Water quality — Determination of sodium and potassium — Part 2: Determination of potassium by atomic absorption spectrometry

The principle of the method specified is adding a caesium chloride solution to the sample as an ionization suppressant, aspirating the sample directly into the air/acetylene flame of an atomic absorption spectrometer and measuring the absorbance at a wavelength of 766,5 nm. Applies to water samples with a mass concentration of potassium in the range from 5 mg/l to 50 mg/l. This range can be extended if other dilution factors are chosen.

Qualité de l'eau — Dosage du sodium et du potassium — Partie 2: Dosage du potassium par spectrométrie d'absorption atomique

Kakovost vode – Določevanje natrija in kalija – 2. del: Določevanje kalija z atomsko absorpcijsko spektrometrijo

General Information

Status
Published
Publication Date
27-Apr-1993
Current Stage
9093 - International Standard confirmed
Start Date
01-Apr-2026
Completion Date
04-Apr-2026

Overview

ISO 9964-2:1993 is an international standard published by ISO, detailing the method to determine potassium concentration in water samples using atomic absorption spectrometry (AAS). This standard is part of a series dedicated to water quality assessment-specifically focusing on the measurement of sodium and potassium ions. The method described applies to raw and drinking water with potassium levels typically ranging from 5 mg/l to 50 mg/l, with flexibility for extension through sample dilution.

The technique involves adding a caesium chloride solution to suppress ionization, followed by direct aspiration of the sample into an air/acetylene-flame atomic absorption spectrometer. The absorbance is measured at a wavelength of 766.5 nm, providing a reliable and precise means of determining dissolved potassium content.

Key Topics

  • Scope and Range: Applicable to raw and drinking water samples containing 5 mg/l to 50 mg/l of potassium, with potential for broader application if samples are diluted accordingly.
  • Principle of Method: Utilizes caesium chloride as an ionization suppressant and flame atomic absorption spectrometry for detection and quantification.
  • Sample Preparation: Includes guidelines for filtering particulate matter, using appropriate reagents of analytical grade, and ensuring consistency in sample and solution preparation.
  • Calibration: Describes the preparation of calibration solutions and the construction of calibration curves critical for accurate determination.
  • Interferences: Standard acknowledges the minimal interference from common ions in water, contributing to the robustness of the method.
  • Apparatus Requirements: Specifies the need for a compatible atomic absorption spectrometer, appropriate glassware, and stringent cleaning protocols to prevent cross-contamination.
  • Reporting: Outlines requirements for reporting results, including identification of samples, adherence to the method, and documentation of any deviations.

Applications

ISO 9964-2:1993 is widely used in the following contexts:

  • Drinking Water Quality Control: Ensures public health by monitoring potassium levels, which is critical for compliance with regulatory standards.
  • Environmental Monitoring: Applied in the assessment of water bodies to detect potential contamination or natural variations in potassium content.
  • Water Treatment Facilities: Utilized to evaluate the effectiveness of treatment processes and to monitor water chemistry before and after processing.
  • Research Laboratories: Provides a standard methodology for academic and industrial research focused on water chemistry and environmental science.
  • Regulatory Compliance: Supports water suppliers and regulatory agencies in meeting analytical quality standards for routine water analysis.

Related Standards

For comprehensive water quality assessment and analytical best practices, the following related ISO standards are relevant:

  • ISO 9964-1:1992: Water quality - Determination of sodium by atomic absorption spectrometry.
  • ISO 9964-3:1996: Water quality - Determination of sodium and potassium by flame emission spectrometry.
  • ISO 5667-1:1980: Water quality - Sampling - Guidance on the design of sampling programmes.
  • ISO 5667-2:1991: Water quality - Sampling - Guidance on sampling techniques.

Practical Value

Adherence to ISO 9964-2:1993 ensures:

  • Consistency and Reliability: Laboratories worldwide achieve comparable results in potassium determination.
  • Data Quality Improvement: Reduces analytical errors through stringent calibration and sample handling protocols.
  • Regulatory Confidence: Facilitates acceptance of results by regulatory authorities and customers by employing recognized standards.

