SIST EN ISO 10993-2:2006
(Main)Biological evaluation of medical devices - Part 2: Animal welfare requirements (ISO 10993-2:2006)
Biological evaluation of medical devices - Part 2: Animal welfare requirements (ISO 10993-2:2006)
ISO 10993-2:2006 is aimed at those who commission, design and perform tests or evaluate data from animal tests undertaken to assess the biocompatibility of materials intended for use in medical devices, or of the medical devices themselves. It specifies the minimum requirements to be satisfied to ensure and demonstrate that proper provision has been made for the welfare of animals used in animal tests to assess the biocompatibility of materials used in medical devices.
ISO 10993-2:2006 also makes recommendations and offers guidance intended to facilitate future further reductions in the overall number of animals used, refinement of test methods to reduce or eliminate pain or distress in animals, and the replacement of animal tests by other scientifically valid means not requiring animal tests.
It applies to tests performed on living vertebrate animals, other than man, to establish the biocompatibility of materials or medical devices.
Biologische Beurteilung von Medizinprodukten - Teil 2: Tierschutzbestimmungen (ISO 10993-2:2006)
Dieser Teil von ISO 10993 ist für den Personenkreis gedacht, der Versuche in Auftrag gibt, sie plant und durchführt oder Daten aus Tierversuchen auswertet, um die Bioverträglichkeit von Materialien, die in Medizinprodukten verwendet werden sollen oder von Medizinprodukten selbst, zu bewerten. Er legt die Mindestanforderungen fest, die erfüllt werden müssen, um sicherzustellen und nachzuweisen, dass für den Schutz der Tiere, die in Versuchen zur Bewertung der Bioverträglichkeit von in Medizinprodukten zu verwendenden Materialien verwendet werden, die richtigen Vorkehrungen getroffen wurden.
In ihm werden auch Empfehlungen gemacht und eine Anleitung angeboten, deren Ziel eine zukünftige weitere Verminderung der Gesamtzahl der verwendeten Tiere, eine Verbesserung der Prüfverfahren zur Verringerung oder Ausschaltung von Schmerz oder Leiden bei Tieren, und der Ersatz von Tierversuchen durch andere wissenschaftlich valide Mittel ist, bei denen Tierversuche nicht erforderlich sind.
Er gilt für Versuche an lebenden Wirbeltieren, außer dem Menschen, zur Feststellung der Bioverträglichkeit von Materialien oder Medizinprodukten.
Er gilt nicht für Versuche an Wirbellosen und anderen niedrigeren Formen des Lebens, auch nicht (außer für die Vorkehrungen hinsichtlich Tierart, Bezugsquellen, Gesundheitszustand, Pflege und Unterbringung) für Versuche an isolierten Geweben und Organen, die aus auf humane Weise getöteten Wirbeltieren entnommen wurden.
Évaluation biologique des dispositifs médicaux - Partie 2: Exigences relatives a la protection des animaux (ISO 10993-2:2006)
L'ISO 10993-2:2006 s'adresse aux personnes qui commandent, conçoivent et réalisent des essais ou qui évaluent les données provenant d'expérimentations animales entreprises pour évaluer la biocompatibilité des matériaux destinés à être utilisés dans des dispositifs médicaux ou des dispositifs médicaux eux-mêmes. Elle spécifie les exigences minimales à satisfaire afin de garantir et de démontrer que des dispositions appropriées ont été prises concernant la protection des animaux utilisés dans les expérimentations animales à des fins d'évaluation de la biocompatibilité de matériaux utilisés dans les dispositifs médicaux.
L'ISO 10993-2:2006 donne également des recommandations et propose des directives visant à réduire dans l'avenir le nombre total d'animaux utilisés, l'affinement des méthodes d'essai en vue de minimiser ou éliminer la douleur ou la détresse des animaux, ainsi que le remplacement des expérimentations animales par d'autres moyens scientifiquement valables ne nécessitant pas l'expérimentation sur les animaux.
Elle est applicable à l'expérimentation animale effectuée sur des vertébrés vivants, à l'exclusion de l'homme, afin d'établir la biocompatibilité des matériaux ou des dispositifs médicaux.
