Medical electrical equipment — Part 2-85: Particular requirements for the basic safety and essential performance of cerebral tissue oximeter equipment

This document applies to basic safety and essential performance of cerebral tissue oximeter equipment, that employs light at multiple wavelengths to derive a quantitative measure of oxygen saturation of haemoglobin within the volume of tissue sampled under the probe attached to the head. The cerebral tissue oximeter equipment can be based on continuous light, frequency domain or time domain technologies. This document applies to ME equipment used in a hospital environment as well as when used outside the hospital environment, such as in ambulances and air transport. Additional standards may apply to ME equipment for those environments of use. NOTE 1 Cerebral tissue oximeters are sometimes referred to as near infrared spectroscopy equipment in medical literature. Not included within the scope of this document are: invasive tissue or vascular oximeters; oximeters that require a blood sample from the patient; equipment measuring dissolved oxygen; ME equipment, or part thereof, that measures path-length-dependent haemoglobin change. The requirements for functional near-infrared spectroscopy equipment are found in ISO 80601-2-71[4]; ME equipment, or part thereof, that measures arterial saturation based on pulsatile changes in tissue optical properties (SpO2). The requirements for pulse oximeter equipment are found in ISO 80601‑2‑61[3]; ME equipment, or any part thereof, that claims to monitor tissue in parts of the body other than the head. This document also applies to cerebral tissue oximeter equipment, including cerebral tissue oximeter monitors, cerebral tissue oximeter probes and probe cable extenders, that have been remanufactured. If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems only, the title and content of that clause or subclause will say so. If that is not the case, the clause or subclause applies both to ME equipment and to ME systems, as relevant. Hazards inherent in the intended physiological function of ME equipment or ME systems within the scope of this document are not covered by specific requirements in this document except in 201.11 and in 201.7.2.13 and 201.8.4.1 of the general standard. NOTE 2 See also 4.2 of the general standard. This document can also be applied to ME equipment and their accessories used for compensation or alleviation of disease, injury or disability. This document is not applicable to remote or slave (secondary) equipment that displays StO2 values that are located outside of the patient environment. NOTE 3 ME equipment that provides selection between diagnostic and monitoring functions is expected to meet the requirements of the appropriate document when configured for that function.

Appareils électromédicaux — Partie 2-85: Exigences particulières pour la sécurité de base et les performances essentielles des oxymètres pour tissu cérébral

Le présent document s'applique à la sécurité de base et aux performances essentielles des oxymètres pour tissu cérébral, qui utilisent la lumière à plusieurs longueurs d'onde afin d'obtenir une mesure quantitative de la saturation en oxygène de l'hémoglobine dans le volume de tissu échantillonné sous le capteur fixé à la tête. L'oxymètre pour tissu cérébral peut s'appuyer sur des technologies de lumière continue, de domaine fréquentiel ou de domaine temporel. Le présent document s'applique aux appareils EM utilisés en environnement hospitalier ou en dehors de l'environnement hospitalier, par exemple dans des ambulances ou lors d'un transport aérien. D'autres normes peuvent s'appliquer aux appareils EM pour ces environnements d'utilisation. NOTE 1 Les oxymètres pour tissu cérébral sont parfois appelés « appareils de spectroscopie dans le proche infrarouge » ou « appareils NIRS » (near infrared spectroscopy) dans la littérature médicale. Les éléments suivants ne relèvent pas du domaine d'application du présent document : les oxymètres tissulaires ou vasculaires invasifs ; les oxymètres nécessitant un échantillon sanguin du patient ; les appareils de mesure de l'oxygène dissous ; les appareils EM, ou parties de ceux-ci, qui mesurent le changement de l'hémoglobine en fonction de la longueur du trajet. Les exigences relatives aux appareils de spectroscopie dans le proche infrarouge figurent dans l'ISO 80601-2-71 [4] ; les appareils EM, ou parties de ceux-ci, qui mesurent la saturation artérielle en fonction des changements pulsatiles dans les propriétés optiques des tissus (SpO2). Les exigences relatives aux oxymètres de pouls figurent dans l'ISO 80601-2-61 [3] ; les appareils EM, ou parties de ceux-ci, qui revendiquent la surveillance des tissus dans des parties du corps autres que la tête. Le présent document s'applique également aux oxymètres pour tissu cérébral, y compris les moniteurs d'oxymètres pour tissu cérébral, les capteurs d'oxymètres pour tissu cérébral et les câbles de raccordement du capteur, qui ont fait l'objet d'une remise à neuf. Si un article ou un paragraphe est spécifiquement destiné à être appliqué uniquement aux appareils EM ou uniquement aux systèmes EM, le titre et le contenu dudit article ou paragraphe l'indiqueront. Si cela n'est pas le cas, l'article ou le paragraphe s'applique à la fois aux appareils EM et aux systèmes EM, selon le cas. Les dangers inhérents à la fonction physiologique prévue des appareils EM ou des systèmes EM dans le cadre du domaine d'application du présent document ne sont pas couverts par des exigences spécifiques contenues dans le présent document, à l'exception de 201.11, 201.7.2.13 et 201.8.4.1 de la norme générale. NOTE 2 Voir également 4.2 de la norme générale. Le présent document peut également être appliqué aux appareils EM et leurs accessoires utilisés pour l'atténuation d'une maladie, la compensation ou l'atténuation d'une blessure ou d'une incapacité. Le présent document ne s'applique pas aux appareils à distance ou esclaves (secondaires) qui affichent les valeurs de StO2 et qui sont situés en dehors de l'environnement du patient. NOTE 3 Il est attendu que les appareils EM qui offrent le choix entre des fonctions de diagnostic et de surveillance respectent les exigences du document approprié lorsqu'ils sont configurés pour cette fonction.

General Information

Status
Published
Publication Date
25-Mar-2021
Current Stage
6060 - International Standard published
Start Date
26-Mar-2021
Due Date
24-Feb-2021
Completion Date
26-Mar-2021
Ref Project

Relations

Buy Standard

Standard
ISO 80601-2-85:2021 - Medical electrical equipment
English language
107 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 80601-2-85:Version 13-okt-2020 - Medical electrical equipment
English language
108 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 80601-2-85:2021 - Appareils électromédicaux
French language
128 pages
sale 15% off
Preview
sale 15% off
Preview
Standard
ISO 80601-2-85:Version 28-nov-2020 - Appareils électromédicaux
French language
126 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

INTERNATIONAL ISO
STANDARD 80601-2-85
First edition
2021-03
Medical electrical equipment —
Part 2-85:
Particular requirements for the basic
safety and essential performance of
cerebral tissue oximeter equipment
Appareils électromédicaux —
Partie 2-85: Exigences particulières pour la sécurité de base et les
performances essentielles des oxymètres pour tissu cérébral
Reference number
ISO 80601-2-85:2021(E)
©
ISO 2021

---------------------- Page: 1 ----------------------
ISO 80601-2-85:2021(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2021
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2021 – All rights reserved

---------------------- Page: 2 ----------------------
ISO 80601-2-85:2021(E)
Contents Page

Foreword . vi
Introduction . vii
201.1 Scope, object and related standards . 1
201.1.1 * Scope . 1
201.1.2 Object . 2
201.1.3 Collateral standards . 2
201.1.4 Particular standards . 3
201.2 Normative references . 4
201.3 Terms and definitions . 5
201.4 General requirements . 10
201.4.3.101 * Additional requirements for essential performance . 11
201.4.102  Additional requirements for acceptance criteria . 11
201.4.103 Additional requirements for cerebral tissue oximeter equipment, parts and
accessories . 11
201.5 General requirements for testing of ME equipment . 12
201.6 Classification of ME equipment and ME systems . 12
201.7 ME equipment identification, marking and documents . 12
201.7.1.101 Information to be supplied by the manufacturer . 12
201.7.2.3 Consult accompanying documents . 12
201.7.2.9.101 IP classification . 12
201.7.2.101 Additional requirements for marking on the outside of ME equipment parts . 13
201.7.4.3 Units of measurement . 13
201.7.9.2 Instructions for use . 13
201.7.9.2.1.101 Additional general requirements. 13
201.7.9.2.2.101 Additional requirements for warnings and safety notices . 15
201.7.9.2.9.101 Additional requirements for operating instructions . 15
201.7.9.2.14.101 Additional requirements for accessories, supplementary equipment, used
material . 15
201.7.9.3.1.101 * Additional general requirements . 16
201.8 Protection against electrical hazards from ME equipment . 16
201.8.3.101 Additional requirements for classification of applied parts . 16
201.8.5.5.1.101  Defibrillation protection . 16
201.8.7.4.7.101  Additional requirements for measurement of the patient leakage current . 16
201.9 Protection against mechanical hazards of ME equipment and ME systems . 17
201.10 Protection against unwanted and excessive radiation hazards . 17
201.10.4 Lasers . 17
201.11 Protection against excessive temperatures and other hazards . 17
201.11.1.2.2  Applied parts not intended to supply heat to a patient . 17
201.11.6.5.101 * Additional requirements for ingress of water or particulate matter into
the ME equipment or ME system. 18
201.11.6.7 Sterilization of ME equipment or ME system . 18
201.11.8.101  Additional requirements for interruption of the power supply/supply mains
to ME equipment . 18
© ISO 2021 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO 80601-2-85:2021(E)
201.11.8.101.1 Technical alarm condition for power supply failure . 18
201.11.8.101.2 Settings and data storage following short interruptions or automatic
switchover . 19
201.11.8.101.3 Operation following long interruptions . 19
201.12 Accuracy of controls and instruments and protection against hazardous outputs. 19
201.12.1.101 * StO accuracy of cerebral tissue oximeter equipment . 19
2
201.12.1.101.1 * Specification . 19
201.12.1.101.2 * Data collection for determination of StO accuracy . 21
2
201.12.1.101.3 * Data analysis for determination of StO accuracy . 22
2
201.12.1.101.4 Characteristics of the study used for determination of StO accuracy . 23
2
201.12.4 Protection against hazardous output . 23
201.12.4.101  * Data update period . 23
201.12.4.102  * Signal inadequacy . 23
201.13 Hazardous situations and fault conditions for ME equipment . 24
201.13.101 Detection of probe faults and probe cable extender faults . 24
201.14 Programmable electrical medical systems (PEMS) . 24
201.15 Construction of ME equipment . 24
201.15.3.5.101 * Additional requirements for rough handling . 25
201.15.3.5.101.1 * Shock and vibration (robustness) . 25
201.15.3.5.101.2 * Shock and vibration for a transit-operable cerebral tissue oximeter
during operation . 26
201.15.101 Mode of operation. 27
201.16 ME systems. 27
201.17 Electromagnetic compatibility of ME equipment and ME systems. 27
201.101 * Cerebral tissue oximeter probes and probe cable extenders . 27
201.101.1 General . 27
201.101.2 Labelling . 28
201.102 Functional connection . 28
201.102.1 General . 28
201.102.2 * Connection to an electronic health record or integrated clinical environment . 28
201.102.3 Connection to a distributed alarm system . 28
202 Electromagnetic disturbances — Requirements and tests . 29
202.4.3.1 Configurations . 29
202.5.2.2.1 Requirements applicable to all ME equipment and ME systems . 29
202.8.1.101 Additional general requirements . 29
202.8.2 Patient physiological simulation . 29
206 Usability . 30
208 General requirements, tests and guidance for alarm systems in medical electrical
equipment and medical electrical systems. 30
208.6.1.2.101 * Additional requirements for alarm condition priority . 30
208.6.5.4.101 * Additional requirements for default alarm preset . 31
208.6.8.5.101 Additional requirements for alarm signal inactivation states, indication and
access . 31
211 Requirements for medical electrical equipment and medical electrical systems used
in the home healthcare environment . 31
212 Requirements for medical electrical equipment and medical electrical systems used
in the emergency medical services environment . 31
iv © ISO 2021 – All rights reserved

