Packaging for terminally sterilized medical devices - Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) - Requirements and test methods

This draft European Standard provides test methods and values for paper, used in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) used as sterile barrier systems and/or packaging systems that are intended to maintain sterility of terminally sterilized medical devices to the point of use.
The need for a protective packaging may be determined by the manufacturer and the user.
This part of EN 868 only introduces performance requirements and test methods that are specific to the products covered by this part of EN 868 but does not add or modify the general requirements specified in EN ISO 11607-1.
As such, the particular requirements in 4.2 can be used to demonstrate compliance with one or more but not all of the requirements of EN ISO 11607-1.
When additional materials are used inside the sterile barrier system in order to ease the organization, drying or aseptic presentation (e.g. inner wrap, container filter indicators, packing lists, mats, instrument organizer sets, tray liners or an additional envelope around the medical device) then other requirements, including the determination of the acceptability of these materials during validation activities, may apply.
The materials specified in this part of EN 868 are intended for single use only.
NOTE   Applicable sterilization methods are specified by the manufacturer.

Verpackungen für in der Endverpackung zu sterilisierende Medizinprodukte - Teil 3: Papier zur Herstellung von Papierbeuteln (festgelegt in EN 868-4) und zur Herstellung von Klarsichtbeuteln und -schläuchen (festgelegt in EN 868-5) - Anforderungen und Prüfverfahren

Dieser europäische Norm-Entwurf enthält Prüfverfahren und Werte für Papier zur Herstellung von Papierbeuteln (festgelegt in EN 868 4) und zur Herstellung von Klarsichtbeuteln und  schläuchen (festgelegt in EN 868 5), die dazu vorgesehen sind, als Sterilbarrieresysteme und/oder Verpackungssysteme die Sterilität von in der Endverpackung sterilisierten Medizinprodukten bis zum Anwendungszeitpunkt zu erhalten.
Die Erfordernis für eine Schutzverpackung darf durch den Hersteller und den Anwender bestimmt werden.
Dieser Teil von EN 868 enthält lediglich Leistungsanforderungen und Prüfverfahren, die spezifisch für die Produkte nach diesem Teil von EN 868 sind, ergänzt oder modifiziert jedoch nicht die in EN ISO 11607 1 festgelegten allgemeinen Anforderungen.
Die besonderen Anforderungen nach 4.2 können daher verwendet werden, um Übereinstimmung mit einer oder mehreren Anforderung(en) - aber nicht mit allen Anforderungen - der EN ISO 11607 1 nachzuweisen.
Werden zusätzliche Materialien im Sterilbarrieresystem verwendet, um die Organisation, die Trocknung oder die aseptische Präsentation zu erleichtern (z. B. Innenverpackung, Filter für Sterilisier¬behälter, Indikatoren, Verpackungslisten, Unterlagen, Einsätze für Instrumente, Fächer oder eine zusätzliche Umhüllung für das Medizinprodukt), dann können andere Anforderungen, einschließlich der Bestimmung der Eignung dieser Materialien im Rahmen von Validierungstätigkeiten, gelten.
Die nach diesem Teil von EN 868 festgelegten Materialien sind nur für den einmaligen Gebrauch vorgesehen.
ANMERKUNG   Anwendbare Sterilisationsverfahren werden durch den Hersteller festgelegt.

Matériaux et systèmes d'emballage pour les dispositifs médicaux stérilisés au stade terminal - Partie 3 : Papier utilisé dans la fabrication des sacs en papier (spécifiés dans l'EN 868-4) et dans la fabrication de sachets et gaines (spécifiés dans l'EN 868-5) - Exigences et méthodes d'essai

