Abstract

1.1   This document specifies the requirements and the relevant tests for automatically controlled sterilizers employing ethylene oxide (EO) gas as the sterilant, either as a pure gas or a mixture with other gases (for example carbon dioxide or nitrogen) being used for the sterilization of medical devices and their accessories.
1.2   This document specifies requirements for ethylene oxide sterilizers (EO sterilizers) for:
—   the performance and design of sterilizers intended to deliver a process capable of sterilizing medical devices;
—   the equipment and controls of these sterilizers, needed for operation, control and monitoring of the sterilization processes;
—   the test equipment and test procedures used to verify the sterilizer performance specified by this document.
1.3   This document does not cover sterilizers which employ the injection of EO or mixtures containing EO directly into packages or into a flexible chamber.

Status
Not Published
Publication Date
22-Dec-2026
Current Stage
5060 - Closure of Vote - Formal Approval
Start Date
22-Oct-2026
Completion Date
22-Oct-2026

Buy Documents

EN 1422:2026 is a CEN European Standard for automatically controlled ethylene oxide (EO) sterilizers used to sterilize medical devices and their accessories. It specifies the design, control, performance tests and documentation for EO sterilizers that use EO as a pure gas or as a mixture with other gases. The standard is for manufacturers, test laboratories, installers, users and buyers who need to verify that an EO sterilizer can deliver a controlled low-temperature sterilization process.

What does EN 1422:2026 specify?

EN 1422:2026 specifies requirements and relevant tests for EO sterilizers used for medical devices and their accessories. It covers the sterilizer itself - its design, controls, monitoring, recording and test methods - rather than validation and routine control of the EO sterilization process, which is addressed in EN ISO 11135:2014.

The scope also excludes sterilizers that inject EO or EO mixtures directly into packages or into a flexible chamber. In practice, this means the standard is aimed at chamber-based EO sterilizers with automatic control and defined test methods for performance verification.

The document is organized into Clauses 1 to 12, followed by Annexes A to ZA.

AnnexWhat it covers
Annex ATest programme for type tests, works tests, acceptance test upon installation and IQ provisions
Annex BLeak test cycle
Annex CSterilizer chamber profile testing
Annex DMicrobiological performance test for EO sterilizers
Annex EEnvironmental aspects
Annex FIllustrations of the interrelationship between control and recording
Annex GAdditional information on protective measures
Annex HTest instrumentation
Annex IExample stages used in an EO sterilization cycle and guidance on their purpose
Annex ZARelationship to Regulation (EU) 2017/745

What are the key requirements of EN 1422:2026?

EN 1422:2026 sets out a complete framework for safe design, automatic control, performance verification and user information for EO sterilizers. The main practical message is that the sterilizer must be designed, tested and documented as a controlled system, not just as a pressure vessel with gas supply.

Safety, risk and usability

Clause 5 begins with risk assessment, risk control and usability. EN 1422:2026 requires the design to be based on risk assessment, with measures against unauthorized access and with usability engineering applied to improve safe operation. That matters because EO handling, door operation, maintenance access and software behavior all affect operator safety.

Clause 5.1.2 links the sterilizer to external protective measures, including electrical, mechanical, fire, burn and software-related hazards. Clause 5.1.3 then requires risk analysis for the specific EO sterilizer design and its features. In practice, a manufacturer has to show that hazards have been identified and reduced before the equipment is released.

Chamber, doors and process hardware

Clause 5.2 covers the sterilizer chamber, including dimensions, doors, interlocks, chamber integrity, pressure vessels and chamber temperature control. The chamber must have one or two doors, the doors must not open during an operating cycle, and double-ended sterilizers need clear door-status indication on both sides. This is important because the load must stay contained during processing and the operator must know when access is safe.

Clause 5.3 covers pipework, vaporizers, the evacuation system, framework and panelling, and loading equipment. EO-carrying pipework must be designed to avoid condensate build-up, vaporizers must be protected against excessive temperature and pressure, and the vacuum system must remove air, EO gas, inert gas and steam. In practice, these requirements are about reliable gas delivery, effective removal and avoiding leaks or trapped condensate that could affect the cycle.

Clause 5.5 requires test connections that allow access for test equipment without harming chamber integrity. This matters because chamber testing, profile measurement and installation checks depend on repeatable access points that do not compromise the sterilizer.

Control, monitoring and records

Clause 6 requires an automatic controller with pre-set programmes, control and monitoring functions, failure handling and status indication. The sterilizer must control and monitor key cycle variables, detect failures, prevent unsafe states and prevent unauthorized changes to programmed parameters during a cycle. For users, this means the cycle should be repeatable and faults should be visible rather than hidden.

