EN ISO 10328:2026
(Main)Prosthetics - Structural testing of lower-limb prostheses - Requirements and test methods (ISO 10328:2026)
- Abstract
This document specifies procedures for static and cyclic strength tests on lower-limb prostheses (see NOTE 2) which typically produce compound loadings by the application of a single test force. The compound loads in the test sample relate to the peak values of the components of loading which normally occur at different instants during the stance phase of walking.
This document specifies Test Ranges (R) by specifying locomotion profiles for the cyclic test in relation to the intended use. According to the concept of the tests of this document, each sample of lower limb prosthetic device or structure submitted for test is, nevertheless, free to develop its individual performance under load.
The tests described in this document comprise:
principal static and cyclic tests for all components;
a separate static test in torsion for all components;
separate static and cyclic tests on ankle-foot devices and foot units for all ankle-foot devices as single components including ankle units or ankle attachments and all foot units as single components;
a separate static ultimate strength test in maximum knee flexion on knee joints and associated parts for all knee units or knee-shin-assemblies and adjacent components that normally provide the flexion stop on a complete prosthesis;
optional separate tests on knee locks for all mechanisms which lock the knee joint in the extended position of the knee unit or knee-shin-assembly.
The tests described in this document apply to specific types of ankle-disarticulation prostheses (see NOTE 2), to transtibial (below-knee), knee-disarticulation and transfemoral (above-knee) prostheses and to the distal (lower) part of hip-disarticulation and hemi-pelvectomy prostheses (see NOTE 3). The test geometry described in this standard is only intended to be used up to a level of 150 mm above the knee-joint centre.
This document is suitable for the assessment of the conformity of lower limb prosthetic devices or structures with the strength requirements specified in ISO 22523:2006, 4.4 (see NOTE 1). Lower limb prosthetic devices or structures on the market, which have demonstrated their compliance with the strength requirements specified in ISO 22523:2006, 4.4 through submission to the relevant tests of ISO 22523:2006, need not be retested to this document.
NOTE 1 The tests can be performed on complete structures, on part structures or on individual components.
NOTE 2 The tests only apply to ankle-disarticulation prostheses which include (foot) components of prosthetic ankle-foot devices taken from the normal production line.
NOTE 3 The distal part comprises the knee unit, the ankle-foot device and all parts between. Tests on hip units are described in ISO 15032.
- Status
- Not Published
- Publication Date
- 08-Dec-2026
- Technical Committee
- CEN/TC 293 - Assistive products and accessibility
- Current Stage
- 6055 - CEN Ratification completed (DOR) - Publishing
- Start Date
- 03-Nov-2026
- Due Date
- 03-Nov-2026
- Completion Date
- 03-Nov-2026
EN ISO 10328:2026 is a European standard published by CEN and aligned with ISO 10328 on structural testing of lower-limb prostheses. It specifies how to carry out static and cyclic strength tests on prosthetic devices and structures, so manufacturers, test laboratories and conformity assessors can check whether lower-limb prostheses meet the strength requirements used for market acceptance and other conformity work.
What does EN ISO 10328:2026 specify?
EN ISO 10328:2026 specifies test procedures, loading conditions and reporting rules for lower-limb prostheses and related components. It covers the geometry of the test set-up, the test forces, the sample preparation, the pass/fail criteria and the way compliance is claimed.
The document is not a guide for choosing a prosthesis for an individual amputee. It is intended for structural testing of devices and parts under defined laboratory conditions, with a stated warning that misuse for prescription can create a safety risk.
It applies to ankle-disarticulation prostheses that use foot components from normal production, transtibial, knee-disarticulation and transfemoral prostheses, and the distal part of hip-disarticulation and hemi-pelvectomy prostheses. The test geometry is intended for use only up to 150 mm above the knee-joint centre.
The main body runs through the scope, references, terms, symbols, strength categories, coordinate systems, test loading conditions, values of loads and cycles, compliance, test samples, responsibilities, the submission document, equipment, accuracy, test principles, the principal and separate test procedures, logging, reporting, classification and final compliance rules.
| Annex | What it covers |
|---|---|
| Annex A | Internal loads and their effects, including reference lines and load components |
| Annex B | Background data for principal cyclic test loading conditions and loading levels |
| Annex C | Guidance on the alternative static ultimate strength loading rate |
| Annex D | Records for the test laboratory/facility log |
| Annex E | Background on loading profiles used for ankle-foot device and foot unit cyclic testing |
| Annex G | Additional P2 loading level data for principal and separate tests |
| Annex H | Alternative test geometry based on recent biomechanical studies |
| Annex ZA | Relationship with Regulation (EU) 2017/745 and presumption of conformity |
What are the key requirements of EN ISO 10328:2026?
