Abstract

This document specifies a method for the determination of the boiling range distribution of petroleum products by capillary gas chromatography using flame ionization detection. The document is applicable to materials having a vapour pressure low enough to permit sampling at ambient temperature and a boiling range of at least 100 °C. The document is applicable to distillates with initial boiling points (IBP) above 100 °C and final boiling points (FBP) below 750 °C, for example, middle distillates and lubricating base stocks.
The test method is not applicable for the analysis of petroleum or petroleum products containing low molecular weight components (for example naphtha’s, reformates, gasolines) or middle distillates like Diesel and Jet fuel.
Petroleum or petroleum products containing blending components which contain heteroatoms (for example alcohols, ethers, acids, or esters) or residue are not to be analysed by this test method.
NOTE   For the purposes of this document, the terms “% (m/m)” and “% (V/V)” are used to represent respectively the mass fraction and the volume fraction.
WARNING — The use of this document can involve hazardous materials, operations and equipment. This document does not purport to address all of the safety problems associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use.

Status
Not Published
Publication Date
24-Feb-2027
Current Stage
5020 - Submission to Vote - Formal Approval
Start Date
05-Nov-2026
Due Date
19-Aug-2027
Completion Date
19-Aug-2027

Buy Documents

FprEN 15199-1 is a draft European Standard from CEN for determining the boiling range distribution of petroleum products by capillary gas chromatography with flame ionization detection. It is intended for laboratories and technical users working with middle distillates and lubricating base oils that can be sampled at ambient temperature and have a boiling range of at least 100 °C. The document tells users how to sample, calibrate, run the test, calculate the distribution, and report the result.

What does FprEN 15199-1 specify?

FprEN 15199-1 specifies a gas chromatographic method for boiling range distribution in petroleum products, focused on middle distillates and lubricating base oils. It sets the product range, excludes several material types, and lays out the document structure from principle through test report.

The scope limits the method to materials with vapour pressure low enough for ambient sampling, an initial boiling point above 100 °C, and a final boiling point below 750 °C. It excludes low molecular weight products such as naphtha, reformates and gasolines, and it also excludes petroleum products containing heteroatom blending components such as alcohols, ethers, acids or esters, as well as residue.

The document is organized into clauses on scope, normative references, terms and definitions, principle, reagents and materials, apparatus, sampling, apparatus preparation, sample preparation, calibration, procedure, visual inspection, calculation, expression of results, precision, and test report. It also has three normative annexes.

AnnexWhat it covers
Annex ACalculation procedure, including blank subtraction, start and end of sample elution, normalization, and conversion from retention time to boiling point
Annex BSystem performance check, including column resolution, detector response, and peak skewness
Annex CBoiling points of n-alkanes used for calibration

What are the key requirements of FprEN 15199-1?

FprEN 15199-1 is built around a controlled chromatographic sequence: prepare the sample correctly, verify the system, run blank and calibration checks, then convert retention time data into boiling range distribution. In practice, the method depends on stable baseline behavior, verified calibration, and consistent sample handling.

Sampling and sample preparation

FprEN 15199-1 requires sampling in accordance with EN ISO 3170 or EN ISO 3171. Samples are to be stored in glass or metal containers, because plastic containers can contaminate the product through leaching.

The sample is prepared as a 2 % to 3 % mass concentration solution in carbon disulfide (CS). The vial is capped immediately after transfer. For a laboratory, this means the dilution, solvent choice, and container material all affect result quality.

Calibration and system readiness

FprEN 15199-1 recommends daily calibration checks before sample analysis. The first run of the day shall not be a blank, reference material, or sample, because injector carryover can affect the result.

The calibration mixture uses n-alkanes and polyethylene wax solution, and the retention time versus boiling point curve is built from those compounds. Annex C provides the boiling points of the n-alkanes used for that curve. This is what links chromatographic time to boiling point.

