Standard Practice for Testing Variable-Wavelength Photometric Detectors Used in Liquid Chromatography

SIGNIFICANCE AND USE
Although it is possible to observe and measure each of the several characteristics of a detector under different and unique conditions, it is the intent of this practice that a complete set of detector specifications should be obtained under the same operating conditions. It should also be noted that to completely specify a detector’ capability, its performance should be measured at several sets of conditions within the useful range of the detector. The terms and tests described in this practice are sufficiently general that they may be used regardless of the ultimate operating parameters.
Linearity and response time of the recorder or other readout device used should be such that they do not distort or otherwise interfere with the performance of the detector. This requires adjusting the gain, damping, and calibration in accordance with the manufacturer’directions. If additional electronic filters or amplifiers are used between the detector and the final readout device, their characteristics should also first be established.
SCOPE
1.1 This practice covers the testing of the performance of a variable-wavelength photometric detector (VWPD) used as the detection component of a liquid-chromatographic (LC) system operating at one or more wavelengths in the range 190 to 800 nm. Many of the measurements are made at 254 nm for consistency with Practice E 685. Measurements at other wavelengths are optional.
1.2 This practice is intended to describe the performance of the detector both independently of the chromatographic system (static conditions) and with flowing solvent (dynamic conditions).
1.3 For general liquid chromatographic procedures, consult Refs (1-9).
1.4 For general information concerning the principles, construction, operation, and evaluation of liquid-chromatography detectors, see Refs (10, 11)in addition to the sections devoted to detectors in Refs  (1-7).
1.5 The values stated in SI units are to be regarded as standard.
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
28-Feb-2006
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E1657-98(2006) - Standard Practice for Testing Variable-Wavelength Photometric Detectors Used in Liquid Chromatography
English language
8 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E1657–98(Reapproved2006)
Standard Practice for
Testing Variable-Wavelength Photometric Detectors Used in
Liquid Chromatography
This standard is issued under the fixed designation E1657; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope E685 Practice for Testing Fixed-Wavelength Photometric
Detectors Used in Liquid Chromatography
1.1 This practice covers the testing of the performance of a
variable-wavelength photometric detector (VWPD) used as the
3. Terminology
detection component of a liquid-chromatographic (LC) system
3.1 Definitions:
operating at one or more wavelengths in the range 190 to 800
3.1.1 absorbance calibration—the procedure that verifies
nm. Many of the measurements are made at 254 nm for
that the absorbance scale is correct within 65%.
consistency with Practice E685. Measurements at other wave-
3.1.2 drift—the average slope of the noise envelope ex-
lengths are optional.
pressed in absorbance units per hour (AU/h) as measured over
1.2 This practice is intended to describe the performance of
a period of 1 h.
the detector both independently of the chromatographic system
3.1.3 dynamic—under conditions of a flow rate of 1.0
(static conditions) and with flowing solvent (dynamic condi-
mL/min.
tions).
3.1.4 linear range—of a VWPD,therangeofconcentrations
1.3 For general liquid chromatographic procedures, consult
of a test substance in a test solvent over which the ratio of
Refs (1-9).
response of the detector versus concentration of test substance
1.4 For general information concerning the principles, con-
is constant to within 5 % as determined from the linearity plot
struction, operation, and evaluation of liquid-chromatography
specified in 7.1.2 and illustrated in Fig. 1. The linear range
detectors, see Refs (10, 11) in addition to the sections devoted
should be expressed as the ratio of the upper limit of linearity
to detectors in Refs (1-7).
obtained from the plot to either a) the lower linear concentra-
1.5 The values stated in SI units are to be regarded as
tion, or b) the minimum detectable concentration, if the
standard.
minimum detectable concentration is greater than the lower
1.6 This standard does not purport to address all of the
linear concentration.
safety concerns, if any, associated with its use. It is the
3.1.5 long-term noise—the maximum amplitude in AU for
responsibility of the user of this standard to establish appro-
all random variations of the detector signal of frequencies
priate safety and health practices and determine the applica-
between 6 and 60 cycles per hour (0.1 and 1.0 cycles per min).
bility of regulatory limitations prior to use.
3.1.5.1 Discussion—Itrepresentsnoisethatcanbemistaken
2. Referenced Documents for a late-eluting peak. This noise corresponds to the observed
3 noise only and may not always be present.
2.1 ASTM Standards:
3.1.6 minimum detectability— of a VWPD, that concentra-
tion of a specific solute in a specific solvent that results in a
detector response corresponding to twice the static short-term
This practice is under the jurisdiction of ASTM Committee E13 on Molecular
noise.
Spectroscopy and is the direct responsibility of Subcommittee E13.19 on Chroma-
tography.