---------------------- Page: 4 ----------------------
ISO 80601-2-85:2021(E)
Annex C (informative) Guide to marking and labelling requirements for ME equipment and
ME systems . 32
Annex D (informative) Symbols on marking. 36
Annex AA (informative) Particular guidance and rationale . 37
Annex BB (informative) Skin temperature at the cerebral tissue oximeter probe . 48
Annex CC (informative) Determination of accuracy . 50
Annex DD (informative) Characteristics of a tissue haemoglobin phantom for the verification
of the accuracy of cerebral tissue oximeter equipment . 56
Annex EE (informative) Guideline for evaluating and documenting StO accuracy in human
2
subjects . 66
Annex FF (informative) Functional testers for cerebral tissue oximeter equipment . 72
Annex GG (informative) Concepts of ME equipment response time . 75
Annex HH (normative) Data interface requirements . 80
Annex II (informative) Comparison of methods of performance evaluation . 84
Annex JJ (informative) Reference to the IMDRF essential principles and labelling guidances . 89
Annex KK (informative) Reference to the essential principles . 92
Annex LL (informative) Reference to the general safety and performance requirements . 95
Annex MM (informative) Terminology — alphabetized index of defined terms . 98
Bibliography . 102

© ISO 2021 – All rights reserved v

---------------------- Page: 5 ----------------------
ISO 80601-2-85:2021(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national
standards bodies (ISO member bodies). The work of preparing International Standards is normally
carried out through ISO technical committees. Each member body interested in a subject for which a
technical committee has been established has the right to be represented on that committee.
International organizations, governmental and non-governmental, in liaison with ISO, also take part in
the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all
matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see
www.iso.org/iso/foreword.html.
This document was prepared jointly by Technical Committee ISO/TC 121, Anaesthetic and respiratory
equipment, Subcommittee SC 3, Respiratory devices and related equipment used for patient care, and
Technical Committee IEC/TC 62, Electrical equipment in medical practice, Subcommittee 62D,
Electromedical equipment, in collaboration with the European Committee for Standardization (CEN)
Technical Committee CEN/TC 215, Respiratory and anaesthetic equipment, in accordance with the
Agreement on technical cooperation between ISO and CEN (Vienna Agreement).
A list of all parts in the ISO and IEC 80601 series can be found on the ISO and IEC websites.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
vi © ISO 2021 – All rights reserved

---------------------- Page: 6 ----------------------
ISO 80601-2-85:2021(E)
Introduction
The estimation of blood oxygen saturation in the brain tissue by cerebral tissue oximetry equipment is
increasingly used in many areas of medicine. This document covers basic safety and essential
performance requirements achievable within the limits of existing technology.
Annex AA contains a rationale for some of the requirements. It is included to provide additional insight
into the reasoning of the committees that led to a requirement and into the hazards that the
requirement addresses.
Annex BB is a literature review and provides recommendations relevant to determining the maximum
safe temperature of the interface between a cerebral tissue oximeter probe and a patient's tissue.
Annex CC discusses both the formulae used to evaluate the StO accuracy of cerebral tissue oximeter
2
equipment measurements, and the names that are assigned to those formulae.
Annex DD presents guidance on using in-vitro methods (phantoms) for verification of StO accuracy of
2
cerebral tissue oximeter equipment.
Annex EE presents a guideline for an in-vivo (human subjects) controlled desaturation study for the
verification of StO accuracy of cerebral tissue oximeter equipment.
2
Annex FF is a description of functional testers for use with cerebral tissue oximeter equipment.
Annex GG describes concepts of cerebral tissue oximeter equipment response time.
Annex HH describes data interface requirements.
Annex II is a comparison between human desaturations (in-vivo) and tissue haemoglobin phantom
desaturations (in-vitro) for assessing StO accuracy.
2
In this document, the following print types are used:
— requirements and definitions: roman type;
— Instructions, test specifications and terms defined in Clause 3 of the general standard, in this document
or as noted: italic type;
— informative material appearing outside of tables, such as notes, examples and references: in smaller
type; normative text of tables is also in a smaller type.
In referring to the structure of this document, the term
— “clause” means one of the numbered divisions within the table of contents, inclusive of all
subdivisions (e.g. Clause 201.7 includes subclauses 201.7.1, 201.7.2) and
— “subclause” means a numbered subdivision of a clause (e.g. 201.7.1, 7.2 and 201.7.2.1 are all
subclauses of Clause 201.7).
References to clauses within this document are preceded by the term “Clause” followed by the clause
number. References to subclauses within this document are by number only.
© ISO 2021 – All rights reserved vii

---------------------- Page: 7 ----------------------
ISO 80601-2-85:2021(E)
In this document, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
For the purposes of this document, the auxiliary verb:
— “shall” means that conformance with a requirement or a test is mandatory for conformance with
this document;
— “should” means that conformance with a requirement or a test is recommended but is not
mandatory for conformance with this document; and
— “may” is used to describe permission (e.g. a permissible way to achieve conformance with a
requirement or test);
— "can" is used to describe a possibility or capability; and
— "must" is used to express an external constraint.
Annex C contains a guide to the marking and labelling requirements in this document.
Annex D contains a summary of the symbols referenced in this document.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title indicates
that there is guidance or rationale related to that item in AA.

viii © ISO 2021 – All rights reserved

---------------------- Page: 8 ----------------------
INTERNATIONAL STANDARD ISO 80601-2-85:2021(E)

Medical electrical equipment —
Part 2-85:
Particular requirements for the basic safety and essential
performance of cerebral tissue oximeter equipment
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows.
NOTE The general standard is IEC 60601-1:2005+AMD1:2012+AMD2:2020.
201.1.1 * Scope
Replacement:
This document applies to basic safety and essential performance of cerebral tissue oximeter equipment,
that employs light at multiple wavelengths to derive a quantitative measure of oxygen saturation of
haemoglobin within the volume of tissue sampled under the probe attached to the head. The cerebral
tissue oximeter equipment can be based on continuous light, frequency domain or time domain
technologies. This document applies to ME equipment used in a hospital environment as well as when
used outside the hospital environment, such as in ambulances and air transport. Additional standards
may apply to ME equipment for those environments of use.
NOTE 1 Cerebral tissue oximeters are sometimes referred to as near infrared spectroscopy equipment in medical
literature.
Not included within the scope of this document are:
 invasive tissue or vascular oximeters;
 oximeters that require a blood sample from the patient;
 equipment measuring dissolved oxygen;
 ME equipment, or part thereof, that measures path-length-dependent haemoglobin change. The
[4]
requirements for functional near-infrared spectroscopy equipment are found in ISO 80601-2-71 ;
 ME equipment, or part thereof, that measures arterial saturation based on pulsatile changes in
tissue optical properties (SpO ). The requirements for pulse oximeter equipment are found in
2
[3]
ISO 80601-2-61 ;
 ME equipment, or any part thereof, that claims to monitor tissue in parts of the body other than the
head.
This document also applies to cerebral tissue oximeter equipment, including cerebral tissue oximeter
monitors, cerebral tissue oximeter probes and probe cable extenders, that have been remanufactured.
© ISO 2021 – All rights reserved 1

---------------------- Page: 9 ----------------------
ISO 80601-2-85:2021(E)
If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems
only, the title and content of that clause or subclause will say so. If that is not the case, the clause or
subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the
scope of this document are not covered by specific requirements in this document except in 201.11 and
in 201.7.2.13 and 201.8.4.1 of the general standard.
NOTE 2 See also 4.2 of the general standard.
This document can also be applied to ME equipment and their accessories used for compensation or
alleviation of disease, injury or disability.
This document is not applicable to remote or slave (secondary) equipment that displays StO values
2
that are located outside of the patient environment.
NOTE 3 ME equipment that provides selection between diagnostic and monitoring functions is expected to meet
the requirements of the appropriate document when configured for that function.
201.1.2 Object
Replacement:
The object of this document is to establish particular basic safety and essential performance
requirements for cerebral tissue oximeter equipment [as defined in 201.3.202] and its accessories.
NOTE 1 Accessories are included because the combination of the cerebral tissue oximeter monitor and the
accessories needs to be adequately safe. Accessories can have a significant impact on the basic safety or essential
performance of cerebral tissue oximeter equipment.
NOTE 2 This document has been prepared to address the relevant International Medical Device Regulators Forum
(IMDRF) essential principles and labelling guidances as indicated in Annex JJ.
NOTE 3 This document has been prepared to address the relevant essential principles of safety and performance of
ISO 16142-1:2016 as indicated in Annex KK.
NOTE 4 This document has been prepared to address the relevant general safety and performance requirements of
[20]
European regulation (EU) 2017/745 as indicated in Annex LL.
201.1.3 Collateral standards
Addition:
This document refers to those applicable collateral standards that are listed in Clause 2 of the general
standard and Clause 201.2 of this document.
IEC 60601-1-2:2014+AMD1:2020, IEC 60601-1-6:2010+AMD1:2013+AMD2:2020,
IEC 60601-1-8:2006+AMD1:2012+AMD2:2020, IEC 60601-1-11:2015+AMD1:2020 and
IEC 60601-1-12:2014+AMD1:2020 apply as modified in Clauses 202, 206, 208, 211 and 212
respectively. IEC 60601-1-3 does not apply. All other published collateral standards in the IEC 60601-1
series apply as published.
2 © ISO 2021 – All rights reserved