Le présent projet de Norme européenne fournit des méthodes d’essai et des valeurs applicables au papier utilisé dans la fabrication des sacs en papier (spécifiés dans l’EN 868-4) et des sachets et gaines (spécifiés dans l’EN 868-5) utilisés comme systèmes de barrière stérile et/ou comme systèmes d’emballage destinés à maintenir l’état de stérilité des dispositifs médicaux, stérilisés au stade terminal, jusqu’à leur utilisation.
La nécessité d’un emballage de protection peut être déterminée par le fabricant et l’utilisateur.
La présente partie de l’EN 868 présente uniquement les exigences de performance et les méthodes d’essai spécifiques aux produits couverts par la présente partie de l’EN 868, mais elle n'ajoute pas ou ne modifie pas les exigences générales spécifiées dans l’EN ISO 11607-1.
En conséquence, les exigences particulières mentionnées en 4.2 peuvent être utilisées pour démontrer la conformité à une ou plusieurs exigences de l’EN ISO 11607-1, mais pas à la totalité de ces exigences.
En cas d'utilisation de matériaux complémentaires à l’intérieur du système de barrière stérile de manière à faciliter le rangement, le séchage ou la présentation aseptique (par exemple, enveloppe interne, filtre de conteneur, indicateurs, inventaires d’emballage, matelas de support silicone ou autres plastiques, kits d’organisation des instruments médicaux, fonds de protection de panier ou une enveloppe supplémentaire contenant le dispositif médical), d’autres exigences peuvent alors s’appliquer, y compris la détermination de l’acceptabilité de ces matériaux lors des activités de validation.
Les matériaux spécifiés dans la présente partie de l’EN 868 sont destinés exclusivement à un usage unique.
NOTE   Les méthodes de stérilisation applicables sont spécifiées par le fabricant.

Embalaža za končno sterilizirane medicinske pripomočke - 3. del: Papir za izdelavo papirnatih vrečk (specifikacija EN 868-4) in papir za izdelavo vrečk in zvitkov (specifikacija EN 868-5) - Zahteve in preskusne metode

Ta osnutek evropskega standarda podaja preskusne metode in vrednosti za papir za izdelavo papirnatih vrečk (specifikacija EN 868-4) ter papir za izdelavo vrečk in zvitkov (specifikacija EN 868-5), ki se uporablja kot sterilni pregradni sistemi in/ali sistemi embalaže, ki so namenjeni za vzdrževanje sterilnosti končno steriliziranih medicinskih pripomočkov do njihove uporabe.
Potrebo po zaščitni embalaži lahko določita proizvajalec in uporabnik.
Ta del standarda EN 868 le uvaja zahteve delovanja in preskusne metode, ki se uporabljajo za proizvode, zajete v tem delu standarda EN 868, vendar ne dodaja ali spreminja splošnih zahtev, podanih v standardu EN ISO 11607-1.
Posebne zahteve iz točke 4.2 se lahko uporabljajo za dokazovanje skladnosti z eno ali več zahtevami, vendar ne z vsemi zahtevami standarda EN ISO 11607-1.
Ko se znotraj sterilnega pregradnega sistema uporabljajo dodatni materiali za lažjo organizacijo, sušenje ali aseptičnost (npr. notranji ovoj, filtrirni vsebniki, kazalniki, embalažni listi, podloge, kompleti organizatorjev za inštrumente, obloge za pladnje ali dodatni ovoj okoli medicinskega pripomočka), se lahko uporabljajo druge zahteve, vključno z ugotavljanjem sprejemljivosti teh materialov med preverjanjem.
Materiali, določeni v tem delu standarda ISO 868, so namenjeni le enkratni uporabi.

General Information

Status
Published
Public Enquiry End Date
29-Oct-2015
Publication Date
27-Feb-2017
Technical Committee
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
15-Feb-2017
Due Date
22-Apr-2017
Completion Date
28-Feb-2017

