SIST EN ISO 16061:2021
(Main)Instruments for use in association with non-active surgical implants - General requirements (ISO 16061:2021)
Instruments for use in association with non-active surgical implants - General requirements (ISO 16061:2021)
This document specifies the general requirements for instruments to be used in association with nonactive
surgical implants. These requirements apply to instruments when they are manufactured and
when they are supplied after refurbishment.
NOTE In this document, unless otherwise specified, the term “instrument” refers to an instrument for use in
association with non-active surgical implants.
This document also applies to instruments which can be connected to power-driven systems, but it
does not apply to the power-driven systems themselves.
With regard to safety, this document gives the requirements for intended performance, design
attributes, materials, design evaluation, manufacture, sterilization, packaging, and information
supplied by the instrument manufacturer, hereafter referred to as the manufacturer.
This document is not applicable to instruments associated with dental implants, transendodontic and
transradicular implants and ophthalmic implants.
Instrumente, die in Verbindung mit nichtaktiven chirurgischen Implantaten verwendet werden - Allgemeine Anforderungen (ISO 16061:2021)
Dieses Dokument legt die allgemeinen Anforderungen für Instrumente, die in Verbindung mit nichtaktiven chirurgischen Implantaten verwendet werden, fest. Diese Anforderungen gelten für Instrumente während ihrer Herstellung und wenn sie nach Aufarbeitung geliefert werden.
ANMERKUNG Sofern nicht anders angegeben bezeichnet der Begriff „Instrument“ in diesem Dokument ein Instrument, das in Verbindung mit nichtaktiven chirurgischen Implantaten verwendet wird.
Dieses Dokument ist auch für Instrumente anzuwenden, die mit Antriebseinheiten verbunden werden können, aber es gilt nicht für die Antriebseinheiten selbst.
In Bezug auf die Sicherheit legt dieses Dokument die Anforderungen für die vorgesehene Funktion, Konstruktionsmerkmale, Werkstoffe, Designbewertung, Herstellung, Sterilisation, Verpackung und durch den Instrumentenhersteller bereitgestellte Informationen, im Weiteren Hersteller genannt, fest.
Dieses Dokument ist nicht für Instrumente anzuwenden, die in Verbindung mit Dentalimplantaten, transendodontischen und transradikulären Implantaten sowie ophthalmologischen Implantaten verwendet werden.
Instruments à utiliser en association avec les implants chirurgicaux non actifs - Exigences générales (ISO 16061:2021)
Le présent document spécifie les exigences générales relatives aux instruments à utiliser en association avec les implants chirurgicaux non actifs. Ces exigences s'appliquent aux instruments fournis après fabrication ou aux instruments fournis après remise en état.
NOTE Dans le présent document, sauf indication contraire, le terme «instrument» désigne un instrument destiné à être utilisé en association avec les implants chirurgicaux non actifs.
Le présent document s'applique également aux instruments pouvant être connectés à des systèmes motorisés, mais il ne s'applique pas aux systèmes eux-mêmes.
En matière de sécurité, le présent document donne les exigences relatives aux performances prévues, aux caractéristiques de conception, aux matériaux, à l'évaluation de la conception, à la fabrication, à la stérilisation, à l'emballage et aux informations fournies par le fabricant d'instruments, ci-après appelé fabricant.
Le présent document ne s'applique pas aux instruments associés aux implants dentaires, aux implants transendodontiques et transradiculaires, et aux implants ophtalmiques.
