SIST EN ISO 10993-3:2026
(Main)Biological evaluation of medical devices - Part 3: Evaluation of genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity (ISO 10993-3:2026)
- Abstract
This document specifies strategies for risk estimation and evaluation of biological harms with respect to:
genotoxicity;
carcinogenicity;
reproductive toxicity; and
developmental toxicity.
This document is applicable when the need to evaluate a medical device for potential genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity has been established.
This document is not applicable to active pharmaceutical ingredients of device-drug combination products or biological components of device-biologic combination products which are covered by regulations.
NOTE Guidance on selecting relevant biological effects for medical devices is covered in ISO 10993-1.
- Status
- Published
- Public Enquiry End Date
- 26-Jun-2025
- Publication Date
- 23-Sep-2026
- Technical Committee
- VAZ - Healthcare
- Current Stage
- 6060 - National Implementation/Publication (Adopted Project)
- Start Date
- 17-Sep-2026
- Due Date
- 22-Nov-2026
- Completion Date
- 24-Sep-2026
SIST EN ISO 10993-3:2026 is a European Standard published by CEN that is identical to ISO 10993-3:2026. It specifies strategies for risk estimation and evaluation of genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity for medical devices, once that evaluation has been established as needed. It is aimed at medical device materials and constituents, not at active pharmaceutical ingredients of device-drug combination products or biological components of device-biologic combination products covered by regulations.
What does SIST EN ISO 10993-3:2026 specify?
SIST EN ISO 10993-3:2026 specifies how to assess four biological harms for medical devices: genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity. The document is organized around a risk-based decision process, supported by test methods, sample preparation guidance and reporting requirements.
The main body is arranged into clauses 1 to 8:
- Scope
- Normative references
- Terms and definitions
- Assessment strategies
- Genotoxicity assessment
- Carcinogenicity assessments
- Evaluation of reproductive toxicity and developmental toxicity
- Test report
The annexes give the detailed technical support. Some are informative guidance, while Annex B is normative for genotoxicity testing.
| Annex | What it covers |
|---|---|
| Annex A | Test sample preparation method for genotoxicity testing |
| Annex B | Tests for genotoxicity |
| Annex C | Carcinogenicity tests in animals |
| Annex D | Genotoxicity assessment of nanomaterials |
| Annex E | Recipe for S9 mix |
| Annex F | Endocrine disruptors |
| Annex G | Evaluation of reproductive toxicity and developmental toxicity |
| Annex ZA | Relationship between this European standard and Regulation (EU) 2017/745 |
What are the key requirements of SIST EN ISO 10993-3:2026?
SIST EN ISO 10993-3:2026 is built around justification, chemical characterization and case-by-case evaluation. In practice, the standard expects the user to explain why a hazard is relevant, how exposure is estimated, and why a test is or is not needed.
Clause 4 - choose the evaluation strategy case by case
Clause 4 requires the evaluation to consider patient population, expected contact type, exposure duration and the worst-case quantity of material. It also requires review of chemical constituents, including manufacturing residues, degradation products and metabolites, together with existing information and prior use of equivalent materials.
This matters because the standard is not a simple test checklist. A manufacturer must show why a device does, or does not, need further biological evaluation, and must justify any decision to waive in vitro or in vivo testing.
Clause 4 also allows toxicological risk assessment to end the evaluation if chemical characterization shows exposure within acceptable levels under ISO 10993-17. That means analytical data can replace unnecessary biological testing when the risk is already understood.
Clause 5 - address genotoxicity through testing or risk assessment
Clause 5 says genotoxic risk should be assessed either by toxicological risk assessment based on chemical characterization or by the standard in vitro genotoxicity battery. If a device releases nanomaterials, additional evaluation is required, and Annex D gives the special approach.
The practical point is that genotoxicity is not limited to one assay. Annex B adapts the bacterial reverse mutation test and in vitro mammalian cell assays for medical devices, while also covering in vivo follow-up and the logic for selection of test samples, controls and dose ranges.
Clause 5 also says the extraction vehicle must be compatible with the test system, and extraction should follow ISO 10993-12 unless otherwise justified. That matters because a poorly chosen extract can distort the result and understate the hazard.