Clauses 6.5 to 6.7 define the instrumentation and records for temperature, pressure, time, EO concentration and relative humidity (RH) where used. The sterilizer must indicate and record the relevant cycle parameters, and the records must allow later evaluation. In practice, this is what supports batch documentation, troubleshooting and review of whether the sterilizer actually ran as intended.

Clause 6.4 sets out how the controller responds to faults, power failure and other failures. If a fault occurs, the load is treated as non-sterile until the defined recovery or release conditions are met. This is a critical practical rule for release decisions and for avoiding accidental use of an unprocessed load.

Services and emissions

Clause 7 covers the services and local environment needed for operation, including the sterilant supply, electrical supply, water, steam, vacuum, drains, lighting, compressed air, air or inert gases and ventilation. The user must be informed about site requirements, and the sterilizer must operate within the specified service conditions. This matters during installation because an EO sterilizer can only perform as intended if the utility quality and supply are right.

Clause 7.2 requires controlled EO supply, while Clause 7.10 deals with air and inert gases and the prevention of microbial recontamination. Clause 8 then covers electromagnetic emissions, noise, exhaust emissions and heat emission. In practice, these clauses address the impact of the sterilizer on nearby equipment, operators and the installation environment.

Performance tests and cycle stages

Clause 10 defines the performance checks that demonstrate the sterilizer can do its job. These include chamber integrity, attainment of physical conditions, microbiological performance, EO removal, aeration, load dryness and loading configuration. Annexes B, C and D provide the corresponding methods. This is the core section for type testing, works testing and verification of claimed cycle performance.

Annex I shows the stages that may appear in an EO sterilization cycle, such as air removal, conditioning, EO exposure, holding, EO removal, desorption, air admission, cycle complete and aeration. The annex is informative, but it is useful for understanding how the cycle is structured and what each stage is meant to achieve.

Information, marking and packaging

Clause 11 requires pre-purchase information, marking and labelling, instructions for use and disposal information. Clause 12 covers packaging for transport, storage and installation. In practice, the standard makes the supplier responsible for telling the buyer what the sterilizer is intended to do, what services it needs, how it should be used and how it should be maintained.

What terms does EN 1422:2026 define?

  • Access device - A key, code or tool used to enter restricted parts of the equipment.
  • Aeration - The stage in which EO or its reaction products, or both, are removed from the product until the required level is reached.
  • Automatic controller - The device that drives the equipment through the cycle stages in the programmed order.
  • Biological indicator (BI) - A test system containing viable microorganisms used to challenge the sterilization process.
  • Calibration - The operation that relates a measurement standard to an indication so a measurement result can be obtained from it.
  • Conditioning - The pre-exposure treatment used to bring the load to the required temperature, RH or other process condition.
  • Cycle parameter - A cycle value, including its tolerance, used for control, monitoring, indication and recording.
  • Fault - A condition in which one or more process or cycle parameters are outside the specified limits and need special handling by the controller.

Who uses EN 1422:2026?

EN 1422:2026 is used by manufacturers of EO sterilizers, test laboratories, installers, purchasers, maintenance teams and quality managers. It is also relevant to users operating EO sterilizers in health care facilities or industrial settings.

The introduction says EO sterilizers can be used for medical, dental, pharmaceutical, veterinary, industrial or related purposes. In practice, people use the standard to design equipment, plan type and works tests, check installation readiness, review technical documentation and confirm that the sterilizer can support the intended EO process.

What changed in EN 1422:2026 from the previous edition?

EN 1422:2026 supersedes EN 1422:2014. The foreword states that the main text has been restructured to align with ISO/TS 22421:2021 and harmonized with the current revisions of EN 17180 and EN 14180.

The foreword also states that references were updated, including Table E.1 and the bibliography, that some definitions were added and most were adapted with reference to EN ISO 11139:2018/A1:2024, that a separate clause on protective measures was added for referencing EN IEC 61010-2-040:2021, that a new Annex G was added, and that the control and monitoring requirements were merged into a new clause with illustrative explanations in Annex F.

Which standards are used with EN 1422:2026?

EN 1422:2026 relies on a number of companion standards for design, testing, packaging, safety and documentation.

