Measurement of fluid flow by means of pressure differential devices inserted in circular cross-section conduits running full - Part 1: General principles and requirements (ISO 5167-1:2022)

This document defines terms and symbols and establishes the general principles for methods of measurement and computation of the flow rate of fluid flowing in a conduit by means of pressure differential devices (orifice plates, nozzles, Venturi tubes, cone meters, and wedge meters) when they are inserted into a circular cross-section conduit running full. This document also specifies the general requirements for methods of measurement, installation and determination of the uncertainty of the measurement of flow rate.
ISO 5167 (all parts) is applicable only to flow that remains subsonic throughout the measuring section and where the fluid can be considered as single-phase. It is not applicable to the measurement of pulsating flow.

Durchflussmessung von Fluiden mit Drosselgeräten in voll durchströmten Leitungen mit Kreisquerschnitt - Teil 1: Allgemeine Grundlagen und Anforderungen (ISO 5167-1:2022)

Dieses Dokument legt Begriffe und Symbole fest und gibt die allgemeinen Grundsätze für Mess- und Berechnungsverfahren für den Durchfluss von Fluiden in voll durchströmten Leitungen mit Kreisquerschnitt mittels in die Leitung eingebauten Drosselgeräten (Blenden, Düsen, Venturirohre, Konus-Durchflussmesser und Keil-Durchflussmesser) an. Dieses Dokument legt auch die allgemeinen Anforderungen an Messverfahren, den Einbau und die Ermittlung der Messunsicherheit für die Durchflussmessung fest.
ISO 5167 (alle Teile) ist ausschließlich für Strömungen anwendbar, die in allen Messquerschnitten im Unter-schallbereich liegen und bei denen das Fluid als einphasig betrachtet werden kann. ISO 5167 (alle Teile) ist nicht anwendbar für die Messung von pulsierenden Strömungen.

Mesurage de débit des fluides au moyen d'appareils déprimogènes insérés dans des conduites en charge de section circulaire - Partie 1: Principes généraux et exigences générales (ISO 5167-1:2022)

Le présent document définit des termes et symboles et établit les principes généraux pour le mesurage et le calcul du débit des fluides dans une conduite au moyen d’appareils déprimogènes (diaphragmes, tuyères, tubes de Venturi, cônes de mesure et débitmètres à coin) insérés dans des conduites en charge de section circulaire. Le présent document spécifie aussi les exigences générales en ce qui concerne les méthodes de mesure, l’installation des appareils et la détermination de l’incertitude de la mesure de débit.
L’ISO 5167 (toutes les parties) est applicable uniquement à un écoulement qui reste subsonique dans tout le tronçon de mesurage et où le fluide peut être considéré comme monophasique. Elle n’est pas applicable au mesurage d’un écoulement pulsé.

Merjenje pretoka fluida na osnovi tlačne razlike, povzročene z napravo, vstavljeno v polno zapolnjen vod s krožnim prerezom – 1. del: Splošna načela in zahteve (ISO 5167-1:2022)

Ta dokument opredeljuje izraze in simbole ter določa splošna načela za metode merjenja in izračunavanja hitrosti pretoka fluida skozi vod na osnovi tlačne razlike, povzročene z napravo (zaslonke, šobe, Venturijeve cevi, stožčasti merilniki in klinasti merilniki),
vstavljeno v polno zapolnjen vod s krožnim prerezom. Ta dokument določa tudi splošne zahteve za
merilne metode, namestitev in določitev negotovosti merjenja pretoka.
Standard ISO 5167 (vsi deli) se uporablja samo za pretok, ki je v celotnem predelu merjenja podzvočen in pri katerem se lahko fluid obravnava kot enofazen. Ne uporablja se za merjenje utripajočega pretoka.

General Information

Status
Published
Public Enquiry End Date
30-Sep-2021
Publication Date
17-Jul-2022
Technical Committee
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
13-Jul-2022
Due Date
17-Sep-2022
Completion Date
18-Jul-2022

Relations

Standard
SIST EN ISO 5167-1:2022 - BARVE
English language
51 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day
Draft
oSIST prEN ISO 5167-1:2021 - BARVE
German language
49 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)


SLOVENSKI STANDARD
01-september-2022
Nadomešča:
SIST EN ISO 5167-1:2004
Merjenje pretoka fluida na osnovi tlačne razlike, povzročene z napravo, vstavljeno
v polno zapolnjen vod s krožnim prerezom – 1. del: Splošna načela in zahteve (ISO
5167-1:2022)
Measurement of fluid flow by means of pressure differential devices inserted in circular
cross-section conduits running full - Part 1: General principles and requirements (ISO
5167-1:2022)
Durchflussmessung von Fluiden mit Drosselgeräten in voll durchströmten Leitungen mit
Kreisquerschnitt - Teil 1: Allgemeine Grundlagen und Anforderungen (ISO 5167-1:2022)
Mesurage de débit des fluides au moyen d'appareils déprimogènes insérés dans des
conduites en charge de section circulaire - Partie 1: Principes généraux et exigences
générales (ISO 5167-1:2022)
Ta slovenski standard je istoveten z: EN ISO 5167-1:2022
ICS:
17.120.10 Pretok v zaprtih vodih Flow in closed conduits
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

EN ISO 5167-1
EUROPEAN STANDARD
NORME EUROPÉENNE
June 2022
EUROPÄISCHE NORM
ICS 17.120.10 Supersedes EN ISO 5167-1:2003
English Version
Measurement of fluid flow by means of pressure
differential devices inserted in circular cross-section
conduits running full - Part 1: General principles and
requirements (ISO 5167-1:2022)
Mesurage de débit des fluides au moyen d'appareils Durchflussmessung von Fluiden mit Drosselgeräten in
déprimogènes insérés dans des conduites en charge de voll durchströmten Leitungen mit Kreisquerschnitt -
section circulaire - Partie 1: Principes généraux et Teil 1: Allgemeine Grundlagen und Anforderungen (ISO
exigences générales (ISO 5167-1:2022) 5167-1:2022)
This European Standard was approved by CEN on 14 June 2022.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and
United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2022 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 5167-1:2022 E
worldwide for CEN national Members.