Clause 6 - assess both genotoxic and non-genotoxic carcinogenicity
Clause 6 requires carcinogenicity to be considered for both genotoxic and non-genotoxic mechanisms. The standard prefers alternative approaches, especially chemical characterization and toxicological risk assessment, over default in vivo testing.
This is important because animal carcinogenicity studies are difficult to run for medical devices. The standard notes that implantation or extract-based studies often need large numbers of devices or concentrated extracts, so they are generally not justified when other evidence can answer the question.
Clause 7 - determine when reproductive and developmental toxicity evaluation is relevant
Clause 7 identifies when reproductive toxicity and developmental toxicity should be considered. The triggers include prolonged or long-term contact with reproductive organs or gametes, contact with embryos or foetuses, early childhood use with known or presumed developmental toxicants, absorbable or in situ-curing materials near reproductive organs, and new materials not previously used for a similar intended use.
Clause 7 also separates the two endpoints clearly. That matters because the exposure scenario for reproductive toxicity is not the same as for developmental toxicity, and the decision on what to evaluate depends on the device use pattern and the exposed population.
Clause 7 and Annex G - use risk assessment first, then testing if needed
Clause 7 says reproductive and developmental risk should first be addressed by toxicological risk assessment based on chemical characterization. Exposure estimates may be strengthened with leachables studies, simulated-use studies or release kinetic studies.
If risk remains unacceptable or the data are too limited, Annex G points to biological test methods and to common OECD study designs, including prenatal developmental toxicity and one-generation, two-generation and extended one-generation reproductive studies. The practical value is that the standard helps users select a study design that matches the question being asked.
Clause 8 - document the work clearly
Clause 8 sets out the minimum test report content. The report needs the standard used, the test sample description, method rationale, analytical methods, quality-practice statement where applicable, S9 mix details, controls, statistical treatment, deviations, interpretation and follow-up details.
This matters for audits, regulatory review and technical files. The report must show not just the result, but how the decision was reached.
What terms does SIST EN ISO 10993-3:2026 define?
- Carcinogenicity test - a test to determine cancer outcomes after long-term or life-span exposure of animals.
- Developmental toxicity evaluation - methods used to determine adverse effects on the developing organism during prenatal development or after birth up to sexual maturation.
- Genotoxicity test - a test in cells, bacteria, yeast, fungi, animals or other biological systems to determine whether gene mutations, chromosomal changes or other genetic damage are caused by the test sample.
- Reproductive toxicity evaluation - methods used to evaluate potential effects on male or female reproductive function during any stage of development or fertility.
- Sham-incubated vehicle - a vehicle control exposed to the same extraction time and temperature as the test sample extract.
- Threshold of toxicological concern (TTC) - an exposure level for constituents below which there is no appreciable risk to human health.
Who uses SIST EN ISO 10993-3:2026?
SIST EN ISO 10993-3:2026 is used by medical device manufacturers, toxicology laboratories, regulatory affairs teams, risk managers, quality teams and test houses working on biological evaluation. It is also relevant to suppliers of device materials and components that may release residues, degradation products or leachables.
In practice, these users apply the standard when deciding whether to run genotoxicity, carcinogenicity, reproductive toxicity or developmental toxicity work, how to prepare extracts, how to interpret chemical characterization, and how to document the conclusion in a biological safety or risk management file.
What changed in SIST EN ISO 10993-3:2026 from the previous edition?
SIST EN ISO 10993-3:2026 supersedes EN ISO 10993-3:2014. The ISO edition also cancels and replaces ISO 10993-3:2014 and ISO/TR 10993-33:2015.
The main changes stated for this edition are:
- the document has been entirely revised to emphasize evaluation rather than testing
- chemical characterization and toxicological risk assessment have been added as an approach
- the follow-up evaluation after a positive in vitro genotoxicity test has changed, and the old flowchart has been removed
- reproductive toxicity and developmental toxicity are now described separately
- the annex on cellular transformation has been deleted
- ISO/TR 10993-33 has been merged into this document as Annex B on genotoxicity test methods
Which standards are used with SIST EN ISO 10993-3:2026?
