Standard Test Method for Obtaining Colorimetric Data From a Visual Display Unit by Spectroradiometry

SIGNIFICANCE AND USE
The most fundamental method for obtaining CIE tristimulus values or other color coordinates for describing the colors of visual display units (VDUs) is by the use of spectroradiometric data. (See CIE No. 18 and 63.) These data are used by summation together with numerical values representing the 1931 CIE Standard Observer and normalized to Km, the maximum spectral luminous efficacy function.
The special requirements for characterizing VDUs possessing narrow or discontinuous spectra are presented and discussed. Modifications to the requirements of Practice E308 are given to correct for the unusual nature of narrow or discontinuous sources.
SCOPE
1.1 This test method prescribes the instrumental measurements required for characterizing the color and brightness of VDUs.
1.2 This test method is specific in scope rather than general as to type of instrument and object.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
31-Oct-2011
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E1336-11 - Standard Test Method for Obtaining Colorimetric Data From a Visual Display Unit by Spectroradiometry
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E1336-11 - Standard Test Method for Obtaining Colorimetric Data From a Visual Display Unit by Spectroradiometry
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1336 − 11
Standard Test Method for
Obtaining Colorimetric Data From a Visual Display Unit by
1
Spectroradiometry
This standard is issued under the fixed designation E1336; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
The fundamental procedure for characterizing the color and luminance of a visual display unit
(VDU) is to obtain the spectroradiometric data under specified measurement conditions, and from
these data to compute CIE chromaticity coordinates and absolute luminance values based on the 1931
CIE Standard Observer.The considerations involved and the procedures to be used to obtain precision
colorimetric data for this purpose are contained in this test method. The values and procedures for
computing CIE chromaticity coordinates are contained in Practice E308.The procedures for obtaining
spectroradiometric data are contained in Test Method E1341. This test method includes some
modifications to the procedures given in Practice E308 that are necessary for computing the absolute
luminance values of VDUs. This procedure is intended to be generally applicable to any VDU device,
including but not limited to cathode ray tubes (CRT), liquid crystal displays (LCD), and electrolu-
minescent displays (ELD).
1. Scope E308 PracticeforComputingtheColorsofObjectsbyUsing
the CIE System
1.1 This test method prescribes the instrumental measure-
E1341 Practice for Obtaining Spectroradiometric Data from
ments required for characterizing the color and brightness of
Radiant Sources for Colorimetry
VDUs.
2.2 CIE Publications:
1.2 This test method is specific in scope rather than general 3
Publication CIE No. 18 Principles of Light Measurements
as to type of instrument and object. 3
Publication CIE No. 15.2 Colorimetry, 2nd ed., 1986
1.3 The values stated in SI units are to be regarded as Publication CIE No. 63 Spectroradiometric Measurement of
3
standard. No other units of measurement are included in this Light Sources, 1984
standard.
2.3 IEC Publications:
Publication No. 441 Photometric and Colorimetric Methods
1.4 This standard does not purport to address all of the
of Measurement of the Light Emitted by a Cathode-Ray
safety concerns, if any, associated with its use. It is the
4
Tube Screen, 1974
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
3. Terminology
bility of regulatory limitations prior to use.
3.1 The definitions of appearance terms in Terminology
2. Referenced Documents
E284 are applicable to this test method.
2
2.1 ASTM Standards:
4. Summary of Test Method
E284 Terminology of Appearance
4.1 Procedures are given for obtaining spectroradiometric
data and for the calculation of CIE tristimulus values and other
1
This test method is under the jurisdiction of ASTM Committee E12 on Color
color coordinates to describe the colors of VDUs. Modifica-
andAppearance and is the direct responsibility of Subcommittee E12.06 on Display,
tions to the standard calculation procedures of Practice E308
Imaging and Imaging Colorimetry.
Current edition approved Nov. 1, 2011. Published November 2011. Originally
approved in 1991. Last previous edition approved in 2003 as E1336 – 96 (2003).
3
DOI: 10.1520/E1336-11. Available from U.S. National Committee of the CIE (International Commission
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or on Illumination), C/o Thomas M. Lemons, TLA-Lighting Consultants, Inc., 7 Pond
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM St., Salem, MA 01970, http://www.cie-usnc.org.
