ISO 4259-1:2026
(Main)Petroleum and related products — Precision of measurement methods and results — Part 1: Determination of precision data in relation to methods of test
- Abstract
This document specifies the methodology for the design, planning, and execution of an interlaboratory study (ILS) and calculation of precision estimates of a test method specified by the study. In particular: it defines the relevant statistical terms, it specifies the procedures to be adopted in the planning and execution of an ILS to determine the precision of a test method, and it specifies the method of calculating the precision from the results of such a study. The procedures in this document have been designed specifically for petroleum and petroleum related products, which are normally considered as homogeneous. However, the procedures described in this document can also be applied to other types of homogeneous products.
- Status
- Published
- Publication Date
- 08-Jun-2026
- Current Stage
- 9092 - International Standard to be revised - 90 Review
- Start Date
- 29-Sep-2026
Buy Documents
ISO 4259-1:2026 - Petroleum and related products — Precision of measurement methods and results — Part 1: Determination of precision data in relation to methods of test
ISO 4259-1:2026 - Produits pétroliers et connexes — Fidélité des méthodes de mesure et de leurs résultats — Partie 1: Détermination des valeurs de fidélité relatives aux méthodes d'essai
ISO 4259-1:2026 is an ISO International Standard for petroleum and related products that sets out how to design, plan, run, and analyse an interlaboratory study (ILS) to determine precision data for a test method. It defines the statistical terms used, the study sequence, and the calculation of repeatability and reproducibility from the study results. The procedures are written for petroleum and petroleum related products treated as homogeneous, and they can also be applied to other homogeneous products.
What does ISO 4259-1:2026 specify?
ISO 4259-1:2026 specifies the methodology for an ILS used to establish precision estimates for a test method. It covers the planning of the study, the execution of the study, and the calculation of precision from the results.
The document is organized into clauses for scope, normative references, terms and definitions, planning stages, statistical treatment of ILS results, analysis of variance and precision estimates, and the R/r ratio. The annexes provide sample-size rules, notation and tests, worked examples, transformation models, regression analysis, rounding rules, and the GESD outlier technique.
| Annex | What it covers |
|---|---|
| Annex A | Determination of the number of samples required |
| Annex B | Rationale for the number of laboratories and samples needed to reach 30 degrees of freedom |
| Annex C | Notation, sums of squares, variances, and tests |
| Annex D | Worked Bromine Number example and statistical tables |
| Annex E | Types of dependence and corresponding transformations |
| Annex F | Weighted linear regression analysis |
| Annex G | Rules for rounding |
| Annex H | GESD technique for multiple outliers |
| Annex I | Glossary |
What are the key requirements of ISO 4259-1:2026?
ISO 4259-1:2026 is built around a practical sequence: prepare the method, run a pilot study if needed, plan and execute the ILS, then analyse the results statistically and express repeatability and reproducibility. The requirements matter because the final precision statement is only as sound as the study design and the treatment of outliers, missing values, and transformations.
Prepare the method before the ILS
ISO 4259-1:2026 requires a draft method of test that contains all details needed to carry out the test and report the results. Any condition that could alter the results has to be specified.
In practice, this means the precision study should not be used to compensate for an incomplete method. The document expects the method to be clear enough that laboratories can follow the same procedure before precision is measured.
Use a pilot study when needed
ISO 4259-1:2026 calls for a pilot study with at least two laboratories when a preliminary check is needed. The pilot study is used to verify the operation of the test, check how well operators follow the instructions, assess sample precautions, and estimate precision roughly.
The pilot study uses at least two samples and at least 12 laboratory/sample combinations, with each sample tested twice by each laboratory under repeatability conditions. That matters because it reveals practical problems early, before the full ILS is run.
Plan the ILS with enough laboratories and samples
ISO 4259-1:2026 requires at least six participating laboratories, and recommends eight or more so the final precision is based on a broader user population. It also requires at least six samples, chosen to cover the measurement range and the types of materials to which the method applies.
The document also ties the study design to degrees of freedom. It calls for at least 30 degrees of freedom for both repeatability and reproducibility, and it gives a table and formulae for choosing the number of samples from the number of laboratories and from pilot-study variance ratios. In practice, this is what makes the precision estimates stable enough to use in a test method.
Control sample leverage and material scope
ISO 4259-1:2026 asks the planner to assess the leverage of each planned sample and to keep it within the stated limit. The sample set should cover the property range at roughly equidistant intervals and should not be over-dependent on one extreme sample.
This matters because an unbalanced sample set can distort the estimated relationship between precision and level. The document also notes that separate ILS work should be considered when different material types show different precision behaviour.
Execute the ILS under controlled conditions
ISO 4259-1:2026 assigns one coordinator to the whole ILS, from distribution of the method and samples to the final review of results. The coordinator must know the method but must not personally take part in the testing.
The samples are to be blind coded, and the laboratories are told how to handle and store them, the order in which to test them, the reporting format, the time window, and the requirement to work under normal conditions with routine qualified operators. The blind coding and random test order matter because they help keep repeat results independent and reduce bias from knowing what is being tested.
Screen and analyse the ILS results statistically
ISO 4259-1:2026 recommends validated software and statistical guidance for the data treatment. The statistical workflow starts with a GESD pre-screen for gross outliers, followed by checks for dependence on level, uniformity of repeatability and reproducibility, rejection of complete sample data when needed, estimation of missing values, and a laboratory-level outlier test.
The document then moves to analysis of variance, calculation of repeatability and reproducibility, and expression of the precision estimates. It also includes rules for confirming whether a transformation is appropriate and whether a sample is unduly influential. In practice, this is the part that turns raw laboratory data into usable precision numbers.
State the method scope and reporting limits
ISO 4259-1:2026 requires the test method to state its scope based on the achieved r and R values, the measurement range, and the reporting resolution. It also requires reporting limits so that valid single results can be recognized within the method scope.
The document defines the lower and upper scope limits using the retained sample means and the achievable result limits, and it sets acceptable single-result limits around the scope boundaries. That matters because users need to know when a reported result is still inside the range supported by the precision study.
Check the R/r ratio
ISO 4259-1:2026 uses the R/r ratio as an indicator of how much laboratory bias contributes to reproducibility. If the ratio exceeds 4, the situation is to be brought to the attention of the test method developers.
