Chemical analysis of ferrous materials - Inductively coupled plasma optical emission spectrometric analysis of unalloyed and low alloyed steels - Determination of Si, Mn, P, Cu, Ni, Cr, Mo, Sn and Al (soluble) [Routine method]

This European Standard specifies an inductively coupled plasma optical emission spectrometry routine method for the analysis of unalloyed and low alloyed steels, whose iron content shall be at least 95 %. This standard differs from the similar standard EN 10351:2011 in that it is optimised for the determination of silicon. This method is applicable to the elements listed in Table 1 within the ranges shown. The sample preparation described may not completely dissolve samples having a combination of high chromium and substantial carbon. Incomplete dissolution may also affect the determination of manganese and molybdenum in these samples. For this reason, the scope of the method is limited to chromium contents ≤ 0,9 %, whereas the scope of EN 10351 covers a range of up to 1,6 % chromium.

Chemische Analyse von Eisenwerkstoffen - Analyse von unlegierten und niedrig legierten Stähle mittels optischer Emissionsspekrometrie mit indukiv gekoppeltem Plasma - Bestimung von Si, Mn, P, Cu, Ni, Cr, Mo, Sn und Al (lösung) [Routineverfahren]

Diese Europäische Norm legt ein Verfahren der optischen Emissionsspektrometrie mit induktiv gekoppeltem Plasma (ICP Emissionsspektrometrie) fest, das routinemäßig zur Analyse unlegierter und niedrig legierter Stähle anzuwenden ist, die einen Eisengehalt von mindestens 95 % haben. Diese Norm weicht von der Norm EN 10351:2011 ab; EN 10351:2011 ist für die Bestimmung von Si optimiert.
Das Verfahren ist auf die in Tabelle 1 aufgeführten Elemente innerhalb der angegebenen Bereiche anwendbar.
Die beschriebene Herstellung von Proben kann Proben, die sich aus einem hohen Chrom- und erheblichem Kohlenstoff¬gehalt zusammensetzen, nicht vollständig lösen. Diese unvollständige Lösung kann außerdem die Bestimmung von Mangan sowie Molybdän in diesen Proben beeinflussen. Aus diesem Grund ist der Anwendungsbereich dieses Verfahrens auf einen Chromanteil von  0,9 % beschränkt, wogegen der Anwendungsbereich der EN 10351 einen Bereich  1,6 % Cr abdeckt.
Tabelle 1 — Anwendungsbereiche
Element   Massenanteil
%
   min.   max.
Si   0,020   0,45
Mn   0,005   1,40
P   0,005   0,10
Cu   0,005   0,60
Ni   0,010   2,00
Cr   0,010   0,90
Mo   0,005   0,60
Sn   0,010   0,10
ANMERKUNG   Für Zinn siehe Absatz 11, Anmerkung 2.
Alle angegebenen Bereiche können erweitert oder (nach einer Validierung) so angepasst werden, dass auch andere Massenanteile zu bestimmen sind, sofern die zu analysierenden Proben einen Eisenanteil von mehr als 95 % haben.
Das Verfahren darf auch dann angewendet, wenn weitere Elemente enthalten sind. Diese Elemente und ihre Massenanteile sollten jedoch unter Berücksichtigung möglicher Störungen, der Empfindlichkeit, der Auflösung und der Linearitätskriterien für jedes ICP Emissionsspektrometer und für jede Wellenlänge sorgfältig überprüft werden.
Entsprechend der Empfindlichkeit des jeweiligen ICP Emissionsspektrometers kann es notwendig sein, geeignete Verdünnungen der Kalibrier  und der Untersuchungsprobenlösungen anzuwenden.
Obwohl das beschriebene Verfahren zur „Mehrelementbestimmung“ angewendet wird, müssen nicht unbedingt alle im Anwendungsbereich angegebenen Elemente gleichzeitig bestimmt werden. Die Mess¬bedingungen sind in jedem Prüflaboratorium in Abhängigkeit von den Leistungsparametern der verfügbaren Prüfgeräte zu optimieren.

