IEC/DIS 60601-2-26
(Main)Medical electrical equipment
Medical electrical equipment
Appareils électromédicaux
General Information
Standards Content (sample)
DRAFT INTERNATIONAL STANDARD
IEC/DIS 60601-2-26
ISO/TC 121/SC 3 Secretariat: ANSI
Voting begins on: Voting terminates on:
2017-07-07 2017-09-29
Medical electrical equipment —
Part 2-26:
Particular requirements for the basic safety and essential
performance of electroencephalographs
Appareils électromédicaux —
Partie 2-26: Titre manque
ICS: 11.140
THIS DOCUMENT IS A DRAFT CIRCULATED
FOR COMMENT AND APPROVAL. IT IS
THEREFORE SUBJECT TO CHANGE AND MAY
This document is circulated as received from the committee secretariat.
NOT BE REFERRED TO AS AN INTERNATIONAL
STANDARD UNTIL PUBLISHED AS SUCH.
IN ADDITION TO THEIR EVALUATION AS
BEING ACCEPTABLE FOR INDUSTRIAL,
This draft is submitted to a parallel vote in ISO and in IEC.
TECHNOLOGICAL, COMMERCIAL AND
USER PURPOSES, DRAFT INTERNATIONAL
STANDARDS MAY ON OCCASION HAVE TO
BE CONSIDERED IN THE LIGHT OF THEIR
POTENTIAL TO BECOME STANDARDS TO
WHICH REFERENCE MAY BE MADE IN
Reference number
NATIONAL REGULATIONS.
IEC/DIS 60601-2-26:2017(E)
RECIPIENTS OF THIS DRAFT ARE INVITED
TO SUBMIT, WITH THEIR COMMENTS,
NOTIFICATION OF ANY RELEVANT PATENT
RIGHTS OF WHICH THEY ARE AWARE AND TO
PROVIDE SUPPORTING DOCUMENTATION. IEC 2017
---------------------- Page: 1 ----------------------
62D/1488/CDV – 2 – IEC CDV 60601-2-26 © IEC:2017
1 CONTENTS
2 FOREWORD .......................................................................................................................... 3
3 INTRODUCTION .................................................................................................................... 5
4 201.1 Scope, object and related standards ............................................................................. 6
5 201.2 Normative references ................................................................................................... 7
6 201.3 Terms and definitions ................................................................................................... 8
7 201.4 General requirements ................................................................................................... 9
8 201.5 General requirements for testing of ME EQUIPMENT ........................................................ 9
9 201.6 Classification of ME EQUIPMENT and ME SYSTEMS ......................................................... 10
10 201.7 ME EQUIPMENT identification, marking and documents ................................................. 10
11 201.8 Protection against electrical HAZARDS from ME EQUIPMENT ........................................... 11
12 201.9 Protection against MECHANICAL HAZARDS of ME EQUIPMENT and ME SYSTEMS ................. 16
13 201.10 Protection against unwanted and excessive radiation HAZARDS ................................. 17
14 201.11 Protection against excessive temperatures and other HAZARDS ................................. 17
15 201.12 Accuracy of controls and instruments and protection against hazardous outputs ...... 18
16 201.13 HAZARDOUS SITUATIONS and fault conditions .............................................................. 23
17 201.14 PROGRAMMABLE ELECTRICAL MEDICAL SYSTEMS (PEMS) ................................................. 23
18 201.15 Construction of ME EQUIPMENT ................................................................................... 23
19 201.16 ME SYSTEMS .............................................................................................................. 23
20 201.17 ELECTROMAGNETIC COMPATIBILITY of ME EQUIPMENT and ME SYSTEMS .......................... 23
21 202 Electromagnetic disturbances – Requirements and tests ............................................... 23
22 206 USABILITY ...................................................................................................................... 29
23 Annexes .............................................................................................................................. 29
24 Annex AA (informative) Particular guidance and rationale ................................................... 30
25 Index of defined terms used in this particular standard ......................................................... 33
27 Figure 201.101 – Test of protection against the effects of defibrillation (common mode) ...... 13
28 Figure 201.102 – Test of protection against the effects of defibrillation (differential mode) ... 15
29 Figure 201.103 – Application of the test voltage between LEAD WIRES to test the energy
30 delivered by the defibrillator ................................................................................................. 16
31 Figure 201.104 – General test circuit .................................................................................... 20
32 Figure 201.105 – Test circuit for noise and common mode rejection (see 201.12.1.103 and
33 201.12.1.105) ....................................................................................................................... 23
34 Figure 202.101 – Test layout for radiated and conducted EMISSION test and radiated immunity