The document relies on several standards and guidelines to make its evaluation process work:
- ISO 10993-1 - general principles for biological safety evaluation within a risk management process
- ISO 10993-2 - animal welfare requirements
- ISO 10993-12 - sample preparation and reference materials
- ISO 10993-17 - toxicological risk assessment of medical device constituents
- ISO 10993-18 - chemical characterization used throughout the assessment strategy
- ISO 10993-16 - toxicokinetic study design for degradation products and leachables
- ISO 14971 - medical device risk management
- OECD 471, 473, 487, 490 and 476 - genotoxicity test methods
- OECD 414, 415, 416, 421, 422, 443 and 43 - reproductive and developmental toxicity study designs
- OECD 451 and 453 - carcinogenicity study designs
- ISO/TR 10993-22 - nanomaterials guidance used in the nanomaterial annex
What does the SIST EN ISO 10993-3:2026 document contain?
SIST EN ISO 10993-3:2026 contains practical guidance for choosing between toxicological risk assessment and biological testing, plus annexes that support specific test types and materials. Annex A gives sample preparation methods for genotoxicity testing, including methods for suspensions, extracts and optional evaporated residues.
Annex B is the core genotoxicity section and sets out the modified bacterial and mammalian test approach, including dose selection, controls, metabolic activation and follow-up evaluation. Annex C explains why carcinogenicity testing of medical devices is difficult and how animal studies are approached when they are used.
Annex D focuses on released nanomaterials, Annex E provides an example S9 mix recipe, and Annex F connects endocrine disruptors to reproductive and developmental concerns. Annex G summarizes common OECD study designs for developmental toxicity and reproductive toxicity, while Annex ZA links the standard to the General Safety and Performance Requirements of Regulation (EU) 2017/745.
Relations
- Effective Date
- 01-Nov-2026
- Effective Date
- 16-Sep-2026
- Effective Date
- 16-Sep-2026
- Effective Date
- 16-Sep-2026
Get Certified
Connect with accredited certification bodies for this standard

BSI Group
BSI (British Standards Institution) is the business standards company that helps organizations make excellence a habit.

TÜV Rheinland
TÜV Rheinland is a leading international provider of technical services.

TÜV SÜD
TÜV SÜD is a trusted partner of choice for safety, security and sustainability solutions.
Sponsored listings
Frequently Asked Questions
SIST EN ISO 10993-3:2026 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Biological evaluation of medical devices - Part 3: Evaluation of genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity (ISO 10993-3:2026)". This standard covers: This document specifies strategies for risk estimation and evaluation of biological harms with respect to: genotoxicity; carcinogenicity; reproductive toxicity; and developmental toxicity. This document is applicable when the need to evaluate a medical device for potential genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity has been established. This document is not applicable to active pharmaceutical ingredients of device-drug combination products or biological components of device-biologic combination products which are covered by regulations. NOTE Guidance on selecting relevant biological effects for medical devices is covered in ISO 10993-1.
This document specifies strategies for risk estimation and evaluation of biological harms with respect to: genotoxicity; carcinogenicity; reproductive toxicity; and developmental toxicity. This document is applicable when the need to evaluate a medical device for potential genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity has been established. This document is not applicable to active pharmaceutical ingredients of device-drug combination products or biological components of device-biologic combination products which are covered by regulations. NOTE Guidance on selecting relevant biological effects for medical devices is covered in ISO 10993-1.
SIST EN ISO 10993-3:2026 is classified under the following ICS (International Classification for Standards) categories: 11.100.20 - Biological evaluation of medical devices. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN ISO 10993-3:2026 has the following relationships with other standards: It is inter standard links to SIST EN ISO 10993-3:2015, SIST-TS CEN/TS 15277:2006, SIST EN 455-3:2007, SIST EN 60601-1:2007/A2:2021. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST EN ISO 10993-3:2026 is associated with the following European legislation: EU Directives/Regulations: 2017/745; Standardization Mandates: M/575, M/575 Amd 2. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
SIST EN ISO 10993-3:2026 is available in PDF format for immediate download after purchase. The document can be added to your cart and obtained through the secure checkout process. Digital delivery ensures instant access to the complete standard document.
