EN 14625:2024/prA1
(Amendment)Ambient air - Standard method for the measurement of the concentration of ozone by ultraviolet photometry
- Abstract
This European Standard specifies a continuous measurement method for the determination of the concentrations of ozone present in ambient air based on the ultraviolet photometric measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate ultraviolet photometric analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use.
The method is applicable to the determination of the concentration of ozone present in ambient air up to 500 μg/m3.
This concentration range represents the certification range for ozone for the type approval test.
NOTE 1 Other ranges may be used for measurement systems applied at rural locations monitoring ecosystems.
NOTE 2 When the standard is used for other purposes than Directive 2008/50/EC, the ranges and uncertainty requirements may not apply.
The method covers the determination of ambient air concentrations of ozone in zones classified as rural areas, urban and urban-background areas.
The results are expressed in μg/m3 (at 20 °C and 101,3 kPa).
NOTE 3 500 μg/m3 of O3 corresponds to 250 nmol/mol of O3 at 20 °C and 101,3 kPa.
This standard contains information for different groups of users.
Clauses 5 to 7 and Annexes B and C contain general information about the principles of ozone measurement by ultraviolet photometric analyser and sampling equipment.
Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of ozone analysers. These sections contain information about:
− type-approval test conditions, test procedures and test requirements;
− analyser performance requirements;
− evaluation of the type-approval test results;
− evaluation of the uncertainty of the measurement results of the ozone analyser based on the type-approval test results.
Clauses 9 to 11 and Annexes F and G are directed towards monitoring networks performing the practical
measurements of ozone in ambient air. These sections contain information about:
− initial installation of the analyser in the monitoring network and acceptance testing;
− ongoing quality assurance/quality control;
− calculation and reporting of measurement results;
− evaluation of the uncertainty of measurement results under practical monitoring conditions.
- Status
- Not Published
- Publication Date
- 03-Aug-2028
- Technical Committee
- CEN/TC 264 - Air quality
- Current Stage
- 4020 - Submission to enquiry - Enquiry
- Start Date
- 22-Oct-2026
- Due Date
- 23-Apr-2027
- Completion Date
- 22-Oct-2026
EN 14625:2024/prA1 is a CEN draft amendment to the European Standard for measuring ozone in ambient air by ultraviolet photometry. It updates the method used to assess ozone analysers for type approval and for routine monitoring at fixed sites. The document is aimed at test houses, laboratories and monitoring networks that work with ambient-air ozone measurements.
What does EN 14625:2024/prA1 specify?
EN 14625:2024/prA1 amends the standard method for continuous measurement of ozone in ambient air using the ultraviolet photometric measuring principle. The underlying method applies up to 500 μg/m3, and the results are expressed at 20 °C and 101.3 kPa.
The standard covers rural, urban and urban-background areas. It also ties the method to the data quality requirements in Annex I of Directive 2008/50/EC for fixed-site monitoring and to the sampling, calibration and quality-assurance conditions needed in practice.
| Part | What it covers |
|---|---|
| Clauses 5 to 7 and Annexes B and C | Basic principles of ozone measurement and sampling equipment |
| Clause 8 and Annex E | Type-approval testing of ozone analysers, including performance criteria and uncertainty evaluation |
| Clauses 9 to 11 and Annexes F and G | Installation in monitoring networks, acceptance testing, QA/QC, calculation and reporting, and practical uncertainty evaluation |
| Annex D | Procedure for applying test gas |
What are the key requirements of EN 14625:2024/prA1?
EN 14625:2024/prA1 mainly corrects and tightens technical points that affect analyser approval and field use. The changes matter because they affect how an analyser is tested, installed, checked and reported.
Type-approval testing and performance criteria
Clause 8 and Annex E are updated for type-approval work. Table 1 is corrected for residual at zero and for long-term and short-term drift, so laboratories must use the revised acceptance values when judging analyser performance.
Clause 8.4.2.3 now points to Table 4 for the purity of zero air and the air feeding the ozone generator. In practice, this means the test gases and gas supply must meet the stated quality before the analyser can be qualified.
Clause 8.4.6 revises the lack-of-fit test for the calibration function. The concentrations are applied in a fixed order across the certification range, with a tolerance at each level and a waiting period for stable readings. That makes the calibration test repeatable and reduces ambiguity in the response check.
Sampling path and installation checks
Clause 6.3.2 updates the particle filter note, and Clause 6.6 tightens the sampling pump to manifold requirement to below 1.0%. For users, this means the sampling path must be built and checked so that it does not distort the ozone reading.
Clause 9.7.3 renames the test from testing the sample manifold to testing the sampling system. That shift is practical: operators need to treat the full sample path as one measurement chain, not just one component.
Routine monitoring, checks and uncertainty
Clause 9.6.1.2 updates the zero and span check wording, and Clause 8.5.4 clarifies the long-term drift requirement at span. These changes matter for routine QA/QC because they define when an analyser needs attention or corrective action.