Instrument setup and column control

FprEN 15199-1 requires a gas chromatograph with flame ionization detection, a linear temperature program, and a sample inlet system of either programmable temperature vaporizer (PTV) or cold on-column (COC) type. The column resolution shall be between 2 and 4, and the carrier gas flow shall remain constant over the temperature range.

The method also calls for column bleed compensation to maintain a stable baseline. In practice, this matters because drifting baselines can distort the start and end points of the chromatogram and change the final boiling point.

Performance check

FprEN 15199-1 uses Annex B to check system performance. The checks cover column resolution with the C50 and C52 peaks, detector response with a binary gravimetric blend, and peak skewness for the C50 peak.

These checks are not just formalities. They tell the user whether the instrument is behaving consistently enough for valid sample analysis, and whether troubleshooting is needed before testing continues.

Running the analysis

FprEN 15199-1 requires a solvent blank before the first sample and then after every five samples. The blank and sample baselines are merged, and the run is acceptable only if the end of the sample chromatogram merges with the blank baseline.

The document also requires visual inspection of the chromatograms at 5 times expansion. That helps the analyst confirm the start and end of the area of interest and verify that the solvent and sample signals are properly separated.

Calculation and reporting

FprEN 15199-1 uses Annex A for the calculation protocol. The method works slice by slice: subtract the blank, correct for baseline offset, sum the corrected area, and convert cumulative area percentages into retention times and then boiling points.

The initial boiling point is reported at 0.5 % and the final boiling point at 99.5 %. Intermediate values are reported at intervals of not less than 1 % mass fraction, with temperatures rounded to the nearest 1 °C and percentages to the nearest 1 % (m/m). For a buyer or quality manager, this gives a standardized boiling distribution profile that can be compared from one lab to another.

Precision

FprEN 15199-1 includes repeatability and reproducibility values based on an inter-laboratory study. The precision section states that the study was carried out with helium as carrier gas, and that nitrogen or hydrogen are allowed but the precision in this clause cannot be applied to them.

This matters because carrier gas choice can change bias. Users who switch away from helium need to treat the precision data with caution.

What terms does FprEN 15199-1 define?

  • Initial boiling point (IBP) - the temperature corresponding to the retention time at which 0.5 % of the total sample area has been reached.
  • Final boiling point (FBP) - the temperature corresponding to the retention time at which 99.5 % of the total sample area has been reached.
  • Area slice - the integrated detector signal over a specified retention time interval.
  • Corrected area slice - an area slice after subtraction of the corresponding blank slice.
  • Total sample area - the cumulative corrected area from the initial area point to the final area point after complete elution.
  • Net area - the cumulative area count for the sample minus the cumulative area count for the blank.
  • Recovery - the ratio of the sample signal to the reference material signal, corrected for dilution, weights, and light ends if applicable.

Who uses FprEN 15199-1?

FprEN 15199-1 is used by laboratories that test petroleum products, especially those handling middle distillates and lubricating base stocks. It is also relevant to quality control staff, production laboratories, and technical buyers who need a comparable boiling range distribution for product release or incoming inspection.

Instrument operators use it to set up the GC, check baseline behavior, and verify calibration. Quality managers use it to compare laboratory results, document deviations, and keep the test report consistent across batches and sites.

What changed in FprEN 15199-1 from the previous edition?

FprEN 15199-1 will supersede EN 15199-1:2020. The foreword states two main changes: allowance of alternative carrier gases, and alignment of the column resolution requirement in the text with ASTM D6352.

Which standards are used with FprEN 15199-1?

  • EN ISO 3170 - manual sampling of hydrocarbon liquids, used when samples are taken by hand.
  • EN ISO 3171 - automatic pipeline sampling of petroleum liquids, used for inline or automated sampling.
  • EN 15199-2 - the companion part for heavy distillates and residual fuels with final boiling points above 750 °C.
  • EN 15199-3 - the companion part for crude oils, including boiling range distribution and recovery up to C100/C120.
  • EN 15199-4 - the companion part for light fractions of crude oil up to n-nonane.
  • EN ISO 3924 - cited as another SIMDIS method and as a source of a suitable alkane calibration mixture.
  • IP 480 and ASTM D6352 - methods the document says it is based on and kept aligned with.

