specifies the minimum requirements which an orga-
laboratory test.
nization should meet to qualify as a participant for the
interlaboratory test. If a pretest is to be included, see details in
1
This guide is under the jurisdiction ofASTM Committee E05 on Fire Standards
Section 13.
and is the direct responsibility of Subcommittee E05.15 on Furnishings and
Contents.
1.4 Thisfirestandardcannotbeusedtoprovidequantitative
Current edition approved Oct. 1, 2008. Published November 2008. Originally
measures.
approved in 2004. Last previous edition approved in 2004 as E2335-04. DOI:
10.1520/E2335-08.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
---------------------- Page: 1 ----------------------
E2335 − 08
1.5 This standard does not purport to address all of the 4.3 This guide is not intended to be used as a tool to qualify
safety concerns, if any, associated with its use. It is the or accredit laboratories to perform any tests. Refer to the
responsibility of the user of this standard to establish appro- Referenced Documents for further guidance on accreditation
priate safety and health practices and determine the applica- requirements.
bility of regulatory requirements prior to use.
5. Qualifications of a Laboratory Monitor
2. Referenced Documents
5.1 The laboratory monitor should demonstrate a level of
2
2.1 ASTM Standards:
expertise about the test method for which the interlaboratory
E176 Terminology of Fire Standards
test will be done.
E691 Practice for Conducting an Interlaboratory Study to
5.2 Technicalunderstandingofthetestmethodisimperative
Determine the Precision of a Test Method
in order to understand the principles of the measurements to be
E1537 Test Method for Fire Testing of Upholstered Furni-
made during the test procedure.
ture
5.3 The laboratory monitor should demonstrate objectivity.
E1822 Test Method for Fi
...
This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
An American National Standard
Designation:E2335–04 Designation:E2335–08
Standard Guide for
1
Laboratory Monitors
This standard is issued under the fixed designation E 2335; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
This guide is for use by laboratory monitors who observe interlaboratory testing performed by
participatingorganizationsthatvalidatethelegitimacyoftestmethodsandalsoestablishabaseofdata
fromwhichprecisionandbiasstatementsaretobewritten.Interlaboratorytestingisdonetodetermine
the validity of a test standard. Such testing requires diligence by adhering to the protocol established
for performing the test in order to minimize variability which often exists between laboratories.
Because laboratories differ in size, organization, management, personnel, facilities and environment
there is always a potential for introducing variability in test results.
Laboratory monitors confirm that the participating facilities adhere to the test protocol as closely as
possible, thereby improving the chances of a successful interlaboratory study. The laboratory monitor
observes and assesses but does not perform the functions of an auditor or trainer; the laboratory
monitor does not provide accreditation of a laboratory. Laboratory monitors should demonstrate
expertise in both the content and the intent of the test method. The laboratory monitor is responsible
to render a report to the appointing subcommittee. It is incumbent upon the laboratory monitor to
maintain objectivity when working with participating laboratories.
Full-scale product fire tests are expensive, especially full-scale tests of bedding and home
furnishings. The development of this standard is intended to ensure that the precision and bias
developed for E5 standards demonstrates a high level of statistical validity.
Full-scale product fire tests are expensive, especially full-scale tests of bedding and home
furnishings. The development of this standard is intended to ensure that the precision and bias
developed for E05 standards demonstrates a high level of statistical validity.
This guide does not contain procedures for analysis of the test data obtained from an interlaboratory
study. The user is directed to Practice E 691, if the laboratory study includes six laboratories or more,
and to Practice E 2653 if the number of laboratories in the interlaboratory study is at least three, but
less than six.
Using a laboratory monitor as part of an interlaboratory test program is optional. However, if a
monitor is used, all of the laboratories in the project are to be included.
1. Scope
1.1 This guide provides a general outline, for use by a laboratory monitor, to assess the qualifications of a laboratory which has
requested to participate in a specific ASTM interlaboratory test.
1.2 The preliminary assessment is based on observations made before initiation of any interlaboratory tests.
1.3 The subcommittee, which appoints the laboratory monitor, specifies the minimum requirements which an organization
should meet to qualify as a participant for the interlaboratory test. If a pretest is to be included, see details in Section 13.
1.4 This fire standard can not be used to provide quantitative measures.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
requirements prior to use.
1
This guide is under the jurisdiction of ASTM Committee E05 on Fire Standards and is the direct responsibility of Subcommittee E05.15 on Furnishings and Contents
.
Current edition approved Jan 1, 2004. Published February 2004.
Current edition approved Oct. 1, 2008. Published November 2008. Originally approved in 2004. Last previous edition approved in 2004 as E 2335-04.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
---------------------- Page: 1 ----------------------
E2335–08
2. Referenced Documents
2
2.1 ASTM Standards:
E 176 Terminology of Fire Standards
E 691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
E 1537 Test Method for Fire Testing of Upholstered Furnitu
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.