F40.01 - Test Methods
Test Methods
General Information
SIGNIFICANCE AND USE
5.1 This test method is intended for the determination of chromium, bromine, cadmium, mercury, and lead, in homogeneous polymeric materials. The test method may be used to ascertain the conformance of the product under test to manufacturing specifications. Typical time for a measurement is 5 to 10 min per specimen, depending on the specimen matrix and the capabilities of the EDXRF spectrometer.
SCOPE
1.1 This test method describes an energy dispersive X-ray fluorescence (EDXRF) spectrometric procedure for identification and quantification of chromium, bromine, cadmium, mercury, and lead in polymeric materials.
1.2 This test method is not applicable to determine total concentrations of polybrominated biphenyls (PBB), polybrominated diphenyl ethers (PBDE) or hexavalent chromium. This test method cannot be used to determine the valence states of atoms or ions.
1.3 This test method is applicable for a range from 20 mg/kg to approximately 1 wt % for chromium, bromine, cadmium, mercury, and lead in polymeric materials.
1.4 This test method is applicable for homogeneous polymeric material.
1.5 The values stated in SI units are to be regarded as the standard. Values given in parentheses are for information only.
1.6 This test method is not applicable to quantitative determinations for specimens with one or more surface coatings present on the analyzed surface; however, qualitative information may be obtained. In addition, specimens less than infinitely thick for the measured X rays, must not be coated on the reverse side or mounted on a substrate.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard12 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This test method may be used for quantitative determinations of Pb in painted and unpainted articles such as toys, children’s products, and other consumer products. Typical test time for quantification of Pb in homogenous samples is 1 to 3 min; and typical test time for quantification of Pb in paint is 4 to 8 min.
SCOPE
1.1 This test method uses energy dispersive X-ray fluorescence (EDXRF) spectrometry for detection and quantification of lead (Pb) in paint layers, similar coatings, or substrates and homogenous materials. The following material types were tested in the interlaboratory study for this standard test method: ABS plastic, polyethylene, polypropylene, PVC, glass, zinc alloy, wood, and fabric.
1.2 This technique may also be commonly referred to as High Definition X-ray Fluorescence (HDXRF) or Multiple Monochromatic Beam EDXRF (MMB-EDXRF).
1.3 This test method is applicable for the products and materials described in 1.1 for a Pb mass fraction range of 14 to 1200 mg/kg for uncoated samples and 30 to 450 mg/kg for coated samples, as specified in Table 1 and determined by an interlaboratory study using representative samples
1.4 Ensure that the analysis area of the sample is visually uniform in appearance and at least as large as the X-ray excitation beam at the point of sample excitation.
1.5 For coating analysis, this test method is limited to paint and similar coatings. Metallic coatings are not covered by this test method.
1.6 X-ray Nomenclature—This standard names X-ray lines using the IUPAC convention with the Siegbahn convention in parentheses.
1.7 There are no known ISO equivalent methods to this standard.
1.8 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.10 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard7 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Tin-based solder alloys are commonly used to manufacture electrical and electronic goods. The elements lead, cadmium, mercury, antimony and bismuth are often declarable substances in solder materials. This test method provides a means of determining the listed declarable substances, as well as other minor and trace constituents, in tin-based solder alloys.
4.2 Two methods of dissolving tin-based solder alloys are given in this standard. The first method uses open-vessel hydrofluoric and nitric acid room temperature digestions; the second method employs closed-vessel nitric and hydrofluoric acid microwave digestions, both for use only with ICP-AES instruments equipped with a hydrofluoric acid resistant sample introduction system.
4.3 The method of preparing calibration solutions uses 1000 mg/kg single element reference material solutions, and uses matching concentrated acids for both the calibration solutions and the sample solutions.
4.4 This test method is intended for use by laboratories experienced with the set-up, calibration and analysis of samples using ICP-AES.
SCOPE
1.1 This test method covers procedures for the analysis of tin-based solder alloys for minor and trace elements using inductively-coupled plasma atomic emission spectrometry (ICP-AES) instrumentation.
