prEN ISO 13366-2
(Main)Milk - Enumeration of somatic cells - Part 2: Flow cytometry method (ISO/DIS 13366-2:2026)
- Abstract
This document gives guidance on the operating conditions for counting the total number of somatic cells, in both raw and chemically preserved milk, using flow cytometers.
The guidance is applicable to the counting of somatic cells in raw cow milk. The guidance is also applicable to raw milk of other species, such as goat, sheep and buffalo, if the specified prerequisites are met.
- Status
- Not Published
- Publication Date
- 05-Dec-2027
- Technical Committee
- CEN/TC 302 - Milk and milk products - Methods of sampling and analysis
- Current Stage
- 4060 - Closure of enquiry - Enquiry
- Start Date
- 07-Sep-2026
- Due Date
- 07-Sep-2026
- Completion Date
- 07-Sep-2026
prEN ISO 13366-2 is a CEN draft standard for enumerating somatic cells in milk by the flow cytometry method. It gives guidance on the operating conditions for counting total somatic cells in raw milk and chemically preserved milk, and it is intended for laboratories and dairy operators that need routine somatic cell counts, calibration, and quality control. The guidance applies directly to raw cow milk and also to raw goat, sheep and buffalo milk when the stated prerequisites are met.
What does prEN ISO 13366-2 specify?
prEN ISO 13366-2 specifies how to run flow cytometer-based counting of total somatic cells in milk, with attention to sample handling, calibration, routine checks, precision, and reporting. Sampling itself is outside the method and the document points users to ISO 707 | IDF 50 for sampling guidance.
The document is organized into clauses on scope, normative references, terms and definitions, principle, factors affecting results, calibration, sampling, determination, routine performance checks, species-specific remarks, precision, and test report. It also includes one informative annex.
| Annex | What it covers |
|---|---|
| Annex A | Direct labelling (DNA stain) for total cell counting in milk using flow cytometry, including a non-dedicated flow cytometer approach |
What are the key requirements of prEN ISO 13366-2?
Clause 4 sets out the measurement principle: the milk sample is mixed with buffer and stain, then passed through the counting section of a flow cytometer where stained particles generate electrical pulses that are processed into a cell count. In practice, this means the result depends on correct reagent dosing, mixing, optical detection, and signal discrimination.
Clause 5 focuses on everything that can change the count before analysis. The sample bottle should be suitable for transport, samples should be cooled promptly, freezing should be avoided, and chemical preservation should only be used with validated preservatives and colour tracers. For the reader, this means the pre-analytical chain is part of the method - poor handling can lower the count or distort the signal.
Clause 5 also distinguishes between sample types and practical risks. Bulk milk needs thorough mixing because cells can concentrate in the upper layer, while milk from individual animals should be representative of the whole milking when the aim is a whole-milking count. Problem samples such as clotted, dirty, high-erythrocyte, colostrum, late lactation or sour milk may require special care or avoidance.
Clause 6 covers calibration and reference materials. Reference materials should be produced under a quality system meeting ISO 17034, and the document recognizes certified reference material (CRM), secondary reference material (SRM), and in-house reference material (IRM). Calibration may be built from prepared milk samples made by adding leucocyte suspension, microfiltration, or centrifugation, and the calibration should cover the relevant counting range.
Clause 6 also requires a linearity check and regular calibration verification. The instrument should be checked for linear response in the working range, and the calibration should be checked at least monthly. In practice, a laboratory should not assume that one calibration is permanent - it needs ongoing confirmation that the instrument still tracks the reference values correctly.
Clause 9 defines the routine quality checks that keep the method under control. It covers blank checks, carry-over checks, pilot checks, repeatability, intralaboratory reproducibility, and external comparisons through proficiency testing. These checks matter because they reveal contamination, sample-to-sample influence, drift, and day-to-day stability problems before they affect reported milk results.
Clause 10 gives species-specific remarks. Goat milk, sheep milk and buffalo milk may need special attention because cell size, fat, protein, or background signals can differ from cow milk. The document allows cow-milk calibration to be used for goat and sheep milk on some instruments when the relevant range is covered, but only if the instrument response remains suitable for that milk type.
Clause 11 provides precision information for repeatability, intralaboratory reproducibility, and reproducibility. The values are presented for cow milk, with a note that other species may be less favourable. For buyers and laboratories, this section is useful when setting acceptance criteria, comparing instruments, or judging whether a result is consistent.
