Abstract

This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle.
This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use.
The method is applicable to the determination of the mass concentration of carbon monoxide present in ambient air up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for the type approval test.
NOTE 1 Other ranges may be used depending on the levels present in ambient air.
NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply.
The method covers the determination of ambient air concentrations of carbon monoxide in zones classified as rural areas, urban-background areas and traffic-orientated locations and locations influenced by industrial sources.
The results are expressed in mg/m3 (at 20 °C and 101,3 kPa).
NOTE 3 100 mg/m3 of CO corresponds to 86 μmol/mol of CO.
This standard contains information for different groups of users.
Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment.
Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about:
− type-approval test conditions, test procedures and test requirements;
− analyser performance requirements;
− evaluation of the type-approval test results;
− evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type approval test results.
Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about:
− initial installation of the analyser in the monitoring network and acceptance testing;
− ongoing quality assurance/quality control;
− calculation and reporting of measurement results;
− evaluation of the uncertainty of measurement results under practical monitoring conditions.

Status
Not Published
Publication Date
03-Aug-2028
Technical Committee
CEN/TC 264 - Air quality
Current Stage
4020 - Submission to enquiry - Enquiry
Start Date
22-Oct-2026
Due Date
23-Apr-2027
Completion Date
22-Oct-2026

Buy Documents

Draft

EN 14626:2025/oprA1:2026

English language (9 pages)
Preview
Preview
e-Library read for
×1 day

EN 14626:2024/prA1 is a CEN draft amendment to EN 14626:2024 for measuring carbon monoxide in ambient air by non-dispersive infrared spectroscopy. It updates the method used to select and assess analysers, and it clarifies how the procedure is applied in type approval testing and in fixed-site monitoring networks. The document is aimed at laboratories, test houses and ambient-air monitoring operators.

What does EN 14626:2024/prA1 specify?

EN 14626:2024/prA1 modifies the European Standard EN 14626:2024, which specifies a continuous measurement method for carbon monoxide in ambient air using the non-dispersive infrared spectroscopic measuring principle. It also covers analyser performance, suitability for fixed-site use, sampling, calibration and quality assurance for monitoring data.

The scope applies to ambient-air measurements in rural, urban-background, traffic-oriented and industrial-influenced locations. The method is tied to the certification range used for type approval testing, and the results are expressed as mass concentration.

This draft amendment is organized as a set of targeted changes to Clause 1, Clause 3, Clauses 6, 8 and 9, and Annexes D, E and F.

AnnexWhat it covers
Annex BGeneral information on the principles of carbon monoxide measurement
Annex CGeneral information on sampling equipment
Annex DTest-gas application procedure and related formulas
Annex EType-approval testing for carbon monoxide analysers
Annex FMonitoring-network use, including uncertainty evaluation

What are the key requirements of EN 14626:2024/prA1?

EN 14626:2024/prA1 mainly corrects and clarifies the technical checks used to qualify an analyser and to operate it in a monitoring network.

Type approval and performance checks

Clause 8 and Annex E are the core for laboratories and test houses. They cover type-approval test conditions, test procedures, analyser performance requirements, evaluation of the test results, and evaluation of measurement uncertainty from the test outcome. In practice, this is the part used to decide whether a carbon monoxide analyser is suitable for official ambient-air monitoring.

Updated performance criteria

Clause 8.2 and Table 1 are revised so that the performance criteria stay aligned with the document’s method language. The criteria cover drift, linearity and sensitivity to operating influences. For users, this means the analyser has to be checked not only for CO response, but also for stability and sensitivity to installation and operating conditions.

Sampling path requirements

Clause 6.5 on the sampling pump to manifold is corrected. The change tightens the stated requirement and fixes a cross-reference. In practice, this matters because the sampling path can affect measurement quality before the gas even reaches the analyser.

Linearity and drift testing

Clause 8.4.6 revises the procedure for testing lack of fit of the calibration function, including the applied concentration sequence and the need to wait for a stable reading after each change. Clause 8.4.4 and Clause 8.5.4 correct the treatment of short-term and long-term drift. For laboratories, this means the analyser has to demonstrate stable behaviour across the calibration range and over time.

Sensitivity to operating conditions

Clause 8.4.8 and Clause 8.4.10 update the checks for sensitivity to sample gas temperature and electrical voltage. These checks matter because ambient-air stations often run unattended, so the analyser has to remain reliable when environmental or power conditions vary.

Monitoring-network operation

Clauses 9 to 11 and Annex F cover installation in the monitoring network, acceptance testing, ongoing QA/QC, calculation and reporting of results, and uncertainty under practical monitoring conditions. This is the part operators use after the analyser has passed type approval.

What terms does EN 14626:2024/prA1 define?

  • Ambient air - outdoor air in the troposphere, excluding workplaces where health and safety at work apply and areas not regularly open to the public.
  • Certification range - the concentration range used for type approval testing of the analyser.
  • Type approval test - the qualification testing used to show that an analyser meets the method’s performance requirements.
  • Fixed site - a monitoring location where the analyser is installed for routine ambient-air measurements.
  • Sampling manifold - the gas path that delivers sample gas to the analyser and can influence test and operating results.
  • QA/QC - quality assurance/quality control, the routine checks that keep monitoring data reliable.
  • Combined standard uncertainty - the uncertainty value built from several contributing components, used in Annex F.