  • EN ISO 11135:2014 - Covers development, validation and routine control of EO sterilization processes. EN 1422:2026 points users to it for process validation and routine control.
  • EN IEC 61010-2-040:2021 - Gives the particular safety requirements for sterilizers and washer-disinfectors used to treat medical materials. EN 1422:2026 uses it for protective measures and safety-related design.
  • EN ISO 14971:2019 - Supports risk management for medical devices. EN 1422:2026 uses risk analysis to shape the sterilizer design and fault handling.
  • EN ISO 12100:2010 - Supports machinery risk assessment and risk reduction. It is relevant to mechanical and usability-related hazards.
  • EN IEC 61326-1:2021 - Covers EMC requirements for measurement, control and laboratory equipment. EN 1422:2026 uses it for emissions and immunity.
  • EN ISO 11138-1:2017 and EN ISO 11138-2:2017 - Define requirements for biological indicators, including those for EO processes. They support Annex D testing.
  • EN ISO 11607-1:2020, EN 868-4:2017 and EN 868-5:2018 - Cover sterile barrier systems and packaging for terminally sterilized medical devices. They are used in microbiological performance testing.
  • EN ISO 10993-7:2008 - Covers EO residuals in medical devices and is used for aeration and residual EO considerations.
  • EN 13445-1:2021 to EN 13445-5:2021+A1:2024 and EN 13445-8:2021 - Support design, fabrication, inspection and testing of unfired pressure vessels.
  • EN 764-7:2002 and EN 14222:2021 - Support pressure equipment safety systems and steam boiler requirements where those components are used.
  • EN 60073:2002, EN ISO 15223-1:2021, EN ISO 3166-1:2020, ISO 639:2023, ISO 8601-1:2019, ISO 8000-1:2022 and ISO 8573-1:2010 - Support indicators, symbols, country and language codes, date-time formats, data quality and compressed air quality.
  • EN 61511-1:2017 and EN 62366-1:2015 - Support functional safety and usability engineering.

What does the EN 1422:2026 document contain?

EN 1422:2026 contains a recommended test programme in Annex A, including type tests, works tests, acceptance testing upon installation and IQ-related provisions. It also contains normative test methods for leak testing, chamber profile testing, microbiological performance testing and test instrumentation in Annexes B, C, D and H.

The document includes practical guidance annexes on environmental aspects, control-recording relationships and example cycle stages. Annex F uses illustrations to explain how control and recording can be arranged, while Annex I shows the stages of an EO cycle and what each stage is for.

Annex ZA maps the standard to the General Safety and Performance Requirements of Regulation (EU) 2017/745, and also to related machinery requirements where applicable. For a buyer, manufacturer or auditor, that annex is the bridge between the technical clauses and the EU regulatory framework.

Relations

Effective Date
30-Aug-2023

Buy Documents

Get Certified

Connect with accredited certification bodies for this standard

BSI Group

BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

UKAS United Kingdom Verified

TÜV Rheinland

TÜV Rheinland is a leading international provider of technical services.

DAKKS Germany Verified

TÜV SÜD

TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.

DAKKS Germany Verified

Sponsored listings

Frequently Asked Questions

EN 1422:2026 is a draft published by the European Committee for Standardization (CEN). Its full title is "Sterilizers for medical purposes - Ethylene oxide sterilizers - Requirements and test methods". This standard covers: 1.1 This document specifies the requirements and the relevant tests for automatically controlled sterilizers employing ethylene oxide (EO) gas as the sterilant, either as a pure gas or a mixture with other gases (for example carbon dioxide or nitrogen) being used for the sterilization of medical devices and their accessories. 1.2 This document specifies requirements for ethylene oxide sterilizers (EO sterilizers) for: — the performance and design of sterilizers intended to deliver a process capable of sterilizing medical devices; — the equipment and controls of these sterilizers, needed for operation, control and monitoring of the sterilization processes; — the test equipment and test procedures used to verify the sterilizer performance specified by this document. 1.3 This document does not cover sterilizers which employ the injection of EO or mixtures containing EO directly into packages or into a flexible chamber.

1.1 This document specifies the requirements and the relevant tests for automatically controlled sterilizers employing ethylene oxide (EO) gas as the sterilant, either as a pure gas or a mixture with other gases (for example carbon dioxide or nitrogen) being used for the sterilization of medical devices and their accessories. 1.2 This document specifies requirements for ethylene oxide sterilizers (EO sterilizers) for: — the performance and design of sterilizers intended to deliver a process capable of sterilizing medical devices; — the equipment and controls of these sterilizers, needed for operation, control and monitoring of the sterilization processes; — the test equipment and test procedures used to verify the sterilizer performance specified by this document. 1.3 This document does not cover sterilizers which employ the injection of EO or mixtures containing EO directly into packages or into a flexible chamber.

EN 1422:2026 is classified under the following ICS (International Classification for Standards) categories: 11.080.10 - Sterilizing equipment. The ICS classification helps identify the subject area and facilitates finding related standards.

EN 1422:2026 has the following relationships with other standards: It is inter standard links to EN 1422:2014. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

EN 1422:2026 is associated with the following European legislation: EU Directives/Regulations: 2017/745; Standardization Mandates: M/575, M/575 Amd 2. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.