By following this ISO guideline, organizations and laboratories can confidently report water potassium concentrations, contributing to safer drinking water and effective environmental management.

Buy Documents

Standard

ISO 9964-2:1993 - Water quality — Determination of sodium and potassium — Part 2: Determination of potassium by atomic absorption spectrometry Released:4/28/1993

English language (4 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

ISO 9964-2:1993 - Qualité de l'eau — Dosage du sodium et du potassium — Partie 2: Dosage du potassium par spectrométrie d'absorption atomique Released:4/28/1993

French language (4 pages)
sale 15% off
Preview
sale 15% off
Preview
Standard

ISO 9964-2:1993 - Qualité de l'eau — Dosage du sodium et du potassium — Partie 2: Dosage du potassium par spectrométrie d'absorption atomique Released:4/28/1993

French language (4 pages)
sale 15% off
Preview
sale 15% off
Preview

Get Certified

Connect with accredited certification bodies for this standard

CIS Institut d.o.o.

Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

SA Slovenia Verified

Kiwa BDA Testing

Building and construction product certification.

RVA Netherlands Verified

Kmetijski inštitut Slovenije

Agricultural Institute of Slovenia. Soil testing, plant health, agricultural product analysis.

SA Slovenia Verified

Sponsored listings

Frequently Asked Questions

ISO 9964-2:1993 is a standard published by the International Organization for Standardization (ISO). Its full title is "Water quality — Determination of sodium and potassium — Part 2: Determination of potassium by atomic absorption spectrometry". This standard covers: The principle of the method specified is adding a caesium chloride solution to the sample as an ionization suppressant, aspirating the sample directly into the air/acetylene flame of an atomic absorption spectrometer and measuring the absorbance at a wavelength of 766,5 nm. Applies to water samples with a mass concentration of potassium in the range from 5 mg/l to 50 mg/l. This range can be extended if other dilution factors are chosen.

The principle of the method specified is adding a caesium chloride solution to the sample as an ionization suppressant, aspirating the sample directly into the air/acetylene flame of an atomic absorption spectrometer and measuring the absorbance at a wavelength of 766,5 nm. Applies to water samples with a mass concentration of potassium in the range from 5 mg/l to 50 mg/l. This range can be extended if other dilution factors are chosen.

ISO 9964-2:1993 is classified under the following ICS (International Classification for Standards) categories: 13.060.50 - Examination of water for chemical substances. The ICS classification helps identify the subject area and facilitates finding related standards.

ISO 9964-2:1993 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


SLOVENSKI STANDARD
01-februar-2000
.DNRYRVWYRGH±'RORþHYDQMHQDWULMDLQNDOLMD±GHO'RORþHYDQMHNDOLMD]
DWRPVNRDEVRUSFLMVNRVSHNWURPHWULMR
Water quality -- Determination of sodium and potassium -- Part 2: Determination of
potassium by atomic absorption spectrometry
Qualité de l'eau -- Dosage du sodium et du potassium -- Partie 2: Dosage du potassium
par spectrométrie d'absorption atomique
Ta slovenski standard je istoveten z: ISO 9964-2:1993
ICS:
13.060.50 3UHLVNDYDYRGHQDNHPLþQH Examination of water for
VQRYL chemical substances
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