1.2 These test procedures were validated for the analytes and mass fractions listed below.
Element
Validated Mass Fraction
Range, mg/kg
Lead
115 to 965
Cadmium
25 to 60
Mercury
5 to 530
Antimony
85 to 1330
Bismuth
80 to 210
Arsenic
95 to 360
Silver
4000 to 42100
Cobalt
0.5 to 60
Iron
15 to 115
Chromium
0.5 to 1.5
Copper
3000 to 30600
Indium
25 to 115
Nickel
5 to 150
Phosphorus
10 to 110
Selenium
1 to 30
Zinc
2 to 160
Aluminum
1 to 3
1.3 The procedures appear in the following order:
Procedure
Section
Internal Standardization
8
Calibration Solution Preparations
9
Preparation of Sample and Validation Solutions
10
Calibration
11
Analysis Procedure
12
1.4 The values stated in SI units are to be regarded as the standard. Any other values are for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard8 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This test method provides for analysis of Pb in applied paint using measurement times on the order of several minutes. It can be used to determine whether the sample of applied paint has an areal mass of Pb either substantially less than a specification limit, and therefore does not exceed it, or substantially above the specified limit, and therefore exceeds it.
5.2 If the value obtained with this test method falls close to a specification limit, a more precise test method may be required to positively determine whether Pb content does or does not exceed the specified limit.
SCOPE
1.1 This test method describes an energy dispersive X-ray fluorescence (EDXRF) procedure for determining the areal mass of Pb in mass per unit area in paint and similar coatings on common substrates of toys and consumer products, such as plastic, wood, steel, aluminum, zinc alloys or fabric.
1.2 This test method is applicable for homogeneous, single layer paint or similar coatings. The method does not apply to metallic coatings.
1.3 This test method is applicable for a range of Pb mass per unit area from 0.36 μg/cm2 to approximately 10 μg/cm2 for Pb in paint and similar coatings applied on common substrates. The lower limit of this test method is between 0.36 and 0.75 μg/cm2 depending on the nature of the substrate. Based on the results obtained during the interlaboratory study (ASTM Report F40-1004), it is estimated that the applicable range of this method can be extended up to 50 μg/cm2.
1.4 The values stated in SI units are to be regarded as standard. Values given in parentheses are for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard15 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Waste glass is currently recycled into various consumer products. This test method has been developed as a tool for evaluation of heavy metals in glass to satisfy reporting requirements for maximum allowable content for some applications.
5.2 The ranges within which this test method is quantitative are given in Table 1.
5.3 For amounts of the analyte elements outside the ranges in Table 1, this test method provides screening results. That is, it provides an unambiguous indication that each element can be described as present in an amount greater than the scope upper limit or that the amount of the element can be described as less than the scope lower limit with a high degree of confidence.
Note 2: In general, when a quantitative result is obtained, the analyst can make a clear decision as to whether a material is suitable for the intended purpose. When the contents of elements of interest are outside the quantitative range, the analyst can still make a decision whether the amount is too high or whether additional analyses are required.
5.4 These methods can be applied to glass beads, plate glass, float glass, fiber glass, or ground glass. This test method has been validated for the ranges of matrix compositions that are summarized in Table 2.
5.5 Detection limits, sensitivity, and element ranges will vary with matrices, detector type, and other instrument conditions and parameters.
5.6 All analytes are determined as the element and reported as such. These include all elements listed in Table 1. This test method may be applicable to other glass matrices, additional elements, and wider concentration ranges provided the laboratory is able to validate the broadened scope of this test method.
SCOPE
1.1 This test method covers field portable X-ray fluorescence (XRF) spectrometric procedures for analyses of arsenic and lead in glass compositions using field portable energy dispersive XRF spectrometers.
1.2 The mass fraction range of arsenic within which this test method is quantitative is given in Table 1. Scope limits were determined from the interlaboratory study results using the approach given in Practice E1601.
1.3 The mass fraction range for which lead was tested is given in Table 1. However, lead results cannot be considered quantitative on the basis of single-sample results because the precision performance is not good enough to allow laboratories to compare results in a quantitative manner.
