Standard Test Method for Temperature Calibration of Thermomechanical Analyzers

SIGNIFICANCE AND USE
Thermomechanical analyzers are employed in their various modes of operation (penetration, expansion, flexure, etc.) to characterize a wide range of materials. In most cases, the value to be assigned in thermomechanical measurements is the temperature of the transition (or event) under study. Therefore, the temperature axis (abscissa) of all TMA thermal curves must be accurately calibrated either by direct reading of a temperature sensor or by adjusting the programmer temperature to match the actual temperature over the temperature range of interest.
SCOPE
1.1 This test method describes the temperature calibration of thermomechanical analyzers from − 50 to 1100 °C. (See Note 1.)
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard is similar to ISO 11359–1 but addresses a larger temperature range and utilizes additional calibration materials.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 7 and Note 9.

General Information

Status
Historical
Publication Date
31-Aug-2008
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM E1363-08 - Standard Test Method for Temperature Calibration of Thermomechanical Analyzers
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
REDLINE ASTM E1363-08 - Standard Test Method for Temperature Calibration of Thermomechanical Analyzers
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1363 − 08
StandardTest Method for
1
Temperature Calibration of Thermomechanical Analyzers
This standard is issued under the fixed designation E1363; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 4. Summary of Test Method
1.1 This test method describes the temperature calibration 4.1 Anequationisdevelopedforthelinearcorrelationofthe
of thermomechanical analyzers from−50 to 1100 °C. (See experimentally observed program temperature and the actual
Note 1.)
melting temperature for known melting standards. This is
accomplished through the use of a thermomechanical analyzer
1.2 The values stated in SI units are to be regarded as
with a penetration probe to obtain the onset temperatures for
standard. No other units of measurement are included in this
twomeltingpointstandards.Analternate,one-pointmethodof
standard.
temperature calibration, is also given for use over very narrow
1.3 This standard is similar to ISO 11359–1 but addresses a
temperature ranges. (See Note 2.)
larger temperature range and utilizes additional calibration
NOTE1—Thistestmethodmaybeusedforcalibratingthermomechani-
materials.
cal analyzers at temperatures outside this range of temperature. However,
1.4 This standard does not purport to address all of the
the accuracy of the calibration will be no better than that of the
safety concerns, if any, associated with its use. It is the
temperature standards used.
NOTE 2—It is possible to develop a more elaborate method of
responsibility of the user of this standard to establish appro-
temperature calibration using multiple (more than two) fusion standards
priate safety and health practices and determine the applica-
and quadratic regression analysis. Since most modern instruments are
bility of regulatory limitations prior to use. Specific precau-
capable of heating rates which are essentially linear in the region of use,
tionary statements are given in Section 7 and Note 9.
the procedure given here is limited to a two-point calibration.
2. Referenced Documents
5. Significance and Use
2
2.1 ASTM Standards:
5.1 Thermomechanical analyzers are employed in their
E473Terminology Relating to Thermal Analysis and Rhe-
various modes of operation (penetration, expansion, flexure,
ology
etc.) to characterize a wide range of materials. In most cases,
the value to be assigned in thermomechanical measurements is
2.2 Other Standards:
the temperature of the transition (or event) under study.
11359–1 Thermomechanical Analysis (TMA)-Part 1: Gen-
3
Therefore, the temperature axis (abscissa) of all TMAthermal
eral Principles
curvesmustbeaccuratelycalibratedeitherbydirectreadingof
a temperature sensor or by adjusting the programmer tempera-
3. Terminology
turetomatchtheactualtemperatureoverthetemperaturerange
3.1 Definitions:
of interest.
3.1.1 The terminology relating to thermal analysis appear-
ing inTerminology E473 shall be considered applicable to this
6. Apparatus
document.
6.1 Thermomechanical Analyzer (TMA), The essential in-
strumentation required to provide the minimum thermome-
chanical analytical or thermodilatometric capability for this
1
ThistestmethodisunderthejurisdictionofASTMCommitteeE37onThermal
method includes:
Measurements and is the direct responsibility of Subcommittee E37.10 on
Fundamental, Statistical and Mechanical Properties. 6.1.1 A Rigid Specimen Holder or Platform, of inert, low
-1 -1
Current edition approved Sept. 1, 2008. Published September 2008. Originally
expansivitymaterial(<1µmm K )tocenterthespecimenin
approved in 1990. Last previous edition approved in 2003 as E1363–03. DOI:
the furnace and to fix the specimen to mechanical ground.
10.1520/E1363-08.
2
6.1.2 A Rigid (expansion compression, flexure, tensile, etc)
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
-1 -1
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Probe, of inert, low expansivity material (< 1 µm m K ) that
Standards volume information, refer to the standard’s Document Summary page on
contacts with the specimen with an applied compressive or
the ASTM website.
3
tensileforce.Forthistestmethodtheuseofapenetrationprobe
Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
4th Floor, New York, NY 10036, http://www.ansi.org. is recommended.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1363 − 08
A
TABLE 1 Recommended Melting Temperature Standards
6.1.3 A Sensing Element—linearoveraminimumrangeof2
Mel
...