4
Standards volume information, refer to the standard’s Document Summary page on Available from International Electrotechnical Commission (IEC), 3 rue de
the ASTM website. Varembé, Case postale 131, CH-1211, Geneva 20, Switzerland, http://www.iec.ch.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1336 − 11
are described.
1

---------------------- Page: 2 ----------------------
E1336 − 11
5. Significance and Use 7.2.2 Monochromatic emission sources such as a low-
pressure mercury arc lamp or tunable laser should also be
5.1 The most fundamental method for obtaining CIE tris-
available for use in calibrating the wavelength scale.
timulus values or other color coordinates for describing the
7.2.3 The electrical supplies for the calibration sources
colors of visual display units (VDUs) is by the use of
should be of the constant current type. Th
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation:E1336–96(Reapproved2003) Designation:E1336–11
Standard Test Method for
Obtaining Colorimetric Data From a Visual Display Unit by
1
Spectroradiometry
This standard is issued under the fixed designation E1336; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
The fundamental procedure for characterizing the color and luminance of a visual display unit
(VDU) is to obtain the spectroradiometric data under specified measurement conditions, and from
these data to compute CIE chromaticity coordinates and absolute luminance values based on the 1931
CIE Standard Observer.The considerations involved and the procedures to be used to obtain precision
colorimetric data for this purpose are contained in this test method. The values and procedures for
computing CIE chromaticity coordinates are contained in Practice E308.The procedures for obtaining
spectroradiometric data are contained in Test Method E1341. This test method includes some
modifications to the procedures given in Practice E308 that are necessary for computing the absolute
luminance values ofVDUs.This procedure is intended to be generally applicable to anyVDU device,
including but not limited to cathode ray tubes (CRT), liquid crystal displays (LCD), and electrolu-
minescent displays (ELD).
1. Scope
1.1 This test method prescribes the instrumental measurements required for characterizing the color and brightness of VDUs.
1.2 This test method is specific in scope rather than general as to type of instrument and object.
1.3The values stated in SI units are to be regarded as the standard.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2
2.1 ASTM Standards:
E284 Terminology of Appearance
E308 Practice for Computing the Colors of Objects by Using the CIE System
E1341 Practice for Obtaining Spectroradiometric Data from Radiant Sources for Colorimetry
2.2 CIE Publications:
3
Publication CIE No. 18 Principles of Light Measurements
3
Publication CIE No. 15.2 Colorimetry, 2nd ed., 1986
3
Publication CIE No. 63 Spectroradiometric Measurement of Light Sources, 1984
2.3 IEC Publications:
1
This test method is under the jurisdiction of ASTM Committee E12 on Color and Appearance and is the direct responsibility of Subcommittee E12.06 on Image Based
Color Measurement.
Current edition approved July 10, 2003. Published July 2003. Originally approved in 1991. Last previous edition approved in 1996 as E1336–96. DOI:
10.1520/E1336-96R03.
Current edition approved Nov. 1, 2011. Published November 2011. Originally approved in 1991. Last previous edition approved in 2003 as E1336 – 96 (2003). DOI:
10.1520/E1336-11.
2
For referencedASTM standards, visit theASTM website, www.astm.org, or contactASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
3
Available from U.S. National Committee of the CIE (International Commission on Illumination), C/o Thomas M. Lemons, TLA-Lighting Consultants, Inc., 7 Pond St.,
Salem, MA 01970, http://www.cie-usnc.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
E1336–11
Publication No. 441 Photometric and Colorimetric Methods of Measurement of the Light Emitted by a Cathode-Ray Tube
4
Screen, 1974
3. Terminology
3.1 The definitions of appearance terms in Terminology E284 are applicable to this test method.
4. Summary of Test Method
4.1 Procedures are given for obtaining spectroradiometric data and for the calculation of CIE tristimulus values and other color
coordinates to describe the colors of VDUs. Modifications to the standard calculation procedures of Practice E308 are described.
5. Significance and Use
5.1 The most fundamental method for obtaining CIE tristimulus values or other color coordinates for describing the colors of
visu
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.