Abstract

This document specifies requirements for single-use burettes to be used as subcomponents within sterilised infusion and transfusion sets for single use under gravity feed conditions or with pressure infusion apparatus. NOTE 1 Annex B provides explanations about the history of the development of the standard and summarises the different arguments discussed within ISO/TC 76 during the collaboration of this document. NOTE 2 Annex C provides recommendations regarding sustainability. NOTE 3 Annex D provides information on attributive and variable testing.

Status
Published
Publication Date
19-Nov-2026
Current Stage
6060 - International Standard published
Start Date
20-Nov-2026
Completion Date
20-Nov-2026

Buy Documents

Standard

ISO 8536-5:2026 - Infusion equipment for medical use — Part 5: Burette infusion sets for single use, gravity feed

Release Date:20-Nov-2026
English language (14 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

ISO 8536-5:2026 - Matériel de perfusion à usage médical — Partie 5: Appareils non réutilisables de perfusion à burette, à alimentation par gravité

Release Date:20-Nov-2026
French language (14 pages)
sale 15% off
Preview
sale 15% off
Preview

ISO 8536-5:2026 is an International Standard from the International Organization for Standardization (ISO) for single-use burettes used as subcomponents in sterilised infusion and transfusion sets. It specifies design, scale, material, physical, chemical and biological requirements, plus a graduation-accuracy test method, for burettes used under gravity feed conditions or with pressure infusion apparatus.

What does ISO 8536-5:2026 specify?

ISO 8536-5:2026 specifies requirements for single-use burettes as components within infusion and transfusion equipment for medical use. The document is focused on the burette itself, while the companion set standards cover the rest of the infusion or transfusion device.

The document is organised into clauses on scope, normative references, terms and definitions, general requirements, materials, physical requirements, chemical requirements, and biological requirements. It also includes four annexes that add a test method, background and rationale, sustainability guidance, and guidance on attributive and variable testing.

AnnexWhat it covers
Annex ATest method for burette scale graduation accuracy
Annex BRationale, guidance and history of development
Annex CSustainability recommendations
Annex DAttributive and variable testing

What are the key requirements of ISO 8536-5:2026?

ISO 8536-5:2026 is built around a few practical controls that affect how the burette is designed, verified, and used. The main points are the tube design, the graduated scale, the material requirements, and the flow and leakage performance of the assembled burette.

Design and intended function

Clause 4 requires a burette made of rigid or semi-rigid plastic that allows the fluid level to be seen. ISO 8536-5:2026 also requires filtered air-venting above the top graduation mark and the ability to receive fluid from the main container or from the injection site/Luer activated valve (LAV), then be closed off as a separate self-vented reservoir. In practice, this means the burette must work as a controllable holding chamber, not just as a connector.

If a product claims light protection, ISO 8536-5:2026 applies the light-resistance requirements of ISO 8536-15 to the graduated burette and attached drip chamber. That matters because light-sensitive fluids need the same protection across the parts that actually contain and deliver the fluid.

Graduated scale accuracy and layout

Clause 4 sets requirements for the graduated scale, including numbering, line spacing, zero position, and the overall scale length to the nominal capacity line. The scale accuracy is tested by Annex A, and any variation in the scale has to be assessed for risk using ISO 14971 and for usability using IEC 62366-1. In practice, manufacturers need to prove that the reading is reliable and that any alternative scale layout does not create use errors.

Materials and biological safety

Clause 5 requires the materials to meet the physical, chemical, and biological requirements in Clauses 6, 7, and 8, and to be assessed under ISO 10993-1. This means material choice is not only a design issue but also part of the biological safety evaluation for the device.

Physical performance

Clause 6 covers particulate contamination, tensile strength, leakage, the air-inlet device, filters, and the drip chamber/drip tube. The document requires manufacturing and handling steps that minimise particulate load, and it points to ISO 8536-4 or ISO 8536-10 for particulate testing of the final set. Leakage requirements depend on whether the burette is for gravity use only or for both gravity and pressure applications, so intended use affects verification.