This document is not anASTM standard and is intended only to provide the user of anASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation:E1363–03 Designation:E1363–08
Standard Test Method for
1
Temperature Calibration of Thermomechanical Analyzers
This standard is issued under the fixed designation E1363; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 Thistestmethoddescribesthetemperaturecalibrationofthermomechanicalanalyzersfrom−50to1100°C.(SeeNote21.)
1.2Computer or electronic based instruments, techniques, or data treatment equivalent to this test method may be used.
NOTE1—Users of this test method are advised that all such instruments or techniques may not be equivalent. It is the responsibility of the user of this
test method to determine the necessary equivalency prior to use.
1.3 SI units are the standard.
1.4This standard is similar to ISO 11359–1
1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3 This standard is similar to ISO 11359–1 but addresses a larger temperature range and utilizes additional calibration
materials.
1.5
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use. Specific precautionary statements are given in Section 7 and Note 10Note 9.
2. Referenced Documents
2
2.1 ASTM Standards:
E473 Terminology Relating to Thermal Analysis and Rheology
2.2 Other Standards:
3
11359–1 Thermomechanical Analysis (TMA)-Part 1: General Principles
3. Terminology
3.1 Definitions:
3.1.1 The terminology relating to thermal analysis appearing in E473 shall be considered applicable to this document.
4. Summary of Test Method
4.1 An equation is developed for the linear correlation of the experimentally observed program temperature and the actual
melting temperature for known melting standards. This is accomplished through the use of a thermomechanical analyzer with a
penetration probe to obtain the onset temperatures for two melting point standards.An alternate, one-point method of temperature
calibration, is also given for use over very narrow temperature ranges. (See Note 32.)
NOTE2—This 1—This test method may be used for calibrating thermomechanical analyzers at temperatures outside this range of temperature.
However, the accuracy of the calibration will be no better than that of the temperature standards used.
NOTE3—It 2—It is possible to develop a more elaborate method of temperature calibration using multiple (more than two) fusion standards and
quadratic regression analysis. Since most modern instruments are capable of heating rates which are essentially linear in the region of use, the procedure
given here is limited to a two-point calibration.
5. Significance and Use
5.1 Thermomechanical analyzers are employed in their various modes of operation (penetration, expansion, flexure, etc.) to
characterize a wide range of materials. In most cases, the value to be assigned in thermomechanical measurements is the
1
This test method is under the jurisdiction ofASTM Committee E37 onThermal Measurements and is the direct responsibility of Subcommittee E37.01 onThermalTest
Methods and Recommended Practices.
Current edition approved March 10, 2003.Sept. 1, 2008. PublishedApril 2003.September 2008. Originally approved in 1990. Last previous edition approved in 19972003
as E1363–97.E1363–03.
2
ForreferencedASTMstandards,visittheASTMwebsite,www.astm.org,orcontactASTMCustomerServiceatservice@astm.org.For Annual Book of ASTM Standards
, Vol 14.02.volume information, refer to the standard’s Document Summary page on the ASTM website.
3
Available from American National Standards Institute, 11 W. 42nd St., 13th Floor, New York, NY 10036.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
E1363–08
temperature of the transition (or event) under study. Therefore, the temperature axis (abscissa) of all TMA thermal curves must
beaccuratelycalibratedeitherbydirectreadingofathermocoupletemperaturesensororbyadjustingtheprogrammertemperature
to match the actual temperature over the temperature range of interest.
6. Apparatus
6.1 Therm
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.