oSIST prEN 196-18:2026
(Main)Methods of testing cement - Part 18: Determination of whiteness or colour of cement and clinker
- Abstract
This document describes a method for the determination of whiteness and the colour of common cements and clinkers.
This method can be used to determine the whiteness of white cements defined in the EN 197-1.
This document does not apply to colour measurements on cement-based formulated products or cement-based products.
The method is based on the measurement of the CIE XYZ trichromatic coordinates and describes the mathematical procedure for converting these coordinates into the CIE 1976 (L*a*b*) colour space.
NOTE Although the possible values of the L* parameter range from 0 to 100, this document applies to the range between 70 and 100 (photopic vision).
- Status
- Not Published
- Public Enquiry End Date
- 22-Oct-2026
- Technical Committee
- CAA - Mineral binders and masonry
- Current Stage
- 4020 - Public enquire (PE) (Adopted Project)
- Start Date
- 06-Aug-2026
- Due Date
- 24-Dec-2026
oSIST prEN 196-18:2026 is a CEN draft European Standard in the Methods of testing cement series. It specifies a reference method for determining the whiteness and colour of common cements and clinkers by measuring Commission Internationale de l’Éclairage (CIE) XYZ trichromatic coordinates and converting them to CIE 1976 (Lab*) colour space. It is intended for laboratories, producers, and quality-control users working with cement and clinker, including white cements defined in EN 197-1.
What does oSIST prEN 196-18:2026 specify?
oSIST prEN 196-18:2026 specifies how to measure whiteness and colour for common cements and clinkers. It also states that the method can be used for white cements defined in EN 197-1, but not for cement-based formulated products or cement-based products.
The document is organised into clauses on scope, normative references, terms and definitions, general testing requirements, apparatus, sample preparation, procedure, calculation, reporting, and repeatability/reproducibility. Annex A is informative and gives the conversion from XYZ to Lab*.
What are the key requirements of oSIST prEN 196-18:2026?
Test frequency and precision control
oSIST prEN 196-18:2026 requires duplicate testing when colour is not part of a statistically controlled series, a single test as the minimum for a controlled series, and two tests in case of dispute. That matters because the document ties the number of measurements to how the results will be used and challenged.
The standard also defines repeatability and reproducibility as standard deviations. In practice, this tells the laboratory how to judge whether two results are close enough to accept.
Apparatus and measurement setup
oSIST prEN 196-18:2026 requires a laboratory balance, a reflectance spectrophotometer, a caliper, a mill, an optical container, and a metal ring for pellet preparation. The spectrophotometer must cover 400 nm to 700 nm and be capable of expressing results in Lab* under specified illuminant and observer conditions.
This matters because colour results depend on the measuring geometry and the instrument setup. The document allows equivalent colour-measuring instruments only if their accuracy can be demonstrated.
Sample preparation
oSIST prEN 196-18:2026 gives separate preparation routes for cement and clinker. Cement is reduced, mixed, and protected from moisture; clinker is ground to a specified Blaine fineness using EN 196-6 before testing.
The standard offers two measurement sample forms: a pressed pellet or a tamped layer in an optical container. In both cases, the measurement surface must be free of cracks or air voids, and the sample should be measured quickly after preparation.
Measurement and calibration
oSIST prEN 196-18:2026 requires the spectrophotometer and the sample to be at the same laboratory temperature and protected from ambient light. The instrument must be switched on and calibrated before use, following the operating instructions.
The colour reading records XYZ values from the spectrophotometer. The method then converts those values to Lab*, which is the output used for reporting and comparison.
Calculation and reporting
oSIST prEN 196-18:2026 requires the mean of three measurements to be reported, with X, Y, Z and L* rounded to one decimal place and a* and b* to two decimal places. If results differ by more than twice the repeatability standard deviation, the test must be repeated.
The test report must identify the sample, cite the document, state how the sample was prepared, give the mean XYZ and Lab* values, note any deviations, and include the test date. That gives users a traceable record of how the colour result was obtained.
What terms does oSIST prEN 196-18:2026 define?
- CIE standard illuminant D65 - a daylight-like reference illuminant with a correlated colour temperature of about 6500 K.
- CIE 1931 colorimetric observer - the 2° observer used for colour perception in a small field of view.
- CIE 1976 (Lab*) colour space - the colour space used to express lightness and chromatic coordinates from XYZ values.
- Whiteness CIE 1976 - the CIE colourimetric attribute that expresses how close a material appears to ideal white under specified conditions.