Instrumenti, ki se uporabljajo pri neaktivnih kirurških vsadkih (implantatih) - Splošne zahteve (ISO 16061:2021)
General Information
Relations
Standards Content (Sample)
SLOVENSKI STANDARD
SIST EN ISO 16061:2021
01-julij-2021
Nadomešča:
SIST EN ISO 16061:2015
Instrumenti, ki se uporabljajo pri neaktivnih kirurških vsadkih (implantatih) -
Splošne zahteve (ISO 16061:2021)
Instruments for use in association with non-active surgical implants - General
requirements (ISO 16061:2021)
Instrumente, die in Verbindung mit nichtaktiven chirurgischen Implantaten verwendet
werden - Allgemeine Anforderungen (ISO 16061:2021)Instruments à utiliser en association avec les implants chirurgicaux non actifs -
Exigences générales (ISO 16061:2021)Ta slovenski standard je istoveten z: EN ISO 16061:2021
ICS:
11.040.30 Operacijski instrumenti in Surgical instruments and
materiali materials
SIST EN ISO 16061:2021 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
---------------------- Page: 1 ----------------------SIST EN ISO 16061:2021
---------------------- Page: 2 ----------------------
SIST EN ISO 16061:2021
EN ISO 16061
EUROPEAN STANDARD
NORME EUROPÉENNE
March 2021
EUROPÄISCHE NORM
ICS 11.040.40; 11.040.99 Supersedes EN ISO 16061:2015
English Version
Instruments for use in association with non-active surgical
implants - General requirements (ISO 16061:2021)
Instruments à utiliser en association avec les implants Instrumente, die in Verbindung mit nichtaktiven
chirurgicaux non actifs - Exigences générales (ISO chirurgischen Implantaten verwendet werden -
16061:2021) Allgemeine Anforderungen (ISO 16061:2021)This European Standard was approved by CEN on 5 January 2021.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and
United Kingdom.EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 16061:2021 E
worldwide for CEN national Members.---------------------- Page: 3 ----------------------
SIST EN ISO 16061:2021
EN ISO 16061:2021 (E)
Contents Page
European foreword ....................................................................................................................................................... 3
---------------------- Page: 4 ----------------------SIST EN ISO 16061:2021
EN ISO 16061:2021 (E)
European foreword
This document (EN ISO 16061:2021) has been prepared by Technical Committee ISO/TC 150 "Implants
for surgery" in collaboration with Technical Committee CEN/TC 285 “Non-active surgical implants” the
secretariat of which is held by DIN.This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by September 2021, and conflicting national standards
shall be withdrawn at the latest by September 2021.Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN ISO 16061:2015.According to the CEN-CENELEC Internal Regulations, the national standards organizations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,
Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland,
Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of
North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the
United Kingdom.Endorsement notice
The text of ISO 16061:2021 has been approved by CEN as EN ISO 16061:2021 without any modification.
---------------------- Page: 5 ----------------------SIST EN ISO 16061:2021
---------------------- Page: 6 ----------------------
SIST EN ISO 16061:2021
INTERNATIONAL ISO
STANDARD 16061
Fourth edition
2021-03
Instruments for use in association
with non-active surgical implants —
General requirements
Instruments à utiliser en association avec les implants chirurgicaux
non actifs — Exigences générales
Reference number
ISO 16061:2021(E)
ISO 2021
---------------------- Page: 7 ----------------------
SIST EN ISO 16061:2021
ISO 16061:2021(E)
COPYRIGHT PROTECTED DOCUMENT
© ISO 2021
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting
on the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address
below or ISO’s member body in the country of the requester.ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii © ISO 2021 – All rights reserved
---------------------- Page: 8 ----------------------
SIST EN ISO 16061:2021
ISO 16061:2021(E)
Contents Page
Foreword ........................................................................................................................................................................................................................................iv
Introduction ................................................................................................................................................................................................................................vi
1 Scope ................................................................................................................................................................................................................................. 1
2 Normative references ...................................................................................................................................................................................... 1
3 Terms and definitions ..................................................................................................................................................................................... 2
4 Intended performance ................................................................................................................................................................................... 2
5 Design attributes .................................................................................................................................................................................................. 3
6 Selection of materials ...................................................................................................................................................................................... 4
7 Design evaluation ................................................................................................................................................................................................ 4
7.1 General ........................................................................................................................................................................................................... 4
7.2 Pre-clinical evaluation ...................................................................................................................................................................... 4
7.3 Clinical evaluation ................................................................................................................................................................................ 5
7.4 Post-market surveillance ............................................................................................................................................................... 5
8 Manufacture .............................................................................................................................................................................................................. 6
9 Sterilization ............................................................................................................................................................................................................... 6
