Standard Practice for Identification of Instrumental Methods of Color or Color-Difference Measurement of Materials

SIGNIFICANCE AND USE
4.1 The options available in methods for the measurement of color or color-difference are many. These involve choices in: (1) specimens, (2) geometric and spectral properties of instruments, (3) calibration bases for standards used, (4) procedure for sample handling including conditioning, (5) procedure for taking data, and (6) equations for converting instrumental data to final results. Once the measurements have been made, it is essential to document what has been done for the purpose of interlaboratory comparisons, or for future use. A sample form is provided in Fig. 1 to record identifying information applicable to any instrumental method of color or color-difference measurement.  
4.2 Refer to Guide E179, Practices E991, E1164, E1345, E1708, E1767, E2152, and E2194 and Test Methods D5386, D6166, E1247, E1331, E1347, E1348, and E1349, for specific details of measurements.
SCOPE
1.1 This practice covers the documentation of instrumental measurement of color or color difference for current communication or for future reference. The practice is applicable to instrumental measurements of materials where color is seen by reflected, transmitted or emitted light and any combinations of one or more of these processes. The practice is recommended for documentation of methodology in interlaboratory color-measurement programs.  
1.2 Providing an adequate identification of an instrumental measure of color or color-difference involves documenting the metadata necessary for archiving and future use of the measurement data collected. The metadata can be divided in five parts:  
1.2.1 Nature and source of available samples and the form of specimens actually measured,  
1.2.2 Instrumental conditions of measurement, including instrument geometrical and spectral conditions of measurement,  
1.2.3 Standards used,  
1.2.4 Data acquisition procedure, and  
1.2.5 Color scales employed.  
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-May-2022
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E805-22 - Standard Practice for Identification of Instrumental Methods of Color or Color-Difference Measurement of Materials
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E805-22 - Standard Practice for Identification of Instrumental Methods of Color or Color-Difference Measurement of Materials
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E805 −22
Standard Practice for
Identification of Instrumental Methods of Color or Color-
1
Difference Measurement of Materials
This standard is issued under the fixed designation E805; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
2
2.1 ASTM Standards:
1.1 This practice covers the documentation of instrumental
D156 Test Method for Saybolt Color of Petroleum Products
measurement of color or color difference for current commu-
(Saybolt Chromometer Method)
nication or for future reference. The practice is applicable to
D1209 Test Method for Color of Clear Liquids (Platinum-
instrumental measurements of materials where color is seen by
Cobalt Scale)
reflected, transmitted or emitted light and any combinations of
D1500 Test Method forASTM Color of Petroleum Products
one or more of these processes. The practice is recommended
(ASTM Color Scale)
for documentation of methodology in interlaboratory color-
D1535 Practice for Specifying Color by the Munsell System
measurement programs.
D1544 Test Method for Color of Transparent Liquids (Gard-
1.2 Providing an adequate identification of an instrumental
ner Color Scale)
measure of color or color-difference involves documenting the
D2244 Practice for Calculation of Color Tolerances and
metadata necessary for archiving and future use of the mea-
Color Differences from Instrumentally Measured Color
surement data collected. The metadata can be divided in five
Coordinates
parts:
D5386 Test Method for Color of Liquids Using Tristimulus
1.2.1 Nature and source of available samples and the form Colorimetry
of specimens actually measured,
D6166 Test Method for Color of Pine Chemicals and Re-
lated Products (Instrumental Determination of Gardner
1.2.2 Instrumental conditions of measurement, including
Color)
instrument geometrical and spectral conditions of
E179 Guide for Selection of Geometric Conditions for
measurement,
Measurement of Reflection and Transmission Properties
1.2.3 Standards used,
of Materials
1.2.4 Data acquisition procedure, and
E259 Practice for Preparation of Pressed Powder White
1.2.5 Color scales employed.
Reflectance Factor Transfer Standards for Hemispherical
and Bi-Directional Geometries
1.3 This standard does not purport to address all of the
E284 Terminology of Appearance
safety concerns, if any, associated with its use. It is the
E308 PracticeforComputingtheColorsofObjectsbyUsing
responsibility of the user of this standard to establish appro-
the CIE System
priate safety, health, and environmental practices and deter-
E313 Practice for Calculating Yellowness and Whiteness
mine the applicability of regulatory limitations prior to use.