Analyse chimique des matériaux sidérurgiques - Analyse des aciers non alliés et faiblement alliés par spectrométrie d'émission optique avec source à plasma induit - Détermination de Si, Mn, P, Cu, Ni, Cr, Mo, Sn et Al (soluble) [Méthode de routine]

La présente Norme européenne spécifie une méthode de routine, par spectrométrie d'émission optique avec source à plasma induit, pour l’analyse des aciers non alliés et faiblement alliés ayant une teneur en fer d’au moins 95 %. La présente norme diffère de la norme similaire EN 10351:2011 dans la mesure où elle est optimisée pour la détermination du silicium.
Cette méthode s’applique aux éléments listés dans le Tableau 1, dans les domaines de teneur indiqués.
La préparation d’échantillon décrite ne peut pas conduire à la mise en solution complète d’échantillons ayant à la fois une teneur élevée en chrome et une teneur substantielle en carbone. Cette mise en solution incomplète peut aussi affecter les déterminations du manganèse et du molybdène dans ces échantillons. Pour cette raison, le domaine d’application de la méthode est limité aux teneurs en chrome  0,9 %, alors que le domaine d’application de l’EN 10351 couvre un domaine allant jusqu’à 1,6 % de chrome.
Tableau 1 — Domaines d’application
Elément   Teneur (fraction en masse) %
   min.   max.
Si   0,020   0,45
Mn   0,005   1,40
P   0,005   0,10
Cu   0,005   0,60
Ni   0,010   2,00
Cr   0,010   0,90
Mo   0,005   0,60
Sn   0,010   0,10
NOTE   Pour l’étain, voir NOTE 2 à l’Article 11.
Dans tous les cas, les domaines spécifiés peuvent être étendus ou adaptés (après validation) pour la détermination d’autres teneurs, à condition que la teneur en fer des échantillons concernés soit supérieure à 95 %.
D'autres éléments peuvent être inclus. Il convient toutefois de contrôler, avec la plus grande attention, ces éléments et leurs teneurs, en tenant compte des interférences éventuelles, de la sensibilité, de la résolution et des critères de linéarité de chaque instrument et chaque longueur d'onde.
Aussi, en fonction de la sensibilité de chaque instrument, des dilutions appropriées des solutions d’étalonnage et des solutions d’échantillon peuvent être nécessaires.
En outre, même si la méthode décrite est « multiélémentaire », il n’est pas absolument nécessaire de déterminer simultanément tous les éléments de son domaine d’application. Les conditions de mesurage doivent être optimisées par chaque laboratoire, en fonction des performances de chaque appareil disponible.

Kemična analiza železovih zlitin - Analiza nelegiranih in malolegiranih jekel z optično emisijsko spektrometrijo z induktivno sklopljeno plazmo - Določevanje Si, Mn, P, Cu, Ni, Cr, Mo, Sn in topnega Al [rutinska metoda]

Ta evropski standard določa rutinsko metodo za analizo nelegiranih in malolegiranih jekel z optično emisijsko spektrometrijo z induktivno sklopljeno plazmo, pri čemer mora biti vsebnost železa v jeklu vsaj 95 %. Ta standard se od podobnega standarda EN 10351:2011 razlikuje v tem, da je optimiziran za določevanje silicija. Ta metoda se uporablja za elemente iz preglednice 1 v prikazanih razponih. Z opisano pripravo vzorca se morda vzorci s kombinacijo visoke vsebnosti kroma in znatne vsebnosti ogljika ne bodo raztopili v celoti. Nepopolno raztapljanje lahko vpliva tudi na določevanje mangana in molibdena v teh vzorcih. Zato je področje uporabe metode omejeno na vsebnost kroma ≤ 0,9 %, medtem ko področje uporabe standarda EN 10351 zajema razpon do vključno 1,6 % kroma.

General Information

Status
Published
Publication Date
12-Dec-2013
Technical Committee
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
05-Dec-2013
Due Date
09-Feb-2014
Completion Date
13-Dec-2013