What does the FprEN 15199-1 document contain?

FprEN 15199-1 contains the full test workflow from principle to reporting, supported by chromatographic figures, worked calculation procedures, and precision data. Clause 4 explains the separation principle, while Clauses 5 and 6 specify reagents, materials, and apparatus, including a gas chromatograph with FID and typical operating conditions in Table 2.

The document includes chromatogram figures for a typical sample, a calibration mixture, baseline behavior, and peak shape. Annex A contains the calculation formulas for blank correction, start and end of sample elution, normalization, and conversion from retention time to boiling point.

Annex B gives the system performance checks used for routine readiness and troubleshooting. Annex C provides the n-alkane boiling points needed to build the calibration curve.

Relations

Effective Date
03-Sep-2025

Buy Documents

Get Certified

Connect with accredited certification bodies for this standard

Element Materials Technology

Materials testing and product certification.

UKAS United Kingdom Verified

ABS Group Brazil

ABS Group certification services in Brazil.

CGCRE Brazil Verified

ABS Quality Evaluations Inc.

American Bureau of Shipping quality certification.

ANAB United States Verified

Sponsored listings

Frequently Asked Questions

FprEN 15199-1 is a draft published by the European Committee for Standardization (CEN). Its full title is "Petroleum products - Determination of boiling range distribution by gas chromatography method - Part 1: Middle distillates and lubricating base oils". This standard covers: This document specifies a method for the determination of the boiling range distribution of petroleum products by capillary gas chromatography using flame ionization detection. The document is applicable to materials having a vapour pressure low enough to permit sampling at ambient temperature and a boiling range of at least 100 °C. The document is applicable to distillates with initial boiling points (IBP) above 100 °C and final boiling points (FBP) below 750 °C, for example, middle distillates and lubricating base stocks. The test method is not applicable for the analysis of petroleum or petroleum products containing low molecular weight components (for example naphtha’s, reformates, gasolines) or middle distillates like Diesel and Jet fuel. Petroleum or petroleum products containing blending components which contain heteroatoms (for example alcohols, ethers, acids, or esters) or residue are not to be analysed by this test method. NOTE For the purposes of this document, the terms “% (m/m)” and “% (V/V)” are used to represent respectively the mass fraction and the volume fraction. WARNING — The use of this document can involve hazardous materials, operations and equipment. This document does not purport to address all of the safety problems associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use.

This document specifies a method for the determination of the boiling range distribution of petroleum products by capillary gas chromatography using flame ionization detection. The document is applicable to materials having a vapour pressure low enough to permit sampling at ambient temperature and a boiling range of at least 100 °C. The document is applicable to distillates with initial boiling points (IBP) above 100 °C and final boiling points (FBP) below 750 °C, for example, middle distillates and lubricating base stocks. The test method is not applicable for the analysis of petroleum or petroleum products containing low molecular weight components (for example naphtha’s, reformates, gasolines) or middle distillates like Diesel and Jet fuel. Petroleum or petroleum products containing blending components which contain heteroatoms (for example alcohols, ethers, acids, or esters) or residue are not to be analysed by this test method. NOTE For the purposes of this document, the terms “% (m/m)” and “% (V/V)” are used to represent respectively the mass fraction and the volume fraction. WARNING — The use of this document can involve hazardous materials, operations and equipment. This document does not purport to address all of the safety problems associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use.

FprEN 15199-1 is classified under the following ICS (International Classification for Standards) categories: 75.080 - Petroleum products in general. The ICS classification helps identify the subject area and facilitates finding related standards.

FprEN 15199-1 has the following relationships with other standards: It is inter standard links to EN 15199-1:2020. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.