Standards Content (Sample)
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 3.1 carcinogenicity test
- 3.2 developmental toxicity evaluation
- 3.3 genotoxicity test
- 3.4 reproductive toxicity evaluation
- 3.5 sham-incubated vehicle
- 3.6 threshold of toxicological concern
- 4 Assessment strategies
- 4.1 General
- 4.2 Other considerations
- 5 Genotoxicity assessment
- 5.1 General
- 5.2 Testing strategy
- 5.3 Test sample preparation
- 6 Carcinogenicity assessments
- 6.1 General
- 6.2 Evaluation strategy
- 7 Evaluation of reproductive toxicity and developmental toxicity
- 7.1 General
- 7.2 Evaluation methods
- 8 Test report
- Annex A
- Annex B
- 5 Absent
- 6 Obscured by particulates
- 7 Non-interfering precipitate
- 8 Interfering precipitate
- Annex C
- Annex D
- Annex E
- 1 M phosphate buffer (pH 7,4) 1,0 ml
- Annex F
- Annex G
- Bibliography
Introduction
The basis for biological evaluation of medical devices is often empirical and driven by relevant concerns for human safety. The risk of serious and irreversible effects, such as cancer or second-generation abnormalities, is of particular public concern. It is inherent in the provision of safe medical devices that such risks be minimized to the greatest extent feasible. The assessment of mutagenic, carcinogenic, reproductive toxicity and developmental toxicity hazards is an essential component of the control of these risks. Not all test methods for the assessment of genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity are equally well developed, nor is their validity well established for the testing of medical devices. Significant issues with test sample size and preparation, scientific understanding of disease processes and test validation can be cited as limitations of available methods. Since the previous edition of this document, many genotoxicity test methods have been updated with revised OECD guidelines. However, these generally provide clearer recommendations but little alteration in test methods. Scientifically sound alternatives to the proposed testing can be acceptable insofar as they address relevant matters of safety assessment of the medical device.
In the selection of tests needed to evaluate a particular medical device, a careful assessment of expected human uses and potential interactions with biological systems is important, particularly in such areas as reproductive toxicity and developmental toxicity.
This document presents test methods and evaluation strategies for the identification of specific biological harms. Testing is not always necessary or appropriate in evaluating biological risks associated with exposure to medical device materials but, where it is appropriate, it is important that maximum test sensitivity is achieved.
In view of the multitude of possible outcomes and the importance of factors such as extent of exposure, species differences and mechanical or physical considerations, risk assessment is typically performed on a [7] case-by-case basis. Suggestions for risk consideration and integration of ISO 10993-17 and ISO 10993-18 are provided.
SIST EN ISO 10993-3:2026
SIST EN ISO 10993-3:2026 International Standard ISO 10993-3:2026(en) Biological evaluation of medical devices — Part 3:
Evaluation of genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity
1 Scope
This document specifies strategies for risk estimation and evaluation of biological harms with respect to:
- genotoxicity;
- carcinogenicity;
- reproductive toxicity; and
- developmental toxicity.
This document is applicable when the need to evaluate a medical device for potential genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity has been established. This document is not applicable to active pharmaceutical ingredients of device-drug combination products or biological components of device-biologic combination products which are covered by regulations. NOTE Guidance on selecting relevant biological effects for medical devices is covered in ISO 10993-1.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. ISO 10993-1:2025, Biological evaluation of medical devices — Part 1: Requirements and general principles for the evaluation of biological safety within a risk management process ISO 10993-2, Biological evaluation of medical devices — Part 2: Animal welfare requirements ISO 10993-12:2021, Biological evaluation of medical devices — Part 12: Sample preparation and reference materials ISO 10993-17, Biological evaluation of medical devices — Part 17: Toxicological risk assessment of medical device constituents ISO 14971:2019, Medical devices — Application of risk management to medical devices OECD 471:2020, OECD Guideline for the Testing of Chemicals — Bacterial Reverse Mutation Test OECD 473:2016, OECD Guideline for the Testing of Chemicals — In vitro Mammalian Chromosome Aberration Test OECD 487:2016, OECD Guideline for the Testing of Chemicals — In vitro Mammalian Cell Micronucleus Test OECD 490:2016, OECD Guideline for the Testing of Chemicals — In vitro Mammalian Cell Gene Mutation Tests Using the Thymidine Kinase Gene
SIST EN ISO 10993-3:2026 ISO 10993-3:2026(en)
3 Terms and definitions
For the purposes of this document, the terms and definitions given in ISO 10993-1, ISO 10993-12 and the following apply.
