EN ISO 11137-1:2015
(Main)Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices (ISO 11137-1:2006, including Amd 1:2013)
Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices (ISO 11137-1:2006, including Amd 1:2013)
ISO 11137-1:2006 specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices. Although the scope of ISO 11137-1:2006 is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other products and equipment.
ISO 11137-1:2006 covers radiation processes employing irradiators using the radionuclide 60Co or 137Cs, a beam from an electron generator or a beam from an X-ray generator.
ISO 11137-1:2006 does not:
specify requirements for development, validation and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease;detail specified requirements for designating a medical device as sterile;specify a quality management system for the control of all stages of production of medical devices;specify requirements for occupational safety associated with the design and operation of irradiation facilities;specify requirements for the sterilization of used or reprocessed devices.
Sterilisation von Produkten für die Gesundheitsfürsorge - Strahlen - Teil 1: Anforderungen an die Entwicklung, Validierung und Lenkung der Anwendung eines Sterilisationsverfahrens für Medizinprodukte
1.1 Dieser Teil von ISO 11137 legt Anforderungen für die Entwicklung, Validierung und Lenkung der Anwendung eines Strahlensterilisationsverfahrens für Medizinprodukte fest.
ANMERKUNG Obgleich der Anwendungsbereich von diesem Teil von ISO 11137 auf Medizinprodukte begrenzt ist, legt er Anforderungen fest und liefert Anleitungen, die für andere Produkte und Ausrüstungen gelten können.
Durch diesen Teil von ISO 11137 erfasste Strahlensterilisationsverfahren wenden Bestrahlungsanlagen an,
a) die die Radionuklide 60Co und 137Cs verwenden,
b) in denen beschleunigte Elektronen erzeugt werden
oder
c) in denen durch beschleunigte Elektronen X-Rays (Röntgenstrahlen) genutzt werden.
1.2 Dieser Teil von ISO 11137 legt keine Anforderungen an die Entwicklung, Validierung und Lenkung der Anwendung eines Verfahrens zur Inaktivierung der Verursacher spongiformer Enzephalopathien wie Scrapie, bovine spongiforme Enzephalopathie und Creutzfeldt-Jakob-Krankheit fest. In bestimmten Ländern sind spezifische Empfehlungen für die Behandlung möglicherweise mit diesen Erregern kontaminierter Materialien erarbeitet worden.
ANMERKUNG Siehe zum Beispiel ISO 22442 1, ISO 22442 2 und ISO 22442 3.
1.2.1 Dieser Teil von ISO 11137 legt keine in den Einzelheiten festgelegte Anforderung zur Kennzeichnung eines Medizinprodukts als steril fest.
ANMERKUNG Zu beachten sind nationale oder regionale Anforderungen für die Kennzeichnung von Medizinprodukten als „steril“. Siehe zum Beispiel EN 556 1 oder ANSI/AAMI ST67.
1.2.2 Dieser Teil von ISO 11137 legt kein Qualitätsmanagementsystem für die Lenkung der Anwendung aller Stufen der Herstellung von Medizinprodukten fest.
ANMERKUNG Es ist keine Anforderung von diesem Teil von ISO 11137, dass bei der Herstellung ein vollständiges Qualitätsmanagementsystem vorhanden ist, aber die als Minimum erforderlichen Elemente eines Qualitätsmanagement-systems zur Lenkung der Anwendung des Sterilisationsverfahrens werden an geeigneten Stellen im Text als normative Verweisungen angeführt (siehe im Besonderen Abschnitt 4). Zu beachten sind die Normen über Qualitäts-managementsysteme (siehe ISO 13485), die alle Stufen der Herstellung oder Aufbereitung von Medizinprodukten lenken einschließlich des Sterilisations¬verfahrens. Regionale und nationale Bestimmungen für die Bereitstellung von Medizinprodukten könnten die Durchsetzung eines vollständigen Qualitätsmanagementsystems und die Beurteilung dieses Systems durch eine dritte Seite fordern.