Clause 9.5.5 removes one item from the data-treatment list after exceedance of performance criteria, and Annexes D, E and F correct formulas used for test gas application, electric voltage and combined standard uncertainty. In practice, this affects how monitoring data, test results and uncertainty statements are calculated and documented.
What terms does EN 14625:2024/prA1 define?
ambient air - Outdoor air in the troposphere, excluding workplaces as defined by Directive 89/654/EEC, where health and safety at work provisions apply, and to which the public does not have regular access.
Who uses EN 14625:2024/prA1?
EN 14625:2024/prA1 is used by test houses and laboratories that perform type-approval testing of ozone analysers. It is also used by operators of monitoring networks that install analysers, run acceptance tests, carry out calibration and quality control, calculate results and report uncertainty.
Environmental air-quality teams use it when they need measurements that support Directive 2008/50/EC. Engineers, quality managers and buyers use it to choose analysers, plan sampling systems and check that the measurement chain matches the standard.
What changed in EN 14625:2024/prA1 from the previous edition?
EN 14625:2024/prA1 is an amendment to EN 14625:2024 and corrects several technical points. The amendment:
- removes the note in Clause 1 about discontinuing older reference analysers;
- revises the definition of ambient air in Clause 3;
- updates wording in Clauses 6.3.2 and 6.6, including the sampling pump to manifold limit;
- corrects performance criteria in Table 1 of Clause 8.2;
- changes the test-gas purity requirement in Clause 8.4.2.3;
- revises the calibration sequence in Clause 8.4.6;
- clarifies the long-term drift statement in Clause 8.5.4;
- deletes Clause 9.5.4;
- changes the zero-drift wording in Clause 9.6.1.2;
- renames Clause 9.7.3;
- corrects formulas in Annexes D, E and F.
Which standards are used with EN 14625:2024/prA1?
EN 14625:2024/prA1 refers to Council Directive 2008/50/EC for the data quality requirements in Annex I and for the source of the original ambient-air definition. It also refers to Directive 89/654/EEC in the revised definition of ambient air.
What does the EN 14625:2024/prA1 document contain?
EN 14625:2024/prA1 contains targeted amendments to clauses on scope, terms and definitions, sampling hardware, performance criteria, test gases, calibration behaviour, drift, field testing and uncertainty. It also updates formulas in Annexes D, E and F.
The document does not restate the full ozone method; it edits the existing standard where the technical corrections are needed. For implementers, the most important parts are the revised acceptance criteria, the calibration test sequence, the sampling-system checks and the uncertainty formulas.
Relations
- Effective Date
- 03-Jun-2026
Get Certified
Connect with accredited certification bodies for this standard
CIS Institut d.o.o.
Personal Protective Equipment (PPE) certification body. Notified Body NB-2890 for EU Regulation 2016/425 PPE.

Kiwa BDA Testing
Building and construction product certification.
Kmetijski inštitut Slovenije
Agricultural Institute of Slovenia. Soil testing, plant health, agricultural product analysis.
Sponsored listings
Frequently Asked Questions
EN 14625:2024/prA1 is a draft published by the European Committee for Standardization (CEN). Its full title is "Ambient air - Standard method for the measurement of the concentration of ozone by ultraviolet photometry". This standard covers: This European Standard specifies a continuous measurement method for the determination of the concentrations of ozone present in ambient air based on the ultraviolet photometric measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate ultraviolet photometric analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use. The method is applicable to the determination of the concentration of ozone present in ambient air up to 500 μg/m3. This concentration range represents the certification range for ozone for the type approval test. NOTE 1 Other ranges may be used for measurement systems applied at rural locations monitoring ecosystems. NOTE 2 When the standard is used for other purposes than Directive 2008/50/EC, the ranges and uncertainty requirements may not apply. The method covers the determination of ambient air concentrations of ozone in zones classified as rural areas, urban and urban-background areas. The results are expressed in μg/m3 (at 20 °C and 101,3 kPa). NOTE 3 500 μg/m3 of O3 corresponds to 250 nmol/mol of O3 at 20 °C and 101,3 kPa. This standard contains information for different groups of users. Clauses 5 to 7 and Annexes B and C contain general information about the principles of ozone measurement by ultraviolet photometric analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of ozone analysers. These sections contain information about: − type-approval test conditions, test procedures and test requirements; − analyser performance requirements; − evaluation of the type-approval test results; − evaluation of the uncertainty of the measurement results of the ozone analyser based on the type-approval test results. Clauses 9 to 11 and Annexes F and G are directed towards monitoring networks performing the practical measurements of ozone in ambient air. These sections contain information about: − initial installation of the analyser in the monitoring network and acceptance testing; − ongoing quality assurance/quality control; − calculation and reporting of measurement results; − evaluation of the uncertainty of measurement results under practical monitoring conditions.