Contents Page
European foreword . 3

European foreword
This document (EN ISO 5167-1:2022) has been prepared by Technical Committee ISO/TC 30
"Measurement of fluid flow in closed conduits" in collaboration with Technical Committee CEN/SS F05
“Measuring Instruments” the secretariat of which is held by CCMC.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by December 2022, and conflicting national standards
shall be withdrawn at the latest by December 2022.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN ISO 5167-1:2003.
Any feedback and questions on this document should be directed to the users’ national standards
body/national committee. A complete listing of these bodies can be found on the CEN website.
According to the CEN-CENELEC Internal Regulations, the national standards organizations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,
Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland,
Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of
North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the
United Kingdom.
Endorsement notice
The text of ISO 5167-1:2022 has been approved by CEN as EN ISO 5167-1:2022 without any
modification.
INTERNATIONAL ISO
STANDARD 5167-1
Third edition
2022-06
Measurement of fluid flow by means of
pressure differential devices inserted
in circular cross-section conduits
running full —
Part 1:
General principles and requirements
Mesurage de débit des fluides au moyen d'appareils déprimogènes
insérés dans des conduites en charge de section circulaire —
Partie 1: Principes généraux et exigences générales
Reference number
ISO 5167-1:2022(E)
ISO 5167-1:2022(E)
© ISO 2022
All rights reserved. Unless otherwise specified, or required in the context of its implementation, no part of this publication may
be reproduced or utilized otherwise in any form or by any means, electronic or mechanical, including photocopying, or posting on
the internet or an intranet, without prior written permission. Permission can be requested from either ISO at the address below
or ISO’s member body in the country of the requester.
ISO copyright office
CP 401 • Ch. de Blandonnet 8
CH-1214 Vernier, Geneva
Phone: +41 22 749 01 11
Email: copyright@iso.org
Website: www.iso.org
Published in Switzerland
ii
ISO 5167-1:2022(E)
Contents Page
Foreword .v
Introduction . vi
1 Scope . 1
2 Normative references . 1
3 Terms and definitions . 1
3.1 Pressure measurement . 2
3.2 Primary devices . . 2
3.3 Flow . 3
4 Symbols and subscripts . 6
4.1 Symbols . 6
5 Principle of the method of measurement and computation . 7
5.1 Principle of the method of measurement . 7
5.2 Method of determination of the required diameter ratio for the selected standard
primary device . 8
5.3 Computation of flow rate . . 8
5.4 Determination of density, pressure and temperature . 8
5.4.1 General . 8
5.4.2 Density . 9
5.4.3 Static pressure . 9
5.4.4 Temperature . 9
5.5 Differential pressure flow measurement system . 10
5.5.1 General . 10
5.5.2 Primary device . 11
5.5.3 Impulse lines and transmitters .12
5.5.4 Impulse line isolation valves and valve manifolds .12
5.5.5 Flow computer . .12
5.6 Differential pressure flow measurement system design considerations .12
5.6.1 Flow rate turndown and stacked transmitters .12
5.6.2 Meter calibration .12
5.6.3 Permanent pressure loss . 13
5.6.4 Diagnostics and meter verification . 14
5.6.5 Overall uncertainty of differential pressure metering system . 14
6 General requirements for the measurements .14
6.1 Primary device . 14
6.2 Nature of the fluid. 15
6.3 Flow conditions . 15
7 Installation requirements .15
7.1 General . 15
7.2 Minimum upstream and downstream straight lengths . 17
7.3 General requirement for flow conditions at the primary device . 17
7.3.1 Requirement . 17
7.3.2 Swirl-free conditions . 17
7.3.3 Good velocity profile conditions . 17
7.4 Flow conditioners . 17
7.4.1 Compliance testing . 17
7.4.2 Specific test .20
8 Uncertainties on the measurement of flow rate .20
8.1 General . 20
8.2 Definition of uncertainty .20
8.3 Practical computation of the uncertainty . 21
8.3.1 Component uncertainties . 21
iii
ISO 5167-1:2022(E)
8.3.2 Practical working formula . 21
Annex A (informative) Iterative computations .24
Annex B (informative) Examples of values of the pipe wall uniform equivalent roughness, k .26
a
Annex C (informative) Flow conditioners and flow straighteners .27
Annex D (informative) Differential pressure transmitters, flow range and turndown .29
Annex E (informative) Example of uncertainty calculation for a differential pressure device .36
Annex F (informative) Permanent pressure loss example .40
Bibliography .42
iv
ISO 5167-1:2022(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards
bodies (ISO member bodies). The work of preparing International Standards is normally carried out
through ISO technical committees. Each member body interested in a subject for which a technical
committee has been established has the right to be represented on that committee. International
organizations, governmental and non-governmental, in liaison with ISO, also take part in the work.
ISO collaborates closely with the International Electrotechnical Commission (IEC) on all matters of
electrotechnical standardization.
The procedures used to develop this document and those intended for its further maintenance are
described in the ISO/IEC Directives, Part 1. In particular, the different approval criteria needed for the
different types of ISO documents should be noted. This document was drafted in accordance with the
editorial rules of the ISO/IEC Directives, Part 2 (see www.iso.org/directives).
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. ISO shall not be held responsible for identifying any or all such patent rights. Details of
any patent rights identified during the development of the document will be in the Introduction and/or
on the ISO list of patent declarations received (see www.iso.org/patents).
Any trade name used in this document is information given for the convenience of users and does not
constitute an endorsement.
For an explanation of the voluntary nature of standards, the meaning of ISO specific terms and
expressions related to conformity assessment, as well as information about ISO's adherence to
the World Trade Organization (WTO) principles in the Technical Barriers to Trade (TBT), see
www.iso.org/iso/foreword.html.
ISO 5167-1 was prepared by Technical Committee ISO/TC 30, Measurement of fluid flow in closed conduits,
Subcommittee SC 2, Pressure differential devices, in collaboration with the European Committee for
Standardization (CEN) Technical Committee CEN/SS F05, Measuring instruments, in accordance with
the Agreement on technical cooperation between ISO and CEN (Vienna Agreement).
This third edition cancels and replaces the second edition (ISO 5167-1:2003), which has been technically
revised
The main changes are as follows:
— improved consistency between ISO 5167-1 to ISO 5167-6 (some items that were new in ISO 5167-5
and ISO 5167-6 have been moved to this document);
— a primary element has been set as part of a differential pressure metering system;
— a short section on diagnostics and CBM (Condition Based Monitoring) has been included;
— a limitation on the use of the 5 % 2° rule for an acceptable profile has been noted;
— improved text about uncertainty calculation and an example in Annex E has been provided;
— annexes on turndown and permanent pressure loss have been included.
A list of all parts in the ISO 5167 series can be found on the ISO website.
Any feedback or questions on this document should be directed to the user’s national standards body. A
complete listing of these bodies can be found at www.iso.org/members.html.
v
ISO 5167-1:2022(E)
Introduction
ISO 5167, consisting of six parts, covers the geometry and method of use (installation and operating
conditions) of orifice plates, nozzles, Venturi tubes, cone meters and wedge meters when they are
inserted in a conduit running full to determine the flow rate of the fluid flowing in the conduit. It also
gives necessary information for calculating the flow rate and its associated uncertainty.
ISO 5167 (all parts) is applicable only to pressure differential devices in which the flow remains
subsonic throughout the measuring section and where the fluid can be considered as single-phase, but
is not applicable to the measurement of pulsating flow. Furthermore, each of these devices can only be
used uncalibrated within specified limits of pipe size and Reynolds number, or alternatively they can be
used across their calibrated range.
ISO 5167 (all parts) deals with devices for which direct calibration experiments have been made,
sufficient in number, spread and quality to enable coherent systems of application to be based on
their results and coefficients to be given with certain predictable limits of uncertainty. ISO 5167 also
provides methodology for bespoke calibration of differential pressure meters.
The devices introduced into the pipe are called primary devices. The term primary device also includes
the pressure tappings. All other instruments or devices required to facilitate the instrument readings
are known as secondary devices, and the flow computer that receives these readings and performs
the algorithms is known as a tertiary device. ISO 5167 covers primary devices; secondary devices (see
ISO 2186) and tertiary devices will be mentioned only occasionally.
Aspects of safety are not dealt with in ISO 5167-1 to ISO 5167-6. It is the responsibility of the user to
ensure that the system meets applicable safety regulations.
Additional documents that may provide assistance include:
— ISO/TR 3313;
— ISO/TR 9464;
— ISO/TR 12767;
— ISO/TR 15377.
vi
INTERNATIONAL STANDARD ISO 5167-1:2022(E)
Measurement of fluid flow by means of pressure
differential devices inserted in circular cross-section
conduits running full —
Part 1:
General principles and requirements
1 Scope
This document defines terms and symbols and establishes the general principles for methods of
measurement and computation of the flow rate of fluid flowing in a conduit by means of pressure
differential devices (orifice plates, nozzles, Venturi tubes, cone meters, and wedge meters) when they
are inserted into a circular cross-section conduit running full. This document also specifies the general
requirements for methods of measurement, installation and determination of the uncertainty of the
measurement of flow rate.
ISO 5167 (all parts) is applicable only to flow that remains subsonic throughout the measuring section
and where the fluid can be considered as single-phase. It is not applicable to the measurement of
pulsating flow.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
ISO 4006, Measurement of fluid flow in closed conduits — Vocabulary and symbols
ISO 5167 (all parts), Measurement of fluid flow by means of pressure differential devices inserted in circular
cross-section conduits running full
ISO 5168, Measurement of fluid flow — Procedures for the evaluation of uncertainties
ISO/IEC Guide 98-3, Uncertainty of measurement — Part 3: Guide to the expression of uncertainty in
measurement (GUM: 1995)
3 Terms and definitions
For the purposes of this document, the terms, definitions and symbols given in ISO 4006 and the
following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— ISO Online browsing platform: available at https:// www .iso .org/ obp
— IEC Electropedia: available at https:// www .electropedia .org/
ISO 5167-1:2022(E)
3.1 Pressure measurement
3.1.1
wall pressure tapping
annular slot or circular hole drilled in the wall of a conduit in such a way that the edge of the hole is
flush with the internal surface of the conduit
Note 1 to entry: The pressure tapping is usually a circular hole but in certain cases may be an annular slot.
3.1.2
static pressure
p
pressure which can be measured by connecting a pressure-measuring device to a wall pressure tapping
(3.1.1)
Note 1 to entry: Only the value of the absolute static pressure is considered in ISO 5167 (all parts).
3.1.3
differential pressure
DP
Δp
difference between the (static) pressures measured at the wall pressure tappings, one of which is on
the upstream side and the other of which is on the downstream side of a primary device [or in the
throat for a throat-tapped nozzle, a Venturi nozzle (3.2.4) or a Venturi tube (3.2.5)], inserted in a straight
pipe through which flow occurs, when any difference in height between the upstream and downstream
tappings has been taken into account
Note 1 to entry: In ISO 5167 (all parts) the term “differential pressure” is used only if the pressure tappings are in
the positions specified for each standard primary device.
3.1.4
pressure ratio
τ
ratio of the absolute (static) pressure at the downstream pressure tapping to the absolute (static)
pressure at the upstream pressure tapping
3.1.5
vena contracta
location in a fluid stream where the diameter of the stream is smallest
3.2 Primary devices
3.2.1
orifice
throat opening of minimum cross-sectional area of a primary device
3.2.2
orifice plate
thin plate in which a circular opening has been machined
Note 1 to entry: Standard orifice plates are described as “thin plate” and “with sharp square edge”, because the
thickness of the plate is small compared with the diameter of the measuring section and because the upstream
edge of the orifice (3.2.1) is sharp and square.
3.2.3
nozzle
device which consists of a convergent inlet connected to a cylindrical section generally called the
“throat”
ISO 5167-1:2022(E)
3.2.4
Venturi nozzle
device which consists of a convergent inlet which is a standardized ISA 1932 nozzle connected to a
cylindrical part called the “throat”, which is itself connected to an expanding section called the
“divergent” which is conical
3.2.5
Venturi tube
device which consists of a convergent inlet which is conical connected to a cylindrical part called the
“throat”, which is itself connected to an expanding section called the “divergent” which is conical
3.2.6
cone meter
device which consists of a cone-shaped restriction held in the centre of the pipe with the nose of the
cone upstream
3.2.7
wedge meter