Who uses EN 14626:2024/prA1?

EN 14626:2024/prA1 is used by test houses and laboratories that perform type-approval testing of carbon monoxide analysers. It is also used by monitoring networks that install analysers, run acceptance tests, carry out QA/QC and report ambient-air results.

It is relevant to air-quality engineers, environmental monitoring teams, regulators and buyers selecting an analyser for official ambient-air use. Manufacturers and suppliers also use it to design and validate instruments against the required performance checks.

What changed in EN 14626:2024/prA1 from the previous edition?

EN 14626:2024/prA1 modifies EN 14626:2024 and addresses critical errors identified in that edition.

The changes stated in the document include:

  • removal of the note in the scope about discontinuing older reference analysers
  • revision of the ambient air definition
  • correction of the sampling pump to manifold text and threshold
  • clarification of the performance criteria reference in Clause 8.2
  • corrections to Table 1 for drift-related criteria
  • replacement of the test procedure figure in Clause 8.4.3.2
  • revision of the short-term drift text
  • revision of the linearity test procedure
  • correction of the sensitivity checks for sample gas temperature and electrical voltage
  • correction of the long-term drift formula and variable definitions
  • updates in Clause 9.5, including deletion of one clause and a data-treatment correction
  • corrections in Annex D, Annex E and Annex F

Which standards are used with EN 14626:2024/prA1?

The document is linked to Council Directive 2008/50/EC, which provides the air-quality framework and the data quality context for ambient-air monitoring.

It also uses Directive 89/654/EEC in the revised definition of ambient air, to exclude workplaces where health and safety at work apply.

What does the EN 14626:2024/prA1 document contain?

EN 14626:2024/prA1 contains clause-by-clause corrections rather than a full rewrite. It changes the scope note, one formal definition, several performance criteria, the test procedure figure, drift wording and formulas, plus the annex material for test-gas application, type testing and uncertainty evaluation.

It also updates the text used by laboratory users and monitoring-network operators, so the same method can be applied consistently in type approval and in field measurements. The annex changes are especially relevant for calculations, test setup and result treatment.

Relations

Effective Date
03-Jun-2026

Buy Documents

Draft

EN 14626:2025/oprA1:2026

English language (9 pages)
Preview
Preview
e-Library read for
×1 day

Get Certified

Connect with accredited certification bodies for this standard

CIS Institut d.o.o.

Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

SA Slovenia Verified

Kiwa BDA Testing

Building and construction product certification.

RVA Netherlands Verified

Kmetijski inštitut Slovenije

Agricultural Institute of Slovenia. Soil testing, plant health, agricultural product analysis.

SA Slovenia Verified

Sponsored listings

Frequently Asked Questions

EN 14626:2024/prA1 is a draft published by the European Committee for Standardization (CEN). Its full title is "Ambient air - Standard method for the measurement of the concentration of carbon monoxide by non-dispersive infrared spectroscopy". This standard covers: This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use. The method is applicable to the determination of the mass concentration of carbon monoxide present in ambient air up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for the type approval test. NOTE 1 Other ranges may be used depending on the levels present in ambient air. NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply. The method covers the determination of ambient air concentrations of carbon monoxide in zones classified as rural areas, urban-background areas and traffic-orientated locations and locations influenced by industrial sources. The results are expressed in mg/m3 (at 20 °C and 101,3 kPa). NOTE 3 100 mg/m3 of CO corresponds to 86 μmol/mol of CO. This standard contains information for different groups of users. Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about: − type-approval test conditions, test procedures and test requirements; − analyser performance requirements; − evaluation of the type-approval test results; − evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type approval test results. Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about: − initial installation of the analyser in the monitoring network and acceptance testing; − ongoing quality assurance/quality control; − calculation and reporting of measurement results; − evaluation of the uncertainty of measurement results under practical monitoring conditions.

This European Standard specifies a continuous measurement method for the determination of the concentration of carbon monoxide present in ambient air based on the non-dispersive infrared spectroscopic measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate non-dispersive infrared spectroscopic analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use. The method is applicable to the determination of the mass concentration of carbon monoxide present in ambient air up to 100 mg/m3 carbon monoxide. This concentration range represents the certification range for the type approval test. NOTE 1 Other ranges may be used depending on the levels present in ambient air. NOTE 2 When the standard is used for other purposes than for measurements required by Directive 2008/50/EC, the ranges and uncertainty requirements may not apply. The method covers the determination of ambient air concentrations of carbon monoxide in zones classified as rural areas, urban-background areas and traffic-orientated locations and locations influenced by industrial sources. The results are expressed in mg/m3 (at 20 °C and 101,3 kPa). NOTE 3 100 mg/m3 of CO corresponds to 86 μmol/mol of CO. This standard contains information for different groups of users. Clauses 5 to 7 and Annexes B, C and D contain general information about the principles of carbon monoxide measurement by non-dispersive infrared spectroscopic analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of carbon monoxide analysers. These sections contain information about: − type-approval test conditions, test procedures and test requirements; − analyser performance requirements; − evaluation of the type-approval test results; − evaluation of the uncertainty of the measurement results of the carbon monoxide analyser based on the type approval test results. Clauses 9 to 11 and Annex F are directed towards monitoring networks performing the practical measurements of carbon monoxide in ambient air. These sections contain information about: − initial installation of the analyser in the monitoring network and acceptance testing; − ongoing quality assurance/quality control; − calculation and reporting of measurement results; − evaluation of the uncertainty of measurement results under practical monitoring conditions.