EN 1422:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)


SLOVENSKI STANDARD
oSIST prEN 1422:2025
01-junij-2025
Sterilizatorji za uporabo v medicini - Sterilizatorji z etilenoksidom - Zahteve in
preskusne metode
Sterilizers for medical purposes - Ethylene oxide sterilizers - Requirements and test
methods
Sterilisatoren für medizinische Zwecke - Ethylenoxid-Sterilisatoren - Anforderungen und
Prüfverfahren
Stérilisateurs à usage médical - Stérilisateurs à l'oxyde d'éthylène - Exigences et
méthodes d'essai
Ta slovenski standard je istoveten z: prEN 1422
ICS:
11.080.10 Sterilizacijska oprema Sterilizing equipment
oSIST prEN 1422:2025 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

oSIST prEN 1422:2025
oSIST prEN 1422:2025
DRAFT
EUROPEAN STANDARD
prEN 1422
NORME EUROPÉENNE
EUROPÄISCHE NORM
April 2025
ICS 11.080.10 Will supersede EN 1422:2014
English Version
Sterilizers for medical purposes - Ethylene oxide
sterilizers - Requirements and test methods
Stérilisateurs à usage médical - Stérilisateurs à l'oxyde Sterilisatoren für medizinische Zwecke - Ethylenoxid-
d'éthylène - Exigences et méthodes d'essai Sterilisatoren - Anforderungen und Prüfverfahren
This draft European Standard is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee
CEN/TC 102.
If this draft becomes a European Standard, CEN members are bound to comply with the CEN/CENELEC Internal Regulations
which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.

This draft European Standard was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.

Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.

EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2025 CEN All rights of exploitation in any form and by any means reserved Ref. No. prEN 1422:2025 E
worldwide for CEN national Members.