device which consists of a wedge-shaped restriction
3.2.8
diameter ratio
β
square root of the ratio of the area of the throat of the
primary device to the internal area of the measuring pipe upstream of the primary device
Note 1 to entry: In ISO 5167-2 and ISO 5167-3 the diameter ratio is the ratio of the diameter of the throat of the
primary device to the internal diameter of the measuring pipe upstream of the primary device.
Note 2 to entry: In ISO 5167-4, where the primary device has a cylindrical section upstream, having the same
diameter as that of the pipe, the diameter ratio is the ratio of the throat diameter to the diameter of this cylindrical
section at the plane of the upstream pressure tappings.
3.2.9
carrier ring
device which is used to hold the primary element in the centre of the pipe and may incorporate the
pressure tappings
3.3 Flow
3.3.1
flow rate
rate of flow
q
mass or volume of fluid passing through the primary device per unit time
3.3.1.1
mass flow rate
rate of mass flow
q
m
mass of fluid passing through the primary device per unit time
3.3.1.2
volume flow rate
rate of volume flow
q
V
volume of fluid passing through the primary device per unit time
Note 1 to entry: In the case of volume flow rate, it is necessary to state the pressure and temperature at which the
volume is referenced.
ISO 5167-1:2022(E)
3.3.2
Reynolds number
Re
dimensionless parameter expressing the ratio between the inertia and viscous forces
3.3.2.1
pipe Reynolds number
Re
D
dimensionless parameter expressing the ratio between the inertia and viscous forces in the upstream
pipe
VD 4q
1 m
Re ==
D
νμπ D
3.3.2.2
throat Reynolds number
Re
d
dimensionless parameter expressing the ratio between the inertia and viscous forces in the orifice or
throat of the primary device
Re
D
Re =
d
β
Note 1 to entry: When an orifice plate is used the throat Reynolds number is sometimes called the orifice
Reynolds number.
3.3.3
isentropic exponent
κ
ratio of the relative variation in pressure to the corresponding relative variation in density under
elementary reversible adiabatic (isentropic) transformation conditions
Note 1 to entry: The isentropic exponent κ appears in the different formulae for the expansibility [expansion]
factor ε and varies with the nature of the gas and with its temperature and pressure.
Note 2 to entry: There are many gases and vapours for which no values for κ have been published so far,
particularly over a wide range of pressure and temperature. In such a case, for the purposes of ISO 5167 (all parts),
the ratio of the specific heat capacity at constant pressure to the specific heat capacity at constant volume of
ideal gases can be used in place of the isentropic exponent.
3.3.4
Joule Thomson coefficient
isenthalpic temperature-pressure coefficient
μ
JT
rate of change of temperature with respect to pressure at constant enthalpy:
∂T
μ =
JT
∂p
H
Note 1 to entry: The Joule Thomson coefficient varies with the nature of the gas and with its temperature and
pressure and can be calculated.
Note 2 to entry: An approximation for the Joule Thomson coefficient for some natural gases is given in
ISO/TR 9464:2008, 5.1.5.4.4.
ISO 5167-1:2022(E)
3.3.5
discharge coefficient
C
coefficient, defined for an incompressible fluid flow, which relates the actual flow rate to the theoretical
flow rate through a device, and is given by the formula for incompressible fluids
q 1−β
m
C =
Ap2Δ ρ
t 1
Note 1 to entry: Calibration of standard primary devices by means of incompressible fluids (liquids) shows
that the discharge coefficient is dependent only on the Reynolds number for a given primary device in a given
installation.
The numerical value of C for any individual differential pressure meter is the same for different installations
whenever such installations are geometrically similar and the flows are characterised by identical Reynolds
numbers.
The formulae for the numerical values of C given in ISO 5167 (all parts) are based on data determined
experimentally.
The uncertainty in the value of C can be reduced by flow calibration in a suitable laboratory.
Note 2 to entry: The quantity 11/ −β is called the “velocity of approach factor”, and C is called the
1−β
“flow coefficient”.
3.3.6
expansibility [expansion] factor
ε
coefficient used to take into account the compressibility of the fluid
q 1−β
m
ε =
AC 2Δpρ
t 1
Note 1 to entry: Calibration of a given primary device by means of a compressible fluid (gas) shows that the
following ratio is dependent on the value of the Reynolds number as well as on the values of the pressure ratio
and the isentropic exponent of the gas:
q 1−β
m
Ap2Δ ρ
t 1
The method adopted for representing these variations consists of multiplying the discharge coefficient, C, of the
primary device considered, as determined by direct calibration carried out with liquids for the same value of the
Reynolds number, by the expansibility [expansion] factor, ε.
The expansibility factor, ε, is equal to unity when the fluid is considered incompressible (liquid) and is less than
unity when the fluid is compressible (gaseous).
This method is possible because experiments show that ε is practically independent of the Reynolds number
and, for a given diameter ratio of a given primary device, ε only depends on the pressure ratio and the isentropic
exponent.
The numerical values of ε for orifice plates given in ISO 5167-2 and for cone meters given in ISO 5167-5 are based
on data determined experimentally. For nozzles (see ISO 5167-3), Venturi tubes (see ISO 5167-4) and wedge
meters (see ISO 5167-6) they are based on the thermodynamic general formula applied to isentropic expansion.
ISO 5167-1:2022(E)
3.3.7
arithmetical mean deviation of the roughness profile
Ra
arithmetical mean deviation from the mean line of the profile being measured
Note 1 to entry: The mean line is such that the sum of the squares of the distances between the effective surface
and the mean line is a minimum. In practice, Ra can be measured with standard equipment for machined surfaces
but can only be estimated for rougher surfaces of pipes. See also ISO 21920-3.
Note 2 to entry: For pipes, the uniform equivalent roughness k may also be used. This value can be determined
a
experimentally (see 7.1.5) or taken from tables (see Annex B).
4 Symbols and subscripts
4.1 Symbols
Table 1 — Symbols
a
Symbol Quantity Dimension SI unit
2 2
A Area of throat L m
t
C Discharge coefficient dimensionless —
2 −2 −1 −1
C Molar-heat capacity at constant pressure ML T Θ mol J/(mol⋅K)
m,p
Diameter of orifice (or throat) of primary device under
d L m
working conditions
Upstream internal pipe diameter (or upstream diameter of a
D L m
classical Venturi tube) under working conditions
2 −2 −1
H Enthalpy ML T mol J/mol
k Coverage factor dimensionless —
k Uniform equivalent roughness L m
a
Pressure loss coefficient (the ratio of the pressure loss, Δϖ,
K to the dynamic pressure, ρV /2), also known as the minor dimensionless —
loss coefficient
l Pressure tapping spacing L m
L Relative pressure tapping spacing: L = l/D dimensionless —
−1 −2
p Absolute static pressure of the fluid ML T Pa
−1
q Mass flow rate MT kg/s
m
3 −1 3
q Volume flow rate L T m /s
V
R Radius L m
2 −2 −1 −1
R Universal gas constant ML T Θ mol J/(mol⋅K)
u
Ra Arithmetical mean deviation of the (roughness) profile L m
Re Reynolds number dimensionless —
Re Throat Reynolds number dimensionless —
d
Re Pipe Reynolds number dimensionless —
D
t Temperature of the fluid Θ °C
T Absolute (thermodynamic) temperature of the fluid Θ K
c c
u Standard uncertainty
a
M = mass, L = length, T = time, Θ = temperature
b
γ is the ratio of the specific heat capacity at constant pressure to the specific heat capacity at constant volume.
For ideal gases, the ratio of the specific heat capacities and the isentropic exponent have the same value (see 3.3.3). These
values depend on the nature of the gas.
The dimensions and units are those of the corresponding quantity.
c
ISO 5167-1:2022(E)
Table 1 (continued)
a
Symbol Quantity Dimension SI unit
′ Relative standard uncertainty dimensionless —
u
c c
U Expanded uncertainty
′ Relative expanded uncertainty dimensionless —
U
−1
V Mean axial velocity of the fluid in the pipe LT m/s
Z Compressibility factor dimensionless —
β Diameter ratio dimensionless —
b
γ Ratio of specific heat capacities dimensionless —
−1 −2
Δp Differential pressure: Δp = p − p ML T Pa
1 2
−1 −2
Δp Pressure loss across a flow conditioner ML T Pa
c
−1 −2
Δϖ Pressure loss across a primary device ML T Pa
ε Expansibility [expansion] factor dimensionless —
b
κ Isentropic exponent dimensionless —
λ Friction factor dimensionless —
−1 −1
μ Dynamic viscosity of the fluid ML T Pa⋅s
−1 2
μ Joule Thomson coefficient M LT Θ K/Pa
JT
2 −1 2
v Kinematic viscosity of the fluid: v = μ/ρ L T m /s
Relative pressure loss (the ratio of the pressure loss to the
ξ dimensionless —
differential pressure)
−3 3
ρ Density of the fluid ML kg/m
τ Pressure ratio: τ = p /p dimensionless —
2 1
ϕ Total angle of the divergent section dimensionless rad
a
M = mass, L = length, T = time, Θ = temperature
b
γ is the ratio of the specific heat capacity at constant pressure to the specific heat capacity at constant volume.
For ideal gases, the ratio of the specific heat capacities and the isentropic exponent have the same value (see 3.3.3). These
values depend on the nature of the gas.
The dimensions and units are those of the corresponding quantity.
c
5 Principle of the method of measurement and computation
5.1 Principle of the method of measurement
The principle of the method of measurement is based on the installation of a primary device into a
pipeline in which a fluid is running full. The installation of the primary device causes a static pressure
difference between the upstream side and the throat or downstream side of the device. The flow rate
can be determined from the measured value of this pressure difference and the knowledge of the
thermodynamic conditions, fluid properties, meter geometry and meter characteristics. It is assumed
that an uncalibrated differential pressure meter is within the geometric and Reynolds number range
required for the ISO discharge coefficient prediction to be valid. Alternatively, it is assumed that a
bespoke calibrated differential pressure meter is to be used within its calibration range.
The mass flow rate can be determined, since it is related to the differential pressure within the
uncertainty limits stated in ISO 5167, using Formula (1):
C
q = ερAp2Δ (1)
mt 1
1−β
NOTE For practical implementation, this formula is expanded upon as Formula (1) of ISO 5167-2, ISO 5167-3,
ISO 5167-4, ISO 5167-5 and ISO 5167-6.
ISO 5167-1:2022(E)
Similarly, the value of the volume flow rate can be calculated using Formula (2):
q
m
q = (2)
V
ρ
where ρ is the fluid density at the temperature and pressure for which the volume is stated.
5.2 Method of determination of the required diameter ratio for the selected standard
primary device
In practice, when determining the diameter ratio of a primary element to be installed in a given pipeline,
C and ε used in Formula (1) are, in general, not precisely known. Hence the following shall be selected a
priori:
— the type of primary device to be used;
— a flow rate and the corresponding desired value of the differential pressure.
The related values of q and Δp are then inserted in Formula (1), rewritten in the form of Formula (3):
m
4q
Cεβ
m
= (3)
πΔDp2 ρ
1−β
in which the diameter ratio of the selected primary device can be determined by iteration (see Annex A).
For a given flow rate, the uncertainty of the discharge coefficient and that of the predicted differential
pressure are directly linked, because the discharge coefficient is proportional to the reciprocal of the
square root of the differential pressure. Consequently, care shall be taken when determining β that
the maximum differential pressure does not exceed the upper range limit of the transmitter. This is of
particular importance where the uncertainty of the discharge coefficient is large.
5.3 Computation of flow rate
Computation of the flow rate, which is a purely arithmetic process, is performed by replacing the
different terms on the right-hand side of Formula (1) by their numerical values.
C may be dependent on Re, which is itself dependent on q . In such cases the final value of C, and hence
m
of q , is obtained by iteration. See Annex A for guidance regarding the choice of the iteration procedure
m
and initial estimates.
The dimensions used in the formulae are the values of the dimensions at the working conditions.
Measurements taken at any other conditions should be corrected for any possible expansion or
contraction of the primary device and the pipe due to the values of the temperature and pressure of the
fluid during the measurement.
NOTE For corrections due to thermal expansion or contraction see ISO/TR 9464:2008 5.1.6.1.3 and 5.2.6.4.2.
It is necessary to know the density and the viscosity of the fluid at working conditions. In the case
of a compressible fluid, it is also necessary to know the isentropic exponent of the fluid at working
conditions.
5.4 Determination of density, pressure and temperature
5.4.1 General
Any method of determining reliable values of the density, static pressure and temperature of the fluid is
acceptable if it does not interfere with the distribution of the flow in any way at the cross-section where
measurement is made.
ISO 5167-1:2022(E)
5.4.2 Density
It is necessary to know the density of the fluid at the upstream pressure tapping; it can either be
measured directly or be calculated from an appropriate equation of state from a knowledge of the
absolute static pressure, absolute temperature and composition of the fluid at that location.
NOTE ISO/TR 9464:2008, 6.4.2 provides a method for correcting density measured downstream of a device
to upstream conditions.
5.4.3 Static pressure
The static pressure of the fluid shall be measured by means of an individual wall pressure tapping, or
several such tappings interconnected, or by means of carrier ring tappings if they are permitted for the
measurement of differential pressure in that tapping plane for the particular primary device.