Buy Standard

Standard
EN 10355:2014
English language
39 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Chemische Analyse von Eisenwerkstoffen - Analyse von unlegierten und niedrig legierten Stähle mittels optischer Emissionsspekrometrie mit indukiv gekoppeltem Plasma - Bestimung von Si, Mn, P, Cu, Ni, Cr, Mo, Sn und Al (lösung) [Routineverfahren]Analyse chimique des matériaux sidérurgiques - Analyse des aciers non alliés et faiblement alliés par spectrométrie d'émission optique avec source à plasma induit - Détermination de Si, Mn, P, Cu, Ni, Cr, Mo, Sn et Al (soluble) [Méthode de routine]Chemical analysis of ferrous materials - Inductively coupled plasma optical emission spectrometric analysis of unalloyed and low alloyed steels - Determination of Si, Mn, P, Cu, Ni, Cr, Mo, Sn and Al (soluble) [Routine method]77.080.20JeklaSteels77.040.30Kemijska analiza kovinChemical analysis of metalsICS:Ta slovenski standard je istoveten z:EN 10355:2013SIST EN 10355:2014en,fr,de01-januar-2014SIST EN 10355:2014SLOVENSKI
STANDARD



SIST EN 10355:2014



EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EN 10355
August 2013 ICS 77.040.30 English Version
Chemical analysis of ferrous materials - Inductively coupled plasma optical emission spectrometric analysis of unalloyed and low alloyed steels - Determination of Si, Mn, P, Cu, Ni, Cr, Mo and Sn, following dissolution with nitric and sulphuric acids [Routine method]
Analyse chimique des matériaux ferreux - Analyse des aciers non alliés et faiblement alliés par spectrométrie d'émission optique avec source à plasma induit - Détermination de Si, Mn, P, Cu, Ni, Cr, Mo et Sn, après mise en solution par les acides nitrique et sulfurique [Méthode de routine]
Chemische Analyse von Eisenwerkstoffen - Analyse von unlegierten und niedrig legierten Stählen mittels optischer Emissionsspektrometrie mit induktiv gekoppeltem Plasma -Bestimmung von Si, Mn, P, Cu, Ni, Cr, Mo und Sn nach Lösen in Salpeter- und Schwefelsäure [Routineverfahren] This European Standard was approved by CEN on 29 June 2013.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre:
Avenue Marnix 17,
B-1000 Brussels © 2013 CEN All rights of exploitation in any form and by any means reserved worldwide for CEN national Members. Ref. No. EN 10355:2013: ESIST EN 10355:2014



EN 10355:2013 (E) 2 Contents Page Foreword . 3 1 Scope . 4 2 Normative references . 4 3 Principle. 5 4 Reagents . 5 5 Apparatus . 7 6 Sampling . 7 7 Sample solution preparation . 7 8 Calibration process . 9 9 Determination . 11 10 Expression of results . 14 11 Precision . 14 12 Test report . 17 Annex A (informative)
Plasma optical emission spectrometer — Suggested performance criteria to be checked . 18 Annex B (normative)
Synoptic of the operations related to Clause 9 . 22 Annex C (informative)
Composition of the test samples used for the validation precision test . 23 Annex D (informative)
Detailed results obtained from the validation precision test . 24 Annex E (informative)
Graphical representation of the precision data . 32 Bibliography. 39
SIST EN 10355:2014



EN 10355:2013 (E) 3 Foreword This document (EN 10355:2013) has been prepared by Technical Committee ECISS/TC 102 “Methods of chemical analysis for iron and steel”, the secretariat of which is held by SIS. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by February 2014, and conflicting national standards shall be withdrawn at the latest by February 2014. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights. According to the CEN-CENELEC Internal Regulations, the national standards organisations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, Former Yugoslav Republic of Macedonia, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United Kingdom. SIST EN 10355:2014