The air-inlet device has to meet ISO 8536-4 and ISO 24072, while the filter has to suit the intended application - either a fluid filter or a filter for blood and blood components. The drip chamber must allow continuous observation of drops and support the expected drop delivery performance. For users, that means the drip system has to be consistent enough for controlled administration and practical priming.

Chemical and biological requirements

Clauses 7 and 8 direct the user to ISO 1135-4 or ISO 8536-4, as applicable. ISO 8536-5:2026 therefore relies on the companion set standards for chemical and biological safety details rather than restating them all here.

Who uses ISO 8536-5:2026?

ISO 8536-5:2026 is used by manufacturers of burettes and by manufacturers of infusion and transfusion sets that include a burette component. It is also relevant to quality, regulatory, and test laboratories that verify graduated-scale accuracy, leakage, particulate contamination, and material suitability.

The standard is most relevant in the infusion and transfusion equipment sector, including products intended for gravity feed and pressure infusion applications. The rationale notes that burettes are often used where small, well-controlled delivery is needed, including paediatric, neonate, and critically ill care.

What changed in ISO 8536-5:2026 from the previous edition?

ISO 8536-5:2026 cancels and replaces ISO 8536-5:2004 and is described as technically revised. The foreword states that the main changes are a scope change to treat burettes as components of infusion and transfusion devices, a wider scope for transfusion applications, revised requirements for the light resistance, graduated scale, particulate contamination, air-inlet, leakage, and drop former, and the removal of clauses on labelling, packaging, and disposal.

The foreword also states that Annexes A, B, and C were added.

Which standards are used with ISO 8536-5:2026?

Referenced standardWhat it contributes
ISO 1135-4Requirements for transfusion sets for single use, gravity feed, including blood-related filter and safety requirements where applicable
ISO 8536-4Requirements for infusion sets for single use, gravity feed, including several physical, chemical, and biological requirements
ISO 8536-8Upstream leakage requirements for burettes used with pressure infusion applications
ISO 8536-10Testing for particulate matter when the burette is used as an add-on accessory in pressure infusion equipment
ISO 8536-15Light-protective infusion set requirements used for light-resistant burettes and attached drip chambers
ISO 10993-1Biological evaluation of the materials in a medical device within a risk management process
ISO 14971Risk management for medical devices, used for scale variations and design changes
ISO 24072Bacterial retention test method for the air-inlet filter
IEC 62366-1Usability engineering, used where scale or design variations need usability assessment

What does the ISO 8536-5:2026 document contain?

ISO 8536-5:2026 contains a typical design figure with nomenclature, a graduated-scale table, and example scale layouts. It also includes a normative test method in Annex A for checking graduation accuracy by filling, dispensing, and gravimetrically verifying the delivered volume.

Annex B explains the rationale behind the requirements and the development history. Annex C gives sustainability guidance, and Annex D explains attributive and variable testing so users can choose an appropriate verification approach.

Relations

Effective Date
12-Feb-2026
Effective Date
16-Dec-2023
Effective Date
02-Dec-2023

Buy Documents

Standard

ISO 8536-5:2026 - Infusion equipment for medical use — Part 5: Burette infusion sets for single use, gravity feed

Release Date:20-Nov-2026
English language (14 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

ISO 8536-5:2026 - Matériel de perfusion à usage médical — Partie 5: Appareils non réutilisables de perfusion à burette, à alimentation par gravité

Release Date:20-Nov-2026
French language (14 pages)
sale 15% off
Preview
sale 15% off
Preview

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

ISO 8536-5:2026 is a standard published by the International Organization for Standardization (ISO). Its full title is "Infusion equipment for medical use — Part 5: Burette infusion sets for single use, gravity feed". This standard covers: This document specifies requirements for single-use burettes to be used as subcomponents within sterilised infusion and transfusion sets for single use under gravity feed conditions or with pressure infusion apparatus. NOTE 1 Annex B provides explanations about the history of the development of the standard and summarises the different arguments discussed within ISO/TC 76 during the collaboration of this document. NOTE 2 Annex C provides recommendations regarding sustainability. NOTE 3 Annex D provides information on attributive and variable testing.