That is a practical warning sign: when R is much larger than r, the spread between laboratories is not mainly random noise. The method may need more harmonization between laboratories.
What terms does ISO 4259-1:2026 define?
ISO 4259-1:2026 defines the statistical language used to design and interpret the study. These are the terms that matter most when reading the precision tables and the annexes.
- analysis of variance - a technique that breaks total variance into component factors, which is the basis for the precision calculations.
- accepted reference value - an agreed reference value for a property, often derived from an interlaboratory study.
- blind coding - giving each sample a different number so the operator cannot identify it.
- interlaboratory study (ILS) - a study run by multiple laboratories to estimate repeatability and reproducibility from their results.
- precision - the closeness of agreement between results obtained under stated conditions.
- repeatability - the expected closeness of two independent results in the same laboratory, by the same operator, with the same equipment, over a short time.
- reproducibility - the expected closeness of two independent results from different laboratories under different normal operating conditions.
- outlier - a result far enough from the others to be treated as not part of the set.
Who uses ISO 4259-1:2026?
ISO 4259-1:2026 is used by laboratories that develop, validate, or run test methods for petroleum and related products. It is also used by method developers, interlaboratory study coordinators, statisticians, and quality control personnel who need precision data for conformity work.
In practice, the document supports tasks such as designing an ILS, checking whether a test method can be followed consistently, identifying outliers, estimating repeatability and reproducibility, and writing the precision clause in a test method. It is especially relevant where commercial petroleum products are assessed by standard laboratory test methods for quality control and conformity to specifications.
What changed in ISO 4259-1:2026 from the previous edition?
ISO 4259-1:2026 is the second edition and cancels and replaces ISO 4259-1:2017. It also incorporates ISO 4259-1:2017/Amd 1:2019 and ISO 4259-1:2017/Amd 2:2020.
The main changes stated in the document are:
- the sufficient number of samples in the ILS has been modified to at least six
- Formula (2) has been modified so that no sample has leverage exceeding 4/n
- the explanation for choosing 30 as the minimum degrees of freedom has been improved
Which standards are used with ISO 4259-1:2026?
ISO 4259-1:2026 refers to a small set of related standards and practices that support its study design and data treatment.
- ISO 5725-2 - provides the basic method for determining repeatability and reproducibility of a standard measurement method, and is the direct normative reference.
- ISO 4259-2:2026 - the companion part in the ISO 4259 series; it covers the interpretation and application of precision data in relation to methods of test.
- ASTM D3244 - gives a practice for using test data to determine conformance with specifications, and is cited as related background.
- ASTM D6300 - provides a practice for determining precision and bias data for petroleum products and lubricants.
- ASTM D7778 - gives guidance for conducting an interlaboratory study to determine the precision of a test method.
- ASTM D7915 - gives the practice for applying the GESD technique to identify multiple outliers.
What does the ISO 4259-1:2026 document contain?
ISO 4259-1:2026 contains the statistical machinery needed to turn an ILS into precision data. It includes the rules for sample size, the notation used in the calculations, and the tests used to screen outliers and judge uniformity.
The worked material in Annex D uses Bromine Number data to show how the procedures are applied in practice, from transformation selection through variance analysis and precision estimation. The document also contains tables of critical values for Cochran’s criterion, Hawkins’ test, and the normal distribution, plus the formulas used for weighted regression, leverage assessment, and rounding.
For a reader implementing the standard, the most useful parts are the study-design rules in Clauses 4 and 5, the precision calculations in Clause 6, and Annexes C, D, E, F, G, and H as the working toolbox.
Relations
- Effective Date
- 01-Jul-2026
- Effective Date
- 12-Feb-2026
- Effective Date
- 17-Feb-2024
- Effective Date
- 17-Feb-2024
- Effective Date
- 17-Feb-2024
Buy Documents
ISO 4259-1:2026 - Petroleum and related products — Precision of measurement methods and results — Part 1: Determination of precision data in relation to methods of test
ISO 4259-1:2026 - Produits pétroliers et connexes — Fidélité des méthodes de mesure et de leurs résultats — Partie 1: Détermination des valeurs de fidélité relatives aux méthodes d'essai
Get Certified
Connect with accredited certification bodies for this standard

Element Materials Technology
Materials testing and product certification.
ABS Group Brazil
ABS Group certification services in Brazil.

ABS Quality Evaluations Inc.
American Bureau of Shipping quality certification.
Sponsored listings
Frequently Asked Questions
ISO 4259-1:2026 is a standard published by the International Organization for Standardization (ISO). Its full title is "Petroleum and related products — Precision of measurement methods and results — Part 1: Determination of precision data in relation to methods of test". This standard covers: This document specifies the methodology for the design, planning, and execution of an interlaboratory study (ILS) and calculation of precision estimates of a test method specified by the study. In particular: it defines the relevant statistical terms, it specifies the procedures to be adopted in the planning and execution of an ILS to determine the precision of a test method, and it specifies the method of calculating the precision from the results of such a study. The procedures in this document have been designed specifically for petroleum and petroleum related products, which are normally considered as homogeneous. However, the procedures described in this document can also be applied to other types of homogeneous products.
This document specifies the methodology for the design, planning, and execution of an interlaboratory study (ILS) and calculation of precision estimates of a test method specified by the study. In particular: it defines the relevant statistical terms, it specifies the procedures to be adopted in the planning and execution of an ILS to determine the precision of a test method, and it specifies the method of calculating the precision from the results of such a study. The procedures in this document have been designed specifically for petroleum and petroleum related products, which are normally considered as homogeneous. However, the procedures described in this document can also be applied to other types of homogeneous products.
ISO 4259-1:2026 is classified under the following ICS (International Classification for Standards) categories: 75.080 - Petroleum products in general. The ICS classification helps identify the subject area and facilitates finding related standards.
