ASTM E1454-02(2008)
(Guide)Standard Guide for Data Fields for Computerized Transfer of Digital Ultrasonic Testing Data (Withdrawn 2013)
Standard Guide for Data Fields for Computerized Transfer of Digital Ultrasonic Testing Data (Withdrawn 2013)
SIGNIFICANCE AND USE
The primary use of this guide is to provide a standardized approach for the data file to be used for the transfer of digital ultrasonic data from one user to another where the two users are working with dissimilar ultrasonic systems. This guide describes the contents, both required and optional, for an intermediate data file that can be created from the native format of the ultrasonic system on which the data was collected and that can be converted into the native format of the receiving ultrasonic or data analysis system. The development of translator software to accomplish these data format conversions is being addressed under a separate effort; this will include specific items needed for the data transfer, for example, language used, memory requirements and intermediate specification, including detailed data formats and structures. Ths guide will also be useful in the archival storage and retrieval of ultrasonic data as either a data format specifier or as a guide to the data elements that should be included in the archival file.
Although the recommended field listing includes more than 120 items, only about one third of those are regarded as essential and marked with Footnote C in Table 1. Fields so marked must be addressed in the data base. The other recommended fields provide additional information that a user will find helpful in understanding the ultrasonic examination result. These header field items will, in most cases, make up only a very small part of an ultrasonic examination file. The actual stream of ultrasonic data that make up the image will take up the largest part of the data base. Since an ultrasonic image file will normally be large, the concept of data compression will be considered in many cases. Compressed data should be noted, along with a description of the compression method, as indicated in Field No. 122.
This guide describes the structure of a data file for all of the ultrasonic information collected in a single scan. Some systems re...
SCOPE
1.1 This guide provides a listing and description of the fields that are recommended for inclusion in a digital ultrasonic examination data base to facilitate the transfer of such data. This guide is prepared for use particularly with digital image data obtained from ultrasonic scanning systems. The field listing includes those fields regarded as necessary for inclusion in the data base (as indicated by Footnote C in Table 1); these fields, so marked, are regarded as the minimum information necessary for a transfer recipient to understand the data. In addition, other optional fields are listed as a remainder of the types of information that may be useful for additional understanding of the data, or applicable to a limited number of applications.
1.2 It is recognized that organizations may have in place an internal format for the storage and retrieval of ultrasonic examination data. This guide should not impede the use of such formats since it is probable that the necessary fields are already included in such internal data bases, or that the few additions can be made. The numerical listing indicated in this guide is only for convenience; the specific numbers carry no inherent significance and are not a part of the data file.
1.3 The types of ultrasonic examination systems that appear useful in relation to this guide include those described in Practices E 114, E 214 and E 1001. Many of the terms used are defined in Terminology E 1013 and E 1316. The search unit parameters used in this guide follow from those used in Guide E 1065.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
WITHDRAWN RATIONALE
This guide provided a listing and description of the fields that were recommended fo...
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: E1454 − 02(Reapproved 2008)
Standard Guide for
Data Fields for Computerized Transfer of Digital Ultrasonic
Testing Data
This standard is issued under the fixed designation E1454; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 2. Referenced Documents
2.1 ASTM Standards:
1.1 Thisguideprovidesalistinganddescriptionofthefields
E114 Practice for Ultrasonic Pulse-Echo Straight-Beam
that are recommended for inclusion in a digital ultrasonic
Contact Testing
examination data base to facilitate the transfer of such data.