INTERNATIONAL
ISO
STANDARD
9964-2
First edition
1993-05-01
Water quality - Determination of sodium
and potassium -
Part 2:
Determination of potassium by atomic
absorption spectrometry
- Dosage du sodium et du potassium -
Quakt6 de I ’eau
Partie 2: Dosage du potassium par spectromhtrie d ’absorption atomique
Reference number
ISO 99642:1993(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide
federation of national Standards bodies (ISO member bodies). The work
of preparing International Standards is normally carried out through ISO
technical committees. Esch member body interested in a subject for
which a technical committee has been established has the right to be
represented on that committee. International organizations, governmental
and non-governmental, in liaison with ISO, also take part in the work. ISO
collaborates closely with the International Electrotechnical Commission
(1 EC) on all matters of electrotechnical standardization.
Draft International Standards adopted by the technical committees are
circulated to the member bodies for voting. Publication as an International
Standard requires approval by at least 75 % of the member bodies casting
a vote.
International Standard ISO 9964-2 was prepared by Technical Committee
ISO/TC 147, Water quality, Sub-Committee SC 2, Physical, Chemical,
biochemical methods.
ISO 9964 consists of the following Parts, under the general title Water
quality - Determination o f sodium and potassium:
- Part 1: Determination of sodium by atomic absorption spectrometry
- Part 2: Determination of potassium by atomic absorption spec-
trome try
- Part 3: Determination of sodium and po tassium flame emission
bY
spectrometry
0 ISO 1993
Al9 rights reserved. No part 0% this publication may be reproduced or utilized in any ferm or
by any means, electronie or mechanical, including photocopying and microfilm, without per-
mission in writing from the publisher.
International Organization for Standardization
Case Postale 56 l CH-l 211 Geneve 20 l Switzerland
Printed in Switzerland
ii
INTERNATIONAL STANDARD ISO 9964=2:1993(E]
Water quality - Determination of sodium and
potassium -
Part 2:
Determination of potassium by atomic absorption
spectrometry
ISO 5667-1: 1980, Water quality - Sampling -
1 Scope
Part 1: Guidance on the design of sampling pro-
grammes.
1.1 Field of application
ISO 5667-2: 1991, Water quality - Sampling -
Part 2: Guidance on sampling techniques.
This International Standard specifies a method for the
determination of dissolved potassium by flame atomic
absorption spectrometry (AAS). lt is intended for the
3 Principle
analysis of raw and drinking waters.
Addition of caesium chloride Solution to the Sample
The method is applicable to water samples with a
as an ionization suppressant. Aspiration of the Sample
mass concentration of potassium in the range from
directly into the airlacetylene flame of an atomic ab-
5 mg/1 to 50 mg/l. This range tan be extended to
sorption spectrometer. Measurement of the
lower or higher limits if dilution factors are Chosen
absorbance at a wavelength of 766,5 nm.
which are different from the factor specified in
clause 8.
4 Reagents
1.2 Interferences
During the analysis, use only reagents of recognized
analytical grade and only deionized water or water of
The ions normally present in raw and drinking waters
equivalent purity.
do not interfere with the determination of potassium
by atomic absorption spectrometry.
4.1 Hydrochlorit acid, c(HCI) x 11 mol/l,
p = 1,18 g/ml.
2 Normative references
4.2 Nitrit acid, c(HNO$ N 16 mol/l, p = 1,41 g/ml.
The following Standards contain provisions which,
through reference in this text, constitute provisions
4.3 Caesium chloride, (CsCI), solution.
of this part of ISO 9964. At the time of publication, the
editions indicated were valid. All Standards are subject
Dissolve 25 g of caesium chloride in a Solution of
to revision, and Parties to agreements based on this
50 ml of hydrochloric acid (4.1) and 450 ml of water,
part of ISO 9964 are encouraged to investigate the
and dilute to 1 litre with water in a one-mark
possibility of applying the most recent editions of the
volumetric flask.
Standards indicated below. Members of IEC and ISO
maintain registers of currently valid International One litre of this Solution contains approximately 20 g
Standards. of Cs.
Nitrit acid (4.2) may be used instead of
NOTE 1
7 Procedure
hydrochloric acid (4.1).
7.1 Preparation of test samples for
presentation to the spectrometer
4.4 Potassium, stock solution.
In a 1 000 ml one-mark volumetric flask, dissolve
7.1.1 Filter samples containing particulate matter
1,907 g -f: 0,005 g of potassium chloride (previously
through an acid-washed filter of pore size 0,45 km to
dried for at least 1 h at 140 “C & 10 “C) in water and
prevent clogging of the nebulizer and burner Systems.
make up to the mark.
(The acid used for washing the filter should have the
same concentration and be of the same type as the
Store the Solution, which is stable for at least
acid used in the Sample preparation.)
6 months, in a polyeth
...


