Standard Specification for Illuminators Used for Viewing Industrial Radiographs

ABSTRACT
This specification provides the recommended minimum requirements for industrial radiographic illuminators used for viewing industrial radiographic films using transmitted light sources. The function of the illuminator is to provide sufficient illumination and viewing capabilities for the purpose of identification and interpretation of radiographic images. The luminance of the transmitted light shall not be less than 30 candelas per square metre for film densities equal to or less than 2.5 optical density and not less than 10 candelas per square meter for film densities greater than 2.5 optical density. If the illuminator has a diffusing screen, the light shall be sufficiently divergent so that both eyes of the observer receive rays from all parts of the screen. The maximum illuminator luminance, and divergence and diffusion of light shall be tested to meet the requirements prescribed.
SCOPE
1.1 The function of the illuminator is to provide sufficient illumination and viewing capabilities for the purpose of identification and interpretation of radiographic images. This specification provides the recommended minimum requirements for industrial radiographic illuminators used for viewing industrial radiographic films using transmitted light sources.  
1.2 The illuminator has to ensure the same safety for personnel, or users of any electric apparatus, as specified by electrical standards applicable in the country in which the illuminator is used.  
1.3 Units—The values stated in SI units are to be regarded as standard. The values given in parentheses after SI units are provided for information only and are not considered standard.  
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.  
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

General Information

Status
Published
Publication Date
31-Oct-2021
Current Stage
Ref Project

Buy Standard

Technical specification
ASTM E1390-21 - Standard Specification for Illuminators Used for Viewing Industrial Radiographs
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview
Technical specification
REDLINE ASTM E1390-21 - Standard Specification for Illuminators Used for Viewing Industrial Radiographs
English language
3 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
Designation:E1390 −21
Standard Specification for
1
Illuminators Used for Viewing Industrial Radiographs
This standard is issued under the fixed designation E1390; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope 3.1.1 diffusing screen, n—the screen that scatters the light
such that the light appears the same brightness from any
1.1 The function of the illuminator is to provide sufficient
viewing angle.
illumination and viewing capabilities for the purpose of iden-
3.1.2 viewing screen, n—the port through which the light is
tification and interpretation of radiographic images.This speci-
projected.
fication provides the recommended minimum requirements for
industrial radiographic illuminators used for viewing industrial
3.2 For additional definitions of terms used in this
radiographic films using transmitted light sources.
specification, see Terminology E1316.
1.2 The illuminator has to ensure the same safety for
4. Ordering Information
personnel, or users of any electric apparatus, as specified by
electrical standards applicable in the country in which the
4.1 This specification is intended to be used by the manu-
illuminator is used.
facturers and purchasers of radiographic illuminators.
Requirements, if imposed on manufacturers, should be estab-
1.3 Units—The values stated in SI units are to be regarded
lished by contractual agreement or appropriate purchase docu-
as standard. The values given in parentheses after SI units are
ment.
provided for information only and are not considered standard.
1.4 This standard does not purport to address all of the
5. Materials and Manufacture
safety concerns, if any, associated with its use. It is the
5.1 General—The illuminator shall consist of a housing
responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter- with one or more of the sides containing a viewing screen
illuminated from the inside of the housing. The viewing screen
mine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accor- may also be the diffusing screen. There shall be thermal
protection to prevent overheating, and subsequent damage to
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the the radiographs placed on the viewing screen. The housing or
Development of International Standards, Guides and Recom- system may or may not require ventilation. A rheostat or
mendations issued by the World Trade Organization Technical suitable electrical circuit shall be provided to vary the light
Barriers to Trade (TBT) Committee. intensity.
2. Referenced Documents 6. Physical Properties
2
2.1 ASTM Standards: 6.1 General—The illuminator shall be manufactured of
E1316 Terminology for Nondestructive Examinations materials deemed suitable to withstand the environmental
conditions encountered under normal operating conditions.
3. Terminology
6.2 Viewing Screen—The viewing screen shall be easy to
clean and made of material which is resistant to scratches. The
3.1 Definitions:
size of the screen shall allow the user to view the radiograph
without excessive glare. If the illuminator is to be used for
viewing radiographs of various sizes, masks of various sizes
1
This specification is under the jurisdiction of ASTM Committee E07 on
and configurations should be provided. Alternately, an adjust-
Nondestructive Testing and is the direct responsibility of Subcommittee E07.01 on
Radiology (X and Gamma) Method.
able aperture may be used.
Current edition approved Nov. 1, 2021. Published November 2021. Originally
6.3 Color of Light—The color of the light used to illuminate
approved in 1990. Last previous edition approved in 2016 as E1390 – 12(2016).
DOI: 10.1520/E1390-21.
the radiograph shall be white, that is, color temperature
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
between 5000 and 6250 °K. However, illuminators using
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
non-white or “colored” light may be used if they have been
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website. recommended by the film manufacturers.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 -------------
...