Clause 12 states what the test report should contain. A report should identify the sample, note the sampling method if known, give the sampling date and sample type, identify the method used, record any operating details that may have influenced the result, and state the result in thousands of cells per millilitre or the final quoted result after repeatability checking.
What terms does prEN ISO 13366-2 define?
- reference method - the somatic cell counting method described in ISO 13366-1 | IDF 148-1, used as the benchmark for comparison and traceability.
- somatic cells - cells that give fluorescence above the defined threshold after DNA staining in their nuclei, with the count expressed in cells per millilitre.
- certified reference material (CRM) - a reference material with certified values used to anchor calibration and traceability.
- secondary reference material (SRM) - a reference material prepared by an external supplier and traceable to CRMs.
- in-house reference material (IRM) - a laboratory-prepared reference material whose traceability is maintained through CRMs or SRMs.
- direct total somatic cell counts (dSCC) - the alternative method in Annex A that uses direct DNA staining and a non-dedicated flow cytometer.
Who uses prEN ISO 13366-2?
prEN ISO 13366-2 is used by dairy laboratories, milk-testing laboratories, and quality teams that measure somatic cell counts in routine milk analysis. It is also relevant to milk producers, processors, and buyers who depend on SCC results for raw milk monitoring, acceptance decisions, and quality control.
Instrument operators use it to set up flow cytometers, prepare samples, and run blanks, pilot samples, and repeatability checks. Calibration and reference material suppliers, as well as proficiency testing providers, also use it because the document ties the method to CRM/SRM/IRM traceability and interlaboratory comparison.
What changed in prEN ISO 13366-2 from the previous edition?
This third edition replaces the second edition ISO 13366-2 | IDF 148-2:2006 and is technically revised. The foreword states these main changes:
- the outdated rotating disk technique has been deleted;
- guidance has been added on building a robust metrological system, including use of a certified reference material (CRM) and assignment of values to secondary reference material;
- control of the ratio/reagent has been removed;
- Annex A adds a method for total somatic cell counts (dSCC) in milk using a non-dedicated flow cytometer and commercially available non-proprietary chemicals.
Which standards are used with prEN ISO 13366-2?
| Reference | What it contributes |
|---|---|
| ISO 8196-1 | Analytical attributes for alternative milk analysis methods and overall accuracy concepts |
| ISO 8196-2 | Calibration and quality control in the dairy laboratory, including control charts and calibration practice |
| ISO 8196-3 | Protocol for evaluating and validating alternative quantitative methods, including intralaboratory reproducibility |
| ISO 13366-1 | The microscopic reference method for somatic cell counting and the basis for traceable reference values |
| ISO 17034 | Competence requirements for producers of reference materials |
| ISO/IEC 17043:2023 | Requirements for proficiency testing providers used in external comparisons |
| ISO 707 | Guidance on sampling, which Clause 7 points to for sample collection practice |
| IDF 128-1, IDF 128-2, IDF 128-3, IDF 148-1 | IDF counterparts of the ISO milk-analysis standards used alongside the method |
What does the prEN ISO 13366-2 document contain?
The document contains clause-level guidance for running somatic cell counts by flow cytometry, including the factors that affect results, calibration routes, routine control checks, species-specific remarks, and reporting rules. It also includes tables for calibration ranges and precision, plus formulas for carry-over and linearity evaluation.
Annex A adds a worked alternative approach for direct labelling with a DNA stain. That annex describes the sample-preparation sequence, instrument settings, reagents, gating approach, calculation of results, and a preliminary precision comparison, which makes it useful for method development or instrument adaptation.
Relations
- Effective Date
- 14-May-2025
- Effective Date
- 14-May-2025
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

Bureau Veritas
Bureau Veritas is a world leader in laboratory testing, inspection and certification services.

DNV
DNV is an independent assurance and risk management provider.
Sponsored listings
Frequently Asked Questions
prEN ISO 13366-2 is a draft published by the European Committee for Standardization (CEN). Its full title is "Milk - Enumeration of somatic cells - Part 2: Flow cytometry method (ISO/DIS 13366-2:2026)". This standard covers: This document gives guidance on the operating conditions for counting the total number of somatic cells, in both raw and chemically preserved milk, using flow cytometers. The guidance is applicable to the counting of somatic cells in raw cow milk. The guidance is also applicable to raw milk of other species, such as goat, sheep and buffalo, if the specified prerequisites are met.