INTERNATIONAL
ISO
STANDARD
9964-2
First edition
1993-05-01
Water quality - Determination of sodium
and potassium -
Part 2:
Determination of potassium by atomic
absorption spectrometry
- Dosage du sodium et du potassium -
Quakt6 de I ’eau
Partie 2: Dosage du potassium par spectromhtrie d ’absorption atomique
Reference number
ISO 99642:1993(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide
federation of national Standards bodies (ISO member bodies). The work
of preparing International Standards is normally carried out through ISO
technical committees. Esch member body interested in a subject for
which a technical committee has been established has the right to be
represented on that committee. International organizations, governmental
and non-governmental, in liaison with ISO, also take part in the work. ISO
collaborates closely with the International Electrotechnical Commission
(1 EC) on all matters of electrotechnical standardization.
Draft International Standards adopted by the technical committees are
circulated to the member bodies for voting. Publication as an International
Standard requires approval by at least 75 % of the member bodies casting
a vote.
International Standard ISO 9964-2 was prepared by Technical Committee
ISO/TC 147, Water quality, Sub-Committee SC 2, Physical, Chemical,
biochemical methods.
ISO 9964 consists of the following Parts, under the general title Water
quality - Determination o f sodium and potassium:
- Part 1: Determination of sodium by atomic absorption spectrometry
- Part 2: Determination of potassium by atomic absorption spec-
trome try
- Part 3: Determination of sodium and po tassium flame emission
bY
spectrometry
0 ISO 1993
Al9 rights reserved. No part 0% this publication may be reproduced or utilized in any ferm or
by any means, electronie or mechanical, including photocopying and microfilm, without per-
mission in writing from the publisher.
International Organization for Standardization
Case Postale 56 l CH-l 211 Geneve 20 l Switzerland
Printed in Switzerland
ii
INTERNATIONAL STANDARD ISO 9964=2:1993(E]
Water quality - Determination of sodium and
potassium -
Part 2:
Determination of potassium by atomic absorption
spectrometry
ISO 5667-1: 1980, Water quality - Sampling -
1 Scope
Part 1: Guidance on the design of sampling pro-
grammes.
1.1 Field of application
ISO 5667-2: 1991, Water quality - Sampling -
Part 2: Guidance on sampling techniques.
This International Standard specifies a method for the
determination of dissolved potassium by flame atomic
absorption spectrometry (AAS). lt is intended for the
3 Principle
analysis of raw and drinking waters.
Addition of caesium chloride Solution to the Sample
The method is applicable to water samples with a
as an ionization suppressant. Aspiration of the Sample
mass concentration of potassium in the range from
directly into the airlacetylene flame of an atomic ab-
5 mg/1 to 50 mg/l. This range tan be extended to
sorption spectrometer. Measurement of the
lower or higher limits if dilution factors are Chosen
absorbance at a wavelength of 766,5 nm.
which are different from the factor specified in
clause 8.
4 Reagents
1.2 Interferences
During the analysis, use only reagents of recognized
analytical grade and only deionized water or water of
The ions normally present in raw and drinking waters
equivalent purity.
do not interfere with the determination of potassium
by atomic absorption spectrometry.
4.1 Hydrochlorit acid, c(HCI) x 11 mol/l,
p = 1,18 g/ml.
2 Normative references
4.2 Nitrit acid, c(HNO$ N 16 mol/l, p = 1,41 g/ml.
The following Standards contain provisions which,
through reference in this text, constitute provisions
4.3 Caesium chloride, (CsCI), solution.
of this part of ISO 9964. At the time of publication, the
editions indicated were valid. All Standards are subject
Dissolve 25 g of caesium chloride in a Solution of
to revision, and Parties to agreements based on this
50 ml of hydrochloric acid (4.1) and 450 ml of water,
part of ISO 9964 are encouraged to investigate the
and dilute to 1 litre with water in a one-mark
possibility of applying the most recent editions of the
volumetric flask.
Standards indicated below. Members of IEC and ISO
maintain registers of currently valid International One litre of this Solution contains approximately 20 g
Standards. of Cs.
Nitrit acid (4.2) may be used instead of
NOTE 1
7 Procedure
hydrochloric acid (4.1).
7.1 Preparation of test samples for
presentation to the spectrometer
4.4 Potassium, stock solution.
In a 1 000 ml one-mark volumetric flask, dissolve
7.1.1 Filter samples containing particulate matter
1,907 g -f: 0,005 g of potassium chloride (previously
through an acid-washed filter of pore size 0,45 km to
dried for at least 1 h at 140 “C & 10 “C) in water and
prevent clogging of the nebulizer and burner Systems.
make up to the mark.
(The acid used for washing the filter should have the
same concentration and be of the same type as the
Store the Solution, which is stable for at least
acid used in the Sample preparation.)
6 months, in a polyethylene bottle.
NOTE 3 Instead of filtration, the particulate matter may
One litre of this Solution contains 1 000 mg of K.
be removed from Suspension by using a centrifuge.
Alternatively, use a commercially available, ready-
made Solution.
7.1.2 Take a number of 100 ml one-mark volumetric
flasks according to the number of samples being an-
alysed. To each of these flasks, add 10 ml of the
caesium chloride Solution (4.3).
4.5 Potassium, Standard solution.
Pipette 10 ml of the potassium stock Solution (4.4)
7.1.3 Pipette, 10 ml of the Sample into each
...