9.1 Instruments supplied sterile ....................................................................................................................................................... 6
9.2 Instruments supplied non-sterile ........................................................................................................................................... 6
9.3 Instruments that are resterilizable ....................................................................................................................................... 6
10 Packaging ..................................................................................................................................................................................................................... 6
10.1 Protection from damage in transport, storage and handling.......................................................................... 6
10.2 Maintenance of sterility in transport, storage and handling ........................................................................... 7
11 Information supplied by the manufacturer ............................................................................................................................. 7
11.1 General ........................................................................................................................................................................................................... 7
11.2 Marking on instruments ................................................................................................................................................................. 8
11.3 Label ................................................................................................................................................................................................................. 8
11.4 Instructions for use ............................................................................................................................................................................. 9
11.5 Additional information for instruments with a measuring function ....................................................10
Annex A (informative) Examples of typical applications of instruments to be used in
association with non-active surgical implants and materials found acceptable forinstrument manufacture ...........................................................................................................................................................................11
Bibliography .............................................................................................................................................................................................................................13
© ISO 2021 – All rights reserved iii---------------------- Page: 9 ----------------------
SIST EN ISO 16061:2021
ISO 16061:2021(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www .iso .org/ directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www .iso .org/ patents).Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to
the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see
www .iso .org/ iso/ foreword .html.This document was prepared by Technical Committee ISO/TC 150, Implants for surgery, in collaboration
with the European Committee for Standardization (CEN) Technical Committee CEN/TC 285, Non-active
surgical implants, in accordance with the Agreement on technical cooperation between ISO and CEN
(Vienna Agreement).This fourth edition cancels and replaces the third edition (ISO 16061:2015), which has been technically
revised. The main changes compared to the previous edition are as follows:— A requirement to include intended purpose has been added in the list of items to be included when
establishing the intended performance of the instrument.— The list of design attributes in Clause 5 has been reorganized and several new attributes have added
to the list.— The selection of materials to be used in the instrument has been based on a risk analysis and the
clause now includes a list of the minimum factors to be considered in the risk analysis.
— The requirement for pre-clinical evaluation has been expanded and includes the requirement for
testing and biological evaluation of the final instrument.— A clinical evaluation of the instrument has been added as a requirement in all cases. However, if the
pre-clinical evaluation demonstrates the safety and intended performance of the instrument in the
conditions of intended use, the results of the pre-clinical evaluation will satisfy the requirement for
the clinical evaluation.— A new requirement for post-market surveillance has been added to Clause 7.
— The requirements in Clause 11 have been reorganized and clarified to reflect current practice and
to reference ISO 17664:2017, Clause 6 for instructions for applicable processing step (i.e. cleaning,
disinfection, drying, packaging, and sterilization) that need to be carried out by someone other than
the manufacturer.iv © ISO 2021 – All rights reserved
---------------------- Page: 10 ----------------------
SIST EN ISO 16061:2021
ISO 16061:2021(E)
— Annex A has been simplified to provide more consistent guidance on selection of material using a
risk-based approach. The stainless-steel grade material characteristic tables have been removed.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www .iso .org/ members .html.© ISO 2021 – All rights reserved v
---------------------- Page: 11 ----------------------
SIST EN ISO 16061:2021
ISO 16061:2021(E)
Introduction
This document provides a method of addressing the fundamental principles outlined in ISO/TR 14283
as they apply to instruments to be used in association with non-active surgical implants. It also provides
a method that can be used to demonstrate compliance with applicable regulatory requirements
relevant to the general safety and performance of medical devices as they apply to instruments used in
association with non-active surgical implants.There are three levels of standards dealing with instruments to be used in association with non-active
surgical implants. They are as follows, with level 1 being the highest.— Level 1: general requirements for instruments to be used in association with non-active surgical
implants.— Level 2: particular requirements for families of instruments to be used in association with non-
active surgical implants.— Level 3: specific requirements for types of instruments to be used in association with non-active
surgical implants.Level 1 standards include this document which contains requirements that apply to all instruments to
be used in association with non-active surgical implants, ISO 14630, which contains the requirement
for non-active surgical implants and ISO 14708-1, which contains requirements for active implants.
They also anticipate that there are additional requirements in the level 2 and level 3 standards.
Level 2 standards apply to a more restricted set or family of instruments, such as those designed for use
with non-active surgical implants used in neurosurgery, cardiovascular surgery, or joint replacement.
Level 3 standards apply to specific types of instruments within a family of instruments used in
association with non-active surgical implants, such as hip joints or arterial stents.