oSIST prEN 1422:2025
prEN 1422:2025 (E)
Contents Page
European foreword . 6
Introduction . 7
1 Scope . 8
2 Normative references . 8
3 Terms and definitions . 10
4 General. 17
4.1 Product definition. 17
4.2 Equipment development . 17
4.3 Calibration . 18
5 Equipment design and construction . 18
5.1 Safety and security . 18
5.1.1 General. 18
5.1.2 Protective measures . 19
5.1.3 Risk control and usability . 21
5.2 Sterilizer chamber . 21
5.2.1 Dimensions. 21
5.2.2 Doors and interlocks of the sterilizer chamber . 22
5.2.3 Chamber integrity . 23
5.2.4 Pressure vessels . 24
5.2.5 Uniformity of conditions, temperature control and insulating the sterilizer chamber . 24
5.3 Further functional components . 24
5.3.1 Pipework and fittings . 24
5.3.2 Vaporizers . 25
5.3.3 Evacuation system . 25
5.3.4 Framework and panelling . 26
5.3.5 Loading Equipment . 27
5.4 Materials . 27
5.4.1 Chamber materials . 27
5.4.2 Other materials used in the construction of the sterilizer . 27
5.5 Test connections . 28
5.6 Vibration . 29
5.7 User interfaces . 29
5.8 Transport . 30
6 Indicating, monitoring, controlling and recording devices . 31
6.1 General. 31
6.1.1 General. 31
6.1.2 Pre-set programme . 31
6.1.3 Automatic controller . 31
6.2 Automatic control . 31
6.3 Control and monitoring system . 33
6.4 Failure . 34
6.4.1 General. 34
6.4.2 Fault . 35
6.4.3 Power failure . 36
oSIST prEN 1422:2025
prEN 1422:2025 (E)
6.4.4 Other failures . 36
6.5 Instrumentation . 36
6.5.1 General . 36
6.5.2 Temperature measuring devices . 37
6.5.3 Pressure measuring devices . 39
6.5.4 Time measuring devices . 40
6.5.5 Sterilizing agent control and measuring devices . 40
6.5.6 Relative humidity (RH) sensors . 41
6.6 Indicating devices . 41
6.6.1 General . 41
6.6.2 Cycle parameter indicating devices . 41
6.6.3 Cycle parameter indications . 43
6.6.4 Status indicators and indications . 43
6.6.5 Operating cycle counter . 44
6.7 Recorders . 45
6.7.1 General . 45
6.7.2 Records . 46
6.7.3 Analogue presentation of records . 47
6.7.4 Digital records . 47
6.8 Operating cycles . 48
6.8.1 General . 48
6.8.2 Maintenance leak test . 48
6.8.3 Sterilization cycles . 49
7 Services and local environment . 51
7.1 General . 51
7.2 Sterilizing agent and sterilant . 52
7.3 Electrical supply . 53
7.4 Water . 53
7.5 Steam . 54
7.6 Vacuum . 54
7.7 Drains . 54
7.8 Lighting . 55
7.9 Compressed air . 55
7.10 Air and inert gases. 55
7.11 Ventilation . 56
8 Emissions . 56
8.1 Electromagnetic emissions . 56
8.2 Noise . 56
8.3 Exhaust emissions . 57
8.4 Heat emission . 58
9 Test instrumentation . 58
10 Performance and assessment . 58
10.1 General . 58
10.2 Chamber integrity . 59
10.3 Attainment of conditions; physical parameters . 60
10.3.1 Heating of the sterilizer chamber internal surfaces . 60
10.3.2 Temperature profile for an empty sterilizer chamber . 60
10.3.3 Pressure profile and pressure change . 60
10.3.4 Sterilizing agent and sterilant . 61
10.4 Microbiological performance . 61
10.5 EO removal (flushing) . 61
oSIST prEN 1422:2025
prEN 1422:2025 (E)
10.6 Aeration . 61
10.7 Load dryness . 61
10.8 Loading configuration . 62
11 Information to be supplied . 62
11.1 General. 62
11.2 Information to be made available prior to purchase . 62
11.3 Marking and labelling . 64
11.4 Labelling . 65
11.5 Instructions for use . 65
12 Packaging . 67
Annex A (informative)  Test programme . 68
A.1 General. 68
A.2 Type test . 68
A.3 Works test . 68
A.4 Tests programme . 68
A.5 Acceptance test upon installation . 70
A.6 Installation qualification (IQ) provisions . 70
Annex B (normative) Leak test cycle. 71
Annex C (normative) Sterilizer chamber profile testing . 72
C.1 Sterilizer chamber internal surfaces . 72
C.2 Empty sterilizer chamber . 72
Annex D (normative) Microbiological performance test for EO sterilizers . 74
D.1 General. 74
D.2 Apparatus . 74
D.3 Procedure . 75