Relations

Buy Standard

Standard
EN 868-3:2017
English language
19 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day
Draft
prEN 868-3:2015
English language
19 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 868-3:2017
01-april-2017
1DGRPHãþD
SIST EN 868-3:2009
(PEDODåD]DNRQþQRVWHULOL]LUDQHPHGLFLQVNHSULSRPRþNHGHO3DSLU]DL]GHODYR
SDSLUQDWLKYUHþN VSHFLILNDFLMD(1 LQSDSLU]DL]GHODYRYUHþNLQ]YLWNRY
VSHFLILNDFLMD(1 =DKWHYHLQSUHVNXVQHPHWRGH
Packaging for terminally sterilized medical devices - Part 3: Paper for use in the
manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches
and reels (specified in EN 868-5) - Requirements and test methods
Verpackungen für in der Endverpackung zu sterilisierende Medizinprodukte - Teil 3:
Papier zur Herstellung von Papierbeuteln (festgelegt in EN 868-4) und zur Herstellung
von Klarsichtbeuteln und -schläuchen (festgelegt in EN 868-5) - Anforderungen und
Prüfverfahren
Matériaux et systèmes d'emballage pour les dispositifs médicaux stérilisés au stade
terminal - Partie 3 : Papier utilisé dans la fabrication des sacs en papier (spécifiés dans
l'EN 868-4) et dans la fabrication de sachets et gaines (spécifiés dans l'EN 868-5) -
Exigences et méthodes d'essai
Ta slovenski standard je istoveten z: EN 868-3:2017
ICS:
11.080.30 Sterilizirana embalaža Sterilized packaging
55.080 9UHþH9UHþNH Sacks. Bags
SIST EN 868-3:2017 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

SIST EN 868-3:2017

---------------------- Page: 2 ----------------------

SIST EN 868-3:2017


EN 868-3
EUROPEAN STANDARD

NORME EUROPÉENNE

February 2017
EUROPÄISCHE NORM
ICS 11.080.30 Supersedes EN 868-3:2009
English Version

Packaging for terminally sterilized medical devices - Part
3: Paper for use in the manufacture of paper bags
(specified in EN 868-4) and in the manufacture of pouches
and reels (specified in EN 868-5) - Requirements and test
methods
Emballages des dispositifs médicaux stérilisés au stade Verpackungsmaterialien für in der Endverpackung zu
terminal - Partie 3: Papier utilisé dans la fabrication sterilisierende Medizinprodukte - Teil 3: Papier zur
des sacs en papier (spécifiés dans l'EN 868-4) et dans Herstellung von Papierbeuteln (festgelegt in EN 868-4)
la fabrication de sachets et gaines (spécifiés dans l'EN und zur Herstellung von Klarsichtbeuteln und -
868-5) - Exigences et méthodes d'essai schläuchen (festgelegt in EN 868-5) - Anforderungen
und Prüfverfahren
This European Standard was approved by CEN on 4 December 2016.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.

This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and United Kingdom.





EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2017 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 868-3:2017 E
worldwide for CEN national Members.

---------------------- Page: 3 ----------------------

SIST EN 868-3:2017
EN 868-3:2017 (E)
Contents Page
European foreword . 3
Introduction . 4
1 Scope . 5
2 Normative references . 5
3 Terms and definitions . 6
4 Requirements . 6
Annex A (informative) Details of significant technical changes between this European
Standard and the previous edition . 9
Annex B (normative) Method for the determination of water repellency . 10
B.1 Apparatus . 10
B.2 Reagent . 10
B.3 Procedure. 10
B.4 Repeatability and reproducibility. 11
B.5 Test report . 11
Annex C (normative) Method for the determination of pore size . 12
C.1 Principle . 12
C.2 Test liquid . 12
C.3 Apparatus . 12
C.4 Preparation of test specimens . 14
C.5 Procedure. 14
C.6 Result . 15
C.6.1 Calculation and expression of results . 15
C.6.2 Derivation of formula for calculation of equivalent pore radius . 15
C.7 Repeatability and reproducibility. 16
C.8 Test report . 16
Annex D (informative) Repeatability and reproducibility of test methods . 17
Bibliography . 19