This document specifies requirements for single-use burettes to be used as subcomponents within sterilised infusion and transfusion sets for single use under gravity feed conditions or with pressure infusion apparatus. NOTE 1 Annex B provides explanations about the history of the development of the standard and summarises the different arguments discussed within ISO/TC 76 during the collaboration of this document. NOTE 2 Annex C provides recommendations regarding sustainability. NOTE 3 Annex D provides information on attributive and variable testing.

ISO 8536-5:2026 is classified under the following ICS (International Classification for Standards) categories: 11.040.20 - Transfusion, infusion and injection equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

ISO 8536-5:2026 has the following relationships with other standards: It is inter standard links to FprEN ISO 8536-5, ISO 8404:2021, ISO 8536-5:2004. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

ISO 8536-5:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


International
Standard
ISO 8536-5
Third edition
Infusion equipment for medical
2026-11
use —
Part 5:
Burette infusion sets for single use,
gravity feed
Matériel de perfusion à usage médical —
Partie 5: Appareils non réutilisables de perfusion à burette, à
alimentation par gravité
Reference number
© ISO 2026
All rights reserved.
This ISO publication is protected by copyright and is owned by ISO and/or its licensors.
The content of this ISO publication is provided under licence, not sold. Use is subject to the applicable licence terms issued by ISO,
an ISO member body, or an authorized third-party distributor.
Except as required for implementation or expressly permitted by a separate licence, no part of this ISO publication may be
reproduced, distributed, modified, used, or made available in any form or by any means – electronic or mechanical, including
photocopying, scanning, recording, or posting on internal or external digital platforms.
Any use beyond the scope of the granted rights is prohibited and may result in legal action.
ISO copyright office
CP 401 • CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
Contents Page
Foreword .iv
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
4 General requirements . 2
4.1 Design .2
4.2 Graduated scale . .3
4.3 Numbering of scales .4
4.4 Overall length of scale to nominal capacity line .4
4.5 Variation of scale .4
5 Materials . 5
6 Physical requirements . 5
6.1 Particulate contamination .5
6.2 Tensile strength .5
6.3 Leakage .5
6.4 Air-inlet device .5
6.5 Fluid filter/filter for blood and blood components .5
6.6 Drip chamber and drip tube .5
7 Chemical requirements . 5
8 Biological requirements . 6
Annex A (normative) Test method for burette scale graduation accuracy . 7
Annex B (informative) Rationale, guidance and history of the development of this document . 8
Annex C (informative) Sustainability .11
Annex D (informative) Attributive and variable testing .13
Bibliography . 14

iii
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out through
ISO technical committees. Each member body interested in a subject for which a technical committee
has been established has the right to be represented on that committee. International organizations,
governmental and non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely
with the International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are described
in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the different types
of ISO document should be noted. This document was drafted in accordance with the editorial rules of the
ISO/IEC Directives, Part 2 (see www.iso.org/directives).
ISO draws attention to the possibility that the implementation of this document may involve the use of (a)
patent(s). ISO takes no position concerning the evidence, validity or applicability of any claimed patent
rights in respect thereof. As of the date of publication of this document, ISO had not received notice of (a)
patent(s) which may be required to implement this document. However, implementers are cautioned that
this may not represent the latest information, which may be obtained from the patent database available at
www.iso.org/patents. ISO shall not be held responsible for identifying any or all such patent rights.
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and expressions
related to conformity assessment, as well as information about ISO's adherence to the World Trade
Organization (WTO) principles in the Technical Barriers to Trade (TBT), see www.iso.org/iso/foreword.html.
This document was prepared by Technical Committee ISO/TC 76, Transfusion, infusion and injection, and blood
processing equipment for medical and pharmaceutical use, in collaboration with the European Committee for
Standardization (CEN) Technical Committee CEN/TC 205, Non-active medical devices, in accordance with the
Agreement on technical cooperation between ISO and CEN (Vienna Agreement).
This third edition cancels and replaces the second edition (ISO 8536-5:2004), which has been technically
revised.
The main changes are as follows:
— a change of scope to consider burettes as components of infusion sets;
— a scope expansion to include additional applications of burettes in transfusion applications;
— addition or revision of clauses concerning light-resistance, the graduated scale, particulate contamination,
air-inlet, leakage, and drop former requirements;
— removal of clauses concerning labelling, packaging, and disposal;
— addition of Annexes A, B, and C.
A list of all parts in the ISO 8536 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.

