Medical electrical equipment - Medical electron accelerators - Functional performance characteristics

This International Standard applies to medical ELECTRON ACCELERATORS when used, for therapy purposes, in human medical practice. This standard applies to medical ELECTRON ACCELERATORS which deliver a RADIATION BEAM of either X-RADIATION or ELECTRON RADIATION with NOMINAL ENERGIES in the range 1 MeV to 50 MeV at maximum ABSORBED DOSE RATES between 0,001 Gy s-1 and 1 Gy s-1 at 1 m from the RADIATION SOURCE and at NORMAL TREATMENT DISTANCES between 50 cm and 200 cm from the RADIATION SOURCE. The present standard describes measurements and test procedures to be performed by the MANUFACTURER at the design and construction stage of a medical ELECTRON ACCELERATOR but does not specify ACCEPTANCE TESTS to be performed after installation at the purchaser's site. The accompanying report, IEC 60977, however, does suggest that many of the test procedures are appropriate for ACCEPTANCE TESTS. The measurement conditions described in the present standard differ from those previously in use. This applies particularly to the PHANTOM position for measurements and the measurement of distances from the ISOCENTRE. These new conditions should be substituted for and not be added to previous methods. This standard specifies test procedures for the determination and disclosure of functional performance characteristics, knowledge of which is deemed necessary for proper application and use of a medical ELECTRON ACCELERATOR and which are to be declared in the ACCOMPANYING DOCUMENTS together with the greatest deviation or variation to be expected under specific conditions in NORMAL USE. A format for presentation of functional performance values is given in Annex A. It is recognized that inaccuracies in the test methods must be allowed for when assessing performance. However, it was not felt to be advisable to combine the errors into an overall performance tolerance but to keep them separate in the expectation that more accurate test methods will be evolved. It is not intended that this standard should in any way inhibit the future development of new designs of equipment which may have operating modes and parameters different from those described herein, provided that such equipment achieves equivalent levels of performance for the TREATMENT of PATIENTS. Except where otherwise stated this standard assumes that the medical ELECTRON ACCELERATORS have an ISOCENTRIC GANTRY. Where the equipment is non-isocentric, the description of performance and test methods may need to be suitably adapted.

Medizinische elektrische Geräte - Medizinische Elektronenbeschleuniger - Apparative Qualitätsmerkmale

Appareils électromédicaux - Accélérateurs médicaux d'électrons - Caractéristiques fonctionnelles de performance

S'applique aux accélérateurs médicaux d'électrons utilisés à des fins thérapeutiques, en médecine humaine. Cette norme s'applique aux accélérateurs médicaux d'électrons délivrant un faisceau de rayonnement soit rayonnement X soit rayonnement électronique avec des énergies nominales comprises dans l'intervalle de 1 MeV à 50 MeV avec un débit de dose absorbée maximum de 0,001 Gy s-1 à 1 Gy s-1 à 1 m de la source de rayonnement et à une distance normale de traitement entre 50 cm et 200 cm de la source de rayonnement. La présente norme décrit des mesures et des méthodes d'essai à réaliser par le fabricant au moment de la conception et de la construction d'un accélérateur médical d'électrons mais elle ne spécifie pas les essais de réception devant être effectués après l'installation sur le site de l'acheteur. Cependant, le rapport d'accompagnement CEI 60977 suggère que de nombreuses méthodes d'essai sont applicables aux essais de réception. Les conditions de mesure décrites dans la présente norme ne sont pas identiques à celles précédemment en usage. Cela s'applique en particulier à la position du fantôme pour les mesures et à la mesure des distances à partir de l'isocentre. Il convient que ces nouvelles conditions remplacent les précédentes et ne s'y ajoutent pas. Cette norme spécifie les méthodes d'essai pour la détermination et la déclaration des caractéristiques fonctionnelles dont la connaissance est nécessaire pour une application et une utilisation correctes d'un accélérateur médical d'électrons. Elles doivent être déclarées dans les documents d'accompagnement ainsi que les écarts ou variations maximum auxquels on peut s'attendre dans les conditions définies d'utilisation normale. Un modèle pour la présentation des valeurs des caractéristiques fonctionnelles est donné à l'Annexe A. Il est nécessaire de tenir compte de l'imprécision des méthodes de mesure dans la détermination des caractéristiques. Cependant, il n'a pas semblé souhaitable d'inclure les erreurs dans une tolérance globale, mais de les séparer, espérant que des méthodes de mesure plus précises soient mises au point. Il n'est pas dans le but de cette norme d'empêcher, de quelque façon que ce soit le développement futur d'équipements de conception nouvelle qui pourraient avoir des modes de fonctionnement et des paramètres différents de ceux décrits ci-après, à condition que de tels équipements conduisent à des caractéristiques fonctionnelles de niveau équivalent pour le traitement des patients. Sauf indication contraire, la présente norme s'applique aux accélérateurs médicaux d'électrons avec support isocentrique. Si l'équipement est non isocentrique, la description des caractéristiques et les méthodes d'essai peuvent nécessiter d'être convenablement adaptées. Cette deuxième édition annule et remplace la première édition publiée en 1989 dont elle constitue une révision technique. Cette deuxième édition comprend l'ajout de normes de performances et de méthodes d'essai en rapport avec les technologies nouvelles suivantes: - techniques de transmission du faisceau dynamiques, telles que la radiothérapie cinétique, la radiothérapie de conformation avec modulation d'intensité de dose (IMRT), la radiothérapie assistée par imagerie médicale (IGRT), et les faisceaux avec filtres en coin programmables (PWF); - radiotherapie stéréotaxique (SRT)/ radiochirurgie stéréotaxique (SRS); - utilisation de dispositifs d'imagerie électroniques. Il convient de lire cette norme ainsi que la CEI TR 60977 conjointement à la CEI 60601-2-1.