Biološko ovrednotenje medicinskih pripomočkov – 2. del: Zahteve za varstvo živali (ISO 10993-2:2006)
General Information
- Status
- Withdrawn
- Publication Date
- 31-Aug-2006
- Withdrawal Date
- 05-Jan-2023
- Technical Committee
- VAZ - Healthcare
- Current Stage
- 9900 - Withdrawal (Adopted Project)
- Start Date
- 06-Jan-2023
- Due Date
- 29-Jan-2023
- Completion Date
- 06-Jan-2023
Relations
- Effective Date
- 01-Sep-2006
- Effective Date
- 01-Feb-2023
Frequently Asked Questions
SIST EN ISO 10993-2:2006 is a standard published by the Slovenian Institute for Standardization (SIST). Its full title is "Biological evaluation of medical devices - Part 2: Animal welfare requirements (ISO 10993-2:2006)". This standard covers: ISO 10993-2:2006 is aimed at those who commission, design and perform tests or evaluate data from animal tests undertaken to assess the biocompatibility of materials intended for use in medical devices, or of the medical devices themselves. It specifies the minimum requirements to be satisfied to ensure and demonstrate that proper provision has been made for the welfare of animals used in animal tests to assess the biocompatibility of materials used in medical devices. ISO 10993-2:2006 also makes recommendations and offers guidance intended to facilitate future further reductions in the overall number of animals used, refinement of test methods to reduce or eliminate pain or distress in animals, and the replacement of animal tests by other scientifically valid means not requiring animal tests. It applies to tests performed on living vertebrate animals, other than man, to establish the biocompatibility of materials or medical devices.
ISO 10993-2:2006 is aimed at those who commission, design and perform tests or evaluate data from animal tests undertaken to assess the biocompatibility of materials intended for use in medical devices, or of the medical devices themselves. It specifies the minimum requirements to be satisfied to ensure and demonstrate that proper provision has been made for the welfare of animals used in animal tests to assess the biocompatibility of materials used in medical devices. ISO 10993-2:2006 also makes recommendations and offers guidance intended to facilitate future further reductions in the overall number of animals used, refinement of test methods to reduce or eliminate pain or distress in animals, and the replacement of animal tests by other scientifically valid means not requiring animal tests. It applies to tests performed on living vertebrate animals, other than man, to establish the biocompatibility of materials or medical devices.
SIST EN ISO 10993-2:2006 is classified under the following ICS (International Classification for Standards) categories: 11.100.20 - Biological evaluation of medical devices. The ICS classification helps identify the subject area and facilitates finding related standards.
SIST EN ISO 10993-2:2006 has the following relationships with other standards: It is inter standard links to SIST EN ISO 10993-2:2000, SIST EN ISO 10993-2:2023. Understanding these relationships helps ensure you are using the most current and applicable version of the standard.
SIST EN ISO 10993-2:2006 is associated with the following European legislation: EU Directives/Regulations: 93/42/EEC; Standardization Mandates: M/023. When a standard is cited in the Official Journal of the European Union, products manufactured in conformity with it benefit from a presumption of conformity with the essential requirements of the corresponding EU directive or regulation.
You can purchase SIST EN ISO 10993-2:2006 directly from iTeh Standards. The document is available in PDF format and is delivered instantly after payment. Add the standard to your cart and complete the secure checkout process. iTeh Standards is an authorized distributor of SIST standards.
Standards Content (Sample)
SLOVENSKI STANDARD
01-september-2006
1DGRPHãþD
SIST EN ISO 10993-2:2000
%LRORãNRRYUHGQRWHQMHPHGLFLQVNLKSULSRPRþNRY±GHO=DKWHYH]DYDUVWYRåLYDOL
,62
Biological evaluation of medical devices - Part 2: Animal welfare requirements (ISO
10993-2:2006)
Biologische Beurteilung von Medizinprodukten - Teil 2: Tierschutzbestimmungen (ISO
10993-2:2006)
Évaluation biologique des dispositifs médicaux - Partie 2: Exigences relatives a la
protection des animaux (ISO 10993-2:2006)
Ta slovenski standard je istoveten z: EN ISO 10993-2:2006
ICS:
11.100.20 %LRORãNRRYUHGQRWHQMH Biological evaluation of
PHGLFLQVNLKSULSRPRþNRY medical devices
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD
EN ISO 10993-2
NORME EUROPÉENNE
EUROPÄISCHE NORM
July 2006
ICS 11.100.20 Supersedes EN ISO 10993-2:1998
English Version
Biological evaluation of medical devices - Part 2: Animal welfare
requirements (ISO 10993-2:2006)
Évaluation biologique des dispositifs médicaux - Partie 2: Biologische Beurteilung von Medizinprodukten - Teil 2:
Exigences relatives à la protection des animaux (ISO Tierschutzbestimmungen (ISO 10993-2:2006)
10993-2:2006)
This European Standard was approved by CEN on 14 July 2006.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European
Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national
standards may be obtained on application to the Central Secretariat or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation
under the responsibility of a CEN member into its own language and notified to the Central Secretariat has the same status as the official
versions.