1.2.3 Dieser Teil von ISO 11137 fordert nicht den Einsatz biologischer Indikatoren für die Validierung oder Lenkung der Strahlensterilisation oder dass für die Produktfreigabe eine Prüfung auf Sterilität nach einem Arzneibuch durchgeführt wird.
1.2.4 Dieser Teil von ISO 11137 legt keine Anforderungen an den Arbeitsschutz fest, die mit dem Design und dem Betrieb von Bestrahlungsanlagen zusammenhängen.
ANMERKUNG Es wird auch darauf aufmerksam gemacht, dass in einigen Ländern Vorschriften bestehen, die Sicherheitsanforderungen für den mit der Bestrahlung zusammenhängenden Arbeitsschutz festlegen.
1.2.5 Dieser Teil von ISO 11137 legt keine Anforderungen für die Sterilisation gebrauchter oder wiederaufbereiteter Medizinprodukte fest.
Stérilisation des produits de santé - Irradiation - Partie 1: Exigences relatives à la mise au point, à la validation et au contrôle de routine d'un procédé de stérilisation pour les dispositifs médicaux (ISO 11137-1:2006, y compris Amd 1:2013)
L'ISO 11137-1:2006 spécifie les exigences relatives à la mise au point, à la validation et au contrôle de routine d'un procédé de stérilisation par irradiation pour les dispositifs médicaux. Bien que le domaine d'application de la présente partie de l'ISO 11137 se limite aux dispositifs médicaux, elle spécifie les exigences et fournit des directives qui peuvent être applicables à d'autres produits et équipements.
L'ISO 11137-1:2006 couvre les procédés d'irradiation utilisant des irradiateurs, qui utilisent les radionucléides 60Co ou 137Cs, un faisceau à partir d'un générateur d'électrons ou un faisceau à partir d'un générateur de rayons X.
L'ISO 11137-1:2006
ne spécifie pas les exigences pour la mise au point, la validation et le contrôle de routine d'un procédé pour l'inactivation des agents responsables des encéphalopathies spongiformes telles que la tremblante du mouton (scrapie en anglais), l'encéphalopathie spongiforme bovine et la maladie de Creutzfeldt-Jakob. Des recommandations spécifiques ont été formulées dans certains pays pour le traitement des matériaux potentiellement contaminés par ces agents,ne détaille pas les exigences spécifiées pour désigner un dispositif médical stérile,ne spécifie pas de système de management de la qualité pour le contrôle de toutes les étapes de production des dispositifs médicaux,n'exige ni l'utilisation d'indicateurs biologiques pour la validation ou la surveillance de la stérilisation par radiation, ni qu'un essai de stérilité défini dans la pharmacopée soit réalisé pour la libération du produit.ne spécifie pas d'exigences relatives à la sécurité du travail associée à la conception et au fonctionnement des installations d'irradiation.ne spécifie pas d'exigences relatives à la stérilisation de dispositifs usagés ou retraités.
Sterilizacija izdelkov za zdravstveno nego - Sevanje - 1. del: Zahteve za razvoj, validacijo in rutinsko kontrolo sterilizacijskih postopkov za medicinske pripomočke (ISO 11137-1:2006, vključno z dopolnilom Amd 1:2013)
Standard ISO 11137-1:2006 določa zahteve za razvoj, validacijo in rutinsko kontrolo sterilizacijskega postopka s sevanjem za medicinske pripomočke. Čeprav se je obseg standarda ISO 11137-1: 2006 omejena na medicinske pripomočke, določa zahteve in smernice, ki se lahko uporabljajo za druge izdelke in opremo.
Standard ISO 11137-1:2006 zajema postopke sevanja z iradiatorji z uporabo radionuklidov 60Co ali 137Cs, žarka iz generatorja elektronov ali žarka iz generatorja rentgenskega sevanja.