EN ISO 10328:2026 turns the strength of a prosthetic device into a set of defined laboratory tests and classified outcomes. The most important requirements are the assignment of a Test Range (R), the use of defined test configurations, the selection of the correct sample type, and the completion of the relevant static and cyclic tests.
How are strength categories and Test Range (R) assigned?
EN ISO 10328:2026 requires the manufacturer or submitter to assign a Test Range (R) from the written intended use of the device. That assignment is the basis for choosing the test loading level and the loads used in the laboratory.
Clause 5 groups the document's strength checks into four categories: proof strength, ultimate strength, fatigue strength, and static strength in torsion. This matters in practice because the type of test chosen depends on whether the device is being checked for occasional severe loads, a gross single event, repeated walking loads, or twisting resistance.
The compliance claim must state both the test loading level and the Test Range (R). In other words, a pass is always tied to a defined use case, not to a general statement that the product is "strong enough" in all situations.
How is the test geometry defined?
EN ISO 10328:2026 uses two three-dimensional coordinate systems, one for right-sided and one for left-sided application. Clause 6 defines the axes, the top, knee, ankle and bottom reference planes, the reference points, the load line and the effective ankle-joint and knee-joint centres.
This geometry matters because the same single test force is meant to create compound internal loads that resemble what happens during walking. If the alignment is wrong, the internal bending and twisting loads change, and the test no longer represents the intended loading case.
Clause 6 also defines offsets, combined offsets and effective lever arms. These values are used to position the load line and to set the sample up in the worst-case alignment position.
What tests must a sample undergo?
EN ISO 10328:2026 includes principal static and cyclic tests for all components, a separate static torsion test for all components, separate tests on ankle-foot devices and foot units, a separate static ultimate strength test in maximum knee flexion for knee joints and associated parts, and optional separate tests on knee locks.
The key point is that the required combination of tests depends on what kind of prosthetic structure is submitted. Clause 9 and Table 14 link the sample type to the tests that are needed to claim compliance.
This is practical for laboratories because a complete prosthesis, a partial structure, and a single ankle-foot device are not tested in the same way. The standard tells the lab which parts are to be included, which parts may be omitted, and which loading configurations apply.
What does the document require for sample preparation?
EN ISO 10328:2026 requires test samples to include the parts normally fitted, to be taken from the normal production line where applicable, and to receive an indelible, unique and traceable identification. Cosmetic parts are omitted unless they contribute to the structure.
Clause 10 also requires the worst-case alignment position to be used for all tests. If a sample can be assembled in more than one prosthetic arrangement, it must be set up in the most adverse one possible.
Clause 11 assigns responsibility for assembly, torque settings, replacement parts during cycling, submission documents and the attachment of end fittings or extension pieces. This matters because the test result depends not only on the component design but also on how it is assembled and aligned.
What do the tests actually measure?
EN ISO 10328:2026 uses static proof tests, static ultimate strength tests and cyclic tests to determine whether a sample can carry defined loads without failing or exceeding permitted deformation. Clause 15 explains that the static tests represent severe single events, while the cyclic tests represent repeated walking loads.
The ultimate strength test also evaluates shock absorption capacity from the force-deflection curve. That is important because some structures may fail safely by deforming and absorbing energy rather than breaking abruptly.
The separate torsion test checks relative angular movement and slippage between clamped parts in both twisting directions. The ankle-foot and foot unit tests apply heel and forefoot loading, first statically and then cyclically, so the device is checked under the loading pattern used in walking.
What controls accuracy and comparability?
EN ISO 10328:2026 requires the test equipment, alignment jigs and measuring devices to be calibrated at least annually and whenever a part is replaced. Clause 14 also sets accuracy targets for dimensions, angles, forces, moments, frequency and displacement.
This matters because small errors in alignment or load magnitude can change the outcome of a structural test. The standard therefore limits how far the measured set-up may deviate from the intended geometry and loading.
What is needed to claim compliance?
EN ISO 10328:2026 requires a prescribed number of samples from the allowed batch to meet the relevant requirements. The test report must identify the laboratory, the submitter, the test samples, the test loading level, the Test Range (R), the dates and the result.
Clauses 20 and 21 define the classification and identifier format as ISO 10328 - P - R - m kg, where m is the maximum body mass stated in the written instructions. This gives a clear way to link the tested load level to the intended user group and to the product marking or instructions for use.
