SIST EN 14626:2025/oprA1:2026
(Amendment)Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy
- Abstract
This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle.
This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use.
The method is applicable to the determination of the mass concentration of carbon monoxide present in ambient air up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for the type approval test.
NOTE 1 Other ranges may be used depending on the levels present in ambient air.
NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply.
The method covers the determination of ambient air concentrations of carbon monoxide in zones classified as rural areas, urban-background areas and traffic-orientated locations and locations influenced by industrial sources.
The results are expressed in mg/m3 (at 20 °C and 101,3 kPa).
NOTE 3 100 mg/m3 of CO corresponds to 86 μmol/mol of CO.
This standard contains information for different groups of users.
Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment.
Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about:
− type-approval test conditions, test procedures and test requirements;
− analyser performance requirements;
− evaluation of the type-approval test results;
− evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type approval test results.
Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about:
− initial installation of the analyser in the monitoring network and acceptance testing;
− ongoing quality assurance/quality control;
− calculation and reporting of measurement results;
− evaluation of the uncertainty of measurement results under practical monitoring conditions.
- Status
- Not Published
- Public Enquiry End Date
- 02-Jan-2027
- Technical Committee
- KAZ - Air quality
- Current Stage
- 4020 - Public enquire (PE) (Adopted Project)
- Start Date
- 03-Nov-2026
- Due Date
- 23-Mar-2027
SIST EN 14626:2025/oprA1:2026 is the Slovenian adoption of the CEN draft amendment EN 14626:2024/prA1:2026 for measuring carbon monoxide in ambient air by non-dispersive infrared spectroscopy. It updates the method used to assess analyser performance, type approval, and suitability for fixed-site ambient-air monitoring. The standard is aimed at test houses, laboratories, and monitoring networks that measure carbon monoxide (CO) in outdoor air.
What does SIST EN 14626:2025/oprA1:2026 specify?
SIST EN 14626:2025/oprA1:2026 amends the European standard for continuous measurement of carbon monoxide in ambient air. The method applies up to 100 mg/m3 CO, which is the certification range for type approval testing, and results are expressed in mg/m3 at 20 °C and 101.3 kPa.
The standard covers rural areas, urban-background areas, traffic-orientated locations, and locations influenced by industrial sources. It also links the measurement method to the data quality needs of Directive 2008/50/EC [1] for fixed-site use.
| Part of the document | What it covers |
|---|---|
| Clauses 5 to 7 and Annexes B, C and D | General principles of carbon monoxide measurement and sampling equipment |
| Clause 8 and Annex E | Type-approval testing, analyser performance, result evaluation and uncertainty from test results |
| Clauses 9 to 11 and Annex F | Installation in monitoring networks, quality assurance/quality control, calculation and reporting, and practical uncertainty |
What are the key requirements of SIST EN 14626:2025/oprA1:2026?
SIST EN 14626:2025/oprA1:2026 mainly corrects and tightens the parts of EN 14626 that control type approval and field use. The important point in practice is that analysers, test houses and monitoring networks must follow the revised clauses, tables, figures and formulas rather than the earlier wording.
Scope and installation
- Clause 3 updates the definition of ambient air so it excludes workplaces as defined by Directive 89/654/EEC. In practice, the method is for outdoor ambient-air monitoring, not workplace air.
- Clause 6.5 tightens the sampling pump to manifold requirement and changes the referenced clause. This matters because the sampling train must not add a noticeable bias before the gas reaches the analyser.
Type approval testing
- Clause 8.2 and Table 1 set the performance characteristics used for type approval. The amendment updates the criteria for long-term drift, short-term drift and sensitivity to sample gas temperature and electrical voltage, so approval testing has to check the analyser against the revised limits.
- Clause 8.4.6 revises the lack-of-fit test for the calibration function. It now uses a defined sequence of concentration levels across the certification range and requires a stable reading after each change, which means the analyser is checked across the range rather than at only a few points.
- Clause 8.4.4 now states the short-term drift requirement at span level. This is the stability check that shows whether the analyser remains steady over a 12-hour period.
Routine monitoring and uncertainty
- Clause 8.5.4 revises the long-term drift calculation, and Annex F revises the combined standard uncertainty formula. In practice, laboratories and operators need the updated expressions when they calculate measurement quality.
- Clauses 9.5.1 and 9.5.5 update field-use references and data treatment after performance criteria are exceeded, and Clause 9.5.4 is deleted. This matters for monitoring networks because acceptance testing and follow-up actions must match the revised procedure set.
What terms does SIST EN 14626:2025/oprA1:2026 define?
- Ambient air - Outdoor air in the troposphere, excluding workplaces as defined by Directive 89/654/EEC, and places with regular public access.