---------------------- Page: 10 ----------------------
ISO 80601-2-85:2021(E)
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards define basic safety and essential performance
requirements, and may modify, replace or delete requirements contained in the general standard,
including the collateral standards, as appropriate for the particular ME equipment under consideration.
A requirement of a particular standard takes priority over the general standard or the collateral
standards.
For brevity, IEC 60601-1:2005+AMD1:2012+AMD2:2020 is referred to in this document as the general
standard. Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this document corresponds to those of the general
standard with the prefix “201” (e.g. 201.1 in this document addresses the content of Clause 1 of the
general standard) or applicable collateral standard with the prefix “2xx” where xx is the final digits of
the collateral standard document number (e.g. 202.4 in this document addresses the content of Clause 4
of the IEC 60601-1-2 collateral standard, 208.4 in this document addresses the content of Clause 4 of
the IEC 60601-1-8 collateral standard, etc.). The changes to the text of the general standard are
specified by the use of the following words:
"Replacement" means that the clause or subclause of the general standard or applicable collateral
standard is replaced completely by the text of this document.
"Addition" means that the text of this document is additional to the requirements of the general
standard or applicable collateral standard.
"Amendment" means that the clause or subclause of the general standard or applicable collateral
standard is amended as indicated by the text of this document.
Clauses, subclauses or figures that are additional to those of the general standard are numbered starting
from 201.101. However, due to the fact that definitions in t
...

FINAL
INTERNATIONAL ISO/FDIS
DRAFT
STANDARD 80601-2-85
ISO/TC 121/SC 3
Medical electrical equipment —
Secretariat: ANSI
Voting begins on:
Part 2-85:
2020-10-23
Particular requirements for the basic
Voting terminates on:
safety and essential performance of
2020-12-18
cerebral tissue oximeter equipment
Appareils électromédicaux —
Partie 2-85: Exigences particulières pour la sécurité de base et les
performances essentielles des oxymètres pour tissu cérébral
This draft is submitted to a parallel vote in ISO and in IEC.
ISO/CEN PARALLEL PROCESSING
RECIPIENTS OF THIS DRAFT ARE INVITED TO
SUBMIT, WITH THEIR COMMENTS, NOTIFICATION
OF ANY RELEVANT PATENT RIGHTS OF WHICH
THEY ARE AWARE AND TO PROVIDE SUPPOR TING
DOCUMENTATION.
IN ADDITION TO THEIR EVALUATION AS
Reference number
BEING ACCEPTABLE FOR INDUSTRIAL, TECHNO-
ISO/FDIS 80601-2-85:2020(E)
LOGICAL, COMMERCIAL AND USER PURPOSES,
DRAFT INTERNATIONAL STANDARDS MAY ON
OCCASION HAVE TO BE CONSIDERED IN THE
LIGHT OF THEIR POTENTIAL TO BECOME STAN-
DARDS TO WHICH REFERENCE MAY BE MADE IN
©
NATIONAL REGULATIONS. ISO 2020

---------------------- Page: 1 ----------------------
ISO/FDIS 80601-2-85:2020(E)

COPYRIGHT PROTECTED DOCUMENT
© ISO 2020
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2020 – All rights reserved

---------------------- Page: 2 ----------------------
ISO/FDIS 80601-2-85:2020(E)
Contents Page

Foreword . vi
Introduction . vii
201.1 Scope, object and related standards . 1
201.1.1 * Scope . 1
201.1.2 Object . 2
201.1.3 Collateral standards . 2
201.1.4 Particular standards . 3
201.2 Normative references . 4
201.3 Terms and definitions . 5
201.4 General requirements . 10
201.4.3.101 * Additional requirements for essential performance . 11
201.4.102  Additional requirements for acceptance criteria . 11
201.4.103 Additional requirements for cerebral tissue oximeter equipment, parts and
accessories . 11
201.5 General requirements for testing of ME equipment . 12
201.6 Classification of ME equipment and ME systems . 12
201.7 ME equipment identification, marking and documents . 12
201.7.1.101 Information to be supplied by the manufacturer . 12
201.7.2.3 Consult accompanying documents . 12
201.7.2.9.101 IP classification . 12
201.7.2.101 Additional requirements for marking on the outside of ME equipment parts . 13
201.7.4.3 Units of measurement . 13
201.7.9.2 Instructions for use . 13
201.7.9.2.1.101 Additional general requirements. 13
201.7.9.2.2.101 Additional requirements for warnings and safety notices . 15
201.7.9.2.9.101 Additional requirements for operating instructions . 15
201.7.9.2.14.101 Additional requirements for accessories, supplementary equipment, used
material . 15
201.7.9.3.1.101 * Additional general requirements . 16
201.8 Protection against electrical hazards from ME equipment . 16
201.8.3.101 Additional requirements for classification of applied parts . 16
201.8.5.5.1.101  Defibrillation protection . 16
201.8.7.4.7.101  Additional requirements for measurement of the patient leakage current . 16
201.9 Protection against mechanical hazards of ME equipment and ME systems . 17
201.10 Protection against unwanted and excessive radiation hazards . 17
201.10.4 Lasers . 17
201.11 Protection against excessive temperatures and other hazards . 17
201.11.1.2.2  Applied parts not intended to supply heat to a patient . 17
201.11.6.5.101 * Additional requirements for ingress of water or particulate matter into
the ME equipment or ME system. 18
201.11.6.7 Sterilization of ME equipment or ME system . 18
201.11.8.101  Additional requirements for interruption of the power supply/supply mains
to ME equipment . 18
© ISO 2020 – All rights reserved iii

---------------------- Page: 3 ----------------------
ISO/FDIS 80601-2-85:2020(E)
201.11.8.101.1 Technical alarm condition for power supply failure . 18
201.11.8.101.2 Settings and data storage following short interruptions or automatic
switchover . 19
201.11.8.101.3 Operation following long interruptions . 19
201.12 Accuracy of controls and instruments and protection against hazardous outputs. 19
201.12.1.101 * StO accuracy of cerebral tissue oximeter equipment . 19
2
201.12.1.101.1 * Specification . 19
201.12.1.101.2 * Data collection for determination of StO accuracy . 21
2
201.12.1.101.3 * Data analysis for determination of StO accuracy . 22
2
201.12.1.101.4 Characteristics of the study used for determination of StO accuracy . 23
2
201.12.4 Protection against hazardous output . 23
201.12.4.101  * Data update period . 23
201.12.4.102  * Signal inadequacy . 23
201.13 Hazardous situations and fault conditions for ME equipment . 24
201.13.101 Detection of probe faults and probe cable extender faults . 24
201.14 Programmable electrical medical systems (PEMS) . 24
201.15 Construction of ME equipment . 24
201.15.3.5.101 * Additional requirements for rough handling . 25
201.15.3.5.101.1 * Shock and vibration (robustness) . 25
201.15.3.5.101.2 * Shock and vibration for a transit-operable cerebral tissue oximeter
during operation . 26
201.15.101 Mode of operation. 27
201.16 ME systems. 27
201.17 Electromagnetic compatibility of ME equipment and ME systems. 27
201.101 * Cerebral tissue oximeter probes and probe cable extenders . 27
201.101.1 General . 27
201.101.2 Labelling . 28
201.102 Functional connection . 28
201.102.1 General . 28
201.102.2 * Connection to an electronic health record or integrated clinical environment . 28
201.102.3 Connection to a distributed alarm system . 28
202 Electromagnetic disturbances — Requirements and tests . 29
202.4.3.1 Configurations . 29
202.5.2.2.1 Requirements applicable to all ME equipment and ME systems . 29
202.8.1.101 Additional general requirements . 29
202.8.2 Patient physiological simulation . 29
206 Usability . 30
208 General requirements, tests and guidance for alarm systems in medical electrical
equipment and medical electrical systems. 30
208.6.1.2.101 * Additional requirements for alarm condition priority . 30
208.6.5.4.101 * Additional requirements for default alarm preset . 31
208.6.8.5.101 Additional requirements for alarm signal inactivation states, indication and
access . 31
211 Requirements for medical electrical equipment and medical electrical systems used
in the home healthcare environment . 31
212 Requirements for medical electrical equipment and medical electrical systems used
in the emergency medical services environment . 31
iv © ISO 2020 – All rights reserved

---------------------- Page: 4 ----------------------
ISO/FDIS 80601-2-85:2020(E)
Annex C (informative) Guide to marking and labelling requirements for ME equipment and
ME systems . 32
Annex D (informative) Symbols on marking. 36
Annex AA (informative) Particular guidance and rationale . 37
Annex BB (informative) Skin temperature at the cerebral tissue oximeter probe . 48
Annex CC (informative) Determination of accuracy . 50
Annex DD (informative) Characteristics of a tissue haemoglobin phantom for the verification
of the accuracy of cerebral tissue oximeter equipment . 56
Annex EE (informative) Guideline for evaluating and documenting StO accuracy in human
2
subjects . 66
Annex FF (informative) Functional testers for cerebral tissue oximeter equipment . 72
Annex GG (informative) Concepts of ME equipment response time . 75
Annex HH (normative) Data interface requirements . 80
Annex II (informative) Comparison of methods of performance evaluation . 84
Annex JJ (informative) Reference to the IMDRF essential principles and labelling guidances . 89
Annex KK (informative) Reference to the essential principles . 92
Annex LL (informative) Reference to the general safety and performance requirements . 95
Annex MM (informative) Terminology — alphabetized index of defined terms . 98
Bibliography . 102

© ISO 2020 – All rights reserved v

---------------------- Page: 5 ----------------------
ISO/FDIS 80601-2-85:2020(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national
standards bodies (ISO member bodies). The work of preparing International Standards is normally
carried out through ISO technical committees. Each member body interested in a subject for which a
technical committee has been established has the right to be represented on that committee.
International organizations, governmental and non-governmental, in liaison with ISO, also take part in
the work. ISO collaborates closely with the International Electrotechnical Commission (IEC) on all
matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to the
World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see
www.iso.org/iso/foreword.html.
This document was prepared jointly by Technical Committee ISO/TC 121, Anaesthetic and respiratory
equipment, Subcommittee SC 3, Respiratory devices and related equipment used for patient care, and
Technical Committee IEC/TC 62, Electrical equipment in medical practice, Subcommittee 62D,
Electromedical equipment, in collaboration with the European Committee for Standardization (CEN)
Technical Committee CEN/TC 215, Respiratory and anaesthetic equipment, in accordance with the
Agreement on technical cooperation between ISO and CEN (Vienna Agreement).
A list of all parts in the ISO and IEC 80601 series can be found on the ISO and IEC websites.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
vi © ISO 2020 – All rights reserved