EN 10355:2013 (E) 4 1 Scope This European Standard specifies an inductively coupled plasma optical emission spectrometry routine method for the analysis of unalloyed and low alloyed steels, whose iron content shall be at least 95 %. This standard differs from the similar standard EN 10351:2011 in that it is optimised for the determination of silicon. This method is applicable to the elements listed in Table 1 within the ranges shown. The sample preparation described may not completely dissolve samples having a combination of high chromium and substantial carbon. Incomplete dissolution may also affect the determination of manganese and molybdenum in these samples. For this reason, the scope of the method is limited to chromium contents ≤ 0,9 %, whereas the scope of EN 10351 covers a range of up to 1,6 % chromium.
Table 1 — Application ranges
Element Mass fraction % min. max. Si 0,020 0,45 Mn 0,005 1,40 P 0,005 0,10 Cu 0,005 0,60 Ni 0,010 2,00 Cr 0,010 0,90 Mo 0,005 0,60 Sn 0,010 0,10
NOTE For tin, see NOTE 2 under Clause 11. In all cases, the ranges specified can be extended or adapted (after validation) for the determination of other mass fractions, provided that the iron content in the samples under concern is above 95 %. Other elements may be included. However such elements and their mass fractions should be carefully checked, taking into account the possible interferences, the sensitivity, the resolution and the linearity criteria of each instrument and each wavelength. Depending also on the sensitivity of each instrument, suitable dilutions of the calibration and the test sample solutions may be necessary. Moreover, even if the method described is "multi elemental", it is not absolutely necessary to carry out the determination of all the elements of its scope simultaneously. The measurement conditions have to be optimised by each laboratory, depending on the performances of each apparatus available. 2 Normative references The following documents, in whole or in part, are normatively referenced in this document and are indispensable for its application. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. EN ISO 648, Laboratory glassware — Single-volume pipettes (ISO 648) EN ISO 1042, Laboratory glassware — One-mark volumetric flasks (ISO 1042) SIST EN 10355:2014



EN 10355:2013 (E) 5 3 Principle Dissolution of a test portion with a nitric-sulphuric acids mixture. Addition of an ammonium peroxodisulphate solution and boiling. After cooling, addition of a hydrogen peroxide solution. After suitable dilution and, if necessary, addition of an internal reference element, nebulisation of the solution into an inductively coupled plasma optical emission spectrometer and measurement of the intensity of the emitted light from each element (including, where relevant, the intensity of the internal reference element). 4 Reagents During the analysis, use only reagents of recognised analytical grade and only distilled water or water of equivalent purity. The same reagents should be used for the preparation of calibration solutions and of sample solutions. 4.1 Hydrochloric acid, HCl (ρ20 = 1,19 g/ml). 4.2 Hydrochloric acid, solution 1 + 1. Add 500 ml of hydrochloric acid (4.1) to 500 ml of water. 4.3 Nitric acid, HNO3 (ρ20 = 1,33 g/ml). 4.4 Nitric acid, solution 1 + 1. Add 500 ml of nitric acid (4.3) to 500 ml of water. 4.5 Sulphuric acid, H2SO4 (ρ20 = 1,84 g/ml). 4.6 Hydrogen peroxide, H2O2 30 % (mass fraction) solution. 4.7 Hydrogen peroxide, H2O2 3 % (mass fraction) solution. Dilute 50 ml of hydrogen peroxide (4.6) with water to 500 ml. 4.8 Nitric-sulphuric acids mixture. Add to approximately 500 ml of water 50 ml of nitric acid (4.3) and 50 ml of sulphuric acid (4.5) and mix. After cooling, dilute to 1 l with water and mix. 4.9 Ammonium peroxodisulphate solution, 20 g/l. Dissolve 10 g of ammonium peroxodisulphate in 500 ml water. 4.10 Chromium 1 g/l standard solution. Weigh (0,5 ± 0,001) g of chromium (99,99 % purity) and transfer into a 250 ml beaker. Add 40 ml of hydrochloric acid (4.1) and heat gently until chromium is completely dissolved. After cooling, transfer the solution quantitatively into a 500 ml one-mark volumetric flask. Dilute to the mark with water and mix well. 1 ml of this solution contains 1 mg of Cr. SIST EN 10355:2014