What terms does EN ISO 10328:2026 define?
- Proof strength - A static load that represents an occasional severe event and should still allow the prosthetic device or structure to function as intended.
- Ultimate strength - A static load representing a gross single event; the device may survive it but could be unusable afterwards.
- Fatigue strength - A cyclic load that the prosthetic device or structure can withstand for a defined number of cycles.
- Batch - The set of test samples submitted together to a laboratory to demonstrate conformity.
- Shock absorption capacity - The ability of a specimen to absorb energy by deflection without a proportional rise in force.
- Test Range (R) - The loading range assigned from the written intended use, used to select the test level and classify the result.
- Offset - The f- or o-coordinate used to position the load line relative to the reference planes.
Who uses EN ISO 10328:2026?
EN ISO 10328:2026 is used by prosthetics manufacturers, component suppliers, test laboratories, conformity assessment bodies and quality managers. They use it to design test plans, prepare specimens, set alignment, run structural strength tests and document results.
It is also relevant to regulatory and technical teams working on medical devices, because Annex ZA links the standard to Regulation (EU) 2017/745. In practice, that means it can support claims of conformity where the standard is cited for that purpose.
What changed in EN ISO 10328:2026 from the previous edition?
The document background states that the loading-rate range for the related static ultimate strength tests is now normative at 100 N/s to 10 kN/s, replacing the narrower range used in ISO 10328:2016. The annexes also add data for a lower P2 loading level and provide an alternative test geometry based on recent biomechanical studies.
Which standards are used with EN ISO 10328:2026?
- ISO 8549-1 - Provides the vocabulary used for prosthetics and orthotics terms in this document.
- ISO 7000 - Supplies registered graphical symbols used on equipment and identifiers.
- IEC 60417 - Supplies graphical symbols for use on equipment.
- ISO 22523:2006 - Gives broader requirements and test methods for external limb prostheses and external orthoses; Clause 5 uses it for the strength requirement reference.
- ISO 22675:2024 - Covers testing of ankle-foot devices and foot units, which this document references for related components.
- ISO 15032 - Named in a note for hip unit testing, which is relevant when considering related prosthetic parts outside this document's main scope.
What does the EN ISO 10328:2026 document contain?
EN ISO 10328:2026 contains definitions of coordinate systems, reference planes, reference points, offsets and test loads, along with tables of forces, moments and cycle counts. It also includes formulas for calculating internal loads and offsets, and figures that illustrate the right- and left-sided test arrangements.
The core procedures are laid out as step-by-step methods for the principal static proof test, the principal static ultimate strength test, the principal cyclic test, the separate torsion test, the separate ankle-foot and foot unit tests, and the maximum knee flexion test. These procedures are the parts used for laboratory testing, certification work and compliance demonstrations.
The annexes add background data, test-log guidance, alternative geometry options and the link to EU medical device regulation. Annex D is particularly practical for labs because it summarises what needs to be recorded in the test log and the test report.
Relations
- Consolidated By
ISO 10328:2026 - Prosthetics — Structural testing of lower-limb prostheses — Requirements and test methods - Effective Date
- 12-Feb-2026
- Effective Date
- 03-May-2023
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
EN ISO 10328:2026 is a draft published by the European Committee for Standardization (CEN). Its full title is "Prosthetics - Structural testing of lower-limb prostheses - Requirements and test methods (ISO 10328:2026)". This standard covers: This document specifies procedures for static and cyclic strength tests on lower-limb prostheses (see NOTE 2) which typically produce compound loadings by the application of a single test force. The compound loads in the test sample relate to the peak values of the components of loading which normally occur at different instants during the stance phase of walking. This document specifies Test Ranges (R) by specifying locomotion profiles for the cyclic test in relation to the intended use. According to the concept of the tests of this document, each sample of lower limb prosthetic device or structure submitted for test is, nevertheless, free to develop its individual performance under load. The tests described in this document comprise: principal static and cyclic tests for all components; a separate static test in torsion for all components; separate static and cyclic tests on ankle-foot devices and foot units for all ankle-foot devices as single components including ankle units or ankle attachments and all foot units as single components; a separate static ultimate strength test in maximum knee flexion on knee joints and associated parts for all knee units or knee-shin-assemblies and adjacent components that normally provide the flexion stop on a complete prosthesis; optional separate tests on knee locks for all mechanisms which lock the knee joint in the extended position of the knee unit or knee-shin-assembly. The tests described in this document apply to specific types of ankle-disarticulation prostheses (see NOTE 2), to transtibial (below-knee), knee-disarticulation and transfemoral (above-knee) prostheses and to the distal (lower) part of hip-disarticulation and hemi-pelvectomy prostheses (see NOTE 3). The test geometry described in this standard is only intended to be used up to a level of 150 mm above the knee-joint centre. This document is suitable for the assessment of the conformity of lower limb prosthetic devices or structures with the strength requirements specified in ISO 22523:2006, 4.4 (see NOTE 1). Lower limb prosthetic devices or structures on the market, which have demonstrated their compliance with the strength requirements specified in ISO 22523:2006, 4.4 through submission to the relevant tests of ISO 22523:2006, need not be retested to this document. NOTE 1 The tests can be performed on complete structures, on part structures or on individual components. NOTE 2 The tests only apply to ankle-disarticulation prostheses which include (foot) components of prosthetic ankle-foot devices taken from the normal production line. NOTE 3 The distal part comprises the knee unit, the ankle-foot device and all parts between. Tests on hip units are described in ISO 15032.