2

---------------------- Page: 4 ----------------------

SIST EN 868-3:2017
EN 868-3:2017 (E)
European foreword
This document (EN 868-3:2017) has been prepared by Technical Committee CEN/TC 102 “Sterilizers
and associated equipment for processing of medical devices”, the secretariat of which is held by DIN.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by August 2017, and conflicting national standards shall
be withdrawn at the latest by August 2017.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN 868-3:2009.
Annex A provides details of significant technical changes between this European Standard and the
previous edition.
EN 868 consists of the following parts, under the general title Packaging for terminally sterilized medical
devices:
— Part 2: Sterilization wrap — Requirements and test methods;
— Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture
of pouches and reels (specified in EN 868-5) — Requirements and test methods;
— Part 4: Paper bags — Requirements and test methods;
— Part 5: Sealable pouches and reels of porous materials and plastic film construction — Requirements
and test methods;
— Part 6: Paper for low temperature sterilization processes — Requirements and test methods;
— Part 7: Adhesive coated paper for low temperature sterilization processes — Requirements and test
methods;
— Part 8: Re-usable sterilization containers for steam sterilizers conforming to EN 285 — Requirements
and test methods;
— Part 9: Uncoated nonwoven materials of polyolefines — Requirements and test methods;
— Part 10: Adhesive coated nonwoven materials of polyolefines — Requirements and test methods.
In addition, ISO/TC 198 “Sterilization of health care products” in collaboration with CEN/TC 102
“Sterilizers and associated equipment for processing of medical devices” has prepared the
EN ISO 11607 series “Packaging for terminally sterilized medical devices”. The EN ISO 11607 series
specifies general requirements for materials, sterile barrier systems and packaging systems (Part 1)
and validation requirements for forming, sealing and assembly processes (Part 2).
According to the CEN-CENELEC Internal Regulations, the national standards organisations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria,
Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia,
France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta,
Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
3

---------------------- Page: 5 ----------------------

SIST EN 868-3:2017
EN 868-3:2017 (E)
Introduction
The EN ISO 11607 series consists of two parts under the general title “Packaging for terminally
sterilized medical devices”. Part 1 of this series specifies general requirements and test methods for
materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are
intended to maintain sterility of terminally sterilized medical devices to the point of use. Part 2 of this
series specifies validation requirements for forming, sealing and assembly processes.
General requirements for all types of sterile barrier systems are provided by EN ISO 11607-1.
The EN 868 series can be used to demonstrate compliance with one or more of the requirements
specified in EN ISO 11607-1.
CEN/TC 102/WG 4 also appreciates the initiatives of CEN with regard to the minimization of adverse
environmental impacts by standards. It was agreed that this subject should be given priority during the
next edition of the EN ISO 11607 series that is the basic reference for all parts of the EN 868 series.
4

---------------------- Page: 6 ----------------------

SIST EN 868-3:2017
EN 868-3:2017 (E)
1 Scope
This European Standard specifies test methods and values for paper, used in the manufacture of paper
bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) used
as sterile barrier systems and/or packaging systems that are intended to maintain sterility of terminally
sterilized medical devices to the point of use.
Other than the general requirements as specified in EN ISO 11607-1 and EN ISO 11607-2 this part of
EN 868 specifies materials, test methods and values that are specific to the products covered by this
European Standard.
The materials specified in this part of EN 868 are intended for single use only.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated
references, the latest edition of the referenced document (including any amendments) applies.
EN 20187, Paper, board and pulps - Standard atmosphere for conditioning and testing and procedure for
monitoring the atmosphere and conditioning of samples (ISO 187)
EN ISO 535, Paper and board - Determination of water absorptiveness - Cobb method (ISO 535)
EN ISO 536, Paper and board - Determination of grammage (ISO 536)
EN ISO 1924-2, Paper and board - Determination of tensile properties - Part 2: Constant rate of elongation
method (20 mm/min) (ISO 1924-2)
EN ISO 1974, Paper - Determination of tearing resistance - Elmendorf method (ISO 1974)
EN ISO 2758, Paper - Determination of bursting strength (ISO 2758)
EN ISO 11607-1:2009+A1:2014, Packaging for terminally sterilized medical devices - Part 1:
Requirements for materials, sterile barrier systems and packaging systems (ISO 11607-
1:2006+AMD1:2014)
ISO 2470-2, Paper, board and pulps — Measurement of diffuse blue reflectance factor — Part 2: Outdoor
daylight conditions (D65 brightness)
ISO 3689, Paper and board — Determination of bursting strength after immersion in water
ISO 3781, Paper and board — Determination of tensile strength after immersion in water
ISO 5636-3, Paper and board — Determination of air permeance (medium range) — Part 3: Bendtsen
method
ISO 6588-2:2012, Paper, board and pulps — Determination of pH of aqueous extracts — Part 2: Hot
extraction
ISO 8601, Data elements and interchange formats — Information interchange — Representation of dates
and times
ISO 9197, Paper, board and pulps — Determination of water-soluble chlorides
5