ISO 4259-1:2026 has the following relationships with other standards: It is inter standard links to EN ISO 4259-2:2026, EN ISO 4259-1:2026, ISO 4259-1:2017, ISO 4259-1:2017/Amd 2:2020, ISO 4259-1:2017/Amd 1:2019. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
ISO 4259-1:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 analysis of variance
- 3.2 accepted reference value
- 3.3 between laboratory variance
- 3.4 bias
- 3.5 blind coding
- 3.6 degrees of freedom
- 3.7 determination
- 3.8 interlaboratory study
- 3.9 known value
- 3.10 mean
- 3.11 mean square
- 3.12 normal distribution
- 3.13 operator
- 3.14 outlier
- 3.15 precision
- 3.16 random error
- 3.17 repeatability
- 3.18 reproducibility
- 3.19 result
- 3.20 standard deviation
- 3.21 sum of squares
- 3.22 true value
- 3.23 variance
- 4 Stages in the planning of an interlaboratory study for the determination of the
- 4.1 General
- 4.2 Preparing a draft method of test
- 4.3 Planning a pilot study with at least two laboratories
- 4.4 Planning the ILS
- 4.5 Executing the ILS
- 5 Statistical treatment of ILS results
- 5.1 General
- 5.2 Pre-screen using GESD technique
- 5.3 Transformation of data and outlier tests
- 5.4 Rejection of complete data (from all laboratories) for a sample
- 5.5 Estimating missing or rejected values
- 5.6 Rejection test for outlying laboratories
- 5.7 Confirmation of selected transformation
- 6 Analysis of variance, calculation and expression of precision estimates
- 6.1 General
- 6.2 Analysis of variance
- 6.3 Expectation of mean squares and calculation of precision estimates
- 6.4 Expression of precision estimates of a method of test
- 6.5 Specification of scope for the test method
- 6.6 Reporting limits instruction for the test method
- 7 R/r ratio
- Annex A
- Annex B
- 2 SP1
- Annex C
- Annex D
- Annex E
- 1. D = C(m + B) Y = ln(x + B) Care shall be taken if
- 2. D = Cm y = x The fitted line will pass
- 3. D = C(m + B ) y = (x + B ) The fitted line will not pass
- 6. DC y = arctan 22 pass through the origin. If
- Annex F
- 1 S 2 S 2 S S
- 1 S 2 S 2 S S
- Annex G
- Annex H
- Annex I
- 6.3.2 at least one result
- Bibliography
Introduction
For purposes of quality control and to check conformity with specifications, the properties of commercial petroleum products are assessed by standard laboratory test methods. Two or more measurements of the same property of a specific sample by a specific test method, or, by different test methods that purport to measure the same property, will not usually give exactly the same result. It is therefore necessary to take proper account of this fact, by arriving at statistically based estimates of the precision for a method, i.e. an objective measure of the degree of agreement expected between two or more results obtained in specified circumstances.
The ISO 4259 series encompasses both the derivation of precision estimates and the application of precision [4] data. The ISO 4259 series combines the information in ASTM D6300 regarding the determination of the [3] precision estimates and the information in ASTM D3244 for the utilization of test data. Careful investigations should be carried out before applying this document to products for which the assumption of homogeneity can be questioned.
[1] A glossary of the variables used in this document and ISO 4259-2 is included for information as Annex I.
International Standard ISO 4259-1:2026(en) Petroleum and related products — Precision of measurement methods and results — Part 1:
Determination of precision data in relation to methods of test
1 Scope
This document specifies the methodology for the design, planning, and execution of an interlaboratory study (ILS) and calculation of precision estimates of a test method specified by the study. In particular:
- it defines the relevant statistical terms,
- it specifies the procedures to be adopted in the planning and execution of an ILS to determine the precision of a test method, and
- it specifies the method of calculating the precision from the results of such a study. The procedures in this document have been designed specifically for petroleum and petroleum related products, which are normally considered as homogeneous. However, the procedures described in this document can also be applied to other types of homogeneous products.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 5725-2, Accuracy (trueness and precision) of measurement methods and results — Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
3.1 analysis of variance
technique that enables the total variance of a method to be broken down into its component factors
3.2 accepted reference value
agreed-upon reference value for a specific property of a material determined using an accepted reference method and protocol, e.g. derived from an interlaboratory study (3.8)
ISO 4259-1:2026(en)
3.3 between laboratory variance
component of the total variance (3.23) attributable to the difference between the means (3.10) of different laboratories Note 1 to entry: When results obtained by more than one laboratory are compared, the scatter is usually wider than when the same number of tests is carried out by a single laboratory, and there is some variation between means obtained by different laboratories. These give rise to the between laboratory variance which is that component of the overall variance due to the difference in the means obtained by different laboratories. Note 2 to entry: There is a corresponding definition for between operator variance. Note 3 to entry: The term “between laboratory” is often shortened to “laboratory” when used to qualify representative parameters of the dispersion of the population of results, for example as “laboratory variance”.
3.4 bias
of a test method> difference between the population mean of test results (3.19) from a very large number of different laboratories for the property of a material obtained using a specific test method versus the accepted reference value (3.2) for the property where this is available Note 1 to entry: See Note 1 to entry in 3.12 for an interpretation of “population mean of test results”.
3.5 blind coding
assignment of a different number to each sample so that no other identification or information on any sample is given to the operator (3.13)
3.6 degrees of freedom
divisor used in the calculation of variance (3.23) Note 1 to entry: This definition applies strictly only in the simplest cases. Definitions for more complex cases are beyond the scope of this document.
3.7 determination
process of carrying out the series of operations specified in a test method, whereby a single value is obtained
3.8 interlaboratory study
ILS study specifically designed to estimate the repeatability and reproducibility of a test method achieved at a fixed point in time by multiple laboratories through the statistical analysis of their test results (3.19) obtained on aliquots prepared from multiple materials
3.9 known value
quantitative value for a property that can be theoretically derived or calculated by the preparation of the sample Note 1 to entry: The known value does not always exist, for example for empirical tests such as flash point.
3.10 mean
sum of a set of results (3.19) divided by the number of results
3.11 mean square
sum of squares (3.21) divided by the degrees of freedom (3.6)
ISO 4259-1:2026(en)
3.12 normal distribution
probability distribution of a continuous random variable, x, such that, if x is any real number, the probability density is as shown in Formula (1):
1 1 x
fx exp , x (1)
2
()2
Note 1 to entry: In the context of modelling a distribution of test results, μ is the population mean (3.10), or true value (3.22) of the property as determined by a specific test method; σ is the standard deviation (3.20) of the normal distribution used to describe the distribution of an infinite number of test results obtained using the same test method by an infinite number of laboratories (σ > 0).