This document specifies procedures for static and cyclic strength tests on lower-limb prostheses (see NOTE 2) which typically produce compound loadings by the application of a single test force. The compound loads in the test sample relate to the peak values of the components of loading which normally occur at different instants during the stance phase of walking. This document specifies Test Ranges (R) by specifying locomotion profiles for the cyclic test in relation to the intended use. According to the concept of the tests of this document, each sample of lower limb prosthetic device or structure submitted for test is, nevertheless, free to develop its individual performance under load. The tests described in this document comprise: principal static and cyclic tests for all components; a separate static test in torsion for all components; separate static and cyclic tests on ankle-foot devices and foot units for all ankle-foot devices as single components including ankle units or ankle attachments and all foot units as single components; a separate static ultimate strength test in maximum knee flexion on knee joints and associated parts for all knee units or knee-shin-assemblies and adjacent components that normally provide the flexion stop on a complete prosthesis; optional separate tests on knee locks for all mechanisms which lock the knee joint in the extended position of the knee unit or knee-shin-assembly. The tests described in this document apply to specific types of ankle-disarticulation prostheses (see NOTE 2), to transtibial (below-knee), knee-disarticulation and transfemoral (above-knee) prostheses and to the distal (lower) part of hip-disarticulation and hemi-pelvectomy prostheses (see NOTE 3). The test geometry described in this standard is only intended to be used up to a level of 150 mm above the knee-joint centre. This document is suitable for the assessment of the conformity of lower limb prosthetic devices or structures with the strength requirements specified in ISO 22523:2006, 4.4 (see NOTE 1). Lower limb prosthetic devices or structures on the market, which have demonstrated their compliance with the strength requirements specified in ISO 22523:2006, 4.4 through submission to the relevant tests of ISO 22523:2006, need not be retested to this document. NOTE 1 The tests can be performed on complete structures, on part structures or on individual components. NOTE 2 The tests only apply to ankle-disarticulation prostheses which include (foot) components of prosthetic ankle-foot devices taken from the normal production line. NOTE 3 The distal part comprises the knee unit, the ankle-foot device and all parts between. Tests on hip units are described in ISO 15032.
EN ISO 10328:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.40 - Implants for surgery, prosthetics and orthotics; 11.180.10 - Aids and adaptation for moving. The ICS classification helps identify the subject area and facilitates finding related standards.