E214 Practice for Immersed Ultrasonic Testing by the Re-
This guide is prepared for use particularly with digital image
flection Method Using Pulsed Longitudinal Waves (With-
data obtained from ultrasonic scanning systems. The field
drawn 2007)
listing includes those fields regarded as necessary for inclusion
E1001 Practice for Detection and Evaluation of Discontinui-
in the data base (as indicated by Footnote C in Table 1); these
ties by the Immersed Pulse-Echo Ultrasonic Method
fields, so marked, are regarded as the minimum information
Using Longitudinal Waves
necessary for a transfer recipient to understand the data. In
E1013 Terminology Relating to Computerized Systems
addition, other optional fields are listed as a remainder of the
(Withdrawn 2000)
types of information that may be useful for additional under-
E1065 Guide for Evaluating Characteristics of Ultrasonic
standing of the data, or applicable to a limited number of
Search Units
applications.
E1316 Terminology for Nondestructive Examinations
1.2 It is recognized that organizations may have in place an
internal format for the storage and retrieval of ultrasonic 3. Significance and Use
examinationdata.Thisguideshouldnotimpedetheuseofsuch
3.1 The primary use of this guide is to provide a standard-
formats since it is probable that the necessary fields are already
ized approach for the data file to be used for the transfer of
included in such internal data bases, or that the few additions
digital ultrasonic data from one user to another where the two
can be made. The numerical listing indicated in this guide is
users are working with dissimilar ultrasonic systems. This
only for convenience; the specific numbers carry no inherent
guide describes the contents, both required and optional, for an
significance and are not a part of the data file.
intermediatedatafilethatcanbecreatedfromthenativeformat
of the ultrasonic system on which the data was collected and
1.3 The types of ultrasonic examination systems that appear
that can be converted into the native format of the receiving
useful in relation to this guide include those described in
ultrasonic or data analysis system. The development of trans-
Practices E114, E214, and E1001. Many of the terms used are
lator software to accomplish these data format conversions is
defined in Terminology E1013 and E1316. The search unit
being addressed under a separate effort; this will include
parameters used in this guide follow from those used in Guide
specific items needed for the data transfer, for example,
E1065.
language used, memory requirements and intermediate
1.4 This standard does not purport to address all of the
specification, including detailed data formats and structures.
safety concerns, if any, associated with its use. It is the
Ths guide will also be useful in the archival storage and
responsibility of the user of this standard to establish appro-
retrieval of ultrasonic data as either a data format specifier or
priate safety and health practices and determine the applica-
as a guide to the data elements that should be included in the
bility of regulatory limitations prior to use.
archival file.
1 2
This guide is under the jurisdiction of ASTM Committee E07 on Nondestruc- For referenced ASTM standards, visit the ASTM website, www.astm.org, or
tive Testing and is the direct responsibility of Subcommittee E07.11 on Digital contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Imaging and Communication in Nondestructive Evaluation (DICONDE). Standards volume information, refer to the standard’s Document Summary page on
Current edition approved July 1, 2008. Published September 2008. Originally the ASTM website.
published in 1992. Last previous edition approved in 2002 as E1454 - 02. DOI: The last approved version of this historical standard is referenced on
10.1520/E1454-02R08. www.astm.org.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
E1454 − 02 (2008)
3.2 Although the recommended field listing includes more pulse echo thickness data may be recorded at the same time.
than 120 items, only about one third of those are regarded as These data may be stored in separate image planes; see Field
essential and marked with Footnote C in Table 1. Fields so
No. 102. In other systems, complete digitized waveforms may
marked must be addressed in the data base. The other recom-
be recorded at each inspection point. It is recognized that the
mended fields provide additional information that a user will
complete examination record may contain several files, for
find helpful in understanding the ultrasonic examination result.
example, for the same examination method in different object
These header field items will, in most cases, make up only a
areas, with or without image processing, for different exami-
very small part of an ultrasonic examination file. The actual
nation methods (through-transmission, pulse-echo, radiologic,
stream of ultrasonic data that make up the image will take up
infrared, etc.) collected during the same or during different
the largest part of the data base. Since an ultrasonic image file
scan sessions, and for variations within a single method
will normally be large, the concept of data compression will be
(frequency change, etc.). Information about the existence of
considered in many cases. Compressed data should be noted,
other images/examination records for the examined object
along with a description of the compression method, as
should be noted in the appropriate fields.Asingle image plane
indicated in Field No. 122.
may be one created by overlaying or processing results for
3.3 This guide describes the structure of a data file for all of
multiple examination approaches, for example data fusion. For
the ultrasonic information collected in a single scan. Some
such images, the notes sections must clearly state how the
systems record multiple inspection results during a single scan.
image for this file was created.