ISO and IEC maintain terminology databases for use in standardization at the following addresses:
- ISO Online browsing platform: available at https:// www .iso .org/ obp
- IEC Electropedia: available at https:// www .electropedia .org/
3.1 carcinogenicity test
test to determine cancer outcomes after long term or life-span exposure of animals Note 1 to entry: Outcomes are typically based on tumorigenic potential and other factors.
3.2 developmental toxicity evaluation
methods to determine adverse effects on the developing organism during prenatal development or postnatally to the time of sexual maturation
3.3 genotoxicity test
test using cells, bacteria, yeast, fungi, animals or other biological systems to determine whether gene mutations, chromosomal alterations or other genetic changes are caused by the test sample
3.4 reproductive toxicity evaluation
methods to evaluate the potential effects of test samples on male or female reproductive function during any stage of development or fertility
3.5 sham-incubated vehicle
vehicle control exposed to the same extraction time and temperature as the test sample extract
3.6 threshold of toxicological concern
TTC level of exposure for constituents, below which there would be no appreciable risk to human health Note 1 to entry: The constituent is defined as a chemical or compound present in or on a finished medical device or material(s) of construction. Constituents can be intentionally present (e.g. an additive such as an antioxidant) or unintentionally present (e.g. an impurity).
[11] [SOURCE: ISO/TS 21726:2019, 3.5 , modified — Note 1 to entry has been added.]
4 Assessment strategies
4.1 General
ISO 10993-1 indicates circumstances where the potential for genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity are relevant biological harms for consideration in an overall biological safety evaluation. Evaluation of the biological harms shall consider the following factors:
- whether assessment is necessary based on patient population (e.g. life expectancy, sensitive population) and the anticipated impact of test results on risk management judgements,
- whether assessment is necessary based on the worst-case quantity, type of contact and duration of exposure,
SIST EN ISO 10993-3:2026 ISO 10993-3:2026(en) NOTE Quantity can be influenced by device size and number used per patient. - an analysis of the chemical constituents of the device material(s), including manufacturing process residues and degradation products or metabolites, to identify causes of concern. The identification can be based on chemical-specific toxicity data if available; otherwise, structure-activity relationships or previous demonstration of relevant toxicity in the chemical class can be used, - existing information relevant to the genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity evaluation of the medical device, and - previous use of equivalent (see ISO 10993-1:2025, 6.7) materials and processing in relevant applications. The decision to waive in vitro genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity testing shall be justified. The decision to conduct in vivo testing shall be justified. [7] If exposure to constituents identified in accordance with ISO 10993-18 is determined to be within acceptable levels based on a toxicological risk assessment conducted in accordance with ISO 10993-17, no additional testing needs to be performed to address the relevant risks.
4.2 Other considerations
Toxicity evaluation can be warranted for additional states of the device, such as wear debris generated from the device, absorbable materials, or materials that cure in situ (e.g. cements, adhesives, pre-polymer mixtures) unless toxicological risk assessment based on material information determines no cause for concern. For guidance on test sample preparation for in situ curing devices, see ISO 10993-12. NOTE Material information can include chemical composition, degradation products, debris size and morphology, pre-polymerization, post-polymerization products, etc.
5 Genotoxicity assessment
5.1 General
Assessment of genotoxic risk should be made:
a) by conducting toxicological risk assessment based on chemical characterization according to [7] ISO 10993-17 and ISO 10993-18 ; or b) by performing genotoxicity testing in the standard test battery according to 5.2.2. In certain cases (e.g. a material which has not been used in a legally marketed device for its intended use), both toxicological risk assessment of medical device constituents based on chemical characterization and genotoxicity testing of a medical device can be appropriate.
NOTE 1 Genotoxicity testing can also be appropriate when chemical characterization data is insufficient for a toxicological risk assessment.
If the device is made with and releases nanomaterials, additional evaluation shall be considered. [9] NOTE 2 See ISO/TR 10993-22 and Annex D for more information.