This document gives guidance on the operating conditions for counting the total number of somatic cells, in both raw and chemically preserved milk, using flow cytometers. The guidance is applicable to the counting of somatic cells in raw cow milk. The guidance is also applicable to raw milk of other species, such as goat, sheep and buffalo, if the specified prerequisites are met.
prEN ISO 13366-2 is classified under the following ICS (International Classification for Standards) categories: 67.100.10 - Milk and processed milk products. The ICS classification helps identify the subject area and facilitates finding related standards.
prEN ISO 13366-2 has the following relationships with other standards: It is inter standard links to EN ISO 13366-2:2006, EN ISO 13366-2:2006/AC:2007. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
prEN ISO 13366-2 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 reference method
- 3.2 somatic cells
- 4 Principle
- 5 Factors affecting the results of measurements
- 5.1 Sample bottles
- 5.2 Sampling
- 5.3 Preservation
- 5.4 Sample storage and transport
- 5.5 Interfering substances
- 5.6 Sample quality at analysis
- 5.7 Chemicals
- 5.8 Instrument condition
- 6 Calibration
- 6.1 Reference materials
- 6.2 Calibration procedure
- 7 Sampling
- 8 Determination
- 9 Checks on performance in routine operation
- 9.1 Blank checks
- 9.2 Carry-over effect
- 9.3 Pilot checks
- 9.4 Additional instrument surveillance
- 9.5 Repeatability
- 9.6 Intralaboratory reproducibility
- 9.7 External comparisons
- 10 Specific remarks for the use with milk from different species
- 10.1 General
- 10.2 Cow milk
- 10.3 Goat milk
- 10.4 Sheep milk
- 10.5 Buffalo milk
- 11 Precision
- 11.1 Repeatability
- 11.2 Intralaboratory reproducibility
- 11.3 Reproducibility
- 12 Test report
- Annex A
- Bibliography
1 Scope
This document gives guidance on the operating conditions for counting the total number of somatic cells, in both raw and chemically preserved milk, using flow cytometers. The guidance is applicable to the counting of total somatic cells in raw cow milk. The guidance is also applicable to raw milk of other species, such as goat, sheep and buffalo, if the specified prerequisites are met.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 8196-1, | IDF 128-1, Milk — Definition and evaluation of the overall accuracy of alternative methods of milk analysis — Part 1: Analytical attributes of alternative methods ISO 8196-2, | IDF 128-2, Milk — Definition and evaluation of the overall accuracy of alternative methods of milk analysis — Part 2: Calibration and quality control in the dairy laboratory ISO 13366-1, | IDF 148-1, Milk — Enumeration of somatic cells — Part 1: Microscopic method (Reference method) ISO 17034, General requirements for the competence of reference material producers ISO/IEC 17043:2023, Conformity assessment — General requirements for the competence of proficiency testing providers
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
3.1 reference method
method described in ISO 13366-1 | IDF 148-1 for the counting of somatic cells
3.2 somatic cells
those cells that show more than a threshold intensity of fluorescence due to the staining of DNA in their nuclei Note 1 to entry: The number of somatic cells is expressed in cells per millilitre.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en)
4 Principle
Flow cytometers designed for automatically counting the total number of somatic cells in milk have functions for the uptake of reagents and test sample, a mixing section and a counting section. In the mixing section, the test sample is mixed with a buffer and a stain solution. Part of the resulting mixture is transferred to the counting section.
For other flow cytometers, the steps described above can follow the procedure described in (see Annex A). In the mixing section, closely controlled volumes of test sample and buffer/stain solutions are dosed and mixed. Mixing can take place in a cup, a mixing chamber, or in the tubing leading to a flow cell. In the counting section, part of the mixture is placed in the high-speed flow of a surrounding sheath liquid in a capillary flow cell. Through acceleration, the mixture forms a thin string in which the somatic cells are dynamically focused and aligned. This string then passes the objective of a fluorescence microscope. Each stained particle produces an electrical pulse that is filtered, amplified and recorded. The resulting pulse height distribution is electronically processed, whereby discrimination is made between noise signals and pulses that are attributed to stained somatic cells. The discriminator level can either be fixed or dynamic. Some instruments contain two channels in the counting section. In terms of analytical quality assurance, such a situation should be considered equivalent to working with two separate units, so the performance should be evaluated separately for each channel.