EN 10355:2013 (E) 6 4.11 Copper 1 g/l standard solution. Weigh (0,5 ± 0,001) g of copper (99,99 % purity) and transfer into a 250 ml beaker. Dissolve it in 10 ml of nitric acid solution (4.4). Heat gently until the metal is dissolved and then boil until nitrous fumes have been expelled. After cooling, transfer the solution quantitatively into a 500 ml one-mark volumetric flask. Dilute to the mark with water and mix well. 1 ml of this solution contains 1 mg of Cu. 4.12 Manganese 1 g/l standard solution. The manganese used to prepare the solution is released from superficial oxide possibly present by introducing a few grams of metal in a 250 ml beaker containing (150 to 160) ml of water and (15 to 20) ml of sulphuric acid (4.5). Shake and after a few seconds, allow the solution to settle and add water. Repeat the water cleaning several times. Remove the metallic manganese and rinse with acetone. Dry the metal in an oven at 100 °C for 2 min or with a hair dryer. Cool in a desiccator. Weigh (0,5 ± 0,001) g of this purified manganese and transfer into a 250 ml beaker. Dissolve it in 5 ml of hydrochloric acid (4.1) and 10 ml of nitric acid solution (4.4). Heat gently until the metal is dissolved. After cooling, transfer the solution quantitatively into a 500 ml one-mark volumetric flask. Dilute to the mark with water and mix well. 1 ml of this solution contains 1 mg of Mn. 4.13 Molybdenum 1 g/l standard solution. Weigh (0,5 ± 0,001) g of molybdenum (99,99 % purity) and transfer into a 250 ml beaker. Dissolve it in 10 ml of hydrochloric acid (4.1) and 10 ml of nitric acid (4.3). Heat gently until the metal is dissolved. After cooling, transfer the solution quantitatively into a 500 ml one-mark volumetric flask. Dilute to the mark with water and mix well. 1 ml of this solution contains 1 mg of Mo. 4.14 Nickel 1 g/l standard solution. Weigh (0,5 ± 0,001) g of nickel (99,99 % purity) and transfer into a 250 ml beaker. Dissolve it in 10 ml of nitric acid solution (4.4). Heat gently until the metal is dissolved, and then boil until nitrous fumes have been expelled. After cooling, transfer the solution quantitatively into a 500 ml one-mark volumetric flask. Dilute to the mark with water and mix well. 1 ml of this solution contains 1 mg of Ni. 4.15 Phosphorus 1 g/l standard solution. Weigh (2,197 ± 0,001) g of dried potassium dihydrogen phosphate, transfer into a 250 ml beaker and dissolve it with water. Transfer the solution quantitatively into a 500 ml one-mark volumetric flask. Dilute to the mark with water and mix well. 1 ml of this solution contains 1 mg of P. 4.16 Silicon 1 g/l standard solution. Using a suitable plastic beaker, dissolve (3,171 ± 0,001) g of dried ammonium hexafluorosilicate [(NH4)2SiF6] in hot water (~ 80 °C). Allow to cool and transfer the solution quantitatively into a polyethylene 500 ml one-mark volumetric flask, dilute to the mark with water and mix well. SIST EN 10355:2014



EN 10355:2013 (E) 7 1 ml of this solution contains 1 mg of Si. 4.17 Tin 1 g/l standard solution [freshly prepared]. Weigh (0,5 ± 0,001) g of tin (99,99 % purity) and transfer into a 250 ml beaker. Dissolve it in 50 ml of hydrochloric acid (4.1). Heat gently until the metal is dissolved. After cooling, transfer the solution quantitatively into a 500 ml one-mark volumetric flask. Dilute to the mark with water and mix well. 1 ml of this solution contains 1 mg of Sn. NOTE Standard solutions whose preparations are described in 4.10 to 4.17 can be replaced by commercially available standard solutions, provided that they are supplied by a recognised producer and their traceability fully documented. 4.18 Internal reference element solution, 1 g/l. Choose a suitable element to be added as internal reference and prepare a 1 g/l solution. NOTE Elements such as Sc and Y are often used for this purpose. 4.19 Pure iron. The iron selected shall have a content of less than one tenth of the lower limit of each element as given in the scope. 5 Apparatus All volumetric glassware shall be class A and calibrated in accordance with EN ISO 648 or EN ISO 1042, as appropriate. 5.1 Microwave system. 5.2 Gravimetric or volumetric dispensers. 5.3 Optical emission spectrometer, equipped with inductively coupled plasma. This shall be equipped with a nebulisation system. The instrument used will be satisfactory if, after optimising in accordance with the manufacturer’s instructions, it meets the performance criteria given in Annex A. 6 Sampling Sampling shall be carried out in accordance with EN ISO 14284 or with an appropriate national standard for steel. 7 Sample solution preparation 7.1 General The method is optimised for use with automated systems, open microwave digestion units (5.1) and gravimetric dispensers (5.2). NOTE This method was also tested with closed microwave units, after a suitable optimisation of the relevant operating conditions: temperatures and times.
SIST EN 10355:2014