3.13 operator
person who normally and regularly carries out a particular test
3.14 outlier
result (3.19) far enough in magnitude from other results to be considered not a part of the set
3.15 precision
closeness of agreement between the results (3.19) obtained by applying the same test procedure several times on essentially the same materials and under prescribed conditions Note 1 to entry: The smaller the random part of the experimental error, the more precise the procedure.
3.16 random error
component of measurement error that in replicate measurements varies in an unpredictable manner
3.17 repeatability
quantitative expression for the random error (3.16) associated with the difference between two independent results (3.19) obtained under repeatability conditions in the normal and correct operation of the same method, that is expected to be exceeded with an approximate probability of 5 % Note 1 to entry: The representative parameter for the dispersion of the population that can be associated with these results is repeatability standard deviation (3.20) or repeatability variance (3.23). Repeatability refers to the maximum difference attributable to random variation between two results obtained under the state of minimum random variability. Therefore, the period of time during which repeat results are to be obtained should be short enough to exclude time dependent variation, for example, variation caused by environmental changes, or variation associated with multiple calibrations.
Note 2 to entry: The term “repeatability” is not to be confused with the terms “between repeats” or “repeats”. Note 3 to entry: The conditions where independent test results (3.19) are obtained using the same method for test material are considered to be the same in the same laboratory by the same operator (3.13) using the same equipment with short intervals of time. This is also known as the repeatability conditions.
3.18 reproducibility
quantitative expression for the random error (3.16) associated with the difference between two independent results (3.19) obtained under reproducibility conditions in the normal and correct operation of the same method, that is expected to be exceeded with an approximate probability of 5 % Note 1 to entry: The representative parameter of the dispersion of the population that can be associated with these results (3.19) is reproducibility standard deviation (3.20) or reproducibility variance (3.23). Reproducibility refers to the maximum difference attributable to random variation between two results obtained under the state of maximum random variability.
ISO 4259-1:2026(en) Note 2 to entry: The conditions where independent test results (3.19) are obtained using the same method for test material are considered to be the same in different laboratories. A different laboratory means a different operator (3.13), different equipment, different geographic location, and under different supervisory control. This is also known as the reproducibility conditions.
3.19 result
final value obtained by following the complete set of instructions in a test method Note 1 to entry: It is assumed that the result is rounded off according to the procedure specified in Annex G.
3.20 standard deviation
measure of the dispersion of a series of results (3.19) around their mean (3.10), equal to the positive square root of the variance (3.23) and estimated by the positive square root of the mean square (3.11)
3.21 sum of squares
differences between a series of results (3.19) and their mean (3.10)
3.22 true value
for practical purposes, value towards which the average of single results (3.19) obtained by n laboratories tends, as n tends towards infinity Note 1 to entry: Such a true value is associated with the particular method of test. [2] Note 2 to entry: A different and idealized definition is given in ISO 3534-2:2006, 3.2.5 .
3.23 variance
mean of the squares of the deviation of a random variable from its mean (3.10), estimated by the mean square (3.11)
4 Stages in the planning of an interlaboratory study for the determination of the
precision of a test method
4.1 General
The stages in planning an interlaboratory study (ILS) are as follows:
a) preparing a draft method of test (see 4.2); b) planning a pilot study with at least two laboratories if necessary (see 4.3); c) planning the ILS (see 4.4); d) executing the ILS (see 4.5).
4.2 Preparing a draft method of test
The draft method of test shall contain all the necessary details for carrying out the test and reporting the results. Any condition that could alter the results shall be specified in the draft. The ILS shall be designed so that it covers the intended range of the test method (see also 6.5). At this stage, a clause on precision is included in the draft method of the test only as a heading.
ISO 4259-1:2026(en)
4.3 Planning a pilot study with at least two laboratories
A pilot study is necessary for the following reasons:
a) to verify the details in the operation of the test; b) to find out how well operators can follow the instructions of the method, and thus of the ILS; c) to check the precautions regarding samples; d) to estimate approximately the precision of the test.
At least two samples are required, covering the range of results to which the test method is intended to apply; however, at least 12 laboratory/sample combinations shall be included. Each sample is tested twice by each laboratory under repeatability conditions. The samples should be equally distributed across the test method range and should include major product groups covered in the test method scope. If any omissions or inaccuracies in the draft test method are revealed, they shall now be corrected. The results shall be analysed for precision, and bias for sample(s) with accepted reference values. If either is considered to be too large, then alterations to the test method shall be considered.
4.4 Planning the ILS
There shall be at least six participating laboratories, but it is recommended this number be increased to eight or more in order to ensure the final precision is based on at least six laboratories and to ensure the precision statement is more representative of the user population. The number of samples shall be sufficient to adequately represent the types of materials to which the test method shall be applied, to cover the range of the property measured at approximately equidistant intervals, and to give reliability to the precision estimates. At least six samples shall be used in the ILS. In order to correctly estimate precision versus level relationship, it is important that the choice of samples evenly covers the range and materials for the property measured, so that an estimated relationship is not too dependent upon the leverage of a sample with extreme property value. It is strongly recommended that the leverage of each planned sample in the sample set design, lev be i, assessed using Formula (2) based on either its known value, or a single test result. No sample shall have a leverage exceeding 4/n . See Table D.11 for an example of leverage calculation (second column from the right t under heading "lev ").
1 ()xix lev (2)
n n 2 t t ()xx
k k1 where lev is leverage of sample i;
n is total number of planned samples; t x is Napierian logarithm, ln (p ), with p being the planned property level for sample i; i i i x is grand average of all x .
In any event, it is necessary to obtain at least 30 degrees of freedom for both repeatability and reproducibility (see Annex B for additional information). For repeatability, this means obtaining a total of at least 30 pairs of results in the ILS.
For reproducibility, Table A.1 shall be used for the minimum number of samples required in terms of L, P and Q, where L is the number of participating laboratories, and P and Q are the ratios of variance component estimates obtained from the pilot study. Specifically, P is the ratio of the interaction component to the repeats component and Q is the ratio of the laboratories component to the repeats component. Annex B gives the derivation of the formula used. If Q is much larger than P, then 30 degrees of freedom cannot be achieved; the blank entries in Table A.1 correspond to this situation (i.e. when more than 20 samples are required). For these cases, there is likely to be a significant bias between laboratories.