Note 1: The performance of this test method was evaluated using results based on single-sample determinations from specimens composed of glass beads. One laboratory has determined that performance can be significantly improved by basing reported results on the mean of determinations from multiple samples to overcome inherent heterogeneity of elements in glass beads, especially the element lead. Additional information is provided in Section 17 on Precision and Bias.
1.3.1 To obtain quantitative performance, lead results must consist of the average of four or more determinations.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 7 on Hazards.
1.6 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
- Standard9 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This test method may be used for quantitative determinations of Pb in painted and unpainted articles such as toys, children’s products, and other consumer products. Typical test time for quantification of Pb in homogenous samples is 1 to 3 min; and typical test time for quantification of Pb in paint is 4 to 8 min.
SCOPE
1.1 This test method uses energy dispersive X-ray fluorescence (EDXRF) spectrometry for detection and quantification of lead (Pb) in paint layers, similar coatings, or substrates and homogenous materials. The following material types were tested in the interlaboratory study for this standard test method: ABS plastic, polyethylene, polypropylene, PVC, glass, zinc alloy, wood, and fabric.
1.2 This technique may also be commonly referred to as High Definition X-ray Fluorescence (HDXRF) or Multiple Monochromatic Beam EDXRF (MMB-EDXRF).
1.3 This test method is applicable for the products and materials described in 1.1 for a Pb mass fraction range of 14 to 1200 mg/kg for uncoated samples and 30 to 450 mg/kg for coated samples, as specified in Table 1 and determined by an interlaboratory study using representative samples
1.4 Ensure that the analysis area of the sample is visually uniform in appearance and at least as large as the X-ray excitation beam at the point of sample excitation.
1.5 For coating analysis, this test method is limited to paint and similar coatings. Metallic coatings are not covered by this test method.
1.6 X-ray Nomenclature—This standard names X-ray lines using the IUPAC convention with the Siegbahn convention in parentheses.
1.7 There are no known ISO equivalent methods to this standard.
1.8 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard7 pagesEnglish language
- Standard7 pagesEnglish language
- Standard7 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This test method is intended for the determination of chromium, bromine, cadmium, mercury, and lead, in homogeneous polymeric materials. The test method may be used to ascertain the conformance of the product under test to manufacturing specifications. Typical time for a measurement is 5 to 10 min per specimen, depending on the specimen matrix and the capabilities of the EDXRF spectrometer.
SCOPE
1.1 This test method describes an energy dispersive X-ray fluorescence (EDXRF) spectrometric procedure for identification and quantification of chromium, bromine, cadmium, mercury, and lead in polymeric materials.
1.2 This test method is not applicable to determine total concentrations of polybrominated biphenyls (PBB), polybrominated diphenyl ethers (PBDE) or hexavalent chromium. This test method cannot be used to determine the valence states of atoms or ions.
1.3 This test method is applicable for a range from 20 mg/kg to approximately 1 wt % for chromium, bromine, cadmium, mercury, and lead in polymeric materials.
1.4 This test method is applicable for homogeneous polymeric material.
1.5 The values stated in SI units are to be regarded as the standard. Values given in parentheses are for information only.
1.6 This test method is not applicable to quantitative determinations for specimens with one or more surface coatings present on the analyzed surface; however, qualitative information may be obtained. In addition, specimens less than infinitely thick for the measured X rays, must not be coated on the reverse side or mounted on a substrate.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard12 pagesEnglish language
- Standard12 pagesEnglish language
- Standard12 pagesEnglish language
SIGNIFICANCE AND USE
5.1 This test method provides for analysis of Pb in applied paint using measurement times on the order of several minutes. It can be used to determine whether the sample of applied paint has an areal mass of Pb either substantially less than a specification limit, and therefore does not exceed it, or substantially above the specified limit, and therefore exceeds it.
5.2 If the value obtained with this test method falls close to a specification limit, a more precise test method may be required to positively determine whether Pb content does or does not exceed the specified limit.
