oSIST prEN ISO 13366-2:2026
(Main)Milk - Enumeration of somatic cells - Part 2: Flow cytometry method (ISO/DIS 13366-2:2026)
- Abstract
This document gives guidance on the operating conditions for counting the total number of somatic cells, in both raw and chemically preserved milk, using flow cytometers.
The guidance is applicable to the counting of somatic cells in raw cow milk. The guidance is also applicable to raw milk of other species, such as goat, sheep and buffalo, if the specified prerequisites are met.
- Status
- Not Published
- Public Enquiry End Date
- 06-Sep-2026
- Technical Committee
- KŽP - Agricultural food products
- Current Stage
- 4020 - Public enquire (PE) (Adopted Project)
- Start Date
- 30-Jun-2026
- Due Date
- 17-Nov-2026
- Completion Date
- 07-Sep-2026
oSIST prEN ISO 13366-2:2026 is the SIST publication of the ISO/IDF draft on milk somatic cell counting by flow cytometry. It gives guidance on the operating conditions for counting total somatic cells in raw and chemically preserved milk, mainly for raw cow milk and, when the stated prerequisites are met, for goat, sheep and buffalo milk. The document is for laboratories and quality-control users that need to set up, calibrate, monitor, and report this measurement consistently.
What does oSIST prEN ISO 13366-2:2026 specify?
oSIST prEN ISO 13366-2:2026 specifies guidance for the flow cytometric enumeration of somatic cells in milk. It focuses on operating conditions, sample handling, calibration, routine checks, precision, and reporting, rather than on sampling itself.
The scope covers raw milk and chemically preserved milk. It is explicitly applicable to raw cow milk and also to raw goat, sheep, and buffalo milk when the prerequisites in the document are met.
Sampling is not part of the method in this document. For sampling guidance, the standard points users to ISO 707 | IDF 50.
The document is organized into 12 clauses plus one informative annex and a bibliography.
| Annex | What it covers |
|---|---|
| Annex A | Direct labelling (DNA stain) for total cell counting in milk using flow cytometry in milk |
What are the key requirements of oSIST prEN ISO 13366-2:2026?
oSIST prEN ISO 13366-2:2026 is built around controlled sample handling, calibrated instruments, routine performance checks, and documented reporting. In practice, it tells a laboratory what to control so that somatic cell counts stay comparable over time and between instruments.
Sample handling and preservation
Clause 5 says sample bottles should be suitable for transport without loss or damage, and samples should be kept clean, dry, and properly mixed. Bulk milk must be mixed thoroughly before sampling, while samples from individual animals should represent the whole milking when that is the purpose.
The document prefers cooling samples at 4 °C ± 2 °C and avoiding freezing. If chemical preservation is needed, it should be done quickly and validated for the instrument used; the text lists boric acid, sodium azide, bronopol, and potassium dichromate, and warns that sodium azide and potassium dichromate are highly hazardous.
This matters because poor sampling, freezing, or incompatible preservatives can lower or distort the count.
Calibration and traceability
Clause 6 requires reference materials made under a quality system and meeting ISO 17034. It distinguishes certified reference materials (CRMs), secondary reference materials (SRMs), and in-house reference materials (IRMs), all tied back to CRMs for traceability.
Calibration uses at least five samples covering the relevant range, with a monthly check on calibration. The document also requires a linearity check in the working range; if the response bends too much, separate calibrations for different counting ranges may be needed.
This matters because the count must stay proportional to the reference values across the range you actually test.
Routine checks during use
Clause 9 requires blank checks, carry-over checks, pilot milk checks, repeatability checks, intralaboratory reproducibility checks, and external comparisons. The blank check is there to confirm that the flow path is clean; the carry-over check shows whether a high-count sample affects the next one; pilot milk checks track short-term stability.
The document also expects participation in interlaboratory proficiency studies according to ISO 17043 at least twice a year. For a laboratory, this is the practical quality-assurance backbone of the method.
Species-specific use
Clause 10 gives remarks for cow, goat, sheep, and buffalo milk. The standard advises checking for effects from higher fat and protein content and adjusting dilution, stain, buffer, temperature, pass time, or signal processing if needed.
Goat milk needs extra care because its somatic cells are smaller and background noise can be higher. Sheep and buffalo milk may be analyzed under a cow-milk calibration on fluoro-opto-electronic counters if the calibration range covers the relevant cell counts and the instrument is suitable.