FprEN 15199-1 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.

Standards Content (Sample)

Contents

  • 1 Scope
  • 2 Normative references
  • 3 Terms and definitions
  • 3.1 initial boiling point
  • 3.2 final boiling point
  • 3.3 area slice
  • 3.4 corrected area slice
  • 3.5 cumulative corrected area
  • 3.6 total sample area
  • 2 IBP (3.1) X retention time (minutes)
  • 3 FBP (3.2) Y response (pA)
  • 3.7 net area
  • 3.8 recovery
  • 4 Principle
  • 5 Reagents and materials
  • 5.7 Calibration mixture
  • 5.8 Reference materials
  • 6 Apparatus
  • 6.2 Column
  • 6.3 Carrier gas control
  • 7 Sampling
  • 8 Preparation of the apparatus
  • 8.1 Gas chromatograph preparation
  • 8.2 System performance check
  • 9 Sample preparation
  • 10 Calibration
  • 11 Procedure
  • 12 Visual inspection of the chromatograms
  • 13 Calculation
  • 14 Expression of results
  • 15 Precision
  • 15.1 General
  • 15.2 Repeatability
  • 15.3 Reproducibility
  • 16 Test report
  • Annex A
  • Annex B
  • Annex C
  • Bibliography

1 Scope

This document specifies a method for the determination of the boiling range distribution of petroleum products by capillary gas chromatography using flame ionization detection. The document is applicable to materials having a vapour pressure low enough to permit sampling at ambient temperature and a boiling range of at least 100 °C. The document is applicable to distillates with initial boiling points (IBP) above 100 °C and final boiling points (FBP) below 750 °C, for example, middle distillates and lubricating base stocks.

The test method is not applicable for the analysis of petroleum or petroleum products containing low molecular weight components (for example naphtha’s, reformates, gasolines) or middle distillates like Diesel and Jet fuel.

Petroleum or petroleum products containing blending components which contain heteroatoms (for example alcohols, ethers, acids, or esters) or residue are not to be analysed by this test method. NOTE For the purposes of this document, the terms “% (m/m)” and “% (V/V)” are used to represent respectively the mass fraction and the volume fraction.

WARNING — The use of this document can involve hazardous materials, operations and equipment. This document does not purport to address all of the safety problems associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use.

2 Normative references

The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. EN ISO 3170, Hydrocarbon Liquids — Manual Sampling (ISO 3170) EN ISO 3171, Petroleum liquids — Automatic pipeline sampling (ISO 3171)

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:

  • ISO Online browsing platform: available at https://www.iso.org/obp
  • IEC Electropedia: available at https://www.electropedia.org/

3.1 initial boiling point

IBP temperature corresponding to the retention time at which a net area counts equal to 0,5 % of the total sample area (3.6) under the chromatogram is obtained (see Figure 1)

3.2 final boiling point

FBP temperature corresponding to the retention time at which a net area (3.7) counts equal to 99,5 % of the total sample area (3.6) under the chromatogram is obtained (see Figure 1)

oSIST prEN 15199-1:2026 prEN 15199-1:2026 (E)

3.3 area slice

area resulting from the integration of the chromatographic detector signal within a specified retention time interval Note 1 to entry: In area slice mode peak detection parameters are bypassed and the detector signal integral is recorded as area slices of consecutive, fixed duration time interval.

3.4 corrected area slice

area slice (3.3) corrected for baseline offset by subtraction of the exactly corresponding area slice in a previously recorded blank (non-sample) analysis

3.5 cumulative corrected area

accumulated sum of corrected area slices (3.4) from the beginning of the analysis through a given retention time, ignoring any non-sample area for example of solvent

3.6 total sample area

cumulative corrected area (3.5), from the initial area point to the final area point, where the chromatographic signal has returned to baseline after complete sample elution