---------------------- Page: 7 ----------------------

SIST EN 868-3:2017
EN 868-3:2017 (E)
ISO 9198, Paper, board and pulp — Determination of water-soluble sulfates
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN ISO 11607-1:2009+A1:2014
apply.
4 Requirements
4.1 General
For any material, preformed sterile barrier system or sterile barrier system, the requirements of
EN ISO 11607-1 shall apply.
This part of EN 868 only introduces performance requirements and test methods that are specific to the
products covered by this part of EN 868 but does not add or modify the general requirements specified
in EN ISO 11607-1.
As such, the particular requirements in 4.2 can be used to demonstrate compliance with one or more
but not all of the requirements of EN ISO 11607-1.
NOTE 1 Compliance to EN 868–3 does not automatically mean compliance to EN ISO 11607-1.
A confirmation of compliance to EN 868-3 shall contain a statement whether EN ISO 11607-1 is
covered.
NOTE 2 When additional materials are used inside the sterile barrier system in order to ease the organization,
drying or aseptic presentation (e.g. inner wrap, container filter, indicators, packing lists, mats, instrument
organizer sets, tray liners or an additional envelope around the medical device) then other requirements,
including the determination of the acceptability of these materials during validation activities, can apply.
4.2 Performance requirements and test methods
NOTE See Annex D for repeatability and reproducibility of the test methods: pore diameters, sulphate
content, chloride content and water repellency. For information on statement of precision and/or bias,
repeatability and reproducibility of other test methods, see EN ISO 11607-1:2009+A1:2014, Table B.1.
4.2.1 The requirements of 4.2.11 and 4.2.15 do not apply to materials solely used in irradiation
sterilization packaging.
4.2.2 No colour shall leach out of the paper. Compliance shall be tested by visual examination of a hot
aqueous extract prepared in accordance with the method given in ISO 6588-2.
2
4.2.3 The average mass of 1 m of the conditioned paper when tested in accordance with EN ISO 536
shall be within ± 5 % of the nominal value stated by the manufacturer.
4.2.4 The pH of an aqueous extract of the paper shall be not less than 5 or greater than 8 when tested
in accordance with ISO 6588-2, hot extraction method.
4.2.5 The chloride content of the paper, calculated as sodium chloride, shall not exceed 0,05 % when
tested in accordance with ISO 9197 using a hot extract prepared in accordance with ISO 6588-2:2012,
7.2 except that 2 ml of potassium chloride solution is not added.
6

---------------------- Page: 8 ----------------------

SIST EN 868-3:2017
EN 868-3:2017 (E)
4.2.6 The sulphate content of the paper, calculated as sodium sulphate, shall not exceed 0,25 % when
tested in accordance with ISO 9198, using a hot extract prepared in accordance with ISO 6588-2:2012,
7.2 except that 2 ml of potassium chloride solution is not added.
4.2.7 When tested in accordance with ISO 2470-2 the material shall not exhibit an increase in D65
brightness, due to the optical brightener agents, of more than 1 %; calculated as the ratio of the D65
brightness measured with the 420 nm UV-cut-off filter in place to the D65 brightness measured without
420 nm UV-cut-off filter.
2
4.2.8 When exposed at 25 cm from a UV light source, the material shall not have per 0,01 m more than
five fluorescent spots, each having an axi
...