---------------------- Page: 6 ----------------------
ISO/FDIS 80601-2-85:2020(E)
Introduction
The estimation of blood oxygen saturation in the brain tissue by cerebral tissue oximetry equipment is
increasingly used in many areas of medicine. This document covers basic safety and essential
performance requirements achievable within the limits of existing technology.
Annex AA contains a rationale for some of the requirements. It is included to provide additional insight
into the reasoning of the committees that led to a requirement the hazards that the requirement
addresses.
Annex BB is a literature survey and provides recommendations relevant to the determination of the
maximum safe temperature of the interface between a cerebral tissue oximeter probe and a patient's
tissue.
Annex CC discusses both the formulae used to evaluate the StO accuracy of cerebral tissue oximeter
2
equipment measurements, and the names that are assigned to those formulae.
Annex DD presents guidance on using in-vitro methods (phantoms) for verification of StO accuracy of
2
cerebral tissue oximeter equipment.
Annex EE presents a guideline for an in-vivo (human subjects) controlled desaturation study for the
verification of StO accuracy of cerebral tissue oximeter equipment.
2
Annex FF is a description of functional testers for use with cerebral tissue oximeter equipment.
Annex GG describes concepts of cerebral tissue oximeter equipment response time.
Annex HH describes data interface requirements.
Annex II is a comparison between human desaturations (in-vivo) and tissue haemoglobin phantom
desaturations (in-vitro) for assessing StO accuracy.
2
In this document, the following print types are used:
— requirements and definitions: roman type;
— Instructions, test specifications and terms defined in Clause 3 of the general standard, in this document
or as noted: italic type;
— informative material appearing outside of tables, such as notes, examples and references: in smaller
type; normative text of tables is also in a smaller type.
In referring to the structure of this document, the term
— “clause” means one of the numbered divisions within the table of contents, inclusive of all
subdivisions (e.g. Clause 201.7 includes subclauses 201.7.1, 201.7.2) and
— “subclause” means a numbered subdivision of a clause (e.g. 201.7.1, 7.2 and 201.7.2.1 are all
subclauses of Clause 201.7).
© ISO 2020 – All rights reserved vii

---------------------- Page: 7 ----------------------
ISO/FDIS 80601-2-85:2020(E)
References to clauses within this document are preceded by the term “Clause” followed by the clause
number. References to subclauses within this document are by number only.
In this document, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
combination of the conditions is true.
For the purposes of this document, the auxiliary verb:
— “shall” means that conformance with a requirement or a test is mandatory for conformance with
this document;
— “should” means that conformance with a requirement or a test is recommended but is not
mandatory for conformance with this document; and
— “may” is used to describe permission (e.g. a permissible way to achieve conformance with a
requirement or test);
— "can" is used to describe a possibility or capability; and
— "must" is used to express an external constraint.
Annex C contains a guide to the marking and labelling requirements in this document.
Annex D contains a summary of the symbols referenced in this document.
An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title indicates
that there is guidance or rationale related to that item in AA.

viii © ISO 2020 – All rights reserved

---------------------- Page: 8 ----------------------
FINAL DRAFT INTERNATIONAL STANDARD ISO/FDIS 80601-2-85:2020(E)

Medical electrical equipment —
Part 2-85:
Particular requirements for the basic safety and essential
performance of cerebral tissue oximeter equipment
201.1 Scope, object and related standards
Clause 1 of the general standard applies, except as follows.
NOTE The general standard is IEC 60601-1:2005+AMD1:2012+AMD2:2020.
201.1.1 * Scope
Replacement:
This document applies to basic safety and essential performance of cerebral tissue oximeter equipment,
that employs light at multiple wavelengths to derive a quantitative measure of oxygen saturation of
haemoglobin within the volume of tissue sampled under the probe attached to the head. The cerebral
tissue oximeter equipment can be based on continuous light, frequency domain or time domain
technologies. This document applies to ME equipment used in a hospital environment as well as when
used outside the hospital environment, such as in ambulances and air transport. Additional standards
may apply to ME equipment for those environments of use.
NOTE 1 Cerebral tissue oximeters are sometimes referred to as near infrared spectroscopy equipment in medical
literature.
Not included within the scope of this document are:
 invasive tissue or vascular oximeters;
 oximeters that require a blood sample from the patient;
 equipment measuring dissolved oxygen;
 ME equipment, or part thereof, that measures path-length-dependent haemoglobin change. The
[4]
requirements for functional near-infrared spectroscopy equipment are found in ISO 80601-2-71 ;
 ME equipment, or part thereof, that measures arterial saturation based on pulsatile changes in
tissue optical properties (SpO ). The requirements for pulse oximeter equipment are found in
2
[3]
ISO 80601-2-61 ;
 ME equipment, or any part thereof, that claims to monitor tissue in parts of the body other than the
head.
This document also apply to cerebral tissue oximeter equipment, including cerebral tissue oximeter
monitors, cerebral tissue oximeter probes and probe cable extenders, that gave been remanufactured.
© ISO 2020 – All rights reserved 1

---------------------- Page: 9 ----------------------
ISO/FDIS 80601-2-85:2020(E)
If a clause or subclause is specifically intended to be applicable to ME equipment only, or to ME systems
only, the title and content of that clause or subclause will say so. If that is not the case, the clause or
subclause applies both to ME equipment and to ME systems, as relevant.
Hazards inherent in the intended physiological function of ME equipment or ME systems within the
scope of this document are not covered by specific requirements in this document except in 201.11 and
in 201.7.2.13 and 201.8.4.1 of the general standard.
NOTE 2 See also 4.2 of the general standard.
This document can also be applied to ME equipment and their accessories used for compensation or
alleviation of disease, injury or disability.
This document is not applicable to remote or slave (secondary) equipment that displays StO values
2
that are located outside of the patient environment.
NOTE 3 ME equipment that provides selection between diagnostic and monitoring functions is expected to meet
the requirements of the appropriate document when configured for that function.
201.1.2 Object
Replacement:
The object of this document is to establish particular basic safety and essential performance
requirements for cerebral tissue oximeter equipment [as defined in 201.3.202] and its accessories.
NOTE 1 Accessories are included because the combination of the cerebral tissue oximeter monitor and the
accessories needs to be adequately safe. Accessories can have a significant impact on the basic safety or essential
performance of cerebral tissue oximeter equipment.
NOTE 2 This document has been prepared to address the relevant International Medical Device Regulators Forum
(IMDRF) essential principles and labelling guidances as indicated in Annex JJ.
NOTE 3 This document has been prepared to address the relevant essential principles of safety and performance of
ISO 16142-1:2016 as indicated in Annex KK.
NOTE 4 This document has been prepared to address the relevant general safety and performance requirements of
[20]
European regulation (EU) 2017/745 as indicated in Annex LL.
201.1.3 Collateral standards
Addition:
This document refers to those applicable collateral standards that are listed in Clause 2 of the general
standard and Clause 201.2 of this document.
IEC 60601-1-2:2014+AMD1:2020, IEC 60601-1-6:2010+AMD1:2013+AMD2:2020,
IEC 60601-1-8:2006+AMD1:2012+AMD2:2020, IEC 60601-1-11:2015+AMD1:2020 and
IEC 60601-1-12:2014+AMD1:2020 apply as modified in Clauses 202, 206, 208, 211 and 212
respectively. IEC 60601-1-3 does not apply. All other published collateral standards in the IEC 60601-1
series apply as published.
2 © ISO 2020 – All rights reserved

---------------------- Page: 10 ----------------------
ISO/FDIS 80601-2-85:2020(E)
201.1.4 Particular standards
Replacement:
In the IEC 60601 series, particular standards define basic safety and essential performance
requirements, and may modify, replace or delete requirements contained in the general standard,
including the collateral standards, as appropriate for the particular ME equipment under consideration.
A requirement of a particular standard takes priority over the general standard or the collateral
standards.
For brevity, IEC 60601-1:2005+AMD1:2012+AMD2:2020 is referred to in this document as the general
standard. Collateral standards are referred to by their document number.
The numbering of clauses and subclauses of this document corresponds to those of the general
standard with the prefix “201” (e.g. 201.1 in this document addresses the content of Clause 1 of the
general standard) or applicable collateral standard with the prefix “2xx” where xx is the final digits of
the collateral standard document number (e.g. 202.4 in this document addresses the content of Clause 4
of the IEC 60601-1-2 collateral standard, 208.4 in this document addresses the content of Clause 4 of
the IEC 60601-1-8 collateral standard, etc.). The
...

NORME ISO
INTERNATIONALE 80601-2-85
Première édition
2021-03
Appareils électromédicaux —
Partie 2-85:
Exigences particulières pour la
sécurité de base et les performances
essentielles des oxymètres pour tissu
cérébral
Medical electrical equipment —
Part 2-85: Particular requirements for the basic safety and essential
performance of cerebral tissue oximeter equipment
Numéro de référence
ISO 80601-2-85:2021(F)
©
ISO 2021

---------------------- Page: 1 ----------------------
ISO 80601-2-85:2021(F)

DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2021
Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en œuvre, aucune partie de cette
publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut
être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Genève
Tél.: +41 22 749 01 11
E-mail: copyright@iso.org
Web: www.iso.org
Publié en Suisse
ii © ISO 2021 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO 80601-2-85:2021(F)
Sommaire Page
Avant-propos . vi
Introduction . vii
201.1 Domaine d'application, objet et normes connexes . 1
201.1.1 * Domaine d'application . 1
201.1.2 Objet . 2
201.1.3 Normes collatérales . 2
201.1.4 Normes particulières . 3
201.2 Références normatives. 4
201.3 Termes et définitions . 5
201.4 Exigences générales . 12
201.4.3.101 * Exigences supplémentaires pour les performances essentielles . 12
201.4.102 Exigences supplémentaires relatives aux critères d'acceptation . 13
201.4.103 Exigences supplémentaires relatives à l'oxymètre pour tissu cérébral, ainsi que ses
parties et accessoires . 13
201.5 Exigences générales relatives aux essais des appareils EM . 14
201.6 Classification des appareils EM et des systèmes EM . 14
201.7 Identification, marquage et documentation des appareils EM . 14
201.7.1.101 Informations devant être fournies par le fabricant . 14
201.7.2.3 Consultation des documents d'accompagnement . 14
201.7.2.9.101 Classification IP . 15
201.7.2.101 Exigences supplémentaires de marquage sur l'extérieur des parties
d'appareils EM . 15
201.7.4.3 Unités de mesure . 15
201.7.9.2 Instructions d’utilisation . 15
201.7.9.2.1.101 Exigences générales supplémentaires . 16
201.7.9.2.2.101 Exigences supplémentaires pour les avertissements et les consignes de
sécurité . 17
201.7.9.2.9.101 Exigences supplémentaires pour les instructions de fonctionnement . 17
201.7.9.2.14.101 Exigences supplémentaires pour les accessoires, équipements
supplémentaires et fournitures utilisées . 18
201.7.9.3.1.101 * Exigences générales supplémentaires . 18
201.8 Protection contre les dangers d'origine électrique provenant d'appareils EM . 18
201.8.3.101 Exigences supplémentaires pour la classification des parties appliquées . 19
201.8.5.5.1.101 Protection contre la défibrillation. 19
201.8.7.4.7.101 Exigences supplémentaires relatives au mesurage du courant de fuite du
patient . 19
201.9 Protection contre les dangers mécaniques des appareils EM et systèmes EM . 19
201.10 Protection contre les dangers dus aux rayonnements involontaires ou excessifs . 19
201.10.4 Lasers . 19
201.11 Protection contre les températures excessives et les autres dangers . 20
201.11.1.2.2 Parties appliquées non destinées à fournir de la chaleur à un patient . 20
© ISO 2021 – Tous droits réservés
iii

---------------------- Page: 3 ----------------------
ISO 80601-2-85:2021(F)
201.11.6.5.101 * Exigences supplémentaires pour la pénétration d'eau ou de matières
particulaires dans les appareils EM ou les systèmes EM . 21
201.11.6.7 Stérilisation des appareils EM ou des systèmes EM . 21
201.11.8.101 Exigences supplémentaires applicables à la coupure de l'alimentation/du réseau
d'alimentation d'un appareil EM . 21
201.11.8.101.1 Condition d'alarme technique signalant un défaut d'alimentation . 21
201.11.8.101.2 Stockage des paramètres de réglage et des données après des coupures de
courte durée ou un basculement automatique . 22
201.11.8.101.3 Fonctionnement après une coupure de longue durée . 22
201.12 Précision des commandes et des instruments et protection contre les caractéristiques
de sortie dangereuses . 22
201.12.1.101 * Exactitude de la StO d'un oxymètre pour tissu cérébral . 23
2
201.12.1.101.1 * Spécification . 23
201.12.1.101.2 * Collecte de données en vue de la détermination de l'exactitude de la StO . 24
2
201.12.1.101.3 * Analyse de données pour la détermination de l'exactitude de la StO . 26
2
201.12.1.101.4 Caractéristiques de l'étude utilisée pour déterminer l'exactitude de la StO . 27
2
201.12.4 Protection contre les caractéristiques de sortie présentant des risques . 27
201.12.4.101  * Période de mise à jour des données . 27
201.12.4.102 * Inadéquation du signal . 28
201.13 Situations dangereuses et conditions de défaut pour les appareils EM . 28
201.13.101 Détection de défauts du capteur et de défauts du câble de raccordement du capteur 28
201.14 Systèmes électromédicaux programmables (SEMP) . 29
201.15 Construction de l'appareil EM . 29
201.15.3.5.101 * Exigences supplémentaires relatives à une manutention brutale . 29
201.15.3.5.101.1 * Chocs et vibrations (robustesse) . 29
201.15.3.5.101.2 * Chocs et vibrations pour un oxymètre pour tissu cérébral opérationnel en
déplacement pendant son fonctionnement . 30
201.15.101 Mode de fonctionnement . 31
201.16 Systèmes EM . 32
201.17 Compatibilité électromagnétique des appareils EM et des systèmes EM . 32
201.101 * Capteurs d'oxymètre pour tissu cérébral et câbles de raccordement du capteur . 32
201.101.1 Généralités . 32
201.101.2 Étiquetage . 33
201.102 Connexion fonctionnelle . 33
201.102.1 Généralités . 33
201.102.2 * Raccordement à un dossier informatisé de santé ou à un environnement clinique
intégré . 33
201.102.3 Raccordement à un système d'alarme réparti . 33
202 Perturbations électromagnétiques — Exigences et essais . 34
202.4.3.1 Configurations . 34
202.5.2.2.1 Exigences applicables à tous les appareils EM et systèmes EM . 34
202.8.1.101 Exigences générales supplémentaires . 34
202.8.2 Simulation physiologique du patient . 35
206 Aptitude à l'utilisation . 35
208 Exigences générales, essais et recommandations pour les systèmes d'alarme des
appareils et des systèmes électromédicaux . 36
© ISO 2021 – Tous droits réservés
iv

---------------------- Page: 4 ----------------------
ISO 80601-2-85:2021(F)
208.6.1.2.101 * Exigences supplémentaires relatives à la priorité des conditions d'alarme. 36
208.6.5.4.101 * Exigences supplémentaires relatives au préréglage d'alarme par défaut . 36
208.6.8.5.101 Exigences supplémentaires relatives à l'indication et à l'accès aux états de
désactivation du signal d'alarme . 36
211 Exigences pour les appareils électromédicaux et les systèmes électromédicaux utilisés
dans l'environnement des soins à domicile . 36
212 Exigences pour les appareils électromédicaux et les systèmes électromédicaux destinés
à être utilisés dans l'environnement de services médicaux d'urgence . 37
Annexe C (informative) Guide des exigences de marquage et d'étiquetage pour les
appareils EM et les systèmes EM . 38
Annexe D (informative) Symboles de marquage . 43
Annexe AA (informative) Recommandations particulières et justifications . 44
Annexe BB (informative) Température de la peau au niveau du capteur d'oxymètre pour tissu
cérébral . 58
Annexe CC (informative) Détermination de l'exactitude . 60
Annexe DD (informative) Caractéristiques d'un fantôme d'hémoglobine tissulaire pour la
vérification de l'exactitude de l'oxymètre pour tissu cérébral. 67
Annexe EE (informative) Lignes directrices pour l'évaluation et la documentation de
l'exactitude de la StO chez les sujets humains . 80
2
Annexe FF (informative) Contrôleurs de fonctionnement pour les oxymètres pour tissu cérébral87
Annexe GG (informative) Concepts de temps de réponse des appareils EM . 91
Annexe HH (normative) Exigences applicables à l'interface de données . 96
Annexe II (informative) Comparaison des méthodes d'évaluation de la performance . 101
Annexe JJ (informative) Référence aux principes essentiels et aux recommandations
d'étiquetage de l'IMDRF . 108
Annexe KK (informative) Référence aux principes essentiels . 111
Annexe LL (informative)  Référence aux exigences générales de sécurité et de performances 114
Annexe MM (informative) Terminologie — Index alphabétique des termes définis . 117
Bibliographie. 122
© ISO 2021 – Tous droits réservés
v

---------------------- Page: 5 ----------------------
ISO 80601-2-85:2021(F)
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux
de normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général
confiée aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire
partie du comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement
avec la Commission électrotechnique internationale (IEC) en ce qui concerne la normalisation
électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier de prendre note des différents
critères d'approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2
(voir www.iso.org/directives).
L'attention est attirée sur le fait que certains des éléments du présent document peuvent faire l'objet de
droits de propriété intellectuelle ou de droits analogues. L'ISO ne saurait être tenue pour responsable de
ne pas avoir identifié de tels droits de propriété et averti de leur existence. Les détails concernant les
références aux droits de propriété intellectuelle ou autres droits analogues identifiés lors de l'élaboration
du document sont indiqués dans l'Introduction et/ou dans la liste des déclarations de brevets reçues par
l'ISO (voir www.iso.org/brevets).
Les appellations commerciales éventuellement mentionnées dans le présent document sont données pour
information, par souci de commodité, à l’intention des utilisateurs et ne sauraient constituer un
engagement.
Pour une explication de la nature volontaire des normes, la signification des termes et expressions
spécifiques de l'ISO liés à l'évaluation de la conformité, ou pour toute information au sujet de l'adhésion de
l'ISO aux principes de l’Organisation mondiale du commerce (OMC) concernant les obstacles techniques au
commerce (OTC), voir le lien suivant : www.iso.org/iso/fr/avant-propos.
Le présent document a été élaboré conjointement par le comité technique ISO/TC 121, Matériel
d'anesthésie et de réanimation respiratoire, sous-comité SC 3, Appareils respiratoires et équipements
connexes utilisés pour les soins aux patients, et le comité d'études IEC/TC 62, Équipements électriques dans
la pratique médicale, sous-comité 62 D, Appareils électromédicaux, en collaboration avec le comité
technique CEN/TC 215 du Comité européen de normalisation (CEN), Équipement respiratoire et
anesthésique, conformément à l'Accord de coopération technique entre l'ISO et le CEN (Accord de Vienne).
Une liste de toutes les parties de la série ISO et IEC 80601 se trouve sur les sites web de l'ISO et de l'IEC.
Il convient que l’utilisateur adresse tout retour d’information ou toute question concernant le présent
document à l’organisme national de normalisation de son pays. Une liste exhaustive desdits organismes se
trouve à l’adresse www.iso.org/fr/members.html.
© ISO 2021 – Tous droits réservés
vi

