ASTM E969-02(2012)
(Specification)Standard Specification for Glass Volumetric (Transfer) Pipets
Standard Specification for Glass Volumetric (Transfer) Pipets
ABSTRACT
This specification covers volumetric pipets of two classes. Class A, Precision Pipet and Class B, General Purpose. Borosilicate glass for pipets shall conform to the glass requirements specified. Pipets shall be calibrated to deliver the intended capacity. The pipets shall consist in general of a suction tube and a delivery tube separated by a bulb; all three parts shall be permanently attached together. All markings shall be permanent and legible.
SCOPE
1.1 This specification covers volumetric pipets of two classes. Class A, Precision Pipet and Class B, General Purpose. Note 1—Specifications for micropipets are given in Specification E193.
1.2 Product with a stated capacity not listed in this standard may be specified class A tolerance when product conforms to the tolerance range of the next smaller volumetric standard product listed in Table 1.TABLE 1 Requirements for Volumetric (Transfer) Pipet
Nominal
Capacity,
mL
Capacity Tolerance,
mL
Length of Delivery Tube, mm
Inside Diameter at
Capacity Mark, mm
Maximum Distance Between Bulb and Graduation Mark, mm
Minimum Outflow
Time, sA
Color-Coding
Band
Class A
Class B
min
max
min
max
Class A
Class B
0.5
±0.006
±0.012
100
140
1.3
4
60
5
3
black (2)
1
±0.006
±0.012
100
140
2
4
60
8
3
blue
2
±0.006
±0.012
120
150
2
4
60
8
3
orange
3
±0.01
±0.02
130
170
2
4
60
8
5
black
4
±0.01
±0.02
130
170
2
4
60
8
5
red (2)
5
±0.01
±0.02
130
170
2
4
60
8
8
white
6
±0.01
±0.03
150
190
2
4
60
10
8
orange (2)
7
±0.01
±0.03
150
190
2
4
60
10
8
green (2)
8
±0.02
±0.04
150
190
2
4
60
10
8
blue
9
±0.02
±0.04
150
190
2
4
60
10
8
black
10
±0.02
±0.04
150
190
2
4
70
15
8
red
15
±0.03
±0.06
180
220
3.5
5.5
70
25
10
green
20
±0.03
±0.06
230
260
3.5
5.5
70
25
10
yellow
25
±0.03
±0.06
230
260
3.5
5.5
70
25
15
blue
30
±0.03
...
230
260
3.5
5.5
70
25
...
black
40
±0.05
...
230
260
4
6
70
25
...
white
50
±0.05
±0.10
230
260
4
6
70
25
15
red
100
±0.08
±0.16
230
260
5
7
80
30
20
yellow A Maximum outflow time for A or B shall be 60 s.
General Information
Relations
Standards Content (Sample)
NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:E969 −02 (Reapproved 2012)
Standard Specification for
Glass Volumetric (Transfer) Pipets
This standard is issued under the fixed designation E969; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope the vessel and no after-drainage period is allowed. Accuracy
shall be within the limits specified in Table 1.
1.1 This specification covers volumetric pipets of two
classes. ClassA, Precision Pipet and Class B, General Purpose.
4. Design
NOTE 1—Specifications for micropipets are given in Specification
4.1 Shape—The pipets shall consist in general of a suction
E193.
tube and a delivery tube separated by a bulb; all three parts
1.2 Product with a stated capacity not listed in this standard
shall be permanently attached together. Any cross-section of
may be specified class A tolerance when product conforms to
the pipet taken in a plane perpendicular to the longitudinal axis
the tolerance range of the next smaller volumetric standard
shall be circular. The shape shall permit complete emptying
product listed in Table 1.
and thorough cleaning.
4.1.1 Bulb—The shape shall permit complete emptying
2. Referenced Documents
without any hold up, and easy cleaning.
2.1 ASTM Standards:
4.2 Dimensions—Thelengthofthesuctiontubeshallbe150
E193 Specification for Laboratory Glass Micropipets
to190mmandtheminimumwallthicknessofbothsuctionand
E438 Specification for Glasses in Laboratory Apparatus
delivery tubes shall be 0.90 mm. Pipets must comply with the
E920 Specification for Commercially Packaged Laboratory
essential dimensions given in Table 1.
Apparatus
4.3 Delivery Tips—Delivery tips shall be made with a
E921 Specification for Export Packaged Laboratory Appa-
gradual taper of 1.5 to 3 cm. The end of the tip shall be
ratus
E1133 Practice for Performance Testing of Packaged Labo- perpendicular to the longitudinal axis of the tip. The outside
edge of the tip may be bevelled slightly and the end and the
ratory Apparatus for United States Government Procure-
ments bevel shall be ground or fire-polished. Sudden constriction at
the orifice would impair smooth flow characteristics of the
E1157 Specification for Sampling and Testing of Reusable
Laboratory Glassware delivery stream and is not acceptable.
4.3.1 Tempered Tips—May be supplied at option of manu-
3. General Requirements
facturer. The tempered tip, when tested in index oil which
matches the refractive index of the glass being used, shall have
3.1 Borosilicate Glass—Borosilicate glass for pipets shall
a temper between 75 and 220 nm (millimicrons).
conform to the glass requirements of Type 1, Class A or B of
Specification E438.
4.4 Markings—All markings shall be permanent and leg-
ible.
3.2 Calibration—Pipets shall be calibrated to deliver (sym-
4.4.1 Capacity Line—The capacity line shall be a sharply
bol TD) the intended capacity at 20°C. The pipet shall be filled
defined line of uniform width (maximum 0.6 mm) in a plane
about 20 mm above the capacity line. The water is lowered
perpendicular to the vertical axis of the pipet. The line shall be
slowly to the capacity line. Delivery of the contents into a
appliedbyoneofthefollowingmethods:etchedand filled with
receiving vessel is made with the tip in contact with the wall of
a permanent pigment; etched through a vertical colored stripe
fusedintotheglass;byapplicationofastainfiredintotheglass
This specification is under the jurisdiction of ASTM Committee E41 on
without etching; or by application of an enamel fired onto the
Laboratory Apparatus and is the direct responsibility of Subcommittee E41.01 on
glass without etching. The line shall completely encircle the
Apparatus.
tube except when the line is a stain fired into the glass without
Current edition approved Nov. 1, 2012. Published November 2012. Originally
etching. If so stained, the line shall extend a minimum of 90 %
a
...