Indices from Instrumentally Measured Color Coordinates
1.4 This international standard was developed in accor-
E991 Practice for Color Measurement of Fluorescent Speci-
dance with internationally recognized principles on standard-
mens Using the One-Monochromator Method
ization established in the Decision on Principles for the
E1164 PracticeforObtainingSpectrometricDataforObject-
Development of International Standards, Guides and Recom-
Color Evaluation
mendations issued by the World Trade Organization Technical
E1247 Practice for Detecting Fluorescence in Object-Color
Barriers to Trade (TBT) Committee.
Specimens by Spectrophotometry
E1331 Test Method for Reflectance Factor and Color by
1
This practice is under the jurisdiction of ASTM Committee E12 on Color and
Appearance and is the direct responsibility of Subcommittee E12.04 on Color and
2
Appearance Analysis. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved June 1, 2022. Published June 2022. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1981. Last previous edition approved in 2017 as E805 – 12a (2017). Standards volume information, refer to the standard’s Document Summary page on
DOI: 10.1520/E0805-22. the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E805 − 22
Spectrophotometry Using Hemispherical Geometry instrumental data to final results. Once the measurements have
E1345 Practice for Reducing the Effect of Variability of been made, it is essential to document what has been done for
Color Measurement by Use of Multiple Measur
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E805 − 12a (Reapproved 2017) E805 − 22
Standard Practice for
Identification of Instrumental Methods of Color or Color-
1
Difference Measurement of Materials
This standard is issued under the fixed designation E805; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This practice covers the documentation of instrumental measurement of color or color difference for current communication
or for future reference. The practice is applicable to instrumental measurements of materials where color is seen by reflected,
transmitted or emitted light and any combinations of one or more of these processes. The practice is recommended for
documentation of methodology in interlaboratory color-measurement programs.
1.2 Providing an adequate identification of an instrumental measure of color or color-difference involves documenting the
metadata necessary for archiving and future use of the measurement data collected. The metadata can be divided in five parts:
1.2.1 Nature and source of available samples and the form of specimens actually measured,
1.2.2 Instrumental conditions of measurement, including instrument geometrical and spectral conditions of measurement,
1.2.3 Standards used,
1.2.4 Data acquisition procedure, and
1.2.5 Color scales employed.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of whoever uses the user of this standard to consult and establish appropriate safety safety, health, and healthenvironmental
practices and determine the applicability of regulatory limitations prior to use.
1.4 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2
2.1 ASTM Standards:
D156 Test Method for Saybolt Color of Petroleum Products (Saybolt Chromometer Method)
D1209 Test Method for Color of Clear Liquids (Platinum-Cobalt Scale)
D1500 Test Method for ASTM Color of Petroleum Products (ASTM Color Scale)
1
This practice is under the jurisdiction of ASTM Committee E12 on Color and Appearance and is the direct responsibility of Subcommittee E12.04 on Color and
Appearance Analysis.
Current edition approved May 1, 2017June 1, 2022. Published May 2017June 2022. Originally approved in 1981. Last previous edition approved in 20122017 as
E805 – 12a.E805 – 12a (2017). DOI: 10.1520/E0805-12AR17.10.1520/E0805-22.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E805 − 22
D1535 Practice for Specifying Color by the Munsell System
D1544 Test Method for Color of Transparent Liquids (Gardner Color Scale)
D2244 Practice for Calculation of Color Tolerances and Color Differences from Instrumentally Measured Color Coordinates
D5386 Test Method for Color of Liquids Using Tristimulus Colorimetry
D6166 Test Method for Color of Pine Chemicals and Related Products (Instrumental Determination of Gardner Color)
E179 Guide for Selection of Geometric Conditions for Measurement of Reflection and Transmission Properties of Materials
E259 Practice for Preparation of Pressed Powder White Reflectance Factor Transfer Standards for Hemispherical and
Bi-Directional Geometries
E284 Terminology of Appearance
E308 Practice for Computing the Colors of Objects by Using the CIE System
E313 Practice for Calculating Yellowness and Whiteness Indices from Instrumentally Measured Color Coordinates
E991 Practice for Color Measurement of Fluorescent Specimens Using the One-Monochromator Method
E1164 Practice for Obtaining Spectrometric Data for Object-Color Evaluation
E1247 Practice for Detecting Fluorescence in Object-Color Specimens by Spectrophotometry
E1331 Test Method for Reflectance Factor and Color by Spectrophotometry Using Hemispheri
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.