Who uses oSIST prEN 196-18:2026?
oSIST prEN 196-18:2026 is used by cement producers, clinker producers, and testing laboratories that need a repeatable way to compare colour or whiteness. Quality managers use it to set and verify product appearance, and buyers use it when whiteness is part of acceptance or specification.
It is also relevant to anyone checking white cements, since the scope explicitly includes the white cements defined in EN 197-1.
Which standards are used with oSIST prEN 196-18:2026?
| Standard | What it contributes |
|---|---|
| EN 196-6 | The fineness method used when clinker is ground to the required Blaine surface area. |
| EN 196-7 | Sampling and sample preparation for common cement and its constituents. |
| EN ISO/CIE 11664-4 | The definition of the CIE 1976 (Lab*) colour space used for reporting. |
| EN 197-1 | The definition of white cements mentioned in the scope. |
What does the oSIST prEN 196-18:2026 document contain?
oSIST prEN 196-18:2026 contains apparatus requirements, sample-preparation steps, a measurement procedure, calculation rules, reporting rules, and a precision statement. It also includes the repeatability and reproducibility table for L*, a*, b*, X, Y, and Z.
Annex A gives the formulas and reference white constants used to convert measured XYZ values into Lab* coordinates. The body of the document also describes instrument calibration, measurement conditions, and the two accepted sample-preparation formats.
Get Certified
Connect with accredited certification bodies for this standard

DIBt (Deutsches Institut für Bautechnik)
German Institute for Building Technology.
DIN CERTCO
DIN Group product certification.

Aboma Certification B.V.
Specialized in construction, metal, and transport sectors.
Sponsored listings
Frequently Asked Questions
oSIST prEN 196-18:2026 is a draft published by the Slovenian Institute for Standardization (SIST). Its full title is "Methods of testing cement - Part 18: Determination of whiteness or colour of cement and clinker". This standard covers: This document describes a method for the determination of whiteness and the colour of common cements and clinkers. This method can be used to determine the whiteness of white cements defined in the EN 197-1. This document does not apply to colour measurements on cement-based formulated products or cement-based products. The method is based on the measurement of the CIE XYZ trichromatic coordinates and describes the mathematical procedure for converting these coordinates into the CIE 1976 (L*a*b*) colour space. NOTE Although the possible values of the L* parameter range from 0 to 100, this document applies to the range between 70 and 100 (photopic vision).
This document describes a method for the determination of whiteness and the colour of common cements and clinkers. This method can be used to determine the whiteness of white cements defined in the EN 197-1. This document does not apply to colour measurements on cement-based formulated products or cement-based products. The method is based on the measurement of the CIE XYZ trichromatic coordinates and describes the mathematical procedure for converting these coordinates into the CIE 1976 (L*a*b*) colour space. NOTE Although the possible values of the L* parameter range from 0 to 100, this document applies to the range between 70 and 100 (photopic vision).
oSIST prEN 196-18:2026 is classified under the following ICS (International Classification for Standards) categories: 91.100.10 - Cement. Gypsum. Lime. Mortar. The ICS classification helps identify the subject area and facilitates finding related standards.
oSIST prEN 196-18:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 CIE standard illuminant D65
- 3.2 CIE 1931 colorimetric observer
- 3.3 CIE 1976 (L*a*b*) colour space
- 3.4 whiteness CIE 1976
- 4 General requirements for testing
- 4.1 Number of tests
- 4.2 Repeatability and reproducibility
- 5 Apparatus
- 5.1 Balance: laboratory balance, capable of weighing to an accuracy of ± 0,01 g
- 5.2 Reflectance spectrophotometer
- 5.4 Mill, made of corundum or zirconium oxide, with a minimum purity of 99 %
- 6 Preparation of a test sample of cement and clinker
- 6.1 General
- 6.2 Preparation of test sample of cement
- 6.3 Preparation of test sample of clinker
- 7 Procedure
- 7.1 General
- 7.2 Preparation of the test sample as a pressed pellet
- 7.3 Preparation of the test sample for tamping (alternative method)
- 7.4 Preparation of the measuring reflectance spectrophotometer
- 8 Calculation and expression of results
- 8.1 Calculation
- 9 Reporting of results
- 9.1 Expression of results
- 9.2 Test report
- 10 Repeatability and reproducibility
- Annex A
- Bibliography
Introduction
This document specifies the reference method for the determination of cement and clinker colour.
oSIST prEN 196-18:2026 prEN 196-18:2026 (E)
1 Scope
This document describes a method for the determination of whiteness and the colour of common cements and clinkers.