Precision and reporting
Clause 11 gives repeatability, intralaboratory reproducibility, and reproducibility values at several cell-count levels. These values are based on cow-milk proficiency studies, so users should be cautious when applying them to other species.
Clause 12 says the test report should identify the sample, note the sampling method if known, record the sampling date and sample type, state the test method, list any deviations that could affect the result, and give the result in thousands of cells per millilitre or the final quoted result if repeatability was checked. That makes the result usable in audits, customer records, and comparisons between laboratories.
What terms does oSIST prEN ISO 13366-2:2026 define?
oSIST prEN ISO 13366-2:2026 defines the terms it relies on so users can apply the method consistently.
- Reference method - the somatic cell counting method described in ISO 13366-1 | IDF 148-1, used as the basis for comparison and traceability.
- Somatic cells - cells that show fluorescence above the stated threshold after DNA staining; the result is expressed in cells per millilitre.
- Certified reference material (CRM) - a reference material produced by a recognized official organization and used to anchor traceability.
- Secondary reference material (SRM) - a reference material prepared by an external supplier and kept traceable to CRMs.
- In-house reference material (IRM) - a reference material prepared by the laboratory itself and kept traceable to CRMs or SRMs.
- Pilot milk - a control milk with an assigned value, used to check short-term instrument stability during routine work.
Who uses oSIST prEN ISO 13366-2:2026?
oSIST prEN ISO 13366-2:2026 is used by milk-testing laboratories, dairy quality-control teams, and metrology staff who run flow cytometers for somatic cell counts. It is also relevant to laboratories working with raw cow milk and, when suitable prerequisites are met, goat, sheep, and buffalo milk.
Instrument users apply it for calibration, blank checks, carry-over checks, repeatability checks, and control-chart monitoring. Proficiency testing providers and laboratories involved in external comparisons also use it as a quality-assurance reference.
What changed in oSIST prEN ISO 13366-2:2026 from the previous edition?
oSIST prEN ISO 13366-2:2026 states that it cancels and replaces ISO 13366-2 | IDF 148-2:2006, and that the text has been technically revised.
The changes named in the document are:
- deletion of the outdated rotating disk technique;
- added guidance for a more robust metrological system using a CRM and for assigning values to secondary reference material;
- removal of the control of ratio/reagent;
- addition of a method for total somatic cell counts in milk using a non-dedicated flow cytometer and commercially available non-proprietary chemicals in Annex A.
Which standards are used with oSIST prEN ISO 13366-2:2026?
oSIST prEN ISO 13366-2:2026 relies on several related ISO and IDF documents for sampling, calibration, traceability, and proficiency testing.
- ISO 8196-1 | IDF 128-1 - defines and evaluates overall accuracy of alternative milk-analysis methods.
- ISO 8196-2 | IDF 128-2 - gives calibration and quality control in the dairy laboratory; used for calibration checks, pilot milk control charts, and repeatability work.
- ISO 13366-1 | IDF 148-1 - the microscopic reference method for somatic cell counting; used for defining the reference method and characterizing reference values.
- ISO 17034 - sets competence requirements for producers of reference materials.
- ISO/IEC 17043:2023 - sets competence requirements for proficiency testing providers.
- ISO 707 | IDF 50 - provides guidance on sampling.
- ISO 8196-3 | IDF 128-3 - is used for intralaboratory reproducibility and validation of alternative quantitative methods.
What does the oSIST prEN ISO 13366-2:2026 document contain?
oSIST prEN ISO 13366-2:2026 contains practical guidance on instrument principle, sample quality, preservation, calibration, and routine surveillance. It also contains tables for calibration ranges and precision values, plus formulas for linearity and carry-over checks.
The document includes procedures for preparing calibration samples, checking blank and carry-over effects, assigning pilot milk values, and checking repeatability and intralaboratory reproducibility. Annex A adds a worked alternative using direct labelling with a DNA stain, including instrument settings, reagents, sample preparation, analysis, and a calculation example.
For users, this means the document is not just a description of the method - it is also a working guide for setting up the test, controlling the instrument, and documenting the result.
Relations
- Effective Date
- 14-May-2025
- Effective Date
- 14-May-2025
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

Bureau Veritas
Bureau Veritas is a world leader in laboratory testing, inspection and certification services.

DNV
DNV is an independent assurance and risk management provider.