SLOVENSKI STANDARD
oSIST prEN 868-3:2015
01-oktober-2015
(PEDODåD]DNRQþQRVWHULOL]LUDQHPHGLFLQVNHSULSRPRþNHGHO3DSLU]DL]GHODYR
SDSLUQDWLKYUHþN VSHFLILNDFLMD(1 LQSDSLU]DL]GHODYRYUHþNLQQHVNRQþQLK
]YLWNRY VSHFLILNDFLMD(1 =DKWHYHLQSUHVNXVQHPHWRGH
Packaging for terminally sterilized medical devices - Part 3: Paper for use in the
manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches
and reels (specified in EN 868-5) - Requirements and test methods
Verpackungen für in der Endverpackung zu sterilisierende Medizinprodukte - Teil 3:
Papier zur Herstellung von Papierbeuteln (festgelegt in EN 868-4) und zur Herstellung
von Klarsichtbeuteln und -schläuchen (festgelegt in EN 868-5) - Anforderungen und
Prüfverfahren
Matériaux et systèmes d'emballage pour les dispositifs médicaux stérilisés au stade
terminal - Partie 3 : Papier utilisé dans la fabrication des sacs en papier (spécifiés dans
l'EN 868-4) et dans la fabrication de sachets et gaines (spécifiés dans l'EN 868-5) -
Exigences et méthodes d'essai
Ta slovenski standard je istoveten z: prEN 868-3
ICS:
11.080.30 Sterilizirana embalaža Sterilized packaging
oSIST prEN 868-3:2015 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
oSIST prEN 868-3:2015

---------------------- Page: 2 ----------------------
oSIST prEN 868-3:2015

EUROPEAN STANDARD
DRAFT
prEN 868-3
NORME EUROPÉENNE

EUROPÄISCHE NORM

August 2015
ICS 11.080.30 Will supersede EN 868-3:2009
English Version
Packaging for terminally sterilized medical devices - Part 3:
Paper for use in the manufacture of paper bags (specified in EN
868-4) and in the manufacture of pouches and reels (specified in
EN 868-5) - Requirements and test methods
 Verpackungen für in der Endverpackung zu sterilisierende
Medizinprodukte - Teil 3: Papier zur Herstellung von
Papierbeuteln (festgelegt in EN 868-4) und zur Herstellung
von Klarsichtbeuteln und -schläuchen (festgelegt in EN 868-
5) - Anforderungen und Prüfverfahren
This draft European Standard is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee CEN/TC 102.

If this draft becomes a European Standard, CEN members are bound to comply with the CEN/CENELEC Internal Regulations which
stipulate the conditions for giving this European Standard the status of a national standard without any alteration.

This draft European Standard was established by CEN in three official versions (English, French, German). A version in any other language
made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania,
Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United
Kingdom.

Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are aware and to
provide supporting documentation.

Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without notice and
shall not be referred to as a European Standard.


EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Avenue Marnix 17, B-1000 Brussels
© 2015 CEN All rights of exploitation in any form and by any means reserved Ref. No. prEN 868-3:2015 E
worldwide for CEN national Members.

---------------------- Page: 3 ----------------------
oSIST prEN 868-3:2015
prEN 868-3:2015 (E)
Contents Page
Foreword .3
Introduction .4
1 Scope .5
2 Normative references .5
3 Terms and definitions .6
4 Requirements .6
4.1 General .6
4.2 Performance requirements and test methods .6
4.3 Marking .7
4.3.1 Protective packaging .7
4.3.2 Labelling of individual units .8
5 Information to be supplied by the manufacturer .8
Annex A (informative) Details of significant technical changes between this draft European
Standard and the previous edition.9
Annex B (normative) Method for the determination of water repellency . 10
B.1 Apparatus . 10
B.2 Reagent . 10
B.3 Procedure . 10
B.4 Repeatability and reproducibility . 10
B.5 Test report . 11
Annex C (normative) Method for the determination of pore size . 12
C.1 Principle . 12
C.2 Test liquid . 12
C.3 Apparatus . 12
C.4 Preparation of test specimens . 13
C.5 Procedure . 13
C.6 Result . 14
C.6.1 Calculation and expression of results . 14
C.6.2 Derivation of formula for calculation of equivalent pore radius . 14
C.7 Repeatability and reproducibility . 15
C.8 Test report . 15
Annex D (informative) Repeatability and reproducibility of test methods . 17
Bibliography . 19