ISO 4259-1:2026(en) In the absence of pilot study information to permit the use of Table A.1, the number of samples shall be at least six, and chosen such that the number of laboratories times the number of samples is greater than or equal to 42.
When it is known or suspected that different types of materials exhibit different precision functional forms when tested by the test method, consideration should be given to conducting separate ILS for each type of material.
4.5 Executing the ILS
One person shall be responsible for the entire ILS, from the distribution of the texts of the test method and samples to the final appraisal of the results. This person shall be familiar with the test method, but shall not personally take part in the tests.
The text of the test method shall be distributed to all the laboratories in time to allow any queries to be raised before the tests begin. If any laboratory wants to practice the method in advance, than this shall be carried out with samples other than those used in the ILS.
The samples shall be accumulated, subdivided and distributed by the coordinator, who shall also keep a reserve of each sample for emergencies. It is most important that the individual laboratory portions be homogeneous and stable for the property of interest throughout the entire duration of the ILS. Prior to distribution, the ILS sample set shall be blind coded in a manner that preserves the anonymity of the nature of the test material and the expected value of the property. The following information shall be sent with the ILS sample set.
a) Agreed (draft) method of test.
b) Handling and storage requirements for the samples.
c) Order in which the samples shall be tested. A different random order for each laboratory is highly recommended. For large number of laboratories, several unique test orders may be randomly assigned to groups of laboratories, with no more than 4 laboratories per group. d) For statistical reasons, the repeat results shall be obtained independently of each other, i.e. that the second result is not biased by knowledge of the first. This is achieved by blind coding where the repeat for each material in the ILS design is included in the test set sent to ILS participants without disclosing that it is a repeat, with an accompanying statement that a single result shall be obtained on each sample in the test set, in the specified testing order, by the same operator with the same apparatus within a short time. If this blind coding is regarded as infeasible to achieve, then the statement shall state that a pair of results associated with a sample shall be obtained by the same operator with the same apparatus within a short time, without disclosing the nature of the sample. e) Period of time within which all the samples shall be tested. f) Blank form for reporting the results. For each sample, there shall be space for the date of testing, the test results, and any unusual occurrences. The unit of accuracy for reporting the results shall be specified. g) Statement that the test shall be carried out under normal conditions, using qualified operators who carry out this kind of test routinely and that the duration of the test shall be the same as normal. h) A questionnaire requesting information on the conditions used in the application of the test method, e.g. apparatus details, reagents and materials, calibration and verification procedures, quality control procedure, any deviations from either the test method or the instructions supplied, observations and suggestions for future improvement of the test method.
Operators that participated in the pilot study may also participate in the ILS. If their extra experience in testing a few more samples produces a noticeable effect, it serves as a warning that the test method is not satisfactory. They shall be identified in the report of the results so that any effect can be noted. [5] [4] NOTE For additional guidance on the planning and execution of an ILS, consult ASTM D7778 and ASTM D6300 .
ISO 4259-1:2026(en)
5 Statistical treatment of ILS results
5.1 General
Although the procedures described in Clauses 5 and 6 are in a form suitable for hand calculation, it is recommended that these procedures be carried out using an electronic computer with appropriately validated software designed specifically to store and analyse ILS test results based on the procedures of this document. It is also recommended that these procedures be carried out under the guidance of a statistician. [14] NOTE D2PP software was historically used for precision evaluation calculations, however this software's capability does not meet the requirements of this document, as it does not execute the generalized extreme studentized deviation (GESD) or Cook’s Distance.
In 5.2 to 5.7, procedures are specified to achieve the following:
...
Contents
- Introduction
- 1 Domaine d'application
- 2 Références normatives
- 3 Termes et définitions
- 3.1 analyse de la variance
- 3.2 valeur de référence acceptée
- 3.3 variance interlaboratoires
- 3.4 biais
- 3.5 codage à l’aveugle
- 3.6 degrés de liberté
- 3.7 détermination
- 3.8 essai interlaboratoires
- 3.9 valeur connue
- 3.10 moyenne
- 3.11 moyenne des carrés
- 3.12 distribution normale
- 3.13 opérateur
- 3.14 valeur aberrante
- 3.15 fidélité
- 3.16 erreur aléatoire
- 3.17 répétabilité
- 3.18 reproductibilité
- 3.19 résultat
- 3.20 écart-type
- 3.21 somme des carrés
- 3.22 valeur vraie
- 3.23 variance
- 4 Étapes de l'organisation d'un essai interlaboratoires pour la détermination de la
- 4.1 Généralités
- 4.2 Préparation d'un projet de méthode d'essai
- 4.3 Organisation d'une étude pilote avec au moins deux laboratoires
- 4.4 Organisation de l’ILS
- 4.5 Exécution de l’ILS
- 5 Traitement statistique des résultats de l’ILS
- 5.1 Généralités
- 5.2 Prévisualisation selon la technique GESD
- 5.3 Transformation des données et recherche des valeurs aberrantes
- 5.4 Rejet de tous les résultats (de tous les laboratoires) concernant un échantillon
- 5.5 Estimation des valeurs manquantes ou rejetées
- 5.6 Test de rejet des résultats des laboratoires aberrants
- 5.7 Confirmation de la transformation sélectionnée
- 6 Analyse de la variance, calcul et expression des estimations de fidélité
- 6.1 Généralités
- 6.2 Analyse de la variance
- 1 L S S
- 6.3 Espérance des carrés moyens et calcul des estimations de fidélité
- 6.4 Expression de l’estimation de la fidélité dans une méthode d'essai
- 6.5 Spécification du domaine d’application pour la méthode d’essai
- 6.6 Instructions relatives aux limites de déclaration pour la méthode d'essai
- 7 Ratio R/r
- Annexe A
- Annexe B
- Annexe C
- Annexe D
- 1 L S
- Annexe E
- 2. D = Cm y = x ln(D) = b + b ln(m) + b T La ligne sur laquelle s'ajuster va passer
- 5. DC1 y = ln
- Annexe F
- 1 S 2 S 2 S S
- 1 S 2 S 2 S S
- Annexe G
- Annexe H
- Annexe I
- 7.2.2 A limite inférieure de la spécification (17), (B.2)
- 7.2.2 b coefficients (D.6)
- 6.3.2 moins un résultat
- Bibliographie
Introduction
Pour les besoins de contrôle de qualité et pour vérifier leur conformité aux spécifications, les caractéristiques des produits pétroliers commerciaux sont contrôlées au moyen de méthodes d'essai normalisées de laboratoire. Deux ou plusieurs déterminations de la même caractéristique d'un échantillon donné, selon une méthode d'essai spécifique ou selon des méthodes d’essai différentes qui prétendent mesurer la même caractéristique, ne donneront généralement pas exactement le même résultat. Il est donc nécessaire de tenir compte correctement de ce fait, en parvenant à des estimations fondées sur les statistiques de la fidélité d'une méthode, qui constituent une mesure objective du degré de concordance attendu entre deux ou plusieurs résultats obtenus dans des conditions données.