SCOPE
1.1 This test method describes an energy dispersive X-ray fluorescence (EDXRF) procedure for determining the areal mass of Pb in mass per unit area in paint and similar coatings on common substrates of toys and consumer products, such as plastic, wood, steel, aluminum, zinc alloys or fabric.
1.2 This test method is applicable for homogeneous, single layer paint or similar coatings. The method does not apply to metallic coatings.
1.3 This test method is applicable for a range of Pb mass per unit area from 0.36 μg/cm2 to approximately 10 μg/cm2 for Pb in paint and similar coatings applied on common substrates. The lower limit of this test method is between 0.36 and 0.75 μg/cm2 depending on the nature of the substrate. Based on the results obtained during the interlaboratory study (ASTM Report F40-1004), it is estimated that the applicable range of this method can be extended up to 50 μg/cm2.
1.4 The values stated in SI units are to be regarded as standard. Values given in parentheses are for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard14 pagesEnglish language
- Standard14 pagesEnglish language
SIGNIFICANCE AND USE
4.1 Tin-based solder alloys are commonly used to manufacture electrical and electronic goods. The elements lead, cadmium, mercury, antimony and bismuth are often declarable substances in solder materials. This test method provides a means of determining the listed declarable substances, as well as other minor and trace constituents, in tin-based solder alloys.
4.2 Two methods of dissolving tin-based solder alloys are given in this standard. The first method uses open-vessel hydrofluoric and nitric acid room temperature digestions; the second method employs closed-vessel nitric and hydrofluoric acid microwave digestions, both for use only with ICP-AES instruments equipped with a hydrofluoric acid resistant sample introduction system.
4.3 The method of preparing calibration solutions uses 1000 mg/kg single element reference material solutions, and uses matching concentrated acids for both the calibration solutions and the sample solutions.
4.4 This test method is intended for use by laboratories experienced with the set-up, calibration and analysis of samples using ICP-AES.
SCOPE
1.1 This test method covers procedures for the analysis of tin-based solder alloys for minor and trace elements using inductively-coupled plasma atomic emission spectrometry (ICP-AES) instrumentation.
1.2 These test procedures were validated for the analytes and mass fractions listed below.
Element
Validated Mass Fraction
Range, mg/kg
Lead
115 to 965
Cadmium
25 to 60
Mercury
5 to 530
Antimony
85 to 1330
Bismuth
80 to 210
Arsenic
95 to 360
Silver
4000 to 42100
Cobalt
0.5 to 60
Iron
15 to 115
Chromium
0.5 to 1.5
Copper
3000 to 30600
Indium
25 to 115
Nickel
5 to 150
Phosphorus
10 to 110
Selenium
1 to 30
Zinc
2 to 160
Aluminum
1 to 3
1.3 The procedures appear in the following order:
Procedure
Section
Internal Standardization
8
Calibration Solution Preparations
9
Preparation of Sample and Validation Solutions
10
Calibration
11
Analysis Procedure
12
1.4 The values stated in SI units are to be regarded as the standard. Any other values are for information only.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard8 pagesEnglish language
- Standard8 pagesEnglish language
SIGNIFICANCE AND USE
5.1 Waste glass is currently recycled into various consumer products. This test method has been developed as a tool for evaluation of heavy metals in glass to satisfy reporting requirements for maximum allowable content for some applications.
5.2 The ranges within which this test method is quantitative are given in Table 1.
5.3 For amounts of the analyte elements outside the ranges in Table 1, this test method provides screening results. That is, it provides an unambiguous indication that each element can be described as present in an amount greater than the scope upper limit or that the amount of the element can be described as less than the scope lower limit with a high degree of confidence.Note 2—In general, when a quantitative result is obtained, the analyst can make a clear decision as to whether a material is suitable for the intended purpose. When the contents of elements of interest are outside the quantitative range, the analyst can still make a decision whether the amount is too high or whether additional analyses are required.
5.4 These methods can be applied to glass beads, plate glass, float glass, fiber glass, or ground glass. This test method has been validated for the ranges of matrix compositions that are summarized in Table 2.