This document is not an ASTM standard and is intended only to provide the user of an ASTM standard an indication of what changes have been made to the previous version. Because
it may not be technically possible to adequately depict all changes accurately, ASTM recommends that users consult prior editions as appropriate. In all cases only the current version
of the standard as published by ASTM is to be considered the official document.
Designation: E1390 − 12 (Reapproved 2016) E1390 − 21
Standard Specification for
1
Illuminators Used for Viewing Industrial Radiographs
This standard is issued under the fixed designation E1390; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 The function of the illuminator is to provide sufficient illumination and viewing capabilities for the purpose of identification
and interpretation of radiographic images. This specification provides the recommended minimum requirements for Industrial
Radiographic Illuminatorsindustrial radiographic illuminators used for viewing industrial radiographic films using transmitted light
sources.
1.2 The illuminator has to ensure the same safety for personnel, or users of any electric apparatus, as specified by electrical
standards applicable in the country in which the illuminator is used.
1.3 Units—The values stated in SI units are to be regarded as standard. The values given in parentheses after SI units are provided
for information only and are not considered standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety safety, health, and healthenvironmental practices and determine the
applicability of regulatory limitations prior to use.
1.4 Values stated in SI units are to be regarded as the standard. The values given in parentheses are provided for information only.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization
established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued
by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
2. Referenced Documents
2
2.1 ASTM Standards:
E1316 Terminology for Nondestructive Examinations
3. Terminology
3.1 Definitions:
3.1.1 diffusing screen—screen, n—the screen that scatters the light such that the light appears the same brightness from any
viewing angle.
1
This specification is under the jurisdiction of ASTM Committee E07 on Nondestructive Testing and is the direct responsibility of Subcommittee E07.01 on Radiology
(X and Gamma) Method.
Current edition approved Dec. 1, 2016Nov. 1, 2021. Published December 2016November 2021. Originally approved in 1990. Last previous edition approved in 20122016
as E1390 - 12.E1390 – 12(2016). DOI: 10.1520/E1390-12R16.10.1520/E1390-21.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM Standards
volume information, refer to the standard’s Document Summary page on the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1

---------------------- Page: 1 ----------------------
E1390 − 21
3.1.2 viewing screen—screen, n—the port through which the light is projected.
3.2 For additional definitions of terms used in this specification, see Terminology E1316.
4. Ordering Information
4.1 This specification is intended to be used by the manufacturers and purchasers of radiographic illuminators. Requirements, if
imposed on manufacturers, should be established by contractual agreement or appropriate purchase document.
5. Materials and Manufacture
5.1 General—The illuminator shall consist of a housing with one or more of the sides containing a viewing screen illuminated
from the inside of the housing. The viewing screen may also be the diffusing screen. There shall be thermal protection to prevent
overheating, and subsequent damage to the radiographs placed on the viewing screen. The housing or system may or may not
require ventilation. A rheostat or suitable electrical circuit shall be provided to vary the light intensity.
6. Physical Properties
6.1 General—The illuminator shall be manufactured of materials deemed suitable to withstand the environmental conditions
encountered under normal operating conditions.
6.2 Viewing Screen—The viewing screen shall be easy to clean and made of material which is resistant to scratches. The size of
the screen shall allow the user to view the radiograph without excessive glare. If the illuminator is to be used for viewing
...