35 test (see 202.4.3.1) .............................................................................................................. 25
36 Figure 202.102 –Test circuit for HF SURGICAL EQUIPMENT protection measurement according to
37 subclause 202.8.102 ............................................................................................................ 27
38 Figure 202.103 – Test setup for HF SURGICAL EQUIPMENT measurement according to subclause
39 202.8.102 ............................................................................................................................. 28
41 Table 201.101 – Distributed ESSENTIAL PERFORMANCE requirements ....................................... 9
---------------------- Page: 2 ----------------------IEC CDV 60601-2-26 ©IEC: 2017 –3 – 62D/1488/CDV
44 INTERNATIONAL ELECTROTECHNICAL COMMISSION
45 ____________
47 MEDICAL ELECTRICAL EQUIPMENT –
49 Part 2-26: Particular requirements for the basic safety
50 and essential performance of electroencephalographs
53 FOREWORD
54 1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising all
55 national electrotechnical committees (IEC National Committees). The object of IEC is to promote international co-
56 operation on all questions concerning standardization in the electrical and electronic fields. To this end and in addition to
57 other activities, IEC publishes International Standards, Technical Specificati ons, Technical Reports, Publicly Available
58 Specifications (PAS) and Guides (hereafter referred to as “IEC Publication(s)”). Their preparation is entrusted to technical
59 committees; any IEC National Committee interested in the subject dealt with may participate in this preparatory work.
60 International, governmental and non-governmental organizations liaising with the IEC also participate in this preparation.
61 IEC collaborates closely with the International Organization for Standardization (ISO) in accordance wi th conditions
62 determined by agreement between the two organizations.63 2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
64 consensus of opinion on the relevant subjects since each technical committee has representation from all interested IEC
65 National Committees.66 3) IEC Publications have the form of recommendations for international use and are accepted by IEC National Committees
67 in that sense. While all reasonable efforts are made to ensure that the technical content of IEC Publications is accurate,
68 IEC cannot be held responsible for the way in which they are used or for any misinterpretation by any end user.
69 4) In order to promote international uniformity, IEC National Committees undertake to ap ply IEC Publications transparently
70 to the maximum extent possible in their national and regional publications. Any divergence between any IEC Publication
71 and the corresponding national or regional publication shall be clearly indicated in the latter.
72 5) IEC itself does not provide any attestation of conformity. Independent certification bodies provide conformity assessment
73 services and, in some areas, access to IEC marks of conformity. IEC is not responsible for any services carried out by
74 independent certification bodies.75 6) All users should ensure that they have the latest edition of this publication.
76 7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and members of
77 its technical committees and IEC National Committees for any personal injury, property damage or other damage of any
78 nature whatsoever, whether direct or indirect, or for costs (including legal fees) and expenses arising out of the
79 publication, use of, or reliance upon, this IEC Publication or any other IEC Publications.
80 8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
81 indispensable for the correct application of this publication.82 9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of patent rights.
83 IEC shall not be held responsible for identifying any or all such patent rights.
84 International standard IEC 60601-2-26 has been prepared by a Joint Working Group of IEC
85 subcommittee 62D: Electromedical equipment, of IEC technical committee 62: Electrical equipment in
86 medical practice and ISO subcommittee SC3: Lung ventilators and related equipment, of ISO technical
87 committee 121: Anaesthetic and respiratory equipment.88 This fourth edition cancels and replaces the third edition of IEC 60601-2-26 published in 2012. This
89 edition constitutes a technical revision to align with Amendment 1:2012 to IEC 60601-1:2005, new
90 versions of collateral standards and amendments thereto.91 The text of this particular standard is based on the following documents:
FDIS Report on voting
93 Full information on the voting for the approval of this particular standard can be found in the report on
94 voting indicated in the above table.---------------------- Page: 3 ----------------------
62D/1488/CDV – 4 – IEC CDV 60601-2-26 © IEC:2017
95 This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
96 In this standard, the following print types are used:97 – Requirements and definitions: roman type.
98 – Test specifications: italic type.