EN ISO 10328:2026 has the following relationships with other standards: It is inter standard links to ISO 10328:2026, EN ISO 10328:2016. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN ISO 10328:2026 is associated with the following European legislation: EU Directives/Regulations: 2017/745; Standardization Mandates: M/575, M/575 Amd 2. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
EN ISO 10328:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 proof strength
- 3.2 ultimate strength
- 3.3 fatigue strength
- 3.4 batch
- 3.5 shock absorption capacity
- 3.6 test force
- 4 Symbols
- 5 Strength and related performance requirements and conditions of use
- 6 Coordinate systems and test configurations
- 6.1 General
- 6.2 Axes of coordinate systems
- 6.3 Reference planes
- 6.4 Reference points
- 6.5 Test force
- 6.6 Load line
- 6.7 Longitudinal axis of the foot and effective joint centres and centrelines
- 6.8 Reference distances
- 7 Test loading conditions and test loading levels
- 7.1 Test loading conditions
- 7.2 Test loading levels and Test Ranges (R)
- 8 Values of test loads, dimensions and cycles
- 9 Compliance
- 9.1 General
- 9.2 Selection of tests required to claim compliance with this document
- 9.3 Arrangements for tests on samples of prosthetic structures including ankle-foot devices
- 9.4 Number of tests and test samples required to claim compliance with this document
- 9.5 Multiple use of test samples
- 9.6 Testing at particular test loading levels not specified in this document
- 16.3.2.22 (see 16.3.2) loading condition II [7.1.2 b)]
- 10 Test samples
- 10.1 Selection of test samples
- 10.2 Types of test samples
- 10.3 Preparation of test samples
- 10.4 Identification of test samples
- 10.5 Alignment of test samples
- 10.6 Worst-case alignment position of test samples
- 11 Responsibility for test preparation
- 12 Test submission document
- 12.1 General requirements
- 12.2 Information required for test samples
- 12.3 Information required for tests
- 13 Equipment
- 13.1 General
- 13.5 Equipment for the separate static ultimate strength test in maximum knee flexion for
- 14 Accuracy
- 14.1 General
- 14.2 Accuracy of equipment
- 14.3 Accuracy of procedure
- 15 Test principles
- 15.1 General
- 15.2 Static test procedure
- 15.3 Cyclic test procedure
- 16 Test procedures – Principal structural tests
- 16.1 Test loading requirements
- 16.2 Principal static test procedure
- 16.3 Principal cyclic test procedure
- 17 Test procedures — Separate structural tests
- 17.1 Separate static test in torsion
- 17.2 Separate tests on ankle-foot devices and foot units
- 17.3 Separate static ultimate strength test in maximum knee flexion for knee joints and
- 17.4 Separate optional tests on knee locks
- 18 Test laboratory/facility log
- 18.1 General requirements
- 18.2 Specific requirements
- 19 Test report
- 19.1 General requirements
- 19.2 Specific requirements
- 19.3 Options
- 20 Classification and designation
- 20.1 General
- 20.2 Examples of classification and designation
- 21 Compliance with this document
- 21.1 General
- 21.2 Example of identifier layout
- 21.3 Identifier placement
- Annex A
Introduction
Throughout this document, the term prosthesis means an externally applied device used to replace wholly, or in part, an absent limb segment.
As a result of concern in the international community about the need to provide prostheses that are safe in use, and also because of an awareness that test standards would assist the development of better prostheses, a series of meetings was held under the aegis of the International Society for Prosthetics and Orthotics (ISPO). The final one was held in Philadelphia, PA, USA in 1977 at which a preliminary consensus was reached on methods of testing and the required load values. From 1979 onwards this work was continued by ISO Technical Committee 168 leading to the development of ISO 10328:1996. The test procedures may not be applicable to prostheses of mechanical characteristics different from those used in the consensus. During use, a prosthesis is subjected to a series of load actions, each varying individually with time. The test methods specified in this document use static and cyclic strength tests which typically produce compound loadings by the application of a single test force.
The static tests relate to the worst loads generated in any activity. The cyclic tests relate to normal walking activities where loads occur regularly with each step. This document specifies fatigue testing of structural components. The tests specified do not provide sufficient data to predict actual service life. The evaluation of lower-limb prostheses and their components requires controlled field trials in addition to the laboratory tests specified in this document.
The laboratory tests and field trials should be repeated when significant design changes are made to a loadbearing part of a prosthesis. Ideally, additional laboratory tests should be carried out to deal with function, wear and tear, new material developments, environmental influences and user activities as part of the evaluation procedure. There are no standards for such tests, so appropriate procedures will need to be determined.
oSIST prEN ISO 10328:2025
oSIST prEN ISO 10328:2025 DRAFT International Standard ISO/DIS 10328:2025(en) Prosthetics — Structural testing of lower-limb prostheses — Requirements and test methods
1 Scope
WARNING — This document is not suitable to serve as a guide for the selection of a specific lower limb prosthetic device/structure in the prescription of an individual lower limb prosthesis! Any disregard of this warning can result in a safety risk for amputees. This document specifies procedures for static and cyclic strength tests on lower-limb prostheses (see NOTE 2) which typically produce compound loadings by the application of a single test force. The compound loads in the test sample relate to the peak values of the components of loading which normally occur at different instants during the stance phase of walking.
This document specifies Test Ranges (R) by specifying locomotion profiles for the cyclic test in relation to the intended use. According to the concept of the tests of this document, each sample of lower limb prosthetic device / structure submitted for test is, nevertheless, free to develop its individual performance under load. The tests described in this document comprise - principal static and cyclic tests for all components; - a separate static test in torsion for all components; - separate static and cyclic tests on ankle-foot devices and foot units for all ankle-foot devices as single components including ankle units or ankle attachments and all foot units as single components; - a separate static ultimate strength test in maximum knee flexion on knee joints and associated parts for all knee units or knee-shin-assemblies and adjacent components that normally provide the flexion stop on a complete prosthesis; - optional separate tests on knee locks for all mechanisms which lock the knee joint in the extended position of the knee unit or knee-shin-assembly.