La série ISO 4259 regroupe à la fois la détermination d’estimations de fidélité et l’application des données [4] de fidélité. La série ISO 4259 combine les informations de l’ASTM D6300 concernant la détermination [3] d’évaluations de la fidélité et les informations de l’ASTM D3244 pour l’utilisation des données d’essai. Il convient de mener des études approfondies avant d'appliquer le présent document à des produits pour lesquels l'hypothèse d'homogénéité peut être remise en cause.
[1] Un glossaire des variables utilisées dans ce document et dans l'ISO 4259-2 est inclus pour information en Annexe I de ce document.
Norme internationale ISO 4259-1:2026(fr) Produits pétroliers et connexes — Fidélité des méthodes de mesure et de leurs résultats — Partie 1:
Détermination des valeurs de fidélité relatives aux méthodes d'essai
1 Domaine d'application
Le présent document spécifie la méthodologie pour la conception, l’organisation et la réalisation d’un essai interlaboratoires (ILS) et pour le calcul des estimations de fidélité d’une méthode d’essai spécifiée par cet ILS. En particulier:
- il définit les termes statistiques appropriés,
- il spécifie les procédures à suivre dans l'organisation et l’exécution d'un ILS destiné à déterminer la fidélité d'une méthode d'essai, et
- il spécifie la méthode de calcul de la fidélité à partir des résultats d'un tel ILS. Les procédures du présent document ont été conçues spécifiquement pour les produits pétroliers et connexes qui sont normalement considérés homogènes. Les procédures décrites dans le présent document peuvent cependant aussi s’appliquer à d’autres types de produits homogènes.
2 Références normatives
Les documents suivants sont cités dans le texte de sorte qu’ils constituent, pour tout ou partie de leur contenu, des exigences du présent document. Pour les références datées, seule l’édition citée s’applique. Pour les références non datées, la dernière édition du document de référence s'applique (y compris les éventuels amendements).
ISO 5725-2, Exactitude (justesse et fidélité) des résultats et méthodes de mesure — Partie 2: Méthode de base pour la détermination de la répétabilité et de la reproductibilité d'une méthode de mesure normalisée
3 Termes et définitions
Pour les besoins du présent document, les termes et définitions suivants s’appliquent. L’ISO et l’IEC tiennent à jour des bases de données terminologiques destinées à être utilisées en normalisation, consultables aux adresses suivantes:
- ISO Online browsing platform: disponible à l’adresse https:// www .iso .org/ obp
- IEC Electropedia: disponible à l’adresse https:// www .electropedia .org/
3.1 analyse de la variance
technique qui permet de décomposer la variance totale d'une méthode en ses différents facteurs composants
ISO 4259-1:2026(fr)
3.2 valeur de référence acceptée
valeur de référence convenue pour une propriété spécifique d'un produit déterminé en suivant une méthode et un protocole de référence acceptés, c'est-à-dire provenant d’un essai interlaboratoires (3.8)
3.3 variance interlaboratoires
constituant d’une variance (3.23) totale attribuable à la différence entre les moyennes (3.10) des différents laboratoires Note 1 à l'article: Lorsque des résultats obtenus par plus d'un laboratoire sont comparés, la dispersion est normalement plus importante que lorsque le même nombre d'essais est effectué par un seul laboratoire, et il y a des écarts entre les moyennes obtenues par les différents laboratoires. Ces écarts donnent lieu à la variance interlaboratoires qui est l'élément de la variance totale dû aux différences entre les moyennes des différents laboratoires. Note 2 à l'article: Il existe une définition correspondante pour la variance entre opérateurs. Note 3 à l'article: Le terme «interlaboratoires» est souvent abrégé en «laboratoires» lorsqu'il est utilisé pour qualifier des paramètres représentatifs de la dispersion de la population de résultats, par exemple sous forme de «variance laboratoires».
3.4 biais
<d’une méthode d’essai> différence entre la moyenne des résultats (3.19) d’essai d’une population à partir d’un très large nombre de laboratoires différents pour la détermination d’une caractéristique d’un produit obtenue suivant une méthode d’essai spécifique et la valeur de référence acceptée (3.2) de cette caractéristique, lorsque celle-ci est disponible Note 1 à l'article: Voir la Note 1 à l’article en 3.12 pour une interprétation de «moyenne des résultats d’essai d’une population».
3.5 codage à l’aveugle
attribution d'un numéro différent pour chaque échantillon, afin qu'aucune autre identification ou information sur les échantillons ne soit donnée à l'opérateur (3.13)
3.6 degrés de liberté
diviseur utilisé dans le calcul de la variance (3.23) Note 1 à l'article: Cette définition n'est strictement applicable que dans les cas les plus simples. Des définitions pour des cas plus complexes sont en dehors du domaine d’application du présent document.
3.7 détermination
exécution de la série d'opérations prescrites dans une méthode d'essai et permettant d'obtenir une valeur unique
3.8 essai interlaboratoires
ILS étude spécifiquement conçue pour évaluer la répétabilité et la reproductibilité d'une méthode d’essai réalisée à un moment fixé par de multiples laboratoires au moyen de l'analyse statistique de leurs résultats (3.19) obtenus sur des échantillons préparés à partir de multiples produits
3.9 valeur connue
valeur quantitative pour une caractéristique qui peut théoriquement être établie ou calculée à partir de la préparation de l'échantillon Note 1 à l'article: La valeur connue n'existe pas toujours, par exemple dans le cas d'essais empiriques tels que la détermination du point d'éclair.