---------------------- Page: 6 ----------------------
ISO 80601-2-85:2021(F)
Introduction
L'estimation de la saturation en oxygène du sang dans les tissus cérébraux par un appareil d'oxymétrie
pour tissu cérébral est de plus en plus utilisée dans de nombreux domaines de la médecine. Le présent
document couvre les exigences de sécurité de base et de performances essentielles réalisables dans les
limites de la technologie existante.
L'Annexe AA justifie certaines de ces exigences. Elle sert à fournir des indications supplémentaires sur le
raisonnement des comités qui a conduit à l’élaboration d’une exigence face aux dangers dont elle traite.
L'Annexe BB est une revue documentaire qui fournit des recommandations pertinentes pour la
détermination de la température maximale de sécurité de l'interface entre un capteur d'oxymètre pour tissu
cérébral et le tissu d'un patient.
L'Annexe CC traite à la fois des formules utilisées pour évaluer l'exactitude des mesurages de la StO des
2
oxymètres pour tissu cérébral, et des noms attribués à ces formules.
L'Annexe DD présente des recommandations concernant l'utilisation de méthodes in vitro (fantômes) pour
vérifier l'exactitude de la StO des oxymètres pour tissu cérébral.
2
L'Annexe EE présente une ligne directrice pour une étude de désaturation contrôlée in vivo (sur des sujets
humains) destinée à la vérification de l'exactitude de la StO2 des oxymètres pour tissu cérébral.
L'Annexe FF est une description des contrôleurs de fonctionnement à utiliser avec les oxymètres pour tissu
cérébral.
L'Annexe GG décrit les concepts de temps de réponse des oxymètres pour tissu cérébral.
L'Annexe HH décrit les exigences applicables à l'interface de données.
L'Annexe II compare des désaturations chez l'être humain (in vivo) et des désaturations de fantôme
d'hémoglobine tissulaire (in vitro) pour évaluer l'exactitude de la StO .
2
Dans le présent document, les types de polices suivants sont utilisés :
 exigences et définitions : caractères romains ;
 instructions, modalités d'essais et termes définis à l'Article 3 de la norme générale, dans le présent
document ou en note : caractères italiques ;
 éléments informatifs situés hors des tableaux, tels que notes, exemples et références : en petits
caractères ; texte normatif des tableaux : petits caractères également.
Concernant la structure de la présente norme, le terme :
 « article » désigne l'une des sections numérotées dans le sommaire, avec toutes ses subdivisions (par
exemple, l'Article 201.7 inclut les paragraphes 201.7.1, 201.7.2, etc.) ; et
 « paragraphe » désigne une subdivision numérotée d'un article (par exemple, 201.7.1, 7.2 et 201.7.2.1
sont des paragraphes appartenant à l'Article 201.7).
© ISO 2021 – Tous droits réservés
vii

---------------------- Page: 7 ----------------------
ISO 80601-2-85:2021(F)
Dans le présent document, les références à des articles sont précédées du terme « Article » suivi du
numéro de l'article concerné. Dans le présent document, les références aux paragraphes utilisent
uniquement le numéro du paragraphe concerné.
Dans le présent document, la conjonction « ou » est utilisée comme « ou inclusif » ; une affirmation est donc
vraie si une combinaison quelconque des conditions est vraie.
Pour les besoins du présent document :
 « doit » (« shall ») signifie que la conformité à une exigence ou à un essai est obligatoire pour la
conformité au présent document ;
 « il convient » (« should ») signifie que la conformité à une exigence ou à un essai est recommandée,
mais n'est pas obligatoire, pour la conformité au présent document ; et
 « peut » (« may ») est utilisé pour décrire la permission (par exemple, une manière autorisée d'obtenir
la conformité à une exigence ou à un essai) ;
 « peut » (« can ») est utilisé pour décrire une possibilité ou une capacité ; et
 « doit » (« must ») est utilisé pour exprimer une contrainte externe.
L'Annexe C contient un guide pour le marquage et les exigences d'étiquetage dans le présent document.
L'Annexe D contient un résumé des symboles référencés dans le présent document.
Lorsqu'un astérisque (*) est utilisé comme premier caractère devant un titre, au début d'un titre de
paragraphe ou de tableau, il indique l'existence d'une recommandation ou d'une justification à consulter à
l'Annexe AA.
© ISO 2021 – Tous droits réservés
viii

---------------------- Page: 8 ----------------------
NORME INTERNATIONALE ISO 80601-2-85:2021(F)

Appareils électromédicaux — Partie 2-85 : Exigences particulières
pour la sécurité de base et les performances essentielles des
oxymètres pour tissu cérébral
201.1 Domaine d'application, objet et normes connexes
L'Article 1 de la norme générale s'applique, à l'exception de ce qui suit :
NOTE La norme générale est l'IEC 60601-1:2005+AMD1:2012+AMD2:2020.
201.1.1 * Domaine d'application
Remplacement :
Le présent document s'applique à la sécurité de base et aux performances essentielles des oxymètres pour
tissu cérébral, qui utilisent la lumière à plusieurs longueurs d'onde afin d'obtenir une mesure
quantitative de la saturation en oxygène de l'hémoglobine dans le volume de tissu échantillonné sous le
capteur fixé à la tête. L'oxymètre pour tissu cérébral peut s'appuyer sur des technologies de lumière
continue, de domaine fréquentiel ou de domaine temporel. Le présent document s'applique aux
appareils EM utilisés en environnement hospitalier ou en dehors de l'environnement hospitalier, par
exemple dans des ambulances ou lors d'un transport aérien. D'autres normes peuvent s'appliquer aux
appareils EM pour ces environnements d'utilisation.
NOTE 1 Les oxymètres pour tissu cérébral sont parfois appelés « appareils de spectroscopie dans le proche
infrarouge » ou « appareils NIRS » (near infrared spectroscopy) dans la littérature médicale.
Les éléments suivants ne relèvent pas du domaine d'application du présent document :
 les oxymètres tissulaires ou vasculaires invasifs ;
 les oxymètres nécessitant un échantillon sanguin du patient ;
 les appareils de mesure de l'oxygène dissous ;
 les appareils EM, ou parties de ceux-ci, qui mesurent le changement de l'hémoglobine en fonction de
la longueur du trajet. Les exigences relatives aux appareils de spectroscopie dans le proche
[4]
infrarouge figurent dans l'ISO 80601-2-71 ;
 les appareils EM, ou parties de ceux-ci, qui mesurent la saturation artérielle en fonction des
changements pulsatiles dans les propriétés optiques des tissus (SpO ). Les exigences relatives aux
2
[3]
oxymètres de pouls figurent dans l'ISO 80601-2-61 ;
 les appareils EM, ou parties de ceux-ci, qui revendiquent la surveillance des tissus dans des parties
du corps autres que la tête.
© ISO 2021 – Tous droits réservés 1

---------------------- Page: 9 ----------------------
ISO 80601-2-85:2021(F)
Le présent document s'applique également aux oxymètres pour tissu cérébral, y compris les moniteurs
d'oxymètres pour tissu cérébral, les capteurs d'oxymètres pour tissu cérébral et les câbles de raccordement
du capteur, qui ont fait l'objet d'une remise à neuf.
Si un article ou un paragraphe est spécifiquement destiné à être appliqué uniquement aux appareils EM
ou uniquement aux systèmes EM, le titre et le contenu dudit article ou paragraphe l'indiqueront. Si cela
n'est pas le cas, l'article ou le paragraphe s'applique à la fois aux appareils EM et aux systèmes EM, selon
le cas.
Les dangers inhérents à la fonction physiologique prévue des appareils EM ou des systèmes EM dans le
cadre du domaine d'application du présent document ne sont pas couverts par des exigences
spécifiques contenues dans le présent document, à l'exception de 201.11, 201.7.2.13 et 201.8.4.1 de la
norme générale.
NOTE 2 Voir également 4.2 de la norme générale.
Le présent document peut également être appliqué aux appareils EM et leurs accessoires utilisés pour
l'atténuation d'une maladie, la compensation ou l'atténuation d'une blessure ou d'une incapacité.
Le présent document ne s'applique pas aux appareils à distance ou esclaves (secondaires) qui affichent
les valeurs de StO et qui sont situés en dehors de l'environnement du patient.
2
NOTE 3 Il est attendu que les appareils EM qui offrent le choix entre des fonctions de diagnostic et de
surveillance respectent les exigences du document approprié lorsqu'ils sont configurés pour cette fonction.
201.1.2 Objet
Remplacement :
Le présent document est destiné à spécifier des exigences particulières concernant la sécurité de base et
les performances essentielles applicables aux oxymètres pour tissu cérébral [comme défini en 201.3.202]
et à leurs accessoires.
NOTE 1 Les accessoires sont inclus du fait que la combinaison du moniteur d'oxymètre pour tissu cérébral et de
ses accessoires nécessite un niveau de sécurité adéquat. Les accessoires peuvent avoir un impact significatif sur la
sécurité de base ou sur les performances essentielles d'un oxymètre pour tissu cérébral.
NOTE 2 Le présent document a été élaboré pour satisfaire aux principes essentiels et recommandations
d'étiquetage pertinents de l'International Medical Device Regulators Forum (IMDRF), comme indiqué à
l'Annexe JJ.
NOTE 3 Le présent document a été élaboré pour satisfaire aux principes essentiels de sécurité et de performance
pertinents de l'ISO 16142-1:2016, comme indiqué à l'Annexe KK.
NOTE 4 Le présent document a été élaboré pour satisfaire aux exigences générales de sécurité et de
[20]
performances pertinentes du Règlement de l'Union européenne (UE) 2017/745 , comme indiqué à l'Annexe LL.
201.1.3 Normes collatérales
Ajout :
Le présent document fait référence aux normes collatérales applicables figurant dans l'Article 2 de la
norme générale et dans l'Article 201.2 du présent document.
2 © ISO 2021 – Tous droits réservés

---------------------- Page: 10 ----------------------
ISO 80601-2-85:2021(F)
L'IEC 60601-1-2:2014+AMD1:2020, l'IEC 60601-
...

