Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)

SCOPE
1.1 This test method is used to evaluate the resistance of medical face masks to penetration by the impact of a small volume (~2 mL) of a high velocity stream of synthetic blood. Medical face mask pass/fail determinations are based on visual detection of synthetic blood penetration.
1.2 This test method does not apply to all forms or conditions of blood-borne pathogen exposure. Users of the test method must review modes for face exposure and assess the appropriateness of this test method for their specific application.
1.3 This test method primarily addresses the performance of materials or certain material constructions used in medical face masks. This test method does not address the performance of the medical face masks design, construction, or interfaces or other factors with the potential to affect the overall protection offered by the medical face mask and its operation (such as filtration efficiency and pressure drop). Procedures for measuring these properties are contained in Test Methods F 2101 and MIL-M-36954C.
1.4 This test method does not address breathability of the medical face mask materials or any other properties affecting the ease of breathing through the medical face mask. This test method evaluates medical face masks as an item of protective clothing. This test method does not evaluate the performance of medical face masks for airborne exposure pathways or in the prevention of the penetration of aerosolized body fluids deposited on the medical face mask.
1.5 The values stated in SI units or inch-pound units are to be regarded separately as standard. The pressure values stated in each system are not exact equivalents. However, as the corresponding velocities are within 1 % of each other, (see ), reporting of the results in either units is permitted.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

General Information

Status
Historical
Publication Date
14-Sep-2005
Current Stage
Ref Project

Relations

Buy Standard

Standard
ASTM F1862-05 - Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
English language
18 pages
sale 15% off
Preview
sale 15% off
Preview

Standards Content (Sample)

NOTICE: This standard has either been superseded and replaced by a new version or withdrawn.
Contact ASTM International (www.astm.org) for the latest information
Designation:F 1862–05
Standard Test Method for
Resistance of Medical Face Masks to Penetration by
Synthetic Blood (Horizontal Projection of Fixed Volume at a
1
Known Velocity)
This standard is issued under the fixed designation F 1862; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
Workers, primarily those in the health care profession, involved in treating and caring for
individuals injured or sick, can be exposed to biological liquids capable of transmitting disease.These
diseases, which may be caused by a variety of microorganisms, can pose significant risks to life and
health. This is especially true of blood-borne viruses which cause Hepatitis (Hepatitis B Virus (HBV)
and Hepatitis C Virus (HCV)) and Acquired Immune Deficiency Syndrome (AIDS) [Human
Immunodeficiency Virus (HIV)]. Because engineering controls can not eliminate all possible
exposures, attention is placed on reducing the potential of direct skin and mucous membrane contact
through the use of protective clothing that resists penetration (29 CFR Part 1910.1030). This test
method was developed for ranking the synthetic blood penetration resistance performance of medical
ace masks in a manner representing actual use as might occur when the face mask is contacted by a
high velocity stream of blood from a punctured wound.
1. Scope the ease of breathing through the medical face mask. This test
method evaluates medical face masks as an item of protective
1.1 This test method is used to evaluate the resistance of
clothing. This test method does not evaluate the performance
medical face masks to penetration by the impact of a small
ofmedicalfacemasksforairborneexposurepathwaysorinthe
volume (~2 mL) of a high velocity stream of synthetic blood.
prevention of the penetration of aerosolized body fluids depos-
Medicalfacemask pass/faildeterminationsarebasedonvisual
ited on the medical face mask.
detection of synthetic blood penetration.
1.5 The values stated in SI units or inch-pound units are to
1.2 This test method does not apply to all forms or condi-
be regarded separately as standard. The pressure values stated
tions of blood-borne pathogen exposure. Users of the test
in each system are not exact equivalents. However, as the
method must review modes for face exposure and assess the
corresponding velocities are within 1 % of each other, (see
appropriateness of this test method for their specific applica-
X1.4.2), reporting of the results in either units is permitted.
tion.
1.6 This standard does not purport to address all of the
1.3 Thistestmethodprimarilyaddressestheperformanceof
safety concerns, if any, associated with its use. It is the
materialsorcertainmaterialconstructionsusedinmedicalface
responsibility of the user of this standard to establish appro-
masks. This test method does not address the performance of
priate safety and health practices and determine the applica-
the medical face mask’s design, construction, or interfaces or
bility of regulatory limitations prior to use.
other factors with the potential to affect the overall protection
offered by the medical face mask and its operation (such as
2. Referenced Documents
filtration efficiency and pressure drop). Procedures for measur-
2
2.1 ASTM Standards:
ing these properties are contained in Test Methods F 2101 and
D 891 Test Methods for Specific Gravity, Apparent, of
MIL-M-36954C.
Liquid Industrial Chemicals
1.4 This test method does not address breathability of the
D 1331 Test Methods for Surface and Interfacial Tension in
medical face mask materials or any other properties affecting
Solutions of Surface-Active Agents
1
This test method is under the jurisdiction of ASTM Committee F23 on
Protective Clothing and is the direct responsibility of Subcommittee F23.40 on 2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
Biological Hazards.
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Current edition approved Sept. 15, 2005. Published December 2005. Originally
Standards volume information, refer to the standard’s Document Summary page on
approved in 1998. Last previous edition approved in 2000 as F 1862 - 00a.
the ASTM website.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1

---------------------- Page: 1 ----------------------
F 1862–05
E 105 Practice for Probability Sampling of Materials mucous membrane areas of the wearer’s nose and mouth, from
E 171 Specification for Standard Atmospheres for Condi- contact with blood and other body fluids during
...