99 – Informative material appearing outside of tables, such as notes, examples and references: in smaller type. Normative
100 text of tables is also in a smaller type.101 – TERMS DEFINED IN CLAUSE 3 OF THE GENERAL STANDARD, IN THIS PARTICULAR STANDARD OR AS NOTED:
102 SMALL CAPITALS.103 In referring to the structure of this standard, the term
104 – “clause” means one of the seventeen numbered divisions within the table of contents, inclusive of
105 all subdivisions (e.g. Clause 7 includes subclauses 7.1, 7.2, etc.);106 – “subclause” means a numbered subdivision of a clause (e.g. 7.1, 7.2 and 7.2.1 are all subclauses
107 of Clause 7).108 References to clauses within this standard are preceded by the term “Clause” followed by the clause
109 number. References to subclauses within this collateral standard are by number only.
110 In this standard, the conjunctive “or” is used as an “inclusive or” so a statement is true if any
111 combination of the conditions is true.112 The verbal forms used in this standard conform to usage described in Annex H of the ISO/IEC
113 Directives, Part 2. For the purposes of this standard, the auxiliary verb:114 – “shall” means that compliance with a requirement or a test is mandatory for compliance with this
115 standard;116 – “should” means that compliance with a requirement or a test is recommended but is not mandatory
117 for compliance with this standard;118 – “may” is used to describe a permissible way to achieve compliance with a requirement or test.
119 An asterisk (*) as the first character of a title or at the beginning of a paragraph or table title indicates
120 that there is guidance or rationale related to that item in Annex AA.121 A list of all parts of the IEC 60601 series, published under the general title Medical electrical
122 equipment, can be found on the IEC website.123 The committee has decided that the contents of this publication will remain unchanged until the
124 stability date indicated on the IEC web site under "http://webstore.iec.ch" in the data related to the
125 specific publication. At this date, the publication will be126 • reconfirmed,
127 • withdrawn,
128 • replaced by a revised edition, or
129 • amended.
130
131 NOTE The attention of National Committees is drawn to the fact that equipment MANUFACTURERS and testing organizations
132 may need a transitional period following publication of a new, amended or revised IEC or ISO publication in which to make
133 products in accordance with the new requirements and to equip themselves for conducting new or revised tests. It is the
134 recommendation of the committee that the content of this publication be adopted for mandatory implementation nationally not
135 earlier than 3 years from the date of publication.136
---------------------- Page: 4 ----------------------
IEC CDV 60601-2-26 ©IEC: 2017 –5 – 62D/1488/CDV
137 INTRODUCTION
138 This particular standard concerns the BASIC SAFETY and ESSENTIAL PERFORMANCE of
139 ELECTROENCEPHALOGRAPHS. It amends and supplements IEC 60601-1 (third edition, 2005, Amendment
140 1:2012; Corrigendum 1:2012): Medical electrical equipment – Part 1: General requirements for basic
141 safety and essential performance), hereinafter referred to as the general standard.
142 The aim of this fourth edition is to bring this particular standard up to date with reference to the edition
143 3.1 of the general standard and new versions of collateral standards and amendments thereto through
144 technical changes.145 The requirements of this particular standard take priority over those of the general standard.
146 A “General guidance and rationale” for the more important requirements of this particular standard is
147 included in Annex AA. It is considered that knowledge of the reasons for these requirements will not
148 only facilitate the proper application of the standard but will, in due course, expedite any revision
149 necessitated by changes in clinical practice or as a result of developments in technology. However,
150 Annex AA does not form part of the requirements of this standard.151
---------------------- Page: 5 ----------------------
62D/1488/CDV – 6 – IEC CDV 60601-2-26 © IEC:2017
152 MEDICAL ELECTRICAL EQUIPMENT –
153
154 Part 2-26: Particular requirements for the basic safety
155 and essential performance of electroencephalographs
156
157
158
159 201.1 Scope, object and related standards
160 Clause 1 of the general standard applies, except as follows:
161 201.1.1 * Scope
162 Replacement:
163 This particular standard applies to BASIC SAFETY and ESSENTIAL PERFORMANCE of
164 ELECTROENCEPHALOGRAPHS as defined in 201.3.204, hereafter also referred to as ME EQUIPMENT or
165 ME SYSTEM. This standard is applicable to ELECTROENCEPHALOGRAPHS intended for use in professional
166 healthcare facilities, the EMERGENCY MEDICAL SERVICE ENVIRONMENT or the167 HOME HEALTHCARE ENVIRONMENT.
168 This standard does not cover requirements for other equipment used in electroencephalography such
169 as:170 – phono-photic stimulators;
171 – EEG data storage and retrieval;
172 – ME EQUIPMENT particularly intended for monitoring during electro-convulsive therapy.
