Standard Test Method for Transmittance and Color by Spectrophotometry Using Hemispherical Geometry

SIGNIFICANCE AND USE
5.1 The most direct and accessible methods for obtaining the color coordinates of object colors are by instrumental measurement using spectrophotometers or colorimeters with either hemispherical or bidirectional optical measuring systems. This test method provides procedures for such measurement by transmittance spectrophotometry using a hemispherical optical measuring system.  
5.2 This test method is especially suitable for measurement of the following types of specimens (see also Guide E179 and Practice E805):  
5.2.1 Fully transparent specimens (free from turbidity, haze, or translucency), and  
5.2.2 Translucent or hazy specimens, provided that the specimen can be placed flush against the transmission port of the integrating sphere.  
5.3 This test method is not recommended for measurement of retroreflective transparent or translucent specimens, or samples that are fluorescent.
SCOPE
1.1 This test method describes the instrumental measurement of the transmission properties and color of object-color specimens by the use of a spectrophotometer or spectrocolorimeter with a hemispherical optical measuring system, such as an integrating sphere.  
1.2 This test method is generally suitable for all fully transparent specimens without regard for the specimen position relative to the transmission port of the instrument. Translucent specimens, however, must be placed flush against the transmission port of the sphere.  
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
30-Sep-2022
Current Stage
Ref Project