Standard ISO 11137-1:2006 ne določa:
zahtev za razvoj, validacijo in rutinsko kontrolo postopka za inaktivacijo povzročiteljev spongiformne encefalopatije, kot so praskavec, bovina spongiformna encefalopatija in Creutzfeldt-Jakobova bolezen; zahtev za označevanje medicinskega pripomočka kot sterilnega; sistema vodenja kakovosti za nadzor vseh faz proizvodnje medicinskih pripomočkov; zahtev za varnost pri delu, povezanih z načrtovanjem in upravljanjem sredstev za obsevanje; zahtev za sterilizacijo uporabljenih ali predelanih pripomočkov.
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
01-september-2015
1DGRPHãþD
SIST EN ISO 11137-1:2006
SIST EN ISO 11137-1:2006/A1:2013
6WHULOL]DFLMDL]GHONRY]D]GUDYVWYHQRQHJR6HYDQMHGHO=DKWHYH]DUD]YRM
YDOLGDFLMRLQUXWLQVNRNRQWURORVWHULOL]DFLMVNLKSRVWRSNRY]DPHGLFLQVNH
SULSRPRþNH,62YNOMXþQR]GRSROQLORP$PG
Sterilization of health care products - Radiation - Part 1: Requirements for development,
validation and routine control of a sterilization process for medical devices (ISO 11137-
1:2006, including Amd 1:2013)
Sterilisation von Produkten für die Gesundheitsfürsorge - Strahlen - Teil 1:
Anforderungen an die Entwicklung, Validierung und Lenkung der Anwendung eines
Sterilisationsverfahrens für Medizinprodukte
Stérilisation des produits de santé - Irradiation - Partie 1: Exigences relatives à la mise
au point, à la validation et au contrôle de routine d'un procédé de stérilisation pour les
dispositifs médicaux (ISO 11137-1:2006, y compris Amd 1:2013)
Ta slovenski standard je istoveten z: EN ISO 11137-1:2015
ICS:
11.080.01 Sterilizacija in dezinfekcija na Sterilization and disinfection
splošno in general
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD
EN ISO 11137-1
NORME EUROPÉENNE
EUROPÄISCHE NORM
June 2015
ICS 11.080.01 Supersedes EN ISO 11137-1:2006
English Version
Sterilization of health care products - Radiation - Part 1:
Requirements for development, validation and routine control of
a sterilization process for medical devices (ISO 11137-1:2006,
including Amd 1:2013)
Stérilisation des produits de santé - Irradiation - Partie 1: Sterilisation von Produkten für die Gesundheitsfürsorge -
Exigences relatives à la mise au point, à la validation et au Strahlen - Teil 1: Anforderungen an die Entwicklung,
contrôle de routine d'un procédé de stérilisation pour les Validierung und Lenkung der Anwendung eines
dispositifs médicaux (ISO 11137-1:2006, y compris Amd Sterilisationsverfahrens für Medizinprodukte (ISO 11137-
1:2013) 1:2006, einschließlich Amd 1:2013)
This European Standard was approved by CEN on 20 May 2015.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management Centre has the same
status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United
Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2015 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 11137-1:2015 E
worldwide for CEN national Members.
Contents Page
Foreword .3
Annex ZA (informative) Relationship between this European Standard and the Essential
Requirements of EU Directive 90/385/EEC on active implantable medical devices .5
Annex ZB (informative) Relationship between this European Standard and the Essential
Requirements of EU Directive 93/42/EEC on medical devices .6
Annex ZC (informative) Relationship between this European Standard and the Essential
Requirements of EU Directive 98/79/EC on in vitro diagnostic medical devices .7
Foreword
The text of ISO 11137-1:2006, including Amd 1:2013 has been prepared by Technical Committee ISO/TC 198
“Sterilization of health care products” of the International Organization for Standardization (ISO) and has been
taken over as EN ISO 11137-1:2015 by Technical Committee CEN/TC 204 “Sterilization of medical devices”
the secretariat of which is held by BSI.