CEN members are the national standards bodies of Austria, Belgium, Cyprus, Czech Republic, Denmark, Estonia, Finland, France,
Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania,
Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION
EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre: rue de Stassart, 36 B-1050 Brussels
© 2006 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN ISO 10993-2:2006: E
worldwide for CEN national Members.
Foreword
This document (EN ISO 10993-2:2006) has been prepared by Technical Committee ISO/TC 194
"Biological evaluation of medical devices" in collaboration with Technical Committee CEN/TC
206 "Biocompatibility of medical and dental materials and devices", the secretariat of which is
held by NEN.
This European Standard shall be given the status of a national standard, either by publication of
an identical text or by endorsement, at the latest by January 2007, and conflicting national
standards shall be withdrawn at the latest by January 2007.
This document supersedes EN ISO 10993-2:1998.
According to the CEN/CENELEC Internal Regulations, the national standards organizations of
the following countries are bound to implement this European Standard: Austria, Belgium,
Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary,
Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland,
Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
Endorsement notice
The text of ISO 10993-2:2006 has been approved by CEN as EN ISO 10993-2:2006 without any
modifications.
INTERNATIONAL ISO
STANDARD 10993-2
Second edition
2006-07-15
Biological evaluation of medical
devices —
Part 2:
Animal welfare requirements
Évaluation biologique des dispositifs médicaux —
Partie 2: Exigences relatives à la protection des animaux
Reference number
ISO 10993-2:2006(E)
©
ISO 2006
ISO 10993-2:2006(E)
PDF disclaimer
This PDF file may contain embedded typefaces. In accordance with Adobe's licensing policy, this file may be printed or viewed but
shall not be edited unless the typefaces which are embedded are licensed to and installed on the computer performing the editing. In
downloading this file, parties accept therein the responsibility of not infringing Adobe's licensing policy. The ISO Central Secretariat
accepts no liability in this area.
Adobe is a trademark of Adobe Systems Incorporated.
Details of the software products used to create this PDF file can be found in the General Info relative to the file; the PDF-creation
parameters were optimized for printing. Every care has been taken to ensure that the file is suitable for use by ISO member bodies. In
the unlikely event that a problem relating to it is found, please inform the Central Secretariat at the address given below.
© ISO 2006
All rights reserved. Unless otherwise specified, no part of this publication may be reproduced or utilized in any form or by any means,
electronic or mechanical, including photocopying and microfilm, without permission in writing from either ISO at the address below or
ISO's member body in the country of the requester.
ISO copyright office
Case postale 56 • CH-1211 Geneva 20
Tel. + 41 22 749 01 11
Fax + 41 22 749 09 47
E-mail copyright@iso.org
Web www.iso.org
Published in Switzerland
ii © ISO 2006 – All rights reserved
ISO 10993-2:2006(E)
Contents Page
Foreword. iv
Introduction . vi
1 Scope . 1
2 Normative references . 1
3 Terms and definitions. 1
4 Requirements . 3
4.1 General. 3
4.2 Justification for animal tests . 4
4.3 Competence of personnel. 4
4.4 Planning and performance of animal tests .4
4.5 Test strategy — Sequence of in vitro and in vivo tests. 5
4.6 Animal care and accommodation . 5
4.7 Humane endpoints. 6
4.8 Study documentation . 7
4.9 Validity of test results and mutual acceptance of data. 7
Annex A (informative) Rationale for the development of this part of ISO 10993. 8
Annex B (informative) Further suggestions for replacing, reducing and refining animal tests . 12
Bibliography . 13
ISO 10993-2:2006(E)
Foreword
ISO (the International Organization for Standardization) is a worldwide federation of national standards bodies
(ISO member bodies). The work of preparing International Standards is normally carried out through ISO
technical committees. Each member body interested in a subject for which a technical committee has been
established has the right to be represented on that committee. International organizations, governmental and
non-governmental, in liaison with ISO, also take part in the work. ISO collaborates closely with the
International Electrotechnical Commission (IEC) on all matters of electrotechnical standardization.
International Standards are drafted in accordance with the rules given in the ISO/IEC Directives, Part 2.