EN 10355:2013 (E) 8 The method can also be used in a classical way with hot plates and manual additions. 7.2 Test portion Weigh, to the nearest 1 mg, 250 mg of the test sample. 7.3 Preparation of the test solution 7.3.1 Automated system, OPEN units Transfer the test portion (7.2) into a PTFE tube. Add 10 g of the nitric-sulphuric acid mixture (4.8) and cover with a suitable cover (PTFE or some other microwave transparent material, perfectly clean and free from contamination. Screw caps shall be avoided with open microwave units). Use the microwave to heat the sample at about 120 °C for 7 min and then allow to cool for about 10 min. Add 5 g of the ammonium peroxodisulphate solution (4.9), and cover. Heat at about 120 °C for 3 min by means of the microwave system, and then allow to cool for about 10 min. Add 5 g of the hydrogen peroxide solution (4.7) and mix well. Cool for about 10 min. The cooling time may be increased to ensure that gas evolution has ceased before analysing the sample. The addition of 2 g HCl (4.1) may help to prevent the hydrolysis of silicon and tin. If an internal reference element is used, add a suitable quantity of the internal reference solution (4.18), dilute the sample solution to 50 g, weighted to the nearest mg and mix well. If necessary, filter the solution through a dry close-texture filter paper. Reject the first portions of the filtrate. It is recommended to carry out the measurements on freshly prepared solutions, in order to avoid silicon hydrolysis. 7.3.2 Manual procedure Transfer the test portion (7.2) into a 100 ml quartz or glassware conical flask (or in a 100 ml PTFE beaker). NOTE It has been shown that perfectly clean quartz or ordinary glass conical flasks are suitable even for the determination of silicon. Other plastic materials may be used, provided that they are heat resistant and free from contamination. Add 10 ml of the nitric-sulphuric acid mixture (4.8). If beakers are used, cover with a PTFE1) watch glass and heat until gently boiling (the hotplate temperature shall be adjusted in order to allow a gentle boiling of the sample solution). Maintain the boiling for about 7 min to 10 min and then allow to cool for about 10 min. Add 5 ml of the ammonium peroxodisulphate solution (4.9), cover and heat until gently boiling. Maintain the boiling for about 3 min to 5 min and then allow to cool for about 10 min. Add 5 ml of the hydrogen peroxide solution (4.7) and mix well. Cool for about 10 min. The cooling time may be increased to ensure that gas evolution has ceased before analysing the sample.
1) Covers made from other materials can be suitable, provided that these materials are free from contamination. SIST EN 10355:2014



EN 10355:2013 (E) 9 Transfer the sample solution into a 50 ml one-mark volumetric flask and then, if an internal reference element is used, add, with a calibrated pipette, a suitable volume of the internal reference element solution (4.18). The addition of 2 ml HCl (4.1) may help to prevent the hydrolysis of silicon and tin. Dilute to the mark with water and mix well. If necessary, filter the solution through a dry close-texture filter paper. Reject the first portions of the filtrate. It is recommended to carry out the measurements on freshly prepared solutions, in order to avoid silicon hydrolysis. 8 Calibration process 8.1 Steps The calibration process is subdivided into calibration and drift compensation by re-calibration. 8.2 Preparation of the multi-elemental calibration solutions Weight 2,500 g of pure iron (4.19) into a 500 ml beaker and add 100 ml of the nitric-sulphuric acids mixture (4.8). Heat at about 120 °C, for 7 min. Add 50 ml of the ammonium peroxodisulphate solution (4.9), cover and heat at about 120 °C for 3 min. Boil gently and allow to cool. Add 50 ml of the hydrogen peroxide solution (4.7) and mix well. Transfer into a 250 ml one-mark volumetric flask, dilute to the mark with water and mix well. Into each of a series of four 100 ml one-mark volumetric flasks transfer 50,0 ml of this iron solution and, with calibrated pipettes, add the volumes of the mono-elemental standard solutions (4.10 to 4.17) or of commercial available mono-elemental standard solutions shown in Table 2. If an internal reference element is used add, with a calibrated pipette a suitable volume of the internal reference element solution (4.18); this volume shall be double of that added to the sample solution (7.3). The addition of 4 ml HCl (4.1) may help to prevent the hydrolysis of silicon and tin. Dilute to the mark and mix well. It is recommended to carry out the inductively coupled plasma optical emission spectrometric measurements on freshly prepared solutions, in order to avoid silicon hydrolysis. In a fifth 100 ml one-mark volumetric flasks transfer 50,0 ml of the above 10 g/l iron solution. Dilute to the mark and mix well. Use this solution to test whether or not the lowest concentration of each element is high enough to be considered as "different from the zero point". Strong care should be taken in order to ensure that the quantity of acids in the calibration solutions (particularly when commercially available standard solutions are used), in the sample solutions and in the Certified Reference Materials solutions prepared to check the trueness of the calibration remain the same. The calibration curves obtained may be stored within a computer, connected to the inductively coupled plasma optical emission spectrometer. SIST EN 10355:2014