3.1.6.1 Discussion—The static short-term noise is a mea-
Current edition approved March 1, 2006. Published March 2006. Originally
surement of peak-to-peak noise.Astatistical approach to noise
approved in 1994. Last previous edition approved in 2001 as E1657 – 98 (2001).
suggests that a value of three times the rms (root-mean-square)
DOI: 10.1520/E1657-98R06.
noise would insure that any value outside this range would not
The boldface numbers in parentheses refer to the list of references at the end of
this practice.
be noise with a confidence level of greater than 99 %. Since
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
peak-to-peak noise is approximately five times the rms noise
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
(12), the minimum detectability defined in this practice is a
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. more conservative estimate.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
E1657–98 (2006)
mance should be measured at several sets of conditions within
the useful range of the detector. The terms and tests described
in this practice are sufficiently general that they may be used
regardless of the ultimate operating parameters.
4.2 Linearity and response time of the recorder or other
readout device used should be such that they do not distort or
otherwise interfere with the performance of the detector. This
requires adjusting the gain, damping, and calibration in accor-
dance with the manufacturer’s directions. If additional elec-
tronicfiltersoramplifiersareusedbetweenthedetectorandthe
final readout device, their characteristics should also first be
established.
5. Noise and Drift
5.1 Test Conditions—Pure, degassed methanol shall be
used in the sample cell. Air or nitrogen shall be used in the
reference cell if there is one. Nitrogen is preferred where the
presenceofhigh-voltageequipmentmakesitlikelythatthereis
ozone in the air. Protect the entire system from temperature
fluctuations because these will lead to detectable drift.
FIG. 1 Example of Linearity Plot for a Variable-Wavelength
5.1.1 The detector should be located at the test site and
Detector
turned on at least 24 h before the start of testing. Insufficient
warm-upmayresultindriftinexcessoftheactualvalueforthe
3.1.7 response time (speed of output)— the detector, the
detector. The detector wavelength should be set to 254 nm.
time required for the detector output to change from 10 % to
5.2 Methods of Measurement:
90 % of the new equilibrium value when the composition of
5.2.1 Connect a suitable device (see Note 1) between the
the mobile phase is changed in a stepwise manner, within the
pump and the detector to provide at least 75 kPa (500 psi) back
linear range of the detector.
pressure at 1.0 mL/min flow of methanol. Connect a short
3.1.7.1 Discussion—Because the detector volume is very
length(about100mm)of0.25-mm(0.01-in.)internal-diameter
small and the transport rate is not diffusion dependent, the
stainless steel tubing to the outlet tube of the detector to retard
response time is generally fast enough to be unimportant. It is
bubble formation. Connect the recorder to the proper detector
generally comparable to the response time of the recorder and
output channels.
dependent on the response time of the detector electrometer
NOTE 1—Suggested devices include (a)2to4mof 0.1-mm (0.004-in.)
and on the recorder amplifier. Factors that affect the observed
internal-diameter stainless steel tubing, ( b) about 250 mm of 0.25 to 0.5
response time include the true detector response time, elec-
mm(0.01to0.02-in.)internal-diameterstainlesssteeltubingcrimpedwith
tronic filtering, and system band-broadening.
pliersorcutters,or( c)aconstantback-pressurevalvelocatedbetweenthe
3.1.8 short-term noise—the maximum amplitude, peak to
pump and the injector.
peak, inAU for all random variations of the detector signal of
5.2.2 Repeatedly rinse the reservoir and chromatographic
a frequency greater than one cycle per minute.
system, including the detector, with degassed methanol to
3.1.8.1 Discussion—It determines the smallest signal de-
remove from the system all other solvents, any soluble mate-
tectable by a VWPD, limits the precision attainable in quanti-
rial, and any entrained gasses. Fill the reservoir with methanol
tation of trace-level samples, and sets the lower limit on
and pump this solvent through the system for at least 30 min to
linearity. This noise corresponds to the observed noise only.
complete the system cleanup.
3.1.9 static—under conditions of no flow.
5.2.3 Air or nitrogen is used in the reference cell, if any.
3.1.10 wavelength accuracy—the deviation of the observed
Ensure that the cell is clean, free of dust, and completely dry.
wavelength maximum from the maximum of a known test
5.2.4 To perform the static test, cease pumping and allow
substance.
the chromatographic system to stabilize for at least1hat room
3.1.11 wavelength precision—a measure of the ability of a
temperature without flow. Set the attenuator at maximum
VWPD to return to the same spectral position as measured by
sensitivity (lowest attenuation), that is, the setting for the
the reproducibility of absorbance values when the detector is
smallest value of absorbance units full-scale (AUFS). Adjust
reset to a wavelength maximum of a known test substance.
the response time as close as possible to 2 s for a VWPD that
has a variable response time (see Note 2). Record the response
4. Significance and Use
time used. Adjust the detector output to near midscale on the
4.1 Although it is possible to observe and measure each of
readout device. Record at least1hof detector signal under
the several characteristics of a detector under different and
unique conditions, it is the intent of this practice that a
complete set of detector specifications should be obtained
under the same operating conditions. It should also be noted
Distilled-in-glass or liquid-chromatography grade. Complete freedom from
that to completely specify a detector’s capability, its perfor- particles may require filtration, for example, through a 0.45-µm membrane filter.