This European Standard shall be given the status of a national standard, either by publication of an identical
text or by endorsement, at the latest by December 2015, and conflicting national standards shall be withdrawn
at the latest by December 2015.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN ISO 11137-1:2006.
This document has been prepared under a mandate given to CEN by the European Commission and the
European Free Trade Association, and supports essential requirements of EU Directives.
For relationship with EU Directives, see informative Annexes ZA, ZB, ZC, which are an integral part of this
document.
The following referenced documents are indispensable for the application of this document. For undated
references, the edition of the referenced document (including any amendments) listed below applies. For
dated references, only the edition cited applies. However, for any use of this standard within the meaning of
Annex ZA, ZB or ZC, the user should always check that any referenced document has not been superseded
and that its relevant contents can still be considered the generally acknowledged state-of-art.
When an IEC or ISO standard is referred to in the ISO standard text, this should be understood as a
normative reference to the corresponding EN standard, if available, and otherwise to the dated version of the
ISO or IEC standard as listed below.
NOTE The way in which these referenced documents are cited in normative requirements determines the extent (in
whole or in part) to which they apply.
Table — Correlation between normative references and dated EN and ISO standards
Normative references Equivalent dated standard
as listed in Clause 2 of the ISO
EN ISO
standard
ISO 10012-1 EN ISO 10012:2003 ISO 10012:2003
ISO 11137-2 EN ISO 11137-2:2013 ISO 11137-2:2013
ISO 11737-1 EN ISO 11737-1:2006 + AC:2009 ISO 11737-1:2006 + Cor 1:2007
ISO 11737-2 EN ISO 11737-2:2009 ISO 11737-2:2009
ISO 13485 EN ISO 13485:2012 ISO 13485:2003
According to the CEN-CENELEC Internal Regulations, the national standards organizations of the following
countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech
Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece,
Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,
Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom.
Endorsement notice
The text of ISO 11137-1:2006, including Amd 1:2013 has been approved by CEN as EN ISO 11137-1:2015
without any modification.
Annex ZA
(informative)
Relationship between this European Standard and the Essential
Requirements of EU Directive 90/385/EEC on active implantable
medical devices
This European Standard has been prepared under a mandate given to CEN/CENELEC by the European
Commission and the European Free Trade Association to provide a means of conforming to Essential
Requirements of the New Approach Directive 90/385/EEC on active implantable medical devices.
Once this standard is cited in the Official Journal of the European Union under that Directive and has been
implemented as a national standard in at least one Member State, compliance with the normative clauses of
this standard given in Table ZA.1 confers, within the limits of the scope of this standard, a presumption of
conformity with the corresponding Essential Requirements of that Directive and associated EFTA regulations.
NOTE 1 Where a reference from a clause of this standard to the risk management process is made, the risk
management process needs to be in compliance with 90/385/EEC, as amended by 2007/47/EC. This means that risks
have to be reduced ‘as far as possible’, ‘to a minimum’, ‘to the lowest possible level’, ‘minimized’ or ‘removed’, according
to the wording of the corresponding essential requirement.
NOTE 2 The manufacturer’s policy for determining acceptable risk must be in compliance with essential requirements
1, 4, 5, 8, 9 and 10 of the Directive.
NOTE 3 This Annex ZA is based on normative references according to the table of references in the European
foreword, replacing the references in the core text.
NOTE 4 When an Essential Requirement does not appear in Table ZA.1, it means that it is not addressed by this
European Standard.
Table ZA.1 — Correspondence between this European Standard and Directive 90/385/EEC
Clauses of this EN Essential Requirements (ERs) of Qualifying remarks/Notes
Directive 90/385/EEC
4,5,6,7,8,9,10,11,12 7 Only a sterilization process using
ionizing radiation is considered by
this standard.