iv
International Standard ISO 8536-5:2026(en)
Infusion equipment for medical use —
Part 5:
Burette infusion sets for single use, gravity feed
1 Scope
This document specifies requirements for single-use burettes to be used as subcomponents within sterilised
infusion and transfusion sets for single use under gravity feed conditions or with pressure infusion
apparatus.
NOTE 1 Annex B provides explanations about the history of the development of the standard and summarises the
different arguments discussed within ISO/TC 76 during the collaboration of this document.
NOTE 2 Annex C provides recommendations regarding sustainability.
NOTE 3 Annex D provides information on attributive and variable testing.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes
requirements of this document. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
ISO 1135-4, Transfusion equipment for medical use — Part 4: Transfusion sets for single use, gravity feed
ISO 8536-4, Infusion equipment for medical use — Part 4: Infusion sets for single use, gravity feed
ISO 8536-8, Infusion equipment for medical use — Part 8: Infusion sets for single use with pressure infusion
apparatus
ISO 8536-10, Infusion equipment for medical use — Part 10: Accessories for fluid lines for single use with pressure
infusion equipment
ISO 8536-15, Infusion equipment for medical use — Part 15: Light-protective infusion sets for single use
ISO 10993-1, Biological evaluation of medical devices — Part 1: Requirements and general principles for the
evaluation of biological safety within a risk management process
ISO 14971, Medical devices — Application of risk management to medical devices
ISO 24072, Aerosol bacterial retention test method for air-inlet filter on administration devices
IEC 62366-1, Medical devices — Part 1: Application of usability engineering to medical devices
3 Terms and definitions
No terms and definitions are listed in this document.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/

4 General requirements
4.1 Design
4.1.1 A typical design along with the nomenclature to be used for burettes is given in Figure 1.
Key
1 injection site
2 air inlet with air filter and closure
3 graduated burette
4 shut-off valve
5 drip tube
6 drip chamber
7 fluid filter/filter for blood and blood components
a
Optional.
b
See ISO 8536-4 for fluid filter requirements.
c
See ISO 1135-4 for blood filter requirements.
Figure 1 — Example of a burette
4.1.2 The burette shall consist of a tube of rigid or semi-rigid plastics material and shall permit observation
of fluid in the chamber.
4.1.3 The burette shall be provided with filtered air-venting capability located in a position above the top
graduation mark.
4.1.4 The burette shall be capable of receiving fluid from the main container (or from the injection site/
Luer activated valve (LAV) as applicable) and of being closed off and serving as a separate self-vented
reservoir.
4.1.5 For burettes that claim light protection, the requirements for drip chambers specified in ISO 8536-15
shall apply to the graduated burette and attached drip chamber which is denoted as items 3 and 6 in Figure 1.
4.1.6 The nominal volume of the burette shall be designated by the total graduated capacity.