NORME
ISO
INTERNATIONALE 9964-2
Première édition
1993-05-01
Qualité de l’eau - Dosage du sodium et
du potassium -
Partie 2:
ium par spectrométrie
Dosage du potass
d’absorption atom ique
Determination of sodium and potassium -
Water quality -
Part 2: Determination of potassium by atomic absorption spectrometry
Numbro de réf&ence
PS0 9964-2: 1993(F)
Avant-propos
L’ISO (Organisation internationale de normalisation) est une fédération
mondiale d’organismes nationaux de normalisation (comités membres de
I’ISO). L’élaboration des Normes internationales est en général confiée aux
comites techniques de I’ISO. Chaque comite membre interessé par une
etude a le droit de faire partie du comité technique crée a cet effet. Les
organisations internationales, gouvernementales et non gouvernemen-
tales, en liaison avec I’ISO participent également aux travaux. L’ISO colla-
bore etroitement avec la Commission électrotechnique internationale (CEI)
en ce qui concerne la normalisation électrotechnique.
Les projets de Normes internationales adoptes par les comités techniques
sont soumis aux comites membres pour vote. Leur publication comme
Normes internationales requiert l’approbation de 75 % au moins des co-
mites membres votants.
La Norme internationale ISO 9964-2 a ete elaboree par le comité techni-
que lSO/TC 147, Qualite de l’eau, sous-comite SC 2, Méthodes physiques,
chimiques et biochimiques.
L’ISO 9964 comprend les parties suivantes, présentées sous le titre gé-
néral Qualité de l’eau - Dosage du sodium et du potassium:
- Partie 1: Dosage du sodium par spectrométrie d’absorption atomi-
que
- Partie 2: Dosage du potassium par spectrométrie d’absorption a to-
mique
- Partie 3: Dosage du sodium et du potassium par spectrométrie
d ‘emission de flamme
0 ISO 1993
Droits de reproduction reservés. Aucune partie de cette publication ne peut être reproduite
ni utilisee sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie et les microfilms, sans l’accord Acrit de I’editeur.
Organisation internationale de normalisation
Case Postale 56 l CH-l 211 Geneve 20 l Suisse
Imprime en Suisse
ii
NORME INTERNATIONALE
ISO 9964-2: 1993(F)
- Dosage du sodium et du
Qualité de l’eau
potassium -
Partie 2:
Dosage du potassium par spectrométrie d’absorption
atomique
des Normes internationales en vigueur à un moment
1 Domaine d’application
donne.
1.1 Applicabilité
ISO 5667-l : 1980, Qualité de l’eau - Échantillonnage
- Partie 1: Guide général pour l’établissement des
La présente partie de I’ISO 9964 prescrit une me-
programmes d’échantillonnage.
thode de dosage du potassium dissous, par spectro-
métrie d’absorption atomique de flamme. Elle est ISO 5667-2:1991, Qualit de l’eau - Échantillonnage
applicable à l’analyse des eaux brutes et des eaux - Partie 2: Guide général sur les techniques
potables. d %chan tillonnage.
La méthode s’applique aux échantillons d’eau dont la
3 Principe
concentration est comprise entre 5 mg/1 et 10 mg/l.
Cet intervalle peut être étendu à des limites plus fai-
