Standard Test Method for Determination of Additive Elements in Lubricating Oils by EDXRF Analysis

SIGNIFICANCE AND USE
Lubricating oils are formulated with organo-metallic additives, which act, for example, as detergents, antioxidants, antifoaming, or antiwear agents, or a combination thereof. Some of these additives contain one or more of the following elements: magnesium, phosphorus, sulfur, chlorine, calcium, zinc, and molybdenum. This test method provides a means of determining the concentrations of these elements, which in turn provides an indication of the additive content of these oils.
Additive packages are the concentrates that are used to blend lubricating oils.
This test method is primarily intended to be used for the monitoring of additive elements in lubricating oils.
If this test method is applied to lubricating oils with matrices significantly different from the calibration materials specified in this test method, the cautions and recommendations in Section 6 should be observed when interpreting the results.
SCOPE
1.1 This test method covers the quantitative determination of additive elements in unused lubricating oils and additive packages, as shown in Table 1. The pooled limit of quantitation of this test method as obtained by statistical analysis of interlaboratory test results is 0.02% for magnesium, 0.003 % for phosphorus, 0.002 % for sulfur, 0.001 % for chlorine, 0.003 % for calcium, 0.001 % for zinc, and 0.002 % for molybdenum.
1.2 Additive packages require dilution with a contamination free diluent (base oil) prior to analysis. The dilution factor has to be calculated from the expected concentrations to bring the concentrations for all elements into the ranges listed in Table 1.
1.3 Some lubrication oils will contain higher concentrations than the maximum concentrations listed in Table 1. These samples require dilution with a contamination free diluent (base oil) prior to analysis. The dilution factor has to be calculated from the expected concentrations to bring the concentrations for all elements into the ranges listed in Table 1.
1.4 This test method is limited to the use of energy dispersive X-ray fluorescence (EDXRF) spectrometers employing an X-ray tube for excitation in conjunction with the ability to separate the signals of adjacent elements by using a high resolution semiconductor detector.
1.5 This test method uses inter-element correction factors calculated from a fundamental parameters (FP) approach or from another matrix correction method.
1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.6.1 The preferred concentration units are mg/kg or mass %.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
TABLE 1 Elements and Range of Applicability  ElementConcentration Range in mass % Magnesium0.02 to 0.4 Phosphorous0.003 to 0.25 Sulfur0.002 to 1.5 Chlorine0.001 to 0.4 Calcium0.003 to 1.0 Zinc0.001 to 0.25 Molybdenum0.002 to 0.05

General Information

Status
Historical
Publication Date
30-Sep-2011
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D7751-11e1 - Standard Test Method for Determination of Additive Elements in Lubricating Oils by EDXRF Analysis
English language
6 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
´1
Designation: D7751 − 11
StandardTest Method for
Determination of Additive Elements in Lubricating Oils by
1
EDXRF Analysis
This standard is issued under the fixed designation D7751; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1
´ NOTE—Added research report footnote editorially in November 2011.
1. Scope 1.7 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
1.1 This test method covers the quantitative determination
responsibility of the user of this standard to establish appro-
of additive elements in unused lubricating oils and additive
priate safety and health practices and determine the applica-
packages,asshowninTable1.Thepooledlimitofquantitation
bility of regulatory limitations prior to use.
of this test method as obtained by statistical analysis of
interlaboratory test results is 0.02% for magnesium, 0.003 %
2. Referenced Documents
forphosphorus,0.002%forsulfur,0.001%forchlorine,0.003
2
2.1 ASTM Standards:
% for calcium, 0.001 % for zinc, and 0.002 % for molybde-
D4057 Practice for Manual Sampling of Petroleum and
num.
Petroleum Products
1.2 Additivepackagesrequiredilutionwithacontamination
D4177 Practice for Automatic Sampling of Petroleum and
free diluent (base oil) prior to analysis. The dilution factor has
Petroleum Products
to be calculated from the expected concentrations to bring the
D6299 Practice for Applying Statistical Quality Assurance
concentrationsforallelementsintotherangeslistedinTable1.
and Control Charting Techniques to Evaluate Analytical
1.3 Some lubrication oils will contain higher concentrations
Measurement System Performance
than the maximum concentrations listed in Table 1. These D6300 Practice for Determination of Precision and Bias
samples require dilution with a contamination free diluent
Data for Use in Test Methods for Petroleum Products and
(base oil) prior to analysis. The dilution factor has to be Lubricants
calculated from the expected concentrations to bring the
D6792 Practice for Quality System in Petroleum Products
concentrationsforallelementsintotherangeslistedinTable1.
and Lubricants Testing Laboratories
D7343 Practice for Optimization, Sample Handling,
1.4 This test method is limited to the use of energy
Calibration, and Validation of X-ray Fluorescence Spec-
dispersive X-ray fluorescence (EDXRF) spectrometers em-
trometry Methods for Elemental Analysis of Petroleum
ploying an X-ray tube for excitation in conjunction with the
Products and Lubricants
ability to separate the signals of adjacent elements by using a
E1621 Guide for X-Ray Emission Spectrometric Analysis
high resolution semiconductor detector.
3
2.2 ISO Standards:
1.5 This test method uses inter-element correction factors
ISO 4259 Determination and application of precision data in
calculated from a fundamental parameters (FP) approach or
relation to methods of test
from another matrix correction method.
3. Terminology
1.6 The values stated in SI units are to be regarded as
standard. No other units of measurement are included in this
3.1 Definitions:
standard.
3.1.1 energy dispersive X-ray spectrometry, n—XRF spec-
1.6.1 The preferred concentration units are mg/kg or mass
trometry applying energy dispersive selection of radiation.
%.
3.2 Abbreviations:
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
1
This test method is under the jurisdiction of ASTM Committee D02 on contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Petroleum Products and Lubricants and is the direct responsibility of Subcommittee Standards volume information, refer to the standard’s Document Summary page on
D02.03 on Elemental Analysis. the ASTM website.
3
CurrenteditionapprovedOct.1,2011.PublishedNovember2011.DOI:10.1520/ Available from International Organization for Standardization (ISO), 1, ch. de
D7751-11E01. la Voie-Creuse, CP 56, CH-1211 Geneva 20, Switzerland, http://www.iso.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
´1
D7751 − 11
TABLE 1 Elements and Range of Applicability
elements: magnesium, phosphorus, sulfur, chlorine, calcium,
Element Concentration Range in mass % zinc, and molybdenum. This test method provides a means of
Magnesium 0.02 to 0.4 determiningtheconcentrationsoftheseelements,whichinturn
Phosphorous 0.003 to 0.25
provides an indication of the additive content of these oils.
Sulfur 0.002 to 1.5
Chlorine 0.001 to 0.4
5.2 Additive packages are the conce
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.