Standard Test Method for Density and Relative Density of Crude Oils by Digital Density Analyzer

SCOPE
1.1 This test method covers the determination of the density or relative density of crude oils that can be handled in a normal fashion as liquids at test temperatures between 15 and 35°C. This test method applies to crude oils with high vapor pressures provided appropriate precautions are taken to prevent vapor loss during transfer of the sample to the density analyzer.
1.2 This test method was evaluated in round robin testing using crude oils in the 0.75 to 0.95 g/mL range. Lighter crude oil can require special handling to prevent vapor losses. Heavier crudes can require measurements at higher temperatures to eliminate air bubbles in the sample.
1.3 The accepted units of measurement of density are grams per millilitre and kilograms per cubic metre.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific hazard statements are given in Notes 1, 2, and 3.

General Information

Status
Historical
Publication Date
09-Nov-1999
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM D5002-99 - Standard Test Method for Density and Relative Density of Crude Oils by Digital Density Analyzer
English language
5 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
An American National Standard
Designation:D5002–99
Standard Test Method for
Density and Relative Density of Crude Oils by Digital
Density Analyzer
This standard is issued under the fixed designation D 5002; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope D 4057 Practice for Manual Sampling of Petroleum and
Petroleum Products
1.1 This test method covers the determination of the density
D 4177 Practice for Automatic Sampling of Petroleum and
or relative density of crude oils that can be handled in a normal
Petroleum Products
fashion as liquids at test temperatures between 15 and 35°C.
D 4377 Test Method for Water in Crude Oils by Volumetric
Thistestmethodappliestocrudeoilswithhighvaporpressures
Karl Fischer Titration
provided appropriate precautions are taken to prevent vapor
loss during transfer of the sample to the density analyzer.
3. Terminology
1.2 This test method was evaluated in round robin testing
3.1 Definitions:
using crude oils in the 0.75 to 0.95 g/mL range. Lighter crude
3.1.1 density—mass per unit volume at a specified tempera-
oil can require special handling to prevent vapor losses.
ture.
Heavier crudes can require measurements at higher tempera-
3.1.2 relative density—the ratio of the density of a material
tures to eliminate air bubbles in the sample.
at a stated temperature to the density of water at a stated
1.3 The accepted units of measurement of density are grams
temperature.
per millilitre and kilograms per cubic metre.
1.4 This standard does not purport to address all of the
4. Summary of Test Method
safety concerns, if any, associated with its use. It is the
4.1 Approximately 0.7 mLof crude oil sample is introduced
responsibility of the user of this standard to establish appro-
into an oscillating sample tube and the change in oscillating
priate safety and health practices and determine the applica-
frequency caused by the change in the mass of the tube is used
bility of regulatory limitations prior to use. Specific hazard
in conjunction with calibration data to determine the density of
statements are given in 7.4, 7.5, and 7.6.
the sample.
2. Referenced Documents
5. Significance and Use
2.1 ASTM Standards:
5.1 Density is a fundamental physical property that can be
D 941 Test Method for Density and Relative Density (Spe-
used in conjunction with other properties to characterize the
cific Gravity) of Liquids by Lipkin Bicapillary Pycnom-
quality of crude oils.
eter
3 5.2 The density or relative density of crude oils is used for
D 1193 Specification for Reagent Water
theconversionofmeasuredvolumestovolumesatthestandard
D 1217 Test Method for Density and Relative Density
2 temperatures of 15°C or 60°F and for the conversion of crude
(Specific Gravity) of Liquids by Bingham Pycnometer
mass measurements into volume units.
D 1250 Guide for Petroleum Measurement Tables
5.3 The application of the density result obtained from this
D 4052 Test Method for Density and Relative Density of
4 test method, for fiscal or custody transfer accounting calcula-
Liquids by Digital Density Meter
tions, can require measurements of the water and sediment
contents obtained on similar specimens of the crude oil parcel.
This test method is under the jurisdiction of ASTM Committee D-2 on
6. Apparatus
Petroleum Products and Lubricants and is the direct responsibility of Subcommittee
6.1 Digital Density Analyzer—Adigital analyzer consisting
D02.04.OD on Physical Methods.
Current edition approved Nov. 10, 1999. Published December 1999. Originally
of a U-shaped, oscillating sample tube and a system for
published as D 5002 – 89. Last previous edition D 5002 – 94.
electronic excitation, frequency counting, and display. The
Annual Book of ASTM Standards, Vol 05.01.
analyzer must accommodate the accurate measurement of the
Annual Book of ASTM Standards, Vol 11.01.