Key 1 start of elution 4 end of elution

2 IBP (3.1) X retention time (minutes)

3 FBP (3.2) Y response (pA)

Figure 1 — Typical chromatogram

3.7 net area

cumulative area counts for the sample minus the cumulative area count for the blank

oSIST prEN 15199-1:2026 prEN 15199-1:2026 (E)

3.8 recovery

ratio of the cumulative area count of the sample to that of the reference material (external standard) corrected for dilution and material weights combined with the percentage of light ends, if applicable

4 Principle

A test portion is introduced into a gas chromatographic column, which separates hydrocarbons in the order of increasing boiling point. The column temperature is raised at a linear reproducible rate and the area under the chromatogram is recorded throughout the analysis. Boiling points are assigned to the time-axis from a calibration curve obtained by running a mixture of known n-alkanes, covering the sample boiling range, under the same conditions. From these data, the boiling range distribution is obtained.

Several SIMDIS methods are standardized test methods. Each one is dedicated to a certain boiling point range or product.

EN ISO 3924 [1] is limited to products having an initial boiling point greater than 55 °C, a final boiling point lower than 538 °C and having a vapour pressure sufficiently low to permit sampling at ambient temperature.

EN 15199-2 is applicable to materials with initial boiling points (IBP) above 100 °C and final boiling points (FBP) above 750 °C, for example, heavy distillate fuels and residuals. The method is not applicable to bituminous samples.

EN 15199-3 is applicable to crude oils. The boiling range distribution and recovery (3.8) up to C or C 100 120 can be determined.

5 Reagents and materials

Unless otherwise stated, only chemicals of recognized analytical quality shall be used. 5.1 Carrier gas, helium, nitrogen or hydrogen, of at least 99,999 % (V/V) purity. Any oxygen present is removed by a chemical resin filter.

WARNING — Follow the safety instructions from the filter supplier. CAUTION — Helium and nitrogen are compressed gases under high pressure. Hydrogen is an extremely flammable gas under high pressure.

5.2 Hydrogen, grade suitable for flame ionization detectors.

5.3 Compressed air, suitable for flame ionization detectors.

5.4 Alkanes, n-alkanes of at least 98 % (m/m) purity from C to C , C , C , C , C , C , C and C . 5 10 12 14 16 18 20 24 28 NOTE The calibration mixture from EN ISO 3924 [1] is also suitable.

5.5 Polyethylene wax solution .

Polywax is the tradename of fully saturated homopolymers of ethylene supplied by Baker Hughes Holdings LLC. This information is given for the convenience of users of this document and does not constitute an endorsement by CEN of this (these) product(s).

oSIST prEN 15199-1:2026 prEN 15199-1:2026 (E) 5.6 Carbon disulfide, (CS ) purity 99,7 % (V/V) minimum.

WARNING — Extremely flammable and toxic by inhalation.

CAUTION — It is recommended that all work with CS is carried out in an explosion protected fume

cupboard.

Cyclohexane (C H )—(>99 % pure) may be used in place of CS for the preparation of the calibration 6 12 2 mixture. However, the precision of this method is based on calibration mixtures, reference material and samples prepared with CS only.

5.7 Calibration mixture

Dissolve 0,1 g of Polyethylene wax solution (5.5) in 7 ml CS (5.6), warming gently if necessary. Prepare

an equal volume mixture of alkanes (5.4) and add 10 µl to the Polyethylene wax solution. NOTE 1 Commercially available alkane standards are suitable for column performance checks. NOTE 2 The calibration mix is used to determine the column resolution, skewness of the C peak, and retention

time versus boiling point calibration curve.

5.8 Reference materials

5.8.1 A reference material has two functions:

  • External Standard: to determine the recovery of samples by comparing the total sample area (3.6) of the reference material with the total sample area (3.5) of the unknown sample;
  • Boiling Point Distribution Standard: to check the proper functioning of the system by comparing the results with a known boiling point distribution on a routine basis. A typical example is given in (5.8.2).
5.8.2 Reference Material 5010, a reference sample that has been analysed by laboratories

participating in the test method cooperative study. Consensus values for the boiling range distribution of this sample are given in Table 1.