D.4 Interpretation of results . 75
Annex E (informative) Environmental aspects . 77
E.1 Environmental aspects regarding the life cycle of EO sterilizers . 77
E.2 EO . 77
E.3 Environmental impact . 77
Annex F (informative)  Illustrations of the interrelationship between control and recording . 79
F.1 Introduction . 79
F.2 Illustration 1 . 81
F.3 Illustration 2 . 82
F.4 Illustration 3 . 83
Annex G (informative)  Additional information on protective measures . 86
Annex H (normative) Test Instrumentation . 87
Annex I (informative) Example of the stages used in an EO sterilization cycle and guidance on
their purpose. . 89
oSIST prEN 1422:2025
prEN 1422:2025 (E)
I.1 General . 89
I.2 Stage 0: Chamber pre-heating/cooling . 89
I.3 Stage 1: Air removal . 89
I.4 Stage 2: Automatic Leak Rate Test (ALRT) . 89
I.5 Stage 3: conditioning (if used) . 90
I.6 Stage 4: EO exposure . 90
I.7 Stage 5: EO holding time . 90
I.8 Stage 6: EO removal . 90
I.9 Stage 7: Desorption . 90
I.10 Stage 8: Air admission . 90
I.11 Stage 9: Cycle complete. 91
I.12 Stage 10: Aeration . 91
Annex ZA (informative) Relationship between this European Standard and the General Safety
and Performance Requirements of Regulation (EU) 2017/745 aimed to be covered. 92
Bibliography . 114

oSIST prEN 1422:2025
prEN 1422:2025 (E)
European foreword
This document (prEN 1422:2025) has been prepared by Technical Committee CEN/TC 102 “Sterilizers
and associated equipment for processing of medical devices”, the secretariat of which is held by DIN.
This document is currently submitted to the CEN Enquiry.
This document will supersede EN 1422:2014.
In comparison with the previous edition, the following technical modifications have been made:
— structure of the main text has been adapted to the structure of ISO/TS 22421:2021 and harmonized
with the current revisions of EN 17180 and EN 14180;
— references have been updated, including Table E.1 on Environmental Aspects, and the bibliography;
— some definitions of terms have been added, most definitions have been adapted with reference to
EN ISO 11139:2018/A1:2024;
— a separate clause ‘Protective measures’ has been implemented for referencing to individual clauses
of EN IEC 61010-2-040:2021;
— a new Annex G ‘Additional information on protective measures’ has been added;
— requirements on the control and monitoring system have been merged into a new clause and
informative illustrations including some explanations have been provided in a new Annex F.
This document has been prepared under a standardization request addressed to CEN by the European
Commission. The Standing Committee of the EFTA States subsequently approves these requests for its
Member States.
For the relationship with EU Legislation, see informative Annex ZA, which is an integral part of this
document.
In this document, the following verbal forms are used:
— “shall” indicates a requirement;
— “should” indicates a recommendation;
— “may” indicates a permission;
— “can” indicates a possibility or capability; and
— "must" is used to express an external constraint.
In this document notes are used for giving additional information intended to assist the understanding
or use of the text. The document is usable without the notes.
oSIST prEN 1422:2025
prEN 1422:2025 (E)
Introduction
This document provides minimum requirements and test methods for sterilizers performing a low
temperature ethylene oxide (EO) sterilization process.
Ethylene oxide (EO) sterilizers employ EO gas as the sterilant, either as a pure gas or in admixture with
other gases (e.g. carbon dioxide). When used as a pure gas the gas exposure stage is usually carried out
at sub-atmospheric pressures. When used as an admixture the gas exposure stage is usually carried out
at super-atmospheric pressures. Ethylene oxide sterilization processes are primarily used for the

