Industrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement

Includes FprAA:2009 (pr=21593)

Industrielle, wissenschaftliche und medizinische Geräte - Funkstörungen - Grenzwerte und Messverfahren

Appareils industriels, scientifiques et médicaux - Caractéristiques des perturbations radioélectriques - Limites et méthodes de mesure

Industrijska, znanstvena in medicinska (ISM) radiofrekvenčna oprema - Karakteristike občutljivosti za radijske motnje - Mejne vrednosti in merilne metode (CISPR 11:2009, spremenjen)

Ta mednarodni standard velja za industrijsko, znanstveno in medicinsko električno opremo v frekvenčnem razponu od 0 Hz do 400 GHz in za hišne in podobne aparate načrtovane za generiranje in/ali uporabljanje radio-frekvenčne energije. Ta standard zajema zahteve emisij, povezane z radio-frekvenčnimi (RF) motnjami v frekvenčnem razponu od 9kHz do 400 GHz. Meritve je potrebno opraviti samo v frekvenčnem razponu, za katerega so meje določene v Klavzuli 6. Za ISM RF uporabo v pomenu opredelitve, ki se nahaja v radijskih predpisih ITU (glej opredelitev 3.1), ta standard zajema zahteve emisij, povezane z radio-frekvenčnimi motnjami v frekvenčnem razponu od 9kHz do 18 GHz. Zahteve za ISM RF aparat za razsvetljavo in UV sevalnike, ki delujejo na frekvencah znotraj ISM frekvenčnih razponov  določenih z radio uredbami ITU, so vsebovane v tem standardu. Oprema, zajeta z ostalimi CISPR proizvodi in standardi emisij družin proizvodov, so izvzete iz področja tega standarda.

General Information

Status
Withdrawn
Publication Date
22-Nov-2009
Withdrawal Date
31-Aug-2012
Current Stage
9960 - Withdrawal effective - Withdrawal
Start Date
15-Feb-2019
Completion Date
15-Feb-2019

Relations

Buy Standard

Standard
EN 55011:2010
English language
59 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Industrielle, wissenschaftliche und medizinische Geräte - Funkstörungen - Grenzwerte und MessverfahrenAppareils industriels, scientifiques et médicaux - Caractéristiques des perturbations radioélectriques - Limites et méthodes de mesureIndustrial, scientific and medical equipment - Radio-frequency disturbance characteristics - Limits and methods of measurement33.100.10EmisijaEmissionICS:Ta slovenski standard je istoveten z:EN 55011:2009SIST EN 55011:2010en01-februar-2010SIST EN 55011:2010SLOVENSKI
STANDARDSIST EN 55011:20071DGRPHãþD



SIST EN 55011:2010



EUROPEAN STANDARD EN 55011 NORME EUROPÉENNE
EUROPÄISCHE NORM November 2009
CENELEC European Committee for Electrotechnical Standardization Comité Européen de Normalisation Electrotechnique Europäisches Komitee für Elektrotechnische Normung
Central Secretariat: Avenue Marnix 17, B - 1000 Brussels
© 2009 CENELEC -
All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 55011:2009 E
ICS 33.100.10 Supersedes EN 55011:2007 + A2:2007
English version
Industrial, scientific and medical equipment -
Radio-frequency disturbance characteristics -
Limits and methods of measurement (CISPR 11:2009, modified)
Appareils industriels,
scientifiques et médicaux -
Caractéristiques
des perturbations radioélectriques -
Limites et méthodes de mesure (CISPR 11:2009, modifiée)
Industrielle, wissenschaftliche
und medizinische Geräte -
Funkstörungen -
Grenzwerte und Messverfahren
(CISPR 11:2009, modifiziert)
This European Standard was approved by CENELEC on 2009-09-01. CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the Central Secretariat or to any CENELEC member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CENELEC member into its own language and notified to the Central Secretariat has the same status as the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Cyprus, the Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and the United Kingdom.
SIST EN 55011:2010



EN 55011:2009 – 2 – Foreword The text of document CISPR/B/478/FDIS, future edition 5 of CISPR 11, prepared by CISPR SC B, Interference relating to industrial, scientific and medical radio-frequency apparatus, to other (heavy) industrial equipment, to overhead power lines, to high voltage equipment and to electric traction, was submitted to the IEC-CENELEC parallel vote. A draft amendment (FprAA) covering common modifications towards the future edition 5 of CISPR 11 (CISPR/B/478/FDIS), prepared by the Technical Committee CENELEC TC 210, Electromagnetic compatibility (EMC), was submitted to the formal vote. The combined texts were approved by CENELEC as EN 55011 on 2009-09-01. This European Standard supersedes EN 55011:2007 + A2:2007. This EN 55011:2009 got a more transparent structure, introduces another set of particular limits for conducted and radiated disturbances of "heavy duty" general purpose equipment of class A group 1 with a rated input power in excess of 20 kVA, in accordance with the needs of the industries and refers to the full approach in respect of the measurement instrumentation uncertainty specified in CISPR 16-4-4. Furthermore, any kind of "legal statements" were removed from the normative main body of this European Standard. The following dates were fixed: – latest date by which the EN
has to be implemented at national level by publication of an identical national standard or by endorsement
(dop)
2010-09-01 – latest date by which the national standards conflicting with the EN have to be withdrawn
(dow)
2012-09-01 This European Standard has been prepared under a mandate given to CENELEC by the European Commission and the European Free Trade Association and covers essential requirements of EC Directive 2004/108/EC. See Annex ZZ. Annexes ZA, ZB and ZZ have been added by CENELEC. ____________ SIST EN 55011:2010



– 3 – EN 55011:2009 The main content of this standard is based on CISPR Recommendation No. 39/2 given below: RECOMMENDATION No. 39/2 Limits and methods of measurement of electromagnetic disturbance characteristics
of industrial, scientific and medical (ISM) radio-frequency equipment The CISPR CONSIDERING a)
that ISM RF equipment is an important source of disturbance; b) that methods of measuring such disturbanc
...