PROJET
NORME ISO/FDIS
FINAL
INTERNATIONALE 80601-2-85
ISO/TC 121/SC 3
Appareils électromédicaux —
Secrétariat: ANSI
Début de vote:
Partie 2-85:
2020-10-23
Exigences particulières pour la
Vote clos le:
sécurité de base et les performances
2020-12-18
essentielles des oxymètres pour tissu
cérébral
Medical electrical equipment —
Part 2-85: Particular requirements for the basic safety and essential
performance of cerebral tissue oximeter equipment
Ce projet est soumis à un vote parallèle à ISO et à IEC.
LES DESTINATAIRES DU PRÉSENT PROJET SONT
INVITÉS À PRÉSENTER, AVEC LEURS OBSER-
VATIONS, NOTIFICATION DES DROITS DE PRO-
TRAITEMENT PARALLÈLE ISO/CEN
PRIÉTÉ DONT ILS AURAIENT ÉVENTUELLEMENT
CONNAISSANCE ET À FOURNIR UNE DOCUMEN-
TATION EXPLICATIVE.
OUTRE LE FAIT D’ÊTRE EXAMINÉS POUR
ÉTABLIR S’ILS SONT ACCEPTABLES À DES FINS
INDUSTRIELLES, TECHNOLOGIQUES ET COM-
Numéro de référence
MERCIALES, AINSI QUE DU POINT DE VUE
ISO/FDIS 80601-2-85:2020(F)
DES UTILISATEURS, LES PROJETS DE NORMES
INTERNATIONALES DOIVENT PARFOIS ÊTRE
CONSIDÉRÉS DU POINT DE VUE DE LEUR POSSI-
BILITÉ DE DEVENIR DES NORMES POUVANT
SERVIR DE RÉFÉRENCE DANS LA RÉGLEMENTA-
©
TION NATIONALE. ISO 2020

---------------------- Page: 1 ----------------------
ISO/FDIS 80601-2-85:2020(F)

DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2020
Tous droits réservés. Sauf prescription différente ou nécessité dans le contexte de sa mise en œuvre, aucune partie de cette
publication ne peut être reproduite ni utilisée sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie, ou la diffusion sur l’internet ou sur un intranet, sans autorisation écrite préalable. Une autorisation peut
être demandée à l’ISO à l’adresse ci-après ou au comité membre de l’ISO dans le pays du demandeur.
ISO copyright office
Case postale 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Genève
Tél.: +41 22 749 01 11
E-mail: copyright@iso.org
Web: www.iso.org
Publié en Suisse
ii © ISO 2020 – Tous droits réservés

---------------------- Page: 2 ----------------------
ISO/FDIS 80601-2-85:2020(F)
Sommaire Page
Avant-propos . vi
Introduction . vii
201.1 Domaine d'application, objet et normes connexes . 1
201.1.1 * Domaine d'application . 1
201.1.2 Objet . 2
201.1.3 Normes collatérales . 2
201.1.4 Normes particulières . 3
201.2 Références normatives. 4
201.3 Termes et définitions . 5
201.4 Exigences générales . 12
201.4.3.101 * Exigences supplémentaires pour les performances essentielles . 12
201.4.102 Exigences supplémentaires relatives aux critères d'acceptation . 12
201.4.103 Exigences supplémentaires relatives à l'oxymètre pour tissu cérébral, ainsi que ses
parties et accessoires . 13
201.5 Exigences générales relatives aux essais des appareils EM . 13
201.6 Classification des appareils EM et des systèmes EM . 14
201.7 Identification, marquage et documentation des appareils EM . 14
201.7.1.101 Informations devant être fournies par le fabricant . 14
201.7.2.3 Consultation des documents d'accompagnement . 14
201.7.2.9.101 Classification IP . 14
201.7.2.101 Exigences supplémentaires de marquage sur l'extérieur des parties
d'appareils EM . 15
201.7.4.3 Unités de mesure . 15
201.7.9.2 Instructions d’utilisation . 15
201.7.9.2.1.101 Exigences générales supplémentaires . 15
201.7.9.2.2.101 Exigences supplémentaires pour les avertissements et les consignes de
sécurité . 17
201.7.9.2.9.101 Exigences supplémentaires pour les instructions de fonctionnement . 17
201.7.9.2.14.101 Exigences supplémentaires pour les accessoires, équipements
supplémentaires et fournitures utilisées . 18
201.7.9.3.1.101 * Exigences générales supplémentaires . 18
201.8 Protection contre les dangers d'origine électrique provenant d'appareils EM . 18
201.8.3.101 Exigences supplémentaires pour la classification des parties appliquées . 18
201.8.5.5.1.101 Protection contre la défibrillation. 19
201.8.7.4.7.101 Exigences supplémentaires relatives au mesurage du courant de fuite du
patient . 19
201.9 Protection contre les dangers mécaniques des appareils EM et systèmes EM . 19
201.10 Protection contre les dangers dus aux rayonnements involontaires ou excessifs . 19
201.10.4 Lasers . 19
201.11 Protection contre les températures excessives et les autres dangers . 20
201.11.1.2.2 Parties appliquées non destinées à fournir de la chaleur à un patient . 20
© ISO 2020 – Tous droits réservés
iii

---------------------- Page: 3 ----------------------
ISO/FDIS 80601-2-85:2020(F)
201.11.6.5.101 * Exigences supplémentaires pour la pénétration d'eau ou de matières
particulaires dans les appareils EM ou les systèmes EM . 21
201.11.6.7 Stérilisation des appareils EM ou des systèmes EM . 21
201.11.8.101 Exigences supplémentaires applicables à la coupure de l'alimentation/du réseau
d'alimentation d'un appareil EM . 21
201.11.8.101.1 Condition d'alarme technique signalant un défaut d'alimentation . 21
201.11.8.101.2 Stockage des paramètres de réglage et des données après des coupures de
courte durée ou un basculement automatique . 22
201.11.8.101.3 Fonctionnement après une coupure de longue durée . 22
201.12 Précision des commandes et des instruments et protection contre les caractéristiques
de sortie dangereuses . 22
201.12.1.101 * Exactitude de la StO d'un oxymètre pour tissu cérébral . 23
2
201.12.1.101.1 * Spécification . 23
201.12.1.101.2 * Collecte de données en vue de la détermination de l'exactitude de la StO . 24
2
201.12.1.101.3 * Analyse de données pour la détermination de l'exactitude de la StO . 26
2
201.12.1.101.4 Caractéristiques de l'étude utilisée pour déterminer l'exactitude de la StO . 27
2
201.12.4 Protection contre les caractéristiques de sortie présentant des risques . 27
201.12.4.101  * Période de mise à jour des données . 27
201.12.4.102 * Inadéquation du signal . 28
201.13 Situations dangereuses et conditions de défaut pour les appareils EM . 28
201.13.101 Détection de défauts du capteur et de défauts du câble de raccordement du capteur 28
201.14 Systèmes électromédicaux programmables (SEMP) . 29
201.15 Construction de l'appareil EM . 29
201.15.3.5.101 * Exigences supplémentaires relatives à une manutention brutale . 29
201.15.3.5.101.1 * Chocs et vibrations (robustesse) . 29
201.15.3.5.101.2 * Chocs et vibrations pour un oxymètre pour tissu cérébral opérationnel en
déplacement pendant son fonctionnement . 30
201.15.101 Mode de fonctionnement . 31
201.16 Systèmes EM . 32
201.17 Compatibilité électromagnétique des appareils EM et des systèmes EM . 32
201.101 * Capteurs d'oxymètre pour tissu cérébral et câbles de raccordement du capteur . 32
201.101.1 Généralités . 32
201.101.2 Étiquetage . 33
201.102 Connexion fonctionnelle . 33
201.102.1 Généralités . 33
201.102.2 * Raccordement à un dossier informatisé de santé ou à un environnement clinique
intégré . 33
201.102.3 Raccordement à un système d'alarme réparti . 33
202 Perturbations électromagnétiques — Exigences et essais . 34
202.4.3.1 Configurations . 34
202.5.2.2.1 Exigences applicables à tous les appareils EM et systèmes EM . 34
202.8.1.101 Exigences générales supplémentaires . 34
202.8.2 Simulation physiologique du patient . 35
206 Aptitude à l'utilisation . 35
208 Exigences générales, essais et recommandations pour les systèmes d'alarme des
appareils et des systèmes électromédicaux . 36
© ISO 2020 – Tous droits réservés
iv

---------------------- Page: 4 ----------------------
ISO/FDIS 80601-2-85:2020(F)
208.6.1.2.101 * Exigences supplémentaires relatives à la priorité des conditions d'alarme. 36
208.6.5.4.101 * Exigences supplémentaires relatives au préréglage d'alarme par défaut . 36
208.6.8.5.101 Exigences supplémentaires relatives à l'indication et à l'accès aux états de
désactivation du signal d'alarme . 36
211 Exigences pour les appareils électromédicaux et les systèmes électromédicaux utilisés
dans l'environnement des soins à domicile . 36
212 Exigences pour les appareils électromédicaux et les systèmes électromédicaux destinés
à être utilisés dans l'environnement de services médicaux d'urgence . 37
Annexe C (informative) Guide des exigences de marquage et d'étiquetage pour les
appareils EM et les systèmes EM . 38
Annexe D (informative) Symboles de marquage . 43
Annexe AA (informative) Recommandations particulières et justifications . 44
Annexe BB (informative) Température de la peau au niveau du capteur d'oxymètre pour tissu
cérébral . 58
Annexe CC (informative) Détermination de l'exactitude . 60
Annexe DD (informative) Caractéristiques d'un fantôme d'hémoglobine tissulaire pour la
vérification de l'exactitude de l'oxymètre pour tissu cérébral. 67
Annexe EE (informative) Lignes directrices pour l'évaluation et la documentation de
l'exactitude de la StO chez les sujets humains . 80
2
Annexe FF (informative) Contrôleurs de fonctionnement pour les oxymètres pour tissu cérébral87
Annexe GG (informative) Concepts de temps de réponse des appareils EM . 91
Annexe HH (normative) Exigences applicables à l'interface de données . 96
Annexe II (informative) Comparaison des méthodes d'évaluation de la performance . 101
Annexe JJ (informative) Référence aux principes essentiels et aux recommandations
d'étiquetage de l'IMDRF . 108
Annexe KK (informative) Référence aux principes essentiels . 111
Annexe LL (informative)  Référence aux exigences générales de sécurité et de performances 114
Annexe MM (informative) Terminologie — Index alphabétique des termes définis . 117
Bibliographie. 122
© ISO 2020 – Tous droits réservés
v

---------------------- Page: 5 ----------------------
ISO/FDIS 80601-2-85:2020(F)
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux
de normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général
confiée aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire
partie du comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement
avec la Commission électrotechnique internationale (IEC) en ce qui concerne la normalisation
électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier de prendre note des différents
critères d'approbation requis pour les différents types de documents ISO. Le présent document a été
rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2
(voir www.iso.org/directives).
L'attention est attirée sur le fait que certains des éléments du présent document peuvent faire l'objet de
droits de propriété intellectuelle ou de droits analogues. L'ISO ne saurait être tenue pour responsable de
ne pas avoir identifié de tels droits de propriété et averti de leur existence. Les détails concernant les
références aux droits de propriété intellectuelle ou autres droits analogues identifiés lors de l'élaboration
du document sont indiqués dans l'Introduction et/ou dans la liste des déclarations de brevets reçues par
l'ISO (voir www.iso.org/brevets).
Les appellations commerciales éventuellement mentionnées dans le présent document sont données pour
information, par souci de commodité, à l’intention des utilisateurs et ne sauraient constituer un
engagement.
Pour une explication de la nature volontaire des normes, la signification des termes et expressions
spécifiques de l'ISO liés à l'évaluation de la conformité, ou pour toute information au sujet de l'adhésion de
l'ISO aux principes de l’Organisation mondiale du commerce (OMC) concernant les obstacles techniques au
commerce (OTC), voir le lien suivant : www.iso.org/iso/fr/avant-propos.
Le présent document a été élaboré conjointement par le comité technique ISO/TC 121, Matériel
d'anesthésie et de réanimation respiratoire, sous-comité SC 3, Appareils respiratoires et équipements
connexes utilisés pour les soins aux patients, et le comité d'études IEC/TC 62, Équipements électriques dans
la pratique médicale, sous-comité 62 D, Appareils électromédicaux, en collaboration avec le comité
technique CEN/TC 215 du Comité européen de normalisation (CEN), Équipement respiratoire et
anesthésique, conformément à l'Accord de coopération technique entre l'ISO et le CEN (Accord de Vienne).
Une liste de toutes les parties de la série ISO et IEC 80601 se trouve sur les sites web de l'ISO et de l'IEC.
Il convient que l’utilisateur adresse tout retour d’information ou toute question concernant le présent
document à l’organisme national de normalisation de son pays. Une liste exhaustive desdits organismes se
trouve à l’adresse www.iso.org/fr/members.html.
© ISO 2020 – Tous droits réservés
vi