4.2 Graduated scale
4.2.1 The burette shall have either only one scale or more than one identical scale, which shall be graduated
and numbered at least at the intervals given in Table 1. The unit of volume shall be marked on the burette.
NOTE The scale interval can be less (finer) than the scale interval given in Table 1.
4.2.2 Graduation lines shall be of uniform thickness. They shall lie in planes at right angles to the axis of
the burette.
4.2.3 Graduation lines shall be evenly spaced along the longitudinal axis between the zero graduation line
and the line for the total graduated capacity.
4.2.4 When the burette is held vertically, the ends of all graduation lines of similar length shall be vertically
beneath each other.
The lengths of the short graduation lines on each scale are recommended to be approximately half the length
of the long lines. If different graduation line configurations are used, this shall be submitted to usability
evaluation in accordance with IEC 62366-1.
Examples of scales and the numbering of graduation lines are shown in Figure 2.
4.2.5 The zero-position mark on the burette shall be located in a position which compensates for the
volume displaced by any shut-off device, the position of the outlet relative to the bottom of the burette, or
any other feature of the bottom cap which can affect the reading.
4.2.6 The accuracy of the scale shall be tested in accordance with Annex A and shall meet the accuracy
requirements of Table 1.
Table 1 — Volume and scale intervals for burettes
Tolerance on any graduated capacity
Minimum Increment
Less than half
Nominal overall length between
and more Equal to or
Less than
capacity of of scale to Scale interval graduation
than or equal greater than
20 % nominal
burette nominal lines to be
to 20 % half nominal
capacity
capacity mark numbered
nominal capacity
capacity
V
ml  mm ml ml
±(1,5 % of V +1 % of expelled ±5 % of ex-
V < 50 67 2,0 10
volume) pelled volume
±(1,5 % of V +1 % of expelled ±5 % of ex-
V ≥ 50 75 5,0 10
volume) pelled volume
NOTE Expelled volume means all the liquid ejected to the point where the meniscus of the water level is coincident with the
zero mark of the scale.
NOTE The vertical lines are optional.
Figure 2 — Typical graduated scales for use in burettes
4.3 Numbering of scales
4.3.1 Graduation lines shall be numbered at least at the volume of increments given in Table 1.
Examples of scale numbering are shown in Figure 2.
4.3.2 When the burette is held vertically with the scale to the front, the numbers shall appear vertical on
the scale and be ap
...


Norme
internationale
ISO 8536-5
Troisième édition
Matériel de perfusion à usage
2026-11
médical —
Partie 5:
Appareils non réutilisables de
perfusion à burette, à alimentation
par gravité
Infusion equipment for medical use —
Part 5: Burette infusion sets for single use, gravity feed
Numéro de référence
DOCUMENT PROTÉGÉ PAR COPYRIGHT
© ISO 2026
Tous droits réservés.
La présente publication de l’ISO est protégée par le droit d’auteur et demeure la propriété de l’ISO et/ou de ses concédants de
licence.
Le contenu de la présente publication de l’ISO est fourni sous licence, et n’est pas vendu. Son utilisation est soumise aux conditions
de licence applicables définies par l’ISO, un organisme membre de l’ISO, ou un distributeur tiers agréé.
Sous réserve des besoins liés à sa mise en œuvre ou d’une autorisation expresse prévue aux termes d’une licence distincte, aucune
partie de la présente publication de l’ISO ne peut être reproduite, distribuée, modifiée, utilisée ou mise à disposition sous quelque
forme ou par quelque moyen que ce soit – électronique ou mécanique, y compris la photocopie, la numérisation, l’enregistrement
ou la publication sur des plateformes numériques internes ou externes.
Toute utilisation hors du champ d’application des droits accordés est interdite et peut donner lieu à des poursuites judiciaires.
ISO copyright office
Case postale 401 • CH-1214 Vernier, Genève
Tél.: +41 22 749 01 11
E-mail: copyright@iso.org
Web: www.iso.org
Publié en Suisse
ii
Sommaire Page
Avant-propos .iv
1 Domaine d'application . 1
2 Références normatives . 1
3 Termes et définitions . 2
4 Exigences générales . . 2
4.1 Conception .2
4.2 Échelle graduée .3
4.3 Numérotation des échelles.5
4.4 Longueur totale de l'échelle par rapport à la ligne de capacité nominale .5
4.5 Variation de l'échelle .5
5 Matériaux . 5
6 Exigences physiques . 5
6.1 Contamination particulaire .5
6.2 Résistance à la traction .5
6.3 Étanchéité .5
6.4 Dispositif de prise d'air .5
6.5 Filtre à fluide/filtre pour le sang et ses composants .6
6.6 Chambre compte-gouttes et tube compte-gouttes .6
7 Exigences chimiques . 6
8 Exigences biologiques. 6
Annexe A (normative) Essai d'exactitude de graduation d'échelle pour une burette . 7
Annexe B (informative) Justification, recommandations et historique de l'élaboration
du présent document . . 8
Annexe C (informative) Développement durable .11
Annexe D (informative) Essais attributifs et variables .13
Bibliographie . 14

