Standard Test Methods for Flash Point by Small Scale Closed Cup Tester

SCOPE
1.1 These test methods cover procedures for the determination of the flash point by a small scale closed tester. The procedures may be used to determine the actual flash point temperature of a sample or whether a product will or will not flash at a specified temperature (flash/no flash).
1.2 &si-value;
1.3 This standard should be used to measure and describe the properties of materials, products, or assemblies in response to heat and flame under controlled laboratory conditions and should not be used to describe or appraise the fire hazard or fire risk of materials, products, or assemblies under actual fire conditions. However, results of this test may be used as elements of a fire risk assessment which takes into account all of the factors which are pertinent to an assessment of the fire hazard of a particular end useThis standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Warning statements appear throughout. See also the Material Safety Data Sheets for the product being tested.

General Information

Status
Historical
Publication Date
30-Apr-2005
Current Stage
Ref Project

Relations

Effective Date
01-May-2005
Effective Date
01-May-2005

Buy Standard

Standard
ASTM D3828-05 - Standard Test Methods for Flash Point by Small Scale Closed Cup Tester
English language
8 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
An American National Standard
Designation:D3828–05
Designation: 303/83 (88)
Standard Test Methods for
1
Flash Point by Small Scale Closed Cup Tester
This standard is issued under the fixed designation D3828; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision.Anumber in parentheses indicates the year of last reapproval.A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.
INTRODUCTION
The flash point method is generally used for testing a sample at a specific temperature. At a set
temperature, the specimen being tested and the air–vapor mixture above it are close to thermal
equilibrium. Test methods for other flash point equipment operated at a specific temperature are
described in Test Method D3941.
Flash point values are a function of the apparatus design, the condition of the apparatus used, and
the operational procedure carried out. Flash point can therefore only be defined in terms of a standard
test method, and no general valid correlation can be guaranteed between results obtained by different
test methods, or with test apparatus different from that specified.
1. Scope* priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.Warningstatements
1.1 These test methods cover procedures for the determina-
appearthroughout.SeealsotheMaterialSafetyDataSheetsfor
tion of the flash point by a small scale closed tester. The
the product being tested.
procedures may be used to determine the actual flash point
temperature of a sample or whether a product will or will not
2. Referenced Documents
flash at a specified temperature (flash/no flash).
2
2.1 ASTM Standards:
1.2 The values stated in SI units are to be regarded as the
D3941 Test Method for Flash Point by the Equilibrium
standard. The values given in parentheses are for information
Method with a Closed-Cup Apparatus
only.
D6299 Practice forApplying Statistical QualityAssurance
1.3 This standard should be used to measure and describe
Techniques to Evaluate Analytical Measurement System
the properties of materials, products, or assemblies in response
Performance
to heat and flame under controlled laboratory conditions and
3
2.2 ISO Standards:
should not be used to describe or appraise the fire hazard or
Guide34 Quality Systems Guidelines for the Production of
fire risk of materials, products, or assemblies under actual fire
Reference Materials
conditions. However, results of this test may be used as
Guide35 Certification of Reference Materials—General
elements of a fire risk assessment which takes into account all
and Statistical Principles
of the factors which are pertinent to an assessment of the fire
hazard of a particular end use.
3. Terminology
1.4 This standard does not purport to address all of the
3.1 Definitions:
safety concerns, if any, associated with its use. It is the
3.1.1 equilibrium—the vapor above the liquid (specimen)
responsibility of the user of this standard to establish appro-
andtheliquidinaflashpointapparatusspecimencupareatthe
same temperature at the time the ignition source is applied.
1
These test methods are under the jurisdiction of ASTM Committee D02 on
PetroleumProductsandLubricantsandisthedirectresponsibilityofSubcommittee
2
D02.08 on Volatility. For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Current edition approved May 1, 2005. Published May 2005. Originally contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
approved in 1979. Last previous edition approved in 2002 as D3828–02. Standards volume information, refer to the standard’s Document Summary page on
This is also a standard of Energy Institute issued under the fixed designation IP the ASTM website.
3
303. The final number indicates year of last revision. These test methods were Available fromAmerican National Standards Institute (ANSI), 25 W. 43rd St.,
adopted as a joint ASTM-IP standard in 1979. 4th Floor, New York, NY 10036.
*A Summary of Changes section appears at the end of this standard.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
D3828–05
TABLE 1 Calibration of Tester
A,B C A C
Material p-xylene (1.4–Dimethylbenzene) (Warning) n-butanol (Butan-1-ol) (Warning)
Specific gravity, 15.6/15.6°C (60/60°F) 0.860 to 0.866 0.810 to 0.812
Boiling range, °C (°F) 2 (4) maximum including 138.35 (281.03) 2 (4) maximum including 117.5 (243.5)
Freezing point, °C (°F) 52.2 (11.23) minimum −90 (−130) minimum
Flash point, °C (°F) (acceptable range) 25.6 6 0.5 (78 6 1) 36.6 6 0.8 (97.9 6 1.7)