The main task of technical committees is to prepare International Standards. Draft International Standards
adopted by the technical committees are circulated to the member bodies for voting. Publication as an
International Standard requires approval by at least 75 % of the member bodies casting a vote.
Attention is drawn to the possibility that some of the elements of this document may be the subject of patent
rights. ISO shall not be held responsible for identifying any or all such patent rights.
ISO 10993-2 was prepared by Technical Committee ISO/TC 194, Biological evaluation of medical devices.
This second edition cancels and replaces the first edition (ISO 10993-2:1992), which has been technically
revised.
ISO 10993 consists of the following parts, under the general title Biological evaluation of medical devices:
⎯ Part 1: Evaluation and testing
⎯ Part 2: Animal welfare requirements
⎯ Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
⎯ Part 4: Selection of tests for interactions with blood
⎯ Part 5: Tests for in vitro cytotoxicity
⎯ Part 6: Tests for local effects after implantation
⎯ Part 7: Ethylene oxide sterilization residuals
⎯ Part 9: Framework for identification and quantification of potential degradation products
⎯ Part 10: Tests for irritation and delayed-type hypersensitivity
⎯ Part 11: Tests for systemic toxicity
⎯ Part 12: Sample preparation and reference materials
⎯ Part 13: Identification and quantification of degradation products from polymeric medical devices
⎯ Part 14: Identification and quantification of degradation products from ceramics
⎯ Part 15: Identification and quantification of degradation products from metals and alloys
iv © ISO 2006 – All rights reserved
ISO 10993-2:2006(E)
⎯ Part 16: Toxicokinetic study design for degradation products and leachables
⎯ Part 17: Establishment of allowable limits for leachable substances
⎯ Part 18: Chemical characterization of materials
⎯ Part 19: Physico-chemical, morphological and topographical characterization of materials
⎯ Part 20: Principles and methods for immunotoxicology testing of medical devices
ISO 10993-2:2006(E)
Introduction
The goal of the ISO 10993 series of International Standards is the protection of humans in the context of the
use of medical devices.
This part of ISO 10993 supports the goal of the ISO 10993 series by promoting good science through paying
proper regard to maximizing the use of scientifically sound non-animal tests and by ensuring that those animal
tests performed to evaluate the biological properties of materials used in medical devices are conducted
humanely according to recognized ethical and scientific principles.
The application of such humane experimental techniques, including high standards of animal care and
accommodation, both help to ensure the scientific validity of safety testing and enhance the welfare of the
animals used.
vi © ISO 2006 – All rights reserved
INTERNATIONAL STANDARD ISO 10993-2:2006(E)
Biological evaluation of medical devices —
Part 2:
Animal welfare requirements
1 Scope
This part of ISO 10993 is aimed at those who commission, design and perform tests or evaluate data from
animal tests undertaken to assess the biocompatibility of materials intended for use in medical devices, or that
of the medical devices themselves. It specifies the minimum requirements to be satisfied to ensure and
demonstrate that proper provision has been made for the welfare of animals used in animal tests to assess
the biocompatibility of materials used in medical devices.
It also makes recommendations and offers guidance intended to facilitate future further reductions in the
overall number of animals used, refinement of test methods to reduce or eliminate pain or distress in animals,
and the replacement of animal tests by other scientifically valid means not requiring animal tests.
It applies to tests performed on living vertebrate animals, other than man, to establish the biocompatibility of
materials or medical devices.
It does not apply to tests performed on invertebrate animals and other lower forms; nor (other than with
respect to provisions relating to species, source, health status, and care and accommodation) does it apply to
testing performed on isolated tissues and organs taken from vertebrate animals that have been euthanized.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.
ISO 10993-1:2003, Biological evaluation of medical devices — Part 1: Evaluation and testing
3 Terms and definitions
For the purposes of this document, the definitions given in ISO 10993-1 and the following apply.
3.1
alternative method
any test method that replaces an animal test, reduces the numbers of animals used, or refines the procedures
applied
3.2
animal
any live non-human vertebrate, excluding immature forms during the first half of gestation of incubation
ISO 10993-2:2006(E)
3.3
animal test
any use of an animal for scientific purposes
NOTE 1 The definition of an animal test excludes acts of recognized veterinary practice applied for the benefit of an
animal or the group of animals of which it is part; recognized husbandry practices to manage or conserve the animal or the
group of which it is part; marking by methods which cause no more than momentary pain or distress; and euthanasia.