5 Factors affecting the results of measurements
5.1 Sample bottles
Sample bottles should be fit for use; i.e. to transfer test samples from the point of sampling to the laboratory without loss or damage.
Care is to be exercised that sample bottles are leak-proof and that a proper empty volume is left. Too large an empty volume can facilitate churning; too small an empty volume can cause problems with mixing.
5.2 Sampling
5.2.1 General
Sampling materials (i.e. sample bottles, beakers and sampling devices) should be clean and dry. Where automatic samplers are used, these should have been properly validated. Test samples should preferably be cooled immediately after sampling at 4 °C ± 2 °C and be kept at that temperature until counting (see 5.4) rather than being preserved. Freezing should be avoided. If preservation is necessary, suitable means for chemical preservation of test samples are described in 5.3.
5.2.2 Bulk milk samples
Thorough mixing of the raw bulk milk to be sampled is essential. Somatic cells will concentrate in the upper layer in the case of insufficient stirring.
5.2.3 Milk samples from individual animals
The release of somatic cells in the milk during milking is uneven. When the aim is to produce a representative counting result for a whole milking, it is essential that a representative sample of the whole milking is obtained. For diagnostic purposes, a sample of a partial milking may suffice.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en)
5.3 Preservation
If chemical preservation is considered necessary, the test sample (5.2.1) should be preserved as soon as possible, but in any case within 24 h after sampling. In all cases, the test sample should be kept cool (4 °C ± 2 °C) until the addition of the preservative.
WARNING — Sodium azide and potassium dichromate are highly hazardous reagents. Handle with extreme care. Prepare and use the reagents only in a properly equipped laboratory, in a fume hood, and always wear appropriate personal protective equipment (PPE), including gloves, lab coat an eye protection.
Suitable preservatives are the following.
a) Boric acid: its final concentration in the test sample should not exceed 0,6 g/100 ml. Such preserved samples may be stored at between 6 °C and 12 °C for up to a further 24 h. b) Sodium azide: its final concentration in the test sample should not exceed 0,024 g/100 ml. Such preserved samples may be stored at between 2 °C and 10 °C for up to a further 72 h. c) Bronopol (2-bromo-2-nitro-1,3-propanediol): its final concentration in the test sample should not exceed 0,05 g/100 ml. Such preserved samples may be stored at between 2 °C and 12 °C for up to a further 6 d. d) Potassium dichromate: its final concentration in the test sample should not exceed 0,2 g/100 ml. Such preserved samples may be stored at between 2 °C and 12 °C for up to a further 6 d. Accompanying colour tracers found to be suitable are:
- Patent Blue V with a final concentration in the test sample of up to 0,15 mg/100 ml;
- Yellow Orange S (E110) with a final concentration in the test sample of up to 1 mg/100 ml;
- a mixture of Patent Blue V and Eosin B with a final concentration in the test sample of up to 0,03 mg/100 ml and 0,45 mg/100 ml, respectively;
- Methylene blue with a final concentration in the test sample of up to 0,5 mg/100 ml. Other preservatives and colour tracers may be used provided that their effectiveness and conditions for use have been soundly validated.
In all cases, properly validate the absence of interference for the counting equipment concerned before application.
In cases where flow cytometers are combined with milk analysers for the measurement of other test sample components, care should be taken that the applied preservatives and colour tracers do not affect the counting result.
5.4 Sample storage and transport
Unpreserved test samples should be stored at 4 °C ± 2 °C and should be counted within 96 h after the completion of sampling. Avoid freezing the test samples. Storage at higher temperatures and/or over longer time scales may result in considerable loss of countable cells and therefore non-representative values. The measurement of samples after freezing and thawing may result in lower counts (by 10 % to 20 %). The age of the samples at freezing and the type of thawing process can influence the counting result.