5.5 Detection limits, sensitivity, and element ranges will vary with matrices, detector type, and other instrument conditions and parameters.
5.6 All analytes are determined as the element and reported as such. These include all elements listed in Table 1. This test method may be applicable to other glass matrices, additional elements, and wider concentration ranges provided the laboratory is able to validate the broadened scope of this test method.
SCOPE
1.1 This test method covers field portable X-ray fluorescence (XRF) spectrometric procedures for analyses of arsenic and lead in glass compositions using field portable energy dispersive XRF spectrometers.
1.2 The mass fraction range of arsenic within which this test method is quantitative is given in Table 1. Scope limits were determined from the interlaboratory study results using the approach given in Practice E1601.
1.3 The mass fraction range for which lead was tested is given in Table 1. However, lead results cannot be considered quantitative on the basis of single-sample results because the precision performance is not good enough to allow laboratories to compare results in a quantitative manner.Note 1—The performance of this test method was evaluated using results based on single-sample determinations from specimens composed of glass beads. One laboratory has determined that performance can be significantly improved by basing reported results on the mean of determinations from multiple samples to overcome inherent heterogeneity of elements in glass beads, especially the element lead. Additional information is provided in Section 17 on Precision and Bias.
1.3.1 To obtain quantitative performance, lead results must consist of the average of four or more determinations.
1.4 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.5 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Some specific hazards statements are given in Section 7 on Hazards.
- Standard9 pagesEnglish language
SIGNIFICANCE AND USE
This test method may be used for quantitative determinations of Pb in painted and unpainted articles such as toys, children’s products, and other consumer products. Typical test time for quantification of Pb in homogenous samples is 1 to 3 min; and typical test time for quantification of Pb in paint is 4 to 8 min.
SCOPE
1.1 This test method uses energy dispersive X-ray fluorescence (EDXRF) spectrometry for detection and quantification of lead (Pb) in paint layers, similar coatings, or substrates and homogenous materials. The following material types were tested in the interlaboratory study for this standard test method: ABS plastic, polyethylene, polypropylene, PVC, glass, zinc alloy, wood, and fabric.
1.2 This technique may also be commonly referred to as High Definition X-ray Fluorescence (HDXRF) or Multiple Monochromatic Beam EDXRF (MMB-EDXRF).
1.3 This test method is applicable for the products and materials described in 1.1 for a Pb mass fraction range of 14 to 1200 mg/kg for uncoated samples and 30 to 450 mg/kg for coated samples, as specified in Table 1 and determined by an interlaboratory study using representative samples
1.4 Ensure that the analysis area of the sample is visually uniform in appearance and at least as large as the X-ray excitation beam at the point of sample excitation.
1.5 For coating analysis, this test method is limited to paint and similar coatings. Metallic coatings are not covered by this test method.
1.6 X-ray Nomenclature—This standard names X-ray lines using the IUPAC convention with the Siegbahn convention in parentheses.
1.7 There are no known ISO equivalent methods to this standard.
1.8 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard7 pagesEnglish language
SIGNIFICANCE AND USE
This test method may be used for quantitative determinations of Pb in painted and unpainted articles such as toys, children’s products, and other consumer products. Typical test time for quantification of Pb in homogenous samples is 1 to 3 min; and typical test time for quantification of Pb in paint is 4 to 8 min.
SCOPE
1.1 This test method uses energy dispersive X-ray fluorescence (EDXRF) spectrometry for detection and quantification of lead (Pb) in paint layers, similar coatings, or substrates and homogenous materials. The following material types were tested in the interlaboratory study for this standard test method: ABS plastic, polyethylene, polypropylene, PVC, glass, zinc alloy, wood, and fabric.
1.2 This technique may also be commonly referred to as High Definition X-ray Fluorescence (HDXRF) or Multiple Monochromatic Beam EDXRF (MMB-EDXRF).