Addition, à l’échantillon, d’une solution de chlorure de
bles ou plus elevees si on choisit des facteurs de di-
cesium servant de suppresseur d’ionisation. Aspi-
lution differents du facteur de dilution spécifié à
ration directe de l’échantillon dans la flamme
l’article 8.
d’air/acétylène d’un spectromètre d’absorption atomi-
que. Mesure de I’absorbance à une longueur d’onde
1.2 Interférences
de 766,5 nm.
Les ions normalement présents dans les eaux pota-
4 Réactifs
bles et les eaux brutes n’interférent pas avec le do-
sage du potassium par spectrométrie d’absorption
Au cours de l’analyse, utiliser uniquement des reactifs
atomique.
de qualité analytique reconnue et de l’eau déionisée
ou de l’eau de pureté équivalente.
2 Références normatives
4.1 Acide chlorhydrique, C(H~I) a 11 mol/l,
p = 1,18 g/ml.
Les normes suivantes contiennent des dispositions
qui, par suite de la reference qui en est faite, consti-
tuent des dispositions valables pour la présente partie 4.2 Acide nitrique, c(HNO$ a 16 mol/l,
de I’ISO 9964. Au moment de la publication, les édi- p = 1,41 g/ml.
tions indiquées étaient en vigueur. Toute norme est
sujette à révision et les parties prenantes des accords
4.3 Chlorure de cbsium, (CsCI), solution.
fondes sur la présente partie de I’ISO 9964 sont invi-
tees à rechercher la possibilité d’appliquer les Aditions Dissoudre 25 g de chlorure de césium dans une so-
les plus recentes des normes indiquées ci-après. Les lution de 50 ml d’acide chlorhydrique (4.1) et de
membres de la CEI et de I’ISO possédent le registre 450 ml d’eau, et diluer avec de l’eau jusqu’à 1 litre.
ISO 9964=2:1993(F)
niers par addition d’acide chlorhydrique (4.1) ou d’acide
1 litre de cette solution contient environ 20 g de Cs.
nitrique (4.2) pour avoir un pH d’environ 1. Tous les échan-
tillons, étalons et blancs devront contenir la même concen-
NOTE 1 Au lieu d’acide chlorhydrique (4.1), on peut utili-
tration et le meme type d’acide.
ser de l’acide nitrique (4.2).
4.4 Potassium, solution mh.
7 Mode opératoire
Dissoudre dans l’eau, dans une fiole jaugée de
1 000 ml, 1,907 g & 0,005 g de chlorure de potas-
7.1 Préparation des échantillons d’essai
sium (seché pendant au moins lh à
pour présentation au spectromètre
140 “C * 10 OC).
7.1.1 Filtrer les échantillons contenant des matiéres
Compléter au trait repère avec de l’eau.
particulaires, sur un filtre lave par un acide, d’ouver-
Conservée dans un flacon en polyéthyléne, la solution ture de pores de 0,45 prn, pour éviter un colmatage
est stable pendant au moins 6 mois. du pulvérisateur et du brûleur.
1 litre de cette solution contient 1 000 mg de K.
NOTE 3 Au lieu d’une filtration, on peut utiliser une cen-
trifugeuse pour éliminer les matiéres particulaires.
En variante, utiliser une solution prete à l’emploi, du
commerce.
7.1.2 Prendre autant de fioles jaugées de 100 ml
qu’il est nécessaire, compte tenu du nombre
4.5 Potassium, solution étalon.
d’echantillons analyses. Introduire dans cha
...