This relevant Essential
Requirement is only partly
addressed in this European
Standard. Design and packaging
for maintenance of sterility during
transportation and storage are not
covered. Aspects of manufacture
other than those related to
sterilization are not covered.
WARNING — Other requirements and other EU Directives may be applicable to the product(s) falling within
the scope of this Standard.
Annex ZB
(informative)
Relationship between this European Standard and the Essential
Requirements of EU Directive 93/42/EEC on medical devices
This European Standard has been prepared under a mandate given to CEN/CENELC by the European
Commission and the European Free Trade Association to provide a means of conforming to Essential
Requirements of the New Approach Directive 93/42/EEC on medical devices.
Once this standard is cited in the Official Journal of the European Union under that Directive and has been
implemented as a national standard in at least one Member State, compliance with the normative clauses of
this standard given in Table ZB.1 confers, within the limits of the scope of this standard, a presumption of
conformity with the corresponding Essential Requirements of that Directive and associated EFTA regulations.
NOTE 1 Where a reference from a clause of this standard to the risk management process is made, the risk
management process needs to be in compliance with 93/42/EEC, as amended by 2007/47/EC. This means that risks have
to be reduced ‘as far as possible’, ‘to a minimum’, ‘to the lowest possible level’, ‘minimized’ or ‘removed’, according to the
wording of the corresponding essential requirement.
NOTE 2 The manufacturer’s policy for determining acceptable risk must be in compliance with essential requirements
1, 2, 5, 6, 7, 8, 9, 11 and 12 of the Directive.
NOTE 3 This Annex ZA is based on normative references according to the table of references in the European
foreword, replacing the references in the core text.
NOTE 4 When an Essential Requirement does not appear in Table ZA.1, it means that it is not addressed by this
European Standard.
Table ZB.1 — Correspondence between this European Standard and Directive 93/42/EEC
Clauses of this EN Essential Requirements (ERs) of Qualifying remarks/Notes
Directive 93/42/EEC
4,5,6,7,8,9,10,11,12 8.3 Only a sterilization process using
ionizing radiation is considered by
this standard.
This relevant Essential
Requirement is only partly
addressed in this European
Standard. Design and packaging
for maintenance of sterility during
transportation and storage are not
covered. Aspects of manufacture
other than those related to
sterilization are not covered.
4,5,6,7,8,9,10,11,12 8.4 This relevant Essential
Requirement is only partly
addressed in this European
Standard. Aspects of manufacture
other than those related to
sterilization are not covered.
WARNING — Other requirements and other EU Directives may be applicable to the product(s) falling within
the scope of this Standard.
Annex ZC
(informative)
Relationship between this European Standard and the Essential
Requirements of EU Directive 98/79/EC on in vitro diagnostic
medical devices
This European Standard has been prepared under a mandate given to CEN/CENELEC by the European
Commission and the European Free Trade Association to provide a means of conforming to Essential
Requirements of the New Approach Directive 98/79/EC on in vitro diagnostic medical devices.
Once this standard is cited in the Official Journal of the European Union under that Directive and has been
implemented as a national standard in at least one Member State, compliance with the normative clauses of
this standard given in Table ZC.1 confers, within the limits of the scope of this standard, a presumption of
conformity with the corresponding Essential Requirements of that Directive and associated EFTA regulations.
NOTE 1 Where a reference from a clause of this standard to the risk management process is made, the risk
management process needs to be in compliance with 98/79/EC. This means that risks have to be reduced ‘as far as
possible’, ‘to a minimum’, ‘to the lowest possible level’, ‘minimized’ or ‘removed’, according to the wording of the
corresponding essential requirement.
NOTE 2 The manufacturer’s policy for determining acceptable risk must be in compliance with essential requirements
Part A: 1, 2 and 5; Part B: 1.2, 2, 3, 5, 6, and
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.