Medicinska električna oprema - Medicinski elektronski pospeševalniki - Funkcijske tehnične karakteristike (IEC 60976:2007)

General Information

Status
Published
Publication Date
09-Jan-2008
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
09-Jan-2008
Due Date
15-Mar-2008
Completion Date
10-Jan-2008

Relations

Buy Standard

Standard
EN 60976:2008
English language
101 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 60976:2008
01-marec-2008
1DGRPHãþD
SIST EN 60976:2002
SIST EN 60976:2002/A1:2002
0HGLFLQVNDHOHNWULþQDRSUHPD0HGLFLQVNLHOHNWURQVNLSRVSHãHYDOQLNL)XQNFLMVNH
WHKQLþQHNDUDNWHULVWLNH ,(&
Medical electrical equipment - Medical electron accelerators - Functional performance
characteristics (IEC 60976:2007)
Medizinische elektrische Geräte - Medizinische Elektronenbeschleuniger - Apparative
Qualitätsmerkmale (IEC 60976:2007)
Appareils électromédicaux - Accélérateurs médicaux d'électrons - Caractéristiques
fonctionnelles (IEC 60976:2007)
Ta slovenski standard je istoveten z: EN 60976:2007
ICS:
11.040.50
13.280
SIST EN 60976:2008 en,fr
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------

EUROPEAN STANDARD
EN 60976

NORME EUROPÉENNE
December 2007
EUROPÄISCHE NORM

ICS 11.040.50; 13.280 Supersedes EN 60976:1999 + A1:2000


English version


Medical electrical equipment -
Medical electron accelerators -
Functional performance characteristics
(IEC 60976:2007)


Appareils électromédicaux -  Medizinische elektrische Geräte -
Accélérateurs médicaux d'électrons - Medizinische Elektronenbeschleuniger -
Caractéristiques fonctionnelles
Apparative Qualitätsmerkmale
de performance (IEC 60976:2007)
(CEI 60976:2007)




This European Standard was approved by CENELEC on 2007-11-01. CENELEC members are bound to comply
with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard
the status of a national standard without any alteration.

Up-to-date lists and bibliographical references concerning such national standards may be obtained on
application to the Central Secretariat or to any CENELEC member.

This European Standard exists in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CENELEC member into its own language and notified
to the Central Secretariat has the same status as the official versions.

CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Cyprus, the
Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia,
Lithuania, Luxembourg, Malta, the Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain,
Sweden, Switzerland and the United Kingdom.