NOTE 2 The prevention of pain, suffering, distress or lasting harm by the effective use of anaesthesia or analgesia or
other methods of rendering the animal insentient to pain (e.g. decerebration) does not place animal tests outside the
scope of this definition. The administration of anaesthetics, analgesics or other methods of rendering the animal insentient
to pain are considered to constitute an integral part of the animal test.
3.4
competent authority
body designated or recognized by a national government to take responsibility for overseeing, supervising or
regulating animal tests, or the breeding and supply of purpose-bred animals for use on such tests, within the
scope of this part of ISO 10993
3.5
euthanasia
humane killing of an animal by a method causing a minimum of physical and mental suffering
3.6
humane endpoints
pre-determined, specific criteria and measures to be implemented to minimize or terminate pain, suffering or
distress caused by animal tests as soon as the scientific objectives have been met, or when it is realized they
cannot be met, or when the animal welfare problems being encountered are greater than can be justified by
the importance, potential benefits, objectives and nature of the study
3.7
procedural training
prior training and acclimatizing of animals to the interventions to be performed during an animal test, with a
view to minimizing stress to the animal when animal tests are conducted
3.8
protocol
documentation prepared in advance of animal tests being undertaken setting out the justification, rationale
and test method (including scientific and humane endpoints) for the animal tests
3.9
purpose-bred animal
any animal bred with the intention that it be used in animal tests or for other experimental or scientific
purposes
3.10
reduction
reducing to the essential minimum the number of animals used in an animal test to meet a defined scientific
objective
3.11
refinement
sum total of measures taken to safeguard the welfare of the test animals by minimizing any resulting pain,
suffering, distress or lasting harm to the animals that are used
3.12
replacement
any scientifically valid and reasonably and practically available test method that either completely or partially
replaces the use of living vertebrate animals with test methods that have not the potential to cause pain or
distress to animals
2 © ISO 2006 – All rights reserved
ISO 10993-2:2006(E)
3.13
test animal
any animal used in in vivo animal tests, or used to provide tissue for ex vivo or in vitro tests
3.14
validation
formal process by which the reliability and relevance of a test method is established for a particular purpose
4 Requirements
4.1 General
This part of ISO 10993 sets forth essential requirements when animal tests are being considered, planned or
performed for the biological evaluation of materials used in medical devices.
It has been developed to protect the welfare of animals used in the biological evaluation of materials used in
medical devices without compromising, indeed to help to ensure, the scientific validity of the test results and
the risk assessments that shall subsequently be performed.
This part of ISO 10993 focuses on the need to demonstrate that animal welfare is properly considered when
expert judgement has to be exercised in relation to the biological evaluation of medical device materials, and
that the principles of humane experimental technique are demonstrably applied to the design and conduct of
animal tests.
This part of ISO 10993 requires that the need to perform animal tests is justified, and any pain, suffering,
distress or lasting harm that is caused during essential animal tests is minimized.
This part of ISO 10993 sets out essential requirements that safeguard animal welfare by minimizing the pain
and distress caused when animal tests are considered or undertaken by:
a) establishing a framework that reflects the relevant ethical and, in many jurisdictions, the legal
considerations relating to the use of animals for experimental or other scientific purposes;
b) minimizing the number of animal tests by the appropriate use of literature searches, data-sharing,
validated replacement alternatives, and appropriate testing strategies and study designs;
c) minimizing any pain, suffering, distress and lasting harm caused to animals used in tests to evaluate the
biocompatibility of materials used in medical devices by requiring appropriate use of relevant reduction
and refinement alternatives;
d) promoting consistent, high standards of accommodation and care to safeguard both the welfare of the
animals used and the scientific validity and the reproducibility of the data generated.
To these ends the design and conduct of animal tests to evaluate the biocompatibility of materials used in
medical devices shall be formed by, and incorporate, relevant strategies for the replacement, reduction and
refinement of animal tests.
Commissioning animal tests without seeking and obtaining this information, exercising these judgements and
implementing these measures does not comply with the essential requirements of this part of ISO 10993.
NOTE These principles, and the essential requirements of this part of ISO 10993, can also be relevant to animal
tests conducted on medical device materials and medical devices, in other contexts.
ISO 10993-2:2006(E)
4.2 Justification for animal tests
When required to make proper provision to ensure human safety, animal testing to enable the proper
biological characterization of materials used in medical devices is acceptable.