Sponsored listings
Frequently Asked Questions
oSIST prEN ISO 13366-2:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Milk - Enumeration of somatic cells - Part 2: Flow cytometry method (ISO/DIS 13366-2:2026)". This standard covers: This document gives guidance on the operating conditions for counting the total number of somatic cells, in both raw and chemically preserved milk, using flow cytometers. The guidance is applicable to the counting of somatic cells in raw cow milk. The guidance is also applicable to raw milk of other species, such as goat, sheep and buffalo, if the specified prerequisites are met.
This document gives guidance on the operating conditions for counting the total number of somatic cells, in both raw and chemically preserved milk, using flow cytometers. The guidance is applicable to the counting of somatic cells in raw cow milk. The guidance is also applicable to raw milk of other species, such as goat, sheep and buffalo, if the specified prerequisites are met.
oSIST prEN ISO 13366-2:2026 is classified under the following ICS (International Classification for Standards) categories: 67.100.10 - Milk and processed milk products. The ICS classification helps identify the subject area and facilitates finding related standards.
oSIST prEN ISO 13366-2:2026 has the following relationships with other standards: It is inter standard links to SIST EN ISO 13366-2:2007, SIST EN ISO 13366-2:2007/AC:2007. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
oSIST prEN ISO 13366-2:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 reference method
- 3.2 somatic cells
- 4 Principle
- 5 Factors affecting the results of measurements
- 5.1 Sample bottles
- 5.2 Sampling
- 5.3 Preservation
- 5.4 Sample storage and transport
- 5.5 Interfering substances
- 5.6 Sample quality at analysis
- 5.7 Chemicals
- 5.8 Instrument condition
- 6 Calibration
- 6.1 Reference materials
- 6.2 Calibration procedure
- 7 Sampling
- 8 Determination
- 9 Checks on performance in routine operation
- 9.1 Blank checks
- 9.2 Carry-over effect
- 9.3 Pilot checks
- 9.4 Additional instrument surveillance
- 9.5 Repeatability
- 9.6 Intralaboratory reproducibility
- 9.7 External comparisons
- 10 Specific remarks for the use with milk from different species
- 10.1 General
- 10.2 Cow milk
- 10.3 Goat milk
- 10.4 Sheep milk
- 10.5 Buffalo milk
- 11 Precision
- 11.1 Repeatability
- 11.2 Intralaboratory reproducibility
- 11.3 Reproducibility
- 12 Test report
- Annex A
- Bibliography
1 Scope
This document gives guidance on the operating conditions for counting the total number of somatic cells, in both raw and chemically preserved milk, using flow cytometers. The guidance is applicable to the counting of total somatic cells in raw cow milk. The guidance is also applicable to raw milk of other species, such as goat, sheep and buffalo, if the specified prerequisites are met.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 8196-1, | IDF 128-1, Milk — Definition and evaluation of the overall accuracy of alternative methods of milk analysis — Part 1: Analytical attributes of alternative methods ISO 8196-2, | IDF 128-2, Milk — Definition and evaluation of the overall accuracy of alternative methods of milk analysis — Part 2: Calibration and quality control in the dairy laboratory ISO 13366-1, | IDF 148-1, Milk — Enumeration of somatic cells — Part 1: Microscopic method (Reference method) ISO 17034, General requirements for the competence of reference material producers ISO/IEC 17043:2023, Conformity assessment — General requirements for the competence of proficiency testing providers
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
3.1 reference method
method described in ISO 13366-1 | IDF 148-1 for the counting of somatic cells
3.2 somatic cells
those cells that show more than a threshold intensity of fluorescence due to the staining of DNA in their nuclei Note 1 to entry: The number of somatic cells is expressed in cells per millilitre.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en)
4 Principle
Flow cytometers designed for automatically counting the total number of somatic cells in milk have functions for the uptake of reagents and test sample, a mixing section and a counting section. In the mixing section, the test sample is mixed with a buffer and a stain solution. Part of the resulting mixture is transferred to the counting section.
For other flow cytometers, the steps described above can follow the procedure described in (see Annex A). In the mixing section, closely controlled volumes of test sample and buffer/stain solutions are dosed and mixed. Mixing can take place in a cup, a mixing chamber, or in the tubing leading to a flow cell. In the counting section, part of the mixture is placed in the high-speed flow of a surrounding sheath liquid in a capillary flow cell. Through acceleration, the mixture forms a thin string in which the somatic cells are dynamically focused and aligned. This string then passes the objective of a fluorescence microscope. Each stained particle produces an electrical pulse that is filtered, amplified and recorded. The resulting pulse height distribution is electronically processed, whereby discrimination is made between noise signals and pulses that are attributed to stained somatic cells. The discriminator level can either be fixed or dynamic. Some instruments contain two channels in the counting section. In terms of analytical quality assurance, such a situation should be considered equivalent to working with two separate units, so the performance should be evaluated separately for each channel.