iii
Avant-propos
L'ISO (Organisation internationale de normalisation) est une fédération mondiale d'organismes nationaux
de normalisation (comités membres de l'ISO). L'élaboration des Normes internationales est en général
confiée aux comités techniques de l'ISO. Chaque comité membre intéressé par une étude a le droit de faire
partie du comité technique créé à cet effet. Les organisations internationales, gouvernementales et non
gouvernementales, en liaison avec l'ISO participent également aux travaux. L'ISO collabore étroitement avec
la Commission électrotechnique internationale (IEC) en ce qui concerne la normalisation électrotechnique.
Les procédures utilisées pour élaborer le présent document et celles destinées à sa mise à jour sont
décrites dans les Directives ISO/IEC, Partie 1. Il convient, en particulier, de prendre note des différents
critères d'approbation requis pour les différents types de documents ISO. Le présent document
a été rédigé conformément aux règles de rédaction données dans les Directives ISO/IEC, Partie 2
(voir www.iso.org/directives).
L'ISO attire l'attention sur le fait que la mise en application du présent document peut entraîner l'utilisation
d'un ou de plusieurs brevets. L'ISO ne prend pas position quant à la preuve, à la validité et à l'applicabilité
de tout droit de propriété revendiqué à cet égard. À la date de publication du présent document, l'ISO
n'avait pas reçu notification qu'un ou plusieurs brevets pouvaient être nécessaires à sa mise en application.
Toutefois, il y a lieu d'avertir les responsables de la mise en application du présent document que des
informations plus récentes sont susceptibles de figurer dans la base de données de brevets, disponible à
l'adresse www.iso.org/brevets. L'ISO ne saurait être tenue pour responsable de ne pas avoir identifié tout ou
partie de tels droits de brevet.
Les appellations commerciales éventuellement mentionnées dans le présent document sont données pour
information, par souci de commodité, à l'intention des utilisateurs et ne sauraient constituer un engagement.
Pour une explication de la nature volontaire des normes, la signification des termes et expressions
spécifiques de l'ISO liés à l'évaluation de la conformité, ou pour toute information au sujet de l'adhésion de
l'ISO aux principes de l'Organisation mondiale du commerce (OMC) concernant les obstacles techniques au
commerce (OTC), voir www.iso.org/avant-propos.
Le présent document a été élaboré par le comité technique ISO/TC 76, Appareils de transfusion, de perfusion
et d'injection et appareils destinés au traitement du sang à usage médical et pharmaceutique, en collaboration
avec le comité technique CEN/TC 205, Dispositifs médicaux non-actifs, du Comité européen de normalisation
(CEN) conformément à l’Accord de coopération technique entre l’ISO et le CEN (Accord de Vienne).
Cette troisième édition annule et remplace la deuxième édition (ISO 8536-5:2004), qui a fait l'objet d'une
révision technique.
Les principales modifications sont les suivantes:
— une modification du domaine d'application afin de prendre en considération les burettes comme
composants d'appareils de perfusion;
— une extension du domaine d'application pour inclure des applications supplémentaires de burettes dans
les applications de transfusion;
— ajout ou révision de paragraphes concernant les exigences relatives à la résistance à la lumière, à l'échelle
graduée, à la contamination particulaire, à la prise d'air, à l'étanchéité, et au dispositif de formation de
gouttes;
— suppression des paragraphes concernant l'étiquetage, l'emballage, et l'élimination;
— ajout des Annexes A, B et C.
Une liste de toutes les parties de la série ISO 8536 se trouve sur le site web de l'ISO.
Il convient que l'utilisateur adresse tout retour d'information ou toute question concernant le présent
document à l'organisme national de normalisation de son pays. Une liste exhaustive desdits organismes se
trouve à l'adresse www.iso.org/fr/members.html.