a
Flow.
b
Section A-A (upstream) also typical for section B-B (downstream).
Figure 1 — “Triple-T” arrangement
Where four pressure tappings are connected together to give the pressure upstream, downstream or
in the throat of the primary device, it is best that they should be connected together in a “triple-T”
arrangement as shown in Figure 1. The “triple-T” arrangement is often used for measurement with
large Venturi tubes.
It is permissible to link simultaneously one pressure tapping with differential pressure measuring
device(s) and static pressure measuring device(s), provided that these connections do not lead to any
distortion of the differential pressure measurement.
5.4.4 Temperature
5.4.4.1 Temperature measurement requires particular care. The thermometer well or pocket shall
take up as little space as possible to avoid reducing the cross-sectional area of the pipe. Thermometer
probes should have adequate immersion depth to ensure the fluid temperature is measured accurately.
ISO 5167-1:2022(E)
Except where a cone meter is used, the temperature of the fluid shall preferably be measured
downstream of the primary device to avoid disturbance to the flow profile affecting the primary device.
If the thermometer well or pocket is located downstream of the primary device, the distance between it
and the primary device shall be at least equal to 5D (and at most 15D when the fluid is a gas). In the case
of a Venturi tube this distance is measured from the throat
...


SLOVENSKI STANDARD
oSIST prEN ISO 5167-1:2021
01-september-2021
Merjenje pretoka fluida na osnovi tlačne razlike, povzročene z napravo, vstavljeno
v polno zapolnjen vod s krožnim prerezom – 1. del: Splošna načela in zahteve
(ISO/DIS 5167-1:2021)
Measurement of fluid flow by means of pressure differential devices inserted in circular
cross-section conduits running full - Part 1: General principles and requirements
(ISO/DIS 5167-1:2021)
Durchflussmessung von Fluiden mit Drosselgeräten in voll durchströmten Leitungen mit
Kreisquerschnitt - Teil 1: Allgemeine Grundlagen und Anforderungen (ISO/DIS 5167-
1:2021)
Mesure de débit des fluides au moyen d'appareils déprimogènes insérés dans des
conduites en charge de section circulaire - Partie 1: Principes généraux et exigences
générales (ISO/DIS 5167-1:2021)
Ta slovenski standard je istoveten z: prEN ISO 5167-1
ICS:
17.120.10 Pretok v zaprtih vodih Flow in closed conduits
oSIST prEN ISO 5167-1:2021 de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

oSIST prEN ISO 5167-1:2021
oSIST prEN ISO 5167-1:2021
ENTWURF
EUROPÄISCHE NORM
prEN ISO 5167-1
EUROPEAN STANDARD
NORME EUROPÉENNE
Juli 2021
ICS 17.120.10 Vorgesehen als Ersatz für EN ISO 5167-1:2003
Deutsche Fassung
Durchflussmessung von Fluiden mit Drosselgeräten in voll
durchströmten Leitungen mit Kreisquerschnitt - Teil 1:
Allgemeine Grundlagen und Anforderungen (ISO/DIS 5167-
1:2021)
Measurement of fluid flow by means of pressure Mesure de débit des fluides au moyen d'appareils
differential devices inserted in circular cross-section déprimogènes insérés dans des conduites en charge de
conduits running full - Part 1: General principles and section circulaire - Partie 1: Principes généraux et
requirements (ISO/DIS 5167-1:2021) exigences générales (ISO/DIS 5167-1:2021)
Dieser Europäische Norm-Entwurf wird den CEN-Mitgliedern zur parallelen Umfrage vorgelegt. Er wurde vom Technischen
Komitee CEN/SS F05 erstellt.
Wenn aus diesem Norm-Entwurf eine Europäische Norm wird, sind die CEN-Mitglieder gehalten, die CEN-Geschäftsordnung zu
erfüllen, in der die Bedingungen festgelegt sind, unter denen dieser Europäischen Norm ohne jede Änderung der Status einer
nationalen Norm zu geben ist.
Dieser Europäische Norm-Entwurf wurde von CEN in drei offiziellen Fassungen (Deutsch, Englisch, Französisch) erstellt. Eine
Fassung in einer anderen Sprache, die von einem CEN-Mitglied in eigener Verantwortung durch Übersetzung in seine
Landessprache gemacht und dem CEN-CENELEC-Management-Zentrum mitgeteilt worden ist, hat den gleichen Status wie die
offiziellen Fassungen.
CEN-Mitglieder sind die nationalen Normungsinstitute von Belgien, Bulgarien, Dänemark, Deutschland, Estland, Finnland,
Frankreich, Griechenland, Irland, Island, Italien, Kroatien, Lettland, Litauen, Luxemburg, Malta, den Niederlanden, Norwegen,
Österreich, Polen, Portugal, der Republik Nordmazedonien, Rumänien, Schweden, der Schweiz, Serbien, der Slowakei, Slowenien,
Spanien, der Tschechischen Republik, der Türkei, Ungarn, dem Vereinigten Königreich und Zypern.

Die Empfänger dieses Norm-Entwurfs werden gebeten, mit ihren Kommentaren jegliche relevante Patentrechte, die sie kennen,
mitzuteilen und unterstützende Dokumentationen zur Verfügung zu stellen.

Warnvermerk : Dieses Schriftstück hat noch nicht den Status einer Europäischen Norm. Es wird zur Prüfung und Stellungnahme
vorgelegt. Es kann sich noch ohne Ankündigung ändern und darf nicht als Europäischen Norm in Bezug genommen werden.