- Certification range - The concentration range used for type-approval testing, up to 100 mg/m3 CO in this standard.
- Type approval - The test-based process used to show that a carbon monoxide analyser meets the required performance characteristics.
- Fixed site - The specific monitoring location where the analyser is installed and evaluated for suitability.
- Sampling manifold - The part of the sampling system that carries the sample gas to the analyser.
- Combined standard uncertainty - The overall uncertainty obtained by combining the separate uncertainty contributions used in Annex F.
Who uses SIST EN 14626:2025/oprA1:2026?
SIST EN 14626:2025/oprA1:2026 is used by test houses and laboratories that perform type-approval testing of carbon monoxide analysers. It is also used by monitoring networks that install analysers, perform acceptance testing, run ongoing quality assurance/quality control, calculate results, and evaluate measurement uncertainty.
In practice, the main users are:
- analyser manufacturers and suppliers
- test houses and laboratories
- ambient air monitoring network operators
- air quality authorities and technical assessors
- engineers responsible for installation, calibration and reporting
What changed in SIST EN 14626:2025/oprA1:2026 from the previous edition?
SIST EN 14626:2025/oprA1:2026 modifies EN 14626:2024. The foreword says it addresses critical errors identified in the earlier standard.
The amendment:
- removes the note about stopping the use of older reference analysers
- updates the definition of ambient air
- corrects the sampling pump to manifold references and limit
- changes the annex reference in Clause 8.2
- revises Table 1 performance criteria
- replaces Figure 1
- corrects the short-term drift wording
- clarifies the linearity test sequence and stability requirement
- corrects temperature and voltage sensitivity wording
- updates the long-term drift formula
- changes table references in Clause 9.5.1
- deletes Clause 9.5.4
- removes one data-treatment reference in Clause 9.5.5
- corrects formulas in Annexes D, E and F
Which standards are used with SIST EN 14626:2025/oprA1:2026?
SIST EN 14626:2025/oprA1:2026 is based on EN 14626:2024, which it modifies. It also refers to Council Directive 2008/50/EC [1] for the ambient-air data quality context and to Directive 89/654/EEC in the updated definition of ambient air.
What does the SIST EN 14626:2025/oprA1:2026 document contain?
SIST EN 14626:2025/oprA1:2026 contains a European foreword and 19 modification items. It revises text in Clauses 1, 3, 6.5, 8.2, 8.4.3.2, 8.4.4, 8.4.6, 8.4.8, 8.4.10, 8.5.4, 9.5.1, 9.5.4 and 9.5.5, plus Annexes D, E and F.
The technical content focuses on a revised test figure, updated performance tables, corrected formulas, and clarified procedures for type approval and field monitoring.
Relations
- Effective Date
- 01-Dec-2026
Get Certified
Connect with accredited certification bodies for this standard
CIS Institut d.o.o.
Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

Kiwa BDA Testing
Building and construction product certification.
Kmetijski inštitut Slovenije
Agricultural Institute of Slovenia. Soil testing, plant health, agricultural product analysis.
Sponsored listings
Frequently Asked Questions
SIST EN 14626:2025/oprA1:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy". This standard covers: This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use. The method is applicable to the determination of the mass concentration of carbon monoxide present in ambient air up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for the type approval test. NOTE 1 Other ranges may be used depending on the levels present in ambient air. NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply. The method covers the determination of ambient air concentrations of carbon monoxide in zones classified as rural areas, urban-background areas and traffic-orientated locations and locations influenced by industrial sources. The results are expressed in mg/m3 (at 20 °C and 101,3 kPa). NOTE 3 100 mg/m3 of CO corresponds to 86 μmol/mol of CO. This standard contains information for different groups of users. Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about: − type-approval test conditions, test procedures and test requirements; − analyser performance requirements; − evaluation of the type-approval test results; − evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type approval test results. Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about: − initial installation of the analyser in the monitoring network and acceptance testing; − ongoing quality assurance/quality control; − calculation and reporting of measurement results; − evaluation of the uncertainty of measurement results under practical monitoring conditions.
This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use. The method is applicable to the determination of the mass concentration of carbon monoxide present in ambient air up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for the type approval test. NOTE 1 Other ranges may be used depending on the levels present in ambient air. NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply. The method covers the determination of ambient air concentrations of carbon monoxide in zones classified as rural areas, urban-background areas and traffic-orientated locations and locations influenced by industrial sources. The results are expressed in mg/m3 (at 20 °C and 101,3 kPa). NOTE 3 100 mg/m3 of CO corresponds to 86 μmol/mol of CO. This standard contains information for different groups of users. Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about: − type-approval test conditions, test procedures and test requirements; − analyser performance requirements; − evaluation of the type-approval test results; − evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type approval test results. Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about: − initial installation of the analyser in the monitoring network and acceptance testing; − ongoing quality assurance/quality control; − calculation and reporting of measurement results; − evaluation of the uncertainty of measurement results under practical monitoring conditions.
