Standard Practice for Using a Guarded-Hot-Plate Apparatus or Thin-Heater Apparatus in the Single-Sided Mode

SIGNIFICANCE AND USE
This practice provides a procedure for operating the apparatus so that the heat flow, Q′, through the meter section of the auxiliary insulation is small; determining Q′; and, calculating the heat flow, Q, through the meter section of the specimen.
This practice requires that the apparatus have independent temperature controls in order to operate the cold plate and auxiliary cold plate at different temperatures. In the single-sides mode, the apparatus is operated with the temperature of the auxiliary cold plate maintained, as close as possible, to the temperature of the side of the hot plate adjacent to the auxiliary insulation.
Note 4—Ideally, if the temperature difference across the auxiliary insulation is zero and there are no edge heat losses or gains, all of the power input to the meter plate will flow through the specimen. In practice, a small correction is made for heat flow, Q′, through the auxiliary insulation.
The thermal conductance, C′, of the auxiliary insulation must be determined from one or more separate tests using either Test Method C 177, C 1114, or as indicated in 5.4. The values of C′ should be checked periodically, particularly if during regular testing it is not possible to keep the temperature drop across the auxiliary insulation less than 1 % of the temperature drop across the test specimen.
This practice can be used when it is desirable to determine the thermal properties of a single specimen. For example, the thermal properties of a single specimen are used to calibrate a heat-flow-meter apparatus for Test Method C 518. In other cases, there may be only one specimen available.
SCOPE
1.1 This practice covers the determination of the steady-state heat flow through the meter section of a specimen when a guarded-hot-plate apparatus or thin-heater apparatus is used in the single-sided mode of operation.
1.2 This practice provides a supplemental procedure for use in conjunction with either Test Method C 177 or C 1114 for testing a single specimen. This practice is limited to only the single-sided mode of operation, and, in all other particulars, the requirements of either Test Method C 177 or C 1114 apply.Note 1 - Test Methods C 177 and C 1114 describe the use of the guarded-hot-plate and thin-heater apparatus, respectively, for determining steady-state heat flux and thermal transmission properties of flat-slab specimens. In principle, these methods cover both the double- and single-sided mode of operation, and at present, do not distinguish between the accuracies for the two modes of operation. When appropriate, thermal transmission properties shall be calculated in accordance with Practice C 1045.
1.3 This practice requires that the cold plates of the apparatus have independent temperature controls. For the single-sided mode of operation, a (single) specimen is placed between the hot plate and the cold plate. Auxiliary thermal insulation, if needed, is placed between the hot plate and the auxiliary cold plate. The auxiliary cold plate and the hot plate are maintained at essentially the same temperature. Ideally, the heat flow from the meter plate is assumed to flow only through the specimen, so that the thermal transmission properties correspond only to the specimen.Note 2 - The double-sided mode of operation requires similar specimens placed on either side of the hot plate. The cold plates that contact the outer surfaces of these specimens are maintained at essentially the same temperature. The electric power supplied to the meter plate is assumed to result in equal heat flow through the meter section of each specimen, so that the thermal transmission properties correspond to an average for the two specimens.
1.4 This practice does not preclude the use of a guarded-hot-plate apparatus in which the auxiliary cold plate may be either larger or smaller in lateral dimensions than either the test specimen or the cold plate.Note 3 - Most guarded-hot-plate apparat...

General Information

Status
Historical
Publication Date
31-Aug-2007
Technical Committee
Drafting Committee
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM C1044-07 - Standard Practice for Using a Guarded-Hot-Plate Apparatus or Thin-Heater Apparatus in the Single-Sided Mode
English language
8 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation: C1044 – 07
Standard Practice for
Using a Guarded-Hot-Plate Apparatus or Thin-Heater
1
Apparatus in the Single-Sided Mode
This standard is issued under the fixed designation C1044; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 1.4 This practice does not preclude the use of a guarded-
hot-plate apparatus in which the auxiliary cold plate may be
1.1 This practice covers the determination of the steady-
eitherlargerorsmallerinlateraldimensionsthaneitherthetest
state heat flow through the meter section of a specimen when
specimen or the cold plate.
a guarded-hot-plate apparatus or thin-heater apparatus is used
in the single-sided mode of operation.
NOTE 3—Most guarded-hot-plate apparatus are designed for the
2
1.2 This practice provides a supplemental procedure for use
double-sided mode of operation (1). Consequently, the cold plate and the
auxiliary cold plate are the same size and the specimen and the auxiliary
in conjunction with either Test Method C177 or C1114 for
insulation will have the same lateral dimensions, although the thickness
testing a single specimen. This practice is limited to only the
may be different. Some guarded-hot-plate apparatus, however, are de-
single-sidedmodeofoperation,and,inallotherparticulars,the
signed specifically for testing only a single specimen that may be either
requirements of either Test Method C177 or C1114 apply.
larger or smaller in lateral dimensions than that auxiliary insulation or the
auxiliary cold plate.
NOTE 1—Test Methods C177 and C1114 describe the use of the
guarded-hot-plate and thin-heater apparatus, respectively, for determining
1.5 This practice can be used for both low- and high-
steady-state heat flux and thermal transmission properties of flat-slab
temperature conditions.
specimens. In principle, these methods cover both the double- and
1.6 This practice shall not be used when operating an
single-sided mode of operation, and at present, do not distinguish between
apparatus in a double-sided mode of operation with a known
the accuracies for the two modes of operation. When appropriate, thermal
and unknown specimen, that is, with the two cold plates at
transmission properties shall be calculated in accordance with Practice
C1045. similar temperatures so that the temperature differences across
the known and unknown specimens are similar.
1.3 This practice requires that the cold plates of the appa-
1.7 This standard does not purport to address all of the
ratus have independent temperature controls. For the single-
safety concerns, if any, associated with its use. It is the
sidedmodeofoperation,a(single)specimenisplacedbetween
responsibility of the user of this standard to establish appro-
the hot plate and the cold plate.Auxiliary thermal insulation, if
priate safety and health practices and determine the applica-
needed, is placed between the hot plate and the auxiliary cold
bility of regulatory limitations prior to use.
plate. The auxiliary cold plate and the hot plate are maintained
at essentially the same temperature. Ideally, the heat flow from
2. Referenced Documents
the meter plate is assumed to flow only through the specimen,
3
2.1 ASTM Standards:
so that the thermal transmission properties correspond only to
3
C168 Terminology Relating to Thermal Insulation
the specimen.
C177 Test Method for Steady-State Heat Flux Measure-
NOTE 2—The double-sided mode of operation requires similar speci-
ments and Thermal Transmission Properties by Means of
mensplacedoneithersideofthehotplate.Thecoldplatesthatcontactthe
3
the Guarded-Hot-Plate Apparatus
outer surfaces of these specimens are maintained at essentially the same
C518 Test Method for Steady-State Thermal Transmission
temperature. The electric power supplied to the meter plate is assumed to
3
Properties by Means of the Heat Flow Meter Apparatus
result in equal heat flow through the meter section of each specimen, so
C1045 PracticeforCalculatingThermalTransmissionProp-
that the thermal transmission properties correspond to an average for the
3
two specimens.
erties Under Steady-State Conditions
1 2
This practice is under the jurisdiction of ASTM Committee C16 on Thermal The boldface numbers in parentheses refer to the list of references at the end of
Insulation and is the direct responsibility of Subcommittee C16.30 on Thermal this standard.
3
Measurement. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved Sept. 1, 2007. Published September 2007. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
ap
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.