EUROPÄISCHES KOMITEE FÜR NORMUNG
EUROPEAN COMMITTEE FOR STANDARDIZATION

COMITÉ EUROPÉEN DE NORMALISATION

CEN-CENELEC Management-Zentrum: Rue de la Science 23, B-1040 Brüssel
© 2021 CEN Alle Rechte der Verwertung, gleich in welcher Form und in welchem Ref. Nr. prEN ISO 5167-1:2021 D
Verfahren, sind weltweit den nationalen Mitgliedern von CEN
vorbehalten.
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
Inhalt
Seite
Europäisches Vorwort . 4
Vorwort . 5
Einleitung . 6
1 Anwendungsbereich . 8
2 Normative Verweisungen . 8
3 Begriffe . 8
4 Symbole und Indizes . 13
4.1 Symbole . 13
4.2 Indizes . 15
5 Grundlage des Mess- und Berechnungsverfahrens . 15
5.1 Grundlage des Messverfahrens . 15
5.2 Verfahren für die Bestimmung des erforderlichen Durchmesserverhältnisses für das
ausgewählte genormte Primärgerät. 16
5.3 Berechnung des Durchflusses . 16
5.4 Bestimmung von Dichte, Druck und Temperatur . 16
5.4.1 Allgemeines . 16
5.4.2 Dichte . 17
5.4.3 Statischer Druck . 17
5.4.4 Temperatur . 18
5.5 Differenzdruck-Durchflussmesssystem . 18
5.5.1 Allgemeines . 18
5.5.2 Primärgerät . 20
5.5.3 Impulsleitungen und Messaufnehmer . 20
5.5.4 Absperrventile und Ventilblöcke der Impulsleitungen . 20
5.5.5 Mengenumwerter . 20
5.6 Betrachtungen bezüglich der Konstruktion des Differenzdruck-Durchflussmesssystems . 20
5.6.1 Messspanne des Durchflusses und gestapelte Messumformer . 20
5.6.2 Kalibrierung des Durchflussmessers . 21
5.6.3 Bleibender Druckverlust . 21
5.6.4 Diagnostik und Verifizierung des Durchflussmessers . 22
5.6.5 Gesamtunsicherheit eines Differenzdruck-Messsystems . 23
6 Allgemeine Anforderungen an die Messungen . 23
6.1 Primärgerät . 23
6.2 Art des Fluids . 23
6.3 Strömungsbedingungen . 23
7 Anforderungen an den Einbau . 24
7.1 Allgemeines . 24
7.2 Mindestwerte für gerade Rohrlängen im Ein- und Auslauf. 25
7.3 Allgemeine Anforderung an die Strömungsbedingungen am Primärgerät . 26
7.3.1 Anforderung . 26
7.3.2 Drallfreie Bedingungen . 26
7.3.3 Gute Strömungsbedingungen . 26
7.4 Strömungsumformer (siehe auch Anhang C) . 26
7.4.1 Übereinstimmungsprüfung . 26
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
7.4.2 Besondere Prüfung . 28
8 Unsicherheiten bei der Durchflussmessung . 28
8.1 Festlegung der Unsicherheit . 29
8.2 Praktische Berechnung der Unsicherheit . 29
8.2.1 Messunsicherheitskomponenten . 29
8.2.2 Praktische Arbeitsgleichung . 30
Anhang A (informativ) Iterative Berechnungen . 32
Anhang B (informativ) Beispiele für Werte der gleichmäßigen äquivalenten Rohrrauheit k der
Rohrwand . 35
Anhang C (informativ) Strömungsumformer und Strömungsgleichrichter . 36
C.1 Allgemeines . 36
C.2 Strömungsgleichrichter . 36
C.3 Strömungsumformer. 36
C.4 Übereinstimmungsprüfungen . 37
Anhang D (informativ) Differenzdruck-Messumformer, Durchflussbereich und Messspanne . 38
D.1 Differenzdruck-Messumformer . 38
D.2 Messspanne . 39
D.3 Einflussfaktoren für den Durchflussbereich von Differenzdruck-Messgeräten . 39
D.3.1 Allgemeines . 39
D.3.2 Art des Messgeräts . 39
D.3.3 Durchmesserverhältnis . 39
D.3.4 Strukturelle Unversehrtheit . 40
D.3.5 Bleibender Druckverlust . 40
D.3.6 Strömungsbedingungen . 40
D.3.7 Wirkdruckbereich . 40
D.3.8 Einstellbare Durchflussbereiche . 41
D.4 Vergleich von Durchflussmesser-Messspannen . 42
Anhang E (informativ) Beispiel für Unsicherheit . 44
E.1 Aufbau des Durchflussmessers . 44
E.2 Unsicherheit und Empfindlichkeit von Parametern . 44
E.2.1 Allgemeines . 44
E.2.2 Durchflusskoeffizient C . 44
E.2.3 Expansionszahl ε . 45
E.2.4 Wirkdruck Δp . 45
E.2.5 Blendendurchmesser d . 45
E.2.6 Durchmesser des zulaufseitigen Rohres D . 45
E.2.7 Massendichte ρ . 45
E.2.8 Kombinieren der Unsicherheiten . 46
Anhang F (informativ) Beispiel für bleibenden Druckverlust . 47
Literaturhinweise. 49

oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
Europäisches Vorwort
Dieses Dokument (prEN ISO 5167-1:2021) wurde vom Technischen Komitee ISO/TC 30 „Measurement of
fluid flow in closed conduits“ in Zusammenarbeit mit dem Technischen Komitee CEN/SS F05 „Messinstru-
mente“ erarbeitet, dessen Sekretariat von CCMC gehalten wird.
Dieses Dokument ist derzeit zur parallelen Umfrage vorgelegt.
Dieses Dokument wird EN ISO 5167-1:2003 ersetzen.
Anerkennungsnotiz
Der Text von ISO/DIS 5167-1:2021 wurde von CEN als prEN ISO 5167-1:2021 ohne irgendeine Abänderung
genehmigt.
Rückmeldungen oder Fragen zu diesem Dokument sollten an das jeweilige nationale Normungsinstitut des
Anwenders gerichtet werden. Eine vollständige Liste dieser Institute ist auf den Internetseiten von CEN
abrufbar.
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
Vorwort
ISO (die Internationale Organisation für Normung) ist eine weltweite Vereinigung nationaler Normungs-
institute (ISO-Mitgliedsorganisationen). Die Erstellung von Internationalen Normen wird üblicherweise von
Technischen Komitees von ISO durchgeführt. Jede Mitgliedsorganisation, die Interesse an einem Thema hat,
für welches ein Technisches Komitee gegründet wurde, hat das Recht, in diesem Komitee vertreten zu sein.
Internationale staatliche und nichtstaatliche Organisationen, die in engem Kontakt mit ISO stehen, nehmen
ebenfalls an der Arbeit teil. ISO arbeitet bei allen elektrotechnischen Normungsthemen eng mit der
Internationalen Elektrotechnischen Kommission (IEC) zusammen.
Die Verfahren, die bei der Entwicklung dieses Dokuments angewendet wurden und die für die weitere Pflege
vorgesehen sind, werden in den ISO/IEC-Direktiven, Teil 1 beschrieben. Es sollten insbesondere die
unterschiedlichen Annahmekriterien für die verschiedenen ISO-Dokumentenarten beachtet werden. Dieses
Dokument wurde in Übereinstimmung mit den Gestaltungsregeln der ISO/IEC-Direktiven, Teil 2 erarbeitet
(siehe www.iso.org/directives).
Es wird auf die Möglichkeit hingewiesen, dass einige Elemente dieses Dokuments Patentrechte berühren
können. ISO ist nicht dafür verantwortlich, einige oder alle diesbezüglichen Patentrechte zu identifizieren.
Details zu allen während der Entwicklung des Dokuments identifizierten Patentrechten finden sich in der
Einleitung und/oder in der ISO-Liste der erhaltenen Patenterklärungen (siehe www.iso.org/patents).
Jeder in diesem Dokument verwendete Handelsname dient nur zur Unterrichtung der Anwender und bedeutet
keine Anerkennung.
Eine Erläuterung der Bedeutung ISO-spezifischer Benennungen und Ausdrücke, die sich auf Konformitäts-
bewertung beziehen, sowie Informationen über die Beachtung der Grundsätze der Welthandelsorganisation
(WTO) zu technischen Handelshemmnissen (TBT, en: Technical Barriers to Trade) durch ISO enthält der
folgende Link: www.iso.org/iso/foreword.html.
ISO 5167-1 wurde vom Technischen Komitee ISO/TC 30, Measurement of fluid flow in closed conduits,
Unterkomitee SC 2, Pressure differential devices erarbeitet.
Diese dritte Fassung von ISO 5167-1 ersetzt die zweite Fassung (ISO 5167-1:2003), die technisch überarbeitet
wurde.
Die wesentlichen Änderungen im Vergleich zur Vorgängerausgabe sind folgende:
Die Übereinstimmung zwischen Teil 1 bis Teil 6 wurde verbessert (einige Themen, die in Teil 5 und Teil 6 neu
waren, wurden in Teil 1 verschoben).
Ein Primärelement wird als Teil eines Differenzdruck-Durchflussmesssystems eingeführt.
Ein kurzer Absatz zu Diagnostik und CBM (zustandsorientierte Überwachung, en: Condition Based Moni-
toring) ist enthalten.
Eine Begrenzung hinsichtlich der Anwendung der 5%-2°-Regel für ein annehmbares Profil wird angemerkt.
Der Text über die Berechnung der Unsicherheit wurde verbessert und in einem Anhang wird ein Beispiel
bereitgestellt.
Es sind Anhänge zur Messspanne und zum bleibenden Druckverlust enthalten.
Eine Auflistung aller Teile der Normenreihe ISO 5167 kann auf der ISO-Internetseite abgerufen werden.
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
Einleitung
ISO 5167 ist in sechs Teile gegliedert und behandelt die Geometrie und Anwendungsverfahren (Einbau- und
Betriebsbedingungen) von Blenden, Düsen, Venturirohren, Konus- und Keil-Durchflussmessern, die in voll
durchströmten Leitungen eingesetzt sind, um den Durchfluss der Fluidströmung in der Leitung zu bestimmen.
Es werden zudem auch notwendige Informationen für die Berechnung des Durchflusses und der damit
verbundenen Unsicherheit gegeben.
ISO 5167 ist nur auf Drosselgeräte anwendbar, in denen die Strömung in allen Messquerschnitten im
Unterschallbereich bleibt und das Fluid als einphasig betrachtet werden kann; sie ist jedoch nicht auf die
Messung von pulsierenden Strömungen anwendbar. Ferner kann jedes dieser Geräte unkalibriert nur
innerhalb festgelegter Grenzen von Rohrweite und Reynoldszahl eingesetzt werden, oder alternativ über
seinen jeweiligen kalibrierten Bereich.
ISO 5167 behandelt Geräte, bei denen direkte Kalibrierversuche in ausreichender Anzahl, Qualität und
ausreichendem Umfang durchgeführt wurden, damit es bei kohärenten Anwendungssystemen möglich ist,
sich auf deren Ergebnisse und Beiwerte zu stützen, die innerhalb bestimmter vorhersagbarer
Unsicherheitsgrenzen anzugeben sind. ISO 5167 stellt auch eine Methodik für maßgeschneiderte Kalibrierung
von Differenzdruck-Durchflussmessern bereit.
Die in das Rohr eingebauten Geräte werden als Primärgeräte bezeichnet. Die Benennung Primärgerät schließt
auch die Druckentnahmen ein. Alle weiteren Messgeräte oder Geräte, die erforderlich sind, um die Messwert-
bestimmung der Messgeräte zu unterstützen, werden als Sekundärgeräte bezeichnet, und der Mengenumwer-
ter, der diese Messwerte empfängt und die Algorithmen durchführt, ist als Tertiärgerät bekannt. ISO 5167
behandelt Primärgeräte; Sekundärgeräte und Tertiärgeräte werden nur gelegentlich erwähnt.
ISO 5167 ist in die folgenden sechs Teile untergliedert:
a) ISO 5167-1 enthält allgemeine Begriffe, Symbole, Messprinzipien und Anforderungen sowie Mess-
verfahren und Angaben zur Unsicherheit, die in Verbindung mit ISO 5167, Teil 2 bis Teil 6 zu verwenden
sind.
b) ISO 5167-2 legt Anforderungen an Blenden fest, die mit Eck-Druckentnahmen, D- und D/2-Druckent-
nahmen und mit Flansch-Druckentnahmen angewendet werden können.
c) ISO 5167-3 legt Anforderungen an ISA 1932-Düsen , Langradiusdüsen und Venturidüsen fest, die sich in
der Form und Lage der Druckentnahmen voneinander unterscheiden. Langradiusdüsen mit Entnahme-
bohrung am Halsteil sind eingeschlossen.
d) ISO 5167-4 legt klassische Venturirohre fest.
e) ISO 5167-5 legt Konus-Durchflussmesser fest.
f) ISO 5167-6 legt Keil-Durchflussmesser fest.

