Specification for Bores and Cones for Modular Femoral Heads (Withdrawn 2001)

SCOPE
1.1 This specification covers the functional dimensions and tolerances for tapered cones of proximal femoral stems and the bores of mating ceramic and metal heads.  
1.2 This specification does not cover the dimensions of proximal femoral stems other than the cone portion, nor does it cover the dimensions of the ceramic/metal femoral heads, other than the inside tapers.  
1.3 The dimensions stated in SI units are to be regarded as the standard.  
1.4 The purpose of this specification is to codify matching bore and cone combinations for easy identification. Nonetheless, components from different manufacturers should not be mixed because the combinations of tolerances within the family, surface finish, and configuration from the different manufacturers may not have been validated through axial load strengths or fatigue strength.  
1.5 The codification described in this specification represents the most current state of the art at the time of balloting. Future additions can be made but because of the time required to revise the specification, it is recommended that the manufacturer of new combinations label their components with at least the angles until such time that new code letters can be established in accordance with future revisions of this specification.

General Information

Status
Historical
Publication Date
31-Dec-1994
Current Stage
Ref Project

Relations

Buy Standard

Technical specification
ASTM F1636-95e1 - Specification for Bores and Cones for Modular Femoral Heads (Withdrawn 2001)
English language
4 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)


NOTICE: This standard has either been superseded and replaced by a new version or discontinued.
Contact ASTM International (www.astm.org) for the latest information.
e1
Designation: F 1636 – 95
Standard Specification for
Bores and Cones for Modular Femoral Heads
This standard is issued under the fixed designation F 1636; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
e NOTE—Editorial changes were made throughout in December 1996.
1. Scope Surgical Implants (Special Quality)
F 562 Specification for Wrought Cobalt-Nickel-Chromium-
1.1 This specification covers the functional dimensions and
Molybdenum Alloy for Surgical Implant Applications
tolerances for tapered cones of proximal femoral stems and the
F 563 Specification for Wrought Cobalt-Nickel-Chromium-
bores of mating ceramic and metal heads.
Molybdenum-Tungsten-Iron Alloy for Surgical Implant
1.2 This specification does not cover the dimensions of
Applications
proximal femoral stems other than the cone portion, nor does it
F 603 Specification for High-Purity Dense Aluminum Ox-
cover the dimensions of the ceramic/metal femoral heads, other
ide for Surgical Implant Application
than the inside tapers.
F 620 Specification for Titanium 6Al, 4V ELI Alloy Forg-
1.3 The dimensions stated in SI units are to be regarded as
ings for Surgical Implants
the standard.
F 621 Specification for Stainless Steel Forgings for Surgical
1.4 The purpose of this specification is to codify matching
Implants
bore and cone combinations for easy identification. Nonethe-
F 799 Specification for Thermomechanically Processed
less, components from different manufacturers should not be
Cobalt-Chromium-Molybdenum Alloy Forgings for Surgi-
mixed because the combinations of tolerances within the
cal Applications
family, surface finish, and configuration from the different
F 961 Specification for Cobalt-Nickel-Chromium-
manufacturers may not have been validated through axial load
Molybdenum Alloy Forgings for Surgical Implant Appli-
strengths or fatigue strength. Information supplied by the
cations
manufacturer with the implants should include a precaution
F 1108 Specification for Ti-6Al, 4V Alloy Castings for
statement that informs the user (surgeon) that bores and cones
Surgical Implants
with the same code letter from different manufacturers should
not be used without first consulting the supplier of the femoral
3. Terminology
components.
3.1 Definitions:
1.5 The codification described in this specification repre-
3.1.1 nominal neck length—the distance, K, parallel to the
sents the most current state of the art at the time of balloting.
taper axis, between the theoretical gage point and the horizon-
Future additions can be made but because of the time required
tal plane through the center of the head; this distance does not
to revise the specification, it is recommended that the manu-
include standard variants used to adjust for resection level, leg
facturer of new combinations label their components with at
length, and so forth. The K dimension may be plus or minus
least the angles until such time that new code letters can be
(proximal or distal to the location of the theoretical gage Point
established in accordance with future revisions of this specifi-
G).
cation.
3.1.2 residual or free length—the distance, A, parallel to the
2. Referenced Documents cone axis, between the end of the cone and the theoretical gage
point.
2.1 ASTM Standards:
3.1.3 theoretical gage point—the diameter, G, at nominal
F 75 Specification for Cast-Cobalt-ChromiumMolybdenum
tolerance, where the stem cone (external taper) theoretically
Alloy for Surgical Implant Applications
engages the femoral head bore (internal taper).
F 136 Specification for Wrought Titanium 6A1, 4V ELI
Alloy for Surgical Implant Applications
4. Materials
F 138 Specification for Stainless Steel Bar and Wire for
4.1 The cones may be made from any of the implantable
grade materials used for femoral stems including: Specifica-
This specification is under the jurisdiction of ASTM Committee F-4 on Medical
tions F 75, F 136, F 138, F 562, F 563, F 620, F 621, F 799,
and Surgical Materials and Devices and is under the direct responsibility of
F 961, and F 1108.
Subcommittee F04.22 on Arthroplasty.
4.2 The ceramic heads may be made from Specification
Current edition approved Dec. 10, 1995. Published June 1996.
F 603 or a material determined to be substantially equivalent or
Annual Book of ASTM Standards, Vol 13.01.
Copyright © ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.
F 1636
TABLE 1 Cones
superior in mechanical strength, wear properties, and biocom-
patibility. The metal heads may be made from any of the G—Diameter, A—Dimension,
a° Size
mm mm
recognized materials used for femoral stems including:
...