Standard Practice for Calibration of a Liquid-Borne Particle Counter Using an Optical System Based Upon Light Extinction

SCOPE
1.1 This practice covers procedures for calibrating and determining performance of an optical liquid-borne particle counter (LPC) which uses an optical system based upon light extinction measurement. This practice is directed towards determination of accuracy and resolution of the LPC for characterizing the size and number of particles, which have been passed into the sample inlet of the LPC. Consideration of inlet sampling efficiency is not part of this practice.
1.2 The procedures covered in this practice include those to measure sample volume and flow rate, zero count level, particle sizing and counting accuracy, particle sizing resolution, particle counting efficiency, and particle concentration limit.
1.3 The particle size parameter reported in this practice is the equivalent optical diameter based on projected area of calibration particles with known physical properties dispersed in liquid. The manufacturer normally specifies the minimum diameter that can be reported by an LPC; the dynamic range of the LPC being used determines the maximum diameter that can be reported for a single sample. Typical minimum reported diameters are approximately 2 ´m, and a typical dynamic range specification will be approximately from 50 to 1.
1.4 The counting rate capability of the LPC is limited by temporal coincidence of particles in the sensing volume of the LPC and by the saturation level or maximum counting rate capability of the electronic sizing and counting circuitry. Coincidence is defined as the simultaneous presence of more than one particle within the LPC optically defined sensing zone at any time. The coincidence limit is a statistical function of particle concentration in the sample and the sensing zone volume when particle size is insignificant in comparison to the sensing volume dimensions. This limitation may be modified by the presence of particles with dimension so large as to be a significant fraction of the sensing zone dimension. The saturation level rate of the electronic counting circuitry shall be specified by the manufacturer and is normally greater than the LPC recommended maximum counting rate for the particle concentrations used for any portion of this practice.
1.5 Calibration in accordance with all parts of this practice may not be required for routine field calibration of an LPC unless significant changes have occurred in operation of the LPC or major component repairs or replacements have been made. The LPC shall then be taken to a suitable metrology facility for complete calibration. Normal routine field calibration may determine sample flow rate, zero count level, and particle sizing accuracy. The specific LPC functions to be calibrated shall be determined on the basis of agreement between the purchaser and the user. The maximum time interval between calibrations shall be determined by agreement between the purchaser and the user, but shall not exceed twelve months, unless LPC stability for longer periods is verified by measurements in accordance with this practice.
1.6 This standard may involve hazardous materials, operation, and equipment. This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
09-Oct-2000
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM F658-00a - Standard Practice for Calibration of a Liquid-Borne Particle Counter Using an Optical System Based Upon Light Extinction
English language
14 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:F658–00a
Standard Practice for
Calibration of a Liquid-Borne Particle Counter Using an
1
Optical System Based Upon Light Extinction
This standard is issued under the fixed designation F658; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope by the presence of particles with dimension so large as to be a
3
significant fraction of the sensing zone dimension . The
1.1 This practice covers procedures for calibrating and
saturationlevelrateoftheelectroniccountingcircuitryshallbe
determining performance of an optical liquid-borne particle
specified by the manufacturer and is normally greater than the
counter (LPC) which uses an optical system based upon light
LPC recommended maximum counting rate for the particle
extinction measurement. This practice is directed towards
concentrations used for any portion of this practice.
determination of accuracy and resolution of the LPC for
1.5 Calibration in accordance with all parts of this practice
characterizing the size and number of particles, which have
may not be required for routine field calibration of an LPC
been passed into the sample inlet of the LPC. Consideration of
unless significant changes have occurred in operation of the
inlet sampling efficiency is not part of this practice.
LPC or major component repairs or replacements have been
1.2 The procedures covered in this practice include those to
made. The LPC shall then be taken to a suitable metrology
measure sample volume and flow rate, zero count level,
facility for complete calibration. Normal routine field calibra-
particle sizing and counting accuracy, particle sizing resolu-
tion may determine sample flow rate, zero count level, and
tion, particle counting efficiency, and particle concentration
particle sizing accuracy. The specific LPC functions to be
limit.
calibrated shall be determined on the basis of agreement
1.3 The particle size parameter reported in this practice is
between the purchaser and the user. The maximum time
the equivalent optical diameter based on projected area of
intervalbetweencalibrationsshallbedeterminedbyagreement
calibration particles with known physical properties dispersed
betweenthepurchaserandtheuser,butshallnotexceedtwelve
in liquid. The manufacturer normally specifies the minimum
months, unless LPC stability for longer periods is verified by
diameterthatcanbereportedbyanLPC;thedynamicrangeof
measurements in accordance with this practice.
theLPCbeinguseddeterminesthemaximumdiameterthatcan
1.6 This standard may involve hazardous materials, opera-
be reported for a single sample. Typical minimum reported
tion, and equipment. This standard does not purport to address
diametersareapproximately2µm,andatypicaldynamicrange
all of the safety concerns, if any, associated with its use. It is
specification will be approximately from 50 to 1.
the responsibility of the user of this standard to establish
1.4 The counting rate capability of the LPC is limited by
appropriate safety and health practices and determine the
temporal coincidence of particles in the sensing volume of the
applicability of regulatory limitations prior to use.
LPC and by the saturation level or maximum counting rate
capability of the electronic sizing and counting circuitry.
2. Referenced Documents
Coincidence is defined as the simultaneous presence of more
2.1 ASTM Standards:
thanoneparticlewithintheLPCopticallydefinedsensingzone
4
D1193 Specification for Reagent Water
at any time. The coincidence limit is a statistical function of
5
D3195 Practice for Rotameter Calibration
particle concentration in the sample and the sensing zone
E20 Practice for Particle Size Analysis of Particulate Sub-
volume when particle size is insignificant in comparison to the
2
stances in the Range of 0.2 to 75 Micrometres by Optical
sensing volume dimensions . This limitation may be modified
6
Microscopy
1
ThispracticeisunderthejurisdictionofASTMCommitteeE29onParticleand
3
Spray Characterization and is the direct responsibility of Subcommittee E29.02 on Knapp, J. Z. and Abramson, L. R., “A New Coincidence Model for Single
Non-Sieving Methods. Particle Counters. I Theory and Experimental Verification,” Journal of Parenteral
Current edition approved October 10, 2000. Published November 2000. Origi- Science and Technology, Vol 48, 1994, pp. 255-294.
4
nally published as F658–99. Last previous edition F658–00. Annual Book of ASTM Standards, Vol 11.01.
2 5
Jaenicke, R., “The Optical Particle Counter: Cross-Sensitivity and Coinci- Annual Book of ASTM Standards, Vol 11.03.
6
dence,” Journal of Aerosol Science, Vol 3, 1972, pp. 95-111.
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.