For example, through transmission attenuation data as well as
TABLE 1 Field Listing
A B
Field Number Field Name and Description Data Type/Units
Header Information:
C D
1 Intermediate file name Alphanumeric string
C
2 Format revision code Alphanumeric string
C D
3 Format revision date yyyy/mm/dd
C
4 Source file name Alphanumeric string
5 Examination file description notes Alphanumeric string
C D
6 Examining company and location Alphanumeric string
C
7 Examination date yyyy/mm/dd
C
8 Examination time hh:mm:ss
C D
9 Type of examination Alphanumeric string
C D
10 Other examinations performed Alphanumeric string
11 Operator Name Alphanumeric string
C
12 Operator identification code Alphanumeric string
C
13 ASTM, ISO, or other applicable standard inspection specification Alphanumeric string
14 Date of applicable standard yyyy/mm/dd
C
15 Acceptance criteria Alphanumeric string
C D
16 System of units Alphanumeric string
17 Notes Alphanumeric string
Examination System Description:
D
18 Examination system manufacturer(s) Alphanumeric string
C
19 Examination system model Alphanumeric string
20 Examination system serial number Alphanumeric string
Pulser Description:
21 Pulser electronics manufacturer Alphanumeric string
22 Pulser electronics model number Alphanumeric string
D
23 Pulser type Alphanumeric string
24 Pulse repetition frequency Real number, kiloHertz
D
25 Pulse height Alphanumeric string
26 Pulse width Real number, nsec
27 Last calibration date yyyy/mm/dd
28 Notes on pulser section Alphanumeric string
Receiver Description:
29 Receiver electronics manufacturer Alphanumeric string
30 Receiver electronics model Alphanumeric string
D
31 Receiver electronics response center frequency Real number, MHz
D
32 Receiver bandwidth Real number, MHz
33 Fixed receiver gain Real number, dB
34 User selected receiver gain Real number, dB
35 Last calibration date yyyy/mm/dd
Notes on receiver section Alphanumeric string
Gate Description:
37 Number of gates Integer
D
38 Gate type Alphanumeric string
39 Gate synchronization Alphanumeric string
40 Gate start delay Alphanumeric string
41 Gate width Alphanumeric string
42 Gate threshold level Alphanumeric string
43 Notes on gate section Alphanumeric string
Search Unit Description:
44 Transmit search unit manufacturer Alphanumeric string
E1454 − 02 (2008)
TABLE 1 Continued
A B
Field Number Field Name and Description Data Type/Units
45 Transmit search unit model Alphanumeric string
46 Transmit search unit serial number Alphanumeric string
47 Transmit search unit element diameter Real number
48 Measured beam diameter of the Transmit search unit at the examination surface Real number
D
49 Location of measurement of beam diameter of the transmit search unit Alphanumeric string
D
50 Transmit search unit focal length Real number
51 Transmit search unit nominal frequency Real number, MHz
52 Transmit search unit response center frequency Real number, MHz
53 Transmit search unit response bandwidth Real number, MHz
54 Transmit search unit cable type Alphanumeric string
55 Transmit search unit cable length Real number
D
56 Number of values for Transmit search unit digitized waveform Integer
57 Transmit search unit waveform values Real number
58 Notes on Transmit search unit waveform Alphanumeric string
59 Transmit search unit coupling technique and medium Alphanumeric string
60 Receive search unit manufacturer Alphanumeric string
61 Receive search unit model number Alphanumeric string
62 Receive search unit serial number Alphanumeric string
63 Receive search unit element diameter Real number
64 Measured beam diameter of the “receive” search unit at the examination surface Real number
D
65 Location of measurement of beam diameter of the receive search unit Alphanumeric string
D
66 Receive search unit focal length Real number
67 Receive search unit nominal frequency Real number, MHz
68 Receive search unit response center frequency Real number, MHz