sterilization of heat labile products at temperatures less than 60 °C.
EO is a highly reactive chemical which can present a toxicity, flammability or explosive hazard if
incorrectly handled (see Annex E).
The EO sterilizer specified in this document can be used for medical, dental, pharmaceutical, veterinary,
industrial or other related purposes in health care facilities or industrial settings.
The equipment related tests described in this document are reference tests intended for use in
demonstrating conformity with the performance requirements specified in this document. They may be
used in type tests, works tests, in validation and re-validation tests, or in periodic and routine tests
carried out by the user. Validation and routine control of sterilization processes are essential to ensure
their efficacy. This document does not cover validation and routine control of EO sterilization processes.
Such activities are covered in EN ISO 11135:2014.
EO sterilizers should not be assumed to deliver processes effectively inactivating the causative agents of
spongiform encephalopathies such as scrapie, Bovine Spongiform Encephalopathy and Creutzfeldt-Jakob
Disease. Specific recommendations have been produced in particular countries for the processing of
materials potentially contaminated with these agents. See also EN ISO 11135:2014, 1.2.
Planning and design of sterilizers complying with this document should consider not only technical issues
but also the environmental impact from the use of the sterilizer during its life cycle. Environmental
aspects are addressed in Annex E of this document.
By performing tests concurrently and/or in a logical succession, the total number of tests and consumable
test equipment disposals is reduced. As a result, the burden on the environment can be reduced (see also
Annex E).
This document is not intended as a checklist for the suitability of an existing EO sterilizer when assessing
compliance with EN ISO 11135:2014. This standard is not intended to be applied retrospectively.
This document does not cover analytical methods for determining levels of residual EO and/or its
reaction products. Analytical methods for determining levels of residual EO and/or its reaction products
are addressed in ISO 10993-7:2008.
General equipment safety requirements are specified in EN 61010-1:2010, EN IEC 61010-2-040:2021
and EN 60204-1:2018 and can provide valuable information on design options. They are referenced
herein, but not repeated.
Occupational safety is not addressed in this document.
oSIST prEN 1422:2025
prEN 1422:2025 (E)
1 Scope
1.1 This document specifies the requirements and the relevant tests for automatically controlled
sterilizers employing ethylene oxide (EO) gas as the sterilant, either as a pure gas or a mixture with other
gases, being used for the sterilization of medical devices and their accessories.
1.2 This document specifies requirements for ethylene oxide sterilizers (EO sterilizers) for:
— the performance and design of sterilizers intended to deliver a process capable of sterilizing medical
devices;
— the equipment and controls of these sterilizers, needed for operation, control and monitoring of the
sterilization processes;
— the test equipment and test procedures used to verify the sterilizer performance specified by this
document.
1.3 This document does not cover sterilizers which employ the injection of EO or mixtures containing EO
directly into packages or into a flexible chamber.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content
constitutes requirements of this document. For dated references, only the edition cited applies. For
undated references, the latest edition of the referenced document (including any amendments) applies.
EN 764-7:2002, Pressure equipment - Part 7: Safety systems for unfired pressure equipment
EN 868-4:2017, Packaging for terminally sterilized medical devices - Part 4: Paper bags - Requirements and
test methods
EN 868-5:2018, Packaging for terminally sterilized medical devices - Part 5: Sealable pouches and reels of
porous materials and plastic film construction - Requirements and test methods
EN 13445-1:2021, Unfired pressure vessels - Part 1: General
EN 13445-2:2021+A1:2023, Unfired pressure vessels - Part 2: Materials
EN 13445-3:2021, Unfired pressure vessels - Part 3: Design
EN 13445-4:2021+A1:2023, Unfired pressure vessels - Part 4: Fabrication
EN 13445-5:2021+A1:2024, Unfired pressure vessels - Part 5: Inspection and testing
EN 13445-8:2021, Unfired pressure vessels - Part 8: Additional requirements for pressure vessels of
aluminium and aluminium alloys
EN 14222:2021, Stainless steel steam boilers
EN 60073:2002, Basic and safety principles for man-machine interface, marking and identification - Coding
principles for indicators and actuators (IEC 60073:2002)
EN IEC 61010-2-040:2021, Safety requirements for electrical equipment for measurement, control and
laboratory use — Part 2-040: Particular requirements for sterilizers and washer-disinfectors used to treat
medical materials (IEC 61010-2-040:2020)
oSIST prEN 1422:2025
prEN 1422:2025 (E)
EN 61511-1:2017, Functional safety - Safety instrumented systems for the process industry sector - Part 1:
Framework, definitions, system, hardware and application programming Requirements (IEC 61511-1:2016
+ COR1:2016)
EN 62366-1:2015, Medical devices Part 1: Application of usability engineering to medical devices (IEC
62366-1:2015)
EN IEC 61326-1:2021, Electrical equipment for measurement, control and laboratory use - EMC
requirements — Part 1: General requirements (IEC 61326-1:2020)
EN ISO 3166-1:2020, Codes for the representation of names of countries and their subdivisions - Part 1:
Country code (ISO 3166-1:2020)
EN ISO 3746:2010, Acoustics - Determination of sound power levels and sound energy levels of noise sources
using sound pressure - Survey method using an enveloping measurement surface over a reflecting plane (ISO
3746:2010)
EN ISO 10993-7:2008, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization
residuals (ISO 10993-7:2008)
EN ISO 11135:2014, Sterilization of health-care products — Ethylene oxide — Requirements for the
development, validation, and routine control of a sterilization process for medical devices (ISO 11135:2014)
EN ISO 11138-1:2017, Sterilization of health care products - Biological indicators - Part 1: General
requirements (ISO 11138-1:2017)
EN ISO 11138-2:2017, Sterilization of health care products - Biological indicators - Part 2: Biological
indicators for ethylene oxide sterilization processes (ISO 11138-2:2017)
EN ISO 11607-1:2020, Packaging for terminally sterilized medical devices - Part 1: Requirements for
materials, sterile barrier systems and packaging systems (ISO 11607-1:2019)
EN ISO 12100:2010, Safety of machinery - General principles for design- Risk assessment and risk reduction
(ISO 12100:2010)
EN ISO 14971:2019, Medical devices - Application of risk management to medical devices (ISO
14971:2019)
EN ISO 15223-1:2021, Medical devices - Symbols to be used with information to be supplied by the
manufacturer - Part 1: General requirements (ISO 15223-1:2021)
ISO 639:2023, Code for individual languages and language groups
ISO 8000-1:2022, Data quality - Part 1: Overview
ISO 8573-1:2010, Compressed air - Part 1: Contaminants and purity classes
ISO 8601-1:2019, Date and time - Representations for information interchange - Part 1: Basic rules