5.8.3 Binary gravimetric blend, a binary distillate mixture with boiling points ranges that gives a baseline at the start, a baseline between the two peaks and an end time that is as close to the end of the chromatogram as possible (see Figure 2 and Clause B.3). This mixture is used to check the relative response of the two distillates and to check the baselines at the start, middle and end of the chromatogram.

oSIST prEN 15199-1:2026 prEN 15199-1:2026 (E)

Key Y response (pA) X retention time (minutes) Figure 2 — Typical chromatogram of binary gravimetric blend distillate Table 1 — Reference material 5010 Allowable difference % Recovery Accepted reference value 95,5 % confidence interval °C °C IBP 428 9 5 477 3 10 493 3 15 502 3 20 510 3 25 518 4 30 524 4 35 531 4 40 537 4 45 543 4 50 548 5 55 554 4 60 560 4 65 566 4 70 572 4 75 578 5 80 585 4 85 593 4 90 602 4

oSIST prEN 15199-1:2026 prEN 15199-1:2026 (E) Allowable difference % Recovery Accepted reference value 95,5 % confidence interval °C °C 95 616 4 FBP 655 18

6 Apparatus

6.1 Gas chromatograph, with the following performance characteristics. 6.1.1 Flame ionization detector, connected to the column to avoid any cold spots. The detector shall be capable of operating at a temperature at least equivalent to the maximum column temperature employed in the method.

6.1.2 Column temperature programmer, capable of linear programmed temperature operation over

a range of 10 °C above ambient to 450 °C.

6.1.3 Sample inlet system, consisting of a programmable temperature vaporizer (PTV) or cold on-

column (COC) injection port. The maximum temperature of the injection device shall be equal to, or higher than, the final oven temperature. The minimum temperature shall be low enough to prevent sample or solvent flashback, but high enough to allow sample focusing at the front of the column. Table 2 contains the typical operating conditions.

6.2 Column

6.2.1 The capillary column should sit just below the flame tip and it is recommended that the orifice of the jet should be 0,6 mm minimum to prevent frequent blocking with silicones.

6.2.2 Use a metal column, 0,53 µm internal diameter coated with methyl silicone. Commercially

available columns with film thickness (d ) = 0,09 µm (for analysis up to C ) and (d ) = 0,17 µm (for f 120 f analysis up to C ) have been found to be satisfactory.

The column resolution, R, shall be at least 2 and not more than 4 (see Clause B.2). 6.2.3 Use some form of column bleed compensation to obtain a stable baseline. This may be carried out by subtraction of a column bleed profile previously obtained using exactly the same conditions as used for the sample analysis, by injecting the same volume, using solvent for the blank run and sample dilution from one batch taken at the same time, to avoid differences due to contamination. Table 2 — Typical operating conditions for gas chromatograph Unit Specification Column length m 5 Column internal diameter mm 0,53 Column material — Ultimetal Stationary phase — Methyl silicone Film thickness µm 0,09 or 0,17 Initial column temperature °C 35

oSIST prEN 15199-1:2026 prEN 15199-1:2026 (E) Unit Specification Final column temperature °C 430 Program rate °C/min 10 Injector initial temperature °C 100 Injector final temperature °C 430 Program rate °C/min 15 Hold time min 5 Detector temperature °C 450 Detector hydrogen flow (5.2) ml/min 35 Detector air flow (5.3) ml/min 350 Carrier gas — Helium a Carrier gas flow rate ml/min 19 Sample size µl 1,0 Sample concentration % (m/m) 2 Injector — PTV or COC a A carrier gas flow rate up to 25 ml/min may be used to ensure all material elutes before the end of the temperature program.

6.3 Carrier gas control

The gas chromatograph (6.1) shall be able to deliver a constant carrier gas flow over the whole temperature range of the analysis.

...