If chemical characterization of the device identifies genotoxicants which are not easily detected using non- [7] targeted analytical approaches described in ISO 10993-18 (e.g. glutaraldehyde, formaldehyde), targeted analysis of genotoxic residuals shall be conducted. Where compositional profiling or analytical chemistry reveals the presence of chemical constituents with inadequate genotoxicity data and where quantitative structure activity relationship (QSAR) information is inconclusive (see ISO 10993-17), testing of individual chemicals or device components containing the chemical should be considered. In either case, the test sample shall be justified and documented.
SIST EN ISO 10993-3:2026 ISO 10993-3:2026(en)
5.2 Testing strategy
5.2.1 General
No single test is capable of detecting all types of genotoxic agents. Therefore, the usual approach is to conduct a battery of in vitro and, under certain circumstances, also in vivo tests. Genotoxicity tests are designed to detect the two major classes of genetic damage:
- gene mutations (DNA single base pair changes, small deletions or insertions), and - chromosomal damage [structural aberrations such as translocations, small or large deletions and insertions, and gain or loss of whole chromosomes (aneuploidy) or parts of chromosomes]. NOTE Bacterial reverse mutation assays have been shown to detect relevant small-scale mutational changes produced by the majority of genotoxic carcinogens detected by rodent bioassays.
5.2.2 In vitro genotoxic test battery
When genotoxicity testing is performed, it shall be conducted in accordance with Annex B and the test battery shall include:
a) a test for gene mutations in bacteria (see OECD 471), modified for use with medical devices, see Clause B.2, and b) one of the following three validated in vitro mammalian cell tests:
1) an in vitro test with cytogenetic evaluation of chromosomal damage with mammalian cells (see OECD 473), modified for medical devices, in accordance with Clause B.3 and B.4.2, or
2) an in vitro mouse lymphoma tk assay using L5178Y (tk ) cells (see OECD 490), modified for medical devices, including detection of small (slow growing) and large colonies, see Clause B.3 and B.4.1, or 3) an in vitro mammalian cell micronucleus test for chromosomal damage and aneugenicity (see OECD 487), modified for medical devices, see Clause B.3 and B.4.3.
NOTE 1 If the medical device is known to contain nanomaterials testing according to OECD 471 is not recommended, see Annex D.
All tests (bacterial and in vitro mammalian tests) shall be performed in the presence and absence of metabolic activation. Metabolic activation is achieved by addition of a co-factor-supplemented postmitochondrial fraction (S9). NOTE 2 An example for a S9 recipe is given in Annex E.
In vitro mammalian tests shall include both a short-term exposure with and without S9 and a long-term exposure without S9.
NOTE 3 See B.3.2.3 for descriptions of short-and long-term exposure.
Results from both chromosomal aberration assays and the in vitro mouse lymphoma tk assay (MLA) have a relatively high level of congruence for compounds that are regarded as genotoxic but yield negative results in the bacterial reverse mutation assay. The MLA detects the broadest set of genetic damage, including both small-scale and large-scale genetic damages. The micronucleus assay detects chromosomal damage in the form of whole chromosome loss or chromosomal fragments and uniquely detects aneuploidy. The three tests are equally acceptable as an in vitro mammalian genotoxicity test (see OECD 473, OECD 487 and OECD 490). The hypoxanthine guanine phosphoribosyl transferase gene mutation assay (HGPRT/HPRT) method (see [84] OECD 476 ) is not included in the recommended test battery for device testing. This test can be appropriate for medical devices containing nanomaterials (see Annex D).
SIST EN ISO 10993-3:2026 ISO 10993-3:2026(en)
5.2.3 In vivo genotoxicity testing
In vivo genotoxicity tests are not part of the recommended genotoxicity test battery because they are relatively insensitive and not likely to detect genotoxicity at the concentrations of substances generally [73][74] found in medical devices or in medical device extracts . These concentrations can be insufficient to meet the limit dose required by the test methods. However, under certain circumstances, an in vivo test can be useful.
These circumstances can include:
a) assessment of nanomaterial or other particulates suspected of sequestration in particular organs or [9] tissues (see Annex D, Reference [30] and ISO/TR 10993-22 ); or [6] b) when additional relevant factors [e.g. toxicokinetics (see ISO 10993-16 ), organ-specific toxicity] need to be considered.