EN 10355:2013 (E) 10 The trueness of these calibration functions should be verified by measuring at least two suitable Certified Reference Materials or Internal Reference Materials sample solutions (just above the low point and just below the high point of each calibration range). Table 2 — Composition of the multi-elemental calibration solutions
Element Label F1 F2 F3 F4 Si (4.16) 0,1 mg [0,02 %] a 0,25 mg [0,05 %] b 2,25 mg [0,45 %] c 0,5 mg [0,1 %] d Mn (4.12) 7 mg [1,4 %] 7 ml 1 mg [0,2 %] 1 ml 0,025 mg [0,005 %] e 0,1 mg [0,02 %] f P (4.15) 0,025 mg [0,005 %] g 0,05 mg [0,01 %] h 0,25 mg [0,05 %] i 0,5 mg [0,1 %] j Cu (4.11) 0,25 mg [0,05 %] k 3,0 mg [0,6 %] 3 ml 0,025 mg [0,005 %] l 0,5 mg [0,1 %] m Ni (4.14) 10 mg [2,0 %] 10 ml 2,5 mg [0,5 %] 2,5 ml 0,05 mg [0,01 %] n 0,25 mg [0,05 %] o Cr (4.10) 0,05 mg [0,01 %] p 0,25 mg [0,05 %] q 4,5 mg [0,9 %] 4,5 ml 2,5 mg [0,5 %] 2,5 ml Mo (4.13) 0,05 mg [0,01 %] r 0,25 mg [0,05 %] s 3 mg [0,6 %] 3 ml 0,5 mg [0,1 %] t Sn (4.17) 0,5 mg [0,1 %] u 0,05 mg [0,01 %] v 0,25 mg [0,05 %] w 0,1 mg [0,02 %] x a
1 ml of a 0,1 g/l silicon standard solution m 5,0 ml of a 0,1 g/l copper standard solution b
2,5 ml of a 0,1 g/l silicon standard solution n 5,0 ml of a 0,01 g/l nickel standard solution c
22,5 ml of a 0,1 g/l silicon standard solution o 2,5 ml of a 0,1 g/l nickel standard solution d
5,0 ml of a 0,1 g/l silicon standard solution p 2,5 ml of a 0,01 g/l chromium standard solution e
2,5 ml of a 0,01 g/l manganese standard solution q 2,5 ml of a 0,1 g/l chromium standard solution f
10 ml of a 0,01 g/l manganese standard solution r
5,0 ml of a 0,01 g/l molybdenum standard solution g
2,5 ml of a 0,01 g/l phosphorus standard solution s
2,5 ml of a 0,1 g/l molybdenum standard solution h
5,0 ml of a 0,01 g/l phosphorus standard solution t
5,0 ml of a 0,1 g/l molybdenum standard solution i
2,5 ml of a 0,1 g/l phosphorus standard solution u
5,0 ml of a 0,1 g/l tin standard solution j
5,0 ml of a 0,1 g/l phosphorus standard solution v
5,0 ml of a 0,01 g/l tin standard solution k
2,5 ml of a 0,1 g/l copper standard solution w 2,5 ml of a 0,1 g/l tin standard solution l
2,5 ml of a 0,01 g/l copper standard solution x
10,0 ml of a 0,01 g/l tin standard solution
8.3 Re-calibration Drifts in the spectrometer readings shall be corrected by using the re-calibration procedure as described in the manufacturer's instruction manual. For that purpose the synthetic multi-elemental solutions (8.2), used for the calibration or suitable Certified Reference Materials or Internal Reference Materials solutions may be used. It is strongly recommended to carry out a re-calibration of the instrument before the analysis of each set of unknown samples. SIST EN 10355:2014