The tests described in this document apply to specific types of ankle-disarticulation prostheses (see NOTE 2), to transtibial (below-knee), knee-disarticulation and transfemoral (above-knee) prostheses and to the distal (lower) part of hip-disarticulation and hemi-pelvectomy prostheses (see NOTE 3). The test geometry described in this standard is only intended to be used up to a level of 150 mm above the knee-joint centre. This document is suitable for the assessment of the conformity of lower limb prosthetic devices/ structures with the strength requirements specified in 4.4 of ISO 22523:2006 (see NOTE 1). Lower limb prosthetic devices/structures on the market, which have demonstrated their compliance with the strength requirements specified in 4.4 of ISO 22523:2006 through submission to the relevant tests of ISO 22523:2006, need not be retested to this document.
NOTE 1 The tests can be performed on complete structures, on part structures or on individual components. NOTE 2 The tests only apply to ankle-disarticulation prostheses which include (foot) components of prosthetic ankle-foot devices taken from the normal production line.
NOTE 3 The distal part comprises the knee unit, the ankle-foot device and all parts between. Tests on hip units are described in ISO 15032.
oSIST prEN ISO 10328:2025 ISO/DIS 10328:2025(en)
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 7000, Graphical symbols for use on equipment — Registered symbols ISO 8549-1, Prosthetics and orthotics — Vocabulary — Part 1: General terms for external limb prostheses and external orthoses ISO 22675:2024, Prosthetics — Testing of ankle-foot devices and foot units — Requirements and test methods ISO 22523:2006, External limb prostheses and external orthoses — Requirements and test methods IEC 60417, Graphical symbols for use on equipment
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 8549-1 and the following apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
3.1 proof strength
static load representing an occasional severe event, which can be sustained by the prosthetic device/ structure and still allow it to function as intended
3.2 ultimate strength
static load representing a gross single event, which can be sustained by the prosthetic device/structure but which could render it thereafter unusable
3.3 fatigue strength
cyclic load which can be sustained by the prosthetic device/structure for a given number of cycles
3.4 batch
set of test samples of a prosthetic device/structure submitted together to a test laboratory/facility to undertake tests to demonstrate conformity with requirements
3.5 shock absorption capacity
capacity of a specimen to absorb energy by deflection without a proportional increase of force
3.6 test force
force applied to a sample under test
4 Symbols
The designations and symbols of all relevant test forces and moments are listed in Table 1.
oSIST prEN ISO 10328:2025 ISO/DIS 10328:2025(en) Table 1 — Designations and symbols of test forces and moments Designation Symbol Test forces; twisting moments F, F , F ; M 1 2 u Proof test force of end attachments F pa Stabilizing test force F stab Settling test force F set Static proof test force F sp Static proof test force on heel/forefoot F , F 1sp 2sp Static ultimate test force F su Static ultimate test force on heel/forefoot F , F 1su 2su Minimum test force F cmin Maximum test force F cmax Range of pulsating test force F cr Mean test force F cmean Amplitude of pulsating test force F ca Pulsating test force F (t)
Final static test force F fin Minimum test force on heel/forefoot F , F 1cmin 2cmin Maximum test force on heel/forefoot F , F 1cmax 2cmax Range of pulsating test force on heel/forefoot F , F 1cr 2cr Mean test force on heel/forefoot F , F 1cmean 2cmean Amplitude of pulsating test force on heel/forefoot F , F 1ca 2ca Pulsating test force on heel/forefoot F (t), F (t) 1c 2c Final static test force on heel/forefoot F , F 1fin 2fin Stabilizing twisting moment M u-stab Settling twisting moment M u-set Maximum twisting moment M u-max NOTE Further details of the test forces and moments listed are given in Table 4.