69 Receive search unit response bandwidth Real number, MHz
70 Receive search unit cable type Alphanumeric string
71 Receive search unit cable length Real number
D
72 Number of values for “receive” search unit digitized waveform Integer
73 Receive search unit waveform values Real number
74 Notes on Receive search unit waveform Alphanumeric string
75 Receive search unit coupling technique and medium Alphanumeric string
Examined Sample Description:
C
76 Examined sample identification Alphanumeric string
C
77 Examined sample name Alphanumeric string
78 Examined sample description Alphanumeric string
C
79 Examined sample material Alphanumeric string
D
80 Examined sample notes (history, use, etc.) Alphanumeric string
C
81 Number of scan segments for this part Integer
82 Reference sample identification Alphanumeric string
83 Reference sample description Alphanumeric string
84 Reference sample file name/location Alphanumeric string
D
85 Reference sample notes (use, etc.) Alphanumeric string
Coordinate System and Scan Description Machine Coordinate System:
D
86 Machine scan axis Alphanumeric string
87 Machine index axis Alphanumeric string
88 Machine third axis Alphanumeric string
89 Reference for machine coordinate system Alphanumeric string
Part Coordinate System:
D
90 First part axis Alphanumeric string
91 Second part axis Alphanumeric string
92 Third part axis Alphanumeric string
93 Reference for part coordinate system Alphanumeric string
Object Target Points:
C
94 Number of target points Integer
C
95 Description of target point Alphanumeric string
C
96 Coordinate of target point in first part axis Real number
C
97 Coordinate of target point in second part axis Real number
98 Coordinate of target point in third part axis Real number
Data Plane:
99 Description of the plane onto which data will be projected Alphanumeric string
100 Coordinate system notes Alphanumeric string
Examination Parameters:
C
101 Coordinate location number Integer
C D
102 Number of data values per coordinate location Integer
C D
103 Minimum value of test data range or resolution Integer
C D
104 Maximum value of test data range or resolution Integer
C D
105 Engineering units for minimum legal data value Alphanumeric string
C D
106 Engineering units for maximum legal data value Alphanumeric string
C
107 Number of bits to which the original data was digitized Integer
C D
108 Type of data scale Alphanumeric string
C D
109 Size of data step Real number
C D
110 Format of data recording Alphanumeric string
C
111 Number of colors or gray levels used Integer
C
112 Distribution of colors or gray levels Alphanumeric string
Examination Results:
E1454 − 02 (2008)
TABLE 1 Continued
A B
Field Number Field Name and Description Data Type/Units
C D
113 Scan segment number Integer
C
114 Scan segment description Alphanumeric string
115 Scan segment location on part Alphanumeric string
116 Scan segment orientation Alphanumeric string
C
117 Scan pattern description Alphanumeric string
D
118 Annotation Alphanumeric string
C
119 Distance between data sample points Real number
C
120 Interval between data locations in index direction Real number
121 Notes on data intervals Alphanumeric string
122 Notes on data format including notes on any compression techniques used Alphanumeric string
C D
123 Total number of data points Integer
C D
124 Actual stream of ultrasonic data Real numbers
A
Field numbers are for reference only.They do not imply a necessity to include all those fields in any specific database nor do they imply a requirement that fields be used
in this particular order.
B
Units listed first are SI; secondary units are inch-pound (English); see Field No. 16.
C
Denotes essential field for computerization of test results.
D
See Section 5 for further explanation.
4. Description of the Field Listings 5.1.4 Field No. 9: Type of Examination—For example, one
of the following may be used: through-transmission; p
...