As impacted by EN 61511-1:2017/A1:2017.
As impacted by EN 62366-1:2015/AC:2015/A1:2020.
As impacted by EN ISO 10993-7:2008/AC:2009/A1:2022.
As impacted by EN ISO 11135:2014/A1:2019.
As impacted by EN ISO 11607-1:2020/A1:2023.
As impacted by EN ISO 14971:2019/A11:2021.
As impacted by ISO 8601-1:2019/AMD 1:2022.
oSIST prEN 1422:2025
prEN 1422:2025 (E)
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply:
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
— IEC Electropedia: available at https://www.electropedia.org/
— ISO Online browsing platform: available at https://www.iso.org/obp/
3.1
access device
means by which entry to restricted parts of equipment is achieved
Note 1 to entry: This can be by dedicated key, code, or tool.
Note 2 to entry: There is likely to be more than one access device all of which can be different depending on
function.
[SOURCE: EN ISO 11139:2018/A1:2024, 3.4, modified — Notes 1 and 2 to entry added.]
3.2
aeration
part of the sterilization process during which either the sterilizing agent or its reaction products, or both,
desorb from the health care product until predetermined levels are reached
Note 1 to entry: This can be performed within the sterilization chamber or in a separate chamber or room.
[SOURCE: EN ISO 11135: 2014, 3.7, modified — Note 1 to entry added and changed “sterilization cycle”
to “sterilization process”.]
3.3
automatic controller
device that directs the equipment sequentially through required stages of the cycle in response to
programmed cycle parameters
[SOURCE: EN ISO 11139:2018/A1:2024, 3.18]
3.4
biological indicator
test system containing viable microorganisms providing a specified resistance to a specified sterilization
process
[SOURCE: EN ISO 11139:2018/A1:2024, 3.29]
3.5
calibration
operation that, under specified conditions, in a first step, establishes a relation between the quantity
values with measurement uncertainties provided by measurement standards and corresponding
indications with associated measurement uncertainties and, in a second step, uses this information to
establish a relation for obtaining a measurement result from an indication
[SOURCE: EN ISO 11139:2018/A1:2024, 3.31]
oSIST prEN 1422:2025
prEN 1422:2025 (E)
3.6
chamber
part of equipment in which a load is processed
[SOURCE: EN ISO 11139:2018/A1:2024]
3.7
conditioning
treatment of load prior to the exposure phase to attain a specified temperature, relative humidity, or
other process variable throughout the load
[SOURCE: EN ISO 11139:2018/A1:2024, 3.58, modified — “product” is changed to “load”.]
3.8
control
regulation of variables within specified limits
[SOURCE: EN ISO 11139:2018/A1:2024, 3.63]
3.9
cycle complete
message from the automatic controller that the operating cycle has ended successfully
Note 1 to entry: Upon indication of “cycle complete” in the sterilizer, the processed load can need a period of
aeration using a separate operating cycle or in a separate chamber.
[SOURCE: EN ISO 11139:2018/A1:2024, 3.71, modified — Note 1 to entry has been added.]
3.10
cycle parameter
value of a cycle variable including its tolerance used for control, monitoring, indication, and recording of
an operating cycle
Note 1 to entry: See also 3.24, process parameter.
[SOURCE: EN ISO 11139:2018/A1:2024, 3.72, modified — Note 1 to entry added.]
3.11
cycle variable
property used to control, monitor, indicate, or record an operating cycle
Note 1 to entry: Process variables that are used by the automatic controller to control, monitor, indicate or record
are also cycle variables, so a cycle variable can also be a process variable, e.g. time.
Note 2 to entry: Pressure can be a cycle variable.
Note 3 to entry: See also 3.26, process variable.
[SOURCE: EN ISO 11139:2018/A1:2024, 3.74, modified — Note 1, 2 and 3 to entry has been added.]
3.12
double-ended
having separate doors for loading and unloading in separate areas
[SOURCE: EN ISO 11139:2018/A1:2024, 3.92, modified — Note 1 to entry deleted.]
oSIST prEN 1422:2025
prEN 1422:2025 (E)
3.13
exposure stage
cycle stage between the introduction of the sterilizing or disinfecting agent into the chamber and when
its microbicidal effect has become negligible
Note 1 to entry: The exposure stage comprises that part of the process for which microbial lethality is claimed.
[SOURCE: EN ISO 11139:2018/A1:2024, 3.111]
3.14
fault
situation in which one or more of the process or cycle parameters is/are outside its/their specified
tolerance(s)
[SOURCE: EN ISO 11139:2018/A1:2024, 3.116]
3.15
flushing
purging
removing by displacement with a fluid
[SOURCE: EN ISO 11139:2018/A1:2024, 3.121]
3.16
indicate
display a value, condition, or stage of process
[SOURCE: EN ISO 11139:2018/A1:2024, 3.139]
3.17
load
product, equipment, or materials to be processed together within an operation cycle
[SOURCE: EN ISO 11139:2018/A1:2024, 3.155]
3.18
maximum allowable working pressure
MAWP
maximum pressure to which a component is designed to be subjected to and which is the basis for
determining the strength of the component under consideration
Note 1 to entry: The maximum pressure which can be applied to a specified product without compromising safety,
quality or performance can be given in the process specification.
3.19
measuring chain
series of elements of a measuring instrument measuring system, which constitutes the path of the
measurement signal from the input (quantity subject to measurement) to the output (the result of the
measurement)
[SOURCE: EN ISO 11139:2018/A1:2024, 3.165]
oSIST prEN 1422:2025
prEN 1422:2025 (E)
3.20
medical device
instrument, apparatus, implement, machine, appliance, implant, reagent for in vitro use, or software
material, or other similar or related article, intended by the manufacturer to be used, alone or in
combination, for human beings, for one or more of the specific medical purpose(s) of:
— diagnosis, prevention, monitoring, treatment, or alleviation of disease;
— diagnosis, monitoring, treatment, alleviation of, or compensation for an injury;
— investigation, replacement, modification, or support of the anatomy, or of a physiological process;
— supporting or sustaining life;
— control of conception;
— disinfection of medical devices;
— providing information by means of in vitro examination of specimens derived from the human body;
and does not achieve its primary intended action by pharmacological, immunological, or metabolic
means, in or on the human body, but which may be assisted in its intended function by such means
Note 1 to entry: Products which may be considered to be medical devices in some jurisdictions, but not in others
include:
— items
...