5 Strength and related performance requirements and conditions of use
5.1 A lower limb prosthetic device/structure shall have the strength to sustain the loads occurring during use by amputees in the manner intended by the manufacturer for that device according to his written instructions on its intended use in accordance with ISO 22523:2006, 4.4.1. Based on the written instructions for use, the manufacturer/submitter assigns a Test Range (R) that is appropriate to test the strength of the device for the intended use. The manufacturer / supplier is responsible for the instructions for use and the related assignment. National or international classification schemes are independent to the instructions and the related assignment. For the assessment of the strength to sustain loads occurring during use, this document provides a means of determining the four categories of strength. These are listed in Table 2, together with the related performance requirements and the test methods for their verification. 5.2 In order to demonstrate the strength to sustain the loads occurring during use by amputees of a specific lower limb prosthetic device/structure, the following safety concept shall apply. The device/structure shall a) comply with the requirements in 9.1, 9.2 and 9.3 and for a specific test loading level the requirements in 7.2. and b) be used in accordance with the body mass limit specified by the manufacturer in consideration of the intended use of that device (see NOTE).
oSIST prEN ISO 10328:2025 ISO/DIS 10328:2025(en) and c) be used solely for the intended use as described in the instructions for use.. The conditions in a), b) and c) are regarded in both the classification and designation of prosthetic devices/ structures according to Clause 20and their identification according to Clause 21. NOTE The statement of the body mass limit not to be exceeded by amputees is part of the conditions of use to be specified, with justification, by the manufacturer in his written instructions on the intended use of a specific lower limb prosthetic device/structure, taking account of all other factors affecting the loads expected to be exerted on that lower limb prosthetic device/structure by amputees (see Clause B.1). Table 2 — Categories of strength addressed in this document, together with the related performance requirements and test methods for their verification Category of Related performance require- Test method for verification a strength ment Principal static proof test (16.2.1), separately applied in two test configurations, Structure shall sustain static loading by proof test forces at prescribed separate static proof test for ankle-foot devices and foot values for prescribed times units (17.2.3), successively applied in heel and forefoot Proof strength loading Permanent deformation of structure shall not exceed prescribed values in Principal static proof test (16.2.1) any loading condition Principal static ultimate strength test (16.2.2), separately applied in two test configurations, Structure shall sustain static loading separate static ultimate strength test for ankle-foot Ultimate strength by ultimate test forces at prescribed devices and foot units (17.2.4), separately applied in heel values and forefoot loading, separate static ultimate strength test in maximum knee flexion for knee joints and associated parts (17.3) Structure shall sustain successively 1) static loading by maximum test forces at prescribed values for prescribed times; Principal cyclic test (16.3), separately applied in two test configurations, 2) cyclic loading by pulsating test Fatigue strength forces at prescribed values for pre- separate cyclic test for ankle-foot devices and foot units scribed numbers of cycles; (17.2.5), separately applied in heel and forefoot loading, 3) final static loading by final test forces at prescribed values for prescribed times Structure shall sustain static loading Static strength in by static test force at prescribed value torsion for prescribed time Separate static test in torsion (17.1), applied in two op- Security against Relative angular movement between posite directions of twisting slippage of ends of structure shall not exceed clamped compo- prescribed value nents a The performance requirements related to a specific category of strength are specified in full length in an individual subclause following the subclause in which the test method for their verification is specified.
6 Coordinate systems and test configurations
6.1 General
6.1.1 For ease in interpretation and presentation, two test configurations are specified, one for rightsided and a mirror image for left-sided application. This measure makes it possible to apply uniform sign
oSIST prEN ISO 10328:2025 ISO/DIS 10328:2025(en) conventions for corresponding components of loading generated in the load-bearing structures of right and left prostheses or in asymmetrically designed prosthetic components. 6.1.2 Each test configuration shall be defined in a three-dimensional, rectangular coordinate system (see Figure 1), having an origin 0 and containing a geometric system of planes, lines and points (see Figures 2 and 3). 6.1.3 Each test configuration specifies reference parameters both for the position of the line of application of the test force and for the alignment of test samples within the coordinate system.
6.2 Axes of coordinate systems
6.2.1 The axes of each of the coordinate systems are specified in 6.2.2 to 6.2.4 in relation to a prosthesis which is standing on the ground in an upright position.
If a test sample is not in the upright position, the axes of the coordinate system shall be rotated to correspond. 6.2.2 The u-axis extends from the origin 0 of the coordinate systems (see Figure 1) and passes through the effective ankle-joint centre and the effective knee-joint centre (see 6.7.3 and 6.7.6 as well as Figure 6). Its positive direction is upwards (in the proximal direction).
6.2.3 The o-axis extends from the origin 0 perpendicular to the u-axis (see Figure 1) and parallel to the effective knee-joint centreline (see 6.7.5 and Figure 6). Its positive direction is outward (in the lateral direction), which is to the left for a left prosthesis and, with the related mirrored values, to the right for a right prosthesis.