173 If a clause or subclause is specifically intended to be applicable to ME EQUIPMENT only, or to
174 ME SYSTEMS only, the title or content of that clause or subclause will say so. If that is not the case, the
175 clause or subclause applies both to ME EQUIPMENT and to ME SYSTEMS, as follows:
176 The clause or subclause applies to ME EQUIPMENT, as default. If the corresponding safety measure or
177 function is intentionally, by decision of the manufacturer, not completely integrated into the
178 ME EQUIPMENT, but has to be realized in an ME SYSTEM, the manufacturer of the ME EQUIPMENT has to
179 specify in its accompanying documents which functionality and safety requirements have to be
180 provided by the connected ME SYSTEM to fulfil that clause or subclause. The ME SYSTEM, then, has to
181 be verified accordingly.182 HAZARDS inherent in the intended physiological function of ME EQUIPMENT or ME SYSTEMS within the
183 scope of this standard are not covered by specific requirements in this standard except in 7.2.13 and
184 8.4.1 of the general standard.185 NOTE See also 4.2 of the general standard.
186 201.1.2 Object
187 Replacement:
188 The object of this particular standard is to establish particular BASIC SAFETY and
189 ESSENTIAL PERFORMANCE requirements for ELECTROENCEPHALOGRAPHS as defined in 201.3.204.
The general standard is IEC 60601-1:2005, Medical electrical equipment – Part 1: General requirements for basic safety
and essential performance including Amendment 1:2012.---------------------- Page: 6 ----------------------
IEC CDV 60601-2-26 ©IEC: 2017 –7 – 62D/1488/CDV
190 201.1.3 Collateral standards
191 This particular standard refers to those applicable collateral standards that are listed in Clause 2 of
192 the general standard and Clause 201.2 of this particular standard.193 IEC 60601-1-2:2014 and IEC 60601-1-6:2013 apply as modified in Clause 202 and 206, respectively.
194 IEC 60601-1-3, IEC 60601-1-9 and IEC 60601-1-10 do not apply. All other published collateral
195 standards in the IEC 60601-1 series apply as published.196 201.1.4 Particular standards
197 Replacement:
198 In the IEC 60601 series, particular standards may modify, replace or delete requirements contained in
199 the general standard and collateral standards as appropriate for the particular ME EQUIPMENT or
200 ME SYSTEM under consideration, and may add other BASIC SAFETY and ESSENTIAL PERFORMANCE
201 requirements.202 A requirement of a particular standard takes priority over the general standard and collateral
203 standards.204 For brevity, IEC 60601-1 is referred to in this particular standard as the general standard. Collateral
205 standards are referred to by their document number.206 The numbering of clauses and subclauses of this particular standard corresponds to that of the
207 general standard with the prefix “201” (e.g. 201.1 in this standard addresses the content of Clause 1
208 of the general standard) or applicable collateral standard with the prefix “20x” where x is the final
209 digit(s) of the collateral standard document number (e.g. 202.4 in this particular standard addresses
210 the content of Clause 4 of the IEC 60601-1-2 collateral standard, etc.). The changes to the text of the
211 general standard and applicable collateral standards are specified by the use of the following words:
212 "Replacement" means that the clause or subclause of the general standard or applicable collateral
213 standard is replaced completely by the text of this particular standard.214 "Addition" means that the text of this particular standard is additional to the requirements of the
215 general standard or applicable collateral standard.216 "Amendment" means that the clause or subclause of the general standard or applicable collateral
217 standard is amended as indicated by the text of this particular standard.218 Subclauses, figures or tables which are additional to those of the general standard are numbered
219 starting from 201.101. However due to the fact that definitions in the general standard are numbered
220 3.1 through 3.139, additional definitions in this standard are numbered beginning from 201.3.201.
221 Additional annexes are lettered AA, BB, etc., and additional items aa), bb), etc.
222 Subclauses, figures or tables which are additional to those of a collateral standard are numbered
223 starting from 20x, where “x” is the number of the collateral standard, e.g. 202 for IEC 60601-1-2, etc.