2

---------------------- Page: 4 ----------------------
oSIST prEN 868-3:2015
prEN 868-3:2015 (E)
Foreword
This document (prEN 868-3:2015) has been prepared by Technical Committee CEN/TC 102 “Sterilizers for
medical purposes”, the secretariat of which is held by DIN.
This document is currently submitted to the Enquiry.
This document will supersede EN 868-3:2009.
Annex A provides details of significant technical changes between this draft European Standard and the
previous edition.
EN 868 consists of the following parts, under the general title Packaging for terminally sterilized medical
devices:
— Part 2: Sterilization wrap — Requirements and test methods;
— Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of
pouches and reels (specified in EN 868-5) — Requirements and test methods;
— Part 4: Paper bags — Requirements and test methods;
— Part 5: Sealable pouches and reels of porous materials and plastic film construction — Requirements and
test methods;
— Part 6: Paper for low temperature sterilization processes — Requirements and test methods;
— Part 7: Adhesive coated paper for low temperature sterilization processes — Requirements and test
methods;
— Part 8: Re-usable sterilization containers for steam sterilizers conforming to EN 285 — Requirements and
test methods;
— Part 9: Uncoated nonwoven materials of polyolefines — Requirements and test methods;
— Part 10: Adhesive coated nonwoven materials of polyolefines — Requirements and test methods.
In addition, ISO/TC 198 “Sterilization of health care products” in collaboration with CEN/TC 102, Sterilizers for
medical purposes has prepared the EN ISO 11607 series, Packaging for terminally sterilized medical devices.
The EN ISO 11607 series specifies general requirements for materials, sterile barrier systems and packaging
systems (Part 1) and validation requirements for forming, sealing and assembly processes (Part 2).
3

---------------------- Page: 5 ----------------------
oSIST prEN 868-3:2015
prEN 868-3:2015 (E)
Introduction
The EN ISO 11607 series consists of two parts under the general title Packaging for terminally sterilized
medical devices. Part 1 of this series specifies general requirements and test methods for materials,
preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain
sterility of terminally sterilized medical devices to the point of use. Part 2 of this series specifies validation
requirements for forming, sealing and assembly processes.
Every sterile barrier system needs to fulfil the requirements of EN ISO 11607-1.
The EN 868 series can be used to demonstrate compliance with one or more of the requirements specified in
EN ISO 11607-1.
CEN/TC 102/WG 4 also appreciates the initiatives of CEN with regard to the minimization of adverse
environmental impacts by standards. It was agreed that this subject should be given priority during the next
edition of the EN ISO 11607 series that is the basic reference for all parts of the EN 868 series.
4

---------------------- Page: 6 ----------------------
oSIST prEN 868-3:2015
prEN 868-3:2015 (E)
1 Scope
This draft European Standard provides test methods and values for paper, used in the manufacture of paper
bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) used as
sterile barrier systems and/or packaging systems that are intended to maintain sterility of terminally sterilized
medical devices to the point of use.
The need for a protective packaging may be determined by the manufacturer and the user.
This part of EN 868 only introduces performance requirements and test methods that are specific to the
products covered by this part of EN 868 but does not add or modify the general requirements specified in
EN ISO 11607-1.
As such, the particular requirements in 4.2 can be used to demonstrate compliance with one or more but not
all of the requirements of EN ISO 11607-1.
When additional materials are used inside the sterile barrier system in order to ease the organization, drying
or aseptic presentation (e.g. inner wrap, container filter indicators, packing lists, mats, instrument organizer
sets, tray liners or an additional envelope around the medical device) then other requirements, including the
determination of the acceptability of these materials during validation activities, may apply.
The materials specified in this part of EN 868 are intended for single use only.
NOTE Applicable sterilization methods are specified by the manufacturer.
2 Normative references
The following documents, in whole or in part, are normatively referenced in this document and are
indispensable for its application. For dated references, only the edition cited applies. For undated references,
the latest edition of the referenced document (including any amendments) applies.
EN 20187, Paper, board and pulps — Standard atmosphere for conditioning and testing and procedure for
monitoring the atmosphere and conditioning of samples (ISO 187:1990)
EN 20535, Paper and board — Determination of water absorptiveness — Cobb method (ISO 535:1991)
EN 21974, Paper — Determination of tearing resistance (Elmendorf method) (ISO 1974:1990)
EN ISO 536, Paper and board — Determination of grammage (ISO 536)
EN ISO 1924-2, Paper and board — Determination of tensile properties — Part 2: Constant rate of elongation
method (20 mm/min) (ISO 1924-2)
EN ISO 2758, Paper — Determination of bursting strength (ISO 2758)
EN ISO 11607-1:2006, Packaging for terminally sterilized medical devices — Part 1: Requirements for
materials, sterile barrier systems and packaging systems (ISO 11607-1:2006)
ISO 2470-2, Paper, board and pulps — Measurement of diffuse blue reflectance factor — Part 2: Outdoor
daylight conditions (D65 brightness)
ISO 3689, Paper and board — Determination of bursting strength after immersion in water
ISO 3781, Paper and board — Determination of tensile strength after immersion in water
5