EN 10355:2013 (E) 11 9 Determination 9.1 Synoptic The synoptic of the operations described in this paragraph is given in Annex B. 9.2 Adjustment of the spectrometer Start the inductively coupled plasma optical emission spectrometer and let it stabilise in accordance with the manufacturer’s instructions before taking any measurements. At the wavelengths of the analytical lines specified in Table 3, adjust all relevant instrumental parameters, as well as the pre-spraying and the integrating times, according to the instrument manufacturer’s instructions while aspirating (for each element) the highest concentration calibration solution. Table 3 — Wavelengths for determinations related with unalloyed and low alloyed steels Element Wavelength (nm) Si 185,0 or 251,6 or 288,2 Mn 257,6 or 293,1 or 403,08 P 178,2 or 178,3 Cu 217,9 or 324,7 or 327,4 Ni 222,5 or 230,3 or 231,6 Cr 205,6 or 206,6 or 267,7 or 285,0 Mo 202,0 or 204,05 or 281,6 Sn 189,9 or 190,0 Sc 361,4 Y 371,0
Depending on the instrument configuration these parameters may include the outer, intermediate or central gas flow-rates, the torch position, the entrance slits, the exit slits and the photomultiplier tubes voltage. Other wavelengths may be used, provided that interferences, sensitivity, resolution and linearity criteria have been carefully investigated. Prepare the software for measurements of the intensity, and for the calculation of the mean value and relative standard deviation corresponding to each analytical line. Each time the internal reference element is used, prepare the software to calculate the ratio between the intensity of each analyte and the intensity of the internal reference element. 9.3 Spectrometric measurements of the calibration solutions Carry out the spectrometric measurements of the calibration solutions (8.2). A simultaneous spectrometer shall be available for measurements using the internal reference element (ratio mode). For each calibration solution, three to five integrations are necessary in order to calculate the mean intensities or the mean ratioed intensities. SIST EN 10355:2014



EN 10355:2013 (E) 12 Establish the calibration curves using the measured mean intensities or ratioed mean intensities and the corresponding analyte amounts, expressed in percentage (mass fraction). Use software suitable for regression calculations and calculate the corresponding correlation coefficients: these coefficients shall be, at least, 0,999. If this criterion is not conforming, check discrepancies within the multi-elemental standard solutions preparation, or adapt the analytical range(s) either by performing suitable dilutions of the calibration standard solutions or by adapting the analytical range itself. If some calibration curves are second degree mathematical functions within the range of content, the number of multi-elemental calibration solution should be increased. The trueness of these calibration functions shall be verified by measuring at least two suitable Certified Reference Materials (or Internal Reference Materials) sample solutions (just above the low point and just below the high point of each calibration range). The differences between the measured values and the reference values shall not exceed either the interval "Certified value ± 2 times the corresponding standard deviation" or the limits given in Table 4, or when appropriate, the limits fixed by suitable control charts (SPC). If these criteria are not met, repeat the test with a re-prepared set of mono-elemental calibration solutions or adapt the analytical range(s). Table 4 — Trueness control limits Content range % Maximum difference (±) % 0,001 to 0,002 0,000 5 0,002 to 0,005 0,000 9 0,005 to 0,01 0,001 4 0,01 to 0,02 0,002 2 0,02 to 0,05 0,004 0 0,05 to 0,1 0,006 0 0,1 to 0,2 0,009 0 0,2 to 0,5 0,016 0,5 to 1,0 0,025 1,0 to 2,0 0,040
9.4 Analysis 9.4.1 Main steps Carry out the analysis by following the steps described here below. 9.4.2 Re-calibration Start each analytical series by a re-calibration of the instrument. Check the performance of this operation for each element to be determined by measuring at least one Certified Reference Material (CRM 1) or one Internal Reference Material (IRM 1) sample solution, prepared as described in Clause 7. SIST EN 10355:2014



EN 10355:2013 (E) 13 Check the results: the re-calibration will be acceptable if all the results for the Certified Reference Material (or Internal Reference Material) concerned are situated in the interval "Certified Value ± 2σ", where σ is the standard deviation of the certified value. If abnormal deviations appear, restart the re-calibration procedure. This sample solution (CRM 1 or IRM 1) can also be used for all the drift controls to be carried out during the analytical series measurements. 9.4.3 Assessment of trueness
Analyse two other Certifi
...