Annual Book of ASTM Standards, Vol 05.02. sample temperature during measurement or must control the
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
D5002–99
sample temperature as described in 6.2 and 6.5.The instrument laboratory test container and sample volume shall be of
shall be capable of meeting the precision requirements de- sufficient dimensions to allow mixing as described in 8.3.1.
scribed in Test Method D 4052. Mixing is required to obtain a homogeneous sample for
6.2 Circulating Constant-Temperature Bath, capable of analysis.
maintaining the temperature of the circulating liquid constant
8.2 Laboratory Sample—Use only representative samples
to 60.05°C in the desired range. Temperature control can be obtainedasspecifiedinPracticesD 4057orD 4177forthistest
maintained as part of the density analyzer instrument package. method.
6.3 Syringes, at least 2 mL in volume with a tip or an 8.3 Test Specimen—The aliquot of sample obtained from
adapter tip that will fit the inlet of the density analyzer.
the laboratory sample and delivered to the density analyzer
6.4 Flow-Through or Pressure Adapter, for use as an sample tube. The test specimen is obtained as follows:
alternative means of introducing the sample into the density
8.3.1 Mix the sample of crude oil to homogenize any
meter. sediment and water present. The mixing may be accomplished
6.5 Thermometer, calibrated and graduated to 0.1°C, and a
as described in Practice D 4177 or Test Method D 4377.
thermometer holder that can be attached to the instrument for Mixing at room temperature in an open container can result in
setting and observing the test temperature. In calibrating the
the loss of light ends, so mixing in closed, pressurized
thermometer, the ice point and bore corrections should be
containers or at sub-ambient temperatures is recommended.
estimated to the nearest 0.05°C. Precise setting and control of
8.3.2 Draw the test specimen from a properly mixed labo-
the test temperature in the sample tube is extremely important.
ratorysampleusinganappropriatesyringe.Alternatively,ifthe
Anerrorof0.1°Ccanresultinachangeindensityofoneinthe
proper density analyzer attachments and connecting tubes are
fourth significant figure.
used then the test specimen can be delivered directly to the
analyzer’s sample tube from the mixing container.
7. Reagents and Materials
7.1 Purity of Reagents—Reagent grade chemicals shall be 9. Preparation of Apparatus
used in all tests. Unless otherwise indicated it is intended that
9.1 Set up the density analyzer and constant temperature
all reagents shall conform to the specifications of the Commit-
bath following the manufacturer’s instructions.Adjust the bath
tee onAnalytical Reagents of theAmerican Chemical Society,
or internal temperature control so that the desired test tempera-
where such specifications are available. Other grades may be
ture is established and maintained in the sample compartment
used, provided it is first ascertained that the reagent is of
of the analyzer. Calibrate the instrument at the same tempera-
sufficiently high purity to permit its use without lessening the
ture at which the density of the sample is to be measured.
accuracy of the determination.
7.2 Purity of Water—Unless otherwise indicated, references
10. Calibration of Apparatus
to water shall be understood to mean reagent water as defined
10.1 Calibrate the instrument when first setting up and
by Type II of Specification D 1193.
whenever the test temperature is changed. Thereafter, conduct
7.3 Water, redistilled, freshly boiled and cooled reagent
calibration checks at least weekly during routine operation or
water for use as a primary calibration standard.
more frequently as may be dictated by the nature of the crude
7.4 Acetone, for flushing and drying the sample tube.
oils being measured, (see 10.3).
(Warning: Extremely flammable.)
6 10.2 Initial calibration, or calibration after a change in test
7.5 Petroleum Naphtha, for flushing viscous petroleum
temperature, necessitates calculation of the values of the
samples from the sample tube. (Warning: Extremely flam-
Constants A and B from the periods of oscillation, (T),
mable.)
observed when the sample cell contains certified reference
7.6 n-Nonane, n-tridecane or cyclohexane, 99 % purity or
liquids such as air and double-distilled boiled water. Other
better, or similar pure material for which the density is known
calibrating materials such as n-nonane, n-tridecane, cyclohex-
precisely from literature references or by direct determination
ane, and n-hexadecane (for high temperature applications) can
in accordance with Test Method D 941 or D 1217.(Warning:
also be used as appropriate.
Extremely flammable.)