224 The term "this standard" is used to make reference to the general standard, any applicable collateral
225 standards and this particular standard taken together.226 Where there is no corresponding clause or subclause in this particular standard, the clause or
227 subclause of the general standard or applicable collateral standard, although possibly not relevant,
228 applies without modification; where it is intended that any part of the general standard or applicable
229 collateral standard, although possibly relevant, is not to be applied, a statement to that effect is given
230 in this particular standard.231 201.2 Normative references
232 Clause 2 of the general standard applies, except as follows:
---------------------- Page: 7 ----------------------
62D/1488/CDV – 8 – IEC CDV 60601-2-26 © IEC:2017
233 Amendment:
234 IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic safety
235 and essential performance – Collateral standard: Electromagnetic disturbances – Requirements and
236 tests237 Addition:
238 IEC 60601-1-11:2015, Medical electrical equipment – Part 1-11: General requirements for basic safety
239 and essential performance – Collateral standard: Requirements for medical electrical equipment and
240 medical electrical systems used in the home healthcare environment241 IEC 60601-1-12:2014, Medical electrical equipment – Part 1-12: General requirements for basic safety
242 and essential performance – Collateral standard: Requirements for medical electrical equipment and
243 medical electrical systems intended for use in the emergency medical services environment
244 201.3 Terms and definitions245 NOTE An index of defined terms is found beginning on page 33.
246 For the purpose of this document, the terms and definitions given in IEC 60601-1:2012, IEC 60601-1-
247 2:2014, IEC 60601-1-6:2013, IEC 60601-1-11:2015, IEC 60601-1-12:2014 and IEC 60601-2-2:2017
248 apply, except as follows:249 Additional definitions:
250 201.3.201
251 CHANNEL
252 hardware and/or software selection of a particular electroencephalograp hic LEAD for purposes of
253 display, recording, or transmission254 201.3.202
255 ELECTRODE
256 sensor that is applied to the scalp, cerebral cortex, or subdural locations to detect electrical activity of
257 the brain258 201.3.203
259 ELECTROENCEPHALOGRAM
260 EEG
261 presentation (on screen or paper) of the variation with time of voltages taken from ELECTRODES, whose
262 positions are specified263 201.3.204
264 ELECTROENCEPHALOGRAPH
265 ME EQUIPMENT or ME SYSTEM to produce an ELECTROENCEPHALOGRAM
266 201.3.206
267 LEAD WIRE
268 cable connected between an ELECTRODE and either a PATIENT CABLE or the ELECTROENCEPHALOGRAPH
269 201.3.207270 NEUTRAL ELECTRODE
271 reference point for differential amplifiers and/or interference suppression circuits
272 201.3.208273 PATIENT CABLE
274 multiwire cable or junction box used to connect LEAD WIRES to the ELECTROENCEPHALOGRAPH
---------------------- Page: 8 ----------------------IEC CDV 60601-2-26 ©IEC: 2017 –9 – 62D/1488/CDV
275 201.4 General requirements
276 Clause 4 of the general standard applies, except as follows:
277 201.4.3 ESSENTIAL PERFORMANCE
278 Addition:
279 201.4.3.101 Additional ESSENTIAL PERFORMANCE requirements
280 Additional ESSENTIAL PERFORMANCE requirements are found in the subclauses listed in Table 201.101.
281 Table 201.101 – Distributed ESSENTIAL PERFORMANCE requirementsRequirement Subclause
Accuracy of signal reproduction 201.12.1.102
Input dynamic range and differential offset voltage 201.12.1.103
Input noise 201.12.1.104
Frequency response 201.12.1.105
Common mode rejection 201.12.1.106
Indication of inoperable ELECTROENCEPHALOGRAPH 201.12.4.101
282
283 Box Note 1:
284 National committees are respectfully requested to provide feedback which of the requirements in
285 clauses 201.12.1.102 to 201.12.1.106 constitute the essential performance of an
286 ELECTROENCEPHALOGRAPH.287 (A) Just, 201.12.1.102 Accuracy of signal reproduction
288 (B) All currently listed clauses 201.12.1.102 to 201.12.1.106
289 (C) Any other subset of 201.12.1.102 to 201.12.106; please specify
290 Please provide rationale for your answer.
291
292 201.5 General requirements for testing of ME EQUIPMENT
293 Clause 5 of the general standard applies, except as follows:
294 201.5.4 Other conditions
295 Addition:
296 aa) If necessary for the purpose of conducting the test, the INTERNAL ELECTRICAL POWER SOURCE may
297 be replaced by an external battery or d.c. power supply to provide the necessary test voltage.
298 bb) The values used in test circuits, unless otherwise specified, shall have at least an accuracy as
299 given below:300 – resistors: 1 %;
301 – capacitors: 10 %;
302 – inductors: 10 %;
303 – test voltages: 1 %.