6.2.4 The f-axis extends from the origin 0 perpendicular to both the o-axis and the u-axis (see Figure 1). Its positive direction is forward towards the toe (in the anterior direction).
6.3 Reference planes
6.3.1 General
The reference planes (see Figures 2 and 3) shall be parallel planes perpendicular to the u-axis. They are specified in 6.3.2 to 6.3.5.
NOTE The reference planes specified in 6.3.2 to 6.3.5 also contain reference lines which relate to Annex B.
6.3.2 Top reference plane, T
The top reference plane, T, is located at a distance u = u from the origin. It contains the top load application T point P (see 6.4).
T
6.3.3 Knee reference plane, K
The knee reference plane, K, is located at a distance u = u from the origin. It contains the knee load reference K point P (see 6.4) and the effective knee-joint centre (see 6.7.6). K
6.3.4 Ankle reference plane, A
The ankle reference plane, A, is located at a distance u = u from the origin. It contains the ankle load A reference point P (see 6.4) and the effective ankle-joint centre (see 6.7.3). A NOTE Connectors or ankle-joint units, connecting the ankle-foot unit to proximal elements, can be located in positions different to the effective ankle-joint centre.
oSIST prEN ISO 10328:2025 ISO/DIS 10328:2025(en)
6.3.5 Bottom reference plane, B
The bottom reference plane, B, is located at a distance u = u from the origin. It contains the bottom load B application point P (see 6.4).
B Key 1 right 2 left 0 origin f forward o outward u upward Figure 1 — Coordinate systems for right and left-sided application Key 1 right 2 left 3 top reference plane, T 4 knee reference plane, K 5 ankle reference plane, A 6 bottom reference plane, B Figure 2 — Coordinate systems according to Figure 1 with reference planes
oSIST prEN ISO 10328:2025 ISO/DIS 10328:2025(en) Key 1 right leg 2 left leg 3 top reference plane, T 4 load line 5 knee reference plane, K 6 ankle reference plane, A 7 bottom reference plane, B P top load application point T P knee load reference point K P ankle load reference point A P bottom load application point B NOTE This figure illustrates a typical test loading condition representative of the condition of forefoot loading during the stance phase of normal walking. It does not illustrate the test loading conditions defined in 7.1.2. Figure 3 — Specific configuration with u = 0, showing coordinate systems with reference planes B (see Figures 1 and 2), reference lines, reference points and test force, F, for right and left-sided application
oSIST prEN ISO 10328:2025 ISO/DIS 10328:2025(en)
6.4 Reference points
The reference points are the points of intersection of the load line (see 6.6) with the reference planes (see Figure 3). The coordinates of the reference points are as follows:
- top load application point, P ( f , o , u ) T T T T
- knee load reference point, P ( f , o , u ) K K K K
- ankle load reference point, P ( f , o , u ) A A A A
- bottom load application point, P ( f , o , u ) B B B B In the subsequent clauses of this document, the f- and o-coordinates are also referred to as OFFSETS (see also 6.8.1).
6.5 Test force
The test force, F, is a single compressive load applied to the bottom and top load application points P and P B T specified in 6.4.
6.6 Load line
The load line is the line of application of the test force, F. It passes through the reference points specified in 6.4.
6.7 Longitudinal axis of the foot and effective joint centres and centrelines
6.7.1 General
In order to align the test sample within the appropriate coordinate system (see 6.1 to 6.3) it is necessary to locate a) the longitudinal axis of the foot (see 6.7.2); b) the effective ankle-joint centre (see 6.7.3); c) the effective ankle-joint centreline (see 6.7.4); d) the effective knee-joint centreline (see 6.7.5); e) the effective knee-joint centre (see 6.7.6).
If the location of the longitudinal axis of the foot or any effective joint centre or effective joint centreline is not obvious, the manufacturer/submitter shall provide a diagram or instructions, with justification, identifying its location in relation to the test sample.
6.7.2 Longitudinal axis of the foot
Unless otherwise specified by the manufacturer/submitter, the longitudinal axis of the foot shall be taken to pass through the centre of the widest part of the forefoot and equidistant between the medial and lateral borders of the foot at a quarter of the length of the foot from the most posterior part of the foot with the foot placed as specified in 6.7.3.3 and illustrated in Figure 4.
6.7.3 Effective ankle-joint centre
6.7.3.1 Locate the effective ankle-joint centre as described in 6.7.3.2 to 6.7.3.4. See also note in 6.2.2. NOTE The position of a mechanical axle for plantar- and dorsiflexion (if present) is irrelevant to the alignment of the test sample within the appropriate coordinate system.
...