---------------------- Page: 6 ----------------------
ISO/FDIS 80601-2-85:2020(F)
Introduction
L'estimation de la saturation en oxygène du sang dans les tissus cérébraux par un appareil d'oxymétrie
pour tissu cérébral est de plus en plus utilisée dans de nombreux domaines de la médecine. Le présent
document couvre les exigences de sécurité de base et de performances essentielles réalisables dans les
limites de la technologie existante.
L'Annexe AA justifie certaines de ces exigences. Elle sert à fournir des indications supplémentaires sur le
raisonnement des comités qui a conduit à l’élaboration d’une exigence face aux dangers dont elle traite.
L'Annexe BB est une enquête documentaire qui fournit des recommandations pertinentes pour la
détermination de la température maximale de sécurité de l'interface entre un capteur d'oxymètre pour tissu
cérébral et le tissu d'un patient.
L'Annexe CC traite à la fois des formules utilisées pour évaluer l'exactitude des mesurages de la StO des
2
oxymètres pour tissu cérébral, et des noms attribués à ces formules.
L'Annexe DD présente des recommandations concernant l'utilisation de méthodes in vitro (fantômes) pour
vérifier l'exactitude de la StO des oxymètres pour tissu cérébral.
2
L'Annexe EE présente une ligne directrice pour une étude de désaturation contrôlée in vivo (sur des sujets
humains) destinée à la vérification de l'exactitude de la StO des oxymètres pour tissu cérébral.
2
L'Annexe FF est une description des contrôleurs de fonctionnement à utiliser avec les oxymètres pour tissu
cérébral.
L'Annexe GG décrit les concepts de temps de réponse des oxymètres pour tissu cérébral.
L'Annexe HH décrit les exigences applicables à l'interface de données.
L'Annexe II compare des désaturations chez l'être humain (in vivo) et des désaturations de fantôme
d'hémoglobine tissulaire (in vitro) pour évaluer l'exactitude de la StO .
2
Dans le présent document, les types de polices suivants sont utilisés :
⎯ exigences et définitions : caractères romains ;
⎯ instructions, modalités d'essais et termes définis à l'Article 3 de la norme générale, dans le présent
document ou en note : caractères italiques ;
⎯ éléments informatifs situés hors des tableaux, tels que notes, exemples et références : en petits
caractères ; texte normatif des tableaux : petits caractères également.
Concernant la structure de la présente norme, le terme :
⎯ « article » désigne l'une des sections numérotées dans le sommaire, avec toutes ses subdivisions (par
exemple, l'Article 201.7 inclut les paragraphes 201.7.1, 201.7.2, etc.) ; et
⎯ « paragraphe » désigne une subdivision numérotée d'un article (par exemple, 201.7.1, 7.2 et 201.7.2.1
sont des paragraphes appartenant à l'Article 201.7).
© ISO 2020 – Tous droits réservés
vii

---------------------- Page: 7 ----------------------
ISO/FDIS 80601-2-85:2020(F)
Dans le présent document, les références à des articles sont précédées du terme « Article » suivi du
numéro de l'article concerné. Dans le présent document, les références aux paragraphes utilisent
uniquement le numéro du paragraphe concerné.
Dans le présent document, la conjonction « ou » est utilisée comme « ou inclusif » ; une affirmation est donc
vraie si une combinaison quelconque des conditions est vraie.
Pour les besoins du présent document :
⎯ « doit » (« shall ») signifie que la conformité à une exigence ou à un essai est obligatoire pour la
conformité au présent document ;
⎯ « il convient » (« should ») signifie que la conformité à une exigence ou à un essai est recommandée,
mais n'est pas obligatoire, pour la conformité au présent document ; et
⎯ « peut » (« may ») est utilisé pour décrire la permission (par exemple, une manière autorisée d'obtenir
la conformité à une exigence ou à un essai) ;
⎯ « peut » (« can ») est utilisé pour décrire une possibilité ou une capacité ; et
⎯ « doit » (« must ») est utilisé pour exprimer une contrainte externe.
L'Annexe C contient un guide pour le marquage et les exigences d'étiquetage dans le présent document.
L'Annexe D contient un résumé des symboles référencés dans le présent document.
Lorsqu'un astérisque (*) est utilisé comme premier caractère devant un titre, au début d'un titre de
paragraphe ou de tableau, il indique l'existence d'une recommandation ou d'une justification à consulter à
l'Annexe AA.
© ISO 2020 – Tous droits réservés
viii

---------------------- Page: 8 ----------------------
PROJET FINAL DE NORME INTERNATIONALE ISO/FDIS 80601-2-85:2020(F)

Appareils électromédicaux — Partie 2-85 : Exigences particulières
pour la sécurité de base et les performances essentielles des
oxymètres pour tissu cérébral
201.1 Domaine d'application, objet et normes connexes
L'Article 1 de la norme générale s'applique, à l'exception de ce qui suit :
NOTE La norme générale est l'IEC 60601-1:2005+AMD1:2012+AMD2:2020.
201.1.1 * Domaine d'application
Remplacement :
Le présent document s'applique à la sécurité de base et aux performances essentielles des oxymètres pour
tissu cérébral, qui utilisent la lumière à plusieurs longueurs d'onde afin d'obtenir une mesure
quantitative de la saturation en oxygène de l'hémoglobine dans le volume de tissu échantillonné sous le
capteur fixé à la tête. L'oxymètre pour tissu cérébral peut s'appuyer sur des technologies de lumière
continue, de domaine fréquentiel ou de domaine temporel. Le présent document s'applique aux
appareils EM utilisés en environnement hospitalier ou en dehors de l'environnement hospitalier, par
exemple dans des ambulances ou lors d'un transport aérien. D'autres normes peuvent s'appliquer aux
appareils EM pour ces environnements d'utilisation.
NOTE 1 Les oxymètres pour tissu cérébral sont parfois appelés « appareils de spectroscopie dans le proche
infrarouge » ou « appareils NIRS » (near infrared spectroscopy) dans la littérature médicale.
Les éléments suivants ne relèvent pas du domaine d'application du présent document :
⎯ les oxymètres tissulaires ou vasculaires invasifs ;
⎯ les oxymètres nécessitant un échantillon sanguin du patient ;
⎯ les appareils de mesure de l'oxygène dissous ;
⎯ les appareils EM, ou parties de ceux-ci, qui mesurent le changement de l'hémoglobine en fonction de
la longueur du trajet. Les exigences relatives aux appareils de spectroscopie dans le proche
[4]
infrarouge figurent dans l'ISO 80601-2-71 ;
⎯ les appareils EM, ou parties de ceux-ci, qui mesurent la saturation artérielle en fonction des
changements pulsatiles dans les propriétés optiques des tissus (SpO ). Les exigences relatives aux
2
[3]
oxymètres de pouls figurent dans l'ISO 80601-2-61 ;
⎯ les appareils EM, ou parties de ceux-ci, qui revendiquent la surveillance des tissus dans des parties
du corps autres que la tête.
© ISO 2020 – Tous droits réservés 1

---------------------- Page: 9 ----------------------
ISO/FDIS 80601-2-85:2020(F)
Le présent document s'applique également aux oxymètres pour tissu cérébral, y compris les moniteurs
d'oxymètres pour tissu cérébral, les capteurs d'oxymètres pour tissu cérébral et les câbles de raccordement
du capteur, qui ont fait l'objet d'une remise à neuf.
Si un article ou un paragraphe est spécifiquement destiné à être appliqué uniquement aux appareils EM
ou uniquement aux systèmes EM, le titre et le contenu dudit article ou paragraphe l'indiqueront. Si cela
n'est pas le cas, l'article ou le paragraphe s'applique à la fois aux appareils EM et aux systèmes EM, selon
le cas.
Les dangers inhérents à la fonction physiologique prévue des appareils EM ou des systèmes EM dans le
cadre du domaine d'application du présent document ne sont pas couverts par des exigences
spécifiques contenues dans le présent document, à l'exception de 201.11, 201.7.2.13 et 201.8.4.1 de la
norme générale.
NOTE 2 Voir également 4.2 de la norme générale.
Le présent document peut également être appliqué aux appareils EM et leurs accessoires utilisés pour
l'atténuation d'une maladie, la compensation ou l'atténuation d'une blessure ou d'une incapacité.
Le présent document ne s'applique pas aux appareils à distance ou esclaves (secondaires) qui affichent
les valeurs de StO et qui sont situés en dehors de l'environnement du patient.
2
NOTE 3 Il est attendu que les appareils EM qui offrent le choix entre des fonctions de diagnostic et de
surveillance respectent les exigences du document approprié lorsqu'ils sont configurés pour cette fonction.
201.1.2 Objet
Remplacement :
Le présent document est destiné à spécifier des exigences particulières concernant la sécurité de base et
les performances essentielles applicables aux oxymètres pour tissu cérébral [comme défini en 201.3.202]
et à leurs accessoires.
NOTE 1 Les accessoires sont inclus du fait que la combinaison du moniteur d'oxymètre pour tissu cérébral et de
ses accessoires nécessite un niveau de sécurité adéquat. Les accessoires peuvent avoir un impact significatif sur la
sécurité de base ou sur les performances essentielles d'un oxymètre pour tissu cérébral.
NOTE 2 Le présent document a été élaboré pour satisfaire aux principes essentiels et recommandations
d'étiquetage pertinents de l'Internationa
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.