5 Factors affecting the results of measurements
5.1 Sample bottles
Sample bottles should be fit for use; i.e. to transfer test samples from the point of sampling to the laboratory without loss or damage.
Care is to be exercised that sample bottles are leak-proof and that a proper empty volume is left. Too large an empty volume can facilitate churning; too small an empty volume can cause problems with mixing.
5.2 Sampling
5.2.1 General
Sampling materials (i.e. sample bottles, beakers and sampling devices) should be clean and dry. Where automatic samplers are used, these should have been properly validated. Test samples should preferably be cooled immediately after sampling at 4 °C ± 2 °C and be kept at that temperature until counting (see 5.4) rather than being preserved. Freezing should be avoided. If preservation is necessary, suitable means for chemical preservation of test samples are described in 5.3.
5.2.2 Bulk milk samples
Thorough mixing of the raw bulk milk to be sampled is essential. Somatic cells will concentrate in the upper layer in the case of insufficient stirring.
5.2.3 Milk samples from individual animals
The release of somatic cells in the milk during milking is uneven. When the aim is to produce a representative counting result for a whole milking, it is essential that a representative sample of the whole milking is obtained. For diagnostic purposes, a sample of a partial milking may suffice.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en)
5.3 Preservation
If chemical preservation is considered necessary, the test sample (5.2.1) should be preserved as soon as possible, but in any case within 24 h after sampling. In all cases, the test sample should be kept cool (4 °C ± 2 °C) until the addition of the preservative.
WARNING — Sodium azide and potassium dichromate are highly hazardous reagents. Handle with extreme care. Prepare and use the reagents only in a properly equipped laboratory, in a fume hood, and always wear appropriate personal protective equipment (PPE), including gloves, lab coat an eye protection.
Suitable preservatives are the following.
a) Boric acid: its final concentration in the test sample should not exceed 0,6 g/100 ml. Such preserved samples may be stored at between 6 °C and 12 °C for up to a further 24 h. b) Sodium azide: its final concentration in the test sample should not exceed 0,024 g/100 ml. Such preserved samples may be stored at between 2 °C and 10 °C for up to a further 72 h. c) Bronopol (2-bromo-2-nitro-1,3-propanediol): its final concentration in the test sample should not exceed 0,05 g/100 ml. Such preserved samples may be stored at between 2 °C and 12 °C for up to a further 6 d. d) Potassium dichromate: its final concentration in the test sample should not exceed 0,2 g/100 ml. Such preserved samples may be stored at between 2 °C and 12 °C for up to a further 6 d. Accompanying colour tracers found to be suitable are:
- Patent Blue V with a final concentration in the test sample of up to 0,15 mg/100 ml;
- Yellow Orange S (E110) with a final concentration in the test sample of up to 1 mg/100 ml;
- a mixture of Patent Blue V and Eosin B with a final concentration in the test sample of up to 0,03 mg/100 ml and 0,45 mg/100 ml, respectively;
- Methylene blue with a final concentration in the test sample of up to 0,5 mg/100 ml. Other preservatives and colour tracers may be used provided that their effectiveness and conditions for use have been soundly validated.
In all cases, properly validate the absence of interference for the counting equipment concerned before application.
In cases where flow cytometers are combined with milk analysers for the measurement of other test sample components, care should be taken that the applied preservatives and colour tracers do not affect the counting result.
5.4 Sample storage and transport
Unpreserved test samples should be stored at 4 °C ± 2 °C and should be counted within 96 h after the completion of sampling. Avoid freezing the test samples. Storage at higher temperatures and/or over longer time scales may result in considerable loss of countable cells and therefore non-representative values. The measurement of samples after freezing and thawing may result in lower counts (by 10 % to 20 %). The age of the samples at freezing and the type of thawing process can influence the counting result.