NORME ISO
INTERNATIONALE 9964-2
Première édition
1993-05-01
- Dosage du sodium et
Qualité de l’eau
du potassium -
Partie 2:
ium par spectrométrie
Dosage du potass
Ique
d’absorption atom
- Determination of sodium and potassium -
Wa ter quality
Part 2: Determination of potassium by atomic absorption spectrometry
Numhro de réf&ence
PS0 9964-2: 1993(F)
ISO 9964=2:1993(F)
Avant-propos
L’ISO (Organisation internationale de normalisation) est une féderation
mondiale d’organismes nationaux de normalisation (comités membres de
I’ISO). L’élaboration des Normes internationales est en général confiée aux
comites techniques de I’ISO. Chaque comite membre interessé par une
etude a le droit de faire partie du comité technique crée à cet effet. Les
organisations internationales, gouvernementales et non gouvernemen-
tales, en liaison avec I’ISO participent également aux travaux. L’ISO colla-
bore étroitement avec la Commission électrotechnique internationale (CEI)
en ce qui concerne la normalisation électrotechnique.
Les projets de Normes internationales adoptes par les comités techniques
sont soumis aux comites membres pour vote. Leur publication comme
Normes internationales requiert l’approbation de 75 % au moins des co-
mites membres votants.
La Norme internationale ISO 9964-2 a ete élaboree par le comité techni-
que lSO/TC 147, Qualité de l’eau, sous-comite SC 2, Méthodes physiques,
chimiques et biochimiques.
L’ISO 9964 comprend les parties suivantes, présentées sous le titre gé-
néral Qualité de l’eau - Dosage du sodium et du potassium:
- Partie 1: Dosage du sodium par spectrométrie d’absorption atomi-
que
- Partie 2: Dosage du potassium par spectrométrie d’absorption a to-
mique
- Partie 3: Dosage du sodium et du potassium par spectrométrie
d ‘emission de flamme
0 60 1993
Droits de reproduction rkwés. Aucune partie de cette publication ne peut être reproduite
ni utilisbe sous quelque forme que ce soit et par aucun procédé, électronique ou mécanique,
y compris la photocopie et les microfilms, sans l’accord Acrit de Wditeur.
Organisation internationale de normalisation
Case Postale 56 l CH-l 211 Genhe 20 l Suisse
Imprime en Suisse
ii
NORME INTERNATIONALE
ISO 9964-2: 1993(F)
- Dosage du sodium et du
Qualité de l’eau
potassium -
Partie 2:
Dosage du potassium par spectrométrie d’absorption
atomique
des Normes internationales en vigueur à un moment
1 Domaine d’application
donne.
1.1 Applicabilité
ISO 5667-l : 1980, Qualité de l’eau - Échantillonnage
- Partie 1: Guide général pour l’établissement des
La présente partie de I’ISO 9964 prescrit une me-
programmes d’échantillonnage.
thode de dosage du potassium dissous, par spectro-
métrie d’absorption atomique de flamme. Elle est ISO 5667-2:1991, Qualit de l’eau - Échantillonnage
applicable à l’analyse des eaux brutes et des eaux - Partie 2: Guide général sur les techniques
potables. d %chan tillonnage.
La méthode s’applique aux échantillons d’eau dont la
3 Principe
concentration est comprise entre 5 mg/1 et 10 mg/l.
Cet intervalle peut être étendu à des limites plus fai-