Buy Standard

Standard
ASTM E1348-22 - Standard Test Method for Transmittance and Color by Spectrophotometry Using Hemispherical Geometry
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E1348-22 - Standard Test Method for Transmittance and Color by Spectrophotometry Using Hemispherical Geometry
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation: E1348 − 22
Standard Test Method for
Transmittance and Color by Spectrophotometry Using
1
Hemispherical Geometry
This standard is issued under the fixed designation E1348; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope Color Differences from Instrumentally Measured Color
Coordinates
1.1 This test method describes the instrumental measure-
E179Guide for Selection of Geometric Conditions for
ment of the transmission properties and color of object-color
Measurement of Reflection and Transmission Properties
specimens by the use of a spectrophotometer or spectrocolo-
of Materials
rimeterwithahemisphericalopticalmeasuringsystem,suchas
E284Terminology of Appearance
an integrating sphere.
E308PracticeforComputingtheColorsofObjectsbyUsing
1.2 This test method is generally suitable for all fully
the CIE System
transparentspecimenswithoutregardforthespecimenposition
E805Practice for Identification of Instrumental Methods of
relative to the transmission port of the instrument. Translucent
Color or Color-Difference Measurement of Materials
specimens, however, must be placed flush against the trans-
E1164PracticeforObtainingSpectrometricDataforObject-
mission port of the sphere.
Color Evaluation
1.3 The values stated in SI units are to be regarded as
standard. No other units of measurement are included in this 3. Terminology
standard.
3.1 Definitions:
1.4 This standard does not purport to address all of the
3.1.1 The definitions in Guide E179, Terminology E284,
safety concerns, if any, associated with its use. It is the
and Practice E1164 are applicable to this test method.
responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter-
4. Summary of Test Method
mine the applicability of regulatory limitations prior to use.
4.1 Thistestmethodprovidesaprocedureformeasuringthe
1.5 This international standard was developed in accor-
transmittanceoftransmittingspecimensbyusingaspectropho-
dance with internationally recognized principles on standard-
tometer or spectrocolorimeter equipped with a hemispherical
ization established in the Decision on Principles for the
optical measuring system such as an integrating sphere.
Development of International Standards, Guides and Recom-
mendations issued by the World Trade Organization Technical
4.2 This test method includes procedures for calibrating the
Barriers to Trade (TBT) Committee.
instrument and for selecting specimens suitable for precision
measurement.
2. Referenced Documents
4.3 This test method is satisfactory for all fully transparent
2
2.1 ASTM Standards:
specimenswithplaneandparallelsurfaces.Whenpossible,the
D1003Test Method for Haze and Luminous Transmittance
user should select the position of the specimen in the trans-
of Transparent Plastics
mission compartment to provide either essentially total or
D2244Practice for Calculation of Color Tolerances and
essentially regular transmittance, depending on the end use of
the measured data.
4.4 Accurate measurement of translucent specimens re-
1
This test method is under the jurisdiction of ASTM Committee E12 on Color
quires that the specimen be placed flush against the transmis-
and Appearance and is the direct responsibility of Subcommittee E12.02 on
Spectrophotometry and Colorimetry. sion port of the integrating sphere (see Practice E1164, 8.1.4).
Current edition approved Oct. 1, 2022. Published November 2022. Originally
ɛ1
4.5 Most modern spectrophotometers have the capacity to
approved in 1990. Last previous edition approved in 2015 as E1348 – 15 . DOI:
10.1520/E1348-22.
compute the color coordinates of the specimen during the
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
measurement. When this is the case, the user of this test
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
method must select the color scale, illuminant and observer
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. (see the Procedure section in Practice E308, Section 6).
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1348 − 22
reflectance as that of the lining of the integrating sphere.
5. Significance and Use
8.2 Verifytheaccuracyofthemeasuredresultsbymeasure-
5.1 The most direct and acc
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
´1
Designation: E1348 − 15 E1348 − 22
Standard Test Method for
Transmittance and Color by Spectrophotometry Using
1
Hemispherical Geometry
This standard is issued under the fixed designation E1348; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1
ε NOTE—Equation 1 was editorially corrected in September 2016.
1. Scope
1.1 This test method describes the instrumental measurement of the transmission properties and color of object-color specimens
by the use of a spectrophotometer or spectrocolorimeter with a hemispherical optical measuring system, such as an integrating
sphere.
1.2 This test method is generally suitable for all fully transparent specimens without regard for the specimen position relative to
the transmission port of the instrument. Translucent specimens, however, must be placed flush against the transmission port of the
sphere.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2
2.1 ASTM Standards:
D1003 Test Method for Haze and Luminous Transmittance of Transparent Plastics
D2244 Practice for Calculation of Color Tolerances and Color Differences from Instrumentally Measured Color Coordinates
E179 Guide for Selection of Geometric Conditions for Measurement of Reflection and Transmission Properties of Materials
E284 Terminology of Appearance
E308 Practice for Computing the Colors of Objects by Using the CIE System
E805 Practice for Identification of Instrumental Methods of Color or Color-Difference Measurement of Materials
E1164 Practice for Obtaining Spectrometric Data for Object-Color Evaluation
1
This test method is under the jurisdiction of ASTM Committee E12 on Color and Appearance and is the direct responsibility of Subcommittee E12.02 on
Spectrophotometry and Colorimetry.
Current edition approved Nov. 1, 2015Oct. 1, 2022. Published November 2015November 2022. Originally approved in 1990. Last previous edition approved in 20112015
ɛ1
as E1348 – 1115 . DOI: 10.1520/E1348-15E01.10.1520/E1348-22.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1348 − 22
3. Terminology
3.1 Definitions:
3.1.1 The definitions in Guide E179, Terminology E284, and Practice E1164 are applicable to this test method.
4. Summary of Test Method
4.1 This test method provides a procedure for measuring the transmittance of transmitting specimens by using a spectrophotometer
or spectrocolorimeter equipped with a hemispherical optical measuring system such as an integrating sphere.
4.2 This test method includes procedures for calibrating the instrument and for selecting specimens suitable for precision
measurement.
4.3 This test method is satisfactory for all fully transparent specimens with plane and parallel surfaces. When possible, the user
should select the position of the specimen in the transmission compartment to provide either essentially total or essentially regular
transmittance, depending on the end use of the measured data.
4.4 Accurate measurement of translucent specimens requires that the specimen be placed flush against the transmission port of the
integrating sphere (see Practice E1164, 8.1.4).
4.5 Most modern spectrophotometers have the capacity to compute the color coordinates of the specimen during the measurement.
When this is the case,
...