CENELEC
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung

Central Secretariat: rue de Stassart 35, B - 1050 Brussels


© 2007 CENELEC - All rights of exploitation in any form and by any means reserved worldwide for CENELEC members.
Ref. No. EN 60976:2007 E

---------------------- Page: 2 ----------------------

EN 60976:2007 – 2 –
Foreword
The text of document 62C/429/FDIS, future edition 2 of IEC 60976, prepared by SC 62C, Equipment for
radiotherapy, nuclear medicine and radiation dosimetry, of IEC TC 62, Electrical equipment in medical
practice, was submitted to the IEC-CENELEC parallel vote and was approved by CENELEC as EN 60976
on 2007-11-01.
This European Standard supersedes EN 60976:1999 + A1:2000.
EN 60976:2007 includes the addition of performance standards and test methods relating to the following
new technologies:
– dynamic beam delivery techniques, such as
• MOVING BEAM RADIOTHERAPY,
• INTENSITY-MODULATED RADIATION THERAPY (IMRT),
• IMAGE-GUIDED RADIOTHERAPY (IGRT) and
• PROGRAMMABLE WEDGE FIELDS (PWF);
STEREOTACTIC RADIOTHERAPY STEREOTACTIC RADIOSURGERY
– (SRT) / (SRS);
– use of ELECTRONIC IMAGING DEVICES.
This standard, together with IEC/TR 60977, is to be used in conjunction with EN 60601-2-1.
The following dates were fixed:
– latest date by which the EN has to be implemented
at national level by publication of an identical
national standard or by endorsement (dop) 2008-08-01
– latest date by which the national standards conflicting
with the EN have to be withdrawn (dow) 2010-11-01
In this standard, the following print types are used:
– requirements, compliance with which can be tested, and definitions: in roman type;
– explanations, advice, general statements, exceptions and notes: in small roman type;
– test specifications and headings of sub-clauses: in italic type;
TERMS USED THROUGHOUT THIS PARTICULAR STANDARD THAT HAVE BEEN LISTED IN THE INDEX OF DEFINED

TERMS AND DEFINED IN CLAUSE 3, OR IN OTHER STANDARDS: SMALL CAPITALS.
Annex ZA has been added by CENELEC.
__________
Endorsement notice
The text of the International Standard IEC 60976:2007 was approved by CENELEC as a European
Standard without any modification.
__________

---------------------- Page: 3 ----------------------

– 3 – EN 60976:2007

Annex ZA
(normative)

Normative references to international publications
with their corresponding European publications

The following referenced documents are indispensable for the application of this document. For dated
references, only the edition cited applies. For undated references, the latest edition of the referenced
document (including any amendments) applies.

NOTE  When an international publication has been modified by common modifications, indicated by (mod), the relevant EN/HD
applies.

Publication Year Title EN/HD Year

IEC 60580 2000 Medical electrical equipment - Dose area EN 60580 2000
product meters


IEC 60601-1 2005 Medical electrical equipment - EN 60601-1 2006
Part 1: General requirements for basic safety
and essential performance


IEC 60601-2-1 1998 Medical electrical equipment - EN 60601-2-1 1998
A1 2002 Part 2-1: Particular requirements for the A1 2002
safety of electron accelerators in the range of
1 MeV to 50 MeV


IEC/TR 60788 2004 Medical electrical equipment - Glossary of – –
defined terms


1)
IEC 60977 – Medical electrical equipment - Medical – –
electron accelerators in the range of 1 MeV to
50 MeV - Guidelines for functional
performance characteristics


1) 2)
IEC 61217 – Radiotherapy equipment - Coordinates, EN 61217 1996
movements and scales


IEC 61223-1 1993 Evaluation and routine testing in medical – –
imaging departments -
Part 1: General aspects




1)
Undated reference.
2)
Valid edition at date of issue.

---------------------- Page: 4 ----------------------

IEC 60976
Edition 2.0 2007-10
INTERNATIONAL
STANDARD
NORME
INTERNATIONALE
Medical electrical equipment – Medical electron accelerators – Functional
performance characteristics

Appareils électromédicaux – Accélérateurs médicaux d’électrons –
Caractéristiques fonctionnelles de performance

INTERNATIONAL
ELECTROTECHNICAL
COMMISSION
COMMISSION
ELECTROTECHNIQUE
PRICE CODE
INTERNATIONALE
XD
CODE PRIX
ICS 11.040.50; 13.280 ISBN 2-8318-9314-3