SIST EN 14626:2025/oprA1:2026 is classified under the following ICS (International Classification for Standards) categories: 13.040.20 - Ambient atmospheres. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN 14626:2025/oprA1:2026 has the following relationships with other standards: It is inter standard links to SIST EN 14626:2025. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST EN 14626:2025/oprA1:2026 is associated with the following European legislation: EU Directives/Regulations: 2008/50/EC. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
SIST EN 14626:2025/oprA1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
01-december-2026
Zunanji zrak - Standardna metoda za določanje koncentracije ogljikovega
monoksida z nedisperzivno infrardečo spektroskopijo - Dopolnilo A1
Ambient air - Standard method for the measurement of the concentration of carbon
monoxide by non-dispersive infrared spectroscopy
Außenluft - Messverfahren zur Bestimmung der Konzentration von Kohlenstoffmonoxid
mit nicht-dispersiver Infrarot-Photometrie
Air ambiant - Méthode normalisée de mesurage de la concentration en monoxyde de
carbone par spectroscopie à rayonnement infrarouge non dispersif
Ta slovenski standard je istoveten z: EN 14626:2024/prA1
ICS:
13.040.20 Kakovost okoljskega zraka Ambient atmospheres
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
DRAFT
EUROPEAN STANDARD
EN 14626:2024
NORME EUROPÉENNE
EUROPÄISCHE NORM
prA1
October 2026
ICS
English Version
Ambient air - Standard method for the measurement of the
concentration of carbon monoxide by non-dispersive
infrared spectroscopy
Air ambiant - Méthode normalisée de mesurage de la Außenluft - Messverfahren zur Bestimmung der
concentration en monoxyde de carbone par Konzentration von Kohlenstoffmonoxid mit nicht-
spectroscopie à rayonnement infrarouge non dispersif dispersiver Infrarot-Photometrie
This draft amendment is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee CEN/TC 264.
This draft amendment A1, if approved, will modify the European Standard EN 14626:2024. If this draft becomes an amendment,
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for inclusion of
this amendment into the relevant national standard without any alteration.
This draft amendment was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.
Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 14626:2024/prA1:2026 E
worldwide for CEN national Members.
EN 14626 : 2024/prA1 :2026
Contents Page
European foreword . 3
1 Modification to Clause 1 ”Scope” . 4
2 Modification to Clause 3 “Terms and definitions” . 4
3 Modifications to 6.5 “Sampling pump to manifold” . 4
4 Modifications to 8.2, “Relevant performance characteristics and criteria”, . 4
5 Modification to 8.2, “Relevant performance characteristics and criteria”, Table 1 . 5
6 Modification to 8.2, “Relevant performance characteristics and criteria”, Table 1 . 5
7 Modification to 8.4.3.2, “Test procedure” . 5
8 Modification to 8.4.4, “Short-term drift” . 6
9 Modification to 8.4.6, “Lack of fit of linearity of the calibration function” . 6
10 Modification to 8.4.8, ” Sensitivity coefficient to the sample gas temperature” . 6
11 Modification to 8.4.10, ” Sensitivity coefficient to electrical voltage” . 6
12 Modification to 8.5.4, ” Long-term drift”, Formula (15) . 7
13 Modification to 9.5.1, “General” . 7
14 Modification to 9.5.4, “Testing the sampling manifold” . 7
15 Modification to 9.5.5, “Treatment of data after exceedance of performance criteria” 7
16 Modifications to D.1, “Procedure for applying test gas ”, Formula (D.1) and D.2 . 8
17 Modification to E.3.2.5.5, “Treatment of data after exceedance of performance
criteria”. 8
18 Modification to E.4, ” Type testing Requirement c)” . 8
19 Modification to F.2, “Combined standard uncertainty”. 9
European foreword
This document (EN 14626:2024/prA1:2026) has been prepared by Technical Committee CEN/TC 264
“Air quality”, the secretariat of which is held by DIN.
This document is currently submitted to the CEN Enquiry.
This amendment addresses only the critical errors identified in the EN 14626:2024. Errors considered as
minor by CEN/TC 264/WG 12 will be addressed in the subsequent revision of EN 14626.
This document has been prepared under a Standardization Request given to CEN by the European
Commission and the European Free Trade Association, and supports essential requirements of
EU Directive(s) / Regulation(s).
EN 14626 : 2024/prA1 :2026
1 Modification to Clause 1 ”Scope”
Remove the following sentence from Note 4:
“As newer versions of analysers tested under this document become available, discontinue the
...