For the purposes of the ISO 10993 series, animal tests shall only be deemed to be justified when:
⎯ the resulting data are not otherwise available, but are essential to properly characterize the test material
in the context in which it is to be used;
⎯ when no suitable scientifically validated test method not involving the use of living animals is reasonably
and practically available;
⎯ when relevant reduction and refinement strategies have been identified and implemented including, if
appropriate, obtaining test data from manufacturers and suppliers, and literature searches for toxicity and
biocompatibility data.
To avoid unnecessary duplication, before animal tests to evaluate the biocompatibility of materials used in
medical devices are undertaken, a review of available, relevant information on the properties of the test
material shall be undertaken and documented. This shall include taking reasonable steps to enable data
sharing.
Animal tests are deemed to be justified only when:
a) they have been shown to be relevant and reliable for the purposes for which they are undertaken;
b) the resulting data are essential to properly characterize and evaluate the test material in the context in
which it is to be used in medical devices;
c) no scientifically valid test method not requiring the use of living animals is reasonably and practically
available;
d) other relevant and appropriate strategies to minimize the pain, suffering, distress and lasting harm caused
to the animals that are used have been identified and implemented.
4.3 Competence of personnel
Animal tests shall be designed, conducted and interpreted by persons competent to discharge the
responsibilities assigned to them.
Animal tests shall be designed and conducted with the involvement of personnel with expertise in veterinary
science, laboratory animal science, and animal husbandry and care.
Details of how staff are equipped by experience, qualification and training (including continued professional
development) to satisfy these requirements shall be documented.
NOTE Although this part of ISO 10993 does not provide an objective specification, it is considered important that those
involved in the conduct of animal tests demonstrate a caring and respectful attitude to the animals used, i.e., that they
have an appropriate “culture of care”.
4.4 Planning and performance of animal tests
4.4.1 General
The selection and design of animal tests shall be appropriate to meet the specific scientific objectives of the
study whilst minimizing the pain, suffering, distress or lasting harm that might be produced to the test animals.
4 © ISO 2006 – All rights reserved
ISO 10993-2:2006(E)
As stated in 4.2, animal testing shall only be undertaken when the information required is essential to
characterize the test material, is not otherwise available and when no suitable scientifically validated test
method not involving the use of living animals is reasonably and practically available.
Following consideration of relevant and reasonably available potential replacement, reduction and refinement
strategies, and before animal tests are undertaken, principal investigators and/or sponsors shall attest and
document that no other replacement, reduction or refinement strategies are required to minimize the animal
welfare costs of the studies.
NOTE In some instances pilot studies can be required to optimize study design before definitive studies can be
designed and performed.
Where the provisions of the ISO 10993 series of International Standards require or permit that an informed
choice be made from a range of species, stages of development or animal numbers for an animal test, the
decisions taken shall both safeguard the sc
...
The article discusses the SIST EN ISO 10993-2:2006 standard, which provides guidelines for ensuring the welfare of animals used in biocompatibility testing of medical devices. It is intended for individuals involved in commissioning, designing, performing, or evaluating animal tests for medical devices. The standard outlines minimum requirements to ensure proper animal welfare and suggests ways to reduce the number of animals used in testing, minimize pain or distress, and explore alternatives to animal tests.
この記事は、医療機器の材料や使用する医療機器の生体適合性を評価するために行われる動物実験に関与する人々を対象にしています。 SIST EN ISO 10993-2:2006は、医療機器の委託、設計、実施、または評価に関与する人々が、医療機器の材料の生体適合性の評価を行うための動物試験において動物の福祉に適切な配慮がなされることを確保し、証明するための最低限の要件を示しています。また、動物実験の数を減らし、動物の苦痛や苦悩を軽減または排除するためのテスト方法の改善、動物の実験を必要としない科学的に妥当な代替手段の模索を促進するための推奨事項とガイダンスも提供しています。
이 문서는 의료기기에 사용되는 재료나 의료기기의 생물학적 적합성을 평가하기 위해 동물 실험을 진행하는 사람들을 대상으로 한다. SIST EN ISO 10993-2:2006은 의료기기 제작과 평가에 참여하는 사람들이 동물실험을 수행할 때 동물의 복지에 적절히 신경쓰도록 돕기 위한 최소한의 요구사항을 제시한다. 또한 실험동물 수를 줄이고, 동물에게 고통이나 고통을 최소화하거나 제거하기 위한 시험 방법의 개선, 동물 실험 없이도 과학적으로 타당한 대안을 모색하는 데 도움이 되는 지침과 권고를 제공한다.











Questions, Comments and Discussion
Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.
Loading comments...