iv
Norme internationale ISO 8536-5:2026(fr)
Matériel de perfusion à usage médical —
Partie 5:
Appareils non réutilisables de perfusion à burette, à
alimentation par gravité
1 Domaine d'application
Le présent document spécifie les exigences relatives aux burettes non réutilisables à utiliser comme sous-
composants dans les appareils de perfusion et transfuseurs stérilisés non réutilisables dans des conditions
d'alimentation par gravité ou avec un appareil de perfusion sous pression.
NOTE 1 L’Annexe B fournit des explications sur l'historique de l'élaboration de la norme et résume les différents
arguments abordés dans l'ISO/TC 76 pendant la collaboration du présent document.
NOTE 2 L’Annexe C fournit des recommandations relatives à la durabilité.
NOTE 3 L’Annexe D fournit des informations sur les essais attributifs et variables.
2 Références normatives
Les documents suivants sont cités dans le texte de sorte qu’ils constituent, pour tout ou partie de leur
contenu, des exigences du présent document. Pour les références datées, seule l’édition citée s’applique. Pour
les références non datées, la dernière édition du document de référence s'applique (y compris les éventuels
amendements).
ISO 1135-4, Matériel de transfusion à usage médical — Partie 4: Transfuseurs non réutilisables, à alimentation
par gravité
ISO 8536-4, Matériel de perfusion à usage médical — Partie 4: Appareils de perfusion non réutilisables, à
alimentation par gravité
ISO 8536-8, Matériel de perfusion à usage médical — Partie 8: Appareils de perfusion non réutilisables avec des
appareils de perfusion sous pression
ISO 8536-10, Matériel de perfusion à usage médical — Partie 10: Accessoires pour tubulures non réutilisables
avec un matériel de perfusion sous pression
ISO 8536-15, Matériel de perfusion à usage médical — Partie 15: Perfuseurs photoprotecteurs à usage unique
ISO 10993-1, Évaluation biologique des dispositifs médicaux — Partie 1: Exigences et principes généraux pour
l’évaluation de la sécurité biologique au sein d’un processus de gestion des risques
ISO 14971, Dispositifs médicaux — Application de la gestion des risques aux dispositifs médicaux
ISO 24072, Méthode d’essai de rétention bactérienne dans les aérosols pour les filtres d'admission d’air utilisés
sur les dispositifs d’administration
IEC 62366-1, Dispositifs médicaux — Partie 1: Application de l'ingénierie de l'aptitude à l'utilisation aux
dispositifs médicaux
3 Termes et définitions
Aucun terme n’est défini dans le présent document.
L’ISO et l’IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en normalisation,
consultables aux adresses suivantes:
— ISO Online browsing platform: disponible à l’adresse https:// www .iso .org/ obp
— IEC Electropedia: disponible à l’adresse https:// www .electropedia .org/
4 Exigences générales
4.1 Conception
4.1.1 Une conception type ainsi que la nomenclature à utiliser pour les burettes sont données à la Figure 1.
Légende
a
1 site d'injection
2 prise d'air avec filtre à air et obturateur
3 burette graduée
4 système d'arrêt
5 tube compte-gouttes
6 chambre compte-gouttes
7 filtre à fluide/filtre pour le sang et ses composants
a
Facultatif.
b
Voir l'ISO 8536-4 pour les exigences relatives aux filtres à fluide.
c
Voir l'ISO 1135-4 pour les exigences relatives aux filtres pour le sang.
Figure 1 — Exemple d'une burette
4.1.2 La burette doit être un tube en matière plastique rigide ou semi-rigide et doit permettre l'observation
du fluide dans la chambre.
4.1.3 La burette doit être pourvue d'un dispositif de filtrage et d'aération placé au-dessus de la graduation
supérieure.
4.1.4 La burette doit pouvoir recevoir un fluide provenant du réservoir principal (ou du site d'injection/
de la vanne actionnée par luer (LAV) le cas échéant) et être obturée pour servir de réservoir indépendant
autoventilé.
4.1.5 Pour les burettes qui demandent une protection contre la lumière, les exigences relatives aux
chambres compte-gouttes spécifiées dans l'ISO 8536-15 doivent s'appliquer à la burette graduée et à la
chambre compte-gouttes attachée qui sont indiquées en tant qu'élément 3 et 6 sur la Figure 1.
4.1.6 Le volume nominal de la burette doit être désigné par la capacité totale graduée.
4.2 Échelle graduée
4.2.1 La burette ne doit comporter qu'une seule échelle ou plusieurs échelles identiques, qui doivent être
graduées et numérotées au moins aux intervalles donnés dans le Tableau 1. L'unité de volume doit être
marquée sur la burette.
NOTE L'intervalle de graduation peut être inférieur (plus fin) à l'intervalle de graduation donné dans le Tableau 1.
4.2.2 Les lignes de graduation doivent être d'épaisseur uniforme. Elles doivent se trouver dans des plans à
angle droit par rapport à l'axe de la burette.
4.2.3 Les lignes de graduation doivent être régulièrement espacées le long de l'axe longitudinal entre la
ligne de graduation zéro et la ligne pour la capacité totale graduée.
4.2.4 Lorsque la burette est maintenue verticalement, les extrémités de toutes les lignes de graduation de
longueur similaire doivent être situées verticalement l'une au-dessous de l'autre.
Il est recommandé que les longueurs des lignes de graduation courtes sur chaque échelle fassent
approximativement la moitié de la longueur des lignes longues. Si différentes configurations de lignes de
graduation sont utilisées, cela doit être soumis à une évaluation d'utilisabilité selon l'IEC 62366-1.
Des exemples d'échelles et la numérotation des lignes de graduation sont présentés à la Figure 2.
4.2.5 Le repère du zéro sur la burette doit être situé dans une position qui prévoit une compensation pour
le volume déplacé par tout système d'arrêt, par la position de l'orifice de sortie par rapport à la base de la
burette, et par tout autre élément du protecteur susceptible d'avoir une incidence sur la lecture.
4.2.6 L'exactitude de l'échelle doit être soumise à l'essai conformément à l’Annexe A et doit satisfaire aux
exigences d'exactitude du Tableau 1.