---------------------- Page: 5 ----------------------

– 2 – 60976 © IEC:2007
CONTENTS
FOREWORD.6
INTRODUCTION.8

1 Scope.9
2 Normative references .10
3 Terms and definitions .10
4 Environmental conditions.14
4.1 General .14
4.2 Transport and storage .14
4.3 Power supply.14
5 General information to the USER .14
5.1 Functional performance characteristics .14
5.2 Available nominal energies and ABSORBED DOSE RATES .14
5.3 Available RADIATION FIELDS .15
5.4 NORMAL TREATMENT DISTANCE.15
5.5 Available WEDGE X-RAY FIELDS .15
5.6 Available flattening FILTERS .15
5.7 Availability.16
5.8 Influencing quantities .16
5.9 Maintenance.16
5.10 Presentation.16
5.11 Dimensions, clearances, within the RADIATION HEAD, and in the region
RADIATION HEAD to ISOCENTRE, of BEAM LIMITING DEVICES .16
5.12 IMRT .16
6 Standardized test conditions.17
6.1 General .17
6.2 Angle settings .17
6.3 Properties and positioning of the PHANTOM .17
6.4 Positioning of measuring points.17
6.5 RADIATION DETECTORS .17
6.6 STANDARD MEASUREMENT DEPTHS.18
6.7 RADIATION FIELDS .18
6.8 Adjustments during test.18
6.9 Use of RADIOGRAPHIC FILM or alternative imaging method.18
7 DOSE MONITORING SYSTEM .18
7.1 General .18
7.2 Reproducibility .19
7.3 Proportionality.20
7.4 Dependence on angular positions.21
7.5 Dependence on GANTRY rotation.22
7.6 Dependence on the shape of the RADIATION FIELD .22
7.7 Stability of calibration .23
7.8 Stability in MOVING BEAM RADIOTHERAPY.25
8 Depth ABSORBED DOSE characteristics .26
8.1 X-RADIATION.26
8.2 ELECTRON RADIATION .28

---------------------- Page: 6 ----------------------

60976 © IEC:2007 – 3 –
9 Uniformity of RADIATION FIELDS .31
9.1 X-RADIATION.31
9.2 ELECTRON RADIATION .36
9.3 PENUMBRA of RADIATION FIELDS.38
10 Indication of RADIATION FIELDS.39
10.1 X-RADIATION.39
10.2 ELECTRON RADIATION .45
10.3 Geometry and motion speeds of adjustable BLDs for X-RADIATION and
ELECTRON RADIATION .46
10.4 Illuminance and PENUMBRA of the LIGHT FIELD.47
11 Indication of the RADIATION BEAM AXIS.48
11.1 General .48
11.2 Indication on entry to the PATIENT .49
11.3 Indication on exit from the PATIENT .51
12 ISOCENTRE .52
12.1 Displacement of the RADIATION BEAM AXIS from the ISOCENTRE .52
12.2 Indication of the ISOCENTRE .53
13 Indication of distance along the RADIATION BEAM AXIS .53
13.1 Indicating device .53
13.2 Additional indicating device for equipment with variable distance between
RADIATION SOURCE and ISOCENTRE and for non-isocentric equipment .54
14 Zero position of rotational scales .55
14.1 General .55
14.2 Information to the USER .55
14.3 Tests.55
15 Congruence of opposed RADIATION FIELDS.56
15.1 Information to the USER .56
15.2 Test.56
16 Movements of the PATIENT table.57
16.1 General .57
16.2 Vertical movement of the table .57
16.3 ISOCENTRIC rotation of the table.58
16.4 Parallelism of table rotational axes .58
16.5 Rigidity of the table .59
17 ELECTRONIC IMAGING DEVICE (e.g. EPID) .60
17.1 Information to the USER .60
17.2 Tests.62

Annex A (informative)  Format for presentation of functional performances values.75

Index of defined terms .96

Figure 1 – Explanatory diagram for the definition of wedge .64
Figure 2 – The rotary GANTRY.65
Figure 3 – The wall- or floor-mounted GANTRY.66
Figure 4 – The ceiling-mounted GANTRY .67
Figure 5 – Flattened area (shown hatched) within the RADIATION FIELD .68

---------------------- Page: 7 ----------------------

– 4 – 60976 © IEC:2007
Figure 6 – Examples of profiles of ABSORBED DOSE along the major axes or the
diagonal axes .69
Figure 7 – Explanatory diagram for flatness of the ELECTRON FIELD .70
Figure 8 – A possible arrangement of equipment for the measurement of the ISOCENTRE
described in Clause 12 .71
Figure 9 – Test 10.1.1.3.72
Figure 10 – RADIATION HEAD showing X-RADIATION BLDs and ACCESSORIES (see 4.11) .73
Figure 11 – Multi-element BLD RADIATION FIELDS used for measurement of X-RADIATION
PENUMBRA (see 8.3.2).74

