Standard Test Method for Pour Point of Petroleum Products (Robotic Tilt Method)

SIGNIFICANCE AND USE
5.1 The pour point of a petroleum product is an index of the lowest temperature of its utility for certain applications. Flow characteristics, such as pour point, can be critical for the correct operation of lubricating systems, fuel systems, and pipeline operations.
FIG. 1 Schematic of Cooling/Heating Block and Cooling Circulating Bath  
5.2 Petroleum blending operations require precise measurement of the pour point.  
5.3 Test results from this test method can be determined at either 1 or 3°C intervals.  
5.4 This test method yields a pour point in a format similar to Test Method D97 or IP 15, when the 3°C interval results are reported.Note 2—Since some users may wish to report their results in a format similar to Test Method D97 or IP 15 (in 3°C intervals) the precision data were derived for the 3°C intervals. For statements on bias relative to Test Method D97 or IP 15, see the research report.  
5.5 This test method has comparable repeatability and better reproducibility relative to Test Method D97 or IP 15 as measured in the 1998 interlaboratory program (see Section 13).
SCOPE
1.1 This test method covers the determination of the pour point of petroleum products by an automatic instrument that tilts the test jar to detect movement of the surface of the test specimen with an optical device, after being removed from a regulated, stepped-bath cooling jacket.  
1.2 This test method is designed to cover the range of temperatures from −57 to +51°C; however, the range of temperatures included in the 1998 interlaboratory test program only covered the temperature range from −51 to −11°C.  
1.3 Test results from this test method can be determined at either 1 or 3°C testing intervals.  
1.4 This test method is not intended for use with crude oils.Note 1—The applicability of this test method on residual fuel samples has not been verified. For further information on the applicability, refer to 13.4.  
1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
31-Dec-2013
Current Stage
Ref Project