Tableau 1 — Volume et intervalles de graduation des burettes
Tolérance sur toute capacité graduée
Longueur to-
Inférieure à
Incrément
tale minimale
la moitié et Égale ou
Capacité entre les
Inférieure
de l'échelle Intervalle de
supérieure supérieure à
nominale de lignes de
à 20 % de
par rapport au graduation
ou égale la moitié de
la burette graduation à
la capacité
repère de capa-
à 20 % la capacité
numéroter
nominale
cité nominale
de la capacité nominale
nominale
V
ml  mm ml ml
±(1,5 % de V + 1 % du volume ±5 % du vo-
V < 50 67 2,0 10
expulsé) lume expulsé
±(1,5 % de V + 1 % du volume ±5 % du vo-
V ≥ 50 75 5,0 10
expulsé) lume expulsé
Le volume expulsé désigne tout le liquide éjecté au point où le ménisque du niveau d'eau coïncide avec le repère du zéro de
l'échelle.
Volume nominal 50 ml 100 ml 150 ml
NOTE Les lignes verticales sont facultatives.
Figure 2 — Échelles graduées types pour utilisation sur les burettes

4.3 Numérotation des échelles
4.3.1 Les lignes de graduation doivent être numérotées au moins au niveau du volume des incréments
donnés dans le Tableau 1.
Des exemples de numérotation d'échelle sont présentés à la Figure 2.
4.3.2 Lorsque la burette est maintenue verticalement avec l'échelle vers l'avant, les nombres doivent
apparaître verticaux sur l'échelle et être approximativement centrés sur les lignes de graduation auxquelles
ils se rapportent. Les nombres doivent être placés près des traits-repères auxquels ils correspondent, mais
sans les toucher.
4.4 Longueur totale de l'échelle par rapport à la ligne de capacité nominale
La longueur totale de l'échelle doit être telle que donnée dans le Tableau 1.
4.5 Variation de l'échelle
Si cela est rendu nécessaire par une application spécifique,
...