ISO 4259-1:2026(fr)
3.10 moyenne
somme d’une série de résultats (3.19) divisée par le nombre de résultats
3.11 moyenne des carrés
somme des carrés (3.21) divisée par le nombre de degrés de liberté (3.6)
3.12 distribution normale
distribution de probabilité d'une variable aléatoire continue x, telle que si x est un nombre réel quelconque, la densité de probabilité est telle que donnée dans la Formule (1):
1 1 x
fx exp , x (1)
2
()2
Note 1 à l'article: Dans le cadre d’une modélisation d’une distribution des résultats de l’essai, μ est la moyenne (3.10) de la population ou la valeur vraie (3.22) de la caractéristique telle que déterminée par une méthode d’essai spécifique; σ est l'écart-type (3.20) de la distribution normale utilisé pour décrire la distribution d'un nombre infini de résultats d’essai obtenus suivant la même méthode d’essai pratiquée par un nombre infini de laboratoires (σ > 0).
3.13 opérateur
personne qui effectue normalement et régulièrement un essai particulier
3.14 valeur aberrante
résultat (3.19) dont la valeur est suffisamment éloignée des autres résultats pour qu'il ne soit pas considéré comme faisant partie de l'ensemble des résultats
3.15 fidélité
étroitesse de l'accord entre les résultats (3.19) obtenus en appliquant la même procédure d’essai à plusieurs reprises sur essentiellement les mêmes produits et dans des conditions déterminées Note 1 à l'article: Plus la part aléatoire de l'erreur expérimentale est faible, plus la procédure est précise.
3.16 erreur aléatoire
composante de l'erreur de mesure qui dans des mesures dupliquées varie de manière imprédictible
3.17 répétabilité
expression quantitative de l’erreur aléatoire (3.16) associée à la différence entre deux résultats (3.19) indépendants obtenus dans des conditions de répétabilité dans le fonctionnement normal et correct de la même méthode, qui devrait être dépassée avec une probabilité approximative de 5 % Note 1 à l'article: Le paramètre représentatif de la dispersion de la population qui peut être associé à ces résultats est l’écart-type (3.20) de répétabilité ou la variance (3.23) de répétabilité. La répétabilité se réfère à la différence maximale attribuable à la variation aléatoire entre deux résultats obtenus sous l'état de variabilité aléatoire minimale. Par conséquent, le temps pendant lequel des résultats répétés doivent être obtenus sera suffisamment court pour exclure les écarts qui dépendent du temps, par exemple, variations causées par des changements d’environnement ou par des étalonnages multiples.
Note 2 à l'article: Le terme «répétabilité» ne doit pas être confondu avec les termes «entre répétitions» ou «répétitions». Note 3 à l'article: Les conditions dans lesquelles des résultats (3.19) d’essai indépendants sont obtenus en utilisant la même méthode pour un produit d'essai considéré comme le même dans le même laboratoire, par le même opérateur (3.13) utilisant le même équipement dans de courts intervalles de temps sont considérées comme étant identiques. Elles sont aussi appelées «conditions de répétabilité».
ISO 4259-1:2026(fr)
3.18 reproductibilité
expression quantitative de l'erreur aléatoire (3.16) associée à la différence entre deux résultats (3.19) indépendants obtenus dans des conditions de reproductibilité dans le fonctionnement normal et correct de la même méthode, qui devrait être dépassée avec une probabilité approximative de 5 % Note 1 à l'article: Le paramètre représentatif de la dispersion de la population qui peut être associé à ces résultats (3.19) est l’écart-type (3.20) de reproductibilité ou la variance (3.23) de reproductibilité. La reproductibilité se réfère à la différence maximale attribuable à la variation aléatoire entre deux résultats obtenus sous l'état de variabilité aléatoire maximale.
Note 2 à l'article: Les conditions dans lesquelles des résultats (3.19) d’essai indépendants sont obtenus en utilisant la même méthode pour un produit d'essai considéré comme le même dans différents laboratoires, où un laboratoire différent signifie un opérateur (3.13) différent, un équipement différent, un emplacement géographique différent et sous un contrôle de supervision différent, sont considérées comme identiques dans différents laboratoires. Elles sont aussi appelées «conditions de reproductibilité».
3.19 résultat
valeur finale obtenue en suivant le mode opératoire complet d’une méthode d'essai Note 1 à l'article: On suppose que le résultat est arrondi conformément à la procédure spécifiée à l'Annexe G.
3.20 écart-type
mesure de la dispersion d'une série de résultats (3.19) autour de leur moyenne (3.10), égale à la racine carrée positive de la variance (3.23) et estimée par la racine carrée positive de la moyenne des carrés (3.11)
3.21 somme des carrés
différence entre une série de résultats (3.19) et leur moyenne (3.10)
3.22 valeur vraie
pour des besoins pratiques, valeur vers laquelle tend la moyenne des résultats (3.19) individuels obtenus par n laboratoires, lorsque n tend vers l'infini Note 1 à l'article: Une valeur vraie ainsi définie est associée à chaque méthode d'essai particulière. [2] Note 2 à l'article: Une définition différente et idéalisée est donnée dans l'ISO 3534-2:2006, 3.2.5 .
3.23 variance
moyenne des carrés de l'écart d'une variable aléatoire par rapport à sa moyenne (3.10), estimée par la moyenne des carrés (3.11)
4 Étapes de l'organisation d'un essai interlaboratoires pour la détermination de la
fidélité d'une méthode d'essai
4.1 Généralités
Les étapes de l'organisation d'un essai interlaboratoires (ILS) sont les suivantes:
a) préparation d'un projet de méthode d'essai (voir 4.2); b) organisation d'une étude pilote avec au moins deux laboratoires, si nécessaire (voir 4.3); c) organisation de l’ILS (voir 4.4); d) exécution de l’ILS (voir 4.5).
ISO 4259-1:2026(fr)
4.2 Préparation d'un projet de méthode d'essai
Le projet de méthode d’essai doit contenir tous les détails nécessaires pour l'exécution de l'essai et l'expression des résultats. Toute condition susceptible d'avoir une influence sur les résultats doit être spécifiée dans le projet.
L'ILS doit être conçu de manière à couvrir la gamme prévue de la méthode d'essai (voir aussi 6.5). À ce stade, seul le titre de l’article relatif à la fidélité est inclus dans le projet de méthode d’essai.