Relations

Effective Date
01-Jan-2014

Buy Standard

Standard
ASTM D6892-03(2014) - Standard Test Method for Pour Point of Petroleum Products (Robotic Tilt Method)
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: D6892 − 03 (Reapproved 2014)
Standard Test Method for
Pour Point of Petroleum Products (Robotic Tilt Method)
This standard is issued under the fixed designation D6892; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2.2 Energy Institute Standard:
IP15Test Method for Pour Point of Petroleum Products
1.1 This test method covers the determination of the pour
point of petroleum products by an automatic instrument that
3. Terminology
tilts the test jar to detect movement of the surface of the test
specimen with an optical device, after being removed from a 3.1 Definitions:
regulated, stepped-bath cooling jacket. 3.1.1 pour point, n—in petroleum products, the lowest
temperature at which movement of the test specimen is
1.2 This test method is designed to cover the range of
observed under prescribed conditions of test.
temperatures from −57 to +51°C; however, the range of
temperatures included in the 1998 interlaboratory test program 3.2 Definitions of Terms Specific to This Standard:
only covered the temperature range from −51 to −11°C. 3.2.1 no-flow point, n— in petroleum products,thetempera-
ture of the test specimen at which a wax crystal structure or
1.3 Test results from this test method can be determined at
viscosity increase, or both, impedes movement of the surface
either 1 or 3°C testing intervals.
of the test specimen under the conditions of the test.
1.4 This test method is not intended for use with crude oils.
3.2.1.1 Discussion—The no-flow point occurs when, upon
NOTE 1—The applicability of this test method on residual fuel samples
cooling, the formation of wax crystal structures or viscosity
has not been verified. For further information on the applicability, refer to
increase,orboth,hasprogressedtothepointwheretheapplied
13.4.
observation device no longer detects movement under the
1.5 The values stated in SI units are to be regarded as
conditions of the test. The preceding observation temperature
standard. No other units of measurement are included in this
at which flow of the test specimen is last observed is the pour
standard.
point.
1.6 This standard does not purport to address all of the
3.2.2 tilting, vt—technique of movement where the test jar
safety concerns, if any, associated with its use. It is the
in a vertical position is moved towards a horizontal position to
responsibility of the user of this standard to establish appro-
induce specimen movement.
priate safety and health practices and determine the applica-
3.2.2.1 Discussion—When the test jar is tilted and held in a
bility of regulatory limitations prior to use.
horizontal position for 5 s without detection of movement of
the surface of the specimen, this is the no-flow point and the
2. Referenced Documents
test is complete.
2.1 ASTM Standards:
D97Test Method for Pour Point of Petroleum Products 4. Summary of Test Method
D4057Practice for Manual Sampling of Petroleum and
4.1 After insertion of the specimen into the automatic pour
Petroleum Products
point apparatus and initiation of the testing program, the
D4177Practice for Automatic Sampling of Petroleum and
specimen is heated and then cooled according to a prescribed
Petroleum Products
profile. The specimen surface is examined periodically for
movement using an optical camera system mounted on top of
the specimen test jar, while tilting the specimen test jar. The
This test method is under the jurisdiction of ASTM Committee D02 on test jar is removed from the jacketed cooling chamber prior to
Petroleum Products, Liquid Fuels, and Lubricants and is the direct responsibility of
eachexamination.Thelowesttemperature,whenmovementof
Subcommittee D02.07 on Flow Properties.
the surface of the specimen is detected, is recorded as the pour
Current edition approved Jan. 1, 2014. Published February 2014. Originally
point determined by this Test Method D6892.
approved in 2003. Last previous edition approved in 2008 as D6892–03(2008).
DOI: 10.1520/D6892-03R14.
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on Available from Energy Institute, 61 New Cavendish St., London, WIG 7AR,
the ASTM website. U.K., http://www.energyinst.org.uk.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
D6892 − 03 (2014)
FIG. 1 Schematic of Cooling/Heating Block and Cooling Circulating Bath
5. Significance and Use
5.1 Thepourpointofapetroleumproductisanindexofthe
lowest temperature of its utility for certain applications. Flow
characteristics, such as pour point, can be critical for the
correct operation of lubricating systems, fuel systems, and
pipeline operations.
5.2 Petroleum blending operations require precise measure-
ment of the pour point.
5.3 Test results from this test method can be determined at
either 1 or 3°C intervals.
5.4 This test method yields a pour point in a format similar
toTest Method D97 or IP15, when the 3°C interval results are
reported.
NOTE 2—Since some users may wish to report their results in a format
similar to Test Method D97 or IP15 (in 3°C intervals) the precision data
were derived for the 3°C intervals. For statements on bias relative to Test
Method D97 or IP15, see the research report.
FIG. 2 Picture of Apparatus
5.5 Thistestmethodhascomparablerepeatabilityandbetter
reproducibility relative to Test Method D97 or IP15 as
interface, cooling/heating block assembly with cylindrical
measuredinthe1998interlaboratoryprogram(seeSection13).
jacket with an inside diameter of 44.2 to 45.8 mm, and about
6. Apparatus
115 mm in depth to accept the test jar) robotic mechanisms for
lifting,tilting,replacingthetestjar,opticalcamerasystem,and
6.1 Automatic Apparatus —Theautomaticpourpointappa-
a temperature measuring device.
ratus described in this test method (see Fig. 2) consists of a
microprocessor controlled measuring unit that is capable of
6.2 Test Jar—Clear, cylindrical glass, flat bottom
heatingthespecimentoprogrammedtemperatures,coolingthe
(darkened), 31.5 6 0.5 mm inside diameter and 120 62mm
specimen according to programmed cooling profiles, mechani-
height with a wall thickness of 1.25 6 0.25 mm. The jar shall
cally manipulating the test jar according to the programmed
be marked with a line to indicate sample filling height
test procedure, while optically observing the surface of the
corresponding to 45 6 0.5 mL.
specimen for movement, using a camera system mounted on
6.3 Temperature Probe—Capableofmeasurementfrom+70
top of the specimen test jar and recording the temperature of
to −80°C with a resolution of 0.1°C. The temperature probe
the specimen. The apparatus shall be equipped with a user
shall be suspended in the center axis of the test jar and the top
ofthetemperaturesensingzoneimmersedbelowthesurfaceof
the specimen.
The sole source of supply of the Herzog Model MP852 or HCP852 known to
the committee at this time is Walter Herzog, Lauda, Germany. If you are aware of
6.4 Circulating Bath—Refrigeration unit, equipped with a
alternative suppliers, please provide this information to ASTM International
circulating pump, capable of maintaining the liquid cooling
Headquarters.Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. medium at a temperature at least 20°C lower than the lowest
D6892 − 03 (2014)
10. Calibration and Verification
10.1 Ensure that all of the manufacturer’s instructions for
calibrating, checking, and operating the apparatus are fol-
lowed.
10.2 A sample with a well-documented pour point can be
used to verify the performance of the automatic apparatus.
Alternatively, a sample which has been extensively tested in a
pourpointcross-checkprogramcanbeused.Suchverification
materials can also be prepared from intra-company cross
checks.
11. Procedure
11.1 Fillthetestjaruptothemarkedlinewiththespecimen.
When necessary, heat the sample in a water bath or oven until
FIG. 3 Assembled Apparatus
it is just sufficiently fluid to pour into the test jar.
NOTE 3—Residual fuels have been known to be sensitive to thermal
history. In the case where a residual fuel sample is tested, refer to Test
Method D97 for sample treatment.
expected pour point to be measured. The circulating bath is
11.2 Insert the test jar into the apparatus and start the test in
connected to the automatic apparatus through suitable means accordance with the manufacturer’s instructions.
for supplying the liquid cooling medium.
11.3 When the expected pour point is known to be above
−33°C, preselect a starting temperature which is at least 9°C
7. Reagents and Materials
above the expected pour point, but to at least 45°C.
7.1 Bath Cooling Medium—Suitable for use in the circulat-
11.4 When the expected pour point is known to be at or
ing bath (an example is methyl alcohol-anhydrous).
below −33°C, preselect a starting temperature of 45°C.
(Warning— Flammable. Liquid causes eye burns. Vapor
11.5 When the expected pour point is not known, preselect
harmful. May be fatal or cause blindness if swallowed or
a starting temperature of 45°C. When the expected pour point
inhaled.)
is not known and the sample must be heated to allow transfer
7.2 Cleaning Solvents—Suitableforcleaninganddryingthe
into the test jar, preselect a starting temperature corresponding
test jar and temperature measuring device, such as; petroleum
to the preheat temperature. (Warning— Exercise care when
naphthaoracetone.(Warning—Flammable.Liquidcauseseye
selectingstar
...