To address all requirements for a specific instrument, it is advisable that the standard of the lowest
available level be consulted first.Compliance with a level 3 standard is intended to imply compliance with the applicable level 2
standards, if available, and with the applicable level 1 standard.NOTE The requirements in this document correspond to international consensus. Individual or national
standards or regulatory bodies can prescribe other requirements.vi © ISO 2021 – All rights reserved
---------------------- Page: 12 ----------------------
SIST EN ISO 16061:2021
INTERNATIONAL STANDARD ISO 16061:2021(E)
Instruments for use in association with non-active surgical
implants — General requirements
1 Scope
This document specifies the general requirements for instruments to be used in association with non-
active surgical implants. These requirements apply to instruments when they are manufactured and
when they are supplied after refurbishment.NOTE In this document, unless otherwise specified, the term “instrument” refers to an instrument for use in
association with non-active surgical implants.This document also applies to instruments which can be connected to power-driven systems, but it
does not apply to the power-driven systems themselves.With regard to safety, this document gives the requirements for intended performance, design
attributes, materials, design evaluation, manufacture, sterilization, packaging, and information
supplied by the instrument manufacturer, hereafter referred to as the manufacturer.
This document is not applicable to instruments associated with dental implants, transendodontic and
transradicular implants and ophthalmic implants.2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 8601-1, Date and time — Representations for information interchange — Part 1: Basic rules
ISO 10993-1, Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk
management processISO 11135, Sterilization of health-care products — Ethylene oxide — Requirements for the development,
validation and routine control of a sterilization process for medical devicesISO 11137-1, Sterilization of health care products — Radiation — Part 1: Requirements for development,
validation and routine control of a sterilization process for medical devicesISO 11137-2, Sterilization of health care products — Radiation — Part 2: Establishing the sterilization dose
ISO 11137-3, Sterilization of health care products — Radiation — Part 3: Guidance on dosimetric aspects of
development, validation and routine controlISO 11607-1, Packaging for terminally sterilized medical devices — Part 1: Requirements for materials,
sterile barrier systems and packaging systemsISO 11607-2, Packaging for terminally sterilized medical devices — Part 2: Validation requirements for
forming, sealing and assembly processesISO 14155, Clinical investigation of medical devices for human subjects — Good clinical practice
ISO 14937, Sterilization of health care products — General requirements for characterization of a sterilizing
agent and the development, validation and routine control of a sterilization process for medical devices
ISO 14971, Medical devices — Application of risk management to medical devices© ISO 2021 – All rights reserved 1
---------------------- Page: 13 ----------------------
SIST EN ISO 16061:2021
ISO 16061:2021(E)
ISO 17664, Processing of health care products — Information to be provided by the medical device
manufacturer for the processing of medical devicesISO 17665-1, Sterilization of health care products — Moist heat — Part 1: Requirements for the development,
validation and routine control of a sterilization process for medical devicesISO 25424, Sterilization of health care products — Low temperature steam and formaldehyde —
Requirements for development, validation and routine control of a sterilization process for medical devices
ISO 80000-1, Quantities and units — Part 1: General3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 14971 and the following
terms and definitions apply.ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at http:// www .electropedia .org/— ISO Online browsing platform: available at https:// www .iso .org/ obp
3.1
associated implant
specific non-active surgical implant (3.4) associated with a specific instrument (3.3) intended to be used
during a surgical procedure3.2
final instrument
instrument (3.3) that has been subjected to all manufacturing processes for the “to be marketed”
instrument including packaging and if applicable, sterilizationNote 1 to entry: Derived from the definition of “final product” in ISO 10993-1:2018, 3.8 to reinforce that the
instrument to be evaluated is the final product referred to in ISO 10993-1.3.3
instrument
non-active medical device intended for use during surgical procedures related to specific non-active
surgical implants (3.4)Note 1 to entry: Examples of typical applications of instruments to be used in association with non-active surgical
implants are presented in A.1.3.4
non-active surgical implant
surgical implant, the operation of which does not depend on a source of electrical energy or any source
of power other than that directly generated by the human body or gravity[SOURCE: ISO 14630:2012, 3.6, modified — the second term "implant" has been deleted.]
4 Intended performanceThe intended performance of an instrument shall be described and documented by addressing the
following, with particular regard to safety:a) intended purpose;
b) functional characteristics;
c) intended conditions of use.