Siehe ISO 2186:2007, Fluid flow in closed conduits — Connections for pressure signal transmissions between primary
and secondary elements.
Blenden mit Vena-Contracta-Druckentnahmen werden in ISO 5167 nicht behandelt.
ISA ist die Abkürzung für „International Federation of the National Standardizing Associations“, deren Nachfolger
1946 die ISO wurde.
In den USA wird das klassische Venturirohr mitunter „Herschel Venturi tube“ genannt.
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
Aspekte der Sicherheit werden in ISO 5167, Teil 1 bis Teil 6 nicht behandelt. Es liegt in der Verantwortung des
Anwenders sicherzustellen, dass das System den zutreffenden Sicherheitsvorschriften entspricht.
Zusätzliche Dokumente, die Unterstützung bieten dürfen, schließen folgende ein:
— ISO/TR 3313, Measurement of fluid flow in closed conduits — Guidelines on the effects of flow pulsations on
flow-measurement instruments
— ISO/TR 9464, Guidelines for the use of ISO 5167:2003
— ISO/TR 12767, Measurement of fluid flow by means of pressure differential devices — Guidelines on the
effect of departure from the specifications and operating conditions given in ISO 5167
— ISO/TR 15377, Measurement of fluid flow by means of pressure-differential devices — Guidelines for the
specification of orifice plates, nozzles and Venturi tubes beyond the scope of ISO 5167
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
1 Anwendungsbereich
Dieser Teil von ISO 5167 legt Begriffe und Symbole fest und gibt die allgemeinen Grundsätze für Mess- und
Berechnungsverfahren für den Durchfluss von Fluiden in voll durchströmten Leitungen mit Kreisquerschnitt
mittels Drosselgeräten (Blenden, Düsen und Venturirohre, Konus-Durchflussmesser und Keil-Durchfluss-
messer) an. Dieser Teil von ISO 5167 legt auch die allgemeinen Anforderungen an Messverfahren, den Einbau
von Drosselgeräten und die Ermittlung der Messunsicherheit der Durchflussmessung fest. Er legt ferner die
allgemeinen spezifizierten Grenzen für Rohrweite und Reynolds-Zahl fest, bei denen diese Drosselgeräte
anzuwenden sind.
ISO 5167 (alle Teile) gilt nur für Strömungen, die über den gesamten Messquerschnitt im Unterschallbereich
bleiben und bei denen das Fluid als einphasig betrachtet werden kann. Sie gilt nicht für die Messung von
pulsierenden Strömungen.
2 Normative Verweisungen
Die folgenden Dokumente werden im Text in solcher Weise in Bezug genommen, dass einige Teile davon oder
ihr gesamter Inhalt Anforderungen des vorliegenden Dokuments darstellen. Bei datierten Verweisungen gilt
nur die in Bezug genommene Ausgabe. Bei undatierten Verweisungen gilt die letzte Ausgabe des in Bezug
genommenen Dokuments (einschließlich aller Änderungen).
ISO 4006, Measurement of fluid flow in closed conduits — Vocabulary and symbols
ISO 5167-2:20xx, Measurement of fluid flow by means of pressure differential devices inserted in circular cross-
section conduits running full — Part 2: Orifice plates
ISO 5167-3:20xx, Measurement of fluid flow by means of pressure differential devices inserted in circular cross-
section conduits running full — Part 3: Nozzles and Venturi nozzles
ISO 5167-4:20xx, Measurement of fluid flow by means of pressure differential devices inserted in circular cross-
section conduits running full — Part 4: Venturi tubes
ISO 5167-5:20xx, Measurement of fluid flow by means of pressure differential devices inserted in circular cross-
section conduits running full — Part 5: Cone meters
ISO 5167-6:20xx, Measurement of fluid flow by means of pressure differential devices inserted in circular cross-
section conduits running full — Part 6: Wedge meters
ISO 5168:2005, Measurement of fluid flow — Procedures for the evaluation of uncertainties
ISO/IEC Guide 98-3:2008, Uncertainty of measurement — Part 3: Guide to the expression of uncertainty in
measurement (GUM:1995)
3 Begriffe
Für die Anwendung dieses Dokuments gelten die Begriffe nach ISO 4006 und die folgenden Begriffe.
ISO und IEC stellen terminologische Datenbanken für die Verwendung in der Normung unter den folgenden
Adressen bereit:
— ISO Online Browsing Platform: verfügbar unter https://www.iso.org/obp
— IEC Electropedia: verfügbar unter http://www.electropedia.org/
Anmerkung 1 zum Begriff: Die folgenden Definitionen beziehen sich nur auf Benennungen, die in einem bestimmten
Sinne angewendet werden oder bei denen es nützlich zu sein scheint, die Bedeutung hervorzuheben.
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
3.1 Druckmessung
3.1.1
Wand-Druckentnahme
ringförmiger Schlitz oder kreisrunde Bohrung in der Leitungswand, deren Kante mit der Innenwand der
Leitung bündig ist
Anmerkung 1 zum Begriff: Die Druckentnahme ist gewöhnlich eine kreisrunde Bohrung, in bestimmten Fällen kann sie
jedoch auch ein ringförmiger Schlitz sein.
3.1.2
statischer Druck eines Fluids, das durch eine Rohrleitung fließt
p
Druck, der durch Anschluss eines Druckmessgerätes an eine Wand-Druckentnahme gemessen werden kann
Anmerkung 1 zum Begriff: In ISO 5167 (alle Teile) wird nur der Wert des absoluten statischen Drucks betrachtet.
3.1.3
Wirkdruck
DP, en: differential pressure
Δp
Differenz der (statischen) Drücke zwischen den Wand-Druckentnahmen im Einlauf und Auslauf eines Primär-
gerätes (oder im Halsteil einer Düse mit Entnahmebohrung am Halsteil, einer Venturidüse oder eines Venturi-
rohres), das in eine gerade, durchströmte Rohrleitung eingebaut ist, wobei Höhenunterschiede zwischen den
Druckentnahmen im Ein- und Auslauf zu berücksichtigen sind
Anmerkung 1 zum Begriff: In ISO 5167 (alle Teile) wird die Benennung „Wirkdruck“ nur dann verwendet, wenn sich
die Druckentnahmen an den Stellen befinden, die für die jeweiligen Norm-Primärgeräte festgelegt sind.
3.1.4
Druckverhältnis
τ
Verhältnis des absoluten (statischen) Druckes an der Druckentnahme im Auslauf zu dem absoluten
(statischen) Druck an der Druckentnahme im Einlauf
3.1.5
Vena Contracta
Stelle in einem Fluidstrom, an der der Durchmesser des Stroms am geringsten ist
3.2 Primärgeräte
3.2.1
Drosselöffnung
Halsteil
kleinster Öffnungsquerschnitt in einem Primärgerät
Anmerkung 1 zum Begriff: Die Öffnungen in genormten Primärgeräten sind kreisrund und koaxial zur Rohrachse.
3.2.2
Blende
dünne Scheibe mit bearbeiteter kreisförmiger Öffnung
Anmerkung 1 zum Begriff: Genormte Blenden werden als „dünne Scheibe“ und mit „scharfer rechtwinkliger Kante“ der
Drosselöffnung beschrieben, da ihre Dicke im Vergleich zum Durchmesser des Messabschnitts klein ist und weil die
Einlaufkante der Drosselöffnung scharf und rechtwinklig ist.
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
3.2.3
Düse
Gerät, bestehend aus einem sich verengenden Einlauf und einem daran anschließenden zylindrischen Teil, der
im allgemeinen als „Halsteil“ bezeichnet wird
3.2.4
Venturidüse
Gerät, bestehend aus einem sich verengenden Einlauf, bei dem es sich um eine genormte ISA-1932-Düse
handelt, einem sich anschließenden zylindrischen Teil, der im Allgemeinen als „Halsteil“ bezeichnet wird und
der sich wiederum selbst an einen sich konisch erweiternden Auslauf anschließt, der als „Diffusor“ bezeichnet
wird
3.2.5
Venturirohr
Gerät, bestehend aus einem sich konisch verengenden Einlauf, einem sich anschließenden zylindrischen Teil,
der als „Halsteil“ bezeichnet wird und sich wiederum an einen sich konisch erweiternden Auslauf anschließt,
welcher als „Diffusor“ bezeichnet wird
3.2.6
Konus-Durchflussmesser
Gerät, das aus einer konusförmigen Verengung besteht, die in der Mitte des Rohres mit der Konusspitze gegen
die Durchflussrichtung gehalten wird
3.2.7
Keil-Durchflussmesser
Gerät, das eine keilförmige Verengung umfasst
3.2.8
Durchmesserverhältnis
β
Quadratwurzel des Verhältnisses der Fläche
des Halsteils des Primärgerätes zur Innenfläche des Messrohres im Einlauf des Primärgerätes
Anmerkung 1 zum Begriff: In Teil 2 und Teil 3 ist das Durchmesserverhältnis das Verhältnis des Durchmessers des
Halsteils des Primärgerätes zum Innendurchmesser des Messrohrs im Einlauf des Primärgerätes.
Anmerkung 2 zum Begriff: In Teil 4, bei dem das Primärgerät einen zylindrischen Einlauf aufweist, dessen Durch-
messer dem Rohrdurchmesser entspricht, ist das Durchmesserverhältnis das Verhältnis des Halsteil-Durchmessers zum
Durchmesser dieses zylindrischen Abschnitts in der Ebene der zulaufseitigen Druckentnahmen.
3.2.9
Fassungsring
Gerät, das genutzt wird, um das Primärelement in der Mitte des Rohrs zu halten und das die Druckentnahmen
enthalten darf
3.3 Strömung
3.3.1
Durchfluss
Durchflussrate
q
Masse oder Volumen eines Fluids, welches das Primärgerät je Zeiteinheit durchströmt
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
3.3.1.1
Massendurchfluss
Massendurchflussrate
q
m
Masse eines Fluids, welches das Primärgerät je Zeiteinheit durchströmt
3.3.1.2
Volumendurchfluss
Volumendurchflussrate
q
V
Volumen eines Fluids, welches das Primärgerät je Zeiteinheit durchströmt
Anmerkung 1 zum Begriff: Im Fall eines Volumendurchflusses ist die Angabe des Druckes und der Temperatur erfor-
derlich, auf die sich das Volumen bezieht.
3.3.2
Reynolds-Zahl
Re
dimensionsloser Parameter, der das Verhältnis zwischen den Trägheits- und Zähigkeitskräften angibt
3.3.2.1
Reynolds-Zahl, bezogen auf das Rohr
Re
D
dimensionsloser Parameter, der das Verhältnis zwischen den Trägheits- und Zähigkeitskräften im
zulaufseitigen Rohr angibt
𝑉𝑉𝐷𝐷 4𝑞𝑞
1 𝑚𝑚
𝑅𝑅𝑒𝑒 = =
𝐷𝐷
𝜈𝜈 π𝜇𝜇𝐷𝐷
1 1
3.3.2.2
Reynolds-Zahl, bezogen auf die Drosselöffnung oder den Halsteil
Re
d
dimensionsloser Parameter, der das Verhältnis zwischen den Trägheits- und Zähigkeitskräften in der Drossel-
öffnung oder dem Halsteil des Primärgerätes angibt
𝑅𝑅𝑒𝑒
𝐷𝐷
𝑅𝑅𝑒𝑒 =
𝑑𝑑
𝛽𝛽
3.3.3
Isentropenexponent
κ
Verhältnis der relativen Druckänderung zur zugehörigen relativen Dichteänderung bei Vorliegen einer
elementaren, umkehrbar adiabatischen (isentropen) Zustandsänderung
Anmerkung 1 zum Begriff: Der Isentropenexponent κ erscheint in den verschiedenen Gleichungen für die Expansions-
zahl ε und schwankt in Abhängigkeit von der Art des Gases und von dessen Temperatur und Druck.
Anmerkung 2 zum Begriff: Für viele Gase und Dämpfe sind bis jetzt noch keine Werte für κ veröffentlicht worden, vor
allem über einen weiten Bereich für Druck und Temperatur. Für die Anwendung von ISO 5167 (alle Teile) kann in solch
einem Fall das Verhältnis der spezifischen Wärmekapazität bei konstantem Druck zur spezifischen Wärmekapazität bei
konstantem Volumen idealer Gase statt des Isentropenexponenten genutzt werden.
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
3.3.4
Joule-Thomson-Koeffizient
μ
JT
Temperatur-Druck-Koeffizient bei konstanter Enthalpie
Änderungsgeschwindigkeit der Temperatur gegenüber dem Druck bei konstanter Enthalpie
𝜕𝜕𝜕𝜕
𝜇𝜇 = �
JT
𝜕𝜕𝜕𝜕
𝐻𝐻
Anmerkung 1 zum Begriff: Der Joule-Thomson-Koeffizient schwankt in Abhängigkeit von der Art des Gases und dessen
Temperatur und Druck und kann berechnet werden.
Anmerkung 2 zum Begriff: Ein Näherungswert für den Joule-Thomson-Koeffizienten ist für einige natürliche Gase in
ISO/TR 9464, 5.1.5.4.4 angegeben.
3.3.5
Durchflusskoeffizient
C
Kennzahl, die für einen inkompressiblen Fluidstrom festgelegt ist, und den Zusammenhang zwischen dem
tatsächlichen und dem theoretischen Durchfluss durch ein Gerät darstellt, und die durch die Gleichung für
inkompressible Fluide gegeben ist
𝑞𝑞 �1−𝛽𝛽
𝑚𝑚
𝐶𝐶 =
𝐴𝐴 2Δ𝜕𝜕𝜌𝜌