This European Standard specifies a continuous measurement method for the determination of the concentrations of ozone present in ambient air based on the ultraviolet photometric measuring principle. This standard describes the performance characteristics and sets the relevant minimum criteria required to select an appropriate ultraviolet photometric analyser by means of type approval tests. It also includes the evaluation of the suitability of an analyser for use in a specific fixed site so as to meet the data quality requirements as specified in Annex I of Directive 2008/50/EC [1] and requirements during sampling, calibration and quality assurance for use. The method is applicable to the determination of the concentration of ozone present in ambient air up to 500 μg/m3. This concentration range represents the certification range for ozone for the type approval test. NOTE 1 Other ranges may be used for measurement systems applied at rural locations monitoring ecosystems. NOTE 2 When the standard is used for other purposes than Directive 2008/50/EC, the ranges and uncertainty requirements may not apply. The method covers the determination of ambient air concentrations of ozone in zones classified as rural areas, urban and urban-background areas. The results are expressed in μg/m3 (at 20 °C and 101,3 kPa). NOTE 3 500 μg/m3 of O3 corresponds to 250 nmol/mol of O3 at 20 °C and 101,3 kPa. This standard contains information for different groups of users. Clauses 5 to 7 and Annexes B and C contain general information about the principles of ozone measurement by ultraviolet photometric analyser and sampling equipment. Clause 8 and Annex E are specifically directed towards test houses and laboratories that perform type-approval testing of ozone analysers. These sections contain information about: − type-approval test conditions, test procedures and test requirements; − analyser performance requirements; − evaluation of the type-approval test results; − evaluation of the uncertainty of the measurement results of the ozone analyser based on the type-approval test results. Clauses 9 to 11 and Annexes F and G are directed towards monitoring networks performing the practical measurements of ozone in ambient air. These sections contain information about: − initial installation of the analyser in the monitoring network and acceptance testing; − ongoing quality assurance/quality control; − calculation and reporting of measurement results; − evaluation of the uncertainty of measurement results under practical monitoring conditions.
EN 14625:2024/prA1 is classified under the following ICS (International Classification for Standards) categories: 13.040.20 - Ambient atmospheres. The ICS classification helps identify the subject area and facilitates finding related standards.
EN 14625:2024/prA1 has the following relationships with other standards: It is inter standard links to EN 14625:2024. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
EN 14625:2024/prA1 is associated with the following European legislation: EU Directives/Regulations: 2008/50/EC. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
EN 14625:2024/prA1 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
SLOVENSKI STANDARD
01-december-2026
Zunanji zrak - Standardna metoda za določanje koncentracije ozona z
ultravijolično fotometrijo - Dopolnilo A1
Ambient air - Standard method for the measurement of the concentration of ozone by
ultraviolet photometry
Außenluft - Messverfahren zur Bestimmung von Ozon in Luft mit dem UV-Verfahren
Air ambiant - Méthode normalisée de mesurage de la concentration en ozone par
photométrie U.V.
Ta slovenski standard je istoveten z: EN 14625:2024/prA1
ICS:
13.040.20 Kakovost okoljskega zraka Ambient atmospheres
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
DRAFT
EUROPEAN STANDARD
EN 14625:2024
NORME EUROPÉENNE
EUROPÄISCHE NORM
prA1
October 2026
ICS
English Version
Ambient air - Standard method for the measurement of the
concentration of ozone by ultraviolet photometry
Air ambiant - Méthode normalisée de mesurage de la Außenluft - Messverfahren zur Bestimmung von Ozon
concentration en ozone par photométrie U.V. in Luft mit dem UV-Verfahren
This draft amendment is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee CEN/TC 264.
This draft amendment A1, if approved, will modify the European Standard EN 14625:2024. If this draft becomes an amendment,
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for inclusion of
this amendment into the relevant national standard without any alteration.
This draft amendment was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Türkiye and
United Kingdom.
Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.
Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2026 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 14625:2024/prA1:2026 E
worldwide for CEN national Members.
Contents Page
European foreword . 3
1 Modification to Clause 1 ”Scope” . 4
2 Modification to Clause 3 "Terms and definitions" . 4
3 Modifications to 6.3.2, "Particle filter" . 4
4 Modifications to 6.6 “Sampling pump to manifold” . 4
5 Modifications to 8.2, "Relevant performance characteristics and criteria" . 5
6 Modification to 8.4.2.3, "Test gases" . 5
7 Modification to 8.4.6, "Lack of fit of linearity of the calibration function” . 5
8 Modification to 8.5.4, "Long-term drift" . 6
9 Modification to 9.5.4, “Testing the sampling system” . 6
10 Modification to 9.5.5, "Treatment of data after exceedance of performance criteria" . 6
11 Modification to 9.6.1.2, "Performance of zero and span checks" . 6
12 Modification to 9.7.3, "Testing the sample manifold" . 6
13 Modifications to D.1, "Procedure for applying test gas", Formula D.1 & D.2 . 6
14 Modification to E.3.2.5.5, "Electric voltage" . 7
15 Modification to F.2, "Combined standard uncertainty" . 7
European foreword
This document (EN 14625:2024/prC1:2026) has been prepared by Technical Committee CEN/TC 264
“Air quality”, the secretariat of which is held by DIN.
This document is currently submitted to the CEN Enquiry.
This amendment addresses only the critical errors identified in the EN 14625:2024. Errors considered
as minor by CEN/TC 264/WG 12 will be addresse
...