EN 14626:2024/prA1 is classified under the following ICS (International Classification for Standards) categories: 13.040.20 - Ambient atmospheres. The ICS classification helps identify the subject area and facilitates finding related standards.

EN 14626:2024/prA1 has the following relationships with other standards: It is inter standard links to EN 14626:2024. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

EN 14626:2024/prA1 is associated with the following European legislation: EU Directives/Regulations: 2008/50/EC. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

EN 14626:2024/prA1 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


SLOVENSKI STANDARD
01-december-2026
Zunanji zrak - Standardna metoda za določanje koncentracije ogljikovega
monoksida z nedisperzivno infrardečo spektroskopijo - Dopolnilo A1
Ambient air - Standard method for the measurement of the concentration of carbon
monoxide by non-dispersive infrared spectroscopy
Außenluft - Messverfahren zur Bestimmung der Konzentration von Kohlenstoffmonoxid
mit nicht-dispersiver Infrarot-Photometrie
Air ambiant - Méthode normalisée de mesurage de la concentration en monoxyde de
carbone par spectroscopie à rayonnement infrarouge non dispersif
Ta slovenski standard je istoveten z: EN 14626:2024/prA1
ICS:
13.040.20 Kakovost okoljskega zraka Ambient atmospheres
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

DRAFT
EUROPEAN STANDARD
EN 14626:2024
NORME EUROPÉENNE
EUROPÄISCHE NORM
prA1
October 2026
ICS
English Version
Ambient air - Standard method for the measurement of the
concentration of carbon monoxide by non-dispersive
infrared spectroscopy
Air ambiant - Méthode normalisée de mesurage de la Außenluft - Messverfahren zur Bestimmung der
concentration en monoxyde de carbone par Konzentration von Kohlenstoffmonoxid mit nicht-
spectroscopie à rayonnement infrarouge non dispersif dispersiver Infrarot-Photometrie
This draft amendment is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee CEN/TC 264.

This draft amendment A1, if approved, will modify the European Standard EN 14626:2024. If this draft becomes an amendment,
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for inclusion of
this amendment into the relevant national standard without any alteration.

This draft amendment was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.

Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.

EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 14626:2024/prA1:2026 E
worldwide for CEN national Members.

EN 14626 : 2024/prA1 :2026
Contents Page
European foreword . 3
1 Modification to Clause 1 ”Scope” . 4
2 Modification to Clause 3 “Terms and definitions” . 4
3 Modifications to 6.5 “Sampling pump to manifold” . 4
4 Modifications to 8.2, “Relevant performance characteristics and criteria”, . 4
5 Modification to 8.2, “Relevant performance characteristics and criteria”, Table 1 . 5
6 Modification to 8.2, “Relevant performance characteristics and criteria”, Table 1 . 5
7 Modification to 8.4.3.2, “Test procedure” . 5
8 Modification to 8.4.4, “Short-term drift” . 6
9 Modification to 8.4.6, “Lack of fit of linearity of the calibration function” . 6
10 Modification to 8.4.8, ” Sensitivity coefficient to the sample gas temperature” . 6
11 Modification to 8.4.10, ” Sensitivity coefficient to electrical voltage” . 6
12 Modification to 8.5.4, ” Long-term drift”, Formula (15) . 7
13 Modification to 9.5.1, “General” . 7
14 Modification to 9.5.4, “Testing the sampling manifold” . 7
15 Modification to 9.5.5, “Treatment of data after exceedance of performance criteria” 7
16 Modifications to D.1, “Procedure for applying test gas ”, Formula (D.1) and D.2 . 8
17 Modification to E.3.2.5.5, “Treatment of data after exceedance of performance
criteria”. 8
18 Modification to E.4, ” Type testing Requirement c)” . 8
19 Modification to F.2, “Combined standard uncertainty”. 9

European foreword
This document (EN 14626:2024/prA1:2026) has been prepared by Technical Committee CEN/TC 264
“Air quality”, the secretariat of which is held by DIN.
This document is currently submitted to the CEN Enquiry.
This amendment addresses only the critical errors identified in the EN 14626:2024. Errors considered as
minor by CEN/TC 264/WG 12 will be addressed in the subsequent revision of EN 14626.
This document has been prepared under a Standardization Request given to CEN by the European
Commission and the European Free Trade Association, and supports essential requirements of
EU Directive(s) / Regulation(s).
EN 14626 : 2024/prA1 :2026
1 Modification to Clause 1 ”Scope”
Remove the following sentence from Note 4:
“As newer versions of analysers tested under this document become available, discontinue the
...