2 © ISO 2021 – All rights reserved
---------------------- Page: 14 ----------------------
SIST EN ISO 16061:2021
ISO 16061:2021(E)
Instruments shall be evaluated to demonstrate that the intended performance is achieved (see
Clause 7).NOTE Information to support the description of the intended performance can be found in sources such as:
— published standards;— published clinical and scientific literature;
— validated test results.
5 Design attributes
The development of the design attributes of an instrument to meet the performance intended by the
manufacturer shall take into account at least the following:a) role of the instrument in conjunction with the associated non-active surgical implant (e.g.
implantation, positioning, alignment, removal);b) biocompatibility of materials for their intended use, including the influence of material by-products
from manufacturing and chemical residuals;c) physical, mechanical, biological and chemical properties of the instrument materials;
d) potential deterioration of the material characteristics;e) usability;
f) compatibility with intended processing agent (e.g. cleaning agent, disinfectant, and sterilizing
agent) and intended processing conditions (e.g. temperature, pressure, vacuum, humidity, time) for
instruments requiring user processing (e.g. sterilization) or intended to be reused;
g) stability of instrument materials under intended manufacturing conditions (e.g. chemicals,
temperature, pressure, vacuum, humidity, time);h) potential deterioration of the form and/or function of the instrument due to repeated use and
reprocessing;i) ease of cleaning, disinfection (if intended by the manufacturer) and sterilization both by the
manufacturer and the user;j) ease of maintenance;
k) effects of contact between the instrument and the body;
l) effects of contact between the instrument and the non-active surgical implant, and other
instruments;m) shape and dimensions of the instrument, including their possible effects on the body or the non-
active surgical implant;n) wear characteristics of materials and the effect of wear and wear products on the instrument, the
body or the non-active surgical implant;o) insertion, removal, and interconnection of parts;
p) extent of fluid leakage and/or diffusion of substances into or out of instruments;
q) accuracy and stability of the measurement for instruments with a measuring function;
r) reciprocal interference with other devices in the specified use environment;© ISO 2021 – All rights reserved 3
---------------------- Page: 15 ----------------------
SIST EN ISO 16061:2021
ISO 16061:2021(E)
EXAMPLE Compatibility with diagnostic imaging systems such magnetic resonance imaging (MRI)
equipment. See References [35], [37] and [38] for standards related to the hazards associated with the
magnetic resonance environment.s) ability of the instrument or fragment of instrument to be located by means of an external imaging
device [see 11.4, q)];t) compatibility with medicinal substances incorporated into or intended to be used with the
instrument.6 Selection of materials
The selection of materials to be used for the manufacture of instruments shall be based on a risk
analysis, which shall take into account at least the following:a) the properties required for the intended purpose considering factors such as mechanical/
functional requirements, anatomical location, dimensions, geometry, and conditions for use; and,
duration and frequency of use over the intended lifetime of the instrument;b) any intended treatment to the material or to the surface of the instrument (e.g. chemical, electro-
chemical, thermal, mechanical, coating);c) the effects of instrument manufacturing, handling, cleaning, packaging, sterilization, storage, and
processing (if applicable);NOTE 1 For information on processing of health care products, see ISO 17664.
d) possible adverse reactions by the human body and body fluids to the instrument materials;
EXAMPLE 1 Adverse reaction to leachable chemicals, degradation products, additives (e.g. plasticisers
and fillers) and impurities.e) possible adverse reactions between the non-active surgical implant materials and instrument
materials in the presence of body fluids.EXAMPLE 2 Electrochemical corrosion.
NOTE 2 Annex A lists some of the materials that have been found acceptable in certain applications. Reference
to Annex A does not eliminate the need to conduct a risk analysis taking into account the factors outlined above.
A biological evaluation of the final instrument shall be performed and shall form part of the pre-clinical
evaluation (see 7.2).7 Design evaluation
7.1 General
Instruments shall be evaluated in association with the non-active surgical implant they are designed
for, in order to demonstrate that the safety and the intended performance is achieved (see Clause 4).
Safety and intended performance shall be demonstrated by pre-clinical evaluation, clinical evaluation
and post-market surveillance, including appropriate risk management at all stages of the life cycle of
the instrument.NOTE Further information on risk management can be found in ISO 14971.
7.2 Pre-clinical evaluation
Instruments shall undergo pre-clinical evaluation based on a critical review of:
4 © ISO 2021 – All rights reserved
---------------------- Page: 16 ----------------------
SIST
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.