---------------------- Page: 9 ----------------------
62D/1488/CDV – 10 – IEC CDV 60601-2-26 © IEC:2017
304 201.5.8 *Sequence of tests
305 Amendment:
306 Tests called for in 201.8.5.5.1 of this particular standard and in 8.5.5 of the general standard shall be
307 carried out prior to the LEAKAGE CURRENT and dielectric strength tests described in subclauses 8.7 and
308 8.8 of the general standard and prior to the tests specified in subclauses 201.12.1.102 to
309 201.12.1.106 of this particular standard.310 201.6 Classification of ME EQUIPMENT and ME SYSTEMS
311 Clause 6 of the general standard applies, except as follows:
312 201.6.6 Mode of operation
313 Replacement:
314 ELECTROENCEPHALOGRAPHS shall be classified for CONTINUOUS OPERATION.
315 201.7 ME EQUIPMENT identification, marking and documents
316 Clause 7 of the general standard applies, except as follows:
317 201.7.2.1 * Minimum requirements for marking on ME EQUIPMENT and on interchangeable
318 parts319 Addition:
320 If the ELECTROENCEPHALOGRAPH is specified as being protected against the effects of defibrillation:
321 Parts of the ELECTROENCEPHALOGRAPH (for example PATIENT CABLES) specified as being protected
322 against the effects of defibrillation shall be marked with symbol 26 or 27 of Table D.1 in Appendix D of
323 the general standard according to the classification as TYPE BF APPLIED PART or TYPE CF APPLIED PART.
324 201.7.9.2 Instructions for use325 201.7.9.2.2 Warning and safety notices
326 Addition:
327 If protection against the effects of defibrillation is provided (see 201.8.5.5.1), the instructions for use
328 shall include a warning that defibrillator protection requires the use of MANUFACTURER-specified
329 ACCESSORIES, including PATIENT CABLES and LEAD WIRES.330 Additional subclause:
331 201.7.9.2.101 Additional instructions for use
332 The instructions for use shall also include:
333 a) The INTENDED USE/INTENDED PURPOSE including environment of use.
334 Likely misuse should be identified by RISK ANALYSIS and disclosed, if necessary (e.g. ‘not suitable
335 for electro-cerebral inactivity (ECI) determination’).336 b) The procedures necessary for safe operation.
337 c) That conductive parts of ELECTRODES and associated connectors for APPLIED PARTS, including the
338 NEUTRAL ELECTRODE, should not contact other conductive parts including earth.
339 d) Information whether the ELECTROENCEPHALOGRAPH incorporates means to protect the PATIENT
340 against burns when used with HF SURGICAL EQUIPMENT and advice regarding the location of
341 ELECTRODES and LEAD WIRES etc, to reduce the HAZARD of burns in the event of a defect in the
342 neutral electrode connection of the HF SURGICAL EQUIPMENT.---------------------- Page: 10 ----------------------
IEC CDV 60601-2-26 ©IEC: 2017 –11 – 62D/1488/CDV
343 e) * The need for regular testing of the ELECTROENCEPHALOGRAPH and its ACCESSORIES.
344 f) Precautions to take when using a defibrillator on a PATIENT, if APPLIED PARTS not being protected
345 against the effects of defibrillation are being used; a description of how the discharge of a
346 defibrillator affects the ELECTROENCEPHALOGRAPH.347 g) The subsequent operation of the ELECTROENCEPHALOGRAPH after interruption of SUPPLY MAINS
348 exceeding 30 s (see 201.11.8).349 h) Any HAZARD due to simultaneous use of other PATIENT-connected ME EQUIPMENT, for example, a
350 cardiac pacemaker or other electrical stimulators.351 i) Technical specifications for the ELECTROENCEPHALOGRAPH of sufficient detail to allow the OPERATOR
352 to understand what is being measured and any limitations. Minimally this shall include:
353 – accuracy of signal reproduction354 – input dynamic range and maximum offset voltage
355 – noise
356 – frequency range and bandwidth;
357 – common mode rejection
358 – a description of all functions;
359 – a description of waveform displays (if applicable).
360 j) * Any known susceptibilities to electromagnetic phenomena.
361 k) If applicable, limitations of multipurpose channels (e.g. that these channels are not suitable for
362 monitoring and of ECG or EMG) and to which clauses of applicable standards (e.g. IEC 80601-2-
363 xx) they were tested, if any.364 201.8 Protection against electrical HAZARDS from ME EQUIPMENT
365 Clause 8 of the general standard applies, except as follows:
366 201.8.1 Fundamental rule of protection against electrical shock
367 A
...
Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.