Addition, à l’échantillon, d’une solution de chlorure de
bles ou plus elevees si on choisit des facteurs de di-
cesium servant de suppresseur d’ionisation. Aspi-
lution differents du facteur de dilution spécifié à
ration directe de l’échantillon dans la flamme
l’article 8.
d’air/acétylène d’un spectromètre d’absorption atomi-
que. Mesure de I’absorbance à une longueur d’onde
1.2 Interférences
de 766,5 nm.
Les ions normalement présents dans les eaux pota-
4 Réactifs
bles et les eaux brutes n’interférent pas avec le do-
sage du potassium par spectrométrie d’absorption
Au cours de l’analyse, utiliser uniquement des reactifs
atomique.
de qualité analytique reconnue et de l’eau déionisée
ou de l’eau de pureté équivalente.
2 Références normatives
4.1 Acide chlorhydrique, C(H~I) a 11 mol/l,
p = 1,18 g/ml.
Les normes suivantes contiennent des dispositions
qui, par suite de la reference qui en est faite, consti-
tuent des dispositions valables pour la présente partie 4.2 Acide nitrique, c(HNO$ a 16 mol/l,
de I’ISO 9964. Au moment de la publication, les édi- p = 1,41 g/ml.
tions indiquées étaient en vigueur. Toute norme est
sujette à révision et les parties prenantes des accords
4.3 Chlorure de cbsium, (CsCI), solution.
fondes sur la présente partie de I’ISO 9964 sont invi-
tees à rechercher la possibilité d’appliquer les Aditions Dissoudre 25 g de chlorure de césium dans une so-
les plus recentes des normes indiquées ci-après. Les lution de 50 ml d’acide chlorhydrique (4.1) et de
membres de la CEI et de I’ISO possédent le registre 450 ml d’eau, et diluer avec de l’eau jusqu’à 1 litre.
ISO 9964=2:1993(F)
niers par addition d’acide chlorhydrique (4.1) ou d’acide
1 litre de cette solution contient environ 20 g de Cs.
nitrique (4.2) pour avoir un pH d’environ 1. Tous les échan-
tillons, étalons et blancs devront contenir la même concen-
NOTE 1 Au lieu d’acide chlorhydrique (4.1), on peut utili-
tration et le meme type d’acide.
ser de l’acide nitrique (4.2).
4.4 Potassium, solution mh.
7 Mode opératoire
Dissoudre dans l’eau, dans une fiole jaugée de
1 000 ml, 1,907 g & 0,005 g de chlorure de potas-
7.1 Préparation des échantillons d’essai
sium (seché pendant au moins lh à
pour présentation au spectromètre
140 “C * 10 OC).
7.1.1 Filtrer les échantillons contenant des matiéres
Compléter au trait repère avec de l’eau.
particulaires, sur un filtre lave par un acide, d’ouver-
Conservée dans un flacon en polyéthyléne, la solution ture de pores de 0,45 prn, pour éviter un colmatage
est stable pendant au moins 6 mois. du pulvérisateur et du brûleur.
1 litre de cette solution contient 1 000 mg de K.
NOTE 3 Au lieu d’une filtration, on peut utiliser une cen-
trifugeuse pour éliminer les matiéres particulaires.
En variante, utiliser une solution prete à l’emploi, du
commerce.
7.1.2 Prendre autant de fioles jaugées de 100 ml
qu’il est nécessaire, compte tenu du nombre
4.5 Potassium, solution étalon.
d’echantillons analyses. Introduire da
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.