Table 1 – Conditions for testing reproducibility.19
Table 2 – Conditions for testing proportionality of the DOSE MONITORING SYSTEM.20
Table 3 – Conditions for testing dependence of the DOSE MONITORING SYSTEM on
equipment position.21
Table 4 – Conditions for testing dependence of the DOSE MONITORING SYSTEM on
GANTRY rotation.22
Table 5 – Conditions for testing dependence on the shape of the RADIATION FIELD.23
Table 6 – Conditions for testing stability of calibration of the DOSE MONITORING SYSTEM .24
Table 7 – Conditions for testing stability of the DOSE MONITORING SYSTEM in MOVING
BEAM RADIOTHERAPY .26
Table 8 – Conditions for testing depth dose characteristics.27
Table 9 – Conditions for testing depth dose characteristics.29
Table 10 – Conditions for testing stability of PENETRATIVE QUALITY of ELECTRON
RADIATION.29
Table 11 – Flattened area according to Figure 5 .31
Table 12 – Conditions for testing flatness and symmetry of X-RAY FIELDS .32
Table 13 – Conditions for testing deviation of dose distribution of X-RAY FIELDS with
angular position .33
Table 14 – Conditions for testing maximum ABSORBED DOSE ratio in the RADIATION FIELD .34
Table 15 – Conditions for testing WEDGE FACTORS .35
Table 16 – Conditions for testing WEDGE ANGLES .35
Table 17 – Conditions for testing flatness, symmetry, deviation of dose distribution with
angular position, and maximum ABSORBED DOSE ratio of ELECTRON FIELDS .37
Table 18 – Conditions for film calibration .41
Table 19 – Conditions for testing the numerical and the LIGHT FIELD-INDICATION.42
Table 20 – Conditions for testing reproducibility of X-RAY FIELDS .44
Table 21 – Conditions for testing the LIGHT FIELD-INDICATOR for ELECTRON RADIATION .46
Table 22 – Conditions for testing geometry of the BEAM LIMITING SYSTEM .48
Table 23 – Conditions for testing the indication of the RADIATION BEAM AXIS on entry to
the PATIENT .50
Table 24 – Conditions for testing the indication of the RADIATION BEAM AXIS on exit from
the PATIENT .51

---------------------- Page: 8 ----------------------

60976 © IEC:2007 – 5 –
Table 25 – Conditions for testing indication of the ISOCENTRE.53
Table 26 – Conditions for testing vertical movement of the table.58
Table 27 – Conditions for testing ISOCENTRIC rotation of the table .58
Table 28 – Conditions for testing the angulation of rotational axes of the table .59
Table 29 – Conditions for testing lateral rigidity of the table .60