In vivo tests are valid only with data demonstrating the test sample reached the target cells, organs, or tissues. Mutational analysis in target organs of rats or mice can be determined using one of several rodent [27] transgenic mutation assays (see OECD 488 ); chromosomal damage can be assessed in the erythrocyte [24] [28] micronucleus assay (see OECD 474 ), the in vivo comet assay (see OECD 489 ), or the in vivo chromosomal [25] aberration assay (see OECD 475 ). See Clause B.5 for additional considerations for in vivo assays. All animal studies shall be performed in a facility meeting the requirements of ISO 10993-2.
5.2.4 Follow-up evaluation
If in vitro genotoxicity tests are performed in accordance with 5.2.2 and if the tests are negative, the test sample can be considered non-genotoxic and further testing is not necessary. Equivocal results (i.e. elevated results that do not meet criteria for a positive result), can require repeating tests with a modified protocol such as adjusting dilution of the test sample or extract. When any single in vitro test is positive or equivocal, then the following options shall be considered:
a) Identification of genotoxic constituents by chemical characterization (information gathering or analytical chemistry testing) or other testing, with appropriate steps taken to eliminate or manage any identified risks.
b) If confounding factors (e.g. osmolality, pH, cytotoxicity) are suspected to have caused the positive response or equivocal in an initial genotoxicity test, the dosing or test conditions may be modified with a justification to address confounding factors.
c) Presume that a genotoxicity risk exists and manage the risk in accordance with ISO 14971.
5.3 Test sample preparation
Test samples should be extracted in vehicles compatible with the test system (see Annex A). Use of extraction vehicles that are incompatible with the test system requires additional manipulation (e.g. solvent removal) that can alter the composition of the extract and, therefore, affect the results of the test. Use of extraction vehicles that are incompatible with the test system shall be justified. Extraction shall be performed in accordance with ISO 10993-12 unless otherwise justified. Tests can be performed on solutions (e.g. soluble or liquid devices), suspensions (e.g. Method A in Annex A), extracts (e.g. Method B in Annex A) of the finished product, medical device components or individual chemicals from the medical device. If individual chemicals are tested, then the appropriate OECD guidelines shall be followed.
NOTE Some authorities having jurisdiction prefer Method C in Annex A while others do not accept Method C.
SIST EN ISO 10993-3:2026 ISO 10993-3:2026(en)
6 Carcinogenicity assessments
6.1 General
The objective of carcinogenicity studies is to determine tumorigenic potential of medical devices or medical device constituents in animals and to evaluate the relevant risk to patients. Mechanisms of carcinogenesis can be genotoxic or non-genotoxic. Medical devices shall be evaluated for both types of carcinogenicity, preferably by alternative approaches, such as chemical characterization and toxicological risk assessment, that do not use in vivo testing.
Animal testing for carcinogenicity is technically challenging. It requires implantation of many medical devices, representative samples of the medical device, or the application of concentrated medical device extracts to achieve the elevated doses necessary for a valid test in a reasonable number of animals. For these reasons, animal testing is generally not justified when the risks can be adequately assessed by other means. NOTE For additional information on animal carcinogenicity testing, see Annex C.
6.2 Evaluation strategy
6.2.1 General
[7] Carcinogenicity should be evaluated initially using chemical characterization according to ISO 10993-18 and toxicological risk assessment according to ISO 10993-17. When medical devices or materials are not readily amenable to analytical chemistry testing and adequate chemical characterization information is available from information gathering, it is always preferable to use the chemical information and toxicological risk assessment including assessment of available genotoxicity data to evaluate carcinogenicity rather than performing an animal carcinogenicity study.
NOTE 1 Some authorities having jurisdiction can require consideration of carcinogenicity testing for novel materials (e.g. when data are not available to provide an adequate assessment). NOTE 2 Evaluation of solid-state carcinogenesis due to release of particulates or fibres from a medical device can be partially addressed by toxicological risk assessment according to ISO 10993-17. This does not address all the risks. See Annex C for more information.
If the results of toxicological risk assessment indicate the risk of carcinogenicity is acceptable, no further assessment is required.
Carcinogenicity testing shall not be performed when risks can be adequately assessed or managed without generating new carcinogenicity test data or when the benefit-risk analysis obviates the need for a carcinogenicity assessment. If the toxicological risk of carcinogenicity is unacceptable, the risk shall be managed accordingly (see also ISO 14971).
...