4.3 Organisation d'une étude pilote avec au moins deux laboratoires
Une étude pilote est nécessaire pour les raisons suivantes:
a) pour vérifier les détails de l'exécution de l'essai; b) pour déterminer dans quelle mesure les opérateurs peuvent observer correctement les instructions de la méthode, et par suite de l’ILS; c) pour contrôler les instructions concernant les échantillons; d) pour estimer grossièrement la fidélité de l'essai.
Au moins deux échantillons sont nécessaires, couvrant la plage de résultats pour laquelle la méthode d'essai est conçue. Cependant, au moins 12 combinaisons laboratoire/échantillon doivent être incluses. Chaque échantillon est essayé deux fois par chaque laboratoire dans les conditions de répétabilité. Il convient que les échantillons soient également distribués sur l’ensemble de l’intervalle de la méthode d’essai et qu’y soient inclus les groupes de produits majoritaires couverts dans le domaine d’application de la méthode. Si des omissions ou des imprécisions dans le projet de méthode d’essai sont révélées, elles doivent alors être corrigées. Les résultats doivent être analysés sous l'angle du biais et de la fidélité pour des échantillons avec des valeurs de référence acceptées. Si l'un ou l'autre paraît trop important, des modifications à la méthode d’essai doivent être envisagées.
4.4 Organisation de l’ILS
Celui-ci doit recueillir la participation d'au moins six laboratoires, mais il est recommandé que ce soit plutôt huit laboratoires ou plus pour s’assurer que la fidélité finale est établie sur au moins six laboratoires et assurer une meilleure représentativité de la population des utilisateurs. Le nombre d'échantillons doit être suffisant pour représenter convenablement les types de produits auxquels la méthode d’essai doit être appliquée, pour couvrir la plage de la caractéristique à des intervalles équivalents mesurée et rendre sûres les estimations de fidélité. Au moins six échantillons doivent être utilisés dans l’ILS. Afin d’évaluer correctement la relation de la fidélité et du niveau, il est important que le choix des échantillons couvre à intervalle régulier la gamme et les produits pour la caractéristique mesurée, de sorte qu’une relation estimée ne dépende pas trop de l’influence d’un échantillon ayant une valeur extrême.
Il est fortement recommandé que l’influence de chaque échantillon planifié dans la conception de l'ensemble des échantillons, lev , soit évaluée en utilisant la Formule (2) sur la base soit de sa valeur connue, soit d’un
ISO 4259-1:2026(fr) résultat d’essai unique. Aucun échantillon ne doit avoir une influence excédant 4/n . Voir le Tableau D.11 t pour un exemple de calcul d’influence (la deuxième colonne en partant de la droite sous l’en-tête «lev »).
()xix
lev (2)
n n 2 t t ()xx
k k1 où lev est l’influence de l’échantillon i;
n est le nombre total d’échantillons planifiés; t x est le logarithme népérien, ln (p), où p est le niveau de la caractéristique planifié pour i i i l’échantillon i; x est la moyenne générale de tous les x .
Dans tous les cas, il est nécessaire d'obtenir au moins 30 degrés de liberté tant pour la répétabilité que pour la reproductibilité (voir l'Annexe B pour plus d’information). Pour la répétabilité, cela implique d'obtenir un total d'au moins 30 paires de résultats dans l'ILS.
Pour la reproductibilité, le Tableau A.1 doit être utilisé pour déterminer le nombre minimal d'échantillons requis en fonction de L, P et Q, où L est le nombre de laboratoires participants et P et Q sont les rapports des estimations des composantes de variance obtenues dans l’étude pilote. Spécifiquement, P est le rapport de la composante interaction à la composante répétitions et Q est le rapport de la composante laboratoires à la composante répétitions. L'Annexe B donne le calcul de la formule utilisée. Si Q est beaucoup plus grand que P, les 30 degrés de liberté ne peuvent être atteints; les entrées vides dans le Tableau A.1 correspondent à cette situation (c'est-à-dire lorsque plus de 20 échantillons sont nécessaires). Dans ces cas, il y a vraisemblablement un biais significatif entre laboratoires.
En absence d'informations sur l’étude pilote pour permettre l'utilisation du Tableau A.1, le nombre d'échantillons doit être de 6 au moins et choisi de telle façon que le nombre de laboratoires multiplié par le nombre d'échantillons soit supérieur ou égal à 42.
Quand il est connu ou suspecté que les différents types de produits présentent différentes formes fonctionnelles de fidélité avec la méthode d’essai, il est recommandé de considérer l’organisation d’ILS séparés pour chaque type de produits.
4.5 Exécution de l’ILS
Une personne doit être responsable de l’ILS entier, depuis la distribution des textes de la méthode d’essai et des échantillons jusqu'à l'évaluation finale des résultats. Elle doit bien connaître la méthode d’essai, mais ne doit pas prendre part personnellement aux essais.
Le texte de la méthode d’essai doit être diffusé à tous les laboratoires pour leur permettre de soulever d'éventuelles questions avant le début des essais. Si un laboratoire désire pratiquer la méthode à l'avance, cela doit alors être fait sur des échantillons autres que ceux utilisés dans l’ILS. Les échantillons doivent être réunis, divisés et distribués par le coordinateur, qui doit également conserver une réserve de chaque échantillon pour les cas urgents. Il est de la plus haute importance que les parties fractionnées pour chaque laboratoire soient homogènes et stables en ce qui concerne la caractéristique évaluée sur l’ensemble de la durée du programme de l’ILS. Avant la distribution, l'ensemble des échantillons de l'ILS doivent être codés à l’aveugle de telle manière que l'anonymat de la nature du produit de l’essai soit préservé ainsi que la valeur attendue de la caractéristique. Les informations suivantes doivent accompagner l’envoi des échantillons.
a) La méthode ou le projet de méthode choisi pour les essais.
b) Les instructions pour la manipulation et le stockage des échantillons. c) L'ordre dans lequel les échantillons doivent être soumis à essai. Un ordre aléatoire différent pour chaque laboratoire est fortement recommandé. Pour un grand nombre de laboratoires, plusieurs ordres d'essai
ISO 4259-1:2026(fr) uniques peuvent être assignés au hasard à des groupes de laboratoires, avec pas plus de 4 laboratoires par groupe.
...