---------------------- Page: 9 ----------------------

– 6 – 60976 © IEC:2007
INTERNATIONAL ELECTROTECHNICAL COMMISSION
______________

MEDICAL ELECTRICAL EQUIPMENT –
MEDICAL ELECTRON ACCELERATORS –
FUNCTIONAL PERFORMANCE CHARACTERISTICS


FOREWORD
1) The International Electrotechnical Commission (IEC) is a worldwide organization for standardization comprising
all national electrotechnical committees (IEC National Committees). The object of IEC is to promote
international co-operation on all questions concerning standardization in the electrical and electronic fields. To
this end and in addition to other activities, IEC publishes International Standards, Technical Specifications,
Technical Reports, Publicly Available Specifications (PAS) and Guides (hereafter referred to as “IEC
Publication(s)”). Their preparation is entrusted to technical committees; any IEC National Committee interested
in the subject dealt with may participate in this preparatory work. International, governmental and non-
governmental organizations liaising with the IEC also participate in this preparation. IEC collaborates closely
with the International Organization for Standardization (ISO) in accordance with conditions determined by
agreement between the two organizations.
2) The formal decisions or agreements of IEC on technical matters express, as nearly as possible, an international
consensus of opinion on the relevant subjects since each technical committee has representation from all
interested IEC National Committees.
3) IEC Publications have the form of recommendations for international use and are accepted by IEC National
Committees in that sense. While all reasonable efforts are made to ensure that the technical content of IEC
Publications is accurate, IEC cannot be held responsible for the way in which they are used or for any
misinterpretation by any end user.
4) In order to promote international uniformity, IEC National Committees undertake to apply IEC Publications
transparently to the maximum extent possible in their national and regional publications. Any divergence
between any IEC Publication and the corresponding national or regional publication shall be clearly indicated in
the latter.
5) IEC provides no marking procedure to indicate its approval and cannot be rendered responsible for any
equipment declared to be in conformity with an IEC Publication.
6) All users should ensure that they have the latest edition of this publication.
7) No liability shall attach to IEC or its directors, employees, servants or agents including individual experts and
members of its technical committees and IEC National Committees for any personal injury, property damage or
other damage of any nature whatsoever, whether direct or indirect, or for costs (including legal fees) and
expenses arising out of the publication, use of, or reliance upon, this IEC Publication or any other IEC
Publications.
8) Attention is drawn to the Normative references cited in this publication. Use of the referenced publications is
indispensable for the correct application of this publication.
9) Attention is drawn to the possibility that some of the elements of this IEC Publication may be the subject of
patent rights. IEC shall not be held responsible for identifying any or all such patent rights.
International standard IEC 60976 has been prepared by subcommittee 62C: Equipment for
radiotherapy, nuclear medicine and radiation dosimetry, of IEC technical committee 62:
Electrical equipment in medical practice.
This second edition cancels and replaces the first edition published in 1989. It constitutes a
technical revision.
This second edition includes the addition of performance standards and test methods relating
to the following new technologies:
− dynamic beam delivery techniques, such as
• MOVING BEAM RADIOTHERAPY,
• INTENSITY-MODULATED RADIATION THERAPY (IMRT),
• IMAGE-GUIDED RADIOTHERAPY (IGRT) and
• PROGRAMMABLE WEDGE FIELDS;
− STEREOTACTIC RADIOTHERAPY (SRT) / STEREOTACTIC RADIOSURGERY (SRS);

---------------------- Page: 10 ----------------------

60976 © IEC:2007 – 7 –
− use of ELECTRONIC IMAGING DEVICES.
This standard, together with IEC TR 60977, should be read in conjunction with IEC 60601-2-1.
The text of this standard is based on the following documents:
FDIS Report on voting
62C/429/FDIS 62C/433/RVD

Full information on the voting for the approval of this standard can be found in the report on
voting indicated in the above table.
This publication has been drafted in accordance with the ISO/IEC Directives, Part 2.
In this standard, the following print types are used:
– requirements, compliance with which can be tested, and definitions: in roman type;
– explanations, advice, general statements, exceptions and notes: in small roman type;
– test specifications and headings of sub-clauses: in italic type;
– TERMS USED THROUGHOUT THIS PARTICULAR STANDARD THAT HAVE BEEN LISTED IN THE INDEX OF
DEFINED TERMS AND DEFINED IN CLAUSE 3, OR IN OTHER STANDARDS: SMALL CAPITALS.
The committee has decided that the contents of this publication will remain unchanged until
the maintenance result date indicated on the IEC web site under "http://webstore.iec.ch" in
the data related to the specific publication. At this date, the publication will be
• reconfirmed;
• withdrawn;
• replaced by a revised edition, or
• amended.

---------------------- Page: 11 ----------------------

– 8 – 60976 © IEC:2007
INTRODUCTION
Standards containing safety requirements for medical ELECTRON ACCELERATORS have been
published by the IEC, details of which will be found in Clause 2.
The present standard specifies methods of testing and methods of disclosure of functional
performance of medical ELECTRON ACCELERATORS intended for RADIOTHERAPY. It permits a
direct comparison between the performance data of equipment of different MANUFACTURERS.
Since this standard does not contain safety requirements, it has not been numbered in the
IEC 60601 publication series. It describes aspects of functional performance of medical
ELECTRON ACCELERATORS and the way in which they should be presented. It also includes test
methods and conditions suitable for TYPE TESTS. These test methods are suggested test
methods and alternative methods may be equally appropriate, but the specified functional
performance characteristics of medical ELECTRON ACCELERATORS shall be related to these test
methods and conditions. Tests specified in this standard are not necessarily appropriate for
ensuring that any individual medical ELECTRON ACCELERATOR conforms to the declared
functional performance during the course of its working lifetime. Guidance on the values
which may be expected is given in the technical report, IEC 60977.
This International Standard was first published in 1989. With the rapidly increasing
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.