1.3 This test method is applicable for the products and materials described in 1.1 for a Pb mass fraction range of 14 to 1200 mg/kg for uncoated samples and 30 to 450 mg/kg for coated samples, as specified in Table 1 and determined by an interlaboratory study using representative samples
1.4 Ensure that the analysis area of the sample is visually uniform in appearance and at least as large as the X-ray excitation beam at the point of sample excitation.
1.5 For coating analysis, this test method is limited to paint and similar coatings. Metallic coatings are not covered by this test method.
1.6 X-ray Nomenclature—This standard names X-ray lines using the IUPAC convention with the Siegbahn convention in parentheses.
1.7 There are no known ISO equivalent methods to this standard.
1.8 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard7 pagesEnglish language
SIGNIFICANCE AND USE
This test method is intended for the determination of chromium, bromine, cadmium, mercury, and lead, in homogeneous polymeric materials. The test method may be used to ascertain the conformance of the product under test to manufacturing specifications. Typical time for a measurement is 5 to 10 min per specimen, depending on the specimen matrix and the capabilities of the EDXRF spectrometer.
SCOPE
1.1 This test method describes an energy dispersive X-ray fluorescence (EDXRF) spectrometric procedure for identification and quantification of chromium, bromine, cadmium, mercury, and lead in polymeric materials.
1.2 This test method is not applicable to determine total concentrations of polybrominated biphenyls (PBB), polybrominated diphenyl ethers (PBDE) or hexavalent chromium. This test method cannot be used to determine the valence states of atoms or ions.
1.3 This test method is applicable for a range from 20 mg/kg to approximately 1 wt % for chromium, bromine, cadmium, mercury, and lead in polymeric materials.
1.4 This test method is applicable for homogeneous polymeric material.
1.5 The values stated in SI units are to be regarded as the standard. Values given in parentheses are for information only.
1.6 This test method is not applicable to quantitative determinations for specimens with one or more surface coatings present on the analyzed surface; however, qualitative information may be obtained. In addition, specimens less than infinitely thick for the measured X rays, must not be coated on the reverse side or mounted on a substrate.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard9 pagesEnglish language
SIGNIFICANCE AND USE
This test method is intended for the determination of chromium, bromine, cadmium, mercury, and lead, in homogeneous polymeric materials. The test method may be used to ascertain the conformance of the product under test to manufacturing specifications. Typical time for a measurement is 5 to 10 min per specimen, depending on the specimen matrix and the capabilities of the EDXRF spectrometer.
SCOPE
1.1 This test method describes an energy dispersive X-ray fluorescence (EDXRF) spectrometric procedure for identification and quantification of chromium, bromine, cadmium, mercury, and lead in polymeric materials.
1.2 This test method is not applicable to determine total concentrations of polybrominated biphenyls (PBB), polybrominated diphenyl ethers (PBDE) or hexavalent chromium. This test method cannot be used to determine the valence states of atoms or ions.
1.3 This test method is applicable for a range from 20 mg/kg to approximately 1 wt % for chromium, bromine, cadmium, mercury, and lead in polymeric materials.
1.4 This test method is applicable for homogeneous polymeric material.
1.5 The values stated in SI units are to be regarded as the standard. Values given in parentheses are for information only.
1.6 This test method is not applicable to quantitative determinations for specimens with one or more surface coatings present on the analyzed surface; however, qualitative information may be obtained. In addition, specimens less than infinitely thick for the measured X rays, must not be coated on the reverse side or mounted on a substrate.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
- Standard9 pagesEnglish language
Frequently Asked Questions
F40.01 is a Technical Committee within ASTM International. It is named "Test Methods". This committee has published 14 standards.
F40.01 develops ASTM standards in the area of Information technology. Currently, there are 14 published standards from this technical committee.
ASTM is a standardization organization that develops and publishes standards to support industry, commerce, and regulatory requirements.
A Technical Committee (TC) in ASTM is a group of experts responsible for developing international standards in a specific technical area. TCs are composed of national member body delegates and work through consensus to create standards that meet global industry needs. Each TC may have subcommittees (SCs) and working groups (WGs) for specialized topics.