---------------------- Page: 7 ----------------------
oSIST prEN 868-3:2015
prEN 868-3:2015 (E)
ISO 5636-3, Paper and board — Determination of air permeance (medium range) — Part 3: Bendtsen method
ISO 6588-2:2005, Paper, board and pulps — Determination of pH of aqueous extracts — Part 2: Hot
extraction
ISO 8601, Data elements and interchange formats — Information interchange — Representation of dates and
times
ISO 9197, Paper, board and pulps — Determination of water-soluble chlorides
ISO 9198, Paper, board and pulp — Determination of water-soluble sulfates
3 Terms and definitions
For the purposes of this document, the terms and definitions given in EN ISO 11607-1:2006 apply.
4 Requirements
4.1 General
For any material, preformed sterile barrier system or sterile barrier system, the requirements of
EN ISO 11607-1 apply. Statements on compliance with prEN 868-3 shall include statements on compliance
with EN ISO 11607-1.
This part of EN 868 only introduces performance requirements and test methods that are specific to the
products covered by this part of EN 868 but does not add or modify the general requirements specified in
EN ISO 11607-1.
As such, the particular requirements in 4.2 can be used to demonstrate compliance with one or more but not
all of the requirements of EN ISO 11607-1.
NOTE Compliance to prEN 868–3 does not automatically mean compliance to EN ISO 11607-1.
A confirmation of compliance to prEN 868-3 shall contain a statement whether EN ISO 11607-1 is covered.
4.2 Performance requirements and test methods
NOTE See Annex D for repeatability and reproducibility of the test methods: pore diameters, sulfate content, chloride
content and water repellency. For information on statement of precision and/or bias, repeatability and reproducibility of
other test methods, see Table B.1 in EN ISO 11607-1:2009/A1:2014.
4.2.1 The requirements of 4.2.11 and 4.2.15 do not apply to materials solely used in irradiation sterilization
packaging.
4.2.2 No colour shall leach out of the paper. Compliance shall be tested by visual examination of a hot
aqueous extract prepared in accordance with the method given in ISO 6588-2.
2
of the conditioned paper when tested in accordance with EN ISO 536 shall
4.2.3 The average mass of 1 m
be within ± 5 % of the nominal value stated by the manufacturer.
4.2.4 The pH of an aqueous extract of the paper shall be not less than 5 nor greater than 8 when tested in
accordance with ISO 6588-2.
6

---------------------- Page: 8 ----------------------
oSIST prEN 868-3:2015
prEN 868-3:2015 (E)
4.2.5 The chloride content of the paper, calculated as sodium chloride, shall not exceed 0,05 % when
tested in accordance with ISO 9197 using a hot extract prepared in accordance with ISO 6588-2:2005, 7.2
except that 2 ml of potassium chloride solution is not added.
4.2.6 The sulfate content of the paper, calculated as sodium sulfate, shall not exceed 0,25 % when tested in
accordance with ISO 9198, using a hot extract prepared in accordance with ISO 6588-2:2005, 7.2 except that
2 ml of potassium chloride solution is not added.
4.2.7
4.2.7.1 When tested in accordance with ISO 2470-2, the material shall not exhibit an increase in D65
brightness, due to the optical brightener agents, of more than 1 %; calculated as the ratio of the D65
brightness measured with the 420 nm UV-cut-off filter in place to the D65 brightness measured without
420 nm UV-cut-off filter.
2
more
4.2.7.2 When exposed at 25 cm from a UV light source, the material shall not have per 0,01 m
than five fluorescent spots, each having an axis greater than 1 mm.
NOTE The UV light to be used is the one described as per Annex B.
4.2.8 The internal tea
...