5.5 Interfering substances
The use of substances that interfere in the counting should be avoided. Substances known to influence the instrument read-out are preservatives and colour tracers at higher concentrations than specified 5.3.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en)
5.6 Sample quality at analysis
Breakdown of somatic cells (lysis) will result in an increase of smaller cell fragments. The lower intensity of fluorescence after staining of these particles causes a shift in the pulse height distribution to the left. This will hamper proper differentiation from noise pulses and therefore result in a lower count. NOTE In several types of instruments, features are available for evaluating the position and the shape of the pulse height distribution. See the relevant instructions and guidance from the instrument manufacturer. After the processing of problematic samples, the flow path should be checked and possible cleaned. The proper functioning of the instrument should be tested before further use. Possible problematic test samples are a) milk samples from severely infected udders, i.e. with clots; b) milk samples with impurities; c) milk samples with high numbers of erythrocytes; d) colostrum; e) late lactation milk; f) sour milk.
Where possible, analysis of problematic test samples should be avoided.
5.7 Chemicals
All reagents used should be of recognized analytical and bacteriological quality. Water used should be demineralized (remaining conductivity < 10 µS/cm) or water of at least equivalent purity. Manufacturer's instructions for the preparation of the working solutions, the maximum storage time and storage requirements should be followed.
Local conditions regarding the use and discharge of applied chemicals and effluents should be observed.
5.8 Instrument condition
5.8.1 General points of attention are:
a) the functioning of the mixing device and the stirrer, b) possible disturbances at sample intake and in the flow system due to blocking by impurities, clots or fouling in the mixing and incubation units, and c) the condition and the functioning of the light source and the photo multiplier, the gain setting and the signal quality.
5.8.2 A specific point of attention with flow cytometers is the variation in the behaviour of the sample string in the flow cell and the sheath liquid flow. Some instrument manufacturers offer special programme features for checking this, thereby indicating possible solutions in the case of deviations.
6 Calibration
6.1 Reference materials
6.1.1 General
Reference materials should be produced under closely controlled conditions; i.e. working with a quality assurance system and fulfilling the requirements as listed in ISO 17034.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en) Reference materials may be [5] a) certified reference materials (CRMs) as produced by a recognized official organization, see ; b) secondary reference materials (SRMs) as prepared by an external supplier, whereby traceability is kept with CRMs or; c) in-house reference materials (IRMs) as prepared by the laboratory itself, whereby traceability is kept with CRMs, SRMs.
6.1.2 Preparation of calibration samples
6.1.2.1 Preparation by addition of a bovine leucocyte suspension
a) Mix 1 000 ml of sterilized or UHT milk with low somatic cell count with 1 ml of polypropylene 2 000 and 0,4 g of bronopol.
b) Add the required amount of a suitable leucocyte suspension to different portions of the mixture in order to obtain a suitable range of cell counts.
6.1.2.2 Preparation by microfiltration
a) Collect fresh bulk milk and add bronopol to a final mass fraction of 0,02 %. b) Skim the milk in a cream separator to a mass fraction of fat of below 0,1 %. c) Concentrate the skimmed milk 20-fold by applying tangential microfiltration over a membrane with a pore size of 1,4 µm, resulting in a portion with a high cell content (HCM) and a portion with a low cell content (LCM).
d) Mix cream, HCM and LCM in the required quantities so as to obtain 5 to 8 milk portions with a mass fraction of fat of 3,5 % and different levels of cell counts covering the range of interest.
6.1.2.3 Preparation by centrifugation
a) Collect fresh bulk milk and add bronopol to a final mass fraction of 0,02 %. b) Filter the thus-prepared milk over a metal filter with a pore size of 0,5 mm. c) Skim the filtered milk by centrifuging at a radial acceleration of 40g for 10 min so as to obtain cream, skimmed milk and a pellet.
d) Mix the skimmed milk and pellet material in the required quantities so as to obtain 5 to 8 milk portions with different cell count levels covering the range of interest. e) If necessary, add cream so as to obtain portions with a mass fraction of fat of 3,3 % ± 0,3 %.
f) Split the obtained portions into individual samples and heat these at 120 °C and 10 Pa for 3 min.
6.1.3 Assignment of reference values
Reference values should be traceable to a CRM that is characterized also with ISO 13366-1|IDF 148-1. An example on how to build traceability is described in (see [5]).
6.1.4 Storage conditions and keeping time
The shelf life, the storage conditions and the keeping time of the prepared secondary reference materials should be properly validated.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en) When chemical preservation is used, alignment should be sought with the type and concentration of the preservative applied to the test sample (see 5.3).
...