𝑡𝑡 1
Anmerkung 1 zum Begriff: Kalibrierungen von genormten Primärgeräten mittels inkompressibler Fluide (Flüssig-
keiten) zeigen, dass der Durchflusskoeffizient für ein bestimmtes Primärgerät bei bestimmten Einbauverhältnissen nur
von der Reynolds-Zahl abhängt.
Der numerischer Wert von C ist bei jedem einzelnen Differenzdruckmesser für verschiedene Einbauverhältnisse gleich,
wenn diese geometrisch ähnlich sind und die Strömungen identische Reynolds-Zahlen aufweisen.
Die Gleichungen für die Zahlenwerte von C in ISO 5167 (alle Teile) beruhen auf experimentell ermittelten Daten.
Die Messunsicherheit des Wertes von C kann durch eine Durchflusskalibrierung in einem geeigneten Laboratorium ver-
ringert werden.

Anmerkung 2 zum Begriff: Die Größe 1 �1−𝛽𝛽 wird als „Vorgeschwindigkeitsfaktor“ und das Produkt
𝐶𝐶
�1−𝛽𝛽
als „Durchflusszahl“ bezeichnet.
3.3.6
Expansionszahl
ε
Kennzahl, die verwendet wird, um die Kompressibilität des Fluids zu berücksichtigen
𝑞𝑞 1−𝛽𝛽

𝑚𝑚
𝜀𝜀 =
𝐴𝐴𝐶𝐶 2Δ𝜕𝜕𝜌𝜌

𝑡𝑡 1
Anmerkung 1 zum Begriff: Die Kalibrierung eines bestimmten Primärgerätes mit einem kompressiblen Fluid (Gas)
zeigt, dass das Verhältnis
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
𝑞𝑞 �1−𝛽𝛽
𝑚𝑚
𝐴𝐴 2Δ𝜕𝜕𝜌𝜌