E1657–98 (2006)
these conditions, during which time the ambient temperature distance between them. Measure the vertical distance, in AU,
should not change by more than 2°C. between the lines. Calculate the average value over all the
segments. Divide this value by the cell length in centimeters to
NOTE 2—Time constant is converted to response time by multiplying
obtain the static short-term noise.
by the factor 2.2. The effect of electronic filtering on observed noise may
be studied by repeating the noise measurements for a series of response-
5.2.6 Now mark the center of each segment over the 15-min
time settings.
period of the static short-term noise measurement. Draw a
5.2.5 Draw pairs of parallel lines, each pair corresponding
series of parallel lines encompassing these centers, each pair
to between 0.5 and 1 min in length, to form an envelope of all
correspondingto10mininlength,andchoosethatpairoflines
observed random variations over any 15-min period (see Fig.
whose vertical distance apart is greatest (see Fig. 2). Divide
2). Draw the parallel lines in such a way as to minimize the
FIG. 2 Example for the Measurement of the Noise and Drift of a VWD (Chart Recorder Output)
E1657–98 (2006)
6.2 Method of Measurement—Wavelength Accuracy—For
thedeterminationofthe wavelength accuracy of a VWPD, (13)
a solution of a compound with known absorbance maxima is
introduced into the cell. The measured maxima are compared
to the known maxima for the compound. There are several
acceptable compounds and solvents. The following procedure
is recommended (Note 3).
NOTE 3—The recommended procedure is covered under U.S. Patent
4,836,673. The American Society for Testing and Materials takes no
position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are
expressly advised that determination of the validity of any such patent
rights, and the risk of infringement of such rights, are entirely their own
responsibility. Alternative procedures will be considered.
6.2.1 Prepare the test solution. For example, dissolve2gof
6 7
erbiumperchloratehexahydrate in25mLwater. Thenominal
concentration is 0.14 M. Filter the solution with an appropriate
filter to ensure the sample is free of particles.
FIG. 3 Example of Wavelength Accuracy Test Plot
NOTE 4—This can be conveniently done by addng water toa2g vial of
this distance in AU by the cell length in centimeters to obtain
erbiumperchloratehexahydratetodissolvethesolid.Transferthecontents
the static long-term noise.
to a 25 mL volumetric flask and make up to volume with water. While
5.2.7 Draw the pair of parallel lines that minimizes the reasonable care should be observed in transferring the dissolved erbium
perchlorate into the volumetric flask, the final solution is not used
vertical distance separating these lines over the 1 h of mea-
quantitatively.
surement (Fig. 2). The slope of either line is the static drift
expressed in AU/h. 6.2.2 Turn on the detector and allow it to warm up accord-
5.2.8 Set the pump to deliver 1.0 mL/min under the same
ing to the manufacturer’s recommendations. Thoroughly flush
conditionsoftubing,solvent,andtemperatureasin5.2.1-5.2.3. the detector cell with water preferably from the same source as
Allow 15 min for the system to stabilize. Record at least1hof
that to make up the test solution. (If using another test
signal under these flowing conditions, during which time the compound,besuretousethesamesolventasthetestsolution.)
ambient temperature should not change by more than 2°C.
Set the detector wavelength to 250 nm. Zero the absorbance of
5.2.9 Draw pairs of parallel lines, measure the vertical the detector. (Some detectors will automatically zero the
distances, and calculate the dynamic short-term noise follow-
detector after changing wavelengths.) Flush the cell with at
ing the procedure of 5.2.5.
least 1 mL of the erbium test solution. Record the absorbance
5.2.10 Make the measurement for the dynamic long-term
reading. Increase the wavelength by 1 nm. Flush the cell with
noise following the procedure outlined in 5.2.6.
at least 1 mL of water. Zero the absorbance of the detector.
5.2.11 Draw the pair of parallel lines as directed in 5.2.7. Flush the cell with the erbium test solution and record the
The slope of these lines is the dynamic drift.
absorbance. Repeat the procedure in 0.5 to 1.0 nm increments
5.2.12 The actual noise of the system may be larger or until reaching 260 nm.
smaller than the observed values, depending upon the method
6.2.3 Plot absorbance versus wavelength and determine the
of data collection, or signal monitoring of the detector, since maximum absorbance. (See Fig. 3) Compare the calculated
observed noise is a function of the frequency, speed of
maximum to the maximum for erbium perchlorate of 255 nm
response, and bandwidth of the readout device. (see Note 4). Report the nominal and calculat
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.