This method can be used to determine the whiteness of white cements defined in the EN 197-1. This document does not apply to colour measurements on cement-based formulated products or cement-based products.
The method is based on the measurement of the CIE XYZ trichromatic coordinates and describes the mathematical procedure for converting these coordinates into the CIE 1976 (L*a*b*) colour space.
NOTE Although the possible values of the L* parameter range from 0 to 100, this document applies to the range between 70 and 100 (photopic vision).
2 Normative references
The following documents are referred to in the text in such a way that some or all their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
EN 196-6, Methods of testing cement — Part 6: Determination of fineness EN 196-7, Methods of testing cement — Part 7: Methods for sampling and sample preparation of common cement and its constituents EN ISO/CIE 11664-4, Colorimetry – Part 4: CIE 1976 L*a*b* colour space
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply. ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https://www.iso.org/obp/
- IEC Electropedia: available at https://www.electropedia.org/
3.1 CIE standard illuminant D65
reference illuminant having a spectral power distribution within the visible range that corresponds to average daylight, with a correlated colour temperature of approximately 6500 K
3.2 CIE 1931 colorimetric observer
colour perception for stimuli presented as an aperture on a dark background, corresponding to a 2° field of view (foveal vision)
oSIST prEN 196-18:2026 prEN 196-18:2026 (E)
3.3 CIE 1976 (L*a*b*) colour space
opponent-colour space with L* representing lightness, and a* and b* representing the chromatic opponent dimensions, derived from the non-linear compression of the CIE XYZ tristimulus values
Figure 1 — CIE L*a*b* colour space where:
L* = lightness coordinate, with values ranging from 0 (black) to 100 (white). a* = red/green chromatic coordinate, where positive a* indicates redness and negative a* indicates greenness. (see EN ISO/CIE 11664-4) b* = yellow/blue chromatic coordinate, where positive b* indicates yellowness and negative b* indicates blueness. (see EN ISO/CIE 11664-4)
3.4 whiteness CIE 1976
colorimetric attribute (L*a*b*), as defined by the CIE, expressing the perceived closeness of a material to an ideal white under specified illumination and observation conditions
4 General requirements for testing
4.1 Number of tests
Determination of colour, where determination is not part of a series subject to statistical control, shall be carried out in duplicate.
Where the determination is one of a series subject to statistical control, a single test shall be the minimum required.
In the case of a dispute, the number of tests shall be two.
oSIST prEN 196-18:2026 prEN 196-18:2026 (E)
4.2 Repeatability and reproducibility
Repeatability - precision under repeatability conditions where independent test results are obtained with the same method on identical test items (material) in the same laboratory by the same operator using the same equipment within short intervals of time. Reproducibility - precision under reproducibility conditions where test results are obtained with the same method on identical test items (material) in different laboratories with different operators using different equipment.
Repeatability and reproducibility in this European Standard are expressed as repeatability standard deviation and reproducibility standard deviation.
5 Apparatus
5.1 Balance: laboratory balance, capable of weighing to an accuracy of ± 0,01 g
5.2 Reflectance spectrophotometer
The reference reflectance spectrophotometer shall have the following characteristics:
- Minimum spectral measurement range: 400 nm to 700 nm.
- Bandwidth less than or equal to 10 nm.
- Minimum reading interval: 10 nm.
- CIE standard illuminant: D65 or C.
- CIE colorimetric observer: 10°; 2° when the CIE standard illuminant is C.
- Colour space: L*a*b*.
- The instrumental illumination geometry shall be d/8° with included specular component, or circumferential 0°/45° (45°/0°) as specified by the CIE.
NOTE Other types of instruments (e.g. colourimeters) can be used as an alternative, provided that their equivalence in terms of accuracy can be demonstrated.
5.3 Caliper, capable of measuring with an accuracy of ± 0.01 mm.
5.4 Mill, made of corundum or zirconium oxide, with a minimum purity of 99 %
for corundum or 99 % for zirconium oxide NOTE Other grinding media that guarantee the absence of contamination are also acceptable (es. Agate).
oSIST prEN 196-18:2026 prEN 196-18:2026 (E) 5.5 Optical container (e.g. glass or other suitable material). 5.6 Metal ring, for pellet preparation.
6 Preparation of a test sample of cement and clinker
6.1 General
Treat the laboratory sample, taken in accordance with EN 196-7, as follows to obtain a homogeneous test sample.
...