10.2.1 While monitoring the oscillator period, T, flush the
sample tube with petroleum naphtha, followed with an acetone
8. Sampling, Test Specimens, and Test Units
flush and dry with dry air. Continue drying until the display
8.1 Sampling is defined as all the steps required to obtain an
exhibits a steady reading. In cases where saline components
aliquot of the contents of any pipe, tank or other system, and to
can be deposited in the cell, flush with distilled water followed
place the sample into the laboratory test container. The
by acetone and dry air. Contaminated or humid air can affect
the calibration. When these conditions exist in the laboratory,
pass the air used for calibration through a suitable purification
Reagent Chemicals, American Chemical Society Specifications, American
and drying train. In addition, the inlet and outlet ports for the
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
U-tube must be plugged during measurement of the calibration
listed by the American Chemical Society, see Analar Standards for Laboratory
Chemicals, BDH Ltd., Poole, Dorset, U.K., and the United States Pharmacopeia
air to prevent ingress of moist air.
and National Formulary, U.S. Pharmacopeial Convention, Inc. (USPC), Rockville,
10.2.2 Allow the dry air in the U-tube to come to thermal
MD.
equilibrium with the test temperature and record the T-value
Suitable solvent naphthas are marketed under various designations such as
“petroleum ether,” “ligroine,” or “precipitation naphtha.” for air.
D5002–99
10.2.3 Introduce about 0.7 mL of freshly boiled and cooled 10.2.11 To calibrate the instrument to determine relative
double-distilled water into the sample tube from the bottom density,thatis,thedensityofthesampleatagiventemperature
opening using a suitable syringe. The water must be free of referred to the density of water at the same temperature, follow
even the smallest air or gas bubbles. The sample tube shall be 10.2.1-10.2.9, but substitute 1.000 for d in performing the
w
completely full. Allow the water to reach thermal equilibrium calculations described in 10.2.8.
at the test temperature and record the T-value for water and the 10.3 Since some crude oils can be difficult to remove from
test temperature. the sample tube, frequent calibration checks are recommended.
10.2.4 Alternatively introduce one of the hydrocarbon cali- These checks and any subsequent adjustments to Constants A
bration standards and measure the T-value as in 10.2.3. and B can be made if required, without repeating the calcula-
10.2.5 Calculate the density of air at the temperature of test tion procedure.
using the following equation:
NOTE 1—The need for a change in calibration is generally attributable
d 5 0.001293[273/T]@P/760] g/mL (1) to deposits in the sample tube that are not removed by the routine flushing
a
procedure.Although this condition can be compensated for by adjustingA
where:
and B, as described below, it is good practice to clean the tube with warm
T = temperature, K, and chromic acid solution (Warning—Causes severe burns. A recognized
P = barometric pressure, torr. carcinogen.) whenever a major adjustment is required. Chromic acid
solution is the most effective cleaning agent; however, surfactant-type
10.2.6 Determine the density of water at the temperature of
cleaning fluids have also been used successfully.
test by reference to Table 1.
10.2.7 Alternatively record the density at the test tempera- 10.3.1 Flush and dry the sample tube as described in 10.2.1
ture for the hydrocarbon calibrant used in 10.2.4 as obtained and allow the display to reach a steady reading. If the display
from an appropriate reference source or from direct determi- does not exhibit the correct T-value or density for air at the
nation (see 7.6). temperature of test, repeat the cleaning procedure or adjust the
10.2.8 Usingtheobserved T-valuesandthereferencevalues value of Constant B commencing with the last decimal place
for water and air, calculate the values of the ConstantsAand B
until the correct density is displayed.
using the following equations: 10.3.2 If adjustment to Constant B was necessary in 10.3.1
then continue the recalibration by introducing freshly boiled
2 2
A 5 [T 2 T #/[d 2 d # (2)
w a w a
and cooled double-distilled water into the sample tube as
B 5 T 2 ~A 3 d ! (3)
a a described in 10.2.3 and allowing the display to reach a steady
reading. If the instrument has been calibrated to display the
where:
density, adjust the reading to the correct value for water at the
T = observed period of oscillation for cell containing
w
test temperature, (see Table 1) by changing the value of
water,
Constant A, commencing with the last decimal place. If the
T = observed period of oscillation for cell containing air,
a
instrument has been calibrated to display the relative density,
d = density of water at test temperature, and
w
adjust the reading to the value 1.0000.
d = density of air at test temperature.
a
Alternatively, use the T and d values for the other reference
NOTE 2—In applying this periodic calibration procedure, it has been
liquid if one is used.
found that more than one value each for A and B, differing in the fourth
10.2.9 If the instrument is equipped to calculate density decimal place, will yield the correct reading for the density of air and
water. The setting chosen would then be dependent upon whether it was
from the ConstantsAand B and the observed T-value from the
approached from a higher or lower value. The setting selected by this
sample, then enter the constants in the instrument memory in
method could have the effect of altering the fourth place of the reading
accordance with the manufacturer’s instructions.
obtained for a sample.
10.2.10 Check the calibration and adjust if needed by
10.4 Some analyzer models are designed to displ
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.