𝑡𝑡 1
sowohl vom Wert der Reynolds-Zahl als auch von den Werten des Druckverhältnisses und des Isentropenexponenten des
Gases abhängt.
Das für die Darstellung dieser Schwankungen angewandte Verfahren besteht aus der Multiplikation des Durchfluss-
koeffizienten C des betrachteten Primärgerätes, wie er sich bei direkter Kalibrierung mit Flüssigkeiten für den gleichen
Wert der Reynolds-Zahl ergibt, mit der Expansionszahl ε.
Die Expansionszahl ε ist gleich eins, wenn das Fluid als inkompressibel (flüssig) betrachtet wird, und kleiner als eins,
wenn es kompressibel (gasförmig) ist.
Dieses Verfahren ist möglich, weil Versuche zeigen, dass ε praktisch von der Reynolds-Zahl unabhängig ist und für ein
gegebenes Durchmesserverhältnis eines gegebenen Primärgerätes nur vom Druckverhältnis und vom Isentropen-
exponenten abhängt.
Die Zahlenwerte von ε, die für Blenden in ISO 5167-2 und für Konus-Durchflussmesser in ISO 5167-5 angegeben werden,
beruhen auf experimentell ermittelten Daten. Für Düsen (siehe ISO 5167-3), Venturirohre (siehe ISO 5167-4) und
Keil-Durchflussmesser (siehe ISO 5167-6) basieren sie auf der allgemeinen thermodynamischen Gleichung für isentrope
Expansion.
3.3.7
arithmetischer Mittenrauwert
Ra
arithmetischer Mittelwert der Abweichung von der Mittellinie des gemessenen Profils
Anmerkung 1 zum Begriff: Die Mittellinie ist so definiert, dass die Summe der Quadrate der Abstände zwischen der
effektiven Oberfläche und der Mittellinie am geringsten ist. In der Praxis kann 𝑅𝑅𝑅𝑅 mit genormten Einrichtungen für
bearbeitete Oberflächen gemessen, jedoch für rauere Rohroberflächen nur geschätzt werden. Siehe auch ISO 4288.
Anmerkung 2 zum Begriff: Für Rohre kann auch die gleichmäßige äquivalente Rauheit k verwendet werden. Dieser
Wert kann experimentell bestimmt (siehe 7.1.5) oder aus Tabellen (siehe Anhang B) entnommen werden.
4 Symbole und Indizes
4.1 Symbole
Tabelle 1 — Symbole
a
Symbol Größe Dimension SI-Einheit
2 2
𝐴𝐴 Fläche des Halsteils
L m
𝑡𝑡
𝐶𝐶 Durchflusskoeffizient dimensionslos —
2 −2 −1 −1
𝐶𝐶 molare Wärmekapazität bei konstantem Druck J/(mol⋅K)
ML T Θ mol
m,𝑝𝑝
Durchmesser der Drosselöffnung (oder des Halsteils)
𝑑𝑑 L m
eines Primärgerätes unter Betriebsbedingungen
innerer Rohrdurchmesser stromaufwärts (oder zulauf-
𝐷𝐷 seitiger Durchmesser eines klassischen Venturirohres) L m
unter Betriebsbedingungen
2 −2 −1
𝐻𝐻 Enthalpie ML T mol J/mol
𝑘𝑘 gleichmäßige äquivalente Rauheit L m
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
a
Symbol Größe Dimension SI-Einheit
Druckverlust-Koeffizient (Verhältnis des Druck-
verlustes Δ𝜛𝜛 zum dynamischen Druck, ρV /2), auch
𝐾𝐾 dimensionslos —
bekannt als Verlustkoeffizient für geringfügige Verluste
(en: minor loss coefficient)
𝑙𝑙 Abstand der Druckentnahme vom Drosselgerät L m
relativer Abstand der Druckentnahme vom Drossel-
𝐿𝐿 dimensionslos —
gerät: L = l/D
−1 −2
𝜕𝜕 absoluter statischer Druck des Fluids Pa
ML T
−1
𝑞𝑞 Massendurchfluss MT kg/s
𝑚𝑚
3 −1 3
𝑞𝑞 Volumendurchfluss L T m /s
𝑉𝑉
𝑅𝑅 Radius L m
2 −2 −1 −1
𝑅𝑅 universelle Gaskonstante J/(mol⋅K)
ML T Θ mol
𝑢𝑢
𝑅𝑅𝑅𝑅 arithmetischer Mittenrauwert L m
𝑅𝑅𝑒𝑒 Reynolds-Zahl dimensionslos —
𝑅𝑅𝑒𝑒 Reynolds-Zahl des Halsteils dimensionslos —
𝑑𝑑
𝑅𝑅𝑒𝑒 Reynolds-Zahl des Rohrs dimensionslos —
𝐷𝐷
𝑡𝑡 Temperatur des Fluids Θ °C
𝜕𝜕 absolute (thermodynamische)Temperatur des Fluids Θ K
𝑈𝑈′ relative Unsicherheit dimensionslos —
−1
𝑉𝑉 mittlere axiale Geschwindigkeit des Fluids im Rohr LT m/s

𝑍𝑍 Realgasfaktor dimensionslos —

𝛽𝛽 Durchmesserverhältnis: 𝛽𝛽 =𝑑𝑑⁄𝐷𝐷 dimensionslos —
b
𝛾𝛾 Verhältnis der spezifischen Wärmekapazitäten dimensionslos —
c c
𝛿𝛿 absolute Unsicherheit
−1 −2
Δ𝜕𝜕 Wirkdruck: Δ𝜕𝜕 =𝜕𝜕 −𝜕𝜕 Pa
ML T
1 2
Druckverlust eines Strömungsumformers/-gleich-
−1 −2
Δ𝜕𝜕 ML T Pa
𝑐𝑐
richters
−1 −2
Δ𝜛𝜛 Druckverlust eines Primärgerätes ML T Pa
𝜀𝜀 Expansionszahl dimensionslos —
b
𝜅𝜅 Isentropenexponent dimensionslos —
𝜆𝜆 Rohrreibungszahl dimensionslos —
−1 −1
𝜇𝜇 dynamische Viskosität des Fluids Pa⋅s
ML T
−1 2
𝜇𝜇 Joule-Thomson-Koeffizient K/Pa
M LT Θ
JT
2 −1 2

𝜈𝜈 kinematische Viskosität des Fluids: 𝜈𝜈 =𝜇𝜇 𝜌𝜌 L T m /s
relativer Druckverlust (Verhältnis Druckverlust zum
𝜉𝜉 dimensionslos —
Wirkdruck)
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
a
Symbol Größe Dimension SI-Einheit
−3 3
𝜌𝜌 Dichte des Fluids
ML kg/m

𝜏𝜏 Druckverhältnis: 𝜏𝜏 =𝜕𝜕 𝜕𝜕 dimensionslos —
2 1
𝜙𝜙 Gesamtwinkel des Diffusorteils dimensionslos rad
a
M = Masse, L = Länge, T = Zeit, Θ = Temperatur
b
𝛾𝛾 ist das Verhältnis der spezifischen Wärmekapazität bei konstantem Druck zur spezifischen Wärmekapazität bei konstantem
Volumen. Für ideale Gase sind die Werte für das Verhältnis der spezifischen Wärmekapazitäten und für den Isentropen-
exponenten gleich groß (siehe 3.3.3). Diese Werte hängen von der Art des Gases ab.
c
Die Dimensionen und Einheiten sind die der zugehörigen Größe.

4.2 Indizes
Index Bedeutung
1 in der Ebene der zulaufseitigen Druckentnahme
2 in der Ebene der ausgangsseitigen Druckentnahme

5 Grundlage des Mess- und Berechnungsverfahrens
5.1 Grundlage des Messverfahrens
Die Grundlage des Messverfahrens beruht auf dem Einbau eines Primärgerätes in eine von einem Fluid voll
durchströmte Rohrleitung. Der Einbau des Primärgerätes führt zu einer Differenz der statischen Drücke
zwischen der Zulaufseite und dem Halsteil oder der Ausgangsseite des Gerätes. Der Durchfluss kann aus dem
gemessenen Wert dieser Druckdifferenz sowie mit Kenntnis der thermodynamischen Bedingungen,
Fluideigenschaften sowie der Geometrie und Eigenschaften des Messgerätes ermittelt werden. Es wird
vorausgesetzt, dass ein unkalibrierter Differenzdruckmesser innerhalb des erforderlichen geometrischen
Bereichs und des Reynolds-Zahl-Bereiches liegt, damit die Vorhersage des ISO-Durchflusskoeffizienten gültig
ist. Es wird vorausgesetzt, dass ein maßgeschneidert kalibrierter Differenzdruckmesser innerhalb seines
Kalibrierbereiches verwendet wird.
Der Massendurchfluss kann, da er sich auf den Wirkdruck bezieht, innerhalb der in ISO 5167 festgelegten
Grenzen für die Unsicherheit nach Gleichung (1) bestimmt werden:
𝐶𝐶
𝑞𝑞 = 𝜀𝜀𝐴𝐴�2Δ𝜕𝜕𝜌𝜌
(1)
𝑚𝑚 𝑡𝑡 1
�1−𝛽𝛽
ANMERKUNG Für die praktische Umsetzung wird diese Gleichung jeweils als Gleichung (1) von ISO 5167-2,
ISO 5167-3, ISO 5167-4, ISO 5167-5 und ISO 5167-6 erweitert.
Entsprechend kann der Wert des Volumendurchflusses nach Gleichung (2) berechnet werden:
𝑞𝑞
𝑚𝑚
𝑞𝑞 =
(2)
𝑉𝑉
𝜌𝜌
Dabei ist 𝜌𝜌 die Dichte des Fluids bei der Temperatur und dem Druck, für die das Volumen angegeben wird.
oSIST prEN ISO 5167-1:2021
prEN ISO 5167-1:2021 (D)
5.2 Verfahren für die Bestimmung des erforderlichen Durchmesserverhältnisses für das
ausgewählte genormte Primärgerät
Wenn in der Praxis das Durchmesserverhältnis eines Primärelementes, das in eine gegebene Rohrleitung
einzubauen ist, bestimmt wird, dann sind 𝐶𝐶 und ε, die in Gleichung (1) verwe
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.

Loading comments...