5.5 Interfering substances
The use of substances that interfere in the counting should be avoided. Substances known to influence the instrument read-out are preservatives and colour tracers at higher concentrations than specified 5.3.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en)
5.6 Sample quality at analysis
Breakdown of somatic cells (lysis) will result in an increase of smaller cell fragments. The lower intensity of fluorescence after staining of these particles causes a shift in the pulse height distribution to the left. This will hamper proper differentiation from noise pulses and therefore result in a lower count. NOTE In several types of instruments, features are available for evaluating the position and the shape of the pulse height distribution. See the relevant instructions and guidance from the instrument manufacturer. After the processing of problematic samples, the flow path should be checked and possible cleaned. The proper functioning of the instrument should be tested before further use. Possible problematic test samples are a) milk samples from severely infected udders, i.e. with clots; b) milk samples with impurities; c) milk samples with high numbers of erythrocytes; d) colostrum; e) late lactation milk; f) sour milk.
Where possible, analysis of problematic test samples should be avoided.
5.7 Chemicals
All reagents used should be of recognized analytical and bacteriological quality. Water used should be demineralized (remaining conductivity < 10 µS/cm) or water of at least equivalent purity. Manufacturer's instructions for the preparation of the working solutions, the maximum storage time and storage requirements should be followed.
Local conditions regarding the use and discharge of applied chemicals and effluents should be observed.
5.8 Instrument condition
5.8.1 General points of attention are:
a) the functioning of the mixing device and the stirrer, b) possible disturbances at sample intake and in the flow system due to blocking by impurities, clots or fouling in the mixing and incubation units, and c) the condition and the functioning of the light source and the photo multiplier, the gain setting and the signal quality.
5.8.2 A specific point of attention with flow cytometers is the variation in the behaviour of the sample string in the flow cell and the sheath liquid flow. Some instrument manufacturers offer special programme features for checking this, thereby indicating possible solutions in the case of deviations.
6 Calibration
6.1 Reference materials
6.1.1 General
Reference materials should be produced under closely controlled conditions; i.e. working with a quality assurance system and fulfilling the requirements as listed in ISO 17034.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en) Reference materials may be [5] a) certified reference materials (CRMs) as produced by a recognized official organization, see ; b) secondary reference materials (SRMs) as prepared by an external supplier, whereby traceability is kept with CRMs or; c) in-house reference materials (IRMs) as prepared by the laboratory itself, whereby traceability is kept with CRMs, SRMs.
6.1.2 Preparation of calibration samples
6.1.2.1 Preparation by addition of a bovine leucocyte suspension
a) Mix 1 000 ml of sterilized or UHT milk with low somatic cell count with 1 ml of polypropylene 2 000 and 0,4 g of bronopol.
b) Add the required amount of a suitable leucocyte suspension to different portions of the mixture in order to obtain a suitable range of cell counts.
6.1.2.2 Preparation by microfiltration
a) Collect fresh bulk milk and add bronopol to a final mass fraction of 0,02 %. b) Skim the milk in a cream separator to a mass fraction of fat of below 0,1 %. c) Concentrate the skimmed milk 20-fold by applying tangential microfiltration over a membrane with a pore size of 1,4 µm, resulting in a portion with a high cell content (HCM) and a portion with a low cell content (LCM).
d) Mix cream, HCM and LCM in the required quantities so as to obtain 5 to 8 milk portions with a mass fraction of fat of 3,5 % and different levels of cell counts covering the range of interest.
6.1.2.3 Preparation by centrifugation
a) Collect fresh bulk milk and add bronopol to a final mass fraction of 0,02 %. b) Filter the thus-prepared milk over a metal filter with a pore size of 0,5 mm. c) Skim the filtered milk by centrifuging at a radial acceleration of 40g for 10 min so as to obtain cream, skimmed milk and a pellet.
d) Mix the skimmed milk and pellet material in the required quantities so as to obtain 5 to 8 milk portions with different cell count levels covering the range of interest. e) If necessary, add cream so as to obtain portions with a mass fraction of fat of 3,3 % ± 0,3 %.
f) Split the obtained portions into individual samples and heat these at 120 °C and 10 Pa for 3 min.
6.1.3 Assignment of reference values
Reference values should be traceable to a CRM that is characterized also with ISO 13366-1|IDF 148-1. An example on how to build traceability is described in (see [5]).
6.1.4 Storage conditions and keeping time
The shelf life, the storage conditions and the keeping time of the prepared secondary reference materials should be properly validated.
oSIST prEN ISO 13366-2:2026 ISO/DIS 13366-2:2026(en) IDF 148-2:2026(en) When chemical preservation is used, alignment should be sought with the type and concentration of the preservative applied to the test sample (see 5.3).
...



